A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement

Size: px
Start display at page:

Download "A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement"

Transcription

1 European Journal of Clinical Nutrition (1997) 51, 312±317 ß 1997 Stockton Press. All rights reserved 0954±3007/97 $12.00 A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement CD Economos 1, ME Nelson 1, MA Fiatarone 1, GE Dallal 1, SB Heyms eld 2, J Wang 2, S Yasumara 3, RMa 3, AN Vaswani 4, M Russell-Aulet 5 and RN Pierson 2 1 Jean Mayer USDA Human Nutrition Research Center on Aging at Tufts University, Boston, MA; 2 Body Composition Unit at St. Luke's- Roosevelt Hospital, New York, NY; 3 Brookhaven National Laboratory, Upton, NY; 4 Osteoporosis Research Clinic, Winthrop University Hospital, Mineola, NY and 5 The University of Michigan, Ann Arbor, MI, USA Objective: To assess intra- and inter-site soft tissue variability by dual energy X-ray absorptiometry (DXA). Design: Cross-sectional trial. Setting: Three medical research institutions. Subjects: Five humans (in vivo) and four phantoms (in vitro), con gured from two whole body phantoms with arti cial skeletons and thickness overlays. Interventions: Duplicate total-body DXA scans were performed on all subjects at each institution within a 15 d period. Results: All intra-site coef cients of variation (CV) were < 0.5% for total tissue mass, but in vitro and in vivo Cvs were 7.2% and 2.3% for fat mass (FM) and 2.5% and 0.9% for lean mass (LM), respectively. Several totalbody and regional FM and LM measurements were signi cantly different between sites (P <0.05), with percent differences between sites ranging from 2.6±13.3% for FM and from 1.6±13.6% for LM. Site 2 was consistently lower for FM and Site 3 was consistently lower for LM. Conclusions: These results stress the need for both rigorous and standardized cross-calibration procedures for soft tissue measurement by DXA. Sponsorship: This work has been supported in part by NIH Training Grant #T32AG00209, grant P01-DK42618 from the National Institutes of Health, federal funds from the US Department of Agriculture, and Agricultural Research Services contract 53-3K Dr Nelson is currently a Brookdale National Fellow. Descriptors: dual energy X-ray absorptiometry; DXA; densitometry; body composition; lean mass; fat mass. Introduction The nutritional assessment of an individual is enhanced tremendously by the accurate measurement of body composition. Since their derivation in the early 1980s, dual energy projection methods have been used to quantify bone mass and assess fracture risk. Now the prevailing technique, dual energy X-ray absorptiometry (DXA), also measures total-body soft tissue and quanti es the fat and lean components. [The acronym DXA is used throughout this text as recommended by Wilson (Wilson et al, 1990).] DXA is non-invasive and when performed properly it is safe and well tolerated, characterizing it as a promising technique for extensive use in body composition research. However, individual variability in the composition of soft tissue as well as DXAs dif culty in precisely partitioning the soft tissue into lean and fat compartments are recognized obstacles for DXAs routine use (Roubenoff et al, 1993). Therefore, further validation is required before DXA Correspondence: Dr ME Nelson, HNRC, 711 Washington Street, Tufts University, Boston, MA 02111, USA. The contents of this publication do not necessarily re ect the views or policies of the US Department of Agriculture, neither does the mention of trade names, commercial products, or organizations imply endorsement by the US government. Portions of this research were presented at the American Society for Clinical Nutrition Meeting May, 1995 in San Diego, CA. Received 13 September 1996; revised 18 December 1996; accepted 13 January should be accepted as a standard measuring technique for body composition. The reliability of DXA depends upon operator-, subject-, and apparatus-factors. The precision of individual DXA instruments for the assessment of bone is excellent, with coef cients of variation (CV) < 1.0% (Johnson et al, 1991; LeBlanc et al, 1990; Mazess et al, 1989; Nuti et al, 1987; Fuller et al, 1992; Compston et al, 1992). On the other hand, Cvs for soft tissue measurements of up to 3.1% for lean tissue, 8.6% for fat tissue, and 7.9% for percent body fat have been observed (Table 1) Mazess et al, 1990; Johnson et al, 1991; Fuller et al, 1992; Compston et al, 1992; Heyms eld et al, 1989; Kelly et al, 1991; Slosman et al, 1992; Haarbo et al, 1991; Pritchard et al, 1993; Rico et al, 1994). The difference in precision between bone and soft tissue is not surprising since DXA was essentially designed to assess bone and soft tissue is heterogeneous. Nevertheless, assumptions that bone precision extends to soft tissue precision are often made erroneously. DXAs widespread availability and relatively inexpensive operation have made it an appealing technique for body composition assessment in multi-center research trials. Cross-validation studies of instruments at different sites are necessary in order to directly compare body composition data generated by different instruments. However, the few studies in the literature examining the agreement between DXA instruments from the same manufacturer, located at different clinical centers, have only

2 Table 1 Summary of intra-site DXA soft tissue precision A. Short-term 313 Author Instrument Duration Subjects (n) Scans/subject CV (%) Fat Lean % Fat Heyms eld 1989 Lunar DPX 5 days 4 5 Ð Ð 0.50 Mazess 1990 Lunar DPX 5±7 days Fuller 1992 Lunar DPX 1 day Ð Compston 1992 Lundar DPX 1 day Ð Pritchard 1993 Hologic QDR days Kelly 1991 Hologic QDR day 13 2± Rico 1994 Norland XR-26 1±16 weeks 10 2 Ð Economos (current study) B. Long-term Lundar DPX 1 day Lunar DPX 1 day Lunar DPX-L 1 day Author Instrument Duration Subjects (n) Scans/subject CV (%) Fat Lean % Fat Johnson 1991 Lunar DPX 1 and 9 months Ð Haarbo 1991 Lunar DPX 1 and 6 months Slosman 1992 Hologic QDR and 12 months Ð Rico 1994 Norland XR-26 1±24 weeks 8 6 Ð examined bone measurements (Kelly et al, 1989; Blake et al, 1991; Orwoll et al, 1991; Rencken et al, 1991; Trevisan et al, 1992; Economos et al, 1996; Wahner et al, 1994). To date, no studies have reported precision in soft tissue between sites, using the same manufacturer, with the exception of a brief report (Paton et al, 1995). Since it is not appropriate to assume that the level of agreement for bone, using instruments from the same manufacturer at different sites would be the same for soft tissue, we conducted a multi-center comparison of DXAs for in vivo and in vitro soft tissue. In a three-site research project that proposed to pool and analyze data cumulatively, the performance of instruments common to each of the sites was investigated to verify the agreement in measurements of total body weight, total soft tissue mass (TM), fat tissue mass (FM), and lean tissue mass (LM) within and between sites. Methods For a more detailed description of the methods please refer to a previously published paper (Economos et al, 1996). Research sites Within 15 d, all measurements were performed at three research sites. The centers were: (site 1) The Jean Mayer USDA Human Nutrition Research Center on Aging at Tufts University, Boston, MA, (site 2) St. Lukes Roosevelt Hospital Center, New York, NY, and (site 3) Winthrop University Hospital, Mineola, NY. Instruments DXA measurements were made with Lunar DPX (sites 1 and 2), software version 3.6z, and DPX-L (site 3) software version 1.3z, which, according to the manufacturer, are analogous. The differential absorption of two X-ray energies allows two components, differing in electron density, to be determined simultaneously (Mazess et al, 1990). The soft tissue mass of the whole body is determined by measuring the R value, the ration of high energy to low energy attenuation, for each pixel that contains no signi cant bone, and extrapolating for soft tissue content in bone-containing image pixels (Mazess et al, 1990; Peppler and Mazess, 1981). Theoretically, DXA measures three tissues: bone mineral, fat, and lean (fat-free mineral-free tissue). Human subjects (in vivo) and phantoms (in vitro) Five healthy subjects (3M, 2F) varying in size (53.6± 99.5 kg body weight) and age (28±61 y of age) and two sizes of whole body polyethylene Bottle Mannequin Absorption Phantoms (BOMAB), with overlays, from Atlan-Tech, Atlanta, GA. Were used in the study. The phantom sizes approximate a reference man (66.4 kg), and a small reference woman (49.9 kg) (Kramer et al, 1991). The mass attenuation coef cients (m/r) used in calculations of photon (X-ray) penetration in biological materials were similar for the materials used and the tissues that were simulated at photon energies of 40 and 70 kev (Hubbell 1982). When the overlays were lled with water and laid over the water- lled phantoms, they represented a large man of 97.3 kg and a large woman of 78.5 kg and were used to determine whether any changes in attenuation occurred in larger subjects, which may require correction. Protocol Within a two day period at each research site, two totalbody scans were performed in immediate succession, after repositioning of each subject and phantom con guration (with and without overlay). The protocol for obtaining and analyzing data was standardized. For each total-body scan, the total and regional (appendicular and trunk), fat mass (FM), lean mass (LM) (fat-free, mineral-free), total tissue mass (TM) (fat lean masses), and total body weight (fat lean bone masses) were

