IntraVascular UltraSound

Size: px
Start display at page:

Download "IntraVascular UltraSound"

Transcription

1 IntraVascular UltraSound Anton FW van der Steen, Johannes A Schaar, David Goertz, Martijn Frijlink, Nico de Jong Patrick W Serruys and. Nicolaas Bom, Thoraxcentre; Erasmus Medical Center Rotterdam, The Netherlands Interuniversity Cardiology Institute of the Netherlands Introduction Currently, IVUS is, beside OCT the only clinically available technique, capable of providing real time cross-sectional images in vivo. It delivers information that is not available from X-ray angiography. Since the number of available interventional techniques for treatment of atherosclerotic luminal narrowing increases, specific diagnostic information that can aid in the selection becomes increasingly important. For this reason, IVUS is more and more routinely used for guiding interventional procedures and for studying the mechanisms for restenosis i. In the last couple of years the possibilities to access a sufficient amount of RF data at adequate speed have improved significantly. This has given the possibility to study IVUS data in a totally different way. In the traditional way the emphasis was always on imaging itself while RF-analysis also allows tissue characterization, elasticity imaging, flow estimation and enhanced boundary detection. Construction of an IVUS catheter Rotating element In Figure 1, the rotating tip system is schematically illustrated. The driving motor is connected to the proximal end of the catheter. The shaft (1) must be very flexible and contains the electric wires for the transducer. The shaft needs to be flexible; yet it must drive the tip in a predictable way. This is not always possible when the catheter must follow a tortuous path. As a result, the beam deflection on the display may not correspond with the acoustic beam deflection. This causes errors in the image. Capacitive and optical feed-back techniques have been suggested to cope with this problem. The element (3) is positioned in such a way that no transmission pulse effect appears on the display, since the echo travel time to the dome (2) is of sufficient duration. It allows imaging very close to the catheter outer wall since no 'dead zone' is present. The dome must be acoustically transparent. Results obtained in-vitro are often reported without such a dome, thus increasing image quality and avoiding the liquid filling process necessary for acoustic coupling. Figure 1. - Flex-shaft rotating element principle. This contains a rotating shaft (1); acoustically transparent dome (2); and echo element (3) in a catheter tip. The flex shaft is powered by a proximal motor outside the catheter. In some practical applications an external guide wire is added and can be inserted through an added catheter nose part. 1

2 Figure 2. - The rotating mirror principle. This contains the flex shaft (1); The transparent dome (2); the acoustic element (3) and the rotating mirror (4). Rotating mirror The rotating mirror technique is similar to the previously described method. Schematically this method is shown in Figure 2. The flexible shaft (1); the transparent dome (2); the echotransducer (3) are complemented with a mirror. The mirror creates an even shorter dead zone outside the catheter due to the longer acoustic pathway inside the catheter. The non-moving transducer avoids the necessity of rotating electric wires. Acoustic lenses and focussing shapes of the mirror have been described. Electronically switched phased array The catheter shown in Figure 3 contains many small acoustic elements (2) which are positioned cylindrically around the catheter tip. The number of elements may be any practical number such as 16, 32, 64 or 128. The tip may contain an electronic component to reduce the number of electric wires. Figure 3. - Electronically switched phased array catheter tip with integrated circuitry for reduction of the number of wires (1); the elements (2) and a guide wire (3). The construction allows for the introduction of a central guidewire (3). A first prototype, using the same principle but meant for intracardiac imaging was already described in 1972 ii. By introduction of time delays subgroups of elements may together form a "single larger echo transducer". This process can be repeated with any other subgroup. This allows aperture variation and electronic focussing methods. On the other hand, acoustic element geometry is not optimal and a near-field dead zone may exist. In a second principle, all elements are used individually for transmission and reception. Beam forming is created with computer algorithms using a Synthetic Aperture System (SAS). A summary of today s IVUS methods is illustrated in Figure 4. Mostly, the intravascular procedure is applied for decision-making when angiographic data are less conclusive. For obvious reasons, there is a strong urge to combine see and do in interventional procedures. This leads to catheters in which, for instance, an angioplasty balloon is combined with ultrasonic imaging in or close to the balloon. Another combination may provide guidance during stent implantation (Figure 4d). 2

3 Figure 4. - Overview of presently most frequently used IVUS catheters. A) flex shaft rotating element system. B) Identical to A with added guide wire capabilities. C) Phased array system with guide wire and multiplexing electronics. D) catheter combining imaging with stent delivery capabilities. Since intravascular imaging provides accurate geometrical information within the cross-section, combination with other interventional procedures is likely to expand in future. IVUS in the clinic Intravascular coronary ultrasound (IVUS) provides real-time high-resolution images of the vessel wall and lumen. iii The size of IVUS catheters is between 2.9 to 3.5 French. Depending on the distance from the catheter the axial resolution is about 150 microns, the lateral 300 microns. The images appear real time at a frequency of up to 30 frames/sec. Features of the vessel can be detected based on the echogenicity and the thickness of the material. Small structures can be visualized, however only those sized over 160 microns can be estimated accurately. The normal thickness of the media is about μm. IVUS provides some insight into the composition of coronary plaques, among them the vulnerable plaque. Rupture of vulnerable plaques is the main cause of acute coronary syndrome and myocardial infarction. Identification of vulnerable plaque is therefore essential to enable the development of treatment modalities to stabilize such plaque. Because myocardial infarction and its consequences are so important, we must investigate options to identify those areas that will be responsible for future events. A wide variety exists in the stability of coronary atherosclerotic plaques. A plaque may be stable for years, however abrupt disruption of his structure is the main cause of acute coronary syndrome. The susceptibility of plaques to rupture is known to be related to their composition, stress distribution and degree of inflammation. Increasingly it is becoming recognized that, while not fully understood, two factors play a major role in plaque progression and vulnerability: the presence of increased plaque invasion by microvessels (vasa vasorum) and the expression of specific molecules by cells within the plaques. Both issues can be assessed with an ultrasound contrast agent (An ultrasound contrast agent contains small bubbles (~ 3µm), which are proven to be blood tracers) using special signatures of these bubbles under ultrasound insonification. The vulnerable plaque contains certain features that could be diagnosed by various specialized methods. The ideal technique would provide morphological, mechanical and chemical information, however at present, no diagnostic modality providing such all-embracing assessment is available. 3

