Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers

Size: px
Start display at page:

Download "Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers"

Transcription

1 Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers Thierry A Folliguet, MD, François Laborde, MD, Konstantinos Zannis, MD, Gabriel Ghorayeb, MD, Axel Haverich, MD, and Malakh Shrestha, MD Institut Mutualiste Montsouris, Paris, France; Medizinsche Hochschule Hannover, Hannover, Germany Background. The Perceval S bioprosthesis (Sorin Biomedica Cardio Srl, Sallugia, Italy) is a self-expanding valve designed to preserve aortic sinuses and sinotubular junction. We report the midterm results of a prospective, multicenter clinical study evaluating the safety and efficacy of this stented bioprosthesis in patients undergoing aortic valve replacement with or without concomitant procedures. Methods. From January 2007 to September 2011, a total of 208 high-risk patients (mean European system for cardiac operative risk evaluation: years) received a Perceval bioprosthesis in 2 European centers. Median follow up was months and 100% complete, and the total accumulated follow-up was 156 patient-years. Ten patients have reached a 4-year follow-up. Valve function was assessed in all patients. Results. Valve implantation resulted in significant improvement of patients symptoms. Mean preoperative and postoperative gradients were mm Hg and mm Hg, respectively, and preoperative and postoperative mean effective orifice areas were and cm 2. Survival at 12 months was 87.1%, success of implantation was 95%, and freedom from reoperation was 96%. In hospital mortality was 2.4%. During follow-up, 9 patients (4%) required reoperation for paravalvular regurgitation; 7 early and 2 late reoperations. Mean cross-clamp time (CCT) and extracorporeal circulation time (ECT) were, respectively, minutes and minutes, including 45 patients who underwent surgery through ministernotomy. Concomitant coronary bypass was done in 48 patients with mean CCT and ECT minutes. Conclusions. Perceval sutureless is a safe bioprosthesis that can easily be implanted, including by a minimally invasive technique. It provides excellent hemodynamic with significant clinical improvement. Overall, these data confirm the safety and utility of the Perceval bioprosthesis aortic valve replacement for high-risk patients. (Ann Thorac Surg 2012;93:1483 8) 2012 by The Society of Thoracic Surgeons Aortic valve replacement with biologic heart valves is the treatment of choice for aortic valve stenosis when it is symptomatic or with severe aortic stenosis ( 0.6 cm 2 /m 2 ), with left ventricular dysfunction, or for older patients [1]. Many clinical reports have shown excellent reports with pericardial and porcine valves with very low gradient and improved long-term durability, with results up to 20 years [2 12]. However, all these valves are mounted on a stent with a Dacron (DuPont, Wilmington, DE) cuff that allows the prosthesis to be sutured to the aortic annulus. In a small and calcified annulus it can be challenging to insert and a significant residual gradient is frequently observed afterward. Stentless valves have been designed in order to overcome some of the disadvantages of the stented valve [13, 14]. By removing the stent, these bioprostheses provide a greater orifice area and preserve the distensibility of the annulus as well as the sinotubular junction expansibility. Over their stented counterpart they offer a greater effective orifice area and lower hemodynamic Accepted for publication Jan 20, Address correspondence to Dr Folliguet, Department of Cardiac Surgery, L Institut Mutualiste Montsouris, 42 boulevard Jourdan, Paris, France; thierry.folliguet@imm.fr. gradients [15]. However, they are more difficult to insert, with increased cross-clamp time, and their immediate results do not translate in improved longterm results [16, 17]. In addition, some of these valves exhibit late structural valve failure compared with stented valves, with wearing and tearing at the commissures [18 20], therefore, the sutureless Perceval bioprosthesis (Sorin Biomedica Cardio Srl, Sallugia, Italy) was designed in order to obtain the hemodynamic benefits of the stentless valves without the increased difficulty in surgical implantation. The design of the Perceval S prosthesis stems from the intention to offer an alternative to traditional flexible prostheses (stented and stentless biologic valves) using conventional open-heart surgery. As a result of the sutureless implant procedure, patients could benefit from reducing aortic cross-clamp times, with subsequent overall reduction of surgical timing and therefore reduction in related risks by avoiding to pass the stitches through the annulus and sutures knotting, with consequent less risk of tearing the aortic annulus and wall, damaging the bundle of His, or embolizing foreign material in the vascular system [21]. We report our experience with the Perceval S bioprosthesis implanted at 2 different institutions by The Society of Thoracic Surgeons /$36.00 Published by Elsevier Inc doi: /j.athoracsur

2 1484 FOLLIGUET ET AL Ann Thorac Surg SUTURELESS PERCEVAL AVR: RESULTS 2012;93: Material and Methods Patient Population From January 2007 to September 2011, a total of 211 patients (83 males, 128 females; mean age years) were enrolled into this prospective, international, multicenter nonrandomized study. Participants were implanted with the Perceval sutureless bioprosthesis at 2 participating European investigational sites; Institut Mutualiste Montsouris (Paris, France) and Medizinsche Hochschule Hannover (Hannover, Germany). Inclusion and exclusion criteria are listed in Table 1. High-risk patients defined as a logistic EuroSCORE (European system for cardiac operative risk evaluation) greater than 5% requiring isolated aortic valve replacement, with or without concomitant procedures, were eligible to participate. The mean EuroSCORE was The Institutional Review Boards and the local Ethics Committee of each hospital gave approval for this study. All patients gave informed written consent and were informed adequately regarding the investigational status of the device, the associated benefits and risks, and the availability of alternative treatment options. Device Description The Perceval S prosthesis is made of a trileaflet bovine pericardial valve mounted on an expandable metal frame in nitinol (equiatomic alloy of nickel and titanium). The Perceval S prosthesis has the special feature that it does not have to be surgically sutured into the implant site as this function is performed by the armature which, as well as supporting the valve edges, adapts itself to the aortic root to which it is anchored. The armature has 2 cylindrical ring segments, on the proximal ( outflow ring) and distal ( inflow ring) sides, and a double set of posts that couple the 2 rings. The first set comprises straight posts designed to support the valve while the second set is sinusoidal, protruding from the cylindrical geometry to allow the prosthesis to be anchored to the aortic root in the Valsalva sinuses. The armature s superelastic alloy and its architectural design allow the device to be compressed to a diameter less than the working geometry in the phase before positioning at the site. The later release from constraint allows the superelastic materials to self-expand so as to reach a final diameter compatible with the aortic root. Thus, once in place and deployed the device is fixed to the annulus and the aorta, and does not expand or move. The biologic tissue was stabilized with glutaraldehyde, using the manufacturing process already tested with traditional biologic valves produced by Sorin Biomedica Cardio s.r.l. Currently, 3 sizes are available at 21, 23, and 25 mm for annuli ranging from 19 to 24 mm. Surgical Technique; Exposure and Incision in the Aorta The aortic incision is performed at the distal portion at the sinotubular junction to preserve a segment of ascending aorta above the prosthetic valve. The aortic valve should be excised at a position corresponding to the incision line of the native leaflets and the aortic annulus should be decalcified to prepare the implant site. Given the characteristics of the prosthesis, a complete decalcification of the aortic annulus is not necessary, but a regular annular profile is beneficial to aid coupling with the prosthesis and to minimize the risk of paraprosthetic leak. If necessary, a partial septal myectomy is recommended. The measurement of the aortic annulus diameter, after the decalcification procedure, must be carried out using dedicated sizers. Positioning and Release of the Prosthesis To ensure the correct positioning and orientation of the prosthesis, 3 suture threads were used to act as reference for accurate alignment of the inflow portion of the prosthesis with the insertion plane of the native leaflets. For each valve sinus, 1 stitch was positioned immediately 2 to 3 mm below the lowest portion of the native leaflet resection line (Fig 1). On the prosthesis, each thread was passed into a dedicated thread loop located at the midlevel of the inflow ring and aligned to the median part of the prosthetic sinuses. Once the prosthesis is connected to the 3 guidethreads, the release device was introduced into the aorta Table 1. Indications and Contraindications for the Sutureless Perceval S Indications Inclusion Criteria Subjects of age 65 years; Subjects with aortic valve stenosis or steno-insufficiency; Subjects in NYHA functional classes III and IV with the logistic EuroSCORE greater than 5%; Subjects who have agreed to participation in the clinical evaluation and who have signed the informed consent. EuroSCORE European system for cardiac operative risk evaluation; Contraindications Exclusion Criteria Subjects with aneurysmal dilation or dissection of the ascending aortic wall needing surgical correction; Subjects needing nonelective intervention; Subjects with aortic annulus (after procedure for decalcification) of dimensions such that the implantation of a valve of size 21 or 23 mm is not possible (direct intraoperative measurement with sizer); Subjects with congenital bicuspid aortic valve; Subjects with aortic root enlargement, where the ratio between observed and expected diameters (calculated as a function of age and patient body surface area) is 1.3; Subjects with known hypersensitivity to nickel alloys. NYHA New York Heart Association.

