Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect

Size: px
Start display at page:

Download "Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect"

Transcription

1 Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect Robert G. Hauser, MD, FHRS, Deepa McGriff, BS, Linda Kallinen Retel, BS, FHRS From the Minneapolis Heart Institute Foundation, Minneapolis, Minnesota. BACKGROUND The Riata family of implantable cardioverter-defibrillator leads (St Jude Medical, Sylmar, CA) appears prone to a unique failure mechanism whereby the conductor cables wear through the silicone insulation from inside-out and are seen outside the lead body (externalized conductors). OBJECTIVE To assess the extent of Riata lead damage associated with inside-out insulation defects and their clinical consequences. METHODS In September 2011, we searched the U.S. Food and Drug Administration s Manufacturers and User Defined Experience medical device database for reports describing Riata lead failures that had been analyzed by the manufacturer. RESULTS The Manufacturers and User Defined Experience search identified 105 leads that had inside-out insulation defects. Eight- French single-coil Riata leads accounted for a higher-thanexpected proportion (25.7%) of the leads with this defect. A total of 226 insulation defects were found in the 105 leads (2.2 defects per lead), including 143 inside-out defects (1.4 defects per lead). The most common location of insulation defects was distal to the proximal coil (n 108). Twenty-eight leads (26.7%) had insideout insulation defects underneath the shocking coils. Of 43 leads whose cables were assessed for the integrity of the ethylenetetrafluoroethylene cable coating, 22 (51.2%) were found to be abraded, exposing the conductor surfaces. On X-ray radiography or fluoroscopy, 7 leads were found to have externalized cables; 2 of these leads had no electrical abnormalities, while 4 exhibited noise or increased impedance. Inappropriate shocks were experienced by 31 of the 105 patients (29.5%). CONCLUSION Riata leads that have inside-out insulation defects often have multiple defects, including additional inside-out abrasions along the body of the lead and beneath the shocking coils. Eight-French single-coil Riata models may be more prone to externalized cables than are dual-coil and 7-F designs. Externalized cables are but one manifestation of interior insulation damage. Our findings question the durability of the ethylene-tetrafluoroethylene cable coating on exposed cables. KEYWORDS Implantable cardioverter-defibrillator; Lead; Insulation; Complications; Failure ABBREVIATIONS ETFE ethylene-tetrafluoroethylene; FDA Food and Drug Administration; ICD implantable cardioverterdefibrillator; MAUDE Manufacturers and User Facility Device Experience (Heart Rhythm 2012;9: ) 2012 Heart Rhythm Society. All rights reserved. Introduction The Riata and Riata ST family of 8-F and 7-F implantable cardioverter-defibrillator (ICD) leads (St Jude Medical, Sylmar, CA) appear prone to a unique failure mechanism whereby the conductor cables wear through the silicone insulation (inside-out abrasion) and appear outside the lead body (externalized conductors) (Figure 1). 1 5 Recent studies suggest Riata failure rates of 8% 15%, including leads that are electrically intact but that exhibit externalized conductors. 6,7 The leads may function normally because the highvoltage and pace-sense cables are covered with ethylenetetrafluoroethylene (ETFE), which serves as a second insulating barrier (Figure 2). However, the reliability of ETFE on exposed conductors in patients is unknown. While published reports have focused on externalized conductors, little is known of Riata insulation or conductor defects associated with inside-out abrasions. The December Address reprint requests and correspondence: Dr Robert G. Hauser, MD, Minneapolis Heart Institute Foundation, th Str, Suite 500, Minneapolis, MN address: rhauser747@aol.com letter from St Jude Medical 8 acknowledged that it had found insulation defects in Riata and Riata ST leads returned for product analysis, including in-the-pocket lead to pulse generator can abrasion, intravascular or intracardiac lead-to-lead abrasion, and lead abrasion caused by intracardiac and extravascular anatomic structures, such as the tricuspid valve apparatus and the clavicle. The purpose of this study was to assess the extent of Riata lead damage associated with inside-out insulation abrasions and the clinical consequences as reported by the manufacturer to the U.S. Food and Drug Administration (FDA) and available online in the FDA s Manufacturers and User Facility Device Experience (MAUDE) database. Methods Riata and Riata ST leads The 8-F Riata ICD lead was market released in the United States in March 2002, and it was removed from distribution in 2010 after approximately 227,000 leads were sold worldwide, including 144,579 U.S. implants (Table 1). 8 The Riata ST models, which were introduced in 2005, are 7-F-diam /$ -see front matter 2012 Heart Rhythm Society. All rights reserved. doi: /j.hrthm

2 Hauser et al Riata ICD Lead Failure 743 Figure 1 Extracted dual-coil Riata lead showing externalized cables (arrows) in 2 locations between the proximal and distal coils. eter leads. Riata and Riata ST leads were insulated with pure silicone rubber; in 2006, a proprietary silicone-polyurethane copolymer (Optim TM, St Jude Medical, Sylmar, CA) covering was added to Riata ST Optim models. The multilumen construction of these leads is illustrated in Figure 2. The high-voltage and 2 pace-sense cables were extruded with 1.5 mil ( in) of ETFE, and the pace-sense conductor coil was strung through a tube of polytetrafluoroethylene. Definitions Externalized conductors are cables that have breached the outer insulation and are visible outside the lead body (Figure 1); approximately 85% of the externalized conductors are caused by inside-out insulation abrasions, and 15% are associated with outside-in insulation abrasions. 9 Inside-out abrasion is an insulation defect caused by the relative motion of the cables within the lead insulation lumen (Figure 2, arrows); this defect begins as a silicone abrasion and creeps outward to the surface of the lead body. Outside-in abrasion is an insulation defect caused by contact with another lead or anatomic structure; it begins with a surface abrasion. A can abrasion is an insulation defect caused by the lead contact with the pulse generator housing in the pocket; can abrasions can cause short circuiting between a high-voltage cable and the pulse generator. FDA s MAUDE database The MAUDE database contains reports of adverse events involving medical devices. In our experience, 10,11 the vast majority of reports originate from manufacturers, and a small percentage is submitted by user facilities. The FDA requires manufacturers to report adverse events that are communicated to them verbally or in writing, and they must report the results of investigations into the causes of device malfunctions. The relevant data items for each event include (1) model number, (2) report source (eg, hospital or manufacturer), (3) event location, (4) date of event and report, (5) device age, (6) patient outcome, and (7) narrative of event and manufacturers analysis of returned devices. MAUDE reports are available online at html. The FDA updates the database about every 2 months; consequently, there may be a delay between the time a report is filed and when it is posted on the FDA s Web site. Previous reports from this center have focused on MAUDE data for ICD pulse generator and lead failures and deaths associated with lead extraction devices. 10,11 We searched the MAUDE database on September 6, 2011, for Riata lead failures by using the search term inside-out abrasion. We also searched the MAUDE database by using the search term Riata lead failure. All the reports returned by the search were from the manufacturer. Results The MAUDE search found 107 Riata leads that had insideout insulation abrasions (Table 1); 105 of these leads were analyzed by St Jude Medical, and 2 leads were not analyzed because they were abandoned in situ and replaced during surgical revisions after externalized cables were seen on fluoroscopy. The MAUDE search for Riata lead failure returned 721 reports; thus, inside-out abrasions were approximately 15% of Riata lead failures. The single-coil leads (models 7002, 1592, 1582, 1572) accounted for 25.7% of the leads in this study even though they represent only Figure 2 Cross sections of 8-F silicone Riata single-coil (A) and dual-coil (B) high-voltage implantable cardioverter-defibrillator leads. The cables are coated in ethylene-tetrafluoroethylene, which are shown in blue, and the central coil pace-sense conductor with a stylet lumen is encased in a tube of polytetrafluoroethylene, which is shown in green. Arrows indicate location and direction of inside-out abrasion.

