The Complete Supraannular Concept In Vivo Hemodynamics of Bovine and Porcine Aortic Bioprostheses

Size: px
Start display at page:

Download "The Complete Supraannular Concept In Vivo Hemodynamics of Bovine and Porcine Aortic Bioprostheses"

Transcription

1 The Complete Supraannular Concept In Vivo Hemodynamics of Bovine and Porcine Aortic Bioprostheses Daniel J. Ruzicka, MD, MSc; Ina Hettich, MD; Andrea Hutter, MD; Sabine Bleiziffer, MD; Catalin C. Badiu, MD; Robert Bauernschmitt, PhD, MD; Ruediger Lange, PhD, MD; Walter B. Eichinger, PhD, MD Background Complete supraannular placement of an aortic bioprosthesis is one approach to optimize the hemodynamic result of an aortic valve replacement. It is achieved with the combination of a special valve design and the supraannular sewing technique with noneverted mattress sutures. We evaluated 5 bioprostheses designed for complete supraannular placement to assess potential hemodynamic differences caused by factors (eg, valve material) other than implantation position. Methods and Results In 336 patients (mean age, years; 143 women), hemodynamics including mean pressure gradients, effective orifice areas, and indices and incidence of patient-prosthesis mismatch were evaluated 6 months after surgery. Annulus diameter was measured during surgery. Patients received the Carpentier Edwards Perimount Magna (Magna, n 169), the Medtronic Mosaic (Mosaic, n 46), the Mosaic Ultra (Ultra, n 17), the SJM Epic Supra (Epic, n 46), and the Sorin Soprano (Soprano, n 58). For small annulus sizes ( 23 mm), the mean pressure gradients of the Magna ( mm Hg) were significantly lower than the Mosaic ( mm Hg) and the Ultra ( mm Hg), and the Soprano ( mm Hg) was hemodynamically superior to the Ultra. For medium (23 to 24 mm) and large ( 24 mm) annulus sizes, the mean pressure gradients of the Magna were lower than the Epic ( mm Hg versus mm Hg; mm Hg versus mm Hg). Furthermore, in patients with large annulus size, the mean pressure gradients of the Soprano ( mm Hg) were lower compared with the Epic ( mm Hg). Severe patient-prosthesis mismatch was observed more frequently in patients with the Mosaic (12/46; 26.1%) and the Ultra (3/17; 17.6%) prostheses. Conclusions Complete supraannular placement cannot prevent high pressure gradients or patient-prosthesis mismatch thoroughly, but the choice of a bovine prosthesis can optimize hemodynamic performance. (Circulation. 2009; 120[suppl 1]:S139 S145.) Key Words: complete supraannular aortic valve prostheses The complete supraannular design of stented aortic bioprostheses combined with the supraannular sewing technique is one approach to optimize the hemodynamic result after aortic valve replacement. In prostheses with this design, the stent is placed on top of the sewing ring and not beside it. Therefore, the entire valve is positioned above the annulus, and no parts of the stent or sewing ring impair the blood stream. With this concept, the internal valve diameter should theoretically be equal to the tissue annulus diameter and thus achieve an optimal hemodynamic performance with no obstruction of the blood flow. 1 In fact, the effective orifice area of a valve prosthesis is in the order of 35% to 48% of the geometric annulus area that is theoretically available 2 ; thus, the aim of the supraannular concept can only be to gain as much effective orifice area as possible in relation to the patient s annulus size. There are several prostheses for complete supraannular placement available from different manufacturers, all following the same concept but with different design implementations. To evaluate factors other than prosthesis position, which might influence the hemodynamic performance (eg, valve material), we evaluated 5 commercially available bioprostheses for complete supraannular placement (Carpentier Edwards Perimount Magna, 1,3 Medtronic Mosaic, 1,4 9 Medtronic Mosaic Ultra, 10,11 St Jude Medical Epic Supra, 4,7,12 and Sorin Soprano 1,7,8,13 ). In this retrospective study, our focus lies on the hemodynamic in vivo performance (mean pressure gradients, effective orifice area, patient-prosthesis mismatch), geometric parameters, and a literature review concerning hemodynamic data to evaluate differences between prostheses designed according to the complete supraannular concept. From the Department for Cardiovascular Surgery, German Heart Center, Munich, Germany. Presented in part at American Heart Association Scientific Sessions 2008, November 8 12, 2008, New Orleans, La. The online-only Data Supplement is available with this article at Correspondence to Dr Daniel J. Ruzicka, MD, MSc, Siemens AG, Healthcare Sector, Siemensstrasse 1, D Forchheim, Germany. daniel@ruzickas.de 2009 American Heart Association, Inc. Circulation is available at DOI: /CIRCULATIONAHA S139

2 S140 Circulation September 15, 2009 Figure 1. Overview of different complete supraannular bioprostheses. A, Magna; B, Mosaic; C, Ultra; D, Epic; E, Soprano. The discrepancy between valve size labeling by different companies makes comparisons of valves on the basis of labeled valve size meaningless. Therefore, we use as a basis for comparisons as in previous studies the patient s aortic annulus, which is measured with a standardized sizer during surgery. 1,4,8,13,14 Methods Bioprostheses The Carpentier Edwards (CE) Perimount Magna (Magna) valve is a stented bovine pericardial bioprosthesis. It is a modification of the CE Perimount valve and is characterized by glutaraldehyde fixation and a modified stent and sewing-ring design to permit complete supraannular placement. It has been in clinical use since September (Figure 1A). The Medtronic Mosaic (Mosaic) valve is a stented porcine bioprosthesis. The prosthetic tissue is treated with glutaraldehyde, and fixation is done by combining zero-pressure and root-pressure techniques. Amino-oleic acid was applied as an anticalcification treatment. 1,4,5,15 It has been in clinical use in Europe since 1994 (Figure 1B). The Medtronic Mosaic Ultra (Ultra) valve is a modification of the Mosaic valve. The sewing ring of the Mosaic was reduced to adapt to the small aortic root and to maximize the valve-to-annulus ratio. Fixation and anticalcification process are similar to that of the Mosaic valve. It was introduced in our institution in February ,11 (Figure 1C). The SJM Epic Supra (Epic) is a stented porcine tissue valve based on the design of the SJM Epic valve. Modification of the SJM Epic allows a complete supraannular implantation of the Epic Supra. The SJM Epic and the Epic Supra are both based on the design of the Biocor valve, which has a documented 20-year long-term durability. 16 Its design consists of a triple composite construction with matched individual leaflets. The leaflet tissue is fixed under low pressure in glutaraldehyde to promote retention of collagen crimp and elasticity, before applying the Linx anticalcification treatment. The Epic Supra was introduced to the European market in October (Figure 1D). The Sorin Soprano (Soprano) is a stented bovine pericardial bioprosthesis. The tissue is fixed at low pressure with glutaraldehyde and then detoxified using homocysteic acid to neutralize residues of unbound aldehyde groups. The Soprano is based on the Sorin Pericarbon More, which has been in clinical use for more than 15 years. New features of the Soprano are the stent design, allowing complete supraannular implantation, the prosthetic dimensions, and the valve size labeling 1,13 (Figure 1E). Operative Techniques Aortic valve replacement was performed using standard cardiopulmonary bypass with moderate hypothermia at 32 C with cold crystalloid cardioplegic cardiac arrest. After removal of the native aortic valve and decalcification of the aortic annulus and root, the internal diameter of the aortic annulus was routinely measured by inserting a probe (Hegar dilator) into the annulus (unit: 1 mm). With the assumption that the aortic annulus approaches a circular shape, the annulus orifice area was calculated as follows: (Hegar dilator diameter (cm) 0.5) 2. The prosthetic valve size was determined by using the original sizer provided by each manufacturer. All bioprostheses were implanted with pledgeted, interrupted, noneverting mattress sutures. By using ventriculoarterial mattress sutures combined with the special valve design, complete supraannular placement is achieved (Figure 2). Patient Enrollment and Follow-Up Inclusion criteria were aortic valve replacement because of degenerative aortic valve disease with aortic stenosis or a combined aortic lesion, a completed 6-month follow-up examination including echocardiography, and a valid intraoperative annulus measurement. Concomitant procedures were allowed, except of valve replacement in other positions. Patients with aortic valve replacement due to acute endocarditis or isolated aortic regurgitation were excluded as well as patients requiring emergency operations. In total, 336 patients (143 women) were included in this retrospective, nonrandomized study. One hundred sixty-nine patients received the Magna, 46 the Mosaic, 17 the Ultra, 46 the Epic, and 58 the Soprano. The 5 tested valves were not all available at the same time during the whole period of the study. The Mosaic has been used since August The Magna was introduced in January 2001 and the Soprano in September The Epic was used since June 2004 and the Ultra was introduced in February 2007; therefore this is the smallest group. Patients were included consecutively to avoid bias by patient selection. Valves were chosen by the surgeon s preference. All patients gave informed consent, and this study was approved by the local ethics committee (No. 2234/08). Patient demographic data are summarized in online supplemental Table 2. There is no significant difference between the groups concerning aortic cross clamp time and annulus dimension. Echocardiographic Measurements All echocardiographic examinations were performed transthoracally by 3 experienced echocardiographers. Transvalvular flow velocity, pressure gradient, and velocity time integral (VTI) were measured using continuous-wave Doppler. Pulsed-wave Doppler was used for the measurements in the left ventricular outflow tract (LVOT). Figure 2. Intrasupraannular (left) versus complete supraannular (right) position. The positions of pledgets and the noneverting suture technique are visualized.

