Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3

Size: px
Start display at page:

Download "Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3"

Transcription

1 Interchangeability of pencil-beam and fan-beam dual-energy X-ray absorptiometry measurements in piglets and infants 1 3 Winston WK Koo, Mouhanad Hammami, and Elaine M Hockman ABSTRACT Background: Compared with the older pencil-beam (PB) dualenergy X-ray absorptiometry (DXA), the newer fan-beam (FB) DXA has the advantage of faster scan acquisition and greater accuracy of body-composition measurement in small subjects. However, no data exist on the relation between the measurements obtained with these techniques. Objective: The objective of the study was to investigate whether PB and FB DXA measurements in small subjects are interchangeable. Design: PB and FB DXA scans were performed on 26 piglets and 54 infants to examine the relation between the measurements obtained by using the 2 techniques. Results: The correlation between all PB and FB DXA measurements of variables (total weight, bone area, bone mineral content, bone mineral density, and lean and fat masses) approached 1.0, but there were significant differences in absolute values. The extent of the differences varied according to the variable, with the lowest value for total weight (mean difference: 1% for both piglets and infants) and the highest value for bone mineral content (mean difference: 35.3% and 36.7% for piglets and infants, respectively). PB and FB DXA measurements were strongly predictive of each other after adjustment (r 2 = for the piglet data and for the infant data). Conclusion: In small subjects, DXA measurements from PB and FB techniques were strongly predictive of each other, although their absolute values differed. Thus, group comparison of PB and FB DXA data is possible after adjustment of the data from either technique. It is advisable to generate normative data for each technique and to use the same technique throughout longitudinal studies. Am J Clin Nutr 2003;78: KEY WORDS Dual-energy X-ray absorptiometry, pencil-beam, fan-beam, body composition, piglets, infants INTRODUCTION Dual-energy X-ray absorptiometry (DXA) is now generally accepted as the standard for the measurement of bone mass and is the preferred means of measuring soft-tissue composition in small subjects such as human infants (1). Extensive data are available on the older pencil-beam (PB) DXA technique for measuring bone mass and body composition throughout infancy, because its use was independently validated by several research groups who used instruments from the same manufacturer, the same scan acquisition technique with the use of an infant platform, and the same analytic algorithm with infant whole-body software (version 5.64 or later; Hologic Inc, Waltham, MA) (2 5). Advances in DXA technology resulted in the development of a new generation of densitometers in which the PB X-ray source and the single detector of the first-generation instruments were replaced by a fan-beam (FB) X-ray source and a multiple detector array. The new densitometers offer increased scan speed and improved spatial resolution; both attributes have recently been confirmed in a study of piglets (6). These attributes enhance the ability of DXA to measure bone mass and body composition because of the FB technique s faster scan acquisition and improved accuracy in small subjects compared with the PB technique, and they offer the potential for increased use in studies of infants and young children. Instrument effect, as indicated by systematic differences among measurements from instruments based on the same DXA technique (7) and those from instruments based on different DXA techniques (8 13), supports the critical need for cross-calibration studies during the transition from first-generation to secondgeneration DXA instrumentation, such as that which occurs with the addition of new equipment or the replacement of old equipment. This information is needed to eliminate or at least minimize the potential discrepancies or errors that may exist between 2 instruments using different DXA techniques. Furthermore, this information is critical to appropriate interpretation of new data and comparison with the existing data generated from older instruments, in particular with regard to whether new normative data should be generated and whether it is feasible to use these instruments interchangeably during a longitudinal study. In adults and older children, the bone, lean mass (LM), and fat mass (FM) measurements with FB DXA show a systematic relation with data from the PB DXA (8 13). In some reports, equivalency of bone mineral density (BMD), LM, and FM measurements with FB DXA has been confirmed by cross-calibration with other criterion methods including PB DXA, the 4-compartment 1 From the Departments of Pediatrics, Obstetrics and Gynecology (WWKK and MH), and Computing and Information Technology (EMH), Wayne State University, Detroit. 2 Supported in part by Ross Products Division, Abbott Laboratories, Columbus, OH. 3 Reprints not available. Address correspondence to WWK Koo, Hutzel Hospital, Department of Pediatrics, 4707 St Antoine Boulevard, Detroit, MI wkoo@wayne.edu. Received July 18, Accepted for publication March 6, Am J Clin Nutr 2003;78: Printed in USA American Society for Clinical Nutrition

