1 110 International Journal of Cardiovascular Sciences. 2016;29(2): ORIGINAL MANUSCRIPT Prosthesis-Patient Mismatch in Individuals Undergoing Aortic Valve Replacement Ramsés Miotto, Marcos Venicio Garcia Joaquim Instituto de Cardiologia de Santa Catarina São José, SC Brazil Abstract Background: Prosthesis-patient mismatch (PPM) in patients undergoing aortic valve replacement surgery is the subject of many research studies and is associated with worse prognosis in the long term. Objectives: To evaluate the incidence of PPM after aortic valve replacement surgery and characterize the clinical profile of patients undergoing this surgery. Methods: Cross-sectional observational retrospective study of 53 patients hospitalized for surgical treatment of severe aortic valve stenosis from January 2014 to June Three models of bioprosthesis and a metal prosthesis model were used. Indexed effective orifice area (ieoa) was calculated by dividing the effective orifice area provided by the prosthesis manufacturer by the body surface area of the recipient patient. The presence of PPM was defined when ieoa <0.90 cm²/m². Results: The average age of patients was 65,8±9,9. Male sex prevailed. Overall mortality rate was 11.3%. PPM occurred in 32.0% of patients, mostly of which were moderate degree. The prosthesis Biocor showed 70.5% of PPM while the Braile prosthesis showed no case. There was no difference in the reduction of transvalvular aortic gradients in the postoperative period between the groups with and without PPM. Conclusions: The patients profile is similar to that described in the literature, the incidence of PPM is high, varying greatly between the prosthesis models. There was no difference in the reduction of postoperative gradients between the groups with or without PPM. Keywords: Aortic valve stenosis; Heart valve prosthesis implantation; Prosthesis fitting; Prosthesis failure Introduction Aortic stenosis is the most common acquired valvular heart disease and its relevance, in terms of public health, rises progressively with population aging, considering that 2-4% of people older than 70 are affected by the disease. 1-5 The valve replacement surgery is the treatment of choice for aortic stenosis, considering the significant change of the natural evolution as clearly demonstrated by the studies of Schwarz et al. 6 and Pellikka et al. 7 Since the definition of prosthesis-patient mismatch (PPM) by Rahimtoola in 1978, this topic has been the subject of many research studies due to its influence on the prognosis of patients. 8 Prosthesis-patient mismatch is considered when the relationship between the effective orifice area of the valve prosthesis and the body surface area of the patient is reduced compared to the healthy native valve. Particularly, patients undergoing valve replacement surgery through aortic stenosis have a higher incidence of PPM compared to patients with aortic regurgitation, which can be explained by the degeneration and calcification of the aortic annulus and consequent narrowing present in aortic stenosis; aortic regurgitation is usually associated with the enlargement of the aortic annulus. Corresponding author: Ramsés Miotto Rua Adolfo Donato da Silva, s/n Praia Comprida São José, SC Brazil ramsesmiotto@gmail.com DOI: / Manuscript received on March 1, 2016; approved on April 12, 2016; revised on April 28, 2016.
2 Int J Cardiovasc Sci. 2016;29(2): ABBREVIATIONS AND ACRONYMS ECC endogenous creatinine clearance ICSC Instituto de Cardiologia de Santa Catarina IEOA indexed effective orifice area MTG mean transvalvular gradient PG peak gradient PPM prosthesis-patient mismatch PPM in classified into three levels of severity: mild, when the indexed effective orifice area is <0.90 cm²/m² and 0,85 cm²/m²; moderate, when <0.85 cm² and 0,65 cm²/m²; and severe, when <0.65 cm²/m² Prosthesis-patient mismatch is associated with less regression of symptoms and left ventricular hypertrophy, maintenance of elevated transvalvular gradients, decreased durability of prostheses and increased mortality in the long term. 10,12-16 The prostheses used and evaluated in this study were: the bioprosthesis Carpentier-Edwards Perimount model 2900 manufactured by Edwards Lifesciences (Irvine, California, USA); the bioprosthesis Biocor model B30 and the mechanical prosthesis model AJ-501 both manufactured by St Jude Medical (Saint Paul, Minnesota, USA) and the Braile bioprosthesis manufactured by Braile Biomédica (São José do Rio Preto, São Paulo, Brazil). The effective orifice area values taken from the manufacturers technical specifications obtained from in vitro tests conducted by the manufacturers (Chart 1) were used. Chart 1 Effective orifice area in cm² of valve prostheses This study aims to assess the incidence of PPM after aortic valve replacement surgery and characterize the clinical profile of patients undergoing this surgery. Size Prosthesis Carpentier (cm²) Methods Cross-sectional observational retrospective study of 53 patients hospitalized at Instituto de Cardiologia de Santa Catarina (ICSC) for surgical treatment of severe aortic valve stenosis from January 2014 to June This study has been approved by the Research Ethics Committee of the institution under no. CAAE Because this is a retrospective study, Informed Consent Form was not required. The institution s Statistical Services database was used to find all patients undergoing aortic valve replacement in the period cited and 93 individuals were identified. The study included patients older than 18, with severe stenosis of the native aortic valve who underwent aortic valve replacement (with or without coronary artery bypass grafting associated). Patients with other moderate or severe valve disease associated, with moderate or severe double valve lesion, patients with prosthetic valve and those undergoing valve re-replacement were excluded. At the end, there was a sample of 53 patients whose data were collected through the electronic medical records system Micromed and the database of the institution s Echocardiography Service. St Jude Biocor (cm²) NA NA St Jude Standard (cm²) NA Braile (cm²) NA = non-available For data collection, performed by one of the authors, a handwritten sheet establishing criteria and routine to evaluate the medical records was used. Echocardiography tests were performed during hospitalization for cardiologists specializing in echocardiography using the device Vivid E9 with the M5S transducer manufactured by GE (Waukesha, Wisconsin, USA). Ejection fraction was obtained by the biplane method; aortic valve area was obtained by the continuity equation method and the peak and mean gradients were obtained by continuous-wave left parasternal, apical or supraclavicular window Doppler. Body surface area was calculated by the Du Bois 17 formula and the indexed effective orifice area was obtained by dividing the effective orifice area of the prosthesis by the body surface area retrospectively from the anthropometric measurements stated on the patient records. Such measure did not influence the surgeons prosthesis choices.