3 Ruzicka et al Evaluating the Complete Supraannular Concept S141 Table. Hemodynamic Results Bovine Porcine Magna (n 169) Soprano (n 58) Mosaic (n 46) Ultra (n 17) Epic (n 46) Annulus 23 MPG a,b (73) c (22) a (22) b,c (6) (12) EOA (73) g (22) g (21) (6) (12) EOA i (73) (22) (7) (5) (12) Annulus MPG d (n 56) (18) (13) (3) d (20) EOA h (56) (18) h (13) (3) (20) EOA i (56) (18) (8) (3) (19) Annulus 24 MPG e (n 40) f (18) (11) (8) e,f (14) EOA (36) i (18) (11) (8) (14) i EOA i (36) (18) (7) (8) (14) n is shown in parentheses. a P 0.008, b P , c P 0.003, d P 0.001, e P , f P 0.017, g P 0.021, h P 0.046, i P We used following calculations for the parameters: MPG (mm Hg) mean aortic pressure gradient mean LVOT pressure gradient EOA (cm 2 ) (LVOT diameter [cm] 0.5) 2 VTI LVOT /VTI Valve (continuity equation) EOA i (cm 2 /m 2 ) EOA/BSA (BSA calculated by the DuBois formula) EOF EOA (cm 2 )/Anatomic annulus area (cm 2 ) where MPG is mean pressure gradient, EOA is effective orifice area, and BSA is body surface area. Patient-Prosthesis Mismatch Patient-prosthesis mismatch was defined by Blais et al 17 as severe for an EOA i of 0.65 cm 2 /m 2, as moderate for an EOA i 0.65 cm 2 and 0.85 cm 2 /m 2, and as clinically not significant if EOA i was 0.85 cm 2 /m 2. For the calculation of the EOA i we only used EOAs that were measured 6 months after surgery. Geometric Data Geometric data were provided by the manufacturers and are summarized for comparison in online supplemental Table 3. Literature Review We did a complete PubMed literature review and used the software Endnote (Thomson Reuters, New York, NY) for archiving the reviewed literature. The following key word combinations were used for our research: Mosaic hemodynamic, Mosaic Ultra hemodynamic, Epic supra hemodynamic, Perimount Magna hemodynamic, and Sorin Soprano hemodynamic. Statistical Analysis Statistical analysis was performed with the Statistical Package for the Social Science 16 for Windows (SPSS, Inc, Chicago, Ill). Continuous data are presented as mean standard deviation. To compare categorical data, the 2 test was performed. Ordinal data in more than 2 groups were compared with the Kruskal-Wallis test. To detect differences between groups, 2-way ANOVA with Bonferroni correction for multiple testing was applied. Results Mean Pressure Gradients The Magna demonstrated the lowest MPGs and had statistically significant lower gradients than the Mosaic and the Ultra (for annulus diameters 23 mm) or the Epic (for annulus diameters 23 mm). Additionally, in small annulus diameters, the Soprano had significantly lower MPGs than the Ultra and for large annulus sizes it demonstrated lower MPGs than the Epic (Table). Effective Orifice Area The Soprano bioprosthesis showed significantly larger EOAs compared with the Mosaic valve in patients with an annulus size 23 mm and compared with the Epic valve in patients with annulus diameters 23 mm. In addition, the Magna demonstrated significantly larger EOAs than the Mosaic in patients with annulus diameters of 23 to 24 mm (Table). According to the 2-way ANOVA, there was no interaction effect between annulus size and valve type. Indexed EOA and Patient-Prosthesis Mismatch The indexed EOA was comparable in all groups (Table), and no interaction effect was seen between annulus size and valve type. There was a significant difference concerning the occurrence of severe patient-prosthesis mismatch in the small annulus analysis between the different valve types. Severe patient-prosthesis mismatch occurred most frequently in the group with small annulus and Mosaic valve. No valve type avoided mismatch completely (Figure 3). Interaction Effects The 2-way ANOVA showed that annulus size has a significant main effect on MPG (F 3.70, P 0.05). Small annulus sizes had significantly higher MPG values. This was a moderate influence (partial ETA ). The valve type also had a significant main effect on MPG (F , P 0.001), and the influence was higher (partial ETA ), indicating that depending on the valve type MPG defers significantly. According to the 2-way ANOVA, there was no interaction effect between annulus size and valve type.

4 S142 Circulation September 15, 2009 Figure 3. Patient-prosthesis mismatch over all valve prostheses. Severe mismatch is found in all prostheses. Annulus size and prosthesis type also had a significant influence on EOA (for annulus size: P 0.001; for prosthesis size: P 0.001), but no interaction effect could be shown (P 0.14, partial ETA ). Only annulus size had significant influence on EOA i (P 0.001). There was also no interaction effect between annulus size and prosthesis type (P 0.287, partial ETA ). Geometric Data Comparison of the different manufacturers data resulted in different geometric dimensions of internal diameters for prostheses with the same valve size label. Online supplemental Table 3 depicts the geometric data provided by the manufacturers. Literature Overview We analyzed the hemodynamic data given in all previous publications. Ten publications (3 in vitro studies) examined the Magna, 16 studies (1 in vitro) the Mosaic (the Mosaic prosthesis was examined until 2004, as there are more than 40 studies available), 4 studies (4 in vitro) the Ultra, 3 studies (2 in vitro) the Epic, and 6 studies (1 in vitro) the Soprano valve. Online supplemental Table 4 (online-only Data Supplement) gives a complete overview of the PubMed research. Valve Distribution In 51 patients (15%), a prosthesis with an internal orifice diameter (IOD) larger or equal to the annulus diameter could be implanted (online supplemental Table 6). The MPG in this group was significantly lower ( mm Hg versus mm Hg, P 0.001). Discussion Comparing hemodynamic parameters of different heart valve prostheses on the basis of labeled valve size is always associated with the difficulty of nonnormalized valve-size labels. First, the Soprano valve labels do not correlate to other valve labels (even numbers instead of uneven numbers). Second, as shown in online supplemental Table 3, there are differences in geometric dimensions for valves of the same labeled valve size. We addressed this problem in a prior study. 13 Therefore, hemodynamic comparison studies based on labeled valve size will end in misleading results. For example, a comparison of a 23 Magna with a 23 Ultra implies comparison of a prosthesis with an IOD of 22 mm (Magna) to a prosthesis with an internal diameter of 20.5 mm (Ultra). The diameter difference of 1.5 mm results in a different geometric orifice area of 50 mm 2, which might lead to a lower pressure gradient for the valve with the larger IOD. In this study, the patient s aortic annulus is used as a basis for comparisons. It is important to distinguish between the geometric and the effective orifice area of any type of valve prosthesis. The geometric orifice area is derived from the IOD. The kind of leaflet material (pericardial, porcine, or even carbon in mechanical valves) is completely irrelevant in this approach. In contrast, the effective orifice area is an echocardiographically obtained, functional in vivo measurement. Our group demonstrated that the geometric orifice area of a valve