3 314 Table 2 Comparison of tissue weight by scale and by DXA for all sites A. In vivo Scale weight (kg) b DXA weight a (kg) c DXA vs scale correlation Mean d (kg) s.d. d RC Site 1 n ˆ r-0.09; P ˆ Site 2 n ˆ r ˆ 0.99 P ˆ Site 3 n ˆ r ˆ 0.99; P ˆ Values are presented as mean s.d. a Total body weight ˆ fat mass lean mass bone mineral content. b NS (between sites) by ANOVA.#15; c Site 1 > Site 2 and 3 by ANOVA. d ˆ [DXA weight (g) (fat mass lean mass bone mineral content) Ð Scale weight (g)]. Mean d ˆ [( Pd )/n]. s.d. d ˆ Standard deviation of the differences ˆ square root of [( Pd )/n]. RC ˆ 2s.d. d B. In vitro Scale weight (kg) b DXA weight a (kg) c DXA vs scale correlation Mean d (kg) s.d. d RC Site n ˆ r ˆ 0.99; P ˆ Site 2 n ˆ r ˆ 0.99; P ˆ Site 3 n ˆ r ˆ 0.99; P ˆ Values are presented as mean s.d. a Total body weight ˆ fat mass lean mass bone mineral content. b NS (between sites) by ANOVA. c Site 2 > Site 3 by ANOVA. d ˆ [DXA weight (g) (fat mass lean mass bone mineral content) - Scale weight (g)]. Mean d ˆ [( d )/n]. s.d. d ˆ Standard deviation of the differences ˆ square root of [( d )/n]. RC ˆ 2s.d. d. measured. We report lean mass, which does not include the bone mineral, although there were no differences in the analyses whether the bone mineral data were included or excluded. The bone results from the study have been previously reported (Economos et al, 1996). Statistical analysis Total body weight by DXA, calculated from the average of measurements 1 and 2, was compared with total body weight by scale within sites using the Bland and Atlman method (Bland and Atlman, 1986). Between site comparisons of total body weight by DXA and total body weight by scale was done using Analysis of variance (ANOVA) with Tukeys HSD (honesty signi cant difference) for both groups [humans (n ˆ 5) and phantoms (n ˆ 4)]. The signi cance level was set at P < Intra-site reproductiblity was calculated as the coef cient of variation (CV) from duplicate TM, FM, and LM measurements by the formula from Nilas (Nilas et al, 1988). The CV provides a measure that can be compared across studies of measurement precision (Table 1). For inter-site comparison, the site means and standard deviations of the total and regional (appendicular and trunk), and LM were calculated from the average of measurements 1 and 2. ANOVA was used to test for signi cant differences between sites and Tukeys HSD was used to compare sites and adjust for multiple comparisons. Results were considered signi cant if P < The inter-site Cvs for FM and LM were calculated as: (CV ˆ s.d./mean 6 100), where the mean ˆ site 1 site 2 site 3/3. The mean percentage difference (MPD) between sites was calculated for FM and LM as MPD (%) ˆ [(site 1 7 site 2j)=(site 1 site 2/2)] All data were analyzed using the SAS statistical package, version 6.06 (SAS 1990). Results Total body weight comparisons Tables 2a and 2b show the in vivo and in vitro comparisons, both within and between sites, for total body weight by scale and by DXA. Within sites, total body weight by scale and by DXA were highly correlated (r ˆ 0.99, P < ), but the Bland and Altman method showed poor agreement between body weight by scale and by DXA for sites 1 and 3 in vivo and sites 1, 2, and 3 in vitro. There were no signi cant differences between sites for total body weight by scale, but there were signi cant differences in vivo and in vitro in the total body weight by DXA. Intra-site comparisons For total-body TM, Cvs were < 0.5% with ranges of 0.3± 0.6% for in vivo and 0.3±0.7% for in vitro. Total-body FM Cvs ranged from 1.0±3.5% in vivo and 3.1±10.7% in vitro and total body LM Cvs ranged from 0.7±1.1% in vivo and 1.0±3.5% in vitro. Inter-site comparisons Inter-site variability of total FM and LM, expressed as Cvs between the three sites, is depicted in Figure 1. The two human subjects that were at the extreme ends of the body size spectrum (subject 1 ˆ 1 largest and subject 5 ˆ smallest), had the highest inter-site Cvs for FM (1 ˆ 9.2%, 5 ˆ 11.0%) and LM (1 ˆ 2.9% and 5 ˆ 3.6%). There were no trends related to size for phantom precision. Figure 2 shows the in vivo and in vitro comparisons of the FM and LM between sites. The in vivo total FM from site 2 was signi cantly lower than sites 1 and 3 (P < 0.05) by 13.3% and 12.0%, respectively. The in vivo total LM from site 3 was signi cantly lower than sites 1 and 2 (P < 0.05) by 3.5% and 4.9%, respectively, and the in vitro LM from site 3 was signi cantly lower than site 2