3 Ann Thorac Surg FOLLIGUET ET AL 2012;93: SUTURELESS PERCEVAL AVR: RESULTS 1485 Fig 1. Valve design features: button holes. Button holes allow correct axial-rotational positioning in the native aortic root. (LC left coronary; NC noncoronary; RC right coronary.) and parachuted down to the point at which it was blocked by the previously positioned stitches. This positioning was followed by freeing of the inflow portion of the prosthesis, allowing the inflow ring to come in contact with the annulus (Fig 2). After checking that the Perceval S prosthesis was well seated within the aortic root, the guide-threads were removed during balloon inflation. A balloon dilation of the inflow ring was performed after the total deployment of the prosthesis. As the balloon is completely deflated, the catheter can be removed from the operative field. Statistical Methods SAFETY ENDPOINTS. Safety of the Perceval S prosthesis was evaluated based on all cardiovascular complications related to cardiac valve replacement, as defined in reference 22. Each serious adverse event (death, structural deterioration, thrombosis, embolism, hemorrhage, endocarditis, clinically significant hemolysis, and nonstructural dysfunction (including paraprosthetic and intraprosthetic leak) were recorded during hospitalization and follow-up. This analysis was also performed for valve reinterventions, including those that entailed replacement of a valve prosthesis and those that did not result in explants of the prosthesis being assessed. Morbid events were stratified as early ( 30 days) or late ( 30 days) postoperative events. The analysis of death included all deaths and analysis of valve-related deaths. Rates of early postoperative events were calculated as a percentage of patients who experience each event, calculated as the number of events divided by the number of patients. For the late postoperative period, linearized rates were calculated as the number of events per 100 patient-years of patient exposure. Actuarial analysis that includes all events was used to show the probability of freedom from first occurrence of each complication at the end of different intervals. The Kaplan-Meier method was used to produce actuarial curves and comparison of curves among patient groups (eg, sex, New York Heart Association [NYHA] class) was performed with the log-rank test at a 0.05 level of significance. Cox proportional hazard regression was applied to identify any risk factors collected in this study that may affect the incidence of reoperation, explants, or death. Performance Endpoints Immediate postoperative aortic regurgitation was evaluated after valve deployment and termination of extracorporeal circulation. The echocardiographic measurements included: peak and mean transvalvular pressure gradients were derived by using the modified Bernoulli s Fig 2. Valve design features dual collar design, with supraannular and intraannular sealing collar.

4 1486 FOLLIGUET ET AL Ann Thorac Surg SUTURELESS PERCEVAL AVR: RESULTS 2012;93: Table 2. Preoperative Patient Demographics Patient Characteristics (%) n 208 Age [years] Sex Male 67 Female 141 NYHA status III 187 pts (93) IV 21 pts (27) Previous cardiac surgery 6 Previous pace maker 15 EuroSCORE [logistic] Ejection fraction ( ) Diabetes % 58 (28) Hypertension % 142 (68) Chronic obstructive lung 32 (15) disease Body mass index EuroSCORE European system for cardiac operative risk evaluation; NYHA New York Heart Association; pts patients. equation, and the effective orifice area was calculated with the continuity equation. According to the echocardiographic protocol, the hemodynamic effectiveness endpoints recorded were mean and peak P, effective orifice area and indexed effective orifice area, cardiac output, cardiac index, ejection fraction, location, and severity degree of aortic and mitral regurgitations. Hemodynamic data were presented as number of patients, means, standard deviations, and ranges for each continuous parameter. Valve regurgitation was presented as numbers and percentages of patients with each grade of regurgitation. Postoperative echo data were analyzed for time-related variation of the degree of aortic regurgitation. The NYHA functional classification data were presented as the number and percentage of patients in each functional class. Patient improvement after implant of the Perceval S prosthesis was determined by comparison of preoperative and postoperative NYHA functional classifications, through application of the Wilcoxon range test. All statistical analysis was performed using the SAS System for Windows (version 9) (SAS Inc, Cary, NC). Results Patient and Procedure Characteristics Patient demographics and operative details are summarized in Tables 2 and 3. The indication for the surgery included degenerative native aortic valve disease in all patients. The total accumulated follow-up was 156 patient-years, as of the cutoff date for the present report. Three patients enrolled were excluded from the study after signed consent due to a larger annulus ( 25 mm) measured on the echocardiogram, and therefore a total of 208 patients received a Perceval sutureless bioprosthesis. There were no intraprocedural deaths. Forty-four patients received concomitant coronary revascularization. Five patients (2.4%) died during hospital stay. Autopsy, however, revealed no valve-related pathologies; the valve was seated appropriately inside the aortic annulus with patent coronary ostia. Twenty patients died during follow-up. Cumulative freedom from valve-related mortality (96% confidence interval) was 87.1% at 1 year, 82.4% at 2 years, 82% at 3 years, and 69.7% at 4 years (Fig 3). Mean follow-up was days, with a cumulative 156 patient-year. Implantation was successful in 95.6% of patients. In 9 patients immediate paravalvular leak (PVL) was noted during implantation; it was treated in 7 patients with a Perceval bioprosthesis and in 2 patients with a stented bioprosthesis. All these patients had a normal recovery. Perioperative morbidity was recorded for each patient during hospitalization and during follow-up. During follow-up 9 patients (4%) developed PVL, which required reoperations: early in 7 patients (from postoperative day 2 to postoperative day 13) and later in 2 patients; at 163 days and 576 days both developed intraprosthetic leak, 1 due to endocarditis and the other due to a pannus ingrowth on the stent, restricting a leaflet. In the serial echocardiograms performed regularly no other cases of restriction or dysfunction of the leaflets have been detected. Minor PVL not requiring surgical intervention was noted in 5 patients. No increase in the size of aortic insufficiency has been noted during follow-up. Bleeding events occurred in 13 patients (9 early, 4 late) requiring transfusions. Ten episodes of thromboembolism occurred during follow-up; 2 strokes, 1 transient ischemic attack, 3 limb embolism, 2 pulmonary embo- Table 3. Perceval S, Intraoperative Characteristics Intra/Postoperative Data Patients (n 208) Median Sternotomy (n 163) Mini Sternotomy (n 45) With Concomitant CABG (n 48) Isolated AVR Patients (n 160) Cross-clamp time [min] CPB-time [min] Valve size [mm] AVR aortic valve replacement; CABG coronary artery bypass grafting; CPB cardiopulmonary bypass; min minutes.