2 PENCIL-BEAM AND FAN-BEAM DXA MEASUREMENTS 237 model, and multislice computed tomography scans (10, 12). However, a recent study in adults who had a mean (± SD) change in body weight of 5.7 ± 4.5 kg found that both PB and FB DXA measured changes in body composition that were similar to changes measured with the use of total body water, but the data from PB and FB DXA are not interchangeable (13). Whether such similarity exists between PB and FB DXA measurements in infants and in young children is not known, because the body composition of small subjects differs substantially from that of adults and older children, and different software algorithms are needed for scan acquisition and analysis. We therefore aimed to ascertain the relation between the older-generation PB DXA measurements and the newer-generation FB DXA measurements in small subjects and to determine whether it is feasible to cross-calibrate the measurements obtained with the use of these 2 techniques. SUBJECTS AND METHODS Piglets Twenty-six domestic piglets (J&M Farms, Lansing, MI) with a mean (± SD) body weight of 8359 ± 5914 g (range: g) were studied as part of a comprehensive protocol to determine the various aspects of body-composition measurements in a piglet model. Each animal was weighed on an electronic scale (Seca; Toledo Scale Co, Toledo OH) immediately before DXA scanning and then killed at completion of the in vivo study procedures. DXA scans were obtained with FB and PB whole-body densitometers (QDR 4500A and 2000+, respectively; Hologic Inc). Scan acquisition was performed by using the infant whole-body mode according to the manufacturer s recommendations. A 2-platform system (aluminum platform overlying a foam table pad) and an external calibration standard were used in conjunction with the PB densitometer. Each piglet was placed on a cotton blanket in the prone position with the front and hind limbs extended. The long axis of the animal was positioned at the midline of the platform with the snout 5 cm from the cranial end of the platform (2, 6). Each piglet was covered with a cotton blanket and a disposable diaper was used in larger piglets to prevent soiling. The same coverings were used for all scan acquisition. The mean weights of the piglets including coverings were 8803 ± 5861 g (range: g). The piglets were sedated with sodium pentobarbital and sodium thiopental and kept in the same posture for all scan acquisitions. Scan analysis used software vkh6 validated for the FB DXA measurement of small subjects (6) and software v5.71p validated for the PB DXA measurement (2, 4, 5). All scans were checked for movement artifacts during scan acquisition and analysis. Details of all DXA procedures were reported elsewhere (2, 3, 6). With our densitometer, the typical entry radiation exposure during an infant whole-body scan was a maximum dose of 3 Sv (1 Sv = 0.1 mrem) for the PB DXA with the infant platform and of 7 Sv for the FB DXA. The radiation scatter at 66 cm from each densitometer did not differ significantly from background during 10 min of measurement. The animal study was approved by the Animal Investigation Committee at Wayne State University. Infants Fifty-four infants with a mean naked body weight of 7285 ± 3310 g (range: g) were studied at 216 ± 160 d (range: d). Each infant underwent one scan with the use of each densitometer to determine the relation between the data generated by these 2 techniques. This is a cross-sectional design with data from subjects in 4 clinical studies at the same institution during the transition phase in which the DXA measurements were switched over from the PB technique to the FB technique. Each infant wore a diaper and was swaddled in a cotton blanket; a crosscalibration scan was done if the infant remained asleep, and the infant s covering and position were kept as similar as possible for both scans. In older infants, the wearing of lightweight underwear was allowed. All coverings, including a diaper and any clothing, were weighed separately, and the mean total weight at DXA measurement was 7506 ± 3358 g (range: g). A parent and the technician were present during the study. Scan acquisition used the infant whole-body mode and scan analysis with the same software as that used for piglet studies (2, 3, 6). All clinical studies were approved by the Human Investigation Committee at Wayne State University, and written informed consent was obtained from the parent of each subject. Statistical analysis Paired t tests were used to ascertain differences between each specific DXA measurement (bone area, bone mineral content, BMD, FM, LM, and total weight) obtained with the use of the PB and FB techniques. Regression analysis was used to determine the extent of the relation between PB and FB DXA measurements. Given the high precision of the PB (14 16) and FB (6) DXA measurements, a sample size of 30 subjects should be sufficient to determine the true relation between these measurements (17), although reports of previous clinical studies indicated that a much larger group of subjects was needed to better reflect the wide range of body composition in vivo and to optimize the ability to determine the true relation between PB and FB DXA measurements (8 12). Data analyses were performed separately for piglets and infants. All statistical tests were performed with the use of SPSS for WINDOWS (version 10.0; SPSS Inc, Chicago), and P = 0.05 indicated significance. Unless otherwise stated, all values are means ± SDs. RESULTS There was near-perfect consistency between the measurements by these 2 techniques in each subject with respect to the position in the distribution ie, the r values approached 1.0 for all variables measured but the location of the distributions ie, t values obtained from comparisons of the means differed significantly. The extent of difference between PB and FB measurements varied according to the specific variable. PB DXA showed slightly greater total weight and LM for the piglets and greater total weight for the infants than did FB DXA. However, there was a difference of > 30% between the PB and the FB techniques in the measurements of some components of body composition, eg, bone mineral content (Tables 1 and 2). In contrast, DXA measurements obtained with the PB technique were highly predictive of those obtained with the FB technique with adjusted r 2 between and for the piglet data and between and for the infant data (Figures 1 and 2). DISCUSSION One of the major advantages of the FB DXA over the PB DXA in the measurement of bone mass and body composition

3 238 KOO ET AL TABLE 1 Pencil-beam (PB) and fan-beam (FB) dual-energy X-ray absorptiometry (DXA) measurements of weight and body composition in 26 piglets PB FB Percentage difference: (PB FB)PB 1 r DXA total weight (g) 8499 ± ± ± Bone area (cm 2 ) 503 ± ± ± Bone mineral content (g) 150 ± ± ± Bone mineral density (g/cm 2 ) ± ± ± Fat mass (g) 674 ± ± ± Lean mass (g) 7675 ± ± ± P for all comparisons (paired t test), except fat mass (P = 0.71). 2 x ± SD. % in infants and young children is the significant difference in scan acquisition time, ie, a reduction from 15 min to < 3 min in older infants (6). This shorter scan acquisition time results in less potential for movement artifact with the FB DXA (3). The FB DXA algorithm also corrected the underestimation of carcass ash content and the overestimation of fat content that were reported for the PB DXA measurement in small subjects (2, 4, 5) and maintained adequate precision for clinical needs (6). With the gradual withdrawal of industry support for the PB DXA instrument, the newer FB technique will inevitably become the only available DXA technique for body-composition measurement in subjects of all ages and body masses. Thus, every investigator will face at some point the necessity to transition from the PB DXA to the FB DXA technique and to determine the relation between the data generated from the current PB DXA measurements and those from the future standard FB DXA measurements, particularly for ongoing longitudinal studies. We considered the use of phantoms for this cross-calibration study, but it has been shown that the use of a small number of phantoms results in unacceptably large residual variability around the regression line (8), and the use of a large number of phantoms to cover the full range of expected body-composition data would be impractical. The domestic piglet has many similarities to the human infant and is frequently used as a model for body-composition studies (2 6). However, domestic pigs are normally bred for the rapid accumulation of lean tissue, and their rate of body fat accumulation is much slower than that of the human infant. Thus, neither the piglet nor other commonly available growing animals can adequately reflect the complete range of body composition in infants and young children, and this fact makes clinical study in vivo a necessity. The sample size of our clinical study is much smaller than the sample sizes reported for calibration studies in adults (8 10). However, our study design in human infants used a few subjects from each of several clinical studies to obtain body-composition measurements in infants with a wide range of body weights. These data coupled with the consistency between the animal and human data support the adequacy of our study design for a meaningful interpretation of the data generated from these 2 DXA techniques, ie, our study design achieved an adequate sampling of heterogeneous subjects in vivo and contributed to a meaningful crosscalibration study during instrument transition. The radiation exposure from FB DXA scan is the same as that from the PB DXA scan obtained without using the infant platform (2, 18 23) or from a repeat PB DXA procedure if the initial scan was unsatisfactory (22). In any case, the extent of radiation exposure is well within the range of natural background radiation (24). This approach is similar to that in reports from multiple investigators on the use of duplicate DXA scans as a quality-control procedure critical to any clinical study (14, 15, 25 31). In this study, the differences between mean PB and FB DXAmeasured total weights were quite small at 1% for both piglets and infants. However, there were greater differences between other DXA measurement obtained with these techniques: a difference of 5% in mean LM and much greater differences in bone and fat measurements. The large differences between PB and FB DXA-measured bone and fat reflect the correction by FB DXA for the PB DXA underestimation and overestimation of these variables in small subjects (2, 4, 5). The differences in bone area and BMD in part reflect the magnification effect of the FB technique and the different software algorithm used by the PB and FB DXA systems for measurements in small subjects. The homogeneity of FM in piglets explains the lack of difference in absolute values of FM between the 2 techniques. Our data showed that all PB DXA measurements are significantly predictive of the corresponding FB DXA measurements. TABLE 2 Pencil-beam (PB) and fan-beam (FB) dual-energy X-ray absorptiometry (DXA) measurements of weight and body composition in 54 infants PB FB Percentage difference: (PB FB)PB 1 r % DXA total weight (g) 7620 ± ± ± Bone area (cm 2 ) 553 ± ± ± Bone mineral content (g) 159 ± ± ± Bone mineral density (g/cm 2 ) ± ± ± Fat mass (g) 2262 ± ± ± Lean mass (g) 5198 ± ± ± P for all comparisons (paired t test). 2 x ± SD.