5 Ruzicka et al Evaluating the Complete Supraannular Concept S143 prosthesis has no impact on hemodynamic performance (eg, patient-prosthesis mismatch). 1,8,14,18,19 In the present study, we focus on aortic bioprostheses designed for and implanted in a complete supraannular position. The design of a complete supraannular heart valve prosthesis aims to optimize the area available for the bloodstream through the aortic root orifice. Regular intraannular or intrasupraannular positioned prostheses are positioned inside the aortic annulus, which narrows the orifice. By positioning the prosthesis on top of the annulus (complete supraannular position), the stent of the prosthesis is meant not to protrude into the orifice and ideally does not impair the bloodstream (Figure 2). 1,4,8,13 The advantage of complete supraannular placement is of special importance in patients with a small aortic annulus. 1,7,8,20 23 In our collective, the focus was not set on a special annulus size but on a normal distribution of annulus sizes within the patient group. All valves have small differences in design (online supplemental Table 3 and Figure 2) and can be divided into 2 groups, either bovine or porcine tissue valves. Prior comparisons of bovine and porcine heart valves showed a superiority of bovine tissue prostheses, especially in small annulus sizes. 14,24 Those studies only compared 2 different valves, whereas this study includes 2 bovine and 3 porcine prostheses and might help to stress the findings in those prior studies. Discussion of Results Mean Pressure Gradients Our results imply a general tendency toward lower mean pressure gradients and larger effective orifice areas for bovine prostheses compared with porcine prostheses. The largest difference appears to be between the Magna and the Ultra prostheses in small annulus groups (MPG difference, 10 mm Hg). The hemodynamic data show satisfactory pressure gradients for all valves in all annulus sizes, except for the Mosaic Ultra in small annulus sizes. Pressure gradients of the Magna, the Mosaic, and the Soprano are in accord with findings of other groups. 1,3 6,8,13,14,20,22,25 31 There are preliminary in vivo hemodynamic data available regarding the Mosaic Ultra 10 and only 1 in vivo evaluation of the Epic, which was published previously by our group. 4 The in vitro data concerning hemodynamic gradients of the Epic and the Ultra correspond well with our results. 7,11,12,32,33 As described in other studies, the bovine prostheses appear to be superior concerning pressure gradient and effective orifice areas. 1,8,14,24 In this study, the Magna as well as the Soprano showed a clear tendency toward lower pressure gradients and larger EOAs throughout all annulus size groups when compared with porcine prostheses. Patient-prosthesis mismatch is a hot topic in the valve surgery community, and there are still many controversies. In the past, we propagated the method of Blais et al. 17 This might be an arbitrarily defined value to distinguish between mismatch and no mismatch and might lead to a loss of information because of changing the interval scaled data to categorical data. Nevertheless, to be comparable to former studies, we analyzed our data as described and could find patient-prosthesis mismatch in all different valve prostheses. Although the supraannular concept is meant to prevent patient-prosthesis mismatch, in this study patient-prosthesis mismatch was not completely avoidable. Valve Distribution The complete supraannular concept was realized in 15% of our patients. These patients benefit in terms of lower MPGs. Hence, we showed that the complete supraannular concept, if put into practice, results in a better hemodynamic performance of the bioprostheses. To our knowledge, there is no other study showing this result. Literature Review Magna There are 3 in vitro studies 11,32,33 and 7 in vivo studies of the Magna. 1,3,6,8,25 27 The annulus diameter is used in 5 of the 7 in vivo studies as a comparing parameter (online supplemental Table 4). The hemodynamic gradients are low for all annulus sizes; the lowest gradients were described by Botzenhardt in patients with an annulus diameter of 22 to 23 mm (7.3 mm Hg) 27 and the highest in patients with an annulus diameter of 18 to 22 mm (13.2 mm Hg). 8 Mosaic The Mosaic valve is by far the prosthesis that was studied most (online supplemental Table 4). There are 33 studies listed in PubMed; we only added the latest 16 in our table. 1,4,6 9,14,20 22,28 30,32,34 37 Most of the studies are in vivo evaluations, and the lowest pressure gradients were around 10 mm Hg in patients with an annulus of 27 mm. 28 In patients with an annulus 22 mm, different hemodynamic gradients were reported, ranging from 13 mm Hg to 20 mm Hg. 4,6,28 Ultra There are no published clinical data available for the Ultra bioprosthesis except for 1 poster presentation showing MPGs 20 mm Hg in 55% of the patients. 10 These results correspond to our in vivo study demonstrating gradients 20 mm Hg for annulus sizes smaller than 23 mm. There are 4 in vitro studies, 3 of them published by the Frankfurt group 11,32,33 and 1 published by Bottio et al 12 regarding the Mosaic Ultra. Epic We published 1 in vivo evaluation of the Epic recently, 4 but apart from that there are only 2 in vitro studies available, published by an Italian group. 7,12 Soprano Six studies evaluated the Soprano. 1,7,8,13,30,31 Five of these studies are in vivo studies, 1,8,13,30,31 and only 2 study groups 1,8 used annulus diameters to compare the prostheses instead of using the labeled valve size for comparisons. Our results showed satisfactory pressure gradients, and the results were comparable to findings by Botzenhardt. 8 Pavoni 31 found surprisingly low gradients, all 10 mm Hg (for labeled valve sizes between 20 and 24), but only labeled valve sizes were given by the authors. We and other authors who demonstrated in vivo results focused on the systolic function of heart valve prostheses.