3 744 Heart Rhythm, Vol 9, No 5, May 2012 Table 1 Characteristics of explanted and returned Riata and Riata ST leads that St Jude Medical analyzed and found to have insideout insulation abrasions Model Number in this study Insulation Coils/fixation Diameter (F) Year of market release Number of U.S. implants Riata ST 7040, Silicone Dual, passive , Riata ST Silicone Single, active , Riata ST 7000, Silicone Dual, active , Riata 1590, Silicone Dual, active , Riata Silicone Single active Riata Silicone Single, active , Riata Silicone Single, passive Riata 1570, Silicone Dual, passive , Riata 1580, Silicone Dual, active , Total , Percentage of U.S. implants 10.7% of the leads sold in the United States. The average age of the 105 analyzed leads was months (range months). The 2 leads that were not analyzed (aged 85 and 57 months) were abandoned in situ and replaced during surgical revisions after externalized cables were seen on fluoroscopy. Clinical observations and engineering analyses Table 2 provides data for each of the 105 leads that were found to have one or more inside-out insulation abrasions, including those that were discovered incidentally and were not the immediate cause of failure. A total of 226 insulation defects were found, including 143 insideout defects. The most common location of insulation defects was distal to the region of the proximal coil (n 108) (Figure 3). The majority of leads (69 of 105; 65.7%) had multiple insulation defects, and many of these (30 of 69; 43.5%) had typical insulation defects caused by can abrasions. Some insulation abrasions were caused by contact with another lead or vascular or cardiac structures (outside-in abrasion). Twenty-eight of the 105 leads (26.7%) had inside-out insulation defects underneath one or more of the highvoltage shocking coils. Of these, 23 were 8-F Riata and 5 were 7-F Riata ST leads; thus, 25.8% of Riata leads (23 of 89) and 31.3% of Riata ST leads (5 of 16) had inside-out abrasions under a shocking coil (P.76). Other findings included 32 leads with exposed cables or conductors, and 6 leads had melted cables, presumably owing to one or more high voltage shocks. Of the 43 leads that were assessed for the integrity of the ETFE cable coating, 22 (51.2%) were found to be abraded, exposing the conductor. Signs and clinical consequences of lead failure As shown in Table 2, increased pacing thresholds tended to appear earlier than other signs while impedance changes occurred later. Noise and other sensing issues were the most common signs of failure (Table 3) and developed around 5 years. On x-ray radiography or fluoroscopy, 7 leads were found to have externalized cables; 2 of these leads had no electrical abnormalities, while 4 exhibited noise or oversensing and 1 lead had increased impedance. Inappropriate shocks were experienced by 31 of the 105 patients (29.5%). Inappropriate shocks were associated with abraded ETFE cables; 9 of the 22 patients (40.9%) who had abraded cables received inappropriate shocks, while inappropriate shocks occurred in 3 of the 21 patients (14.2%) whose cables were not abraded (P.088). One death was associated with a can abrasion and truncated high-voltage shock. Discussion The results of this study suggest that failed 8-F Riata and 7-F Riata ST leads with inside-out abrasions frequently have multiple insulation defects that are distributed along the length of the lead. These additional insulation defects (both inside-out and outside-in) are often accompanied by other defects, including exposed cables, fractured and melted conductors, and can abrasions. Sensing abnormalities, including noise on the lead, and impedance changes are the most common electrical abnormalities. A third of the patients in this study experienced serious adverse events, primarily inappropriate shocks, and all patients required surgical intervention. Inside-out insulation abrasion resulting in externalized cables is a unique cause of ICD lead failure. It has been suggested that movement of the cables within the lumens produces outward forces that with time (years) disrupts the silicone insulation and releases the cables outside the body of the lead. 8,11 This may explain the frequent occurrence of externalized cables in the region of the tricuspid valve, but it does not explain the other lead damage we found in this study. We hypothesize that once a cable is exposed and is no longer restrained by the insulation it begins to flex and cause collateral damage, which accumulates over time; the lead damage may be greatest within the heart where flexion is most pronounced. A contributing factor may be the tendency of silicone to flow away from pressure points or creep at body temperature, rather than mechanical abrasion alone. Creep is time dependent, and it can result in permanent thinning of the insulation, reducing its tensile strength and resistance to abrasion. 12 Thus the cause of inside-out insulation failure in Riata and Riata ST leads is likely multifactorial, a combination of cable movement or stacking within the lumens,