4 315 Figure 1 Inter-site precision of DXA soft tissue measurements. CV ˆ coef cient of variation. The values represent the fat and lean mass Cvs [(s.d./mean) 6 100] calculated for each individual human (in vivo, n ˆ 5) and phantom (in vitro, n ˆ 4) across the three sites. The means are shown for comparison between in vivo and in vitro results and fat and lean results. The mean inter-site CVs were 7.5 and 7.9% for in vivo and in vitro fat mass, and 2.7% and 4.1% for in vivo and in vitro lean mass. Figure 3 In vivo and in vitro regional soft tissue by DXA. The values represent the means s.e.m. for appendicular (arms legs) and trunk fat and lean tissue in grams for each site, both in vivo and in vitro. ANOVA and Tukeys HSD were used to test for differences between sites (P < 0.05). Figure 2 In vivo and in vitro total body soft tissue by DXA. The values represent the means s.e.b. for total-body fat and lean tissue in grams for each site, both in vivo and in vitro. ANOVA and Tukeys HSD were used to test for differences between sites (P < 0.05). (P < 0.05) by 13.6%. To further examine the differences between sites, Figure 3 shows the regional (appendicular and trunk) comparisons of FM and LM from each site. Discussion An ample amount of work has been done to investigate the intra- and inter-site precision of DXA for bone (Mazess et al, 1990; Johnson et al, 1991; LeBlanc et al, 1990; Mazess et al, 1989; Nuti et al, 1987; Fuller et al, 1992; Compston et al, 1992; Kelly et al, 1991; Economos et al, 1996; Seto et al, 1991). Within the last ve years, quite a few studies examining soft tissue intra-site precision have been published (see Table 1). While most of the total-body lean tissue Cvs would be considered acceptable, fat tissue Cvs are not. Furthermore, regional Cvs are generally discouraging, with reports of regional lean tissue measurements ranging from 1.3±4.2% and regional fat tissue measurements ranging from 2.4±12.6% (Mazess et al, 1990; Johnson et al, 1991; Fuller et al, 1992). The incongruous ndings from these studies do not support the use of this technique for body composition measurement as a gold standard. Compared to the precision of other body composition techniques ( 1% for muscle tissue from computed tomography, 1.5% for lean and fat tissue for hydrodensitometry, 7% for skinfolds, 2% for bioelectric impedance (Nelson et al, 1996)] DXA is comparable to the less sophisticated techniques. Nevertheless, DXA is currently being used as a standard technique for the assessment of body composition and to monitor the conventional change in body composition over time or as a result of an intervention. Unlike bone, standard monitoring of soft tissue precision using phantoms of known composition is relatively limited, and most often small, partial body phantoms are used (Kelly et al, 1991; Haarbo et al, 1991; Jensen et al, 1993; Laskey et al, 1991). The soft tissue CVs in these studies vary dramatically, from 0.3±7.0%, due to the various phantom sizes and thicknesses and the variability of the fat and lean equivalent materials. Using weight stable, whole-body phantoms with invariable body compositions we were able to show that intra-site reproducibility for total-body TM is highly acceptable and similar to results seen in humans. In fact, much of the data supporting DXAs ability to accurately measure soft tissue is based on evidence that total tissue mass by DXA correlates signi cantly to scaled weight (Gotfredsen et al, 1986; Ellis et al, 1994; Johansson et al, 1993; Dawson- Hughes et al, 1992; Svendsen et al, 1991; Going et al, 1993). However, simple correlation analyses can be misleading because they do not address the reproducibility, only the linear association. This is supported by our data, shown in Table 2, where correlationss between DXA weight and scale weight were very strong, nevertheless

5 316 we found signi cant differences between method, indicating disagreement in the results. Still, as made evident by a review of the literature, researchers often draw erroneous conclusions regarding DXAs ability to measure body composition merely based on evidence that total tissue weight reconstruction is satisfactory; a phenomenon which provides no direct evidence about its accuracy for measuring the individual masses of lean, fat, or bone. Our data supports the argument that DXAs ability to predict total tissue weight is constant, but its ability to compartmentalize FM and LM is variable. Over the last few years, a number of studies have begun to investigate soft tissue accuracy by DXA, (Pierson et al, 1991; Withers et al, 1992; Jensen et al, 1993; Ellis et al, 1994; Svendsen et al, 1993; Jebb et al, 1995). Nonetheless, until a number of comparisons using well-developed measurement techniques are repeated, accuracy cannot be con rmed. Based on our calculations using physical tables published by Hubbel (Hubbell, 1982), DXA principles (Mazess et al, 1990), and the experience of others (Kelly et al, 1991; Laskey et al, 1991), we expected the materials used to construct our phantoms to be detected as the same proportion of LM and FM with repeat scanning. However, compared to humans, the LM and FM of the phantoms were poorly reproduced; CVs up to 2.5 for LM and up to 10.7 for FM. Since our results were not consistent with our presumptions, we resolved that these phantoms do not have much potential for use in maintaining intra-site quality control for DXA scanning in the total-body mode. The CVs reported here are clearly un t for use in research and underscores the need for the development of a soft tissue phantom for total-body DXA studies. The question of whether similar DXA instruments at various centers demonstrate acceptable agreement for soft tissue has only once before been reported, which was a letter to the editor (Paton et al, 1995). Despite excellent agreement for TM between identical instruments, the authors reported an average 3.0 kg and 3.9 kg difference for fat and lean tissue, respectively. Our nding that the inter-site precision of FM was less favorable than that of LM is consistent with intra-site ndings by the current study and others (Table 1), which all report better precision for LM than for FM. Signi cant disagreements among instruments, as reported in this cross-sectional study as well as Paton et al, 1995, would make it dif cult to compare the results obtained in different laboratories and would interfere seriously with the use of DXA for analysis of soft tissue in both research and clinical medicine. Furthermore, our results indicate that single-and multicenter intervention studies should be extremely concerned about this variability since the expected changes in FM and LM may be less than the variability within and between sites, which could confound the results from longitudinal studies. It is also important to note that the present day study was performed under very stringent conditions. We used one analyzer, weight-stable humans and whole-body phantoms, internally well-controlled research facilities, and DXA instruments of the same manufacturer, yet there were still discrepancies, indicating that differences in the machines calibration was the likely source of variability. Therefore, the development and availability of appropriate quality control soft tissue phantoms that would allow the con dent interpretation of DXA from serial patient measurements cannot be overstated, especially for situations where soft tissue thicknesses change over time, as in growth, weight loss or gain, illness and disease. Because of the proprietary nature of the DXA it is not possible to fully speculate the reasons for the imprecision of the technique. However, several factors are most likely responsible: The technique was not developed for extreme body thicknesses; partial body phantoms varying in size, thickness and soft tissue equivalent materials have been used to determine and monitor precision; human soft tissue is heterogeneous in nature; and the technique must extrapolate for soft tissue content in 40% of the body area where there are bone containing image pixels. It is imperative for scientists to work closely with the DXA manufacturers to develop this promising technique to its maximum potential. Conclusions Our results show that very similar or identical DXA instruments may produce signi cant inter- and intra-site differences in soft tissue measurements with human subjects and whole-body phantoms. In addition, whole-body phantoms can be used for multi-center cross-comparison studies when the identi cation of differences between sites is essential and when the transfer of human subjects between sites is not feasible. Nevertheless, for intra- and inter-site measurements to be truly comparable, measurements with a group of common in vivo and in vitro phantoms are suggested. These results stress the need for both rigorous and standardized cross-calibration procedures for soft tissue measurement by DXA. AcknowledgementsÐWe would like to thank the members of the Calcium and Bone Metabolism Laboratory at The Jean Mayer USDA HNRC at Tufts University, The Bone Density and Body Composition Laboratories at St. Luke-Roosevelt Hospital and the Bone Laboratory at Winthrop University Hospital. References Blake GM, Tong CM & Fogelman I (1991): Intersite comparison of the Hologic QDR-1000 dual energy x-ray bone densitometer. Br. J. Radiol. 64, 440±446. Bland JM & Altman DGM (1986): Statistical methods for assessing agreement between two methods of clinical measurement. The. Lancet. 2, 307±310. Compston JE, Bhambhani M, Laskey MA, Murphy S & Khaw KT (1992): Body composition and bone mass in post-menopausal women. Clin. Endocrinol. 37, 426±431. Dawson-Hughes B & Harris S. (1992): Regional changes in body composition by time of year in healthy postmenopausal women. Am. J. Clin. Nutr 56, 307±313. Economos CD, Nelson ME, Fiatarone MA, Dallal GE, Heyms eld SB, Russell-Aulet M, Yasumura S, Ma R, Vaswani AN, Pierson RN & Wang J (1996): A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro measurements of bone mineral content and density. J. Bone. Min. Res. 11, 275±285. Ellis KJ, Shypailo RJ, Pratt J & Pond WG (1994): Accuracy of dual-energy x-ray absorptiometry for body composition measurements in children. Am. J. Clin. Nutr. 60, 660±665. Fuller NJ, Jebb SA, Laskey MA, Coward WA & Elia M (1992): Fourcomponent model for the assessment of body composition in humans: comparison with alternative methods, and evaluation of the density and hydration of fat-free mass. Clin. Sci. 82, 687±693. Fuller NJ, Laskey MA & Elia M (1992): Assessment of the composition of major body regions by dual-energy x-ray absorptiometry (DEXA) with special reference to limb muscle. Clin. Physiol. 12, 253±256. Going SB, Massett MP, Hall MC, Bare LA, Root PA, Williams DP & Lohman TG (1993): Detection of small changes in body composition by dual-energy x-ray absorptiometry. Am. J. Clin. Nutr. 57, 845±850. Gotfredsen A, Jensen J, Borg J & Christiansen C (1986): Measurement of lean body mass and total fat using dual photon absorptiometry. Metabolism 35, 88±93. Haarbo J, Gotfredsen A, Hassager C, Christiansen C (1991): Validation of body composition by dual energy X-ray absorptiometry (DEXA). Metabolism. 11, 331±341. Hangartner TN & Johnston CC (1990): In uence of fat on bone measure-