4 PENCIL-BEAM AND FAN-BEAM DXA MEASUREMENTS 239 FIGURE 1. Predictive ability of pencil-beam dual-energy X-ray absorptiometry for fan-beam dual-energy X-ray absorptiometry measurements in piglets showing regression prediction line with 95% CI for the mean. Total weight includes the weight of the covering. Numbers in parentheses = adjusted r 2 and SEE, respectively. Given an almost perfect linear relation between each variable measured by the two techniques, it is expected that the critical ratios and the corresponding P values would be equivalent, whether the data were generated from PB or FB DXA. It is therefore possible to compare the existing PB data with those increasingly being generated by the newer FB DXA technique. Our data from measurements in small subjects suggest that, on a group basis, there is adequate interchangeability between the 2 DXA techniques, although the data from one of these techniques will have to be adjusted. This finding is in contrast with that from a report in adults that changes in body composition associated with a small change in body weight as measured by PB and FB DXA techniques are not interchangeable (13). This difference in findings presumably reflects the differences in the software for adult and infant DXA measurements, with the latter having a better capability for discrimination at low body mass. The acceptability of a conversion factor in the standardization of DXA measurement from PB to FB values, ie, the clinical relevance of this systematic relation, would depend on the magnitude of the residuals associated with the conversion, which varies with the value being compared and with the changes in these values expected under different circumstances. For infants, the consistently excellent predictive ability of PB DXA for FB DXA measurements is reflected in the adjusted r 2 values of 0.999, 0.993, 0.986, and for the prediction of total weight, bone mineral content, LM, and FM, respectively. Conversely, the SEE for the respective FB DXA measurements is extremely small. However, FIGURE 2. Predictive ability of pencil-beam dual-energy X-ray absorptiometry for fan-beam dual-energy X-ray absorptiometry measurements in infants showing regression prediction line with 95% CI for the mean. Total weight includes the weight of the covering. Numbers in parentheses = adjusted r 2 and SEE, respectively. the absolute values from these 2 techniques are not directly interchangeable. Thus, to minimize any errors in the conversion of data between these 2 techniques and to optimize the ability to detect small differences or changes in body-composition measurement, it seems preferable to generate normative data specifically for each DXA technique and to maintain the use of the same technique, whether PB DXA or FB DXA, throughout longitudinal studies. The same conclusions can be drawn for bone area and BMD measurements, although these variables have little clinical relevance because bone area is subjected to changes in the posture of the subject at scan acquisition, and DXA BMD is not recommended for use as a measurement in small growing subjects (32, 33). We conclude that, in growing piglets and infants, there is a systematic relation between PB and FB DXA measurements, although there are significant differences in the absolute values obtained with the use of the 2 techniques. The extent of the differences varies with the DXA variable. Other potential errors due to instrument effect, operators, and techniques of scan acquisition and analysis (3, 7, 34, 35) also must be considered. Thus, to optimize the interpretation of the data obtained from both instruments, it is incumbent on the investigator to generate a cross-calibration equation from a subset of subjects on the specific instruments in use during the transition from the PB DXA technique to the FB DXA technique. MH participated in the study design, data collection, and preparation of the manuscript; EMH participated in the study design, data analysis, and preparation of the manuscript; and WWKK participated in all aspects of this study.