5 Ann Thorac Surg FOLLIGUET ET AL 2012;93: SUTURELESS PERCEVAL AVR: RESULTS 1487 symptoms at the 1-year and 2-year follow-up visits. These data represent a significant (p 0.001) improvement in the functional capacity of most patients undergoing aortic valve replacement with the Perceval S aortic bioprosthesis. Fig 3. Overall survival for Perceval S patients. lisms, and 1 retinal embolism. Pericardial effusion occurred in 4 patients requiring drainage, sepsis other than endocarditis requiring antibiotics occurred in 18 patients, and heart failure requiring inotropic drugs occurred in 5 patients. Endocarditis was diagnosed in 3 patients leading to 2 reoperations, while the third patient was managed medically and responded well to medical treatment. There was 1 case of hemolysis due to PVL. Pacemaker insertion for atrioventricular block was necessary in 16 patients (7%). Hemodynamic Parameters Transthoracic echocardiography was performed and hemodynamic parameters were obtained at the time of hospital discharge and at regularly scheduled outpatient visits at 3 to 6 months and 12 to 13 months. The average mean aortic gradient was measured at mm Hg. The mean effective orifice area was cm 2 and a mean indexed effective orifice area was cm 2 /m 2 (Table 4). No migration or dislodgement of the prosthesis has occurred during further follow-up. Effect of Aortic Valve Replacement on the NYHA Functional Capacity The NYHA functional capacity was evaluated and recorded preoperatively and at all subsequent visits. As shown in Table 2, all patients were categorized as NYHA class III or IV at the preimplantation assessment. By contrast, 82% of patients reported NYHA class I and II Comment Recent published series of conventional aortic valve replacement performed in octogenarians show an operative mortality between 4% and 10% [23 25]. Age itself is not a contraindication to conventional surgery but comorbidities such as low ejection fraction, renal failure, and calcified aorta are major risk factors [24]. In our study, the sutureless Perceval S aortic valve can be implanted after thorough surgical removal of the diseased valve, as done with conventional valve replacement. Mortality remains low, with minimal morbidity despite the increased risk of the population. Our early experience at 2 different institutions in Europe suggests an extremely secure valve positioning with very low gradient, including in a small annulus, and with excellent hemodynamics. This is in comparison with other reported sutureless valves which all have improved hemodynamics and low gradients [26]. Compared with stented bioprostheses this sutureless valve can be implanted with reduced cross-clamp time and extracorporeal time. This can be an advantage in patients requiring an added procedure; such as concomitant coronary artery bypass in older patients. In this case the sizing should be done first and the loading of the valve can be done while bypass is being performed. Once inserted there were no migrations of the valve and minimal complications were encountered during follow-up. Early nonfatal complications were documented in 4%, which is reflected by PVL. This can be a result of either inadequate sizing or due to inappropriate decalcification of the annulus. Due to its design this device has a round circumference that holds by its expandable structure nitinol; therefore, a circular rim of calcium should be left at the annulus level to assure proper anchoring without PVL. An important finding in our study is that when PVL is found intraoperatively or immediately postoperative, it should be immediately corrected. Either a new sutureless of usually different size can be placed, or a sutured bioprosthesis implanted. Older patients with a hypertrophic ventricle do not tolerate significant PVL, as rapidly they develop increased left ventricle diastolic pressure, Table 4. Hemodynamic Parameters During the Study Period, as Measured by Transthoracic Echocardiography Postoperative Variables Preoperative Discharge Post Implantation 3 6 months Post Implantation 1 4 years Mean gradient (mm Hg) Peak gradient (mm Hg) Effective orifice area (cm 2 ) Indexed effective orifice area (cm 2 /m 2 )