4 Elastography and Palpography In 1991, a new technique was introduced to measure the mechanical properties of tissue using ultrasound: elastography. iv The underlying concept is that upon uniform loading, the local relative amount of deformation (strain) of a tissue is related to the local mechanical properties of that tissue. If we apply this concept to determine the local properties of arterial tissue, blood pressure acts as a stressor. At a given pressure difference, soft plaque components will deform more than hard components. Measurement of local plaque deformation in the radial direction can be obtained with ultrasound. For intravascular purposes, a derivate of elastography called palpography may be a suitable tool. v In this approach, one strain value per angle is determined and plotted as a colorcoded contour at the lumen vessel boundary. Since radial strain is obtained, the technique may have the potential to detect regions with elevated stress: increased circumferential stress results in an increased radial deformation of the plaque components. In vitro studies with histological confirmation have shown that there are differences of strain normalized to pressure between fibrous, fibro-fatty and fatty components of the plaque of coronary as well as femoral arteries. vi This difference was mainly evident between fibrous and fatty tissue. The plaque types could not be differentiated by echo-intensity differences on the IVUS echogram. The principal of elastography is as follows: Two IVUS images are made with a short time interval. Due to cardiac pressure there will be a small difference in intraluminal pressure. In case of little difference of pressure the pictures will be almost identical. However, thorough analysis of the ultrasound RF data tells a different story. Proper data processing will allow visualisation of the pressure caused differences. This may be called an elastogram where a colour coding is used for identification of the softness. A principle is illustrated in figure 5. Figure 5. - IVUS elastography: two IVUS images are acquired at two slightly different intraluminal pressures. A deformation image is obtained by data processing. This image displays soft regions in green. Palpography is a further development that is simulating putting your finger inside the vessel and palpating it to feel how hard it is. It is a new name for elastography in only a thin internal layer of the artery. Palpography reveals information that is not seen on IVUS. To differentiate between hard and soft tissue may be important for the detection of an instable plaque that is prone to rupture. Since palpography is based on clinically available IVUS catheters, the technique can be easily introduced into the catheterization laboratory. By acquiring data at the end of the filling phase, when catheter motion is minimal, the quality and reliability of the palpogram is increased. The clinical value of this technique is currently under investigation. 4

5 It is feasible to apply intravascular palpography during interventional catheterization procedures. In a recent study, data were acquired in patients (n=12) during PTCA procedures with echo apparatus equipped with radiofrequency output. The systemic pressure was used to strain the tissue. Significantly higher strain values were found in non-calcified plaques than in calcified plaques. vii Another in vivo validation study in atherosclerotic Yucatan pigs showed that fatty plaques have an increased mean strain value. High-strain spots were also associated with the presence of macrophages, a further feature of vulnerable plaques. viii Figure 6. - Mode vibrations for a bubble with a diameter of 4 µm. Here a mode n = 3 is observed, however a range of modes is observed for a transmit frequency of 3.5 Mhz. Contrast imaging Unfortunately, in smaller vessels and capillaries blood detection is not possible due to low signal strengths from blood, tissue motion effects, and limited resolution (~ mm). Ultrasound contrast agents (UCA, consisting of small (encapsulated) gas bubbles), will increase the reflection of ultrasound by the blood pool, after intravenously administer, and by that make it possible to provide perfusion images and imaging the vascularity in plaques. When a gas bubble in liquid is subjected to pressure variations induced by an acoustic wave it acts as a forced, damped, oscillator, with a resonant frequency that is inversely related to its diameter. If the incident acoustic wave is of sufficient amplitude and of a frequency near to the resonant frequency of the bubble, the induced radial oscillations will become more pronounced and track the incident pressure wave in a nonlinear manner. As a result, re-radiated acoustic energy will be coupled into other frequency bands. The coupling of acoustic energy into the second harmonic frequency region occurs most efficiently when a bubble is insonated at or near its resonant frequency. Subharmonics (at half the insonation frequency) can be efficiently generated at twice the natural resonant frequency. Optical observation of the vibrating bubble with a fast frame camera and acoustic measurements with high frequency ultrasound reveal these nonlinear vibrations. The fast-frame camera Brandaris 128, developed at the Erasmus medical center in collaboration with Twente University, can be used to visualize microbubble agent behavior at high ultrasound frequencies. Designed to investigate microbubble dynamics at conventional ultrasound frequencies, it has a unique capablilty to image bubble dynamics at a frame rate of up to 25 Mega frames-per-second for 128 consecutive frames. For bubble sizes of 4 micrometer figure 6 shows both radial oscillations and surface modes. In this figure only 5 frames from the 128 available frames are plotted, with an interframe time of 100 ns. In vivo Experiments In ultrasound contrast imaging, these nonlinear emissions are the foundation of specific microbubble detection schemes, which are essential for detecting small vessel In vivo. Such In vivo studies were conducted in atherosclerotic rabbit abdominal aortas. Atherosclerosis was initiated using endothelial cell injury procedures followed by a high cholesterol diet ix. Experiments were performed weeks after initiation of atherosclerosis. Example in vivo results are shown in fig. 7 for fundamental and harmonic imaging modes. In fundamental mode (top) the injected contrast can hardly be discriminated from the tissue. In the special contrast mode (harmonic imaging) results 5