3 112 Int J Cardiovasc Sci. 2016;29(2): For the endogenous creatinine clearance (ECC), we used the CKD-EPI 18 formula, which considers the serum creatinine, age, sex and ethnicity. Patients using antihypertensive medication were considered hypertensive; diabetics were considered to be those using oral hypoglycemic medication and/or insulin; dyslipidemia was considered when there was documentation through preoperative laboratory tests and/or use of lipid-lowering medication. The presence of coronary artery disease was determined in case of prior coronary event or previous coronary procedure or coronary angiography, demonstrating stenosis greater than 50%. To define chronic obstructive pulmonary disease, spirometry results were used when available and/or use of bronchodilators and inhaled corticosteroids. Peripheral occlusive vascular disease was defined when there was history of ischemic vascular event, bypass surgery, previous angioplasty or amputation. To estimate operative risk, EuroSCORE II was calculated retrospectively from patient records data through the electronic calculator available at For the electrocardiographic diagnosis of left ventricular hypertrophy, at least one of the following criteria was considered: Sokolow-Lyon, Sokolow-Lyon- Rappaport; Lewis; Cornell, Gubner-Ungerleider and Romhilt-Estes. 19,20 Data for categorical variables were described by absolute count and percentage and analyzed using Fisher s exact test or chi-square test. Data related to continuous variables were expressed as mean and standard deviation. Intra- and intergroup comparison of continuous variables was performed using paired and/or unpaired t-test; p<0.05 was considered significant. Data were analyzed using Microsoft Excel. Results The general characteristics of the group are shown in Table 1. The sample was predominantly male and mean age of 65.8±9.9 years. Most had clinical manifestations of dyspnea (NYHA functional class II and III) and angina pectoris. The predominant subtype of aortic stenosis in these patients was degenerative involvement with calcification. Table 1 General characteristics of the study sample (n=53) Variables Values Age in years (mean±sd) 65.8 ± 9.9 Male gender (%) 62.3 Comorbidities (%) Hypertension 56.6 Diabetes 30.2 Dyslipidemia 39.6 CAD 41.5 PVOD 5.7 COPD 11.3 ECC < 60 ml/min 26.4 Clinical conditions (%) Dyspnea 83.0 Chest pain 60.4 Syncope 14.0 LVH on ECG 82.0 Functional Class (NYHA) (%) Class I 9.6 Class II 30.8 Class III 44.2 Class ignored 15.4 Anthropometry (mean±sd) Body surface area (m²) 1.8 ± 0.2 Weight (kg) 71.0 ± 14.0 Height (cm) ± 32.1 Valve disease etiology (%) Three-valve calcified 75.4 Bicuspid valve 20.8 Rheumatic disease 3.8 EuroSCORE II rating (mean±sd) 2.1 ± 1.6 Isolated valve replacement (n, %) 38/ Valve replacement combined with MR (n,%) 15/ Minutes under CPB (mean±sd) 98.8 ± 40.9 Minutes in aortic clamping (mean±sd) 75 ± 28.9 CAD coronary artery disease; PVOD peripheral occlusive vascular disease; COPD chronic obstructive pulmonary disease; ECC endogenous creatinine clearance; LVH left ventricular hypertrophy; ECG electrocardiogram; NYHA New York Heart Association; CABG coronary artery bypass grafting; CPB cardiopulmonary bypass; SD standard deviation
4 Int J Cardiovasc Sci. 2016;29(2): On average, patients had left ventricular hypertrophy, preserved ejection fraction and high transvalvular gradients, as it can be seen on the preoperative echocardiography on Table 2. Table 2 Preoperative echocardiographic characteristics of the study sample Aortic valve area (cm²) 0.7 ± 0.2 Ejection fraction (%) 57.0 ± 18.9 IVS thickness (mm) 14.3 ± 2.7 LVPW thickness (mm) 13.5 ± 2.4 Peak transvalvular gradient (mmhg) 86.7 ± 26.0 Mean transvalvular gradient (mmhg) 56.1 ± 19.2 Values expressed as mean±standard deviation. IVS interventricular septum; LVPW left ventricular posterior wall Chart 1 specifies the effective orifice areas of the prosthesis according to the manufacturer and the nominal size, obtained from material published by the industry (values obtained by in vitro tests). Table 3 presents hospitalization time and hospital mortality data. The hospitalized patients studied waited 28.4±12.6 days on average to undergo surgery, amounting to an average hospital stay of 45.1±20.2 days. In-hospital mortality was 20.0% for those who underwent combined surgery and 7.9% for those who underwent isolated valve replacement. Table 3 Hospitalization of the study sample Pre-op hospitalization days (mean±sd) 28.4 ± 12.6 Hospitalization days in the ICU postoperatively (mean±sd) 8.1 ± 15.6 Post-op days in the ward (mean±sd) 16.7 ± 17.1 Total hospital stay (mean±sd) 45.1 ± 20.2 Overall hospital mortality (n,%) 6/ Hospital mortality in combined surgery (n,%) 3/ Hospital mortality in isolated valve replacement (n,%) 3/ Pre-op preoperative; Post-op post-operative; SD standard deviation Table 4 presents postoperative echocardiography data of the patients evaluated: none of them presented prosthesis dysfunction or severe paravalvular failure; there was mild paravalvular failure in 15.6% and moderate in 2.2% of patients. Peak and mean transvalvar aortic gradients were 31.5±11.3 mmhg and 17.2±7.5 mmhg, respectively. Table 4 Postoperative echocardiographic characteristics of the study sample Prosthesis dysfunction (%) 0.0 Mild paravalvular failure (%) 15.6 Moderate paravalvular failure (%) 2.2 Severe paravalvular failure (%) 0.0 Peak transvalvular gradient (mmhg) (mean±sd) Mean transvalvar gradient (mmhg) (mean±sd) 31.5 ± ± 7.5 There was no statistically significant difference in the reduction of the postoperative transvalvular gradient among the patients who did or did not present PPM (Table 5). A total of 17 (32.0%) patients presented PPM, of which 3 (5.6%) presented mild PPM, 13 (24.5%) presented moderate PPM and 1 (1.8%), severe PPM. Analyzing by prosthesis, 10 (18.8%) patients had Braile bioprosthesis implanted; 17 (32.0%) received the prosthesis St Jude Biocor, 8 (15.0%) received Carpentier and 18 (34.0%) patients received the metal prosthesis St. Jude Standard. All 53 patients had a statistically significant reduction in peak (PG) and mean (MG) transvalvular gradient compared to the preoperative value (Figure 1). No patient receiving the Braile prosthesis presented PPM. The following presented PPM: for the Biocor prosthesis 12 (70.5%) of the 17 patients; for the Carpentier prosthesis, 3 (37.5%) of the 8 patients; and 2 (11.1%) of the 18 patients who received the metal St. Jude Standard prosthesis. Between Biocor and Carpentier, there was no statistically significant difference in the incidence of PPM. There was a significant difference between Biocor and the metal St. Jude prosthesis (p=0.005) and between Biocor and Braille (p=0.0001).