5 240 KOO ET AL None of the authors had any financial or contractual conflict of interest associated with this study. REFERENCES 1. Koo WWK. Body composition measurements during infancy. Ann N Y Acad Sci 2000;904: Koo WWK, Massom LR, Walters J. Validation of accuracy and precision of dual energy x-ray absorptiometry for infants. J Bone Miner Res 1995;10: Koo WWK, Walters J, Bush AJ. Technical considerations of dual energy x-ray absorptiometry-based bone mineral measurements for pediatric studies. J Bone Miner Res 1995;10: Picaud JC, Rigo J, Nyamugabo K, Milet J, Senterre J. Evaluation of dual-energy X-ray absorptiometry for body-composition assessment in piglets and term human neonates. Am J Clin Nutr 1996;63: Fusch C, Slotboom J, Fuehrer U, et al. Neonatal body composition: dual energy X ray absorptiometry, magnetic resonance imaging, and three dimensional chemical shift imaging versus chemical analysis in piglets. Pediatr Res 1999;46: Koo WWK, Hammami M, Hockman EM. Use of fan beam dual energy x-ray absorptiometry to measure body composition of piglets. J Nutr 2002;132: Hammami M, Picaud JC, Fusch C, Hockman EM, Rigo J, Koo WWK. Phantoms for cross-calibration of dual energy X-ray absorptiometry measurements in infants. J Am Coll Nutr 2002;21: Finkelstein JS, Butler JP, Cleary RL, Neer RM. Comparison of four methods for cross-calibrating dual energy X-ray absorptiometers to eliminate systematic errors when upgrading equipment. J Bone Miner Res 1994;9: Abrahamsen B, Gram J, Hansen TB, Beck-Nielsen H. Cross calibration of QDR 2000 and QDR 1000 dual energy X-ray densitometers for bone mineral and soft tissue measurements. Bone 1995;16: Bouyoucef SE, Cullum ID, Ell PJ. Cross-calibration of a fan-beam X-ray densitometer with a pencil-beam system. Br J Radiol 1996;69: Ellis K, Shypailo RJ. Bone mineral and body composition measurements: cross calibration of pencil-beam and fan-beam dual energy X-ray absorptiometers. J Bone Miner Res 1998;13: Salamone LM, Fuerst T, Visser M, et al. Measurement of fat mass using DEXA: a validation study in elderly adults. J Appl Physiol 2000;89: Tylavski FA, Lohman TG, Dockrell M, et al. Comparison of the effectiveness of 2 dual-energy x-ray absorptiometers with that of total body water and computed tomography in assessing changes in body composition during weight change. Am J Clin Nutr 2003;77: Koo WWK, Bush AJ, Walters J, Carlson SE. Postnatal development of bone mineral status during infancy. J Am Coll Nutr 1998;17: Koo WWK, Walters JC, Hockman EM. Body composition in human infants at birth and postnatally. J Nutr 2000;130: Brunton JA, Weiler HA, Atkinson SA. Improvement in the accuracy of dual energy x-ray absorptiometry for whole body and regional analysis of body composition: validation using piglets and methodological considerations in infants. Pediatr Res 1997;41: Nunnally JC. Psychometric theory. New York: McGraw Hill, 1967: Venkataraman PS, Ahluwalia BW. Total bone mineral content and body composition by x-ray densitometry in newborns. Pediatrics 1992;90: Lapillonne AA, Glorieux FH, Salle BL, et al. Mineral balance and whole body bone mineral content in very-low-birth-weight infants. Acta Paediatr Suppl 1994;405: Lapillonne A, Braillon P, Claris O, Chatelain PG, Delmas PD, Salle BL. Body composition in appropriate and in small for gestational age infants. Acta Paediatr 1997;86: Brunton JA, Saigal S, Atkinson SA. Growth and body composition in infants with bronchopulmonary dysplasia up to 3 months corrected age: a randomized trial of a high-energy nutrient-enriched formula fed after hospital discharge. J Pediatr 1998;133: Wauben IPM, Atkinson SA, Shah JK, Paes B. Growth and body composition of preterm infants: influence of nutrient fortification of mother s milk in hospital and breastfeeding post-hospital discharge. Acta Paediatr 1998;87: Wauben IP, Atkinson SA, Grad TL, Shah JK, Paes B. Moderate nutrient supplementation of mother s milk for preterm infants supports adequate bone mass and short term growth: a randomized, controlled trial. Am J Clin Nutr 1998;67: International Atomic Energy Agency fact sheet. Radiation in everyday life. Internet: (accessed 12 November 2002). 25. Li R, O Connor L, Buckley D, Specker B. Relation of activity levels to body fat in infants 6 to 12 months of age. J Pediatr 1995; 126: Rupich RC, Specker BL, Leiuw-A-Fa M, Ho M. Gender and race differences in bone mass during infancy. Calcif Tissue Int 1996:58: Specker BL, Beck A, Kalkwarf H, Ho M. Randomized trial of varying mineral intake on total body bone mineral accretion during the first year of life. Pediatrics 1997;99:E12. Internet: pediatrics.aappublications.org/cgi/content/full/99/6/e12 (accessed 11 June 2003). 28. Mehta KC, Specker BL, Bartholmey S, Giddens J, Ho ML. Trial on timing of introduction to solids and food type on infant growth. Pediatrics 1998;102: Butte N, Heinz C, Hopkinson J, Wong W, Shypailo R, Ellis K. Fat mass in infants and toddlers: comparability of total body water, total body potassium, total body electrical conductivity, and dual-energy x-ray absorptiometry. J Pediatr Gastroenterol Nutr 1999;29: Butte NF, Hopkinson JM, Wong WW, O Brien Smith E, Ellis KJ. Body composition during the first 2 years of life: an updated reference. Pediatr Res 2000;47: Butte NF, Wong WW, Hopkinson JM, O Brian Smith E, Ellis KJ. Infant feeding mode affects early growth and body composition. Pediatrics 2000;106: Nelson DA, Koo WWK. Interpretation of absorptiometric bone mass measurements in the growing skeleton: issues and limitations. Calcif Tissue Int 1999;65: Rauch F, Schoenau E. Changes in bone density during childhood and adolescence: an approach based on bone s biological organization. J Bone Miner Res 2001:16: Koo MWM, Yang KH, Begeman P, Hammami M, Koo WWK. Prediction of bone strength in growing animals using non-invasive bone mass measurements. Calcif Tissue Int 2001;68: Hammami M, Koo MWM, Koo WWK, Thomas RT, Rakhman D. Regional bone mass measurement from whole body dual energy x-ray absorptiometry scan. J Clin Densitom 2001;4:131 6.

pqct Measurement of Bone Parameters in Young Children

pqct Measurement of Bone Parameters in Young Children Journal of Clinical Densitometry, vol. 3, no. 1, 9 14, Spring 2000 Copyright 2000 by Humana Press Inc. All rights of any nature whatsoever reserved. 0169-4194/00/3:9 14/$11.50 Original Article pqct Measurement

More information

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Stephen J Pintauro, Tim R Nagy, Christa M Duthie, and Michael

More information

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers* JOURNAL OF BONE AND MINERAL RESEARCH Volume 13, Number 10, 1998 Blackwell Science, Inc. 1998 American Society for Bone and Mineral Research Bone Mineral and Body Composition Measurements: Cross-Calibration

More information

Roberto D Sainz and Alessandro Urlando

Roberto D Sainz and Alessandro Urlando Evaluation of a new pediatric air-displacement plethysmograph for body-composition assessment by means of chemical analysis of bovine tissue phantoms 1 3 Roberto D Sainz and Alessandro Urlando ABSTRACT