6 1488 FOLLIGUET ET AL Ann Thorac Surg SUTURELESS PERCEVAL AVR: RESULTS 2012;93: resulting in heart failure. Almost all patients with PVL were reoperated immediately or within a few days after the operation. Only 2 patients required late reoperation at 163 days and 576 days postoperative; 1 due to endocarditis and the other due to a fibrous pannus overgrowth. Another advantage of this bioprosthesis is its ease of insertion in minimal access surgery, as witnessed by a mean cross-clamp time of 33 minutes, which is identical to the sternotomy group. In small annuli (n 32, 23 mm), the gradients remained low as the mean gradient was mm Hg in this subgroup, which remains low compared with stented bioprostheses. In addition, no structural prosthetic deterioration, valve thrombosis, or significant transvalvular aortic regurgitation occurred during the study period. Cumulative freedom from valve-related mortality was 87% and 82% at 1 and 3 years, respectively. There were no cases of tilting or migration once appropriately inserted during the entire follow-up. Also, due to the unique design of the bioprosthesis, ostial blockage of the coronary arteries was never encountered. Further experience is needed to find the potential of the device for either an open or a minimally invasive approach. Furthermore, the sutureless bioprosthesis appears to be ideal for patients with severe calcification of the aortic root and patients requiring concomitant procedures in whom a reduced bypass time is mandatory. At this point, however, careful patient selection and echographic assessment are crucial in choosing the proper size. Correct sizing of the valve is critical to minimize paravalvular leakage and this should be performed with transesophageal echocardiography and intraoperative sizing. For an enlarged aorta with a ratio greater than 1.3, the predicted diameter according to body surface area represents a contraindication for this device. Also, long-term studies are needed to evaluate echographic results of this new prosthesis. References 1. Bonow RO, Carabello BA, Kanu C, et al. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation 2006;114:e Boeken U, Litmathe J, Kurt M, Feindt P, Gams E. Hemodynamic performance and clinical consequences of aortic valve replacement with 21-mm sized pericardial bioprostheses. Thorac Cardiovasc Surg 2005;53: Yankah CA, Pasic M, Musci M, et al. Aortic valve replacement with the Mitroflow pericardial bioprosthesis: durability results up to 21 years. J Thorac Cardiovasc Surg 2008;136: Dunning J, Graham RJ, Thambyrajah J, Stewart MJ, Kendall SW, Hunter S. Stentless vs. stented aortic valve bioprostheses: a prospective randomized controlled trial. Eur Heart J 2007;28: Sjogren J, Gudbjartsson T, Thulin LI. Long-term outcome of the MitroFlow pericardial bioprosthesis in the elderly after aortic valve replacement. J Heart Valve Dis 2006;15: Pomar JL, Jamieson WR, Pelletier LC, Castella M, Germann E, Brownlee RT. Mitroflow pericardial bioprosthesis experience in aortic valve replacement or 60 years of age. Ann Thorac Surg 1998;66(6 Suppl):S Dellgren G, David TE, Raanani E, Armstrong S, Ivanov J, Rakowski H. Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis. J Thorac Cardiovasc Surg 2002;124: Puvimanasinghe JP, Takkenberg JJ, Eijkemans MJ, et al. Prognosis after aortic valve replacement with the Carpentier-Edwards pericardial valve: use of microsimulation. Ann Thorac Surg 2005;80: Pavoni D, Badano LP, Musumeci SF, et al. Results of aortic valve replacement with a new supra-annular pericardial stented bioprosthesis. Ann Thorac Surg 2006;82: Puvimanasinghe JP, Takkenberg JJ, Eijkemans MJ, et al. Comparison of Carpentier-Edwards pericardial and supraannular bioprostheses in aortic valve replacement. Eur J Cardiothorac Surg 2006;29: David TE, Armstrong S, Maganti M. Hancock II bioprosthesis for aortic valve replacement: the gold standard of bioprosthetic valves durability? Ann Thorac Surg 2010;90: Yankah CA, Schubel J, Buz S, Siniawski H, Hetzer R. Seventeen-year clinical results of 1,037 Mitroflow pericardial heart valve prostheses in the aortic position. J Heart Valve Dis 2005;14: Repossini A, Kotelnikov I, Bouchikhi R, et al. Single-suture line placement of a pericardial stentless valve. J Thorac Cardiovasc Surg 2005;130: David TE, Feindel CM, Bos J, Sun Z, Scully HE, Rakowski H. Aortic valve replacement with a stentless porcine aortic valve. A six- year experience. J Thorac Cardiovasc Surg 1994;108: Beholz S, Dushe S, Konertz W. The Freedom SOLO valve: superior hemodynamic results with a new stentless pericardial valve for aortic valve replacement. J Heart Valve Dis 2007;16: Cohen G, Zagorski B, Christakis GT, et al. Are stentless valves hemodynamically superior to stented valves? Longterm follow-up of a randomized trial comparing Carpentier- Edwards pericardial valve with the Toronto Stentless Porcine Valve. J Thorac Cardiovasc Surg 2010;139: Kunadian B, Vijayalakshmi K, Thornley AR, et al. Metaanalysis of valve hemodynamics and left ventricular mass regression for stentless versus stented aortic valves. Ann Thorac Surg 2007;84: David TE, Feindel CM, Bos J, Ivanov J, Armstrong S. Aortic valve replacement with Toronto SPV bioprosthesis: optimal patient survival but suboptimal valve durability. J Thorac Cardiovasc Surg 2008;135: Borger MA, Prasongsukarn K, Armstrong S, Feindel CM, David TE. Stentless aortic valve reoperations: a surgical challenge. Ann Thorac Surg 2007;84: Kobayashi J. Stentless aortic valve replacement: an update. Vasc Health Risk Manag 2011;7: Shrestha M, Folliguet T, Meuris B, et al. Sutureless Perceval S aortic valve replacement: a multicenter, prospective pilot trial. J Heart Valve Dis 2009;18: Akins CW, Miller DC, Turina MI, et al. Guidelines for reporting mortality and morbidity after cardiac valve interventions. Ann Thorac Surg 2008;85: Urso S, Sadaba R, Greco E, et al. One-hundred aortic valve replacements in octogenarians: outcomes and risk factors for early mortality. J Heart Valve Dis 2007;16: Langanay T, Flecher E, Fouquet O, et al. Aortic valve replacement in the elderly: the real life. Ann Thorac Surg 2010;93: Di Eusanio M, Fortuna D, De Palma R, et al. Aortic valve replacement: results and predictors of mortality from a contemporary series of 2256 patients. J Thorac Cardiovasc Surg 2011; 141: Martens S, Ploss A, Sirat S, Miskovic A, Moritz A, Doss M. Sutureless aortic valve replacement with the 3f Enable aortic bioprosthesis. Ann Thorac Surg 2009;87:

Management of Difficult Aortic Root, Old and New solutions

Management of Difficult Aortic Root, Old and New solutions Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult

More information

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic,

Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, Role of Sutureless Valves in the Surgeon s Armamentarium Prof. Dr Malakh Shrestha Vice Chair, Director of Aortic Surgery Cardiothoracic, transplantation and Vascular Surgery Hannover Medical School, Germany

More information

The St. Jude Medical Biocor Bioprosthesis

The St. Jude Medical Biocor Bioprosthesis The St. Jude Medical Biocor Bioprosthesis Clinical Evidence of Long-term Durability Long-term Biocor Experience A Review and Comparative Assessment Long-term Biocor Stented Tissue Valve Studies Twenty-year

More information

16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900

16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 CLINICAL COMMUNIQUé 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 69 The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 69, was introduced into clinical

More information

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden Long-Term Outcome of the Mitroflow Pericardial Bioprosthesis in the Elderly after Aortic Valve Replacement Johan Sjögren, Tomas Gudbjartsson, Lars I. Thulin Department of Cardiothoracic Surgery, Heart

More information

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)

More information

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,

More information

TAVI Versus Suturless Valve In Intermediate Risk Patients

TAVI Versus Suturless Valve In Intermediate Risk Patients TAVI Versus Suturless Valve In Intermediate Risk Patients Walid Abukhudair FRCSc President of Saudi Society for Cardiac Surgeons Head of Cardiac Surgery in KFAFH Background AS is the most frequent cardiac

More information

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for

More information

CLINICAL COMMUNIQUE 16 YEAR RESULTS

CLINICAL COMMUNIQUE 16 YEAR RESULTS CLINICAL COMMUNIQUE 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 Introduction The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 6900, was introduced

