1 Doppler Echocardiography of 119 Normal-functioning St Jude Medical Mitral Valve Prostheses: A Comprehensive Assessment Including Time-velocity Integral Ratio and Prosthesis Performance Index* Joseph F. Malouf, MD, Manfredi Ballo, MD, Heidi M. Connolly, MD, David O. Hodge, MS, Regina M. Herges, BS, Charles J. Mullany, MD, and Fletcher A. Miller, Jr, MD, Rochester, Minnesota The purpose of this study was to provide, in a large number of patients, comprehensive Doppler echocardiographic assessment of normal St Jude Medical mitral valve prosthesis function using Doppler-derived hemodynamic variables, including the mitral valve prosthesis to left ventricular outflow tract time-velocity integral ratio and prosthesis performance index. The pressure half-time was less than 130 milliseconds in all patients, and all but one patient had either a peak early mitral diastolic velocity of 2 m/s or less or a mitral valve prosthesis to left ventricular outflow tract time-velocity integral ratio of less than 2.2. There was a significant (P <.001) negative correlation between the prosthesis performance index and prosthesis size. This negative correlation suggests that there is more efficient use of the in vitro geometric orifice area with smaller prostheses. (J Am Soc Echocardiogr 2005;18: ) Doppler echocardiography is currently the method of choice for assessing normal and abnormal prosthetic valve hemodynamics. Normal Dopplerderived values for mean gradients and effective orifice area (EOA) for different sizes of St Jude Medical (SJM) mitral valve prostheses (mvp) have been reported. 1-4 However, the collective experience is limited to a few relatively small series and case reports. 1-4 A large series is needed because of the large variability in Doppler-derived EOA for each prosthesis size. Moreover, the Doppler indices that have been recently proposed to separate normal from abnormal prosthesis function have not been validated in a large number of patients with normalfunctioning SJM prostheses. Data are limited on the correlation between Doppler-derived EOAs and the actual geometric orifice area (GOA), as provided by the manufacturer. 5,6 From the Divisions of Cardiovascular Diseases (J.F.M., M.B., H.M.C., F.A.M.), Biostatistics (D.O.H., R.M.H.), and Cardiovascular Surgery (C.J.M.), Mayo Clinic. * Nothing in this article implies endorsement of the products of St Jude Medical Inc. Reprint requests: Joseph F. Malouf, MD, Division of Cardiovascular Diseases, Mayo Clinic, 200 First St SW, Rochester, MN /$30.00 Copyright 2005 by the American Society of Echocardiography. doi: /j.echo The purpose of this study was 2-fold: first, to define, in a large number of patients, normal SJM mvp function using all the important Doppler-derived hemodynamic variables that have been described; and, second, to study the correlation between in vivo Doppler-derived EOA and in vitro GOA. METHODS Patient Selection From our institutional cardiac surgical database, we identified 122 patients aged 18 years or older, who underwent isolated mitral valve replacement with an SJM prosthesis between 1993 and 2001, and who had 2-dimensional (2D) and Doppler echocardiographic studies within 4 weeks after operation. Three patients with more than mild aortic regurgitation were excluded. The remaining 119 patients formed the final study population. The physical examination and appearance of the prosthesis by both intraoperative transesophageal echocardiography (TEE) and follow-up transthoracic echocardiography (TTE) were normal in each of these patients. Routine intraoperative TEE and postoperative TTE evaluation of mitral prosthesis function included 2D imaging of the sewing ring and disk occluders to assess morphologic abnormalities, color-flow Doppler imaging to detect flow alteration, and spectral Doppler assessment of prosthesis hemodynamics. This 252
2 Volume 18 Number 3 Malouf et al 253 study was approved by the institutional review board, and all patients had consented to have their medical records accessed. Doppler Data Doppler data were obtained from review of the postoperative TTE reports. Missing measurements were obtained using offline analysis (Digiview Release 3.0, Digisonics, Houston, Tex) by two investigators (M.B., F.A.M.). The prosthesis EOA was measured using both pressure halftime (PHT) and continuity (CON) methods. Prosthesis size ranged from 25 to 33 mm. Because the GOAs and hemodynamics of the 31- and 33-mm prostheses are identical, they were evaluated together. The prosthesis performance index (PPI) was calculated as the ratio of the EOA derived from the CON method to the GOA as provided by the manufacturer (3.09 cm 2, 3.67 cm 2, 4.52 cm 2, and 5.18 cm 2 for the 25-mm, 27-mm, 29-mm, and the 31- or 33-mm prosthesis size, respectively). 2,5 Other calculated variables included mean diastolic gradient, EOA indexed to body surface area, EOA/prosthesis size, mvp to left ventricular (LV) outflow tract (LVOT) time-velocity integral (TVI) ratio (TVI mvp /TVI LVOT ), peak early mitral diastolic velocity (E velocity), and E velocity/tvi LVOT. LV ejection fraction (LVEF) was obtained by visual estimate in 63 patients (53%), D measurements in 47 patients (40%), and M-mode measurements in 9 patients (8%). Standard practice in our laboratory is to average the Doppler measurements of 3 cycles if a patient is in sinus rhythm and 5 cycles or more when a patient is in atrial fibrillation or other irregular rhythm. Statistical Methods Measurements were compared between the CON and PHT methods using a paired t test. Continuous variables were compared among the 4 valve size groups using the analysis of variance procedure. Significant differences were investigated, adjusting for multiple comparisons using the Student-Newman-Keuls procedure. RESULTS Table Overall baseline characteristics (N 119) Characteristic Finding* Range Age, y Female, No. (%) 81 (68)... Heart rate, beats/min Hemoglobin, g/dl Hematocrit, % BSA, m LVEF, % LVEF 50%, No. (%) 28 (24)... SV LVOT, ml PHT, ms MG, mm Hg EVLCTY, m/s TVI mvp,cm EVLCTY/TVI LVOT TVI mvp /TVI LVOT EOA CON, cm EOA PHT, cm IEOA CON, cm 2 /m IEOA PHT, cm 2 /m EOA CON/prosth size ratio EOA PHT/prosth size ratio PPI CON BSA, Body surface area; CON, continuity method; EOA, effective orifice area; EVLCTY, E velocity; IEOA, indexed effective orifice area; LVEF, left ventricular ejection fraction; LVOT, left