More information

Body-composition assessment in infancy: air-displacement plethysmography compared with a reference 4-compartment model 1 4

Body-composition assessment in infancy: air-displacement plethysmography compared with a reference 4-compartment model 1 4 Body-composition assessment in infancy: air-displacement plethysmography compared with a reference 4-compartment model 1 4 Kenneth J Ellis, Manjiang Yao, Roman J Shypailo, Alessandro Urlando, William W

More information

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers European Journal of Clinical Nutrition (1997) 51, 449±454 ß 1997 Stockton Press. All rights reserved 0954±3007/97 $12.00 Body composition analysis by dual energy X-ray absorptiometry in female diabetics

More information

Norland Densitometry A Tradition of Excellence

Norland Densitometry A Tradition of Excellence Norland Densitometry A Tradition of Excellence Norland DXA Bone Density Measurement Osteoporosis is a disease marked by reduced bone strength leading to an increased risk of fractures. About 54 million

More information

special communication

special communication special communication Determining bone and total body mineral content from body density and bioelectrical response spectroscopy STEVEN F. SICONOLFI, 1 RANDAL J. GRETEBECK, 2 WILLIAM W. WONG, 3 SHERIL S.

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data

More information

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? Osteoporos Int (2010) 21:1227 1236 DOI 10.1007/s00198-009-1062-3 ORIGINAL ARTICLE Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? B. Fan & Y. Lu

More information

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in 1 2 In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in adults: the influence of body mass index 3 4 Running title: Precision of the idxa for the measurement of visceral

More information

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017 STRUCTURED EDUCATION REQUIREMENTS Bone Densitometry The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled

More information

Feeding the Small for Gestational Age Infant. Feeding the Small for Gestational Age Infant

Feeding the Small for Gestational Age Infant. Feeding the Small for Gestational Age Infant Feeding the Small for Gestational Age Infant Feeding the Small for Gestational Age Infant What s the right strategy? Infants born small-for-gestational age (SGA) are at higher risk for adult diseases.

More information

Chapter 17: Body Composition Status and Assessment

Chapter 17: Body Composition Status and Assessment Chapter 17: Body Composition Status and Assessment American College of Sports Medicine. (2010). ACSM's resource manual for guidelines for exercise testing and prescription (6th ed.). New York: Lippincott,

More information

Postdischarge nutrition,

Postdischarge nutrition, Postdischarge nutrition, is there a role for human milk? Harrie N. Lafeber MD,Ph.D. Professor of Neonatology, VU university medical center, Amsterdam, NL Ehrencranz et al. Indirect calorimetry 120-130

More information

The Bone Densitometry Examination

The Bone Densitometry Examination The Bone Densitometry Examination The purpose of The American Registry of Radiologic Technologist (ARRT ) Bone Densitometry Examination is to assess the knowledge and cognitive skills underlying the intelligent

More information

Validation of a new pediatric air-displacement plethysmograph for assessing body composition in infants 1 4

Validation of a new pediatric air-displacement plethysmograph for assessing body composition in infants 1 4 Validation of a new pediatric air-displacement plethysmograph for assessing body composition in infants 1 4 Guansheng Ma, Manjiang Yao, Yan Liu, Aiwei Lin, Hui Zou, Alessandro Urlando, William W Wong,

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Bone Densitometry The purpose of the bone densitometry examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

The bone mass concept: problems in short stature

The bone mass concept: problems in short stature European Journal of Endocrinology (2004) 151 S87 S91 ISSN 0804-4643 The bone mass concept: problems in short stature E Schoenau, C Land, A Stabrey, T Remer 1 and A Kroke 1 Children s Hospital, University

More information

DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals

DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals PERFORMANCE DEXA is an advanced technology originally used to, and still capable of assessing bone health

More information

9 Quality Assurance in Bone Densitometry section

9 Quality Assurance in Bone Densitometry section 9 Quality Assurance in Bone Densitometry section Introduction Bone densitometry is frequently used to determine an individual's fracture risk at a particular point in time but may also be used to assess

More information

Discussion Session 2 CHAIR: H.K.A. VISSER CHAPTER 6

Discussion Session 2 CHAIR: H.K.A. VISSER CHAPTER 6 Discussion Session 2 CHAIR: H.K.A. VISSER CHAPTER 6 Elia One of the major considerations with any method for determining body composition is its accuracy. Given your attempt to demonstrate this accuracy

More information

Bone Mineral Density and Its Associated Factors in Naresuan University Staff

Bone Mineral Density and Its Associated Factors in Naresuan University Staff Naresuan University Journal 2005; 13(3): 13-18 13 Bone Mineral Density and Its Associated Factors in Naresuan University Staff Supawitoo Sookpeng *, Patsuree Cheebsumon, Malinee Dhanarun, Thanyavee Pengpan

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT O S T E O P O R O S I S A S S E S S M E N T Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT Identify Patients at Risk The evaluation of bone health has become an essential

More information

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7 STRUCTURED SELF ASSESSMENT CONTENT SPECIFICATIONS SSA LAUNCH DATE: JANUARY 1, 2018 Bone Densitometry The purpose of continuing qualifications requirements (CQR) is to assist registered technologists in

More information

Lunar Prodigy Advance

Lunar Prodigy Advance GE Medical Systems Lunar Prodigy Advance Direct-Digital Densitometry imagination at work Your practice needs to move fast, yet you want peace of mind. A partnership is a journey - expertise, support and

More information

Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry

Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry Estimating the Absorbed Dose to Critical Organs During Dual X-ray Absorptiometry M Mokhtari-Dizaji, PhD 1 A A Sharafi, PhD 2 B Larijani, MD 3 N Mokhlesian, MSc 4 H Hasanzadeh, MS 1 Index terms: Dual X-ray

More information

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Page: 1 of 6 Last Review Status/Date: March 2015 Determine Body Composition Description Using low dose x-rays of two different energy levels, whole body dual x-ray absorptiometry (DXA) measures lean tissue