6 are substantially different. Post-injection, agent was first visualized in the main lumen. By approximately 5 seconds post injection a ring at the boundary of the main lumen could be seen, which can be attributed to the presence of a small amount of more slowly moving agent adjacent to the aortic wall. This effect enabled the lumen boundary to be easily distinguished. There is then an eccentric circumferential region devoid of enhancement, with a thickness larger in the 2 o clock direction than in the 8 o clock direction. Outside this hypoechogenic region, numerous locations of enhancement are observed that are associated with the presence of agent. While towards 10 o clock the enhancement is associated with the vena cava, the other locations are consistent with the detection of microvessels outside the main vessel lumen x. Figure 7. - In vivo results in an atherosclerotic rabbit aorta using decanted Definity TM. Left: normal imaging. Right: 10 second post-injection harmonic mode shows significant adventitial enhancement, consistent with the detection of adventitial microvessels. Scale of images is 12 mm across. The dynamic range of the fundamental and harmonic images are 40 and 25 db respectively. References: i G. S. Mintz, J. J. Popma, K. M. Kent, L. F. Satler, S. C. Wong, M. K. Hong, J. A. Kovach, and M. B. Leon, "Arterial remodeling after coronary angioplasty. A serial intravascular ultrasound study," Circulation, vol. 94, pp , ii Bom N, Lancée CT, Van Egmond FC (1972) An ultrasonic intracardiac scanner. Ultrasonics 10: 72- iii Bom N, Li W, van der Steen AF, Lancee CT, Cespedes EI, Slager CJ, de Korte CL. Intravascular imaging. Ultrasonics 1998; 36: iv Ophir J, Cespedes I, Ponnekanti H, et al. Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging 1991; 13: v Doyley MM, Mastik F, de Korte CL, et al. Advancing intravascular ultrasonic palpation toward clinical applications. Ultrasound Med Biol 2001; 27: vi de Korte CL, Pasterkamp G, van der Steen AF, et al. Characterization of plaque components with intravascular ultrasound elastography in human femoral and coronary arteries in vitro. Circulation 2000; 102: vii de Korte CL, Carlier SG, Mastik F, et al. Morphological and mechanical information of coronary arteries obtained with intravascular elastography; feasibility study in vivo. Eur Heart J 2002; 23: viii de Korte CL, Sierevogel MJ, Mastik F, Strijder C, et al. Identification of atherosclerotic plaque components with intravascular ultrasound elastography in vivo: a Yucatan pig study. Circulation 2002; 105: ix J. A. Schaar, C. L. de Korte, F. Mastik, L. C. van Damme, R. Krams, P. W. Serruys, and A. F. W. van der Steen. Three-dimensional palpography of human coronary arteries. Herz, 30(2): , x Goertz D. E., Frijlink M. E., Tempel D., Van Damme L. C. A., Krams R., Schaar J. A., Ten Cate F. J., Serruys P. W., De Jong N. and Van der Steen A. F. W. Contrast Harmonic Intravascular Ultrasound: a Feasibility Study for Vasa Vasorum Imaging Invest. Radiol 2006;41:

DISens symposium-book

DISens symposium-book 2.7 Ultra Sound Sensors 3D Intravascular Ultrasound Palpography for Vulnerable Plaque Detection Anton F. W. van der Steen, Chris L. de Korte, Johannes A. Schaar, Frits Mastik, Radj A. Baldewsing, Patrick

More information

Invasive Coronary Imaging Modalities for Vulnerable Plaque Detection

Invasive Coronary Imaging Modalities for Vulnerable Plaque Detection Invasive Coronary Imaging Modalities for Vulnerable Plaque Detection Gary S. Mintz, MD Cardiovascular Research Foundation New York, NY Greyscale IVUS studies have shown Plaque ruptures do not occur randomly

More information

Imaging Atheroma The quest for the Vulnerable Plaque

Imaging Atheroma The quest for the Vulnerable Plaque Imaging Atheroma The quest for the Vulnerable Plaque P.J. de Feijter 1. Department of Cardiology 2. Department of Radiology Coronary Heart Disease Remains the Leading Cause of Death in the U.S, Causing

More information

Characterization of plaque components and vulnerability with intravascular ultrasound elastography

Characterization of plaque components and vulnerability with intravascular ultrasound elastography Phys. Med. Biol. 45 (2000) 1465 1475. Printed in the UK PII: S0031-9155(00)06797-X Characterization of plaque components and vulnerability with intravascular ultrasound elastography Chris L de Korte, Anton

More information

IVUS Virtual Histology. Listening through Walls D. Geoffrey Vince, PhD The Cleveland Clinic Foundation

IVUS Virtual Histology. Listening through Walls D. Geoffrey Vince, PhD The Cleveland Clinic Foundation IVUS Virtual Histology Listening through Walls D. Geoffrey Vince, PhD Disclosure VH is licenced to Volcano Therapeutics Grant funding from Pfizer, Inc. Grant funding from Boston-Scientific Most Myocardial

More information

Subscriptions: Information about subscribing to Circulation is online at

Subscriptions: Information about subscribing to Circulation is online at Characterization of Plaque Components With Intravascular Ultrasound Elastography in Human Femoral and Coronary Arteries In Vitro Chris L. de Korte, Gerard Pasterkamp, Anton F. W. van der Steen, Hein A.