6 ments with dual-energy absorptiometry. Bone. and. Mineral. 9, 71±81. Heyms eld SB, Wang J, Heshka S, Keyayias JJ & Pierson RN (1989): Dual-photon absorptiometry: comparison of bone mineral and soft tissue mass measurements in vivo with established methods. Am. J. Clin. Nutr. 49, 1283±1289. Heyms eld SB, Wang J, Kehayias J, Heshka S, Lichtman SB & Pierson RN (1989): Chemical determination of human body density in vivo: relevance to hydrodensitometry. Am. J. Clin. Nutr. 50, 1282±1290. Hubbell JH (1982): Photon mass attenuation and energy-absorption coef cients from 1 kev to 20 MeV. Int. J. Appl. Rad. Isot. 33, 1269± Jebb AS, Goldberg GR, Jennings G & Elia M (1995): Dual-energy X-ray absorptiometry measurements of body composition: effects of depth and tissue thickness, including comparisons with direct analysis. Clin. Sci. 88, 319±324. Jensen MD, Kanaley JA, Roust LR, OBrien PC, Braun JS, Dunn WL & Wahner HW (1993): Assessment of body composition with use of Dual- Energy X-ray Absorptiometry: Evaluation and comparison with other methods. Mayo. Clin. Proc. 68, 867±873. Johansson AG, Forslund A, Sjodin A, Mallmin H, Hambraeus L & Ljunghall S (1993): Determination of body composition ± a comparison of dual-energy x-ray absorptiometry and hydrodensitometry. Am. J. Clin. Nutr. 57, 323±326. Johnson J & Dawson-Hughes B (1991): Precision and stability of Dual- Energy X-ray Absorptiometry Measurements. Calc. Tiss. Inter. 49, 174±178. Kramer GH, Burns L & Noel L (1991): The BRMD BOMAB phantom family. Health. Phys. 61, 895±902. Kelly TL, Richardson TL, Berger N, Steiger P & Stein, JA (1991): Assessment of body composition by dual x-ray absorptiometry using a simultaneously scanned calibration standard. Med. Sci. Sports. Exerc. 23, Suppl. (4), S90 (abstr.). Kelly TL, Slovik DM, Shoenfeld DA & Neer RM (1980): Calibration and standardization of bone mineral densitometers. J. Bone. Miner. Res. 4, 663±669. Laskey MA, Lyttle KD, Flaxman ME & Barber RW (1992): The in uence of tissue depth and composition on the performance of the Lunar dualenergy X-ray absorptiometer whole-body scanning mode. Eur. J. Clin. Nutr 46, 39±45. Laskey MA, Flaxman ME, Barber RW, Trafford S, Hayball MP, Lyttle KD, Crisp AJ & Compston JE (1991): Comparative performance in vitro and in vivo of lunar DPX and Hologic QDR-100 dual energy x-ray absorptiometers. Br. J. Radiol. 64, 1023±1029. LeBlanc AD, Schneider VS, Engelbretson DA & Evans HJ (1990): Precision of regional bone mineral measurements obtained from totalbody scans. J. Nucl. Med. 31, 43±45. Mazess R, Barden HS, Bisek JP & Hanson J (1990): Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue composition. Am. J. Clin. Nutr. 51, 1106±1112. Mazess R, Collick B, Trempe J, Barden H & Hanson J (1989): Performance evaluation of a dual-energy X-ray Bone Densitometer. Calc. Tiss. Inter. 44, 228±232. Nelson ME, Fiatarone MA, Layne JE, Trice I, Economos CD, Fielding RA, Ma R, Pierson, RN & Evans WJ (1996): Analysis of bodycomposition techniques and models for detecting change in soft tissue with strength training. Am. J. Clin. Nutr. 63, 678±686. Nilas L, Hassager C & Christiansen C (1988): Long term precision of dual photon absorptiometry in the lumbar spine in clinical settings. Bone. and. Mineral. 3, 305±315. Nuti R, Righi G, Martini G, Turchetti V, Lepore C & Canaggia A (1987): Methods and clinical applications of total body absorptiometry. J. Nucl. Med. 31, 213±221. Orwoll E, Oviatt S and the Nafarelin bone study group (1991): Longitudinal precision of Dual-Energy X-ray Absorptiometry in a multicenter study. J. Bone. Miner. Res. 6, 191±197. Paton NIJ, Macalian DC, Jebb SA, Pazianes M, Grif n GE (1995): Dualenergy X-ray absorptiometry results differ between machines. Lancet. 346, 899. Peppler WW & Mazess RB (1981): Total body bone mineral and lean body mass by dual-photon absorptiometry. Theory and measurement procedure. Calc. Tiss. Inter. 33, 353±359. Pierson RN, Wang J, Heyms eld SB, Russell-Aulet M & Mazariegos M (1991): Measuring body fat: calibrating the rulers. Intermethod comparisons in 389 normal Caucasian subjects. Am. J. Physiol. 261, (Endocrinol. Metab). 24, E103±E108. Pritchard JE, Nowson CA, Strauss BJ, Carlson JS, Kaymakci B & Wark JD (1993): Evaluation of dual energy X-ray absorptiometry as a method of measurement of body fat. Eur. J. Clin. Nutr. 47, 216±228. Rencken ML, Murano R, Drinkwater BL & Chestnut III, CH (1991): In Vitro comparability of dual energy X-ray Absorptiometry (DEXA) bone densitometers. Calc. Tiss. Inte. 48, 245±248. Rico H, Revilla M, Villa L, Ruiz-Contreras D, Hernandez ER & Alverez de Buergo M (1994): The four-compartment models in body composition data from a study with dual-energy x-ray absorptiometry and nearinfrared interactance on 815 normal subjects. Metabolism. 43, 417±422. Roubenoff R, Kehayias JJ, Dawson-Hughes B & Heyms eld SB (1993): Use of dual-energy x-ray absorptiometry in body composition studies: not yet a gold standard. Am. J. Clin. Nutr. 58, 589±591. SAS (1990): SAS users guide: statistics, SAS Institute Inc.: Cary, NC Seto H, Kageyama M, Nomura K, Kakishita M & Tonami S (1991): Precision of Total-Body and Regional Bone Mineral Measurement by Dual-Energy X-ray Absorptiometry. Radiat. Med. 9, 110±113. Slosman DO, Casez J, Pichard C, Rochat T, Fery F, Rizzoli R, Bonjour JP, Morabia A & Donath A (1992): Assessment of whole-body composition with dual energy x-ray absorptiometry. Radiology. 185, 593±598. Svendsen OL, Haarbo J, Hassager C & Christiansen C (1993): Accuracy of measurements of body composition by dual-energy x-ray absorptiometry in vivo. Am. J. Clin. Nutr. 57, 605±608. Svendsen OL, Haarbo J, Heitmann BL, Gotfredsen A & Christiansen C (1991): Measurement of body fat in elderly subjects by dual-energy x- ray absorptiometry, bioelectrical impedance, and anthropometry. Am. J. Clin. Nutr. 53, 1117±1123. Trevisan C, Gandolini GG, Sibilla P, Penotti M, Caraceni M & Ortolani S (1992): Bone mass measurement by DXA: In uence of analysis procedures and interunit variation. J. Bone. Miner. Res. 7, 1373± Wahner HW, Looker A, Dunn WL, Walters LC, Hauser MF & Novak C (1994): Quality control of bone densitometry in a national healthy survey (NHANES III) using three mobile examination centers. J. Bone. Miner. Res. 9, 951±960. Wilson CR, Collier DB, Carrera GF & Jacobson DR (1990): Acronym for dual-energy x-ray absorptiometry. Radiology. 176, 875±876. Withers RT, Smith DA, Chatterton BE, Schultz CG & Gaffney RD (1992): A comparison of four methods of estimating body composition of male endurance athletes. Eur. J. Clin. Nutr. 46, 773±