4 Hauser et al Riata ICD Lead Failure 745 Table 2 Results of St Jude Medical s engineering analyses of returned Riata and Riata ST leads that were found to have inside-out insulation abrasions as well as the associated clinical signs of failure and patient consequences as reported to the Food and Drug Administration by the manufacturer Model number length Age (mo) Clinical sign of failure Engineering analysis Insulation defects Electrical Other ETFE cable coating Patient consequences 1570/65 32 Noise Multiple Normal Cables exposed 1570/ Impedance, noise Multiple Normal Cables exposed Normal 1570/ Externalized cables Single Cables exposed Partial lead 1580/65 48 Oversensing Single Partial lead 1580/65 59 Oversensing Multiple incl Lead in 2 pieces Normal 1580/ Impedance Multiple Normal 1581/65 53 Loss of capture Multiple Partial lead Normal 7000/65 30 Noise, externalized cables Single Normal Normal 7000/65 40 Abnormal impedance Multiple Normal Clavicular crush 1570/65 57 Abnormal impedance Multiple Cables exposed 1570/65 61 No information Multiple SVC cables exposed Normal 1570/65 69 Oversensing, 1Impedance Multiple Cables exposed Abraded Inappropriate shocks 1570/65 82 Noise Multiple Ring electrode to RV coil contact Abraded Inappropriate shocks 1570/65 96 Noise Single Abrasion below the RV HV coil 1571/65 52 No information Multiple incl 1571/ HV impedance IOA SVC coil SVC and RV coils melted (short circuit) 1572/65 57 Loss of capture Multiple incl IOA involving RV coil Normal Cables exposed 1572/65 90 Aborted shocks, noise Externalized cables 1572/65? Insulation break Multiple Fractured RV shock c 1580/60 48 High threshold Multiple RV cables abraded Abraded 1580/60 60 Oversensing Multiple incl Cables exposed Abraded Inappropriate shocks 1580/60 60 Aborted shocks, noise Multiple incl Abraded 1580/60 70 Abnormal impedance Single Partial lead Normal 1580/60 71 Externalized cables Multiple incl 1580/60 80 Oversensing Multiple incl 1580/65 18 No information Single 1580/65 37 Oversensing, 1Impedance Multiple incl Shorted Abraded 1580/65 48 Noise Single Lead in 2 pieces 1580/65 54 Oversensing Single 1580/65 55 Multiple Partial lead Normal 1580/ Impedance Multiple Cables exposed Abraded Inappropriate shocks 1580/65 62 Loss of capture Single Partial lead 1580/65 65 Noise, insulation defect Multiple Fractures HV cables on x-ray Lead in 2 pieces 1580/65 67 Break Multiple Normal Cables exposed Normal 1580/65 68 Noise Multiple incl Sensing conductors exposed 1580/65 72 No information Multiple 1580/ Threshold, 1Impedance Multiple incl Lead in 2 pieces Normal 1580/65 76 Loss of capture Single

5 746 Heart Rhythm, Vol 9, No 5, May 2012 Table 2 Continued Model number length Age (mo) Clinical sign of failure Engineering analysis Insulation defects Electrical Other ETFE cable coating Patient consequences 1580/65 80 Noise Single Cables exposed Abraded Inappropriate shocks 1580/65 80 Noise, externalized cables Multiple incl 1580/65 96 No information Multiple incl Abraded 1580/65 98 No information Single Normal 1580/ Impedance Multiple incl Cables exposed 1581/60 50 Noise Multiple incl Sensing cable exposed 1581/ Impedance Single Partial lead 1581/ Impedance Single Lead in 2 pieces Normal 1581/ HV impedance Single Melted cables 1581/65 46 No information Single Normal 1581/65 52 No information Multiple incl Melted cables 1581/ Threshold, 2R wave Single Partial lead 1581/65 54 Externalized cables; Multiple incl 1Impedance, noise 1581/65 56 Oversensing Multiple RV cable melted 1581/65 60 Oversensing Multiple incl Abraded 1581/65 63 No information Single 1581/ Impedance Single Clavicular crush 1581/ Impedance Multiple 1Impedance Cables melted 1581/65 77 No information Single Normal 1581/ Impedance Multiple incl Partial lead Abraded 1581/65 79 Short Multiple 1581/65 80 Oversensing Multiple Cables melted 1581/65 82 No information Multiple Normal 1581/ Pacing impedance Multiple incl Abraded 1581/65 98 Loss of sensing and Single Pacing cable and SVC pacing, noise coil contact 1581/65? 1Impedance Multiple incl High impedance Cables melted Death Inappropriate shocks truncated shock 1582/60 55 Noise Single Normal Inappropriate shocks 1582/60 72 No information Single Lead in 2 pieces 1582/60? Oversensing Multiple Cables exposed Abraded Inappropriate shocks 1582/65 15 Insulation break Multiple incl Cables exposed 1582/65 36 Noise 1582/65 55 Noise Multiple Shorted Cables exposed 1582/65 56 Noise Multiple Abraded Inappropriate shocks

6 Hauser et al Riata ICD Lead Failure 747 Table 2 Continued Model number length Age (mo) Clinical sign of failure Engineering analysis Insulation defects Electrical Other ETFE cable coating Patient consequences 1582/65 62 High threshold Multiple incl Cables exposed 1582/65 62 No information Multiple 1582/65 63 Oversensing, noise Multiple Oversensing, noise 1582/65 68 Oversensing Multiple Cables exposed Partial lead 1582/65 72 Oversensing Single Shorted 1582/65 74 No information Multiple Abraded 1582/65 77 X-ray abnormality Multiple Abraded 1582/ Impedance Multiple Cables exposed 1582/65 82 Oversensing Multiple Cables exposed Abraded Inappropriate shocks 1582/ HV impedance Multiple Partial lead Abraded 1582/65 86 Externalized cables, Single abnormal sensing 1582/65 94 Vibratory alert triggered by low impedance Multiple incl Abraded 1582/65? Oversensing Multiple Cables exposed Abraded 1590/60 59 Fracture Multiple Externalized cables No fracture 1590/60 66 Oversensing Single Cables exposed Normal Inappropriate shocks 1591/65 65 Oversensing Multiple incl Cables exposed 1592/65 60 Abnormal impedance Multiple incl Cables exposed 7000/60 33 Noise, oversensing Multiple Shorted 7000/60 42 Oversensing Multiple Insulation abraded owing to friction with another lead 7000/60 53 Oversensing Multiple Abraded 7000/65 33 Noise Multiple External and inside-out abrasions 7000/65 48 Oversensing, externalized Multiple Lead in 2 Cables exposed cables pieces Lead in 2 pieces 7000/65 53 High threshold Single Cables exposed Normal Lead in.2 pieces 7001/ Impedance Single Shorted Cables exposed and melted 7001/65? Fracture Multiple incl Shorted Cables exposed Distal coil fractured Normal Inappropriate shocks 7001/75 23 Insulation break Single 7002/65? Loss of capture Single Normal Cable exposed Normal 7040/65 24 Noise Single Partial lead Abraded Inappropriate shocks 7040/ Impedance, noise Multiple incl 7041/65 55 Undersensing Multiple Cables exposed Normal 1581/65 62 Noise Multiple incl Shorted Multiple inside-out abrasions 1582/65 52 Undersensing Single Shorted Undersensing 1582/65 59 Oversensing Single Shorted 7000/ Impedance Multiple incl Shorted Inappropriate shock ETFE ethylene-tetrafluoroethylene; HV high voltage; IOA inside-out abrasion; SVC Superior Vena Cava; RV right ventricular; incl including.