More information

Optical Coherence Tomography

Optical Coherence Tomography Optical Coherence Tomography Disclosure Information Demetrius Lopes MD The following relationships exist related to this presentation: University Grant/Research Support: Rush University Industry Grant

More information

Added Value of Invasive Coronary Imaging for Plaque Rupture and Erosion

Added Value of Invasive Coronary Imaging for Plaque Rupture and Erosion Assessment of Coronary Plaque Rupture and Erosion Added Value of Invasive Coronary Imaging for Plaque Rupture and Erosion Yukio Ozaki, MD, PhD, FACC, FESC Cardiology Dept., Fujita Health Univ. Toyoake,

More information

Chapter 43 Noninvasive Coronary Plaque Imaging

Chapter 43 Noninvasive Coronary Plaque Imaging hapter 43 Noninvasive oronary Plaque Imaging NIRUDH KOHLI The goal of coronary imaging is to define the extent of luminal narrowing as well as composition of an atherosclerotic plaque to facilitate appropriate

More information

Effectiveness of IVUS in Complex Cases

Effectiveness of IVUS in Complex Cases Effectiveness of IVUS in Complex Cases Satoru Sumituji,M.D. Rinku General Medical Center IVUS is can provide images of the vessel wall and the tissue around the vessel which cannot be viewed by angiography.

More information

Imaging Overview for Vulnerable Plaque: Data from IVUS Trial and An Introduction to VH-IVUS Imgaging

Imaging Overview for Vulnerable Plaque: Data from IVUS Trial and An Introduction to VH-IVUS Imgaging Imaging Overview for Vulnerable Plaque: Data from IVUS Trial and An Introduction to VH-IVUS Imgaging Gary S. Mintz,, MD Cardiovascular Research Foundation New York, NY Today, in reality, almost everything

More information

Original Contribution

Original Contribution doi:10.1016/j.ultrasmedbio.2007.01.018 Ultrasound in Med. & Biol., Vol. 33, No. 8, pp. 1199 1205, 2007 Copyright 2007 World Federation for Ultrasound in Medicine & Biology Printed in the USA. All rights

More information

Introduction What Causes Peripheral Vascular Disease? How Do Doctors Treat Peripheral Vascular Disease?... 9

Introduction What Causes Peripheral Vascular Disease? How Do Doctors Treat Peripheral Vascular Disease?... 9 Patient Information Table of Contents Introduction... 3 What is Peripheral Vascular Disease?... 5 What Are Some of the Symptoms of Peripheral Vascular Disease?... 7 What Causes Peripheral Vascular Disease?...

More information

Anatomy is Destiny, But Physiology is Here Today

Anatomy is Destiny, But Physiology is Here Today Published on Journal of Invasive Cardiology (http://www.invasivecardiology.com) September, 2010 [1] Anatomy is Destiny, But Physiology is Here Today Thu, 9/9/10-10:54am 0 Comments Section: Commentary Issue

More information

1st Department of Cardiology, University of Athens, Hippokration Hospital, Athens, Greece

1st Department of Cardiology, University of Athens, Hippokration Hospital, Athens, Greece Konstantinos Toutouzas, Maria Riga, Antonios Karanasos, Eleftherios Tsiamis, Andreas Synetos, Maria Drakopoulou, Chrysoula Patsa, Georgia Triantafyllou, Aris Androulakis, Christodoulos Stefanadis 1st Department

More information

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016 Ultrasound Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Image Generation Echography A-Mode B-Mode M-Mode 2.5D Ultrasound 3D Ultrasound 4D Ultrasound

More information

Cover Page. Author: Wang, Ancong Title: Automatic quantification of intravascular optical coherence tomography Issue Date:

Cover Page. Author: Wang, Ancong Title: Automatic quantification of intravascular optical coherence tomography Issue Date: Cover Page The handle http://hdl.handle.net/1887/29690 holds various files of this Leiden University dissertation Author: Wang, Ancong Title: Automatic quantification of intravascular optical coherence

More information

Coronary angiography and PCI

Coronary angiography and PCI Coronary arteries Coronary angiography and PCI Samo Granda, Franjo Naji Department of Cardiology Clinical department of internal medicine University clinical centre Maribor Coronary arteries Atherosclerosis

More information

Your heart is a muscular pump about the size of your fist, located

Your heart is a muscular pump about the size of your fist, located How Your Heart Works Your heart is a muscular pump about the size of your fist, located slightly to the left and behind your breastbone. Its function is to pump blood throughout your body. As your heart

More information

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of

More information

Introduction 3. What is Peripheral Vascular Disease? 5. What Are Some of the Symptoms of Peripheral Vascular Disease? 6

Introduction 3. What is Peripheral Vascular Disease? 5. What Are Some of the Symptoms of Peripheral Vascular Disease? 6 Patient Information Table of Contents Introduction 3 What is Peripheral Vascular Disease? 5 What Are Some of the Symptoms of Peripheral Vascular Disease? 6 What Causes Peripheral Vascular Disease? 7 How

More information

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS Jeffrey P. Milligan, Daniel T. Peters, Structural Integrity Associates, Inc., USA Many advances in Non-Destructive Examination

More information

Coronary Artery Thermography

Coronary Artery Thermography Coronary Artery Thermography The 10th Anniversary, Interventional Vascular Therapeutics Angioplasty Summit 2005 TCT Asia Pacific Christodoulos Stefanadis Professor of Cardiology Athens Medical School In

More information

Intravascular Ultrasound

Intravascular Ultrasound May 2008 Beth Israel Deaconess Medical Center Harvard Medical School Intravascular Ultrasound Matthew Altman, HMS III Gillian Lieberman, MD BIDMC Department of Radiology Presentation Overview 1. Patient

More information

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA

CT Imaging of Atherosclerotic Plaque. William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA CT Imaging of Atherosclerotic Plaque William Stanford MD Professor-Emeritus Radiology University of Iowa College of Medicine Iowa City, IA PREVALENCE OF CARDIOVASCULAR DISEASE In 2006 there were 80 million

More information

IVUS Analysis. Myeong-Ki. Hong, MD, PhD. Cardiac Center, Asan Medical Center University of Ulsan College of Medicine, Seoul, Korea

IVUS Analysis. Myeong-Ki. Hong, MD, PhD. Cardiac Center, Asan Medical Center University of Ulsan College of Medicine, Seoul, Korea IVUS Analysis Myeong-Ki Hong, MD, PhD Cardiac Center, Asan Medical Center University of Ulsan College of Medicine, Seoul, Korea Intimal disease (plaque) is dense and will appear white Media is made of

More information

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),

More information

Optimal assessment observation of intravascular ultrasound

Optimal assessment observation of intravascular ultrasound Optimal assessment observation of intravascular ultrasound Katsutoshi Kawamura and Atsunori Okamura Division of Radiology Cardiovascular Center Sakurabashi Watanabe Hospital SAKURABASHI WATANABE Hospital

More information

Underwater Acoustic Measurements in Megahertz Frequency Range.