5 114 Int J Cardiovasc Sci. 2016;29(2): Table 5 Comparison of reduction of gradients between the groups studied Patients with PPM (n=17) Patients without PPM (n=36) p value Pre-op peak grad 89.4 ± ± 26.7 ns Pre-op mean grad 57.6 ± ± 19.5 ns Post-op peak grad 34.3 ± ± 11.7 ns Post-op mean grad 17.1 ± ± 8.1 ns Pre- and post-operative mean peak gradients expressed in mmhg Grad gradient; Pre-op preoperative; Mean grad mean gradient; Post-op post-operative; ns non-significant; PPM prosthesis-patient mismatch Figure 1 Distribution of cases of PPM and pre- and postoperative gradients by prosthesis. PPM prosthesis-patient mismatch; PG peak gradient; MG mean gradient; Pre-op preoperative period; post-op postoperative period. Discussion The profile of the patients in this study is very similar to that found by other authors, 11,14,21-26 but the study sample showed higher percentage of patients with diabetes and chronic kidney disease. The larger number of patients with renal failure may be related to different definition criteria adopted: while in most studies the creatinine value was used, in this study, the endogenous creatinine clearance was used. All patients had a significant reduction of transvalvular gradients in relation to the preoperative period, however, there was no difference between patients with or without PPM. Mean postoperative PG values around 31.5±11.3 mmhg were found. Flameng et al. 14 found a significant difference in postoperative PG among patients with or without PPM, with mean values of 29.0±11.2 mmhg and 22.9±9.3 mmhg, respectively. The study of Hanayama et al. 25 found an average postoperative PG of 23.9±9.8 mmhg and an average MG of 13.6±6.5 mmhg, and showed a significant difference with a lower postoperative MG among patients without PPM. 25 These contradictory results, found in this study, are probably due to the small sample size and the absence
6 Int J Cardiovasc Sci. 2016;29(2): of data on the medical records for postoperative echocardiography. The general hospital mortality found in this study was 11.3%; however, for patients undergoing isolated aortic valve replacement, it was 7.9% lower than in another study conducted in Brazil in 2011, with patients undergoing isolated aortic valve replacement 21 and similar to the overall mortality of 7.2% obtained by Dayan et al. 22 Howell et al. 23 published a study in which they obtained mortality of 3.8% in a sample of 944 patients undergoing isolated aortic valve replacement, 23 which may be even lower according to other authors. 24,25,27 Several studies have shown that aortic valve replacement surgery associated with coronary artery bypass grafting increases mortality. In this study, 20% hospital mortality was found for combined surgery, which can be considered high, compared to the findings in the literature, ranging from 5.0 to 8.2%. 23,27,28 There is conflicting information about the best ieoa calculation methodology. Pibarot et al. 9 affirmed that the measurements from in vitro tests are similar to in vivo findings, suggesting that, for the evaluation of metal prostheses, in vitro measurements should be used, while others suggest that in vitro values are overestimated and which should be based on in vivo tests from echocardiography. 29,30 For the Braile prosthesis, no reference on in vivo hemodynamic performance was found, and in vitro data were used for all prostheses. Interestingly, none of the patients who received the Braile prosthesis presented PPM and the postoperative gradients tended to be smaller, but without statistical significance. Even considering that the EOA supplied by the manufacturers are overestimated, still, for the Biocor, the metal prosthesis St. Jude and Carpentier-Edwards Perimount 2900, there is a high prevalence of PPM, with worse performance found in the Biocor prosthesis. Similarly, this PPM proportion may be underestimated for the same reason. In this study, 32.0% patients with PPM were identified. Only one (1.88%) patient had severe PPM. Kaminishi et al. 24 obtained a total of 8.5% of patients with PPM, however, in their study, they used only mechanical prostheses, stentless models and Carpentier, known to have better hemodynamic profile. 24 Howell et al. 23 concluded that 8.6% of patients had severe PPM, which is much higher than our findings. Both authors used vitro data for the calculations. Other authors have used in vivo measurements to calculate the ieoa, including Blais et al. 11 who found 36.0% of moderate PPM and 2.0% severe PPM. Bleiziffer et al. 30 found 33.4% of moderate PPM and 6.3% of severe PPM. Flameng et al. 14 identified 46.0% of moderate PPM and 4.0% of severe PPM. Yap et al. 31 described 6.6% of severe PPM and Dayan et al. 22 reported 62.8% of moderate PPM and 1.8% of severe PPM. There is a difference in the incidence of PPM in studies with in vivo data compared to those with in vitro data for the definition of PPM, which reinforces the argument that in vitro test values are overestimated. Although the study sample is based on in vitro data, it was very similar to studies using in vivo data, which showed that the problem of PPM is very common, although severe cases are a minority. The problem of PPM is very common in the literature, as well as in the ICSC, where the prevalence is similar to that of major studies. This clinical situation has important prognostic implications and there should be efforts to minimize it, using surgeries to increase the aortic annulus and implant prostheses with a better hemodynamic profile. Some of the medical records searched included no data on the control echocardiography conducted before hospital discharge or during follow-up, which may have affected the calculation of postoperative gradients. However, this did not invalidate the research. The ieoa were calculated based on the technical specifications of the manufacturer of each prosthesis and no patient was submitted to control echocardiography to calculate ieoa using the continuity equation. In anyway, one should try to avoid the PPM by calculating the preoperative prosthesis size based on the specifications disclosed by the suppliers, although they can be obtained using non-standard methodologies and suffer commercial industry bias. 32 Conclusions The general characteristics of patients undergoing surgery are similar to those of other published studies. The incidence of PPM is high, but there are sharp