6 S144 Circulation September 15, 2009 Some authors who published in vitro studies focused on other aspects such as calcification and valve opening and closing characteristics. They identified a faster and more severe calcification in pericardial valves leading to a prolongation of closing time and higher closing volumes (altered diastolic function) but an almost undisturbed systolic function. 11,32,33 If this has any influence on durability, further long-term studies must be done. Thus far, there are not any consistent data showing the superiority of any valve material as far as valve durability is concerned. Even though there are many studies concerning hemodynamics of valve prostheses published, it is difficult to compare the data. Our literature overview shows 2 major factors that cause hampered comparisons of different valve prostheses: first, the use of the labeled valve size instead of the annulus diameter, and second, the different grouping of patients depending on size or annulus diameter. Limitations A clear limitation of our study is the small number of patients. Especially in the Ultra group we had very few patients included because this valve was included in our institutional valve portfolio only recently. This might bias the results on the Ultra. This was a nonrandomized study, and patients were included consecutively into the study. The surgeon chose the valve by his or her preferences, which might also bias the results of our study. All surgeons used the same implantation technique for complete supraannular valves. The only difference between various surgeons could be the perseveration in choosing the largest valve possible for a patient with a given anatomy. Thus, we did an analysis of the relation of annulus diameter and implanted valve size label (online supplemental Table 5) and could demonstrate that most patients had annulus sizes around 23 mm and received prostheses with a valve size label around 23. A further analysis, namely a comparison of annulus diameter and real internal valve diameter (online supplemental Table 6) revealed that the bovine prostheses led to a more frequent implantation of prostheses with larger internal diameter than the annulus diameter, which, we assume, will lead to a beneficial hemodynamic performance. Conclusion In a prior study, we emphasized our patient-centered view not only to compare valve hemodynamics but to answer the question of what should be the ideal prosthesis for a patient with a certain annulus diameter. 8 In a cardiothoracic surgery clinic, the surgeon has the possibility to choose between different prostheses and is confronted with a special situation: Some parameters (such as prosthesis type and implantation position) are changeable and some are given and not changeable (patient s annulus diameter, body surface area), and the patient s preferences according to his or her lifestyle must also be considered. 38 The surgeon must identify the prosthesis that will be the most promising one for the patient (in means of hemodynamics and long-term performance). Keeping this in mind, the implantation of a complete supraannular prosthesis is a good approach to optimize the blood flow through the aortic annulus. Nevertheless, in this study, we showed that even with this concept, high pressure gradients and patient-prosthesis mismatch cannot be prevented completely. The choice of a bovine prosthesis will optimize the systolic hemodynamic performance. Disclosures Dr Bauernschmitt received honoraria for consulting and lectures from Sorin above US$10,000. Dr Bleiziffer received honoraria for lectures from Datascope below US$10,000. Dr Eichinger received lecture fees from St Jude Medical, Inc and Medtronic, Inc both below US$10,000. References 1. Wagner IM, Eichinger WB, Bleiziffer S, Botzenhardt F, Gebauer I, Guenzinger R, Bauernschmitt R, Lange R. Influence of completely supraannular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulus. J Thorac Cardiovasc Surg. 2007;133: Eichinger WB, Botzenhardt F, Guenzinger R, Bleiziffer S, Keithahn A, Bauernschmitt R, Lange R. The effective orifice area/patient aortic annulus area ratio: a better way to compare different bioprostheses? A prospective randomized comparison of the Mosaic and Perimount bioprostheses in the aortic position. J Heart Valve Dis. 2004;13: Borger MA, Nette AF, Maganti M, Feindel CM. Carpentier-Edwards Perimount Magna valve versus Medtronic Hancock II: a matched hemodynamic comparison. Ann Thorac Surg. 2007;83: Ruzicka DJ, Eichinger WB, Hettich IM, Bleiziffer S, Bauernschmitt R, Lange R. Hemodynamic performance of the new St Jude Medical Epic Supra porcine bioprosthesis in comparison to the Medtronic Mosaic on the basis of patient annulus diameter. J Heart Valve Dis. 2008;17: Eichinger WB, Botzenhardt F, Gunzinger R, Kemkes BM, Sosnowski A, Maiza D, Coto EO, Bleese N. European experience with the Mosaic bioprosthesis. J Thorac Cardiovasc Surg. 2002;124: Dalmau MJ, Maria Gonzalez-Santos J, Lopez-Rodriguez J, Bueno M, Arribas A, Nieto F. One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprosthesis in the aortic position: a prospective randomized study. Interactive Cardiovasc Thoracic Surg. 2007;6: Gerosa G, Tarzia V, Rizzoli G, Bottio T. Small aortic annulus: the hydrodynamic performances of 5 commercially available tissue valves. J Thorac Cardiovasc Surg. 2006;131: Botzenhardt F, Eichinger WB, Bleiziffer S, Guenzinger R, Wagner IM, Bauernschmitt R, Lange R. Hemodynamic comparison of bioprostheses for complete supra-annular position in patients with small aortic annulus. J Am Coll Cardiol. 2005;45: Bleiziffer S, Eichinger WB, Wagner I, Guenzinger R, Bauernschmitt R, Lange R. The Toronto root stentless valve in the subcoronary position is hemodynamically superior to the mosaic stented completely supraannular bioprosthesis. J Heart Valve Dis. 2005;14: Buck T. Echocardiographic detection of high pressure gradients and small orifice areas in a novel aortic valve bioprosthesis necessitating termination of utilization. Presented at European Society of Cardiology Congress. 2007; Vienna, Austria. 11. Bakhtiary F, Dzemali O, Steinseiffer U, Schmitz C, Glasmacher B, Moritz A, Kleine P. Opening and closing kinematics of fresh and calcified aortic valve prostheses: an in vitro study. J Thorac Cardiovasc Surg. 2007;134: Bottio T, Tarzia V, Rizzoli G, Gerosa G. The changing spectrum of bioprostheses hydrodynamic performance: considerations on in-vitro tests. Interactive Cardiovasc Thoracic Surg. 2008;7: Eichinger WB, Botzenhardt F, Wagner I, Bleiziffer S, Ruzicka DJ, Guenzinger R, Bauernschmitt R, Lange R. Hemodynamic evaluation of the Sorin Soprano bioprosthesis in the completely supra-annular aortic position. J Heart Valve Dis. 2005;14: Eichinger WB, Botzenhardt F, Keithahn A, Guenzinger R, Bleiziffer S, Wagner I, Bauernschmitt R, Lange R. Exercise hemodynamics of bovine versus porcine bioprostheses: a prospective randomized comparison of the mosaic and Perimount aortic valves. J Thoracic Cardiovasc Surg. 2005;129:

7 Ruzicka et al Evaluating the Complete Supraannular Concept S Gansera B, Botzenhardt F, Gunzinger R, Spiliopoulos K, Angelis I, Kemkes BM. The Mosaic bioprosthesis in the aortic position: seven years results. J Heart Valve Dis. 2003;12: Eichinger WB, Hettich IM, Ruzicka DJ, Holper K, Schricker C, Bleiziffer S, Lange R. Twenty-year experience with the St Jude medical Biocor bioprosthesis in the aortic position. Ann Thoracic Surg. 2008;86: Blais C, Dumesnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108: Bleiziffer S, Eichinger WB, Hettich I, Guenzinger R, Ruzicka D, Bauernschmitt R, Lange R. Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method? Heart (Br Cardiac Society). 2007;93: Bleiziffer S, Eichinger WB, Lange R. Letter by Bleiziffer et al regarding article, Long-term outcomes after valve replacement for low-gradient aortic stenosis: impact of prosthesis-patient mismatch. Circulation. 2006;114:e627 e Tanigawa K, Eishi K, Yamachika S, Hashizume K, Tada S, Yamane K, Izumi K, Takai H, Miura T, Nakaji S. Comparison of the effects of aortic valve replacement using 19-mm Carpentier-Edwards Perimount bioprosthesis and 19-mm Medtronic Mosaic bioprosthesis. Ann Thorac Cardiovasc Surg. 2008;14: Kirsch ME, Tzvetkov B, Vermes E, Pouzet B, Sauvat S, Loisance D. Clinical and hemodynamic performance of the 19-mm Medtronic mosaic bioprosthesis. J Heart Valve Dis. 2005;14: Seitelberger R, Bialy J, Gottardi R, Seebacher G, Moidl R, Mittelbock M, Simon P, Wolner E. Relation between size of prosthesis and valve gradient: comparison of two aortic bioprosthesis. Eur J Cardiothorac Surg. 2004;25: Jamieson WR, Janusz MT, MacNab J, Henderson C. Hemodynamic comparison of second- and third-generation stented bioprostheses in aortic valve replacement. Ann Thoracic Surg. 2001;71(5 Suppl): S282 S Chambers JB, Rajani R, Parkin D, Rimington HM, Blauth CI, Venn GE, Young CP, Roxburgh JC. Bovine pericardial versus porcine stented replacement aortic valves: early results of a randomized comparison of the Perimount and the Mosaic valves. J Thoracic Cardiovasc Surg. 2008;136: Dalmau MJ, Mariagonzalez-Santos J, Lopez-Rodriguez J, Bueno M, Arribas A. The Carpentier-Edwards Perimount Magna aortic xenograft: a new design with an improved hemodynamic performance. Interactive Cardiovasc Thoracic Surg. 2006;5: Totaro P, Degno N, Zaidi A, Youhana A, Argano V. Carpentier-Edwards PERIMOUNT Magna bioprosthesis: a stented valve with stentless performance? J Thoracic Cardiovasc Surg. 2005;130: Botzenhardt F, Eichinger WB, Guenzinger R, Bleiziffer S, Wagner I, Bauernschmitt R, Lange R. Hemodynamic performance and incidence of patient-prosthesis mismatch of the complete supraannular Perimount magna bioprosthesis in the aortic position. Thoracic Cardiovasc Surgeon. 2005;53: Riess FC, Bader R, Cramer E, Hansen L, Kleijnen B, Wahl G, Wallrath J, Winkel S, Bleese N. Hemodynamic performance of the Medtronic Mosaic porcine bioprosthesis up to ten years. Ann Thoracic Surg. 2007; 83: John A, Khan Z, Kuo J, Allen SM, Marchbank AJ, Dalrymple-Hay MJ, Lewis CT, Twine G, Edwards M, Unsworth-White J. A prospective randomized comparison of Medtronic Mosaic and Carpentier- Edwards-SAV in the aortic position: an interim report. J Heart Valve Dis. 2006;15: Nozohoor S, Nilsson J, Luhrs C, Roijer A, Sjogren J. Influence of prosthesis-patient mismatch on diastolic heart failure after aortic valve replacement. Ann Thoracic Surg. 2008;85: Pavoni D, Badano LP, Musumeci SF, Frassani R, Gianfagna P, Mazzaro E, Livi U. Results of aortic valve replacement with a new supra-annular pericardial stented bioprosthesis. Ann Thoracic Surg. 2006;82: Dzemali O, Bakhtiary F, Steinseiffer U, Schmitz C, Glasmacher B, Moritz A, Kleine P. Hemodynamic performance of fresh and calcified aortic valve prostheses in critical low stroke volume. J Heart Valve Dis. 2008;17: Bakhtiary F, Dzemali O, Steinseiffer U, Schmitz C, Glasmacher B, Moritz A, Kleine P. Hydrodynamic comparison of biological prostheses during progressive valve calcification in a simulated exercise situation: an in vitro study. Eur J Cardiothorac Surg. 2008;34: Bakhtiary F, Schiemann M, Dzemali O, Wittlinger T, Doss M, Ackermann H, Moritz A, Kleine P. Stentless bioprostheses improve postoperative coronary flow more than stented prostheses after valve replacement for aortic stenosis. J Thoracic Cardiovasc Surg. 2006;131: Bakhtiary F, Abolmaali N, Dzemali O, Wittlinger T, Doss M, Moritz A, Kleine P. Impact of mechanical and biological aortic valve replacement on coronary perfusion: a prospective, randomized study. J Heart Valve Dis. 2006;15: Borger MA, Carson SM, Ivanov J, Rao V, Scully HE, Feindel CM, David TE. Stentless aortic valves are hemodynamically superior to stented valves during mid-term follow-up: a large retrospective study. Ann Thoracic Surg. 2005;80: Walther T, Lehmann S, Falk V, Metz S, Doll N, Rastan A, Viehweg M, Richter M, Gummert J, Mohr FW. Prospectively randomized evaluation of stented xenograft hemodynamic function in the aortic position. Circulation. 2004;110(11 Suppl 1):II-74 II Bonow RO, Carabello BA, Kanu C, de Leon AC Jr, Faxon DP, Freed MD, Gaasch WH, Lytle BW, Nishimura RA, O Gara PT, O Rourke RA, Otto CM, Shah PM, Shanewise JS, Smith SC Jr, Jacobs AK, Adams CD, Anderson JL, Antman EM, Faxon DP, Fuster V, Halperin JL, Hiratzka LF, Hunt SA, Lytle BW, Nishimura R, Page RL, Riegel B. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (writing committee to revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): developed in collaboration with the Society of Cardiovascular Anesthesiologists: endorsed by the Society for Cardiovascular Angiography and Interventions and the Society of Thoracic Surgeons. Circulation. 2006;114:e84 e231.

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Speaker's name: I have the following potential conflicts of interest to report: Proctorship for Medtronic Agenda Failure modes

More information

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,

More information

Late haemodynamic performance and survival after aortic valve replacement with the Mosaic bioprosthesis

Late haemodynamic performance and survival after aortic valve replacement with the Mosaic bioprosthesis Interactive CardioVascular and Thoracic Surgery 19 (2014) 756 762 doi:10.1093/icvts/ivu238 Advance Access publication 12 July 2014 ORIGINAL ARTICLE ADULTCARDIAC Late haemodynamic performance and survival

More information

Surgery for Acquired Cardiovascular Disease

Surgery for Acquired Cardiovascular Disease Bovine pericardial versus porcine stented replacement aortic valves: Early results of a randomized comparison of the Perimount and the Mosaic valves John B. Chambers, MD, FACC, Ronak Rajani, MD, MRCP,

More information

Aortic valve replacement: is porcine or bovine valve better?

Aortic valve replacement: is porcine or bovine valve better? Interactive CardioVascular and Thoracic Surgery Advance Access published December 4, 2012 Interactive CardioVascular and Thoracic Surgery (2012) 1 13 doi:10.1093/icvts/ivs447 BEST EVIDENCE TOPIC Aortic

More information

Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized study

Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized study European Journal of Cardio-thoracic Surgery 39 (2011) 844 852 www.elsevier.com/locate/ejcts Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of

More information

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,

More information

CoreValve in a Degenerative Surgical Valve

CoreValve in a Degenerative Surgical Valve CoreValve in a Degenerative Surgical Valve Ran Kornowski, MD, FESC, FACC Chairman Department of Cardiology Rabin Medical Center, Petach Tikva, Israel Disclosure Statement of Financial Interest I, Ran Kornowski,

More information

The implantation of bioprostheses is the preferred. Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement

The implantation of bioprostheses is the preferred. Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Shahab Nozohoor, MD, Johan Nilsson, MD, PhD, Carsten Lührs, MD, Anders Roijer, MD, PhD, and Johan Sjögren,

More information

Management of Difficult Aortic Root, Old and New solutions

Management of Difficult Aortic Root, Old and New solutions Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult

More information

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years Surgery for Acquired Cardiovascular Disease Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years W. R. Eric Jamieson, MD, Lawrence H. Burr, MD, Robert T. Miyagishima,

More information

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles

More information

PPM: How to fit a big valve in a small heart

PPM: How to fit a big valve in a small heart PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat

More information

The St. Jude Medical Biocor Bioprosthesis

The St. Jude Medical Biocor Bioprosthesis The St. Jude Medical Biocor Bioprosthesis Clinical Evidence of Long-term Durability Long-term Biocor Experience A Review and Comparative Assessment Long-term Biocor Stented Tissue Valve Studies Twenty-year

More information

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient

More information

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis Acquired Cardiovascular Disease Mykén and Bech-Hansen A 2-year experience of 1712 patients with the Biocor porcine bioprosthesis Pia S. U. Mykén, MD, PhD, a and Odd Bech-Hansen, MD, PhD b Objective: The

More information

Patient/prosthesis mismatch: how to evaluate and when to act?

Patient/prosthesis mismatch: how to evaluate and when to act? Patient/prosthesis mismatch: how to evaluate and when to act? Svend Aakhus, MD, PhD Oslo University Hospital, Norway Disclosures: No conflict of interest Types of aortic valve prostheses (AVR) Mechanical

More information

Prosthetic valve dysfunction: stenosis or regurgitation

Prosthetic valve dysfunction: stenosis or regurgitation Prosthetic valve dysfunction: stenosis or regurgitation Jean G. Dumesnil MD, FRCP(C), FACC, FASE(Hon) Quebec Heart and Lung Institute, Québec, Québec No disclosures Possible Causes of High Gradients in

More information

Comparison of eight prosthetic aortic valves in a cadaver model

Comparison of eight prosthetic aortic valves in a cadaver model Thomas Jefferson University Jefferson Digital Commons Department of Surgery Faculty Papers Department of Surgery July 2007 Comparison of eight prosthetic aortic valves in a cadaver model Benjamin A. Youdelman

More information

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Augusto D Onofrio, MD, Stefano Auriemma, MD, Paolo Magagna, MD, Alessandro Favaro, MD,

More information

Long-term mortality is increased in patients after aortic valve replacement

Long-term mortality is increased in patients after aortic valve replacement Bakhtiary et al Surgery for Acquired Cardiovascular Disease Stentless bioprostheses improve postoperative coronary flow more than stented prostheses after valve replacement for aortic stenosis Farhad Bakhtiary,