More information

Experience with 500 Stentless Aortic Valve Replacements

Experience with 500 Stentless Aortic Valve Replacements Experience with 500 Stentless Aortic Valve Replacements Dimitrios C. Iliopoulos, MD Cardiac Surgeon Ass. Professor of Surgery University of Athens, School of Medicine I declare no conflict of interest

More information

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim 42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim Current Guideline for AR s/p TOF Surgery is reasonable in adults with prior repair of

More information

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis Acquired Cardiovascular Disease Mykén and Bech-Hansen A 2-year experience of 1712 patients with the Biocor porcine bioprosthesis Pia S. U. Mykén, MD, PhD, a and Odd Bech-Hansen, MD, PhD b Objective: The

More information

Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results

Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results Short Communication Bicuspid aortic root spared during ascending aorta surgery: an update of long-term results Marco Russo, Guglielmo Saitto, Paolo Nardi, Fabio Bertoldo, Carlo Bassano, Antonio Scafuri,

More information

The stentless bioprosthesis has many salient features that

The stentless bioprosthesis has many salient features that Aortic Valve Replacement with the Medtronic Freestyle Xenograft Using the Subcoronary Implantation Technique D. Michael Deeb, MD The stentless bioprosthesis has many salient features that make it an attractive

More information

Long-term results (22 years) of the Ross Operation a single institutional experience

Long-term results (22 years) of the Ross Operation a single institutional experience Long-term results (22 years) of the Ross Operation a single institutional experience Authors: Costa FDA, Schnorr GM, Veloso M,Calixto A, Colatusso D, Balbi EM, Torres R, Ferreira ADA, Colatusso C Department

More information

Reconstruction of the intervalvular fibrous body during aortic and

Reconstruction of the intervalvular fibrous body during aortic and Aortic and mitral valve replacement with reconstruction of the intervalvular fibrous body: An analysis of clinical outcomes Nilto C. De Oliveira, MD Tirone E. David, MD Susan Armstrong, MSc Joan Ivanov,

More information

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility

More information

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://www.annalsthoracicsurgery.org/cme/ home. To take the CME activity related to this article, you must have either an STS member

More information

PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic.

PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic. PROVEN PLUS. Introducing the Avalus Aortic Valve by Medtronic. With more than 40 years of heart valve innovations, we took proven valve design concepts and adapted them for excellent implantability for

More information

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve

More information

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Augusto D Onofrio, MD, Stefano Auriemma, MD, Paolo Magagna, MD, Alessandro Favaro, MD,

More information

CoreValve in a Degenerative Surgical Valve

CoreValve in a Degenerative Surgical Valve CoreValve in a Degenerative Surgical Valve Ran Kornowski, MD, FESC, FACC Chairman Department of Cardiology Rabin Medical Center, Petach Tikva, Israel Disclosure Statement of Financial Interest I, Ran Kornowski,

More information

Surgical AVR: Are there any contraindications? Pyowon Park Samsung Medical Center Seoul, Korea

Surgical AVR: Are there any contraindications? Pyowon Park Samsung Medical Center Seoul, Korea Surgical AVR: Are there any contraindications? Pyowon Park Samsung Medical Center Seoul, Korea Contents Decision making in surgical AVR in old age Clinical results of AVR with tissue valve Impact of 19mm

More information

The Ross Procedure: Outcomes at 20 Years

The Ross Procedure: Outcomes at 20 Years The Ross Procedure: Outcomes at 20 Years Tirone David Carolyn David Anna Woo Cedric Manlhiot University of Toronto Conflict of Interest None The Ross Procedure 1990 to 2004 212 patients: 66% 34% Mean age:

More information

Interventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504

Interventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504 Transcatheter valve-in-valve e implantation for aortic bioprosthetic valve dysfunction Interventional procedures guidance Published: 26 September 2014 nice.org.uk/guidance/ipg504 Your responsibility This

More information

Thrombocytopenia in Moderate- to High-Risk Sutureless Aortic Valve Replacement

Thrombocytopenia in Moderate- to High-Risk Sutureless Aortic Valve Replacement Korean J Thorac Cardiovasc Surg 2018;51:172-179 ISSN: 2233-601X (Print) ISSN: 2093-6516 (Online) CLINICAL RESEARCH https://doi.org/10.5090/kjtcs.2018.51.3.172 Thrombocytopenia in Moderate- to High-Risk

More information

MITROFLOW VALSALVA CONDUIT TM

MITROFLOW VALSALVA CONDUIT TM TM Features the only bovine pericardium aortic root surgery system THE REPLICATES THE NATIVE SINUS, NATURALLY 2 TM LIVANOVA S INNOVATIVE IS DESIGNED TO: Provide a reproducible and easy solution for better

More information

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,

More information

The operative mortality rate after redo valvular operations

The operative mortality rate after redo valvular operations Clinical Outcomes of Redo Valvular Operations: A 20-Year Experience Naoto Fukunaga, MD, Yukikatsu Okada, MD, Yasunobu Konishi, MD, Takashi Murashita, MD, Mitsuru Yuzaki, MD, Yu Shomura, MD, Hiroshi Fujiwara,

More information

TAVR for Valve-In-Valve. Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology

TAVR for Valve-In-Valve. Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology TAVR for Valve-In-Valve Brian O Neill Assistant Professor of Medicine Department of Medicine, Section of Cardiology Temple Hearth and Vascular Institute Disclosures: Consultant: Cardiac Assist TAVR for

More information

I will not discuss off label use or investigational use in my presentation.

I will not discuss off label use or investigational use in my presentation. I will not discuss off label use or investigational use in my presentation. Surgical valves Design and Durability Testing Potential Concerns Real Practice 1952-1962 1963-1966 1967-1969 1969-1977 1977-1984

More information

Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival

Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival Effect of Valve Suture Technique on Incidence of Paraprosthetic Regurgitation and 10-Year Survival Sukumaran K. Nair, FRCS (C Th), Gauraang Bhatnagar, MBBS, Oswaldo Valencia, MD, and Venkatachalam Chandrasekaran,

More information

TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair?

TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? Elaine E. Tseng, MD and Marlene Grenon, MD Department of Surgery Divisions of Adult Cardiothoracic and Vascular and Endovascular

More information

AVR with Sutureless Valves State of the Art

AVR with Sutureless Valves State of the Art AVR with Sutureless Valves State of the Art T. Fischlein Department of Cardiac Surgery, Cardiovascular Center Klinikum Nürnberg Paracelsus Medical University Nuremberg, Germany Disclosures Consultant and

More information

Anatomy determines the close vicinity of the sinuses of

Anatomy determines the close vicinity of the sinuses of Aortic Valve Reimplantation According to the David Type I Technique Matthias Karck, MD, and Axel Haverich, MD Department of Thoracic and Cardiovascular Surgery, Hannover Medical School, Hannover, Germany.