7 748 Heart Rhythm, Vol 9, No 5, May 2012 A B C Model/ Defect Type No. Leads Total Defects Number of Defects in Each Lead Segment A B* C Unspecified Location Dual Coil Models 1580, Inside-out Outside-in Unspecified , Inside-out Outside-in Unspecified , Inside-out Outside-in , Inside-out Outside-in , Inside-out 3 3 Outside-in Single Coil Models* 1572, 1582, Inside-out Outside-in Inside-out 1 1 Total Figure 3 Location of Riata and Riata ST insulation defects. Note that the proximal coil is not present in single-coil models, and defects in segment B will be in the lead body for single-coil models. which applies pressure against the silicone, causing creep, thinning, and loss of tensile strength. Eventually, the outer insulation (Figure 1) ruptures and tears proximally and distally, uncovering the cables, which may be displaced outside the lead body. The unrestrained cables then produce additional damage as described previously. As shown in Table 1, single-coil leads accounted for 10.7% (15,492 of 144,579) of 8-F Riata and 7-F Riata ST leads sold in the United States. Thus, a larger-than-expected proportion (27 of 105; 25.7%) of leads in this study (Table 1) were single-coil models (Figure 2A). Recently, St Jude Medical reported that 8-F single-coil Riata leads were more prone to externalized conductors than were other Riata silicone models, including the downsized 7-F Riata ST whose 3 cable pairs were positioned closer to a smaller stylet lumen. 9 The addition of a third cable pair to the single-coil Riata ST leads is a unique design feature because this third cable pair provides no electrical function.

8 Hauser et al Riata ICD Lead Failure 749 Table 3 Signs of Riata and Riata ST lead failures with insideout insulation defects Number Impedance changes Increased impedance Decreased impedance Unspecified impedance abnormality 4 4 Sensing Noise on lead Oversensing Decreased R wave/undersensing 3 3 Pacing Loss of captures 6 7 High threshold 5 6 X-ray/fluoroscopic signs 9 10 Fracture or insulation defect 2 2 Externalized cables 7 8 Other 3 3 Total Percentage A quarter of the 105 leads in this study had inside-out defects under a shocking coil, which prevented the cables from externalizing. These defects were seen equally in 8-F Riata and 7-F Riata ST leads. In addition, two 7-F Riata ST leads had externalized cables. Accordingly, while the 7-F Riata ST design changes may have mitigated the 8-F singlecoil lead s propensity to externalize cables, they have not prevented inside-out insulation defects. A critical decision confronting physicians is how to manage patients who are found to have externalized cables. This study was not designed to answer this question, but certain of our findings should be considered when planning follow-up studies. First, many leads in this study were returned in pieces, suggesting that their removal was difficult. St Jude Medical, in its recent letter to physicians, 9 reported 2 deaths and 1 serious injury associated with the extraction of leads with externalized cables. Physicians who have extracted these leads report the need for larger extraction sheaths, which have been shown to be associated with higher complication rates. Second, it is not known how durable the ETFE is on a cable exposed to blood flow and continuous cardiac motion. The thickness of the ETFE on the cables is only approximately 1.5 mil, and its dielectric strength (nominally 4500 V) can be reduced if abraded. While the ETFE may appear to be electrically intact in low-voltage applications, it may not withstand the energy delivered by a high-voltage shock. In this study, we found abraded cables, meaning that the ETFE had been damaged; the patients who had abraded cables tended to receive inappropriate shocks. Studies are needed to determine whether it is feasible to know with a high degree of certainty that externalized cables are durable over time and capable of delivering effective therapy. This study has certain limitations. Some of the lead damage described herein may have been sustained during extraction. However, the manufacturer s reports did not attribute any of the insulation defects to extraction-related damage except to observe that a lead was returned partially or in sections. Inside-out and outside-in insulation defects are the result of chronic changes in the silicone rather than acute damage. Moreover, we have found that extracted silicone Fidelis leads rarely exhibited insulation damage based on returned product analyses. 11 The purpose of this study was to assess leads that had inside-out insulation defects. Thus we did not analyze all Riata and Riata ST failures in MAUDE. To our knowledge, there are no reports in the medical literature of inside-out abrasions involving St Jude Medical leads that employ Optim insulation. In conclusion, failed 8-F Riata and 7-F Riata ST leads that have been explanted and found to have inside-out insulation defects often exhibit multiple insulation defects, including additional inside-out abrasions along the body of the lead and beneath the shocking coils. Eight-French single-coil Riata models may be more prone to externalized cables than are dual-coil Riata and Riata ST models. Externalized cables may be but one manifestation of interior insulation damage caused by chronic, unrestrained cable movement and possibly the inherent tendency of silicone to creep and tear. Lastly, our findings question the reliability of ETFE to insulate cables in leads with inside-out abrasions, especially those that have been exposed, externalized, or continuously flexed. Clinical studies are urgently needed to determine how best to manage patients who have Riata and Riata ST leads. References 1. Duray GZ, Israel CW, Schmitt J, Hohnloser SH. Implantable cardioverterdefibrillator lead disintegration at the level of the tricuspid valve. Heart Rhythm 2008;5: Jalal Z, Derval N, Ploux S, Bordachar P. Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead. Heart Rhythm 2010; 7: Richards MW, Warren CE, Anderson MH. Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation. Europace 2010;12: Valk S, Luitjen R. Jordaens L. Insulation damage in a shock wire: an unexpected fluoroscopic image. Pacing Clin Electrophysiol 2010;33: Krebsbach A, Alhumaid F, Henrickson CA, Calkins H, Berger RD, Cheng A. Premature failure of a Riata defibrillator lead impedance change or inappropriate sensing: a case report and review of the literature. J Cardiovasc Electrophysiol 2011;22: Kodoth V, Cromie N, Lau E, McEneaney D, Wilson C, Roberts MJ. Riata lead failure: a report from Northern Ireland Riata lead screening programme. Available from: &eevtid 48. Accessed December 4, Erkapic D, Duray GZ, Bauernfeind T, De Rosa S, Hohnloser SH. Insulation defects of thin high-voltage ICD leads: an underestimated problem? J Cardiovasc Electrophysiol 2011;22: Chester KM. Important Product Information: St Jude Medical Riata and Riata ST Silicone Endocardial Leads. Sylmar, CA: St Jude Medical; December 15, Carlson M, Tsung P. Medical Device Advisory. Sylmar, CA: St Jude Medical; November 28, Hauser RG, Katsiyiannis WT, Gornick CC, Almquist AK, Kallinen LM. Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction. Europace 2010;12: Hauser RG, Kallinen LM, Almquist AK, Gornick CC, Katsiyiannis WT. Early failure of a small-diameter high-voltage implantable cardioverter lead. Heart Rhythm 2007;4: Cheremisinoff NP, Cheremisinoff PN, eds. Elastomer Technology Handbook. Harpers Ferry, VA: CRC Press; 1993:66.

Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads

Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads Robert G. Hauser, MD, FHRS, Raed Abdelhadi, MD, Deepa McGriff, BS, Linda Kallinen Retel, BS, FHRS From the

More information

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson

Riata Lead Summit Jan 20, SJM Perspective Dr. Mark Carlson Riata Lead Summit Jan 20, 2012 SJM Perspective Dr. Mark Carlson 1 1 Presentation Outline Background, Characterization of Externalized Conductors Modeling and Testing Diagnostics and Therapeutics Recommendations

More information

Fluoroscopic Grading Scale and Correlation to Lead Extraction in Failed St. Jude Medical Riata Family ICD Leads

Fluoroscopic Grading Scale and Correlation to Lead Extraction in Failed St. Jude Medical Riata Family ICD Leads Fluoroscopic Grading Scale and Correlation to Lead Extraction in Failed St. Jude Medical Riata Family ICD Leads Sunthosh V Parvathaneni MD 1, Christopher R Ellis MD 1, Jason A Rytlewski MD 1, and Jeffrey

More information

Riata Lead Extraction. Thomas D. Callahan, MD, FACC, FHRS

Riata Lead Extraction. Thomas D. Callahan, MD, FACC, FHRS Riata Lead Extraction Thomas D. Callahan, MD, FACC, FHRS Outline Riata lead history Lead design Lead Failure Extraction outcomes Techniques Special considerations Riata Lead Background 8F introduced in

More information

ICD leads survival and troubles in the last 10 years

ICD leads survival and troubles in the last 10 years XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial

More information

The Riata Implantable Cardioverter Defibrillator Lead: Extraction Experience for Conductor Exteriorization and Electrical Malfunction

The Riata Implantable Cardioverter Defibrillator Lead: Extraction Experience for Conductor Exteriorization and Electrical Malfunction The Riata Implantable Cardioverter Defibrillator Lead: Extraction Experience for Conductor Exteriorization and Electrical Malfunction Heath E. Saltzman, MD, Faiz Subzposh, MD, Christine Saari, CRNP, S.

More information

Riata lead extraction- a single centre experience

Riata lead extraction- a single centre experience Riata lead extraction- a single centre experience Rebecca L Noad, Keith W Morrice, Vivek N Kodoth, Carol M Wilson, Michael JD Roberts. Royal Victoria Hospital, Belfast, United Kingdom Background of previous

More information

Fluoroscopic investigation of Riata transvenous defibrillator leads

Fluoroscopic investigation of Riata transvenous defibrillator leads ORIGINAL ARTICLE Cardiology Journal 2015, Vol. 22, No. 1, 57 67 DOI: 10.5603/CJ.a2014.0039 Copyright 2015 Via Medica ISSN 1897 5593 Fluoroscopic investigation of Riata transvenous defibrillator leads Gurjit

More information

SUPPLEMENTAL MATERIAL: ICD Leads: Design, Diagnostics, and. Management

SUPPLEMENTAL MATERIAL: ICD Leads: Design, Diagnostics, and. Management SUPPLEMENTAL MATERIAL: ICD Leads: Design, Diagnostics, and Management Charles Swerdlow MD* and Kenneth A. Ellenbogen, MD * Cedars Sinai Heart Institute and David Geffen School of Medicine at UCLA, Los

More information

Cardiac Rhythm Management Division Product Performance Report April 2011

Cardiac Rhythm Management Division Product Performance Report April 2011 Cardiac Rhythm Management Division Product Performance Report April April Letter from St. Jude Medical As a world leader in the development of state-of-the-art technology for cardiac rhythm management

More information

PUBLICATION SUMMARY PERTAINING TO ABBOTT RIATA LEADS (WITH >100 PATIENTS OR RIATA LEADS UNLESS NOTED)

PUBLICATION SUMMARY PERTAINING TO ABBOTT RIATA LEADS (WITH >100 PATIENTS OR RIATA LEADS UNLESS NOTED) Abbott 15900 Valley View Court Sylmar, CA 91342 USA Main 818 362 6822 Fax 818 364 5814 PUBLICATION SUMMARY PERTAINING TO ABBOTT RIATA LEADS (WITH >100 PATIENTS OR RIATA LEADS UNLESS NOTED) 2016 2017 Publications

More information

Patient Safety: the optimal lead body design

Patient Safety: the optimal lead body design Patient Safety: the optimal lead body design E. Soldati U.O. Malattie Cardiovascolari II Azienda Ospedaliero Universitaria Pisana Advances in Cardiac Arrhythmias Torino, 25-27 Ottobre 2012 Lead Malfunction

More information

that physicians image Riata and Riata ST leads implanted in patients to assess for externalization or other visible insulation abnormalities.

that physicians image Riata and Riata ST leads implanted in patients to assess for externalization or other visible insulation abnormalities. St. Jude Medical Cardiac Rhythm Management Division 15900 Valley View Court Sylmar, CA 91342-3577 USA Tel 818 362 6822 800 423 5611 www.sjm.com Mark Carlson, MD, MA Chief Medical Officer and Sr. VP Clinical

More information

Riata TM Publications

Riata TM Publications St. Jude Medical Implantable Electronic Systems 15900 Valley View Court Sylmar, CA 91342 USA Main 818 362 6822 Fax 818 364 5814 Riata TM Publications Citation Mean follow-up Data From Results Lorvidhaya