Underwater Acoustic Measurements in Megahertz Frequency Range. Underwater Acoustic Measurements in Megahertz Frequency Range. Current State and Prospects of Development in Russia Alexander M. Enyakov,, Many medical applications of underwater acoustic measurements

More information

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal

More information

DISRUPT CAD. Todd J. Brinton, MD Clinical Associate Professor of Medicine Adjunct Professor of Bioengineering Stanford University

DISRUPT CAD. Todd J. Brinton, MD Clinical Associate Professor of Medicine Adjunct Professor of Bioengineering Stanford University DISRUPT CAD A multicenter, prospective, single-arm study of percutaneous Lithoplasty prior to stent implantation in heavily calcified coronary lesions Todd J. Brinton, MD Clinical Associate Professor of

More information

Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013

Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013 Evaluation of Intermediate Coronary lesions: Can You Handle the Pressure? Jeffrey A Southard, MD May 4, 2013 Disclosures Consultant- St Jude Medical Boston Scientific Speaker- Volcano Corporation Heart

More information

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics The Physics of Ultrasound Pipe Organ 10-8000 Emission Dog 452-1080 Man 85-1100 Spectrum Bat 10,000-120,000 Porpoise 7000-120,000 Claus G. Roehrborn Professor and Chairman 10 20 Cycles per second Reception

More information

Detection of carotid plaque neovascularization with Superb Micro-Vascular Imaging

Detection of carotid plaque neovascularization with Superb Micro-Vascular Imaging Detection of carotid plaque neovascularization with Superb Micro-Vascular Imaging Yong Qiang Professor Ultrasonic Department, Beijing An-Zhen Hospital, Capital Medical University, China 1. Background Conventional

More information

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System January 2007 Objectives: After completion of this module, the participant will be able to accomplish the following: Understand

More information

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging Original Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging P. Khamdaengyodtai 1, T. Khamdaeng 1, P. Sakulchangsatjatai 1, N. Kammuang-lue

More information

Report originating from the Frauenhofer-Institute of Construction Physics

Report originating from the Frauenhofer-Institute of Construction Physics IBP Report originating from the Frauenhofer-Institute of Construction Physics Abridged version from IBP-Report BS 113/84 An examination of the principle of operation behind the sonic wave device SK1 and

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

CLINICAL APPLICATIONS OF OPTICAL COHERENCE TOMOGRAPHY. Konstantina P. Bouki, FESC 2 nd Department of Cardiology General Hospital Of Nikea, Pireaus

CLINICAL APPLICATIONS OF OPTICAL COHERENCE TOMOGRAPHY. Konstantina P. Bouki, FESC 2 nd Department of Cardiology General Hospital Of Nikea, Pireaus CLINICAL APPLICATIONS OF OPTICAL COHERENCE TOMOGRAPHY Konstantina P. Bouki, FESC 2 nd Department of Cardiology General Hospital Of Nikea, Pireaus OPTICAL COHERENCE TOMOGRAPHY (OCT) IVUS and OCT IVUS OCT

More information

High resolution ultrasound scanner for skin imaging

High resolution ultrasound scanner for skin imaging High resolution ultrasound scanner for skin imaging Christine Turlat Sales Director Atys medical 17 Parc d Arbora 69510 SOUCIEU EN JARREST Atys company Principle of ultrasound imaging DERMCUP Normal image

More information

Culprit Lesion Remodeling and Long-term (> 5years) Prognosis in Patients with Acute Coronary Syndrome

Culprit Lesion Remodeling and Long-term (> 5years) Prognosis in Patients with Acute Coronary Syndrome Culprit Lesion Remodeling and Long-term (> 5years) Prognosis in Patients with Acute Coronary Syndrome Hiroyuki Okura*, MD; Nobuya Matsushita**,MD Kenji Shimeno**, MD; Hiroyuki Yamaghishi**, MD Iku Toda**,

More information

Coronary Artery Calcium Score

Coronary Artery Calcium Score Coronary Artery Calcium Score August 19, 2014 by Axel F. Sigurdsson MD 174 Comments essential for living organisms. Calcium is a chemical element that is Most of the calcium within the human body is found

More information

Calcium Removal and Plaque Modification in the Era of DEB and Contemporary Stenting for Femoro- Popliteal Disease

Calcium Removal and Plaque Modification in the Era of DEB and Contemporary Stenting for Femoro- Popliteal Disease Calcium Removal and Plaque Modification in the Era of DEB and Contemporary Stenting for Femoro- Popliteal Disease Thomas M. Shimshak, MD Heart and Vascular Center Florida Hospital Heartland Medical Center

More information

Lower Extremity Arterial Disease

Lower Extremity Arterial Disease Lower Extremity Arterial Disease Circulating the Facts About Peripheral Disease Brought to you by the Education Committee of the Society for 1 www.svnnet.org Peripheral Artery Disease (PAD) Many people

More information

Access More Patients. Customize Each Seal.