7 116 Int J Cardiovasc Sci. 2016;29(2): differences between the prosthesis models. Regarding the gradients, there was no significant difference between patients with or without PPM. Potential Conflicts of Interest This study has no relevant conflicts of interest. Sources of Funding This study had no external funding sources. Academic Association This manuscript is part of the Final Term Paper of Ramsés Miotto for the Medical Residency Program in Cardiology at Instituto de Cardiologia de Santa Catarina. References 1. Carabello BA, Paulus WJ. Aortic stenosis. Lancet. 2009;373(9667): Tarasoutchi F, Montera MW, Grinberg M, Barbosa MR, Piñeiro DJ, Sánchez CRM, et al. Diretriz brasileira de valvopatias - SBC 2011 / I Diretriz Interamericana de valvopatias - SIAC Arq Bras Cardiol. 2011;97(5 supl. 1): Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Guyton RA, et al; American College of Cardiology/ American Heart Association Task Force on Practice Guidelines AHA/ACC Guideline for the management of patients with valvular heart disease: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2014;63(22): Erratum in: J Am Coll Cardiol. 2014;63(22): Supino PG, Borer JS, Preibisz J, Bornstein A. The epidemiology of valvular heart disease: a growing public health problem. Heart Fail Clin. 2006;2(4): Freeman RV, Otto CM. Spectrum of calcific aortic valve disease: pathogenesis, disease progression, and treatment strategies. Circulation. 2005;111(24): Schwarz F, Baumann P, Manthey J, Hoffmann M, Schuler G, Mehmel HC, et al. The effect of aortic valve replacement on survival. Circulation. 1982;66(5): Pellikka PA, Nishimura RA, Bailey KR, Tajik AJ. The natural history of adults with asymptomatic, hemodynamically significant aortic stenosis. J Am Coll Cardiol. 1990;15(5): Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation. 1978;58(1): Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36(4): Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92(8): Blais C, Dumesnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108(8): Head SJ, Mokhles MM, Osnabrugge RL, Pibarot P, Mack MJ, Takkenberg JJ, et al. The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising patients with patient-years. Eur Heart J. 2012;33(12): Dumesnil JG, Pibarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13(3): Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121(19): Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg. 1992;6(Suppl 1):S Kandler K, Møller CH, Hassager C, Olsen PS, Lilleør N, Steinbrüchel DA. Patient-prosthesis mismatch and reduction in left ventricular mass after aortic valve replacement. Ann Thorac Surg. 2013;96(1): Du Bois D, Du Bois EF. A formula to estimate the approximate surface area if height and weight be known Nutrition. 1989;5(5): Levey AS, Stevens LA, Schmid CH, Zhang YL, Castro AF 3rd, Feldman HI, et al; CKD-EPI (Chronic Kidney Disease Epidemiology Collaboration). A new equation to estimate glomerular filtration rate. Ann Intern Med. 2009;150(9): Erratum in: Ann Intern Med. 2011;155(6): Povoa R, Souza D. Análise crítica do eletrocardiograma e do ecocardiograma na detecção da hipertrofia ventricular esquerda. Rev Bras Hipertens. 2008;15(2): Matos DIA. Acuidade do eletrocardiograma no diagnóstico de hipertrofia ventricular esquerda. Rev Bras Cardiol. 2010;23(6): Almeida AS, Picon PD, Wender OC. Resultados de pacientes submetidos à cirurgia de substituição valvar aórtica usando próteses mecânicas ou biológicas. Rev Bras Cir Cardiovasc. 2011;26(3): Dayan V, Soca G, Stanham R, Lorenzo A, Ferreiro A. Is patientprosthesis mismatch a predictor of survival or a surrogate marker of co-morbidities in cardiac surgery? Int J Cardiol. 2015;190: Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30(1): Kaminishi Y, Misawa Y, Kobayashi J, Konishi H, Miyata H, Motomura N, et al; Japan Cardiovascular Surgery Database Organization. Patient-prosthesis mismatch in patients with aortic valve replacement. Gen Thorac Cardiovasc Surg. 2013;61(5): Hanayama N, Christakis GT, Mallidi HR, Joyner CD, Fremes SE, Morgan CD, et al. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thorac Surg. 2002;73(6):
8 Int J Cardiovasc Sci. 2016;29(2): Kohsaka S, Mohan S, Virani S, Lee VV, Contreras A, Reul GJ, et al. Prosthesis patient mismatch affects long-term survival after mechanical valve replacement. J Thorac Cardiovasc Surg. 2008;135(5): Emery RW, Krogh CC, Arom KV, Emery AM, Benyo-Albrecht K, Joyce LD, et al. The St. Jude Medical cardiac valve prosthesis: a 25-year experience with single valve replacement. Ann Thorac Surg. 2005;79(3): Rao V, Jamieson WR, Ivanov J, Armstrong S, David TE. Prosthesis-patient mismatch affects survival after aortic valve replacement. Circulation. 2000;102(19 Suppl 3):III Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of indexed aortic prosthetic valve areas calculated by Doppler echocardiography. J Am Coll Cardiol. 1990;16(3): Bleiziffer S, Eichinger WB, Hettich I, Guenzinger R, Ruzicka D, Bauernschmitt R, et al. Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method? Heart. 2007;93(5): Yap CH, Mohajeri M, Yii M. Prosthesis-patient mismatch is associated with higher operative mortality following aortic valve replacement. Heart Lung Circ. 2007;16(4): Cohen RG, Bourne ET. Industry-generated charts for the selection of stented aortic valve prostheses: clinical tool or marketing ploy? Ann Thorac Surg. 2011;91(4):
Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?
Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles
More informationValve prosthesis-patient mismatch (PPM) was first defined
Impact of Valve Prosthesis-Patient Mismatch on Short-Term Mortality After Aortic Valve Replacement Claudia Blais, BSc; Jean G. Dumesnil, MD; Richard Baillot, MD; Serge Simard, MS; Daniel Doyle, MD; Philippe
More informationManagement of Difficult Aortic Root, Old and New solutions
Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult
More informationPatient/prosthesis mismatch: how to evaluate and when to act?