More information

Introduction. Case Report ECG & EP CASES ABSTRACT

Introduction. Case Report ECG & EP CASES ABSTRACT Successful extraction of an implantable cardioverter-defibrillator lead in a patient with pocket infection via the femoral approach with a basket snare Jin-Bae Kim, MD, PhD. Cardiology Division, Department

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt Leadless Pacing Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt The weakest link in Pacemaker system the lead. The more the leads the more the complications Dislodgement Fracture Insulation

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736

Active Fixation Models: 7740, 7741 and 7742 Passive Fixation Models: 7731, 7732, 7735 and 7736 The INGEVITY MRI pacing leads are 6F (2.0 mm) steroid-eluting, endocardial pace / sense leads designed for permanent implantation for either atrial or ventricular applications. INGEVITY MRI is the only

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

Bench testing: ETFE-coated conductors

Bench testing: ETFE-coated conductors St. Jude Medical Implantable Electronic Systems 15900 Valley View Court Sylmar, CA 91342 USA Main 818 362 6822 Fax 818 364 5814 Riata Publications Citation Fischer A, Klehn R. Contribution of ethylenetetrafluoroethylene

More information

Lead Selection and Subcutaneous ICD Considerations. The Future of CIEDs

Lead Selection and Subcutaneous ICD Considerations. The Future of CIEDs Lead Selection and Subcutaneous ICD Considerations The Future of CIEDs February, 2013 Steven P. Kutalek, MD Director, Cardiac Electrophysiology Drexel University College of Medicine New Technologies 2012

More information

The transvenous defibrillation lead was a critical technological

The transvenous defibrillation lead was a critical technological Arrhythmia Management Devices Implantable Cardioverter-Defibrillator Leads Design, Diagnostics, and Management Charles D. Swerdlow, MD; Kenneth A. Ellenbogen, MD The transvenous defibrillation lead was

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Guidelines for Identifying Externalized Conductors on Radiographic Images

Guidelines for Identifying Externalized Conductors on Radiographic Images Guidelines for Identifying Externalized Conductors on Radiographic Images These guidelines for determining the presence of externalized conductors were used by independent physician adjudicators in the

More information

CIEDs Infection: Lead Extraction, First or Last option?

CIEDs Infection: Lead Extraction, First or Last option? CIEDs Infection: Lead Extraction, First or Last option? More CRM Systems & Leads 600,000 new devices annually 1.2 million new leads annually Sources: 1) Medtronic CRDM Product Performance Report, Jan 2009.

More information

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Device Interrogation- Pacemakers, ICD and Loop Recorders Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI Disclosures Consultant: Medtronic Speaker s Bureau: St. Jude Medical

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network Pacing troubleshooting NASPE training Lancashire & South Cumbria Cardiac Network Pacing stimulus present failure to capture Lead dislodgement Early unstable position Late twiddlers syndrome Abnormal ECG

More information

Heart Rhythm Society. Welcome to San Francisco! 2/26/2013. WS #5 ICD Introduction 1. Workshop # 9. ICD Registry Program

Heart Rhythm Society. Welcome to San Francisco! 2/26/2013. WS #5 ICD Introduction 1. Workshop # 9. ICD Registry Program Welcome to San Francisco! Workshop # 9 ICD Registry Program Thursday March 7, 2013 4:00pm 5:30pm Heart Rhythm Society Presenter Disclosure Information Chairman, ICD Registry Steering Committee The following

More information

12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads

12 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads 12 YEARS PROVEN PERFORMANCE Sprint Quattro Defibrillation Leads PROVEN PERFORMANCE. PROVEN RELIABLE. We made intentional choices and took proactive steps from lead design to active product monitoring to

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experience

The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experience ORIGINAL ARTICLE Cardiovascular Disorders https://doi.org/10.3346/jkms.2017.32.10.1610 J Korean Med Sci 2017; 32: 1610-1615 The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experience

More information

Device Implantation: Long-Term Commitments. Fayez Bokhari, MD, FRCPC Director of Cardiology Military Hospital, Jeddah

Device Implantation: Long-Term Commitments. Fayez Bokhari, MD, FRCPC Director of Cardiology Military Hospital, Jeddah Device Implantation: Long-Term Commitments Fayez Bokhari, MD, FRCPC Director of Cardiology Military Hospital, Jeddah Who should Implant ICD? 1. Electrophysiologist 2. Surgeons 3. General Cardiologist with

More information

MEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System

MEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System MEET MICRA ACTUAL SIZE Micra TM Transcatheter Pacing System MEET MICRA The world s smallest pacemaker 1 MINIATURIZED. 93% smaller than modern-day pacemakers 7 Completely self contained within the heart,

More information

UNITED STATES DISTRICT COURT FOR THE DISTRICT OF MINNESOTA. Civil No. CLASS ACTION COMPLAINT

UNITED STATES DISTRICT COURT FOR THE DISTRICT OF MINNESOTA. Civil No. CLASS ACTION COMPLAINT CASE 0:12-cv-01419-PJS-JJK Document 1 Filed 06/14/12 Page 1 of 38 UNITED STATES DISTRICT COURT FOR THE DISTRICT OF MINNESOTA ROBERT L. SATOW, Individually and on Behalf of All Others Similarly Situated,

More information

Complications of Lead Extraction: Prevention and treatment. Maria Grazia Bongiorni, MD, FESC

Complications of Lead Extraction: Prevention and treatment. Maria Grazia Bongiorni, MD, FESC Complications of Lead Extraction: Prevention and treatment Maria Grazia Bongiorni, MD, FESC Director of Cardiovascular Division University Hospital of Pisa (Italy) ourtesy of Dr Eivind Platou Potential

More information

Update on Device Innovation (S-ICD, Wearable, Leadless)

Update on Device Innovation (S-ICD, Wearable, Leadless) Update on Device Innovation (S-ICD, Wearable, Leadless) C. W. Israel Dept. of Medicine - Cardiology Evangelical Hospital Bielefeld J. W. Goethe University Frankfurt Carsten.Israel@em.uni-frankfurt.de Conflicts

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

Lead Extraction: Challenges in our area

Lead Extraction: Challenges in our area Lead Extraction: Challenges in our area More CRM Systems & Leads 600,000 new devices annually 1.2 million new leads annually Sources: 1) Medtronic CRDM Product Performance Report, Jan 2009. 2) Boston Scientific