More information

Clinical Densitometry

Clinical Densitometry Volume 8 Number 3 Fall 2005 ISSN: 1094 6950 Journal of Clinical Densitometry The Official Journal of The International Society for Clinical Densitometry Editor-in-Chief Paul D. Miller, MD HumanaJournals.com

More information

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Policy Number: 6.01.40 Last Review: 4/2018 Origination: 4/2005 Next Review: 4/2019 Policy Blue Cross and Blue Shield of Kansas City

More information

Location of body fat and body size impacts DXA soft tissue measures: a simulation study

Location of body fat and body size impacts DXA soft tissue measures: a simulation study (2008) 62, 553 559 & 2008 Nature Publishing Group All rights reserved 0954-3007/08 $30.00 www.nature.com/ejcn ORIGINAL ARTICLE Location of body fat and body size impacts DXA soft tissue measures: a simulation

More information

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA)

Standard Operating Procedure TCRC Dual-Energy X-ray Absorptiometry (DXA) 1. RELEVANCE a. This SOP outlines the instructions to completing Duel Energy X-Ray Absorptiometry (DXA) including: scanning, analysis, review and filing. 2. SCOPE a. This SOP applies to all TCRC RDs. 3.

More information

Quality Control of DXA System and Precision Test of Radio-technologists

Quality Control of DXA System and Precision Test of Radio-technologists J Bone Metab 2014;21:2-7 http://dx.doi.org/10.11005/jbm.2014.21.1.2 pissn 2287-6375 eissn 2287-7029 Review Article Quality Control of DXA System and Precision Test of Radio-technologists Ho-Sung Kim 1,

More information

New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment

New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment Case 1 New Dual-energy X-ray Absorptiometry Machines (idxa) and Vertebral Fracture Assessment (VFA) History and Examination Your wealthy friend who is a banker brings his 62-year-old mother to your office

More information

Weight Loss and Regain and Effects on Body Composition: The Health, Aging, and Body Composition Study

Weight Loss and Regain and Effects on Body Composition: The Health, Aging, and Body Composition Study Journal of Gerontology: MEDICAL SCIENCES The Author 2009. Published by Oxford University Press on behalf of The Gerontological Society of America. Cite journal as: J Gerontol A Biol Sci Med Sci All rights

More information

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi Clinical Utility of Bone Densitometry Diagnosis (DXA)

More information

CLINICAL DENSITOMETRY

CLINICAL DENSITOMETRY JOURNAL OF CLINICAL DENSITOMETRY THE OFFICIAL JOURNAL OF THE INTERNATIONAL SOCIETY FOR CLINICAL DENSITOMETRY EDITOR-IN-CHIEF CLIFFORD J. ROSEN, MD HUMANA PRESS VOLUME 5 NUMBER 4 WINTER 2002 ISSN: 1094

More information

Objectives. Neonatal Nutrition, Growth and Neurodevelopment. Brain Development through Term Gestation

Objectives. Neonatal Nutrition, Growth and Neurodevelopment. Brain Development through Term Gestation Objectives Neonatal Nutrition, Growth and Neurodevelopment Sara Ramel MD Assistant Professor Division of Neonatology University of Minnesota Children s Hospital sramel@umn.edu Influence of growth and nutrition

More information

Bone Densitometry Equipment Operator

Bone Densitometry Equipment Operator Bone Densitometry Equipment Operator The purpose of the Bone Densitometry Equipment Operator Examination, which is made available to state licensing agencies, is to assess the knowledge and cognitive skills

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 48.2 years Height / Weight: 150.0 cm 72.0 kg Sex / Ethnic: Female

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 43.4 years Height / Weight: 170.0 cm 66.0 kg Sex / Ethnic: Female

More information

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare.

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare. Prodigy from GE Healthcare Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world gehealthcare.com Prodigy High performance, efficient and reliable DXA system

More information

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System

Introducing the future of DXA. Powerful images. Clear answers. Horizon DXA System Introducing the future of DXA Powerful images. Clear answers. Horizon DXA System Hologic turns ideas into innovation. Again. Hologic cares about you and your patients about keeping their bones healthy,

More information

2013 ISCD Combined Official Positions

2013 ISCD Combined Official Positions 2013 ISCD Combined Oicial Positions Oicial Positions of the International Society for Clinical Densitometry The International Society for Clinical Densitometry (ISCD) is a not-for-profit multidisciplinary

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 40.2 years Height / Weight: 158.0 cm 52.0 kg Sex / Ethnic: Female Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 50% 40% 30% 20% 20 30 40 50 60 70 80 90 100

More information

Medical Policy. MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Medical Policy. MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Medical Policy MP 6.01.40 BCBSA Ref. Policy: 6.01.40 Last Review: 09/28/2017 Effective Date: 09/28/2017 Section: Radiology End Date: 09/18/2018 Related Policies 6.01.44 Vertebral Fracture Assessment With

More information

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 24.7 years Height / Weight: 8.0 cm 79.0 kg Sex / Ethnic: Male Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 40% 30% 20% 0% 20 30 40 50 60 70 80 90 00 Centile

More information

2013 ISCD Official Positions Adult

2013 ISCD Official Positions Adult 2013 ISCD Official Positions Adult These are the Official Positions of the ISCD as updated in 2013. The Official Positions that are new or revised since 2007 are in bold type. Indications for Bone Mineral

More information

Copyright is owned by the Author of the thesis. Permission is given for a copy to be downloaded by an individual for the purpose of research and

Copyright is owned by the Author of the thesis. Permission is given for a copy to be downloaded by an individual for the purpose of research and Copyright is owned by the Author of the thesis. Permission is given for a copy to be downloaded by an individual for the purpose of research and private study only. The thesis may not be reproduced elsewhere

More information

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p. Dedication Preface Acknowledgments Continuing Education An Introduction to Conventions in Densitometry p. 1 Densitometry as a Quantitative Measurement Technique p. 2 Accuracy and Precision p. 2 The Skeleton

More information

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Susumu Sato 1), Shinichi Demura 2), Tamotsu Kitabayashi

More information

Effect of hand positioning on idxa precision error for total and regional bone and body

Effect of hand positioning on idxa precision error for total and regional bone and body Effect of hand positioning on idxa precision error for total and regional bone and body composition parameters. Abstract Dual energy X-ray absorptiometry (DXA) body composition measurements are performed

More information

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) Date of Interview/Examination (MM/DD/YYYY): A downloadable PDF

More information

Accuracy of DEXA scanning & other methods for determining BMD.