More information

ORIGINAL ARTICLE. Thorsten Hanke a, *, Efstratios I. Charitos a,, Hauke Paarmann b, Ulrich Stierle a and Hans-H. Sievers a. Abstract INTRODUCTION

ORIGINAL ARTICLE. Thorsten Hanke a, *, Efstratios I. Charitos a,, Hauke Paarmann b, Ulrich Stierle a and Hans-H. Sievers a. Abstract INTRODUCTION European Journal of Cardio-Thoracic Surgery (2013) 1 7 doi:10.1093/ejcts/ezt367 ORIGINAL ARTICLE Haemodynamic performance of a new pericardial aortic bioprosthesis during exercise and recovery: comparison

More information

Valve prosthesis-patient mismatch (PPM) was first defined

Valve prosthesis-patient mismatch (PPM) was first defined Impact of Valve Prosthesis-Patient Mismatch on Short-Term Mortality After Aortic Valve Replacement Claudia Blais, BSc; Jean G. Dumesnil, MD; Richard Baillot, MD; Serge Simard, MS; Daniel Doyle, MD; Philippe

More information

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for

More information

How to Avoid Prosthesis-Patient Mismatch

How to Avoid Prosthesis-Patient Mismatch How to Avoid Prosthesis-Patient Mismatch Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE, FESC Canada Research Chair in Valvular Heart Diseases INSTITUT UNIVERSITAIRE DE CARDIOLOGIE ET DE PNEUMOLOGIE DE QUÉBEC

More information

Stentless aortic valves. Current aspects

Stentless aortic valves. Current aspects Endorsed by proceedings in Intensive Care Cardiovascular Anesthesia EXPERT OPINION HSR Proceedings in Intensive Care and Cardiovascular Anesthesia 2012; 4(2): 77-82 Stentless aortic valves. Current aspects

More information

Copyright by ICR Publishers 2014

Copyright by ICR Publishers 2014 Comprehensive Hemodynamic Performance and Frequency of Patient-Prosthesis Mismatch of the St. Jude Medical Trifecta Bioprosthetic Aortic Valve Ajay Yadlapati 1, Jimmy Diep 3, Mary-Jo Barnes 2, Tristan

More information

PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic.

PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic. PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic. With more than 40 years of heart valve innovations, we took proven valve design concepts and adapted them for excellent implantability for

More information

Echocardiographic Evaluation of Mitral Valve Prostheses

Echocardiographic Evaluation of Mitral Valve Prostheses Echocardiographic Evaluation of Mitral Valve Prostheses Dennis A. Tighe, M.D., FACC, FACP, FASE Cardiovascular Medicine University of Massachusetts Medical School Worcester, MA www.asecho.org 1 Nishimura

More information

Impact of Prosthesis-Patient Mismatch on Long-Term Survival After Aortic Valve Replacement

Impact of Prosthesis-Patient Mismatch on Long-Term Survival After Aortic Valve Replacement Journal of the American College of Cardiology Vol. 53, No. 1, 9 9 by the American College of Cardiology Foundation ISSN 735-197/9/$36. Published by Elsevier Inc. doi:1.116/j.jacc.8.9.22 Valvular Heart

More information

Aortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve implantation

Aortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve implantation Gen Thorac Cardiovasc Surg (2016) 64:435 440 DOI 10.1007/s11748-016-0657-9 CURRENT TOPICS REVIEW ARTICLE Aortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve

More information

Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography

Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler and Dobutamine Stress Echocardiography Akira Sezai, MD, PhD, Yuji Kasamaki, MD, PhD, Keisuke Abe, RMS, Mitsumasa Hata,

More information

Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis

Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis CLINICS 2012;67(1):55-60 DOI:10.6061/clinics/2012(01)09 CLINICAL SCIENCE Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis Luis M. Astudillo, I Orlando

More information

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility

More information

Valvular Heart Disease

Valvular Heart Disease Valvular Heart Disease Roman M. Sniecinski, MD, FASE Associate Professor of Anesthesiology Emory University School of Medicine Learning Objectives Review the major pathophysiology of the most common heart

More information

Experience with 500 Stentless Aortic Valve Replacements

Experience with 500 Stentless Aortic Valve Replacements Experience with 500 Stentless Aortic Valve Replacements Dimitrios C. Iliopoulos, MD Cardiac Surgeon Ass. Professor of Surgery University of Athens, School of Medicine I declare no conflict of interest

More information

Bioprostheses are prone to continuous degeneration

Bioprostheses are prone to continuous degeneration Twenty-Year Experience With the St. Jude Medical Biocor Bioprosthesis in the Aortic Position Walter B. Eichinger, MD, Ina M. Hettich, MD, Daniel J. Ruzicka, MD, Klaus Holper, MD, Carolin Schricker, Sabine

More information

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden Long-Term Outcome of the Mitroflow Pericardial Bioprosthesis in the Elderly after Aortic Valve Replacement Johan Sjögren, Tomas Gudbjartsson, Lars I. Thulin Department of Cardiothoracic Surgery, Heart

More information

Patient prosthesis mismatch after mitral valve replacement: Myth or reality?

Patient prosthesis mismatch after mitral valve replacement: Myth or reality? Patient prosthesis mismatch after mitral valve replacement: Myth or reality? Pasquale Totaro, MD, a and Vincenzo Argano, MD b Objective: Determining the risk of patient prosthesis mismatch after mitral

More information

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction

More information

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://www.annalsthoracicsurgery.org/cme/ home. To take the CME activity related to this article, you must have either an STS member

More information

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background

More information

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)

More information

Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis?

Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? EuroValves 2015, Nice Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE FESC Canada Research Chair in Valvular Heart Diseases Université LAVAL Disclosure

More information

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability SOLO SMART TM The smart way to return to life Native-like performance now with stented-like implantability MANY PATIENTS NEED SUPERIOR HEMODYNAMIC PERFORMANCE TO RETURN TO THEIR NORMAL LIFESTYLE. 2 SOLO

More information

Prospective randomized evaluation of stentless vs. stented aortic biologic prosthetic valves in the elderly at five years

Prospective randomized evaluation of stentless vs. stented aortic biologic prosthetic valves in the elderly at five years doi:10.1510/icvts.008.18136 Interactive CardioVascular and Thoracic Surgery 8 (009) 449 453 www.icvts.org Follow-up papers - Valves Prospective randomized evaluation of stentless vs. stented aortic biologic

More information

TAVR: Echo Measurements Pre, Post And Intra Procedure

TAVR: Echo Measurements Pre, Post And Intra Procedure 2017 ASE Florida, Orlando, FL October 10, 2017 8:00 8:25 AM 25 min TAVR: Echo Measurements Pre, Post And Intra Procedure Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology Echo Lab Associate

More information

The diameter of the aortic valve is in direct proportion

The diameter of the aortic valve is in direct proportion The CarboMedics Top-Hat Supraannular Prosthesis José M. Bernal, MD, Rafael Martin-Duran, MD, José M. Rabasa, MD, and José M. Revuelta, MD Departments of Cardiovascular Surgery and Echocardiography, Hospital

More information

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.