ventricular outflow tract; MG, mean gradient; mvp, mitral valve prosthesis; PHT, pressure half-time; PPI, prosthesis performance index; prosth, prosthesis; SV, stroke volume; TVI, timevelocity integral. *Values are mean SD unless indicated otherwise. Baseline Clinical Characteristics Patient characteristics are listed in the Table. A standard bileaflet SJM prosthesis was implanted in all but one patient who received a 25-mm Hemodynamic-Plus SJM prosthesis (St. Jude Medical, St. Paul, Minn). A total of 99 patients (85%) had their TTE within a week from mitral valve replacement. The rhythm was sinus in 69 patients (58%), atrial fibrillation or flutter (one patient) in 41 patients (34%), and paced or junctional in 9 patients (8%). The indications for mitral valve replacement were mitral regurgitation in 53 patients (45%), mitral stenosis in 33 patients (28%), and mixed stenosis and regurgitation in 33 patients (28%). At the time of echocardiography, mean SD heart rate was 82 14/min (median, 83/min; range, /min). Complete Doppler hemodynamic data were available in 109 patients (92%), and missing data were limited to the TVI LVOT. A total of 51 patients (43%) had either trivial or mild aortic regurgitation, and 30 patients (25%) had trivial to mild mitral prosthesis regurgitation, including one patient with mild periprosthetic regurgitation. Hemodynamic data, grouped by valve size, are detailed in Figures 1 through 6. The EOA and indexed EOA were significantly lower by the CON method than by the PHT method (P.001). E velocity did not exceed 2 m/s in 104 patients (87%) and in all but 1 of the 28 patients with an LVEF less than 50%. TVI mvp /TVI LVOT was less than 2.2 in 91 of the 109 patients (83%) with complete TVI data, including 11 of the 12 patients with an E velocity exceeding 2 m/s. Among the 18 patients with a TVI mvp /TVI LVOT of 2.2 or higher, 8 patients had an LVEF lower than 50%. The PHT was less than 130 milliseconds in all patients, and all but one patient (99%) had either an E velocity of 2 m/s or less or a TVI mvp /TVI LVOT of less than 2.2. The combination of E velocity less than 1.9 m/s, TVI mvp /TVI LVOT less than 2.2, and PHT less than 130
3 254 Malouf et al March 2005 Figure 1 Plot of mean prosthesis gradient versus valve size. Error bars, Mean 2 SD. Figure 3 Plot of mitral valve prosthesis to left ventricular outflow tract time-velocity integral ratio (TVI mvp /TVI L - VOT) versus valve size. Error bars, Mean 2 SD. Figure 2 Plot of E velocity versus valve size. Error bars, Mean 2 SD. milliseconds 11 was found in 67 of the 109 patients (61%) with complete data. The mean mitral gradient exceeded 5 mm Hg in 28 patients (6 mm Hg in 16 patients, 7 mm Hg in 5 patients, 8 mm Hg in 2 patients, 9 mm Hg in 3 patients, and 10 and 12 mm Hg in 1 patient each). Among these patients, the heart rate was 90/min or higher in 14 patients (50%), and 24 patients (86%) had a TVI mvp /TVI LVOT of less than 2.2 (16 patients), LVEF less than 50% (6 patients), or both (two patients). There was a significant negative correlation between severity of anemia, as measured by either the hemoglobin level or hematocrit reading, and the mitral gradient (P.001). Effect of Prosthesis Size Valve size was 25 mm in 7 patients (6%), 27 mm in 30 patients (25%), 29 mm in 38 patients (32%), and 31 or 33 mm in 44 patients (37%). There was no significant difference in the mean mitral gradient Figure 4 Plot of prosthesis effective orifice area (EOA) by continuity method versus valve size. Error bars, Mean 2 SD. among the 4 valve size groups after adjusting for multiple comparisons. There was a trend of increasing EOA by the CON method with increasing valve size, but the mean measurements of the EOA and indexed EOA for the different valve sizes by either the PHT or CON method were not significantly different. The mean PPI decreased with increasing valve size ( , , , and for the 25-mm, 27-mm, 29-mm, and 31- or 33-mm groups, respectively). This difference was significant among the 4 groups (P.001). The 29-mm and 31- or 33-mm groups were significantly different from the 25-mm and 27-mm groups. No other groups were significantly different from one another. There was no significant difference among the different valve sizes with respect to the mean
4 Volume 18 Number 3 Malouf et al 255 Figure 5 Plot of prosthesis effective orifice area (EOA) by continuity method indexed to body surface area versus valve size. Error bars, Mean 2 SD. Figure 6 Plot of prosthesis performance index (PPI) versus valve size. Error bars, Mean 2 SD. TVI mvp /TVI LVOT or the mean E velocity TVI LVOT ratio. DISCUSSION To our knowledge, this is the largest single-center study profiling normal Doppler hemodynamics of SJM mvp. The PHT was less than 130 milliseconds in all patients, and all but one patient had either an E velocity that did not exceed 2 m/s or a TVI mvp / TVI LVOT less than 2.2, regardless of prosthesis size or LV systolic function. The previous collective experience is limited to 4 studies with a total of 102 prostheses of different sizes, 4 and only the study by Bitar et al 3 provided normal values for EOA using the CON equation. The EOA is a measure of the physiologic area occupied by blood flow and not the true anatomic orifice area of the prosthesis. Importantly, EOA, not GOA, determines the mitral diastolic gradient, hence the persistence or progression of pulmonary hypertension. 12 EOA by the PHT method has serious limitations. Dumesnil et al 13 showed that EOA by the CON and not the PHT method correlates best with the in vitro EOA of normal mitral bioprostheses and that EOA by the PHT method tends to grossly overestimate the in vitro EOA. Both our study and that of Bitar et al 3 demonstrate that EOAs for SJM prostheses by the PHT method are significantly larger than those obtained by the CON method, and Bitar et al 3 showed more variability in the EOA by the PHT method. PPI, the ratio between the in vivo mitral prosthesis EOA and the in vitro GOA, has not been well studied. 2,5 In the report by Bitar et al, 3 the mean PPI was less than 0.5, irrespective of prosthesis size. The significant negative correlation we found between the PPI and prosthesis size suggests that smaller prostheses use the in vitro GOA more efficiently. The EOA for the same SJM valve size has been shown to increase with increasing flow rates. 