Patient/prosthesis mismatch: how to evaluate and when to act? Svend Aakhus, MD, PhD Oslo University Hospital, Norway Disclosures: No conflict of interest Types of aortic valve prostheses (AVR) Mechanical
More informationQUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH
QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH 1,2 Radu A. SASCĂU 3 Cristina OLARIU 1,2 Cristian STĂTESCU 1 Internal Medicine Department, Gr.T.Popa University of Medicine and Pharmacy,
More informationValve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal
Valve Disease in Patients With Heart Failure TAVI or Surgery? Miguel Sousa Uva Hospital Cruz Vermelha Lisbon, Portugal I have nothing to disclose. Wide Spectrum Stable vs Decompensated NYHA II IV? Ejection
More informationPatient prosthesis mismatch after mitral valve replacement: Myth or reality?
Patient prosthesis mismatch after mitral valve replacement: Myth or reality? Pasquale Totaro, MD, a and Vincenzo Argano, MD b Objective: Determining the risk of patient prosthesis mismatch after mitral
More informationClinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis
CLINICS 2012;67(1):55-60 DOI:10.6061/clinics/2012(01)09 CLINICAL SCIENCE Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis Luis M. Astudillo, I Orlando
More informationAortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve implantation
Gen Thorac Cardiovasc Surg (2016) 64:435 440 DOI 10.1007/s11748-016-0657-9 CURRENT TOPICS REVIEW ARTICLE Aortic valve replacement and prosthesis-patient mismatch in the era of trans-catheter aortic valve
More informationImpact of Prosthesis-Patient Mismatch on Long-Term Survival After Aortic Valve Replacement
Journal of the American College of Cardiology Vol. 53, No. 1, 9 9 by the American College of Cardiology Foundation ISSN 735-197/9/$36. Published by Elsevier Inc. doi:1.116/j.jacc.8.9.22 Valvular Heart
More informationHow to Avoid Prosthesis-Patient Mismatch
How to Avoid Prosthesis-Patient Mismatch Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE, FESC Canada Research Chair in Valvular Heart Diseases INSTITUT UNIVERSITAIRE DE CARDIOLOGIE ET DE PNEUMOLOGIE DE QUÉBEC
More informationExperience with 500 Stentless Aortic Valve Replacements
Experience with 500 Stentless Aortic Valve Replacements Dimitrios C. Iliopoulos, MD Cardiac Surgeon Ass. Professor of Surgery University of Athens, School of Medicine I declare no conflict of interest
More informationTSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD
TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)
More informationOutcome of Next-Generation Transcatheter Valves in Small Aortic Annuli: A Multicenter Propensity-Matched Comparison
Outcome of Next-Generation Transcatheter Valves in Small Aortic Annuli: A Multicenter Propensity-Matched Comparison Mauri, V. et al.: Circ Cardiovasc Interv. 2017;10:e005013 All trademarks are the property
More informationIncidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves
INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient
More informationPPM: How to fit a big valve in a small heart
PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat
More informationProsthetic valve dysfunction: stenosis or regurgitation
Prosthetic valve dysfunction: stenosis or regurgitation Jean G. Dumesnil MD, FRCP(C), FACC, FASE(Hon) Quebec Heart and Lung Institute, Québec, Québec No disclosures Possible Causes of High Gradients in
More informationLate haemodynamic performance and survival after aortic valve replacement with the Mosaic bioprosthesis
Interactive CardioVascular and Thoracic Surgery 19 (2014) 756 762 doi:10.1093/icvts/ivu238 Advance Access publication 12 July 2014 ORIGINAL ARTICLE ADULTCARDIAC Late haemodynamic performance and survival
More informationThe impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation
Editor s choice Scan to access more free content 1 Division of Cardiac Surgery, The Johns Hopkins Hospital, Baltimore, Maryland, USA 2 Division of Cardiac Surgery, University of Ottawa Heart Institute,
More informationProf. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM
The Patient with Aortic Stenosis and Mitral Regurgitation Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM Aortic Stenosis + Mitral Regurgitation?
More informationHani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz
Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for
More informationCopyright by ICR Publishers 2014
Comprehensive Hemodynamic Performance and Frequency of Patient-Prosthesis Mismatch of the St. Jude Medical Trifecta Bioprosthetic Aortic Valve Ajay Yadlapati 1, Jimmy Diep 3, Mary-Jo Barnes 2, Tristan
More informationHemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized study
European Journal of Cardio-thoracic Surgery 39 (2011) 844 852 www.elsevier.com/locate/ejcts Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of
More informationA 20-year experience of 1712 patients with the Biocor porcine bioprosthesis
Acquired Cardiovascular Disease Mykén and Bech-Hansen A 2-year experience of 1712 patients with the Biocor porcine bioprosthesis Pia S. U. Mykén, MD, PhD, a and Odd Bech-Hansen, MD, PhD b Objective: The
More informationImpact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement
Koene et al. Journal of Cardiothoracic Surgery 2013, 8:96 RESEARCH ARTICLE Open Access Impact of prosthesis-patient mismatch on early and late mortality after aortic valve replacement Bart M Koene 2*,
More informationCARDIACSURGERY TODAY. Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease
VOLUME 1 NUMBER 2 23 CARDIACSURGERY TODAY Commentary and Analysis on Advances in the Surgical Treatment of Cardiac Disease EDITORS-IN-CHIEF Robert W Emery, St Paul, MN, USA Francesco Musumeci, Rome, Italy
More informationAortic Valve Replacement or Heart Transplantation in Patients With Aortic Stenosis and Severe Left Ventricular Dysfunction
Aortic Valve Replacement or Heart Transplantation in Patients With Aortic Stenosis and Severe Left Ventricular Dysfunction L.S.C. Czer, S. Goland, H.J. Soukiasian, S. Gallagher, M.A. De Robertis, J. Mirocha,
More informationImpact of Valve Prosthesis-Patient Mismatch on Left Ventricular Mass Regression Following Aortic Valve Replacement
Impact of Valve Prosthesis-Patient Mismatch on Left Ventricular Mass Regression Following Aortic Valve Replacement Giordano Tasca, MD, Federico Brunelli, MD, Marco Cirillo, MD, Margherita DallaTomba, MD,
More informationReoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment
Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background
More informationProsthesis-Patient Mismatch or Prosthetic Valve Stenosis?