Accuracy of DEXA scanning & other methods for determining BMD. BMD- Measurement Site Accuracy of DEXA scanning & other methods for determining BMD. Ann Larkin In general, densitometry techniques can be performed in either the axial or the appendicular skeleton, depending

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Technical: Patient positioning Performance of the scan Analysis of the data Theoretical: Identification

More information

Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines

Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines International Journal of Obesity (1999) 23, 764±770 ß 1999 Stockton Press All rights reserved 0307±0565/99 $12.00 http://www.stockton-press.co.uk/ijo Short Communication Differences in whole body measurements

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

QDR Series. Discovery and Explorer Advanced Health Assessment

QDR Series. Discovery and Explorer Advanced Health Assessment O S T E O P O R O S I S A S S E S S M E N T QDR Series Discovery and Explorer Advanced Health The Hologic QDR Series bone densitometers combine the proven clinical value of bone mineral density (BMD) measurement

More information

Protein Needs of Preterm Infants: Why Are They So Difficult to Meet?

Protein Needs of Preterm Infants: Why Are They So Difficult to Meet? Protein in the Feeding of Preterm Infants Bhatia J, Shamir R, Vandenplas Y (eds): Protein in Neonatal and Infant Nutrition: Recent Updates. Nestlé Nutr Inst Workshop Ser, vol 86, pp 121 128, (DOI: 10.1159/000442732

More information

1 University of Kansas School of Medicine-Wichita, Department of Pediatrics 2 Wesley Medical Center, Department of Neonatology

1 University of Kansas School of Medicine-Wichita, Department of Pediatrics 2 Wesley Medical Center, Department of Neonatology Impact of on Very Low Birth Weight Infants Siddharthan Sivamurthy, M.D. 1, Carolyn R. Ahlers-Schmidt, Ph.D. 1, Katherine S. Williams, M.Ed. 1, Jared Shaw 2, Paula Delmore, M.S.N. 2, Barry T. Bloom, M.D.

More information

Timing and tempo of first year growth in relation to cardiovascular and metabolic risk profile in early adulthood

Timing and tempo of first year growth in relation to cardiovascular and metabolic risk profile in early adulthood Note: for non-commercial purposes only Timing and tempo of first year growth in relation to cardiovascular and metabolic risk profile in early adulthood Anita Hokken-Koelega Professor of Pediatric Endocrinology

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

BODY COMPOSITION AS MEASURED BY IN VIVO NEUTRON ACTIVATION ANALYSIS

BODY COMPOSITION AS MEASURED BY IN VIVO NEUTRON ACTIVATION ANALYSIS BNL..26819 BODY COMPOSITION AS MEASURED BY IN VIVO NEUTRON ACTIVATION ANALYSIS S. H. Cohn, A. Sawitsky*, D. Vartsky, S. Yasumura, I. Zanzi > W. Gartenhaus and K. J. Ellis Medical Research Center Brookhaven

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

Impact of infant feeding on growth trajectory patterns in childhood and body composition in young adulthood

Impact of infant feeding on growth trajectory patterns in childhood and body composition in young adulthood Impact of infant feeding on growth trajectory patterns in childhood and body composition in young adulthood WP10 working group of the Early Nutrition Project Peter Rzehak*, Wendy H. Oddy,* Maria Luisa

More information

Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery

Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery Body Composition Changes Following a Very Low Calorie Pre-Operative Diet in Patients Undergoing Bariatric Surgery 03 September 2018 Jonathan Sivakumar, Lynn Chong, Salena Ward, Tom R Sutherland, Michael

More information

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status Journal of Clinical Densitometry, vol. 10, no. 3, 239e243, 2007 Ó Copyright 2007 by The International Society for Clinical Densitometry 1094-6950/07/10:239e243/$32.00 DOI: 10.1016/j.jocd.2007.03.002 Original

More information

Lunar idxa. The intelligent DXA. gehealthcare.com

Lunar idxa. The intelligent DXA. gehealthcare.com Lunar idxa The intelligent DXA gehealthcare.com The best of DXA technology for bone and metabolic health assessment With a state-of-the-art design, Lunar idxa offers research-grade image resolution and

More information

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Policy Number: 6.01.40 Last Review: 4/2014 Origination: 4/2005 Next Review: 4/2015 Policy Blue Cross and Blue Shield of Kansas

More information

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation Recent citations - Munawar A Riyadi et al To cite this article:

More information

The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers

The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers Original Article http://dx.doi.org/10.3349/ymj.2012.53.2.289 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 53(2):289-293, 2012 The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density

More information

TOTAL BODY COMPOSITION

TOTAL BODY COMPOSITION Specialists in Joint Replacement, Spinal Surgery, Orthopaedics and Sport Injuries DEXA TOTAL BODY COMPOSITION www.sportsurgeryclinic.com WHY CHOOSE DEXA TOTAL BODY COMPOSITION? DEXA, short for Dual Energy

More information

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition. Original Policy Date 12:2013

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition. Original Policy Date 12:2013 MP 6.01.28 Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Medical Policy Section Radiology Is12:2013sue 12:2012 Original Policy Date 12:2013 Last Review Status/Date reviewed

More information

A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution

A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution Ronenn Roubenoff, Lauren Weiss, Ann McDermott, Tanya Heflin, Gregory J. Cloutier, Michael Wood and

More information

Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients

Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients OBES SURG (2014) 24:1476 1480 DOI 10.1007/s11695-014-1190-5 OTHER Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients Silvia L. Faria

More information

Post Discharge Nutrition. Jatinder Bhatia, MD, FAAP

Post Discharge Nutrition. Jatinder Bhatia, MD, FAAP Post Discharge Nutrition Jatinder Bhatia, MD, FAAP Declaration of potential conflicts of interest Regarding this presentation the following relationships could be perceived as potential conflicts of interest:

More information

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm Interpreting DEXA Scan and the New Fracture Risk Assessment Algorithm Prof. Samir Elbadawy *Osteoporosis affect 30%-40% of women in western countries and almost 15% of men after the age of 50 years. Osteoporosis