3 Therefore, higher flow rates across smaller SJM mitral prostheses may have contributed to a more sustained opening of the valve leaflets compared with larger prostheses, 3 thus, accounting for the decreasing PPI with increasing prosthesis valve size. More data are needed, however. Several studies have attempted to define the Doppler hemodynamic profile that best describes normal SJM prosthesis function. In the study by Panidis et al 1 of 44 normal-functioning SJM prostheses, E velocity did not exceed 2.2 m/s among patients with either 27-mm or 29-mm prostheses and did not exceed 2.5 m/s among patients with 31-mm prostheses. In their study, the mean of the mean gradients was 5 mm Hg or less, regardless of prosthesis size, but peak mean gradients of 7 mm Hg, 8 mm Hg, and 14 mm Hg were recorded for the 27-mm, 29-mm, and 31-mm prostheses, respectively. In the report by Bitar et al, 3 E velocity exceeded 2 m/s in only one of the 40 patients with SJM mitral prostheses, and the mean gradient exceeded 5 mm Hg in only 5 patients, all of whom had a heart rate higher than 90/min. Fernandes et al 11 were first to propose TVI mvp / TVI LVOT as an important clue to mechanical prosthesis obstruction or regurgitation. The TVI mvp can be increased by obstruction, regurgitation, or increased cardiac output. With the latter, the TVI LVOT also increases, keeping the ratio less than 2.2. The study by Fernandes et al 11 included 73 normal mechanical mvp. Although the majority were bileaflet prostheses, the number of SJM prostheses was not specified. They found the combination of E velocity less than 1.9 m/s, TVI mvp /TVI LVOT less than 2.2, and PHT less than 130 milliseconds to be highly predictive of normal prosthesis function, verified by TEE, with less than 2% probability of valve dysfunction. In
5 256 Malouf et al March 2005 contrast, fewer than two-thirds of the patients in our study had all 3 of these characteristics. Potential Limitations This was a retrospective study. Significant (moderate or severe) mitral regurgitation can influence mitral prosthesis forward flow hemodynamics. These prostheses were defined as normally functioning on the basis of normal physical examination findings, the short time between implant and TTE, and normal 2D and color Doppler appearance of the valves by both intraoperative TEE and postoperative TTE. Although we think it is unlikely, significant mitral regurgitation may have been overlooked in a few of these patients by these screening criteria. It was impossible to be certain that the lengths of the cardiac cycle that were used for LVOT and mitral flow measurements were matched for patients with atrial fibrillation. However, measurement of multiple cycles for such patients makes significant error from varying R-R intervals unlikely. Conclusions Ascertaining normal heart prosthesis function remains a challenge. The intent of this study was to provide in vivo data from TTE on a large number of normally functioning SJM mvp, including all the important Doppler variables described to date. The findings can be used for comparison when assessing patients with SJM mitral prostheses. Outlier Doppler data should prompt further evaluation, including TEE, for possible prosthesis dysfunction. In particular, PHT of 130 milliseconds or higher warrants TEE to evaluate bileaflet motion and to check for thrombus, pannus, or both. Either E velocity greater than 2 m/s or TVI mvp /TVI LVOT of 2.2 or higher in the presence of a PHT less than 130 milliseconds warrants consideration of TEE to check for significant prosthetic or periprosthetic regurgitation. REFERENCES 1. Panidis IP, Ross J, Mintz GS. Normal and abnormal prosthetic valve function as assessed by Doppler echocardiography. J Am Coll Cardiol 1986;8: Reisner SA, Harpaz D, Skulski R, Borenstein D, Milo S, Meltzer RS. Hemodynamic performance of four mechanical bileaflet prosthetic valves in the mitral position: an echocardiographic study. Eur J Ultrasound 1998;8: Bitar JN, Lechin ME, Salazar G, Zoghbi WA. Doppler echocardiographic assessment with the continuity equation of St Jude Medical mechanical prostheses in the mitral valve position [erratum in Am J Cardiol 1995;76:642]. Am J Cardiol 1995;76: Rosenhek R, Binder T, Maurer G, Baumgartner H. Normal values for Doppler echocardiographic assessment of heart valve prostheses. J Am Soc Echocardiogr 2003;16: Rashtian MY, Stevenson DM, Allen DT, Yoganathan AP, Harrison EC, Edmiston WA, et al. Flow characteristics of four commonly used mechanical heart valves. Am J Cardiol 1986; 58: Stamm RB, Carabello BA, Mayers DL, Martin RP. Twodimensional echocardiographic measurement of left ventricular ejection fraction: prospective analysis of what constitutes an adequate determination. Am Heart J 1982;104: Rich S, Sheikh A, Gallastegui J, Kondos GT, Mason T, Lam W. Determination of left ventricular ejection fraction by visual estimation during real-time two-dimensional echocardiography. Am Heart J 1982;104: Amico AF, Lichtenberg GS, Reisner SA, Stone CK, Schwartz RG, Meltzer RS. Superiority of visual versus computerized echocardiographic estimation of radionuclide left ventricular ejection fraction. Am Heart J 1989;118: Mueller X, Stauffer JC, Jaussi A, Goy JJ, Kappenberger L. Subjective visual echocardiographic estimate of left ventricular ejection fraction as an alternative to conventional echocardiographic methods: comparison with contrast angiography. Clin Cardiol 1991;14: Enriquez-Sarano M, Tajik AJ, Schaff HV, Orszulak TA, Bailey KR, Frye RL. Echocardiographic prediction of survival after surgical correction of organic mitral regurgitation. Circulation 1994;90: Fernandes V, Olmos L, Nagueh SF, Quinones MA, Zoghbi WA. Peak early diastolic velocity rather than pressure half-time is the best index of mechanical prosthetic mitral valve function. Am J Cardiol 2002;89: Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg 1992;6 (Suppl 1):S Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of mitral prosthetic valvular areas calculated by Doppler echocardiography. Am J Cardiol 1990; 65:
Prosthetic valve dysfunction: stenosis or regurgitation
Prosthetic valve dysfunction: stenosis or regurgitation Jean G. Dumesnil MD, FRCP(C), FACC, FASE(Hon) Quebec Heart and Lung Institute, Québec, Québec No disclosures Possible Causes of High Gradients in
More informationProsthesis-Patient Mismatch or Prosthetic Valve Stenosis?