More information

Stentless aortic valves. Current aspects

Stentless aortic valves. Current aspects Endorsed by proceedings in Intensive Care Cardiovascular Anesthesia EXPERT OPINION HSR Proceedings in Intensive Care and Cardiovascular Anesthesia 2012; 4(2): 77-82 Stentless aortic valves. Current aspects

More information

Transcatheter aortic valve implantation for severe aortic valve stenosis with the ACURATE neo2 valve system: 30-day safety and performance outcomes

Transcatheter aortic valve implantation for severe aortic valve stenosis with the ACURATE neo2 valve system: 30-day safety and performance outcomes All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sale by or on the order of a physician. Indications, contraindications, warnings and instructions

More information

National Institute for Health and Care Excellence IP865/2 Sutureless Aortic Valve Replacement for aortic stenosis

National Institute for Health and Care Excellence IP865/2 Sutureless Aortic Valve Replacement for aortic stenosis National Institute for Health and Care Excellence IP865/2 Sutureless Aortic Valve Replacement for aortic stenosis IPAC 14/06/18: 1 1 NHS professional 2 1 NHS professional ments 1.1 1.1 Current evidence

More information

APOLLO TMVR Trial Update: Case Presentation

APOLLO TMVR Trial Update: Case Presentation APOLLO TMVR Trial Update: Case Presentation Anelechi Anyanwu, MD, MSc, FRCS-CTh Professor and Vice-Chairman Department of Cardiovascular Surgery Icahn School of Medicine at Mount Sinai New York, NY Disclosure

More information

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years Surgery for Acquired Cardiovascular Disease Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years W. R. Eric Jamieson, MD, Lawrence H. Burr, MD, Robert T. Miyagishima,

More information

Outcome of Next-Generation Transcatheter Valves in Small Aortic Annuli: A Multicenter Propensity-Matched Comparison

Outcome of Next-Generation Transcatheter Valves in Small Aortic Annuli: A Multicenter Propensity-Matched Comparison Outcome of Next-Generation Transcatheter Valves in Small Aortic Annuli: A Multicenter Propensity-Matched Comparison Mauri, V. et al.: Circ Cardiovasc Interv. 2017;10:e005013 All trademarks are the property

More information

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Speaker's name: I have the following potential conflicts of interest to report: Proctorship for Medtronic Agenda Failure modes

More information

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT Gideon Cohen, MD Tirone E. David, MD Joan Ivanov, MSc Sue Armstrong, MSc

More information

14 Valvular Stenosis

14 Valvular Stenosis 14 Valvular Stenosis 14-1. Valvular Stenosis unicuspid valve FIGUE 14-1. This photograph shows severe valvular stenosis as it occurs in a newborn. There is a unicuspid, horseshoe-shaped leaflet with a

More information

Kinsing Ko, Thom de Kroon, Najim Kaoui, Bart van Putte, Nabil Saouti. St. Antonius Hospital, Nieuwegein, The Netherlands

Kinsing Ko, Thom de Kroon, Najim Kaoui, Bart van Putte, Nabil Saouti. St. Antonius Hospital, Nieuwegein, The Netherlands Minimal Invasive Mitral Valve Surgery After Previous Sternotomy Without Aortic Clamping: Short- and Long Term Results of a Single Surgeon Single Institution Kinsing Ko, Thom de Kroon, Najim Kaoui, Bart

More information

CROWN PRT. Inservice implantation guide

CROWN PRT. Inservice implantation guide Inservice implantation guide DEVICE DESCRIPTION The Crown PRT aortic pericardial heart valve consists of a single piece of bovine pericardium that is preserved with glutaraldehyde and sewn onto a polyester

More information

ORIGINAL PAPER. The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan

ORIGINAL PAPER. The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan Nagoya J. Med. Sci. 78. 369 ~ 376, 2016 doi:10.18999/nagjms.78.4.369 ORIGINAL PAPER The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan

More information

REDUCING THE IMPACT OF SURGERY AND FACILITATING FASTER RECOVERY

REDUCING THE IMPACT OF SURGERY AND FACILITATING FASTER RECOVERY REDUCING THE IMPACT OF SURGERY AND FACILITATING FASTER RECOVERY AORTIC SOLUTIONS PERCEVAL VALVE PERCEVAL THE BIOLOGICAL VALVE OF CHOICE IN AVR 0914 RE0210195/A AORTIC SOLUTIONS PERCEVAL THE BIOLOGICAL

More information

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background

More information

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction

More information

Dr Nikolaos Baikoussis

Dr Nikolaos Baikoussis Dr Nikolaos Baikoussis Cardiac Surgeon Evangelismos General Hospital of Athens, Greece STS database: any procedure not performed with a full sternotomy (FS) and cardiopulmonary bypass (CPB)..(TAVI) Schmitto

More information

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles

More information

Sotirios N. Prapas, M.D., Ph.D, F.E.C.T.S.

Sotirios N. Prapas, M.D., Ph.D, F.E.C.T.S. CORONARY ARTERY REVASCULARIZATION WITH MILD AORTIC STENOSIS: STRATEGIES OF TREATMENT 9 th ANNUAL MEETING OF THE EAB SOCIETY, Pravets, Bulgaria, 2012 Sotirios N. Prapas, M.D., Ph.D, F.E.C.T.S. Director

More information

How to Perform a Valve Sparing Root Replacement Joseph S. Coselli, M.D.

How to Perform a Valve Sparing Root Replacement Joseph S. Coselli, M.D. How to Perform a Valve Sparing Root Replacement Joseph S. Coselli, M.D. AATS International Cardiovascular Symposium 2017 Session 6: Technical Aspects of Open Surgery on the Aortic Valve Sao Paulo, Brazil

More information

PPM: How to fit a big valve in a small heart

PPM: How to fit a big valve in a small heart PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat

More information

Portico (St. Jude Medical Inc, St.

Portico (St. Jude Medical Inc, St. Review Article Portico Transcatheter Heart Valve Apostolos Tzikas 1,2, Michael Chrissoheris 2, Antonios Halapas 2, Konstantinos Spargias 2 1 Interbalkan European Medical Centre, Thessaloniki, 2 Hygeia

More information

T sors in the following aspects: the porcine aortic valve

T sors in the following aspects: the porcine aortic valve Clinical and Hemodynamic Assessment of the Hancock I1 Bioprosthesis Tirone E. David, MD, Susan Armstrong, MSc, and Zhao Sun, MA Division of Cardiovascular Surgery, The Toronto Hospital and University of

More information

Incidence and Management of Early Implant Failure after Transcatheter Aortic Valve Implantation

Incidence and Management of Early Implant Failure after Transcatheter Aortic Valve Implantation ESC Congress 2010 28 Aug 2010-01 Sep 2010 Stockholm - Sweden Incidence and Management of Early Implant Failure after Transcatheter Aortic Valve Implantation Gian Paolo Ussia, MD Director of Interventional

More information

Indications and Late Results of Aortic Valve Repair

Indications and Late Results of Aortic Valve Repair Indications and Late Results of Aortic Valve Repair Prof. Gebrine El Khoury Department of Cardiovascular and Thoracic Surgery Cliniques St. Luc Brussels, Belgium Aortic Valve Repair Question # 1 Can the