Access More Patients. Customize Each Seal. Access More. Customize Each Seal. The Least Invasive Path Towards Proven Patency ULTRA LOW PROFILE TO EASE ADVANCEMENT The flexible, ultra-low 12F ID Ovation ix delivery system enables you to navigate

More information

Assessment of vulnerable plaque by OCT

Assessment of vulnerable plaque by OCT Assessment of vulnerable plaque by OCT Comparison with histology and possible clinical applications Takashi Akasaka, M.D. Department of Cardiovascular Medicine Wakayama, Japan Identification of vulnerable

More information

Pathology of Coronary Artery Disease

Pathology of Coronary Artery Disease Pathology of Coronary Artery Disease Seth J. Kligerman, MD Pathology of Coronary Artery Disease Seth Kligerman, MD Assistant Professor Medical Director of MRI University of Maryland Department of Radiology

More information

Assessment of plaque morphology by OCT in patients with ACS

Assessment of plaque morphology by OCT in patients with ACS Assessment of plaque morphology by OCT in patients with ACS Takashi Akasaka, M.D. Department of Cardiovascular Medicine Wakayama, Japan Unstable plaque Intima Lipid core Plaque rupture and coronary events

More information

Challenging of contrast agent-free endovascular treatment using 3D imaging

Challenging of contrast agent-free endovascular treatment using 3D imaging AC17-0010 Challenging of contrast agent-free endovascular treatment using 3D imaging Amane Kozuki Department of Cardiology, Osaka Saiseikai Nakatsu Hospital Introduction With advances in devices and techniques,

More information

Basic Physics of Ultrasound and Knobology

Basic Physics of Ultrasound and Knobology WELCOME TO UTMB Basic Physics of Ultrasound and Knobology By Daneshvari Solanki, FRCA Laura B. McDaniel Distinguished Professor Anesthesiology and Pain Medicine University of Texas Medical Branch Galveston,

More information

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426)

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Pulse-Echo Ultrasound Imaging Ultrasound Imaging Lecture 2 Daniel (Dan) Turnbull, Ph.D. Skirball Institute and Dept of Radiology NYU School of Medicine (daniel.turnbull@med.nyu.edu)

More information

Intravascular Ultrasound

Intravascular Ultrasound Scan for mobile link. Intravascular Ultrasound Intravascular ultrasound (IVUS) uses a transducer or probe to generate sound waves and produce pictures of the coronary arteries. IVUS can show the entire

More information

For Personal Use. Copyright HMP 2013

For Personal Use. Copyright HMP 2013 Case Report J INVASIVE CARDIOL 2013;25(2):E39-E41 A Case With Successful Retrograde Stent Delivery via AC Branch for Tortuous Right Coronary Artery Yoshiki Uehara, MD, PhD, Mitsuyuki Shimizu, MD, PhD,

More information

Development of Sensing System for Cerebral Aneurysm Treatment

Development of Sensing System for Cerebral Aneurysm Treatment Technical Paper Development of Sensing System for Cerebral Aneurysm Treatment Yoshitaka NAGANO Hideo FUJIMOTO In recent years, endovascular coil embolization has become a standard treatment option for

More information

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Page 1 of 7 11/18 Candidates are expected to cover all of the content of this syllabus when preparing

More information

Medical Review Approaches to the Diagnosis of Liver Fibrosis

Medical Review Approaches to the Diagnosis of Liver Fibrosis Medical Review Approaches to the Diagnosis of Liver Fibrosis Hiroko Iijima Department of Hepatobiliary and Pancreatic Disease Ultrasound Imaging Center, Hyogo College of Medicine Hiroko Iijima Department

More information

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound ARCHIVES OF ACOUSTICS 35, 4, 653 660 (2010) DOI: 10.2478/v10168-010-0049-6 Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound Wojciech SECOMSKI, Andrzej NOWICKI, Janusz WÓJCIK, Marcin

More information

Ultrasound in Medicine

Ultrasound in Medicine Ultrasound in Medicine Experimental Equipment for Medical Education Universities Colleges Medical Schools Medical and Med-Technical Training Education can befun! WELCOME TO GAMPT Devices and accessories

More information

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING H. J. Shin 1, Y. H. Jang 1, J. R. Kwan 2, H. D. Lee 3 1 INDE System Co., Ltd., Suwon, Kyunggi-do, 440-746,

More information

Measuring the radius of vibrating contrast bubbles in high-speed camera optical images Implementation of repetition rate excitation

Measuring the radius of vibrating contrast bubbles in high-speed camera optical images Implementation of repetition rate excitation Erasmus MC : Student Projects Student Projects We always have lots of interesting student projects available, for students in physics, mechanical and electrical engineering, image processing, optics, cell

More information

Abdominal Ultrasound

Abdominal Ultrasound Abdominal Ultrasound What is Ultrasound Imaging of the Abdomen? What are some common uses of the procedure? How should I prepare? What does the equipment look like? How does the procedure work? How is

More information

38 Vasa Vasorum Imaging

38 Vasa Vasorum Imaging 38 Vasa Vasorum Imaging Ioannis A. Kakadiaris, Sean O Malley, Manolis Vavuranakis, Ralph Metcalfe, Craig J. Hartley, Erling Falk, and Morteza Naghavi Co n t e n t s Key Points Introduction Fundamental

More information

Ultrasound: Past and Present. Lecturer: Dr. John M Hudson, PhD

Ultrasound: Past and Present. Lecturer: Dr. John M Hudson, PhD Ultrasound: Past and Present Lecturer: Dr. John M Hudson, PhD Disclosures 2 No conflicts of interest to declare Course Outline 3 1. Survey of ultrasound physics & applications 2. (Sep 21) 3. (Sep 28) 4.