EuroValves 2015, Nice Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis? Philippe Pibarot, DVM, PhD, FACC, FAHA, FASE FESC Canada Research Chair in Valvular Heart Diseases Université LAVAL Disclosure
More informationEchocardiographic Evaluation of Mitral Valve Prostheses
Echocardiographic Evaluation of Mitral Valve Prostheses Dennis A. Tighe, M.D., FACC, FACP, FASE Cardiovascular Medicine University of Massachusetts Medical School Worcester, MA www.asecho.org 1 Nishimura
More informationA Practical Approach to Prosthetic Valves
A Practical Approach to Prosthetic Valves Bonita Anderson DMU (Cardiac), MApplSc (Med Ultrasound), ACS, AMS, FASE https://doi.org/10.1161/circulationaha.108.778886 Disclosures None 1 Know the Product Know
More information25 different brand names >44 different models Sizes mm
Types of Prosthetic Valves BIOLOGIC STENTED Porcine xenograft Pericardial xenograft STENTLESS Porcine xenograft Pericardial xenograft Homograft (allograft) Autograft PERCUTANEOUS MECHANICAL Bileaflet Single
More informationDoes Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?
Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles
More informationEchocardiographic Evaluation of Aortic Valve Prosthesis
Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co-Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System
More informationEvaluation of St. Jude Medical Valve s Long-term Function by Doppler Echocardiography
Evaluation of St. Jude Medical Valve s Long-term Function by Doppler Echocardiography Akira Sezai, MD, Motomi Shiono, MD, Kenji Akiyama, MD, Yukihiko Orime, MD, Hiroaki Hata, MD, Shinya Yagi, MD, Tomonori
More informationSevere Aortic Stenosis With Low Transvalvular Gradient and Severe Left Ventricular Dysfunction
Severe Aortic Stenosis With Low Transvalvular Gradient and Severe Left Ventricular Dysfunction Result of Aortic Valve Replacement in 52 Patients Heidi M. Connolly, MD; Jae K. Oh, MD; Hartzell V. Schaff,
More informationCases of Abnormal Prosthetic Valves
Cases of Abnormal Prosthetic Valves Sunil Mankad, MD, FACC, FCCP, FASE Associate Professor of Medicine Mayo Clinic College of Medicine Director, Transesophageal Echcoardiography Associate Director, Cardiology
More informationECHOCARDIOGRAPHY DATA REPORT FORM
Patient ID Patient Study ID AVM - - Date of form completion / / 20 Initials of person completing the form mm dd yyyy Study period Preoperative Postoperative Operative 6-month f/u 1-year f/u 2-year f/u
More informationEchocardiographic evaluation of mitral stenosis
Echocardiographic evaluation of mitral stenosis Euroecho 2011 Philippe Unger, MD, FESC Erasme Hospital, ULB, Brussels, Belgium I have nothing to declare EuroHeart Survey Etiology of single native left-sided
More informationIs the Peak-to-Mean Pressure Gradient Ratio Useful for Assessment of Aortic Valve Prosthesis Obstruction? Abstract. Introduction
Original Article Is the Peak-to-Mean Pressure Gradient Ratio Useful for Assessment of Aortic Valve Prosthesis Obstruction? Abstract Maryam Esmaeilzadeh, MD, FACC, FCAPSC *, Ahmad Mirdamadi, MD, Mozhgan
More informationDoppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position
Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position Chee-Siong Soo, MRCP, Mestres Ca, MD, Monica Tay, Joon-Kuan Yeoh, MRCP, Eugene Sim, FRCS, and M. Choo, FRCP,
More informationUncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency
Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency Rahul R. Jhaveri, MD, Muhamed Saric, MD, PhD, FASE, and Itzhak Kronzon, MD, FASE, New York, New York Background: Two-dimensional
More informationEchocardiographic Evaluation of Aortic Valve Prosthesis
Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Co Director, Echocardiography, Director, Interventional Cardiology Research, Beaumont Health System
More informationHemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics
Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial
More informationPROSTHETIC VALVE BOARD REVIEW
PROSTHETIC VALVE BOARD REVIEW The correct answer D This two chamber view shows a porcine mitral prosthesis with the typical appearance of the struts although the leaflets are not well seen. The valve
More informationPatient/prosthesis mismatch: how to evaluate and when to act?