More information

Disease of the aortic valve is frequently associated with

Disease of the aortic valve is frequently associated with Stentless Aortic Bioprosthesis for Disease of the Aortic Valve, Root and Ascending Aorta John R. Doty, MD, and Donald B. Doty, MD Disease of the aortic valve is frequently associated with morphologic abnormalities

More information

Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis

Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis Neal D. Kon, MD,* Robert D. Riley, MD, Sandy M. Adair, RN, Dalane W. Kitzman, MD, and A. Robert

More information

Aortic stenosis is the most common acquired heart valve. The Toronto Root Bioprosthesis: Midterm Results in 186 Patients

Aortic stenosis is the most common acquired heart valve. The Toronto Root Bioprosthesis: Midterm Results in 186 Patients The Toronto Root Bioprosthesis: Midterm Results in 186 Patients Sven Lehmann, MD, Thomas Walther, MD, PhD, Sergey Leontyev, MD, Jörg Kempfert, MD, Jens Garbade, MD, Michael A. Borger, MD, PhD, and Friedrich

More information

CIPG Transcatheter Aortic Valve Replacement- When Is Less, More?

CIPG Transcatheter Aortic Valve Replacement- When Is Less, More? CIPG 2013 Transcatheter Aortic Valve Replacement- When Is Less, More? James D. Rossen, M.D. Professor of Medicine and Neurosurgery Director, Cardiac Catheterization Laboratory and Interventional Cardiology

More information

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose Presenter Disclosure Patrick O. Myers, M.D. No Relationships to Disclose Aortic Valve Repair by Cusp Extension for Rheumatic Aortic Insufficiency in Children Long term Results and Impact of Extension Material

More information

Edwards Transcatheter AVR: Have the Outcomes Changed after CE Approval?

Edwards Transcatheter AVR: Have the Outcomes Changed after CE Approval? Edwards Transcatheter AVR: Have the Outcomes Changed after CE Approval? Update from PARTNER EU and SOURCE Registries T. Lefèvre Disclosure Statement Cardiologist Interventional cardiologist 1 st PABV in

More information

Less Invasive Aortic Valve Surgery

Less Invasive Aortic Valve Surgery Less Invasive Aortic Valve Surgery SCTS Brighton 19th March 2013 Andrew Chukwuemeka MD FRCS Consultant Cardiothoracic Surgeon Hammersmith Hospital Imperial College Healthcare NHS Trust Less Invasive Aortic

More information

Repair or Replacement

Repair or Replacement Surgical intervention post MitraClip Device: Repair or Replacement Saudi Heart Association, February 21-24 Rüdiger Lange, MD, PhD Nicolo Piazza, MD, FRCPC, FESC German Heart Center, Munich, Germany Division

More information

1-YEAR OUTCOMES FROM JOHN WEBB, MD

1-YEAR OUTCOMES FROM JOHN WEBB, MD 1-YEAR OUTCOMES FROM JOHN WEBB, MD ON BEHALF OF THE SAPIEN 3 INVESTIGATORS UNIVERSITY OF BRITISH COLUMBIA VANCOUVER, CANADA Potential conflicts of interest Speaker's name: John Webb I have the following

More information

Late failure of transcatheter heart valves: An open question

Late failure of transcatheter heart valves: An open question Late failure of transcatheter heart valves: An open question A comparison with surgically implanted bioprosthetic heart valves. A. Rashid The Cardiothoracic Centre Liverpool, UK. Conflict of Interest Statement

More information

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated?

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? RM Suri, V Sharma, JA Dearani, HM Burkhart, RC Daly, LD Joyce, HV Schaff Division of Cardiovascular Surgery, Mayo Clinic, Rochester,

More information

TAVR: Echo Measurements Pre, Post And Intra Procedure

TAVR: Echo Measurements Pre, Post And Intra Procedure 2017 ASE Florida, Orlando, FL October 10, 2017 8:00 8:25 AM 25 min TAVR: Echo Measurements Pre, Post And Intra Procedure Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology Echo Lab Associate

More information

Stainless Steel. Cobalt-chromium

Stainless Steel. Cobalt-chromium Sapien is better than Corevalve! Raj R. Makkar, MD Associate Director, Cedars-Sinai Heart Institute Associate Professor, UCLA School of Medicine, Los Angeles Eberhard Grube: Pioneer in the field of TAVR

More information

An anterior aortoventriculoplasty, known as the Konno-

An anterior aortoventriculoplasty, known as the Konno- The Konno-Rastan Procedure for Anterior Aortic Annular Enlargement Mark E. Roeser, MD An anterior aortoventriculoplasty, known as the Konno-Rastan procedure, is a useful tool for the cardiac surgeon. Originally,

More information

Early and One-year Outcomes of Aortic Root Surgery in Marfan Syndrome Patients

Early and One-year Outcomes of Aortic Root Surgery in Marfan Syndrome Patients Early and One-year Outcomes of Aortic Root Surgery in Marfan Syndrome Patients A Prospective, Multi-Center, Comparative Study Joseph S. Coselli, Irina V. Volguina, Scott A. LeMaire, Thoralf M. Sundt, Elizabeth

More information

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability

SOLO SMART. The smart way to return to life. Native-like performance now with stented-like implantability SOLO SMART TM The smart way to return to life Native-like performance now with stented-like implantability MANY PATIENTS NEED SUPERIOR HEMODYNAMIC PERFORMANCE TO RETURN TO THEIR NORMAL LIFESTYLE. 2 SOLO

More information

Tissue vs Mechanical What s the Data??

Tissue vs Mechanical What s the Data?? Biological (Tissue) Valve in a 60 year old patient: Debate Tissue vs Mechanical What s the Data?? Joseph E. Bavaria, MD Immediate-Past President - Society of Thoracic Surgeons (STS) Brooke Roberts-William

More information

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD

2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD 2017 Cardiovascular Symposium CARDIAC SURGERY UPDATE: SMALLER INCISIONS AND LESS COUMADIN DAVID L. SAINT, MD David L Saint M.D. Tallahassee Memorial Hospital Southern Medical Group Division of Cardiothoracic

More information

Clinical material and methods. Copyright by ICR Publishers 2003

Clinical material and methods. Copyright by ICR Publishers 2003 Fourteen Years Experience with the CarboMedics Valve in Young Adults with Aortic Valve Disease Jan Aagaard 1, Jens Tingleff 2, Per V. Andersen 1, Christel N. Hansen 2 1 Department of Cardio-Thoracic and

More information

TAVR SPRING 2017 The evolution of TAVR

TAVR SPRING 2017 The evolution of TAVR TAVR SPRING 2017 The evolution of TAVR Matthew Johnson, MD Disclosers None Evolution of the Balloon- Expandable Transcatheter Valves Cribier 2002 SAPIEN 2006 SAPIEN XT 2009 SAPIEN 3 2013 * Sheath compatibility

More information

Emergency Intraoperative Echocardiography

Emergency Intraoperative Echocardiography Emergency Intraoperative Echocardiography Justiaan Swanevelder Department of Anaesthesia, Glenfield Hospital University Hospitals of Leicester NHS Trust, UK Carl Gustav Jung (1875-1961) Your vision will

More information

Ian T. Meredith AM. MBBS, PhD, FRACP, FCSANZ, FACC, FAPSIC. Monash HEART, Monash Health & Monash University Melbourne, Australia

Ian T. Meredith AM. MBBS, PhD, FRACP, FCSANZ, FACC, FAPSIC. Monash HEART, Monash Health & Monash University Melbourne, Australia Two-Year Outcomes With the Fully Repositionable and Retrievable Lotus Transcatheter Aortic Replacement Valve in 120 High-Risk Surgical Patients With Severe Aortic Stenosis: Results From the REPRISE II

More information

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.