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers European Journal of Clinical Nutrition (1997) 51, 449±454 ß 1997 Stockton Press. All rights reserved 0954±3007/97 $12.00 Body composition analysis by dual energy X-ray absorptiometry in female diabetics

More information

Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines

Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines International Journal of Obesity (1999) 23, 764±770 ß 1999 Stockton Press All rights reserved 0307±0565/99 $12.00 http://www.stockton-press.co.uk/ijo Short Communication Differences in whole body measurements

More information

9 Quality Assurance in Bone Densitometry section

9 Quality Assurance in Bone Densitometry section 9 Quality Assurance in Bone Densitometry section Introduction Bone densitometry is frequently used to determine an individual's fracture risk at a particular point in time but may also be used to assess

More information

Body composition in children and adults by air displacement plethysmography

Body composition in children and adults by air displacement plethysmography European Journal of Clinical Nutrition (1999) 53, 382±387 ß 1999 Stockton Press. All rights reserved 0954±3007/99 $12.00 http://www.stockton-press.co.uk/ejcn Body composition in children and adults by

More information

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers* JOURNAL OF BONE AND MINERAL RESEARCH Volume 13, Number 10, 1998 Blackwell Science, Inc. 1998 American Society for Bone and Mineral Research Bone Mineral and Body Composition Measurements: Cross-Calibration

More information

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13 Stephen J Pintauro, Tim R Nagy, Christa M Duthie, and Michael

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Dual-Energy X-ray Absorptiometry Femur Bone Measurements Examination Survey Years: 2005 to 2006 SAS Transport File: DXXFEM_D.XPT January 2009 NHANES 2005 2006 Data

More information

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status Journal of Clinical Densitometry, vol. 10, no. 3, 239e243, 2007 Ó Copyright 2007 by The International Society for Clinical Densitometry 1094-6950/07/10:239e243/$32.00 DOI: 10.1016/j.jocd.2007.03.002 Original

More information

A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution

A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution A Pilot Study of Exercise Training to Reduce Trunk Fat in Adults With HIV Associated Fat Redistribution Ronenn Roubenoff, Lauren Weiss, Ann McDermott, Tanya Heflin, Gregory J. Cloutier, Michael Wood and

More information

Multi-Compartment Model (MCM) and Body Composition Measurement Techniques

Multi-Compartment Model (MCM) and Body Composition Measurement Techniques White Paper Contents Multi Compartment Model (MCM) 1 Methods comparable to MCM 2 Methods NOT comparable to MCMC 3 References 6 Multi-Compartment Model (MCM) and Body Composition Measurement Techniques

More information

Review Comparison of body composition methods: a literature analysis

Review Comparison of body composition methods: a literature analysis European Journal of Clinical Nutrition (1997) 51, 495±503 ß 1997 Stockton Press. All rights reserved 0954±3007/97 $12.00 Review Comparison of body composition methods: a literature analysis M Fogelholm

More information

Nutritional Methodology-Body Composition

Nutritional Methodology-Body Composition Nutritional Methodology-Body Composition An Evaluation of Dual-Energy X-Ray Absorptiometry and Underwater Weighing to Estimate Body Composition by Means of Carcass Analysis in Piglets 1 Per Elowsson,*

More information

Evaluating body fat in girls and female adolescents: advantages and disadvantages of dual-energy X-ray absorptiometry 1 4

Evaluating body fat in girls and female adolescents: advantages and disadvantages of dual-energy X-ray absorptiometry 1 4 Evaluating body fat in girls and female adolescents: advantages and disadvantages of dual-energy X-ray absorptiometry 1 4 William W Wong, Albert C Hergenroeder, Janice E Stuff, Nancy F Butte, E O Brian

More information

New bioelectrical impedance formula for patients with respiratory insufficiency: comparison to dual-energy X-ray absorptiometry

New bioelectrical impedance formula for patients with respiratory insufficiency: comparison to dual-energy X-ray absorptiometry Eur Respir J 1998; 12: 960 966 DOI: 10.1183/09031936.98.12040960 Printed in UK - all rights reserved Copyright ERS Journals Ltd 1998 European Respiratory Journal ISSN 0903-1936 New bioelectrical impedance

More information

The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers

The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers Original Article http://dx.doi.org/10.3349/ymj.2012.53.2.289 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 53(2):289-293, 2012 The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density

More information

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Page: 1 of 6 Last Review Status/Date: March 2015 Determine Body Composition Description Using low dose x-rays of two different energy levels, whole body dual x-ray absorptiometry (DXA) measures lean tissue

More information

Whole body and regional body composition analysis by dual-energy X-ray absorptiometry in cirrhotic patients

Whole body and regional body composition analysis by dual-energy X-ray absorptiometry in cirrhotic patients European Journal of Clinical Nutrition (1997) 51, 810±814 ß 1997 Stockton Press. All rights reserved 0954±3007/97 $12.00 Whole body and regional body composition analysis by dual-energy X-ray absorptiometry

More information

Anthropometry and methods of body composition measurement for research and eld application in the elderly

Anthropometry and methods of body composition measurement for research and eld application in the elderly (2000) 54, Suppl 3, S26±S32 ß 2000 Macmillan Publishers Ltd All rights reserved 0954±3007/00 $15.00 www.nature.com/ejcn Anthropometry and methods of body composition measurement for research and eld application

More information

J EisenkoÈlbl 1, M Kartasurya 1 and K Widhalm 1 * Introduction

J EisenkoÈlbl 1, M Kartasurya 1 and K Widhalm 1 * Introduction (2001) 55, 423±429 ß 2001 Nature Publishing Group All rights reserved 0954±3007/01 $15.00 www.nature.com/ejcn Original Communication Underestimation of percentage fat mass measured by bioelectrical impedance