Patient/prosthesis mismatch: how to evaluate and when to act? Svend Aakhus, MD, PhD Oslo University Hospital, Norway Disclosures: No conflict of interest Types of aortic valve prostheses (AVR) Mechanical
More informationDoppler Basic & Hemodynamic Calculations
Doppler Basic & Hemodynamic Calculations August 19, 2017 Smonporn Boonyaratavej MD Division of Cardiology, Department of Medicine Chulalongkorn University Cardiac Center, King Chulalongkorn Memorial Hospital
More informationThe Doppler Examination. Katie Twomley, MD Wake Forest Baptist Health - Lexington
The Doppler Examination Katie Twomley, MD Wake Forest Baptist Health - Lexington OUTLINE Principles/Physics Use in valvular assessment Aortic stenosis (continuity equation) Aortic regurgitation (pressure
More informationQUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH
QUANTIFICATION AND PREVENTION TECHNIQUES OF PROSTHESIS-PATIENT MISMATCH 1,2 Radu A. SASCĂU 3 Cristina OLARIU 1,2 Cristian STĂTESCU 1 Internal Medicine Department, Gr.T.Popa University of Medicine and Pharmacy,
More informationManagement of Difficult Aortic Root, Old and New solutions
Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult
More informationPPM: How to fit a big valve in a small heart
PPM: How to fit a big valve in a small heart Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC King Abdulaziz Cardiac Centre National Guard Health Affairs Riyadh, Saudi Arabia GHA meeting Muscat
More informationCopyright by ICR Publishers 2014
Comprehensive Hemodynamic Performance and Frequency of Patient-Prosthesis Mismatch of the St. Jude Medical Trifecta Bioprosthetic Aortic Valve Ajay Yadlapati 1, Jimmy Diep 3, Mary-Jo Barnes 2, Tristan
More informationLV geometric and functional changes in VHD: How to assess? Mi-Seung Shin M.D., Ph.D. Gachon University Gil Hospital
LV geometric and functional changes in VHD: How to assess? Mi-Seung Shin M.D., Ph.D. Gachon University Gil Hospital LV inflow across MV LV LV outflow across AV LV LV geometric changes Pressure overload
More informationTAVR: Echo Measurements Pre, Post And Intra Procedure
2017 ASE Florida, Orlando, FL October 10, 2017 8:00 8:25 AM 25 min TAVR: Echo Measurements Pre, Post And Intra Procedure Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology Echo Lab Associate
More informationNomograms for severity of aortic valve stenosis using peak aortic valve pressure gradient and left ventricular ejection fraction
European Journal of Echocardiography (2009) 10, 532 536 doi:10.1093/ejechocard/jen333 Nomograms for severity of aortic valve stenosis using peak aortic valve pressure gradient and left ventricular ejection
More informationHEMODYNAMIC ASSESSMENT
HEMODYNAMIC ASSESSMENT INTRODUCTION Conventionally hemodynamics were obtained by cardiac catheterization. It is possible to determine the same by echocardiography. Methods M-mode & 2D echo alone can provide
More informationBogdan A. Popescu. University of Medicine and Pharmacy Bucharest, Romania. EAE Course, Bucharest, April 2010
Bogdan A. Popescu University of Medicine and Pharmacy Bucharest, Romania EAE Course, Bucharest, April 2010 This is how it started Mitral stenosis at a glance 2D echo narrow diastolic opening of MV leaflets
More informationDOPPLER HEMODYNAMICS (1) QUANTIFICATION OF PRESSURE GRADIENTS and INTRACARDIAC PRESSURES
THORAXCENTRE DOPPLER HEMODYNAMICS (1) QUANTIFICATION OF PRESSURE GRADIENTS and INTRACARDIAC PRESSURES J. Roelandt DOPPLER HEMODYNAMICS Intracardiac pressures and pressure gradients Volumetric measurement
More informationReverse left atrium and left ventricle remodeling after aortic valve interventions
Reverse left atrium and left ventricle remodeling after aortic valve interventions Alexandra Gonçalves, Cristina Gavina, Carlos Almeria, Pedro Marcos-Alberca, Gisela Feltes, Rosanna Hernández-Antolín,
More informationFavorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position
Favorable Results in Patients with Small Size CarboMedics Heart Valves in the Aortic Position Kazuhito Imanaka, MD, Shinichi Takamoto, MD, and Akira Furuse, MD 2 Hemodynamic performance of the CarboMedics
More informationProf. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM
The Patient with Aortic Stenosis and Mitral Regurgitation Prof. Patrizio LANCELLOTTI, MD, PhD Heart Valve Clinic, University of Liège, CHU Sart Tilman, Liège, BELGIUM Aortic Stenosis + Mitral Regurgitation?
More informationEchocardiographic Evaluation of Aortic Valve Prosthesis
Echocardiographic Evaluation of Aortic Valve Prosthesis Amr E Abbas, MD, FACC, FASE, FSCAI, FSVM, RPVI Director, Interventional Cardiology Research, Beaumont Health System Associate Professor of Medicine,
More informationIndicator Mild Moderate Severe
Indicator Mild Moderate Severe Jet velocity (m/s) 2.0-2.9 3.0-3.9 4.0 Mean gradient (mmhg) < 20 20-39 40 Valve area (cm 2 ) 1.0 Valve area index (cm 2 /m 2 ) 0.6 1 Abnormal AV with Reduced Systolic Opening
More informationValue of echocardiography in chronic dyspnea
Value of echocardiography in chronic dyspnea Jahrestagung Schweizerische Gesellschaft für /Schweizerische Gesellschaft für Pneumologie B. Kaufmann 16.06.2016 Chronic dyspnea Shortness of breath lasting
More informationAortic stenosis (AS) is common with the aging population.