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Publication Summary pertaining to St. Jude Medical Durata Leads (with >100 patients and Durata leads unless noted)

Publication Summary pertaining to St. Jude Medical Durata Leads (with >100 patients and Durata leads unless noted) St. Jude Medical 15900 Valley View Court Main 818 362 6822 Fax 818 364 5814 Sylmar, CA 91342 USA Publication Summary pertaining to St. Jude Medical Durata Leads (with >100 patients and Durata leads unless

More information

Product Performance Report

Product Performance Report Product Performance Report Cardiac Rhythm Management October 9 October 9 Letter from St. Jude Medical As a world leader in the development of state-of-the-art technology for cardiac rhythm management

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

Advances in Arrhythmia and Electrophysiology

Advances in Arrhythmia and Electrophysiology Advances in Arrhythmia and Electrophysiology Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I Charles D. Swerdlow, MD; Samuel J. Asirvatham, MD; Kenneth A. Ellenbogen, MD; Paul A.

More information

Generator and Lead-Related Complications of Implantable Cardioverter Defibrillators

Generator and Lead-Related Complications of Implantable Cardioverter Defibrillators Int Cardiovasc Res J.2014;8(2):66-70 Generator and Lead-Related Complications of Implantable Cardioverter Defibrillators Ahmad Yaminisharif 1, *, Nader Soofizadeh 1, Akbar Shafiee 1, Ali Kazemisaeid 1,

More information

You have a what, inside you?

You have a what, inside you? Costal Emergency Medicine Conference You have a what, inside you? Less than mainstream medical devices encountered in the ED. Eric Ossmann, MD, FACEP Associate Professor Duke University Medical Center

More information

Product Performance Report

Product Performance Report Product Performance Report Cardiac Rhythm Management May May Letter from St. Jude Medical As a world leader in the development of state-of-the-art technology for cardiac rhythm management devices, St.

More information

MEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System

MEET MICRA. Micra TM ACTUAL SIZE. Transcatheter Pacing System MEET MICRA ACTUAL SIZE Micra TM Transcatheter Pacing System MEET MICRA The world s smallest pacemaker 1 MINIATURIZED. 93% smaller than modern-day pacemakers 6 Completely self contained within the heart,

More information

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators:

Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Outcomes of Defibrillator Lead Implants Performed by High Volume Operators vs. Low Volume Operators: Results from the Pacemaker and Implantable Defibrillator Leads Survival Study ( PAIDLESS ) Partially

More information

11 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads

11 YEARS PROVEN PERFORMANCE. Sprint Quattro Defibrillation Leads 11 YEARS PROVEN PERFORMANCE Sprint Quattro Defibrillation Leads PROVEN PERFORMANCE. PROVEN RELIABLE. We made intentional choices and took proactive steps from lead design to active product monitoring to

More information

PRODUCT PERFORMANCE REPORT

PRODUCT PERFORMANCE REPORT PRODUCT PERFORMANCE REPORT MAY 215 Contents Contents 1 Sorin CRM commitment to Quality 6 2 Definitions and methods 7 2.1 General definitions......................................... 7 2.2 Malfunctions.............................................

More information

Aalborg Universitet. The St. Jude Medical Riata defibrillator lead advisory Moesgaard, Jacob

Aalborg Universitet. The St. Jude Medical Riata defibrillator lead advisory Moesgaard, Jacob Aalborg Universitet The St. Jude Medical Riata defibrillator lead advisory Moesgaard, Jacob DOI (link to publication from Publisher): 10.5278/vbn.phd.med.00025 Publication date: 2015 Document Version Publisher's

More information

Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany

Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany Leadless pacemakers and subcutaneous ICD s: will we use them for most of our patients? K.-H. Kuck Asklepios Klinik St. Georg Hamburg, Germany Disclosure Statement Research Grants Consultant / Advisory

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

FINELINE II STEROX EZ Models 4469/4470/4471/4472/4473/4474 Guidant Implantable Pacing Lead

FINELINE II STEROX EZ Models 4469/4470/4471/4472/4473/4474 Guidant Implantable Pacing Lead FINELINE II STEROX EZ Models 4469/4470/4471/4472/4473/4474 Guidant Implantable Pacing Lead CAUTION: Federal Law restricts this device to sale by or on the order of a physician trained or experienced in

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

The reliability of an implantable cardioverter-defibrillator. Original Article

The reliability of an implantable cardioverter-defibrillator. Original Article 1170 Circ Arrhythm Electrophysiol December 2013 1561, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592, 1940, 7000, 7001, 7002, 7010, 7020, 7021, 7022, 7040, 7041, 7042, 7052, 7070, 7071, 7120, 7121,

More information

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD

Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Recent Advances in Pacing and Defibrillation Harish Doppalapudi, MD Harish Doppalapudi, MD Assistant Professor of Medicine Director, Clinical Cardiac Electrophysiology Training Program University of Alabama

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

The reliability of an implantable cardioverter-defibrillator. Original Article

The reliability of an implantable cardioverter-defibrillator. Original Article Original Article Performance of Lead Integrity Alert to Assist in the Clinical Diagnosis of Implantable Cardioverter Defibrillator Lead Failures Analysis of Different Implantable Cardioverter Defibrillator

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

Inappropriate electrical shocks: Tackling the beast

Inappropriate electrical shocks: Tackling the beast ESC Paris 2011 Inappropriate electrical shocks: Tackling the beast Gerhard Hindricks University of Leipzig Heart Center Dept. of Electrophysiology ESC Paris 2011 Inappropriate electrical shocks: Tackling

More information

Product Performance Report

Product Performance Report Product Performance Report Cardiac Rhythm Management April 7 Letter from St. Jude Medical APRIL 7 As world leaders in the development of state-of-the-art technology for cardiac rhythm management (CRM)

More information

Pediatric Pacemaker Implantation Endocardial or Epicardial

Pediatric Pacemaker Implantation Endocardial or Epicardial Pediatric Pacemaker Implantation Endocardial or Epicardial HAITHAM BADRAN, MD, FEHRA CONSULTANT OF INTERVENTIONAL CARDIOLOGY CONSULTANT OF CARDIAC PACING AND ELECTROPHYSIOLOGY LECTURER OF CARDIOLOGY AIN

More information

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART December 2015 This chart encompasses all Medtronic cardiac devices FDA-Approved as MR Conditional and included in the MRI SureScan portfolio. If a device