More information

The Assessment of Body Composition in Health and Disease

The Assessment of Body Composition in Health and Disease The Assessment of Body Composition in Health and Disease Giorgio Bedogni, Paolo Brambilla, Stefano Bellentani and Claudio Tiribelli CHAPTER 3 BODY AND NUTRITIONAL STATUS Nutritional status can be operationally

More information

Determination of Chicken Body Composition Measured by Dual Energy X-Ray Absorptiometry

Determination of Chicken Body Composition Measured by Dual Energy X-Ray Absorptiometry International Journal of Poultry Science 11 (7): 46-468, 1 ISSN 168-8356 Asian Network for Scientific Information, 1 Determination of Chicken Body Composition Measured by Dual Energy X-Ray Absorptiometry

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 29.5 years Height / Weight: 156.0 cm 57.0 kg Sex / Ethnic: Female

More information

Bone Density Measurement in Women

Bone Density Measurement in Women Bone Density Measurement in Women Revised 2005 Scope This guideline defines the medical necessity of bone mineral density (BMD) measurement using dualenergy x-ray absorptiometry (DXA or DEXA), and applies

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Postnatal growth failure Causes, consequences and prevention

Postnatal growth failure Causes, consequences and prevention Postnatal growth failure Causes, consequences and prevention Bielsko March 9, 2014 Ekhard E. Ziegler, M.D. Fomon Infant Nutrition Unit University of Iowa 1995; 26-29 weeks gestation NICHD Growth Observational

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA202 Rad protection class two True/False Indicate whether the sentence or statement is true or false. 1. Secondary radiation comes from scatter and leakage. 2. Grids are considered a protection device.

More information

Lactation, weaning, and calcium supplementation: effects on body composition in postpartum women 1 3

Lactation, weaning, and calcium supplementation: effects on body composition in postpartum women 1 3 Lactation, weaning, and calcium supplementation: effects on body composition in postpartum women 1 3 Karen S Wosje and Heidi J Kalkwarf ABSTRACT Background: Concern that long-term weight retention after

More information

PROFESSIONAL FITNESS & WELLBEING CATALOGUE

PROFESSIONAL FITNESS & WELLBEING CATALOGUE PROFESSIONAL FITNESS & WELLBEING CATALOGUE CONTENTS BENEFITS OF BODY COMPOSITION GOOD, BETTER, BEST TANITA BIA TECHNOLOGY MULTI FREQUENCY BODY COMPOSITION ANALYSER BODY COMPOSITION ANALYSERS WEIGHING SCALES

More information

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA ORIGINAL ARTICLE Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA Leila Amiri 1, Azita Kheiltash 2, Shafieh Movassaghi 1, Maryam Moghaddassi 1, and Leila Seddigh 2 1 Rheumatology

More information

Maternal and Infant Nutrition Briefs

Maternal and Infant Nutrition Briefs Maternal and Infant Nutrition Briefs November/December 2002 A research-based newsletter prepared by the University of California for professionals interested in maternal and infant nutrition Recent Trends

More information

MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Medical Policy MP 6.01.40 BCBSA Ref. Policy: 6.01.40 Last Review: 09/19/2018 Effective Date: 09/19/2018 Section: Radiology Related Policies 6.01.44 Vertebral Fracture Assessment With Densitometry 9.01.502

More information

(1 280 ± 286) g; (1 436 ± 201) g

(1 280 ± 286) g; (1 436 ± 201) g 259 ( 100730) : 34 1 800 g 24 (, 11 ) (, 13 ) 50%, 100 ml / (kg d) FM85, ; (30.6 ± 2.9), (1 280 ± 286) g; (31.6 ± 1.9), (1 436 ± 201) g, 81.6%, 34.1 24.6 d [18.9 vs 17.1 g / (kg d), P = 0.364] (1.16 vs

More information

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men Osteoporos Int (1999) 10:265 270 ß 1999 International Osteoporosis Foundation and National Osteoporosis Foundation Osteoporosis International Original Article Bone Mineral Density and Vertebral Fractures

More information

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry J Bone Metab 2017;24:105-109 https://doi.org/10.11005/jbm.2017.24.2.105 pissn 2287-6375 eissn 2287-7029 Original Article Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

More information

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents British Journal of Nutrition (2003), 90, 987 992 q The Authors 2003 DOI: 10.1079/BJN2003983 Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and

More information

BONE DENSITY AS A SOURCE OF ERROR MEASURING. BODY COMPOSITION WITH THE BOD POD and idxa IN FEMALE RUNNERS THESIS. the Degree Master of Science in the

BONE DENSITY AS A SOURCE OF ERROR MEASURING. BODY COMPOSITION WITH THE BOD POD and idxa IN FEMALE RUNNERS THESIS. the Degree Master of Science in the BONE DENSITY AS A SOURCE OF ERROR MEASURING BODY COMPOSITION WITH THE BOD POD and idxa IN FEMALE RUNNERS THESIS Presented in Partial Fulfillment of the Requirements for the Degree Master of Science in

More information

Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine

Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine HENZ W. WAHNER, M.D., WLLAM L. DUNN, M.Sc, MANUEL L. BROWN, M.D., Section

More information

Twelve percent of children under 2 years1 and 10% of 2 5 year olds are obese

Twelve percent of children under 2 years1 and 10% of 2 5 year olds are obese Fields Insights From Body Composition Studies David Fields, PhD Twelve percent of children under 2 years1 and 10% of 2 5 year olds are obese (National Health and Nutrition Examination Survey; body mass

More information

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years Europe PMC Funders Group Author Manuscript Published in final edited form as: Int J Obes (Lond). 2010 April ; 34(4): 649 655. doi:10.1038/ijo.2009.249. Evaluation of DXA against the four-component model

More information

Use of dual X-ray absorptiometry to measure body mass during short- to medium-term trials of nutrition and exercise interventions

Use of dual X-ray absorptiometry to measure body mass during short- to medium-term trials of nutrition and exercise interventions bs_bs_banner Special Article Use of dual X-ray absorptiometry to measure body mass during short- to medium-term trials of nutrition and exercise interventions Christopher PF Marinangeli and Amira N Kassis

More information