New Insights Into the Progression of Aortic Stenosis Implications for Secondary Prevention Sanjeev Palta, MD; Anita M. Pai, MD; Kanwaljit S. Gill, MD; Ramdas G. Pai, MD Background The risk factors affecting
More informationProsthesis-Patient Mismatch after Mitral Valve Replacement: Comparison of Different Methods of Effective Orifice Area Calculation
Original Article Yonsei Med J 2016 Mar;57(2):328-336 pissn: 0513-5796 eissn: 1976-2437 Prosthesis-Patient Mismatch after Mitral Valve Replacement: Comparison of Different Methods of Effective Orifice Area
More informationAbileaflet mechanical valve is described by a nominal size, which
Surgery for Acquired Cardiovascular Disease Chambers et al Nominal size in six bileaflet mechanical aortic valves: A comparison of orifice size and biologic equivalence John B. Chambers, MD, FACC a Lionel
More informationThe Patient with Atrial Fibrilation
Assessment of Diastolic Function The Patient with Atrial Fibrilation Assoc. Prof. Adriana Ilieşiu, FESC University of Medicine Carol Davila Bucharest, Romania Associated Conditions with Atrial Fibrillation
More informationP = 4V 2. IVC Dimensions 10/20/2014. Comprehensive Hemodynamic Evaluation by Doppler Echocardiography. The Simplified Bernoulli Equation
Comprehensive Hemodynamic Evaluation by Doppler Echocardiography Itzhak Kronzon, MD North Shore LIJ/ Lenox Hill Hospital New York, NY Disclosure: Philips Healthcare St. Jude Medical The Simplified Bernoulli
More informationAortic Stenosis: Spectrum of Disease, Low Flow/Low Gradient and Variants
Aortic Stenosis: Spectrum of Disease, Low Flow/Low Gradient and Variants Martin G. Keane, MD, FASE Professor of Medicine Lewis Katz School of Medicine at Temple University Basic root structure Parasternal
More informationOutcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease
Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve
More informationAortic Stenosis and Perioperative Risk With Non-cardiac Surgery
Aortic Stenosis and Perioperative Risk With Non-cardiac Surgery Aortic stenosis (AS) is characterized as a high-risk index for cardiac complications during non-cardiac surgery. A critical analysis of old
More information2019 Qualified Clinical Data Registry (QCDR) Performance Measures
2019 Qualified Clinical Data Registry (QCDR) Performance Measures Description: This document contains the 18 performance measures approved by CMS for inclusion in the 2019 Qualified Clinical Data Registry
More informationPISA Evaluation of Mitral Regurgitation. Raymond Graber, MD Cardiac Anesthesia Group University Hospitals Case Medical Center 4/07/2011
PISA Evaluation of Mitral Regurgitation Raymond Graber, MD Cardiac Anesthesia Group University Hospitals Case Medical Center 4/07/2011 Introduction Evaluation of MR. What is PISA? Physiologic basis Issues
More informationSurgery for Valvular Heart Disease. Very Long-Term Survival and Durability of Mitral Valve Repair for Mitral Valve Prolapse
Surgery for Valvular Heart Disease Very Long-Term Survival and Durability of Mitral Valve Repair for Mitral Valve Prolapse Dania Mohty, MD; Thomas A. Orszulak, MD; Hartzell V. Schaff, MD; Jean-Francois
More informationAsymptomatic Severe Aortic Stenosis with Left Ventricular Dysfunction: Watchful Waiting or Valve Replacement?
CM&R Rapid Release. Published online ahead of print April 12, 2013 as Perspective Asymptomatic Severe Aortic Stenosis with Left Ventricular Dysfunction: Watchful Waiting or Valve Replacement? Larry E.
More informationIn a previous study of 113 patients with asymptomatic
Outcome of 622 Adults With Asymptomatic, Hemodynamically Significant Aortic Stenosis During Prolonged Follow-Up Patricia A. Pellikka, MD; Maurice E. Sarano, MD; Rick A. Nishimura, MD; Joseph F. Malouf,
More informationComprehensive Echo Assessment of Aortic Stenosis
Comprehensive Echo Assessment of Aortic Stenosis Smonporn Boonyaratavej, MD, MSc King Chulalongkorn Memorial Hospital Bangkok, Thailand Management of Valvular AS Medical and interventional approaches to
More informationThe correlation of AVA measured by transthoracic, transesophageal echocardiography and cardiac CT
The correlation of AVA measured by transthoracic, transesophageal echocardiography and cardiac CT R.Petr, H.Linkova, J.Knot, E.Paskova, J.Daniel, M.Labos Cardiocenter of University Hospital Kralovske Vinohrady
More informationInfluence of Atrial Fibrillation on Outcome Following Mitral Valve Repair
Influence of Atrial Fibrillation on Outcome Following Mitral Valve Repair Eric Lim, MBChB, MRCS; Clifford W. Barlow, DPhil, FRCS; A. Reza Hosseinpour, FRCS; Christopher Wisbey, BA; Kate Wilson, RN, BSc;
More informationHOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY?