EuroValves 2015, Nice Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE FESC Canada Research Chair in Valvular Heart Diseases Université LAVAL Disclosure
More informationA Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision
A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction
More informationHemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves
Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Speaker's name: I have the following potential conflicts of interest to report: Proctorship for Medtronic Agenda Failure modes
More informationCarpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience
SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://www.annalsthoracicsurgery.org/cme/ home. To take the CME activity related to this article, you must have either an STS member
More informationReverse left atrium and left ventricle remodeling after aortic valve interventions
Reverse left atrium and left ventricle remodeling after aortic valve interventions Alexandra Gonçalves, Cristina Gavina, Carlos Almeria, Pedro Marcos-Alberca, Gisela Feltes, Rosanna Hernández-Antolín,
More informationAssessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography
Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler and Dobutamine Stress Echocardiography Akira Sezai, MD, PhD, Yuji Kasamaki, MD, PhD, Keisuke Abe, RMS, Mitsumasa Hata,
More informationThe implantation of bioprostheses is the preferred. Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement
Influence of Prosthesis Patient Mismatch on Diastolic Heart Failure After Aortic Valve Replacement Shahab Nozohoor, MD, Johan Nilsson, MD, PhD, Carsten Lührs, MD, Anders Roijer, MD, PhD, and Johan Sjögren,
More informationAortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon?
Aortic Valve Replacement With 17-mm Mechanical Prostheses: Is Patient Prosthesis Mismatch a Relevant Phenomenon? Andrea Garatti, MD, Francesca Mori, MD, Francesco Innocente, MD, Alberto Canziani, MD, Piervincenzo
More informationHani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.
Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility
More informationThe results of aortic valve (AV) surgery continue to improve
Predictors of Low Cardiac Output Syndrome After Isolated Aortic Valve Surgery Manjula D. Maganti, MSc; Vivek Rao, MD, PhD; Michael A. Borger, MD, PhD; Joan Ivanov, PhD; Tirone E. David, MD Background Low
More informationCLINICAL COMMUNIQUE 16 YEAR RESULTS
CLINICAL COMMUNIQUE 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 Introduction The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 6900, was introduced
More informationShould early elective surgery be performed in patients with severe but asymptomatic aortic stenosis?
European Heart Journal (2002) 23, 1417 1421 doi:10.1053/euhj.2002.3163, available online at http://www.idealibrary.com on Clinical Perspective Should early elective surgery be performed in patients with
More informationIn , three studies described patients
Heart 2001;85:337 341 VALVE DISEASE Should patients with asymptomatic mild or moderate aortic stenosis undergoing coronary artery bypass surgery also have valve replacement for their aortic stenosis? Shahbudin
More informationIndication, Timing, Assessment and Update on TAVI
Indication, Timing, Assessment and Update on TAVI Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Starr- Edwards Mechanical
More informationAortic stenosis (AS) is common with the aging population.
New Insights Into the Progression of Aortic Stenosis Implications for Secondary Prevention Sanjeev Palta, MD; Anita M. Pai, MD; Kanwaljit S. Gill, MD; Ramdas G. Pai, MD Background The risk factors affecting
More information16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900
CLINICAL COMMUNIQUé 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 69 The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 69, was introduced into clinical
More information15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses
ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,
More informationTranscatheter Aortic Valve Implantation in Patients With Severe Aortic Stenosis and Small Aortic Annulus
Journal of the American College of Cardiology Vol. 58, No. 10, 2011 2011 by the American College of Cardiology Foundation ISSN 0735-1097/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2011.05.026
More informationValvular Guidelines: The Past, the Present, the Future
Valvular Guidelines: The Past, the Present, the Future Robert O. Bonow, MD, MS Northwestern University Feinberg School of Medicine Bluhm Cardiovascular Institute Northwestern Memorial Hospital Editor-in-Chief,
More informationInfluence of patient gender on mortality after aortic valve replacement for aortic stenosis
Influence of patient gender on mortality after aortic valve replacement for aortic stenosis Jennifer Higgins, MD, W. R. Eric Jamieson, MD, Osama Benhameid, MD, Jian Ye, MD, Anson Cheung, MD, Peter Skarsgard,
More informationThe St. Jude Medical Biocor Bioprosthesis
The St. Jude Medical Biocor Bioprosthesis Clinical Evidence of Long-term Durability Long-term Biocor Experience A Review and Comparative Assessment Long-term Biocor Stented Tissue Valve Studies Twenty-year
More informationLate incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves
Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves Marc Ruel, MD, MPH a,b Fraser D. Rubens, MD a Roy G. Masters, MD a Andrew L. Pipe, MD a
More information2880 DE AGUSTIN JA et al. Circ J 2018; 82: ORIGINAL ARTICLE doi: /circj.CJ Discongruence Index
2880 DE AGUSTIN JA et al. Circ J 2018; 82: 2880 2886 ORIGINAL ARTICLE doi: 10.1253/circj.CJ-18-0298 Valvular Heart Disease Discongruence Index Simple Indicator to Predict Prosthesis-Patient Mismatch After
More informationThree-Dimensional P3 Tethering Angle at the Heart of Future Surgical Decision Making in Ischemic Mitral Regurgitation
Accepted Manuscript Three-Dimensional P3 Tethering Angle at the Heart of Future Surgical Decision Making in Ischemic Mitral Regurgitation Wobbe Bouma, MD PhD, Robert C. Gorman, MD PII: S0022-5223(18)32805-8
More informationValvular Heart Disease
Valvular Heart Disease Prosthesis-Patient Mismatch Predicts Structural Valve Degeneration in Bioprosthetic Heart Valves Willem Flameng, MD, PhD; Marie-Christine Herregods, MD, PhD; Monique Vercalsteren;
More informationAortic valve implantation using the femoral and apical access: a single center experience.