More information

Surgery for Valvular Heart Disease. Reoperation of Left Heart Valve Bioprostheses According to Age at Implantation

Surgery for Valvular Heart Disease. Reoperation of Left Heart Valve Bioprostheses According to Age at Implantation Surgery for Valvular Heart Disease Reoperation of Left Heart Valve Bioprostheses According to Age at Implantation Vincent Chan, MD, MPH; Tarek Malas, MD; Harry Lapierre, MD; Munir Boodhwani, MMSc, MD;

More information

TAVR for Valve-In-Valve. Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology

TAVR for Valve-In-Valve. Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology TAVR for Valve-In-Valve Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology Temple Hearth and Vascular Institute Disclosures: Consultant: Cardiac Assist TAVR for

More information

Aortic stenosis is the most common acquired heart valve. The Toronto Root Bioprosthesis: Midterm Results in 186 Patients

Aortic stenosis is the most common acquired heart valve. The Toronto Root Bioprosthesis: Midterm Results in 186 Patients The Toronto Root Bioprosthesis: Midterm Results in 186 Patients Sven Lehmann, MD, Thomas Walther, MD, PhD, Sergey Leontyev, MD, Jörg Kempfert, MD, Jens Garbade, MD, Michael A. Borger, MD, PhD, and Friedrich

More information

In Vitro Two-Dimensional Echocardiographic Imaging of a Stented Porcine Bioprosthetic Valve: The Bent Strut Artifact

In Vitro Two-Dimensional Echocardiographic Imaging of a Stented Porcine Bioprosthetic Valve: The Bent Strut Artifact C 2008, the Author Journal compilation C 2008, Wiley Periodicals, Inc. DOI: 10.1111/j.1540-8175.2008.00753.x In Vitro Two-Dimensional Echocardiographic Imaging of a Stented Porcine Bioprosthetic Valve:

More information

The impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation

The impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation Editor s choice Scan to access more free content 1 Division of Cardiac Surgery, The Johns Hopkins Hospital, Baltimore, Maryland, USA 2 Division of Cardiac Surgery, University of Ottawa Heart Institute,

More information

TAVI EN INSUFICIENCIA AORTICA

TAVI EN INSUFICIENCIA AORTICA TAVI EN INSUFICIENCIA AORTICA Cesar Moris Profesor Cardiología Director Departamento del Corazón Hospital Universitario Central de Asturias Universidad de Oviedo OVIEDO -- ESPAÑA CONFLICTO DE INTERESES

More information

Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival

Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival ORIGINAL ARTICLES: ADULT CARDIAC SURGERY: To participate in The Annals of Thoracic Surgery CME Program, please visit http://cme.ctsnetjournals.org. Prosthesis-Patient Mismatch After Aortic Valve Replacement:

More information

CARDIACSURGERY TODAY. Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease

CARDIACSURGERY TODAY. Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease VOLUME 1 NUMBER 2 23 CARDIACSURGERY TODAY Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease EDITORS-IN-CHIEF Robert W Emery, St Paul, MN, USA Francesco Musumeci, Rome, Italy

More information

QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH

QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH 1,2 Radu A. SASCĂU 3 Cristina OLARIU 1,2 Cristian STĂTESCU 1 Internal Medicine Department, Gr.T.Popa University of Medicine and Pharmacy,

More information

25 different brand names >44 different models Sizes mm

25 different brand names >44 different models Sizes mm Types of Prosthetic Valves BIOLOGIC STENTED Porcine xenograft Pericardial xenograft STENTLESS Porcine xenograft Pericardial xenograft Homograft (allograft) Autograft PERCUTANEOUS MECHANICAL Bileaflet Single

More information

Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years

Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years Friedrich-Christian Riess, MD, Ralf Bader, MD, Eva Cramer, MD, Lorenz Hansen, MD, Bèr Kleijnen, MS, Gunther Wahl, MD,

More information

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim 42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim Current Guideline for AR s/p TOF Surgery is reasonable in adults with prior repair of

More information

Transcatheter Aortic Valve Implantation in Patients With Severe Aortic Stenosis and Small Aortic Annulus

Transcatheter Aortic Valve Implantation in Patients With Severe Aortic Stenosis and Small Aortic Annulus Journal of the American College of Cardiology Vol. 58, No. 10, 2011 2011 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2011.05.026

More information

Copyright by ICR Publishers 2015

Copyright by ICR Publishers 2015 A Comprehensive Fluid Dynamic and Geometric Study for an In-Vitro Comparison of Four Surgically Implanted Pericardial Stented Valves Giordano Tasca 1,2, Riccardo Vismara 2,4, Gianfranco Beniamino Fiore

More information

Stainless Steel. Cobalt-chromium

Stainless Steel. Cobalt-chromium Sapien is better than Corevalve! Raj R. Makkar, MD Associate Director, Cedars-Sinai Heart Institute Associate Professor, UCLA School of Medicine, Los Angeles Eberhard Grube: Pioneer in the field of TAVR

More information

Magdalena Erlebach 1, Michael Wottke 1, Marcus-André Deutsch 1, Markus Krane 1, Nicolo Piazza 2, Ruediger Lange 1, Sabine Bleiziffer 1

Magdalena Erlebach 1, Michael Wottke 1, Marcus-André Deutsch 1, Markus Krane 1, Nicolo Piazza 2, Ruediger Lange 1, Sabine Bleiziffer 1 Original Article on TAVI Redo aortic valve surgery versus transcatheter valve-in-valve implantation for failing surgical bioprosthetic valves: consecutive patients in a single-center setting Magdalena

More information

Valvular Heart Disease

Valvular Heart Disease Valvular Heart Disease Prosthesis-Patient Mismatch Predicts Structural Valve Degeneration in Bioprosthetic Heart Valves Willem Flameng, MD, PhD; Marie-Christine Herregods, MD, PhD; Monique Vercalsteren;

More information

CROWN PRT. Inservice implantation guide

CROWN PRT. Inservice implantation guide Inservice implantation guide DEVICE DESCRIPTION The Crown PRT aortic pericardial heart valve consists of a single piece of bovine pericardium that is preserved with glutaraldehyde and sewn onto a polyester

More information

Treatment of Bio-Prosthetic Valve Deterioration Using Transcatheter Techniques

Treatment of Bio-Prosthetic Valve Deterioration Using Transcatheter Techniques Treatment of Bio-Prosthetic Valve Deterioration Using Transcatheter Techniques Pablo Codner, Abid Assali, Hanna Vaknin-Assa, Katia Orvin, Ram Sharony, Leor Perl, Gabriel Greenberg, Marina Kupershmidt,

More information

Echo Evaluation of a Mitral Valve Prostheses Sunday, February 14, :50 2:10 PM 20 min

Echo Evaluation of a Mitral Valve Prostheses Sunday, February 14, :50 2:10 PM 20 min 2016 ASE State of the Art Echocardiography Course Tucson, AZ Echo Evaluation of a Mitral Valve Prostheses Sunday, February 14, 2016 1:50 2:10 PM 20 min 1 M U H A M E D S A R I Ć, M D, P H D D i r e c t

More information

Transcatheter valve-in-valve implantation for degenerated surgical bioprostheses

Transcatheter valve-in-valve implantation for degenerated surgical bioprostheses Review Article Transcatheter valve-in-valve implantation for degenerated surgical bioprostheses Dale J. Murdoch, John G. Webb Centre for Heart Valve Innovation, St. Paul s Hospital, Vancouver, Canada Contributions:

More information

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic,

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, transplantation and Vascular Surgery Hannover Medical School, Germany

More information

The stentless bioprosthesis has many salient features that

The stentless bioprosthesis has many salient features that Aortic Valve Replacement with the Medtronic Freestyle Xenograft Using the Subcoronary Implantation Technique D. Michael Deeb, MD The stentless bioprosthesis has many salient features that make it an attractive

More information

Case. 15-year-old boy with bicuspid AV Severe AR with moderate AS. Ross vs. AVR (or AVP)

Case. 15-year-old boy with bicuspid AV Severe AR with moderate AS. Ross vs. AVR (or AVP) Case 15-year-old boy with bicuspid AV Severe AR with moderate AS Ross vs. AVR (or AVP) AMC case 14-year-old boy with bicuspid AV Severe AS with mild AR Body size Bwt: 55 kg, Ht: 154 cm, BSA: 1.53 m 2 Echocardiography

More information

I will not discuss off label use or investigational use in my presentation.