HOW IMPORTANT ARE THESE ECHO MEASUREMENTS ANYWAY? John D. Carroll, MD Professor, Director of Interventional Cardiology and Co-Medical Director of the Cardiac and Vascular Center, University of Colorado
More informationECHO HAWAII. Role of Stress Echo in Valvular Heart Disease. Not only ischemia! Cardiomyopathy. Prosthetic Valve. Diastolic Dysfunction
Role of Stress Echo in Valvular Heart Disease ECHO HAWAII January 15 19, 2018 Kenya Kusunose, MD, PhD, FASE Tokushima University Hospital Japan Not only ischemia! Cardiomyopathy Prosthetic Valve Diastolic
More informationAffecting the elderly Requiring new approaches. Echocardiographic Evaluation of Hemodynamic Severity. Increasing prevalence Mostly degenerative
Echocardiographic Evaluation of Hemodynamic Severity Steven J. Lester MD, FACC, FRCP(C), FASE Mayo Clinic, Arizona Relevant Financial Relationship(s) None Off Label Usage None A re-emerging public-health
More informationRole of Transesophageal Echocardiography in the Diagnosis of Paradoxical Low Flow, Low Gradient Severe Aortic Stenosis
Original Article Print ISSN 1738-5520 On-line ISSN 1738-5555 Korean Circulation Journal Role of Transesophageal Echocardiography in the Diagnosis of Paradoxical Low Flow, Low Gradient Severe Aortic Stenosis
More informationDobutamine Stress testing In Low Flow, Low EF, Low Gradient Aortic Stenosis Case Studies
Dobutamine Stress testing In Low Flow, Low EF, Low Gradient Aortic Stenosis Case Studies Mitral Regurgitation The New ASE Guidelines: Role of 2D/3D and CMR William A. Zoghbi MD, FASE, MACC Professor and
More informationIncidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves
INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient
More informationEchocardiographic variables associated with mitral regurgitation after aortic valve replacement for aortic valve stenosis
The Egyptian Heart Journal (2013) 65, 135 139 Egyptian Society of Cardiology The Egyptian Heart Journal www.elsevier.com/locate/ehj www.sciencedirect.com ORIGINAL ARTICLE Echocardiographic variables associated
More informationChoose the grading of diastolic function in 82 yo woman
Question #1 Choose the grading of diastolic function in 82 yo woman E= 80 cm/s A= 70 cm/s LAVI < 34 ml/m 2 1= Grade 1 2= Grade 2 3= Grade 3 4= Normal 5= Indeterminate 2018 MFMER 3712003-1 Choose the grading
More informationThe impact of prosthesis patient mismatch after aortic valve replacement varies according to age at operation
Editor s choice Scan to access more free content 1 Division of Cardiac Surgery, The Johns Hopkins Hospital, Baltimore, Maryland, USA 2 Division of Cardiac Surgery, University of Ottawa Heart Institute,
More informationEffect of Patient-Prosthesis Mismatch in Aortic Position on Late-Onset Tricuspid Regurgitation and Clinical Outcomes after Double Valve Replacement
Original Article Yonsei Med J 17 Sep;58(5):968-974 pissn: 0513-5796 eissn: 1976-2437 Effect of Patient-Prosthesis Mismatch in Aortic Position on Late-Onset Tricuspid Regurgitation and Clinical Outcomes
More informationECHOCARDIOGRAPHY. Patient Care. Goals and Objectives PF EF MF LF Aspirational
Patient Care Be able to: Perform and interpret basic TTE and X cardiac Doppler examinations Perform and interpret a comprehensive X TTE and cardiac Doppler examination Perform and interpret a comprehensive
More informationLeft atrial function. Aliakbar Arvandi MD
In the clinic Left atrial function Abstract The left atrium (LA) is a left posterior cardiac chamber which is located adjacent to the esophagus. It is separated from the right atrium by the inter-atrial
More informationTSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD
TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)
More informationWe present the case of an asymptomatic, 75-year-old
Images in Cardiovascular Medicine Asymptomatic Rupture of the Left Ventricle Lech Paluszkiewicz, MD; Stefan Ożegowski, MD; Mohammad Amin Parsa, MD; Jan Gummert, PhD, MD We present the case of an asymptomatic,
More informationIndex. K Knobology, TTE artifact, image resolution, ultrasound, 14
A Acute aortic regurgitation (AR), 124 128 Acute aortic syndrome (AAS) classic aortic dissection diagnosis, 251 263 evolutive patterns, 253 255 pathology, 250 251 classifications, 247 248 incomplete aortic
More informationNouvelles indications/ Nouvelles valves
Nouvelles indications/ Nouvelles valves Sadra Lotus Valve system Boston Scientific A. Tirouvanziam Institut Thorax Nantes SH-31314-AC Sept 2012 Slide 1 of 53 SH-31314-AC Sept 2012 Slide 2 of 53 Building
More informationQuantification of Mitral Stenosis: Planimetry, pressure Half time, Continuity Common Errors
Quantification of Mitral Stenosis: Planimetry, pressure Half time, Continuity Common Errors Christopher J Kramer RDCS Advanced Cardiovascular Services Aurora Health Care Milwaukee, WI No Disclosures Baumgartner,
More informationNikos Kouris a, *, Ignatios Ikonomidis b, Dimitra Kontogianni a, Peter Smith b, Petros Nihoyannopoulos b
Eur J Echocardiography (2005) 6, 435e442 Mitral valve repair versus replacement for isolated non-ischemic mitral regurgitation in patients with preoperative left ventricular dysfunction. A long-term follow-up
More informationOutline. EuroScore II. Society of Thoracic Surgeons Score. EuroScore II
SURGICAL RISK IN VALVULAR HEART DISEASE: WHAT 2D AND 3D ECHO CAN TELL YOU AND WHAT THEY CAN'T Ernesto E Salcedo, MD Professor of Medicine University of Colorado School of Medicine Director of Echocardiography
More informationSurvival After Aortic Valve Replacement for Severe Aortic Stenosis With Low Transvalvular Gradients and Severe Left Ventricular Dysfunction
Journal of the American College of Cardiology Vol. 39, No. 8, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)01759-X
More informationEvaluation of Left Ventricular Diastolic Dysfunction by Doppler and 2D Speckle-tracking Imaging in Patients with Primary Pulmonary Hypertension
ESC Congress 2011.No 85975 Evaluation of Left Ventricular Diastolic Dysfunction by Doppler and 2D Speckle-tracking Imaging in Patients with Primary Pulmonary Hypertension Second Department of Internal
More informationAdult Echocardiography Examination Content Outline
Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,
More informationAsymptomatic Valvular Disease:
Asymptomatic Valvular Disease: Can Echocardiography Help You Decide When to Intervene? Neil J. Weissman, MD MedStar Health Research Inst at MedStar Washington Hospital Center & Professor of Medicine Georgetown
More informationExercise Pulmonary Hypertension predicts the Occurrence of Symptoms in Asymptomatic Degenerative Mitral Regurgitation
Exercise Pulmonary Hypertension predicts the Occurrence of Symptoms in Asymptomatic Degenerative Mitral Regurgitation Julien Magne, PhD, Kim O Connor, MD, Giuseppe Romano, MD, Marie Moonen, MD, Luc A.