Aortic valve implantation using the femoral and apical access: a single center experience. R. Hoffmann, K. Brehmer, R. Koos, R. Autschbach, N. Marx, G. Dohmen Rainer Hoffmann, University Aachen, Germany
More informationEchocardiographic changes after aortic valve replacement: Does the failure rate of mitral valve change? Original Article
Echocardiographic changes after aortic valve replacement: Does the failure rate of mitral valve change? Abstract Arezoo Khosravi (1), Hadi Sheykhloo (2), Reza Karbasi-Afshar (1), Amin Saburi (3) Original
More information: mm 86 mm EF mm
37 Vol. 35, pp. 37 42, 2007 2 3 : 9 6 22 68 40 2003 4 Ejection fraction: EF44 IV 70 mm 86 mm EF46 6 24 mm 4 mm EF 80 60 mm 70 mm Aortic Regurgitation: AR 2 3 AR Aortic Valve Replacement: AVR AR 38 : 68
More informationComparison of eight prosthetic aortic valves in a cadaver model
Thomas Jefferson University Jefferson Digital Commons Department of Surgery Faculty Papers Department of Surgery July 2007 Comparison of eight prosthetic aortic valves in a cadaver model Benjamin A. Youdelman
More informationFavorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position
Favorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position Kazuhito Imanaka, MD, Shinichi Takamoto, MD, and Akira Furuse, MD 2 Hemodynamic performance of the CarboMedics
More informationSuccessful Transfemoral Edwards Sapien Aortic. Valve Implantation in a Patient with Previous. Mitral Valve Replacement
Advanced Studies in Medical Sciences, Vol. 2, 2014, no. 1, 37-45 HIKARI Ltd, www.m-hikari.com http://dx.doi.org/10.12988/asms.2014.31213 Successful Transfemoral Edwards Sapien Aortic Valve Implantation
More informationProsthesis-Patient Mismatch after Mitral Valve Replacement: Comparison of Different Methods of Effective Orifice Area Calculation
Original Article Yonsei Med J 2016 Mar;57(2):328-336 pissn: 0513-5796 eissn: 1976-2437 Prosthesis-Patient Mismatch after Mitral Valve Replacement: Comparison of Different Methods of Effective Orifice Area
More informationTranscatheter Aortic Valve Replacement: Current and Future Devices: How do They Work, Eligibility, Review of Data
Transcatheter Aortic Valve Replacement: Current and Future Devices: How do They Work, Eligibility, Review of Data Echo Florida 2013 Jonathan J. Passeri, M.D. Co-Director, Heart Valve Program Director,
More informationAortic Stenosis: UPDATE Anjan Sinha, MD Krannert Institute of Cardiology
Aortic Stenosis: UPDATE 2010 Anjan Sinha, MD Krannert Institute of Cardiology None Disclosures 67-Year-Old Male Dyspnea and angina Class III heart failure No PND or orthopnea 3/6 late peak SEM Diminished
More informationSotirios N. Prapas, M.D., Ph.D, F.E.C.T.S.
CORONARY ARTERY REVASCULARIZATION WITH MILD AORTIC STENOSIS: STRATEGIES OF TREATMENT 9 th ANNUAL MEETING OF THE EAB SOCIETY, Pravets, Bulgaria, 2012 Sotirios N. Prapas, M.D., Ph.D, F.E.C.T.S. Director
More informationIschemic Mitral Valve Disease: Repair, Replace or Ignore?
Ischemic Mitral Valve Disease: Repair, Replace or Ignore? Fabio B. Jatene Full Professor of Cardiovascular Surgery, Medical School, University of São Paulo, Brazil DISCLOSURE I have no financial relationship
More informationDoppler echocardiography is currently the
Doppler Echocardiography of 119 Normal-functioning St Jude Medical Mitral Valve Prostheses: A Comprehensive Assessment Including Time-velocity Integral Ratio and Prosthesis Performance Index* Joseph F.
More informationTreatment decision in asymptomatic aortic valve stenosis: role of exercise testing
Heart 2001;86:381 386 381 Heart Institute (InCor) of the University of São Paulo and Jaraguá Hospital, Av Juriti 144 Moema, 04520-000 São Paulo, Brazil M C M Amato PJMoVa K E Werner J A F Ramires Correspondence
More informationLong-term Mortality Predictors in Patients with Small Aortic Annulus Undergoing Aortic Valve Replacement with a 19- or 21-mm Bioprosthesis
ORIGINAL ARTICLE Long-term Mortality Predictors in Patients with Small Aortic Annulus Undergoing Aortic Valve Replacement with a 19- or 21-mm Bioprosthesis Jenny Lourdes Rivas de Oliveira 1, MD; Renato
More informationValue of echocardiography in chronic dyspnea
Value of echocardiography in chronic dyspnea Jahrestagung Schweizerische Gesellschaft für /Schweizerische Gesellschaft für Pneumologie B. Kaufmann 16.06.2016 Chronic dyspnea Shortness of breath lasting
More informationTAVR in patients with. End-Stage CKD or in Renal Replacement Therapy:
TAVR in patients with End-Stage CKD or in Renal Replacement Therapy: Special Considerations and Prevention of early Valve Failure Antonios Chalapas, MD, PhD, FESC THV & Hygeia Hospital Heart Team Athens,
More informationTAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair?
TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? Elaine E. Tseng, MD and Marlene Grenon, MD Department of Surgery Divisions of Adult Cardiothoracic and Vascular and Endovascular
More informationCoreValve in a Degenerative Surgical Valve
CoreValve in a Degenerative Surgical Valve Ran Kornowski, MD, FESC, FACC Chairman Department of Cardiology Rabin Medical Center, Petach Tikva, Israel Disclosure Statement of Financial Interest I, Ran Kornowski,
More informationThe use of mitral valve (MV) repair to correct mitral
Outcomes and Long-Term Survival for Patients Undergoing Repair Versus Effect of Age and Concomitant Coronary Artery Bypass Grafting Vinod H. Thourani, MD; William S. Weintraub, MD; Robert A. Guyton, MD;
More informationEarly Surgery in Asymptomatic Severe Aortic Stenosis Pros and Cons
Early Surgery in Asymptomatic Severe Aortic Stenosis Pros and Cons Duk-Hyun Kang, M.D. Division of Cardiology Asan Medical Center Seoul, Korea Background Dilemma of balancing the risks versus benefits
More informationRisk stratification of severe aortic stenosis according to new guidelines: long term outcomes
Original Article Risk stratification of severe aortic stenosis according to new guidelines: long term outcomes Andrea Colli, Eleonora Bizzotto, Laura Besola, Dario Gregori, Francesca Toto, Erica Manzan,
More informationEchocardiographic Evaluation of Aortic Valve Prosthesis
Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co-Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System
More informationTitle of image and video article Sub-Acute Leaflet Thrombosis: A Reversible Cause of Aortic Stenosis
Page 1 of 5 Title of image and video article Sub-Acute Leaflet Thrombosis: A Reversible Cause of Aortic Stenosis Authors Athina Chasapi, Adam Hobbs, Theodore Velissaris & Benoy N Shah. Wessex Cardiac &
More informationProsthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis
Prosthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis George L. Zorn, III On Behalf of the CoreValve US Clinical Investigators
More informationIncidence And Predictors Of Left Bundle Branch Block After Transcatheter Aortic Valve Implantation
Incidence And Predictors Of Left Bundle Branch Block After Transcatheter Aortic Valve Implantation Ömer Aktug 1, MD; Guido Dohmen 2, MD; Kathrin Brehmer 1, MD; Verena Deserno 1 ; Ralf Herpertz 1 ; Rüdiger
More informationSeventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter?