More information

Evolution In Interventional Cardiology. Jawed Polad Jeroen Bosch Hospital s-hertogenbosch The Netherlands

Evolution In Interventional Cardiology. Jawed Polad Jeroen Bosch Hospital s-hertogenbosch The Netherlands Evolution In Interventional Cardiology Jawed Polad Jeroen Bosch Hospital s-hertogenbosch The Netherlands 25 November 2010 Coronary Atherosclerosis Timeline in interventional cardiology Indications for

More information

Dr Emma Chung. Safety first - Physical principles for excellent imaging

Dr Emma Chung. Safety first - Physical principles for excellent imaging Safety first - Physical principles for excellent imaging Dr Emma Chung Lecturer in Medical Physics, University of Leicester Clinical Scientist, University Hospitals of Leicester NHS Trust Thanks to Caroline

More information

Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training & Support

Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training & Support www.ozeninc.com info@ozeninc.com (408) 732 4665 1210 E Arques Ave St 207 Sunnyvale, CA 94085 Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training &

More information

Ambiguity in Detection of Necrosis in IVUS Plaque Characterization Algorithms and SDH as Alternative Solution

Ambiguity in Detection of Necrosis in IVUS Plaque Characterization Algorithms and SDH as Alternative Solution Ambiguity in Detection of Necrosis in IVUS Plaque Characterization Algorithms and SDH as Alternative Solution Amin Katouzian, Ph.D., Debdoot Sheet, M.S., Abouzar Eslami, Ph.D., Athanasios Karamalis, M.Sc.,

More information

Ultrasonic strain imaging made by speckle tracking of B-mode image

Ultrasonic strain imaging made by speckle tracking of B-mode image Ultrasonic strain imaging made by speckle tracking of B-mode image Introduction For centuries, physicians have used palpation to detect abnormal regions of increased elastic modulus (hardness or stiffness)

More information

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions. Lesson 07: Ultrasound Transducers This lesson contains 73 slides plus 16 multiple-choice questions. This lesson was derived from pages 33 through 42 in the textbook: Ultrasound Transducers Ultrasound Transducers

More information

Can IVUS Define Plaque Features that Impact Patient Care?

Can IVUS Define Plaque Features that Impact Patient Care? Can IVUS Define Plaque Features that Impact Patient Care? A Pichard L Satler, K Kent, R Waksman, W Suddath, N Bernardo, N Weissman, M Angelo, D Harrington, J Lindsay, J Panza. Washington Hospital Center

More information

M ost myocardial infarctions and other acute coronary

M ost myocardial infarctions and other acute coronary 556 BASIC RESEARCH Optical coherence tomographic elastography technique for measuring deformation and strain of atherosclerotic tissues J Rogowska, N A Patel, J G Fujimoto, M E Brezinski... See end of

More information

Descriptions of NDT Projects Fall 2004 October 31, 2004

Descriptions of NDT Projects Fall 2004 October 31, 2004 Descriptions of NDT Projects Fall 2004 October 31, 2004 Introduction There are two separate NDT labs in Magister: ULTRA for ultrasound and EDDY for eddy current. Both labs are equipped with mechanical

More information

Cigna - Prior Authorization Procedure List: Radiology & Cardiology

Cigna - Prior Authorization Procedure List: Radiology & Cardiology Cigna - Prior Authorization Procedure List: Radiology & Cardiology Category CPT Code CPT Code Description 93451 Right heart catheterization 93452 Left heart catheterization 93453 Combined right and left

More information

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents. Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm AN ANALYSIS OF THE CONTACT BETWEEN THE STENT AND THE ARTERY USING TUBE HIDROFORMING SIMULATION

More information

Human Systems. Technology - Ultrasounds

Human Systems. Technology - Ultrasounds Human Systems Technology - Ultrasounds What is General Ultrasound Imaging? Ultrasound imaging, also called ultrasound scanning or sonography, involves exposing part of the body to high-frequency sound

More information

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 Principles of Ultrasound Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 None Disclosures Outline Introduction Benefits and Limitations of US Ultrasound (US) Physics

More information

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Transcatheter Aortic Valve Implantation Procedure (TAVI) Page 1 of 5 Procedure (TAVI) Introduction Aortic stenosis (AS) is a common heart valve problem associated with heart failure and death. Surgical valve repair or replacement is recommended if AS patients

More information

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA) Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA) Aurélien Dumenil Pierre Louat Thierry Marchal Michel Rochette 2009 ANSYS, Inc. All rights reserved. 1 ANSYS, Inc. Proprietary Outline Clinical

More information

Calcium is a chemical element that is essential for living organisms.

Calcium is a chemical element that is essential for living organisms. 1 of 8 9/28/2015 9:04 AM Home About me Health and Nutrition Diet General Health Heart Disease August 19, 2014 By Axel F. Sigurdsson MD 259 Comments Like Share 82 Calcium is a chemical element that is essential

More information

Intravascular Ultrasound (IVUS)

Intravascular Ultrasound (IVUS) University Heart Journal Vol. 10, No. 2, July 2014 Intravascular Ultrasound (IVUS) S MOKADDAS HOSSAIN (SADI) Labaid Cardiac Hospital, Dhaka, Bangladesh Address of Correspondence: Dr. S Mokaddas Hossain

More information

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1 Q1.(a) When ultrasound is incident at an interface between two different media some energy is transmitted and some is reflected. The ratio of the reflected energy intensity I r to the incident energy intensity

More information

Ultrasound Physics & Doppler

Ultrasound Physics & Doppler Ultrasound Physics & Doppler Endocrine University 2018 Mark Lupo, MD, FACE, ECNU Objectives Review the essential components of ultrasound physics in neck sonography Demonstrate the importance of ultrasound

More information

Intravascular Ultrasound (IVUS) and Optical Coherence Tomography (OCT)

Intravascular Ultrasound (IVUS) and Optical Coherence Tomography (OCT) Intravascular Ultrasound (IVUS) and Optical Coherence Tomography (OCT) Clare McLaren Great Ormond Street Hospital London Introduction IVUS and OCT supplementary techniques to angiography provide information

More information

Chapter 6: Summary

Chapter 6: Summary Chapter 6: Summary The present thesis describes several clinical studies on the topic of intravenous myocardial contrast echocardiography, and an experimental study on the effect of ultrasound contrast

More information

Animesh Rathore, MD 4/21/17. Penetrating atherosclerotic ulcers of aorta

Animesh Rathore, MD 4/21/17. Penetrating atherosclerotic ulcers of aorta Animesh Rathore, MD 4/21/17 Penetrating atherosclerotic ulcers of aorta Disclosures No financial disclosures Thank You Dr. Panneton for giving this lecture for me. I am stuck at Norfolk with an emergency

More information

Ultrasound imaging is a noninvasive medical test that helps physicians diagnose and treat medical conditions.