More information

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in 1 2 In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in adults: the influence of body mass index 3 4 Running title: Precision of the idxa for the measurement of visceral

More information

Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine

Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine HENZ W. WAHNER, M.D., WLLAM L. DUNN, M.Sc, MANUEL L. BROWN, M.D., Section

More information

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? Osteoporos Int (2010) 21:1227 1236 DOI 10.1007/s00198-009-1062-3 ORIGINAL ARTICLE Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems? B. Fan & Y. Lu

More information

Bioelectrical impedance analysis to assess body composition in obese adult women: The effect of ethnicity

Bioelectrical impedance analysis to assess body composition in obese adult women: The effect of ethnicity International Journal of Obesity (1998) 22, 243±249 ß 1998 Stockton Press All rights reserved 0307±0565/98 $12.00 Bioelectrical impedance analysis to assess body composition in obese adult women: The effect

More information

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes Susumu Sato 1), Shinichi Demura 2), Tamotsu Kitabayashi

More information

Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients

Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients OBES SURG (2014) 24:1476 1480 DOI 10.1007/s11695-014-1190-5 OTHER Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients Silvia L. Faria

More information

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p. Dedication Preface Acknowledgments Continuing Education An Introduction to Conventions in Densitometry p. 1 Densitometry as a Quantitative Measurement Technique p. 2 Accuracy and Precision p. 2 The Skeleton

More information

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years

Evaluation of DXA against the four-component model of body composition in obese children and adolescents aged 5 to 21 years Europe PMC Funders Group Author Manuscript Published in final edited form as: Int J Obes (Lond). 2010 April ; 34(4): 649 655. doi:10.1038/ijo.2009.249. Evaluation of DXA against the four-component model

More information

Precision of measurement and body size in whole-body air-displacement plethysmography

Precision of measurement and body size in whole-body air-displacement plethysmography ]> PAPER (2001) 25, 1161±1167 ß 2001 Nature Publishing Group All rights reserved 0307±0565/01 $15.00 www.nature.com/ijo of measurement and body size in whole-body air-displacement plethysmography JCK Wells

More information

Medical Policy. MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Medical Policy. MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Medical Policy MP 6.01.40 BCBSA Ref. Policy: 6.01.40 Last Review: 09/28/2017 Effective Date: 09/28/2017 Section: Radiology End Date: 09/18/2018 Related Policies 6.01.44 Vertebral Fracture Assessment With

More information

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition Policy Number: 6.01.40 Last Review: 4/2018 Origination: 4/2005 Next Review: 4/2019 Policy Blue Cross and Blue Shield of Kansas City

More information

pqct Measurement of Bone Parameters in Young Children

pqct Measurement of Bone Parameters in Young Children Journal of Clinical Densitometry, vol. 3, no. 1, 9 14, Spring 2000 Copyright 2000 by Humana Press Inc. All rights of any nature whatsoever reserved. 0169-4194/00/3:9 14/$11.50 Original Article pqct Measurement

More information

DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals

DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals DEXA Bone Mineral Density Tests and Body Composition Analysis Information for Health Professionals PERFORMANCE DEXA is an advanced technology originally used to, and still capable of assessing bone health

More information

Clinical Densitometry

Clinical Densitometry Volume 8 Number 3 Fall 2005 ISSN: 1094 6950 Journal of Clinical Densitometry The Official Journal of The International Society for Clinical Densitometry Editor-in-Chief Paul D. Miller, MD HumanaJournals.com

More information

Luís B Sardinha, Timothy G Lohman, Pedro J Teixeira, Dartagnan P Guedes, and Scott B Going

Luís B Sardinha, Timothy G Lohman, Pedro J Teixeira, Dartagnan P Guedes, and Scott B Going Comparison of air displacement plethysmography with dual-energy X-ray absorptiometry and 3 field methods for estimating body composition in middle-aged men 1,2 Luís B Sardinha, Timothy G Lohman, Pedro

More information

Quality Control of DXA System and Precision Test of Radio-technologists

Quality Control of DXA System and Precision Test of Radio-technologists J Bone Metab 2014;21:2-7 http://dx.doi.org/10.11005/jbm.2014.21.1.2 pissn 2287-6375 eissn 2287-7029 Review Article Quality Control of DXA System and Precision Test of Radio-technologists Ho-Sung Kim 1,

More information

Norland Densitometry A Tradition of Excellence

Norland Densitometry A Tradition of Excellence Norland Densitometry A Tradition of Excellence Norland DXA Bone Density Measurement Osteoporosis is a disease marked by reduced bone strength leading to an increased risk of fractures. About 54 million

More information

Accuracy of DEXA scanning & other methods for determining BMD.

Accuracy of DEXA scanning & other methods for determining BMD. BMD- Measurement Site Accuracy of DEXA scanning & other methods for determining BMD. Ann Larkin In general, densitometry techniques can be performed in either the axial or the appendicular skeleton, depending

More information

Bone Densitometry Radiation dose: what you need to know

Bone Densitometry Radiation dose: what you need to know Bone Densitometry Radiation dose: what you need to know John Damilakis, PhD Associate Professor and Chairman University of Crete, Iraklion, Crete, GREECE Estimation of bone status using X-rays Assessment

More information

Bedside methods versus dual energy X-ray absorptiometry for body composition measurement in COPD

Bedside methods versus dual energy X-ray absorptiometry for body composition measurement in COPD Eur Respir J ; 19: 66 631 DOI: 1.1183/931936..796 Printed in UK all rights reserved Copyright #ERS Journals Ltd European Respiratory Journal ISSN 93-1936 Bedside methods versus dual energy X-ray absorptiometry

More information

MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition Medical Policy MP 6.01.40 BCBSA Ref. Policy: 6.01.40 Last Review: 09/19/2018 Effective Date: 09/19/2018 Section: Radiology Related Policies 6.01.44 Vertebral Fracture Assessment With Densitometry 9.01.502

More information

Is percentage body fat differentially related to body mass index in Hispanic Americans, African Americans, and European Americans?

Is percentage body fat differentially related to body mass index in Hispanic Americans, African Americans, and European Americans? Is percentage body fat differentially related to body mass index in Hispanic Americans, African Americans, and European Americans? 1 3 José R Fernández, Moonseong Heo, Steven B Heymsfield, Richard N Pierson

More information

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302) Date of Interview/Examination (MM/DD/YYYY): A downloadable PDF

More information

Bone mineral density in the normal Iranian population: a comparison with American reference data

Bone mineral density in the normal Iranian population: a comparison with American reference data Arch Osteoporos (06) 1:29 DOI 10.1007/s117-006-0005-2 ORIGINAL ARTICLE Bone mineral density in the normal Iranian population: a comparison with American reference data Gholamhossein Ranjbar Omrani & Seyed

More information

Purpose. Methods and Materials

Purpose. Methods and Materials Prevalence of pitfalls in previous dual energy X-ray absorptiometry (DXA) scans according to technical manuals and International Society for Clinical Densitometry. Poster No.: P-0046 Congress: ESSR 2014

More information

Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density in Japanese Adults

Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density in Japanese Adults Tohoku J. Exp. Med., 2007, Measurement 213, 51-61Error Characteristics of Skinfold Caliper 51 Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density

More information

Obesity Increases Precision Errors in Total Body Dual X-ray Absorptiometry Measurements.