Hu et al. Journal of Cardiothoracic Surgery 2014, 9:17 RESEARCH ARTICLE Open Access Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient
More informationSurgery for Valvular Heart Disease. Very Long-Term Survival and Durability of Mitral Valve Repair for Mitral Valve Prolapse
Surgery for Valvular Heart Disease Very Long-Term Survival and Durability of Mitral Valve Repair for Mitral Valve Prolapse Dania Mohty, MD; Thomas A. Orszulak, MD; Hartzell V. Schaff, MD; Jean-Francois
More informationSevere aortic stenosis should be operated before symptom onset CONTRA. Helmut Baumgartner
Severe aortic stenosis should be operated before symptom onset CONTRA Helmut Baumgartner Westfälische Wilhelms-Universität Münster Adult Congenital and Valvular Heart Disease Center Dept. of Cardiology
More informationNew Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology
New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.
More informationA Case of Aortic Valve Replacement with St. Jude Medical Regent Valve (First Implant in Japan)
Case Report A Case of Aortic Valve Replacement with St. Jude Medical Regent Valve (First Implant in Japan) Akira Sezai, MD, 1 Motomi Shiono, MD, 1 Kenji Akiyama, MD, 1 Yukihiko Orime, MD, 1 Mitsumasa Hata,
More informationPredicting functional mitral stenosis after restrictive annuloplasty for ischemic mitral regurgitation
Powered by TCPDF (www.tcpdf.org) This is a provisional PDF only. Copyedited and fully formatted version will be made available soon. ONLINE FIRST ISSN: 1897-5593 e-issn: 1898-018X Predicting functional
More informationAortic Stenosis: Spectrum of Disease, Low Flow/Low Gradient and Variants
Aortic Stenosis: Spectrum of Disease, Low Flow/Low Gradient and Variants Martin G. Keane, MD, FASE Professor of Medicine Lewis Katz School of Medicine at Temple University Basic root structure Parasternal
More informationClinical material and methods. Copyright by ICR Publishers 2007
16847_JHVD_Biancari_3197_(116-121)_r1:Layout 1 21/3/07 17:07 Page 116 Predicting Immediate and Late Outcome after Surgery for Mitral Valve Regurgitation with EuroSCORE Jouni Heikkinen, Fausto Biancari,
More informationSevere left ventricular dysfunction and valvular heart disease: should we operate?
Severe left ventricular dysfunction and valvular heart disease: should we operate? Laurie SOULAT DUFOUR Hôpital Saint Antoine Service de cardiologie Pr A. COHEN JESFC 16 janvier 2016 Disclosure : No conflict
More informationSurgical AVR: Are there any contraindications? Pyowon Park Samsung Medical Center Seoul, Korea
Surgical AVR: Are there any contraindications? Pyowon Park Samsung Medical Center Seoul, Korea Contents Decision making in surgical AVR in old age Clinical results of AVR with tissue valve Impact of 19mm
More informationMechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute
Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Assistant Professor, Georgetown School of Medicine
More informationMichigan Society of Echocardiography 30 th Year Jubilee
Michigan Society of Echocardiography 30 th Year Jubilee Stress Echocardiography in Valvular Heart Disease Moving Beyond CAD Karthik Ananthasubramaniam, MD FRCP (Glas) FACC FASE FASNC Associate Professor
More informationOutcome of elderly patients with severe but asymptomatic aortic stenosis
Outcome of elderly patients with severe but asymptomatic aortic stenosis Robert Zilberszac, Harald Gabriel, Gerald Maurer, Raphael Rosenhek Department of Cardiology Medical University of Vienna ESC Congress
More informationNatural History and Echo Evaluation of Aortic Stenosis
Natural History and Echo Evaluation of Aortic Stenosis Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM AORTIC STENOSIS First valvular disease
More informationMedtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance
Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,
More informationRepeated mitral valve replacement in a patient with extensive annular calcification
CASE REPORT Open Access Repeated mitral valve replacement in a patient with extensive annular calcification Tadashi Kitamura 1,2*, Sachito Fukuda 1, Takahiro Sawada 1, Sumio Miura 1, Ikutaro Kigawa 1,3
More informationImaging in TAVI. Jeroen J Bax Dept of Cardiology Leiden Univ Medical Center The Netherlands Davos, feb 2013
Imaging in TAVI Jeroen J Bax Dept of Cardiology Leiden Univ Medical Center The Netherlands Davos, feb 2013 Research grants: Medtronic, Biotronik, Boston Scientific, St Jude, BMS imaging, GE Healthcare,
More informationChronic Primary Mitral Regurgitation
Chronic Primary Mitral Regurgitation The Case For Early Surgical Intervention William K. Freeman, MD, FACC, FASE DISCLOSURES Relevant Financial Relationship(s) None Off Label Usage None Watchful Waiting......
More informationAffecting the elderly Requiring new approaches. Echocardiographic Evaluation of Hemodynamic Severity. Increasing prevalence Mostly degenerative
Echocardiographic Evaluation of Hemodynamic Severity Steven J. Lester MD, FACC, FRCP(C), FASE Mayo Clinic, Arizona Relevant Financial Relationship(s) None Off Label Usage None A re-emerging public-health
More informationECHO HAWAII. Role of Stress Echo in Valvular Heart Disease. Not only ischemia! Cardiomyopathy. Prosthetic Valve. Diastolic Dysfunction
Role of Stress Echo in Valvular Heart Disease ECHO HAWAII January 15 19, 2018 Kenya Kusunose, MD, PhD, FASE Tokushima University Hospital Japan Not only ischemia! Cardiomyopathy Prosthetic Valve Diastolic
More informationAppropriate Patient Selection or Healthcare Rationing? Lessons from Surgical Aortic Valve Replacement in The PARTNER I Trial Wilson Y.
Appropriate Patient Selection or Healthcare Rationing? Lessons from Surgical Aortic Valve Replacement in The PARTNER I Trial Wilson Y. Szeto, MD on behalf of The PARTNER Trial Investigators and The PARTNER
More information