More informationRotation: Echocardiography: Transthoracic Echocardiography (TTE)
Rotation: Echocardiography: Transthoracic Echocardiography (TTE) Rotation Format and Responsibilities: Fellows rotate in the echocardiography laboratory in each clinical year. Rotations during the first
More informationSubacute left ventricular outflow tract obstruction after transapical closure of a mitral paravalvular leak in the region of the aortomitral curtain
DOI: 10.1111/echo.13357 CASE REPORT Subacute left ventricular outflow tract obstruction after transapical closure of a mitral paravalvular leak in the region of the aortomitral curtain Sathish Chikkabyrappa
More informationCharacteristic Doppler Echocardiographic Pattern of Mitral Inflow Velocity in Severe Aortic Regurgitation
1712 JACC Vol. 14, No. 7 December 1989: 1712-7 Characteristic Doppler Echocardiographic Pattern of Mitral Inflow Velocity in Severe Aortic Regurgitation JAE K. OH, MD, FACC, LIV K. HATLE, MD, LAWRENCE
More informationNew Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology
New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.
More informationRestoration of Sinus Rhythm by the Maze Procedure Halts Progression of Tricuspid Regurgitation After Mitral Surgery
Restoration of Sinus Rhythm by the Maze Procedure Halts Progression of Tricuspid Regurgitation After Mitral Surgery John M. Stulak, MD,* Hartzell V. Schaff, MD, Joseph A. Dearani, MD, Thomas A. Orszulak,
More informationCase Reviews: Hemodynamic Calculations in Valvular Regurgitation
Case Reviews: Hemodynamic Calculations in Valvular Regurgitation Case 5 History: 69-year-old man with orthotopic heart transplant 15 years ago. Inferior MI several years ago. Recurrent CHF. Currently dyspneic
More informationEffect of Heart Rate on Tissue Doppler Measures of E/E
Cardiology Department of Bangkok Metropolitan Administration Medical College and Vajira Hospital, Bangkok, Thailand Abstract Background: Our aim was to study the independent effect of heart rate (HR) on
More informationDevelopment of tricuspid regurgitation late after left-sided valve surgery: A single-center experience with long-term echocardiographic examinations
Imaging and Diagnostic Testing Development of tricuspid regurgitation late after left-sided valve surgery: A single-center experience with long-term echocardiographic examinations Jae-Jin Kwak, MD, a Yong-Jin
More informationTransvalvular pressure gradients (TPG) and valve effective
Assessment of Aortic Valve Stenosis Severity A New Index Based on the Energy Loss Concept Damien Garcia, BEng; Philippe Pibarot, DVM, PhD; Jean G. Dumesnil, MD; Frédéric Sakr, BEng; Louis-Gilles Durand,
More informationPrimary Mitral Regurgitation
EURO VALVE Madrid News from Valves Guidelines 2012: What s new and Why? Primary Mitral Regurgitation Luc A. Pierard, MD, PhD Professor of Medicine Head of the Department of Cardiology Heart Valve Clinic,
More informationA Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision
A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction
More informationAortic Valve Stenosis: When stress TTE and/or TEE is required to make the diagnosis and guide treatment
Aortic Valve Stenosis: When stress TTE and/or TEE is required to make the diagnosis and guide treatment Stefanos Karagiannis MD PhD Cardiologist Director Echocardiology Dpt Athens Medical Center ESC 2017
More informationMitral Valve prolapse: What s new? Which indications of early surgery? Input of new 2017 ESC/EACTS guidelines. Christophe Tribouilloy Amiens, France
Mitral Valve prolapse: What s new? Which indications of early surgery? Input of new 2017 ESC/EACTS guidelines Christophe Tribouilloy Amiens, France I have no financial relationships to disclose related
More informationCongestive Heart Failure or Heart Failure
Congestive Heart Failure or Heart Failure Dr Hitesh Patel Ascot Cardiology Group Heart Failure Workshop April, 2014 Question One What is the difference between congestive heart failure and heart failure?
More informationAssessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler. and dobutamine stress echocardiography
Assessment of the St. Jude Medical Regent Prosthetic Valve by Continuous-Wave Doppler and Dobutamine Stress Echocardiography Akira Sezai, MD, PhD, Yuji Kasamaki, MD, PhD, Keisuke Abe, RMS, Mitsumasa Hata,
More informationEcho Doppler Assessment of Right and Left Ventricular Hemodynamics.
Echo Doppler Assessment of Right and Left Ventricular Hemodynamics. Itzhak Kronzon, MD, FASE, FACC, FESC, FAHA, FACP, FCCP Northwell, Lenox Hill Hospital, New York Professor of Cardiology Hofstra University
More informationAssessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography
American Journal of Hematology 81:563 567 (2006) Assessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography Y.J. Zeng, 1,4 * S.W. Xu,
More informationOrganic mitral regurgitation
The best in heart valve disease Organic mitral regurgitation Ewa Szymczyk Department of Cardiology Medical University of Lodz, Poland I have nothing to declare Organic mitral regurgitation leaflet abnormality
More informationFailure of Guideline Adherence for Intervention in Patients With Severe Mitral Regurgitation
Journal of the American College of Cardiology Vol. 54, No. 9, 2009 2009 by the American College of Cardiology Foundation ISSN 0735-1097/09/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2009.03.079
More informationJournal of the American College of Cardiology Vol. 34, No. 4, by the American College of Cardiology ISSN /99/$20.
Journal of the American College of Cardiology Vol. 34, No. 4, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00341-1 Changes
More informationStress Testing in Valvular Disease
2017 ASE Florida Orlando, FL October 10, 2017 2:40 2:50 PM 10 min Grand Harbor Ballroom South Stress Testing in Valvular Disease Muhamed Sarić MD, PhD, MPA Director of Noninvasive Cardiology Echo Lab Associate
More informationAortic Regurgitation and Aortic Aneurysm - Epidemiology and Guidelines -
Reconstruction of the Aortic Valve and Root - A Practical Approach - Aortic Regurgitation and Aortic Aneurysm Wednesday 14 th September - 9.45 Practice must always be founded on sound theory. Leonardo
More informationJournal of the American College of Cardiology Vol. 34, No. 5, by the American College of Cardiology ISSN /99/$20.
Journal of the American College of Cardiology Vol. 34, No. 5, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00360-5 Hemodynamic
More information