More information

Functional anatomy of the aortic root. ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη

Functional anatomy of the aortic root. ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη Functional anatomy of the aortic root ΔΡΟΣΟΣ ΓΕΩΡΓΙΟΣ Διεσθσνηής Καρδιοθωρακοτειροσργικής Κλινικής Γ.Ν. «Γ. Παπανικολάοσ» Θεζζαλονίκη What is the aortic root? represents the outflow tract from the LV provides

More information

Multicentre clinical study evaluating a novel resheatable self-expanding transcatheter aortic valve system

Multicentre clinical study evaluating a novel resheatable self-expanding transcatheter aortic valve system Multicentre clinical study evaluating a novel resheatable self-expanding transcatheter aortic valve system Preliminary Results: Acute and 1-year Outcomes Ganesh Manoharan, MBBCh, MD, FRCP Consultant Cardiologist

More information

Surgery: The Gold Standard treatment for Aortic Stenosis

Surgery: The Gold Standard treatment for Aortic Stenosis Surgery: The Gold Standard treatment for Aortic Stenosis Andrew Chukwuemeka MD FRCS Consultant Cardiothoracic Surgeon Hammersmith Hospital, Imperial College Healthcare NHS Trust Hon. Senior Lecturer, National

More information

Early Experience of Transcatheter Mitral Valve Replacement Results from the Intrepid Global Pilot Study

Early Experience of Transcatheter Mitral Valve Replacement Results from the Intrepid Global Pilot Study Early Experience of Transcatheter Mitral Valve Replacement Results from the Paul Sorajja, MD for the Investigators Presenter Disclosure Information Within the past 12 months, I or my spouse/partner have

More information

How to Avoid Prosthesis-Patient Mismatch

How to Avoid Prosthesis-Patient Mismatch How to Avoid Prosthesis-Patient Mismatch Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE, FESC Canada Research Chair in Valvular Heart Diseases INSTITUT UNIVERSITAIRE DE CARDIOLOGIE ET DE PNEUMOLOGIE DE QUÉBEC

More information

Acquired Cardiovascular Disease

Acquired Cardiovascular Disease One-year outcomes of the Surgical Treatment of Aortic Stenosis With a Next Generation Surgical Aortic Valve (TRITON) trial: A prospective multicenter study of rapid-deployment aortic valve replacement

More information

Case. 15-year-old boy with bicuspid AV Severe AR with moderate AS. Ross vs. AVR (or AVP)

Case. 15-year-old boy with bicuspid AV Severe AR with moderate AS. Ross vs. AVR (or AVP) Case 15-year-old boy with bicuspid AV Severe AR with moderate AS Ross vs. AVR (or AVP) AMC case 14-year-old boy with bicuspid AV Severe AS with mild AR Body size Bwt: 55 kg, Ht: 154 cm, BSA: 1.53 m 2 Echocardiography

More information

Important Safety Information

Important Safety Information Important Safety Information Edwards SAPIEN 3 Transcatheter Heart Valve with the Edwards Commander Delivery System Indications: The Edwards SAPIEN 3 transcatheter heart valve, model 9600TFX, and the Society

More information

Re-do aortic valve replacement after previous homograft aortic root replacement

Re-do aortic valve replacement after previous homograft aortic root replacement Re-do aortic valve replacement after previous homograft aortic root replacement Jullien Gaer, Toufan Bahrami, Fabio de Robertis, Ahmed Abdulsalam, John Pepper, NHS Foundation Trust, UK Professor Sir Magdi

More information

Indication, Timing, Assessment and Update on TAVI

Indication, Timing, Assessment and Update on TAVI Indication, Timing, Assessment and Update on TAVI Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Starr- Edwards Mechanical

More information

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT Tissue Valve for Aortic and Mitral Valve Replacement ARE MEDTRONIC SURGICAL TISSUE HEART VALVES RIGHT FOR YOU? Medtronic surgical heart valves are for

More information

2/28/2010. Speakers s name: Paul Chiam. I have the following potential conflicts of interest to report: NONE. Antegrade transvenous transseptal route

2/28/2010. Speakers s name: Paul Chiam. I have the following potential conflicts of interest to report: NONE. Antegrade transvenous transseptal route Transcatheter Aortic Valve Implantation Asian perspective Speakers s name: Paul Chiam Paul TL Chiam MBBS, MRCP, FACC I have the following potential conflicts of interest to report: NONE Consultant National

More information

Aortic valve repair: Techniques and Pitfalls. Allan Stewart, MD Columbia University Medical Center New York, NY

Aortic valve repair: Techniques and Pitfalls. Allan Stewart, MD Columbia University Medical Center New York, NY Aortic valve repair: Techniques and Pitfalls Allan Stewart, MD Columbia University Medical Center New York, NY Take Away Points 1. Valve anatomy is essential to assess repair 2. Unique Decisions with Aneurysm/AI

More information

Percutaneous Valve Interventions. Percutaneous Valve Interventions

Percutaneous Valve Interventions. Percutaneous Valve Interventions Percutaneous Valve Interventions Stanton J. Rowe President, Percutaneous Valve Interventions Edwards is Best Positioned to Capitalize on Percutaneous Valve Opportunities #1 global valve replacement and

More information

Decellularization of Aortic Homografts: South American and European Current Experience

Decellularization of Aortic Homografts: South American and European Current Experience Department of Cardiac Surgery Instituto de Neurologia e Cardiologia de Curitiba (INC-Cardio) Decellularization of Aortic Homografts: South American and European Current Experience Francisco Diniz Affonso

More information

Bioprostheses are prone to continuous degeneration

Bioprostheses are prone to continuous degeneration Twenty-Year Experience With the St. Jude Medical Biocor Bioprosthesis in the Aortic Position Walter B. Eichinger, MD, Ina M. Hettich, MD, Daniel J. Ruzicka, MD, Klaus Holper, MD, Carolin Schricker, Sabine

More information

Valve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal

Valve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal Valve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal I have nothing to disclose. Wide Spectrum Stable vs Decompensated NYHA II IV? Ejection

More information