Ultrasound imaging is a noninvasive medical test that helps physicians diagnose and treat medical conditions. CAROTID ULTRASOUND What is Carotid Ultrasound Imaging? Ultrasound imaging, also called ultrasound scanning or sonography, involves exposing part of the body to highfrequency sound waves to produce pictures

More information

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p Unit 3 Maintaining Dynamic Equilibrium Circulatory system The heart Blood Blood Pressure Circulatory Disorders Oct 11 3:27 PM Circulatory System p.304 307 If an organism is small and flat it can transport

More information

Quantitative Imaging of Transmural Vasa Vasorum Distribution in Aortas of ApoE -/- /LDL -/- Double Knockout Mice using Nano-CT

Quantitative Imaging of Transmural Vasa Vasorum Distribution in Aortas of ApoE -/- /LDL -/- Double Knockout Mice using Nano-CT Quantitative Imaging of Transmural Vasa Vasorum Distribution in Aortas of ApoE -/- /LDL -/- Double Knockout Mice using Nano-CT M. Kampschulte 1, M.D.; A. Brinkmann 1, M.D.; P. Stieger 4, M.D.; D.G. Sedding

More information

CPIS So-Yeon Choi, MD., PhD. Department of Cardiology Ajou University School of MedicineSuwon, Korea

CPIS So-Yeon Choi, MD., PhD. Department of Cardiology Ajou University School of MedicineSuwon, Korea So-Yeon Choi, MD., PhD. Department of Cardiology Ajou University School of MedicineSuwon, Korea Coronary Artery Imaging The ideal coronary imaging technology would be capable of identifying not only vessel

More information

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification Whitepaper Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification ACUSON S2000 Ultrasound System Answers for life. Page 1 Tissue Strain Analytics: Virtual Touch Tissue Imaging and Quantification

More information

Rheological, mechanical and failure properties of biological soft tissues at high strains and rates of deformation

Rheological, mechanical and failure properties of biological soft tissues at high strains and rates of deformation Rheological, mechanical and failure properties of biological soft tissues at high strains and rates of deformation Society of Rheology Conference Salt Lake City, Utah October 10, 2007 Martin L. Sentmanat,

More information

Radiologic Imaging Magnetic Resonance Imaging (MRI)

Radiologic Imaging Magnetic Resonance Imaging (MRI) Radiologic Imaging X-ray has always been the golden rule in diagnosing and treating podiatric patients. Unfortunately, for some patients the diagnosis is not as evident. That is when we need to utilize

More information

Use of Nuclear Cardiology in Myocardial Viability Assessment and Introduction to PET and PET/CT for Advanced Users

Use of Nuclear Cardiology in Myocardial Viability Assessment and Introduction to PET and PET/CT for Advanced Users Use of Nuclear Cardiology in Myocardial Viability Assessment and Introduction to PET and PET/CT for Advanced Users February 1 5, 2011 University of Santo Tomas Hospital Angelo King A-V Auditorium Manila,

More information

Quick guide. Core. precision guided therapy system

Quick guide. Core. precision guided therapy system Quick guide Core precision guided therapy system The Philips Volcano imaging system should only be operated by trained personnel. The following information is presented for your convenience and is not

More information

Ultrasound Physics & Terminology

Ultrasound Physics & Terminology Ultrasound Physics & Terminology This module includes the following: Basic physics terms Basic principles of ultrasound Ultrasound terminology and terms Common artifacts seen Doppler principles Terms for

More information

How to evaluate heart disease - Do we need new tools Focus on myocardial circulation

How to evaluate heart disease - Do we need new tools Focus on myocardial circulation How to evaluate heart disease - Do we need new tools Focus on myocardial circulation Alf Inge Larsen Professor department of clinical science University of Bergen Head of Cardiovascular Research Group

More information

Antegrade techniques for CTO recanalization. Dr. George Karavolias, MD, PhD, FESC, FACC Interventional Cardiologist

Antegrade techniques for CTO recanalization. Dr. George Karavolias, MD, PhD, FESC, FACC Interventional Cardiologist Antegrade techniques for CTO recanalization Dr. George Karavolias, MD, PhD, FESC, FACC Interventional Cardiologist can CTOs be reliably opened by PCI? Meta-Analysis of 18,061 Patients Patel V, J Am Coll

More information

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D. Diagnostic Ultrasound Sutiporn Khampunnip, M.D. Definition of Ultrasound Ultrasound is simply sound waves, like audible sound. High-frequency sound and refers to mechanical vibrations above 20 khz. Human

More information

Investigation of metallic surface area of coronary stents

Investigation of metallic surface area of coronary stents Investigation of metallic surface area of coronary stents Dóra KÁROLY 1, Miksa KOVÁCS 1, Andrew Attila TERDIK 1, Eszter BOGNÁR 1,2 1 Department of Materials Science and Engineering, Faculty of Mechanical

More information

Multi-Frequency Processing for Lumen Enhancement with Wideband. Intravascular Ultrasound. A Thesis Presented to the Faculty of the

Multi-Frequency Processing for Lumen Enhancement with Wideband. Intravascular Ultrasound. A Thesis Presented to the Faculty of the Multi-Frequency Processing for Lumen Enhancement with Wideband Intravascular Ultrasound A Thesis Presented to the Faculty of the Biomedical Engineering Department California Polytechnic State University,

More information