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Chapter 80 Implantable Cardiac Devices

Chapter 80 Implantable Cardiac Devices Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the

More information

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness

More information

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients

More information

Tachycardia Devices Indications and Basic Trouble Shooting

Tachycardia Devices Indications and Basic Trouble Shooting Tachycardia Devices Indications and Basic Trouble Shooting Peter A. Brady, MD., FRCP Cardiology Review Course London, March 6 th, 2014 2011 MFMER 3134946-1 Tachycardia Devices ICD Indications Primary and

More information

Advanced ICD Troubleshooting: Part II REVIEW

Advanced ICD Troubleshooting: Part II REVIEW REVIEW Advanced ICD Troubleshooting: Part II CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN From the *Cedars-Sinai Medical Center, Los Angeles, California, and Mayo Clinic, Rochester, Minnesota, USA Failure

More information

MRI ACCESS. Evera MRI SureScan ICD Systems

MRI ACCESS. Evera MRI SureScan ICD Systems ACCESS Evera SureScan ICD Systems ACCESS. 36% of ICD patients are likely to have an ordered over 4 years 3 Introducing Evera, featuring SureScan Technology. ICD Patients are not receiving s 1.4% 36% CONTOURED.

More information

IBHRExam Prep Implanted CRM Device System Radiography

IBHRExam Prep Implanted CRM Device System Radiography Implanted CRM Device System Radiography IBHRExam Prep www.pacericd.com 2 Where do they go? 3 Anatomy diagram 4 Anatomy 5 Pulse generator configurations www.pacericd.com 6 Guidant pacemaker pulse generator

More information

Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D.

Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D. Apparent Failure of Automated Threshold Evaluation: What is the Mechanism? Vikas Kalra, M.B.B.S. Lynne Foreman, R.N. Deepak Bhakta, M.D. Case Summary A 57-year-old man developed severe symptomatic tricuspid

More information

A 'shocking' case rectified

A 'shocking' case rectified A 'shocking' case rectified Journal: Pacing and Clinical Electrophysiology Manuscript ID: PACE-13-00.R1 Wiley - Manuscript type: Device Rounds Date Submitted by the Author: n/a Complete List of Authors:

More information

Chapter 14. Risk of Failure of Transvenous Implantable Cardioverter Defibrillator Leads.

Chapter 14. Risk of Failure of Transvenous Implantable Cardioverter Defibrillator Leads. Risk of Failure of Transvenous Implantable Cardioverter Defibrillator Leads. C. Jan Willem Borleffs, MD 1 ; Lieselot van Erven, MD, PhD 1 ; Rutger J. van Bommel, MD 1 ; Enno T. van der Velde, PhD 1 ; Ernst

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Table S1. Patient Selection Codes, CIED Generator Procedures Code Type Code Description ICD9 Proc 00.51 Implantation of cardiac resynchronization defibrillator, total system [CRT-D]

More information

Physician s Manual. ENDOTAK RELIANCE TM Models 0147/0148/0149. ENDOTAK RELIANCE TM S Models 0127/0128/0129

Physician s Manual. ENDOTAK RELIANCE TM Models 0147/0148/0149. ENDOTAK RELIANCE TM S Models 0127/0128/0129 Physician s Manual ENDOTAK RELIANCE TM Models 0147/0148/0149 ENDOTAK RELIANCE TM S Models 0127/0128/0129 Steroid-Eluting Tined Endocardial Defibrillation Leads RESTRICTED DEVICE: Federal law (USA) restricts

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France

How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France How to treat Cardiac Resynchronization Therapy complications? C. Leclercq Departement of Cardiology Centre Cardio-Pneumologique Rennes, France Presenter Disclosure Information Christophe Leclercq, MD,

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Single- versus Dual-Coil ICD Leads: Does it Matter?

Single- versus Dual-Coil ICD Leads: Does it Matter? Single- versus Dual-Coil ICD Leads: Does it Matter? C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Conflict of Interest Biotronik Boston-Scientific

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

GDT1000 SENSING ACUTE STUDY

GDT1000 SENSING ACUTE STUDY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston

More information

Subcutaneous implantable cardioverter-defibrillator (S-ICD)

Subcutaneous implantable cardioverter-defibrillator (S-ICD) NEW DRUGS AND TECHNOLOGIES IN CARDIOLOGY Cardiology Journal 2011, Vol. 18, No. 3, pp. 326 331 Copyright 2011 Via Medica ISSN 1897 5593 Subcutaneous implantable cardioverter-defibrillator (S-ICD) S. Suave

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Disclosures. Optimal Lead Selection: An Extractor s Guide to Lead Choice and Implant Technique. Extraction Experts Have Learned!

Disclosures. Optimal Lead Selection: An Extractor s Guide to Lead Choice and Implant Technique. Extraction Experts Have Learned! 1 Optimal Lead Selection: An Extractor s Guide to Lead Choice and Implant Technique Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services Wexner Medical

More information

Draft Recommendations Report by the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines

Draft Recommendations Report by the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Draft Report by the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines Embargoed until April 26, 2006 Heart Rhythm Society Draft Report Task Force Mark D. Carlson, MD, MA, Task

More information

EASYTRAK 3 Coronary Venous Steroid-Eluting Bipolar Spiral Fixation Pace/Sense Lead

EASYTRAK 3 Coronary Venous Steroid-Eluting Bipolar Spiral Fixation Pace/Sense Lead Physician s Manual EASYTRAK 3 Coronary Venous Steroid-Eluting Bipolar Spiral Fixation Pace/Sense Lead MODELS: 4524/4525/4527 RESTRICTED DEVICE: Federal law (USA) restricts the sale, distribution, or use

More information

His Bundle Pacing in Bundle Branch Block May 11, 2017

His Bundle Pacing in Bundle Branch Block May 11, 2017 His Bundle Pacing in Bundle Branch Block May 11, 2017 Gopi Dandamudi, MD FHRS System Medical Director, IUH Cardiac EP Program Director, IUH Atrial Fibrillation Center Assistant Professor of Clinical Medicine

More information

Lead extraction. Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013

Lead extraction. Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013 Lead extraction Dr. Mervat Abo El Maaty Professor of Cardiology Ain shams University 2013 Agenda Introduction History of consensus Definitions Complications Indications Lead management environment Extraction

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,900 116,000 120M Open access books available International authors and editors Downloads Our

More information