I will not discuss off label use or investigational use in my presentation. I will not discuss off label use or investigational use in my presentation. Surgical valves Design and Durability Testing Potential Concerns Real Practice 1952-1962 1963-1966 1967-1969 1969-1977 1977-1984

More information

Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study

Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study Federico M. Asch, MD, David Heimansohn, MD, Daniel Doyle, MD, Walter Dembitsky, MD, Francis D. Ferdinand, MD,

More information

Reconstruction of the intervalvular fibrous body during aortic and

Reconstruction of the intervalvular fibrous body during aortic and Aortic and mitral valve replacement with reconstruction of the intervalvular fibrous body: An analysis of clinical outcomes Nilto C. De Oliveira, MD Tirone E. David, MD Susan Armstrong, MSc Joan Ivanov,

More information

Transcatheter Aortic Valve Replacement with Evolut-R

Transcatheter Aortic Valve Replacement with Evolut-R Transcatheter Aortic Valve Replacement with Evolut-R Department of Transcatheter Heart Valves and 2 nd Cardiothoracic Surgery Clinic K. Spargias, M.Chrissoheris, A.Halapas, I. Nikolaou, S.Pattakos Disclosures

More information

Transcatheter Mitral Valve Replacement How Close Are We?

Transcatheter Mitral Valve Replacement How Close Are We? Transcatheter Mitral Valve Replacement How Close Are We? Gregory Pavlides, MD, PhD, FACC, FESC Professor of Medicine Miscia Chair of Interventional Cardiology Director, Cardiac Catheterization Laboratories,

More information

We present the case of an asymptomatic, 75-year-old

We present the case of an asymptomatic, 75-year-old Images in Cardiovascular Medicine Asymptomatic Rupture of the Left Ventricle Lech Paluszkiewicz, MD; Stefan Ożegowski, MD; Mohammad Amin Parsa, MD; Jan Gummert, PhD, MD We present the case of an asymptomatic,

More information

The results of aortic valve (AV) surgery continue to improve

The results of aortic valve (AV) surgery continue to improve Predictors of Low Cardiac Output Syndrome After Isolated Aortic Valve Surgery Manjula D. Maganti, MSc; Vivek Rao, MD, PhD; Michael A. Borger, MD, PhD; Joan Ivanov, PhD; Tirone E. David, MD Background Low

More information

Comparison between three types of stented pericardial aortic valves (Trivalve trial): study protocol for a randomized controlled trial

Comparison between three types of stented pericardial aortic valves (Trivalve trial): study protocol for a randomized controlled trial Azarnoush et al. Trials 2013, 14:413 TRIALS STUDY PROTOCOL Comparison between three types of stented pericardial aortic valves (Trivalve trial): study protocol for a randomized controlled trial Kasra Azarnoush

More information

Long-Term Survival After Bovine Pericardial Versus Porcine Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter?

Long-Term Survival After Bovine Pericardial Versus Porcine Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter? Long-Term Survival After Bovine Pericardial Versus Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter? Asvin M. Ganapathi, MD, Brian R. Englum, MD, Jeffrey E. Keenan, MD, Matthew

More information

PROSTHETIC VALVE BOARD REVIEW

PROSTHETIC VALVE BOARD REVIEW PROSTHETIC VALVE BOARD REVIEW The correct answer D This two chamber view shows a porcine mitral prosthesis with the typical appearance of the struts although the leaflets are not well seen. The valve

More information

Mitral Programme Update

Mitral Programme Update Direct Flow Medical Innovations in the Aortic and Mitral Programmes Mitral Programme Update Azeem Latib MD EMO-GVM Centro Cuore Columbus and San Raffaele Scientific Institute, Milan, Italy Potential conflicts

More information

Disclosures Nothing to disclose

Disclosures Nothing to disclose Aortic Valve Disease Illustrative Cases December 8, 2009 David Stultz, MD, FACC Southwest Cardiology, Inc Handout available at http://www.drstultz.com Disclosures Nothing to disclose Objectives Understand

More information

Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers

Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers Thierry A Folliguet, MD, François Laborde, MD, Konstantinos Zannis, MD, Gabriel Ghorayeb, MD, Axel Haverich, MD, and Malakh

More information

Repair or Replacement

Repair or Replacement Surgical intervention post MitraClip Device: Repair or Replacement Saudi Heart Association, February 21-24 Rüdiger Lange, MD, PhD Nicolo Piazza, MD, FRCPC, FESC German Heart Center, Munich, Germany Division

More information

Influence of patient gender on mortality after aortic valve replacement for aortic stenosis

Influence of patient gender on mortality after aortic valve replacement for aortic stenosis Influence of patient gender on mortality after aortic valve replacement for aortic stenosis Jennifer Higgins, MD, W. R. Eric Jamieson, MD, Osama Benhameid, MD, Jian Ye, MD, Anson Cheung, MD, Peter Skarsgard,

More information

Neal Kleiman, MD Houston Methodist DeBakey Heart and Vascular Institute

Neal Kleiman, MD Houston Methodist DeBakey Heart and Vascular Institute Neal Kleiman, MD Houston Methodist DeBakey Heart and Vascular Institute Despite a 33 fold growth in the first five years, there is still tremendous variability among penetration in different countries

More information

Echocardiographic Evaluation of Aortic Valve Prosthesis

Echocardiographic Evaluation of Aortic Valve Prosthesis Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co-Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System

More information

The Edge-to-Edge Technique f For Barlow's Disease

The Edge-to-Edge Technique f For Barlow's Disease The Edge-to-Edge Technique f For Barlow's Disease Ottavio Alfieri, Michele De Bonis, Elisabetta Lapenna, Francesco Maisano, Lucia Torracca, Giovanni La Canna. Department of Cardiac Surgery, San Raffaele

More information

Outline. EuroScore II. Society of Thoracic Surgeons Score. EuroScore II

Outline. EuroScore II. Society of Thoracic Surgeons Score. EuroScore II SURGICAL RISK IN VALVULAR HEART DISEASE: WHAT 2D AND 3D ECHO CAN TELL YOU AND WHAT THEY CAN'T Ernesto E Salcedo, MD Professor of Medicine University of Colorado School of Medicine Director of Echocardiography

More information

Comprehensive Echo Assessment of Aortic Stenosis

Comprehensive Echo Assessment of Aortic Stenosis Comprehensive Echo Assessment of Aortic Stenosis Smonporn Boonyaratavej, MD, MSc King Chulalongkorn Memorial Hospital Bangkok, Thailand Management of Valvular AS Medical and interventional approaches to

More information

Late failure of transcatheter heart valves: An open question

Late failure of transcatheter heart valves: An open question Late failure of transcatheter heart valves: An open question A comparison with surgically implanted bioprosthetic heart valves. A. Rashid The Cardiothoracic Centre Liverpool, UK. Conflict of Interest Statement

More information

Percutaneous Therapy for Calcific Mitral Valve Disease

Percutaneous Therapy for Calcific Mitral Valve Disease 31 st Annual State of the Art Echocardiography San Diego, CA February 18, 2018 5:00 5:15 PM 15 min Percutaneous Therapy for Calcific Mitral Valve Disease Muhamed Sarić MD, PhD, MPA Director of Noninvasive

More information

The need for right ventricular outflow tract reconstruction

The need for right ventricular outflow tract reconstruction Polytetrafluoroethylene Bicuspid Pulmonary Valve Implantation James A. Quintessenza, MD The need for right ventricular outflow tract reconstruction and pulmonary valve replacement is increasing for many

More information

Reverse left atrium and left ventricle remodeling after aortic valve interventions

Reverse left atrium and left ventricle remodeling after aortic valve interventions Reverse left atrium and left ventricle remodeling after aortic valve interventions Alexandra Gonçalves, Cristina Gavina, Carlos Almeria, Pedro Marcos-Alberca, Gisela Feltes, Rosanna Hernández-Antolín,

More information

Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon?

Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon? Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon? Andrea Garatti, MD, Francesca Mori, MD, Francesco Innocente, MD, Alberto Canziani, MD, Piervincenzo

More information

Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position

Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position Chee-Siong Soo, MRCP, Mestres Ca, MD, Monica Tay, Joon-Kuan Yeoh, MRCP, Eugene Sim, FRCS, and M. Choo, FRCP,

More information