Obesity Increases Precision Errors in Total Body Dual X-ray Absorptiometry Measurements. Obesity Increases Precision Errors in Total Body Dual X-ray Absorptiometry Measurements. 1 K.M.Knapp, 1 J.R.Welsman, 1 S.J.Hopkins, 1 A.Shallcross, 2 I.Fogelman and 2 G.M.Blake 1. University of Exeter,

More information

2013 ISCD Official Positions Adult

2013 ISCD Official Positions Adult 2013 ISCD Official Positions Adult These are the Official Positions of the ISCD as updated in 2013. The Official Positions that are new or revised since 2007 are in bold type. Indications for Bone Mineral

More information

Detecting meaningful body composition changes in athletes

Detecting meaningful body composition changes in athletes Detecting meaningful body composition changes in athletes using dual-energy X-ray absorptiometry Steffi L Colyer 1, Simon P Roberts 1, Jonathan B Robinson 2, Dylan Thompson 1, Keith A Stokes 1, James L

More information

RELIABILITY OF DUAL-ENERGY X-RAY

RELIABILITY OF DUAL-ENERGY X-RAY RELIABILITY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY IN ASSESSING BODY COMPOSITION IN ELITE ATHLETES Alisa Nana BND (Hons) A thesis submitted in fulfilment of the requirements for the degree of Doctor of Philosophy

More information

Location of body fat and body size impacts DXA soft tissue measures: a simulation study

Location of body fat and body size impacts DXA soft tissue measures: a simulation study (2008) 62, 553 559 & 2008 Nature Publishing Group All rights reserved 0954-3007/08 $30.00 www.nature.com/ejcn ORIGINAL ARTICLE Location of body fat and body size impacts DXA soft tissue measures: a simulation

More information

Estimation of body composition from bioelectrical impedance of body segments : comparison with dual-energy X-ray absorptiometry

Estimation of body composition from bioelectrical impedance of body segments : comparison with dual-energy X-ray absorptiometry British Journal of Nutrition (1993), 69, 645-655 645 Estimation of body composition from bioelectrical impedance of body segments : comparison with dual-energy X-ray absorptiometry BY S. P. STEWART1, P.

More information

ESPEN Congress The Hague 2017

ESPEN Congress The Hague 2017 ESPEN Congress The Hague 2017 Paediatric specificities of nutritional assessment Body composition measurement in children N. Mehta (US) 39 th ESPEN Congress The Hague, Netherlands Body Composition Measurement

More information

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition. Original Policy Date 12:2013

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition. Original Policy Date 12:2013 MP 6.01.28 Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Medical Policy Section Radiology Is12:2013sue 12:2012 Original Policy Date 12:2013 Last Review Status/Date reviewed

More information

Fat-free mass index: changes and race/ethnic differences in adulthood

Fat-free mass index: changes and race/ethnic differences in adulthood (2011) 35, 121 127 & 2011 Macmillan Publishers Limited All rights reserved 0307-0565/11 www.nature.com/ijo ORIGINAL ARTICLE Fat-free mass index: changes and race/ethnic differences in adulthood HR Hull

More information

The use of dual-energy X-ray absorptiometry to estimate the dissected composition of lamb carcasses

The use of dual-energy X-ray absorptiometry to estimate the dissected composition of lamb carcasses Meat Science 73 (2006) 249 257 MEAT SCIENCE www.elsevier.com/locate/meatsci The use of dual-energy X-ray absorptiometry to estimate the dissected composition of lamb carcasses J. Mercier a, C. Pomar b,

More information

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents

Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and obese adolescents British Journal of Nutrition (2003), 90, 987 992 q The Authors 2003 DOI: 10.1079/BJN2003983 Evaluation of two foot-to-foot bioelectrical impedance analysers to assess body composition in overweight and

More information

special communication

special communication special communication Determining bone and total body mineral content from body density and bioelectrical response spectroscopy STEVEN F. SICONOLFI, 1 RANDAL J. GRETEBECK, 2 WILLIAM W. WONG, 3 SHERIL S.

More information

Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults

Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults Body Composition Analysis by Air Displacement Plethysmography in Normal Weight to Extremely Obese Adults Kazanna C. Hames 1,2, Steven J. Anthony 2, John C. Thornton 3, Dympna Gallagher 3 and Bret H. Goodpaster

More information

Body-fat measurement in patients with acquired immunodeficiency syndrome: which method should be used?13

Body-fat measurement in patients with acquired immunodeficiency syndrome: which method should be used?13 Original Research Communications-general Body-fat measurement in patients with acquired immunodeficiency syndrome: which method should be used?13 Jack Wang, Donald P Koiler, Man Russell, Santiago Burasiero,

More information

DUAL ENERGY X-RAY ABSORPTIOMETRY (DXA) GOLD STANDARD FOR BONE HEALTH AND BODY COMPOSITION ASSESSMENT

DUAL ENERGY X-RAY ABSORPTIOMETRY (DXA) GOLD STANDARD FOR BONE HEALTH AND BODY COMPOSITION ASSESSMENT DUAL ENERGY X-RAY ABSORPTIOMETRY (DXA) GOLD STANDARD FOR BONE HEALTH AND BODY COMPOSITION ASSESSMENT Prof d-r Slavica Šubeska Stratrova Clinic of Endocrinology, Skopje, Macedonia Definition and classification

More information

Bone Mineral Density and Its Associated Factors in Naresuan University Staff

Bone Mineral Density and Its Associated Factors in Naresuan University Staff Naresuan University Journal 2005; 13(3): 13-18 13 Bone Mineral Density and Its Associated Factors in Naresuan University Staff Supawitoo Sookpeng *, Patsuree Cheebsumon, Malinee Dhanarun, Thanyavee Pengpan

More information

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA Strong Bone Asia 2013 Osteoporosis in ASEAN Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA Chris Schultz Scientist-in-Charge (Bone Densitometry) Royal Adelaide Hospital, Adelaide,

More information

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 24.7 years Height / Weight: 8.0 cm 79.0 kg Sex / Ethnic: Male Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 40% 30% 20% 0% 20 30 40 50 60 70 80 90 00 Centile

More information

Association of Lean Tissue and Fat Mass with Bone Mineral Content in Children and Adolescents

Association of Lean Tissue and Fat Mass with Bone Mineral Content in Children and Adolescents Association of Lean Tissue and Fat Mass with Bone Mineral Content in Children and Adolescents Angelo Pietrobelli,* Myles S. Faith,* Jack Wang,* Paolo Brambilla, Giuseppe Chiumello, and Steven B. Heymsfield*

More information

The Assessment of Body Composition in Health and Disease

The Assessment of Body Composition in Health and Disease The Assessment of Body Composition in Health and Disease Giorgio Bedogni, Paolo Brambilla, Stefano Bellentani and Claudio Tiribelli CHAPTER 3 BODY AND NUTRITIONAL STATUS Nutritional status can be operationally

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Birth Date: 40.2 years Height / Weight: 158.0 cm 52.0 kg Sex / Ethnic: Female Patient ID: Total Body Tissue Quantitation Composition Reference: Total Tissue 50% 40% 30% 20% 20 30 40 50 60 70 80 90 100

More information

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation

Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Development of Bio-impedance Analyzer (BIA) for Body Fat Calculation Recent citations - Munawar A Riyadi et al To cite this article:

More information

Spinal Cord Rehab Program, Toronto Rehabilitation Institute, University Health Network, 520 Sutherland Drive, Toronto, ON, Canada M4G 3V9 2

Spinal Cord Rehab Program, Toronto Rehabilitation Institute, University Health Network, 520 Sutherland Drive, Toronto, ON, Canada M4G 3V9 2 Journal of Obesity Volume 212, Article ID 936, 9 pages doi:1.11/212/936 Clinical Study Bioelectrical Impedance and Dual-Energy X-Ray Absorptiometry Assessments of Changes in Body Composition Following

More information

Selecting regions of interest on intra oral radiographs. for the prediction of bone mineral density

Selecting regions of interest on intra oral radiographs. for the prediction of bone mineral density Selecting regions of interest on intra oral radiographs for the prediction of bone mineral density W.G.M. Geraets a, J.G.C. Verheij a, P.F. van der Stelt a, K. Horner b, C. Lindh c, K. Nicopoulou-Karayianni

More information

European Journal of Endocrinology (1997) ISSN

European Journal of Endocrinology (1997) ISSN European Journal of Endocrinology (1997) 137 167 171 ISSN 0804-4643 Change in C-terminal cross-linking domain of type I collagen in urine, a new marker of bone resorption, during and after gonadotropin-releasing

More information

N. J. FULLER, S. A. JEBB, M. A. LASKEY', W. A. COWARD and M. ELlA INTRODUCTION

N. J. FULLER, S. A. JEBB, M. A. LASKEY', W. A. COWARD and M. ELlA INTRODUCTION Clinical Science (1992) 82, 687693 (Printed in Great Britain) Fourcomponent model for the assessment of body composition in humans: comparison with alternative methods, and evaluation of the density and

More information

Body mass index in screening for adiposity in children and adolescents: systematic evaluation using receiver operating characteristic rv 2

Body mass index in screening for adiposity in children and adolescents: systematic evaluation using receiver operating characteristic rv 2 Body mass index in screening for adiposity in children and adolescents: systematic evaluation using receiver operating characteristic rv 2 Ross Lazarus, Louise Baur, Karen Webb, and Fiona Blyth ABTRACT

More information

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi Clinical Utility of Bone Densitometry Diagnosis (DXA)

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 48.2 years Height / Weight: 150.0 cm 72.0 kg Sex / Ethnic: Female

More information

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1 Patient: Birth Date: 43.4 years Height / Weight: 170.0 cm 66.0 kg Sex / Ethnic: Female

More information

Understanding Body Composition

Understanding Body Composition PowerPoint Lecture Outlines 7 Understanding Body Composition Objectives Define body composition. Explain why the assessment of body size, shape, and composition is useful. Explain how to perform assessments

More information

Estimates of body composition with dual-energy X-ray absorptiometry in adults 1 3

Estimates of body composition with dual-energy X-ray absorptiometry in adults 1 3 Original Research Communications Estimates of body composition with dual-energy X-ray absorptiometry in adults 1 3 Chaoyang Li, Earl S Ford, Guixiang Zhao, Lina S Balluz, and Wayne H Giles ABSTRACT Background:

More information

BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES. Abstract. Introduction. Volume 3, No. 2, 2011, UDC :572.

BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES. Abstract. Introduction. Volume 3, No. 2, 2011, UDC :572. EXERCISE AND QUALITY OF LIFE Volume 3, No. 2, 2011, 43-48 UDC 796.034.6-051:572.087 Research article BODY MASS INDEX AND BODY FAT CONTENT IN ELITE ATHLETES Jelena Popadiã Gaãeša *, Otto Barak, Dea Karaba

More information

Introduction. Methods

Introduction. Methods (2000) 24, 627±632 ß 2000 Macmillan Publishers Ltd All rights reserved 0307±0565/00 $15.00 www.nature.com/ijo Overweight and obese children have low bone mass and area for their weight A Goulding 1 *,

More information

Healthy percentage body fat ranges: an approach for developing guidelines based on body mass index 1 3

Healthy percentage body fat ranges: an approach for developing guidelines based on body mass index 1 3 Healthy percentage body fat ranges: an approach for developing guidelines based on body mass index 1 3 Dympna Gallagher, Steven B Heymsfield, Moonseong Heo, Susan A Jebb, Peter R Murgatroyd, and Yoichi

More information

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition Policy Number: 6.01.40 Last Review: 4/2014 Origination: 4/2005 Next Review: 4/2015 Policy Blue Cross and Blue Shield of Kansas

More information

Techniques for Undertaking Dual-Energy X-Ray Absorptiometry Whole-Body Scans to Estimate Body Composition in Tall and/or Broad Subjects

Techniques for Undertaking Dual-Energy X-Ray Absorptiometry Whole-Body Scans to Estimate Body Composition in Tall and/or Broad Subjects International Journal of Sport Nutrition and Exercise Metabolism, 2012, 22, 313-322 2012 Human Kinetics, Inc. www.ijsnem-journal.com ORIGINAL RESEARCH Techniques for Undertaking Dual-Energy X-Ray Absorptiometry

More information

Measurement of Periprosthetic Bone Density in Hip Arthroplasty Using Dual-energy X-ray Absorptiometry Reproducibility of Measurements

Measurement of Periprosthetic Bone Density in Hip Arthroplasty Using Dual-energy X-ray Absorptiometry Reproducibility of Measurements The Journal of Arthroplasty Vol. 11 No. 4 1996 Measurement of Periprosthetic Bone Density in Hip Arthroplasty Using Dual-energy X-ray Absorptiometry Reproducibility of Measurements R. C. Smart, BSc, MSc,

More information

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men Osteoporos Int (1999) 10:265 270 ß 1999 International Osteoporosis Foundation and National Osteoporosis Foundation Osteoporosis International Original Article Bone Mineral Density and Vertebral Fractures

More information

Pharmacy Management Drug Policy

Pharmacy Management Drug Policy SUBJECT: - Forteo (teriparatide), Prolia (denosumab), Tymlos (abaloparatide) POLICY NUMBER: Pharmacy-35 EFFECTIVE DATE: 9/07 LAST REVIEW DATE: 9/29/2017 If the member s subscriber contract excludes coverage

More information

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA ORIGINAL ARTICLE Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA Leila Amiri 1, Azita Kheiltash 2, Shafieh Movassaghi 1, Maryam Moghaddassi 1, and Leila Seddigh 2 1 Rheumatology

More information

Chapter 17: Body Composition Status and Assessment

Chapter 17: Body Composition Status and Assessment Chapter 17: Body Composition Status and Assessment American College of Sports Medicine. (2010). ACSM's resource manual for guidelines for exercise testing and prescription (6th ed.). New York: Lippincott,

More information

O. Bruyère M. Fossi B. Zegels L. Leonori M. Hiligsmann A. Neuprez J.-Y. Reginster

O. Bruyère M. Fossi B. Zegels L. Leonori M. Hiligsmann A. Neuprez J.-Y. Reginster DOI 10.1007/s00296-012-2460-y ORIGINAL ARTICLE Comparison of the proportion of patients potentially treated with an anti-osteoporotic drug using the current criteria of the Belgian national social security

More information

COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN

COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN COMPARISON OF AIR DISPLACEMENT PLETHYSMOGRAPHY TO HYDROSTATIC WEIGHING FOR ESTIMATING TOTAL BODY DENSITY IN CHILDREN Geo Claros 1, Holly R Hull 1, and David A Fields 1,2 * 1 Department of Health and Exercise

More information

Body composition A tool for nutritional assessment

Body composition A tool for nutritional assessment Body composition A tool for nutritional assessment Ingvar Bosaeus Clinical Nutrition Unit Sahlgrenska University Hospital NSKE Oslo 2012-01-18 Outline What is body composition? What is nutritional assessment?

More information

2013 ISCD Combined Official Positions

2013 ISCD Combined Official Positions 2013 ISCD Combined Oicial Positions Oicial Positions of the International Society for Clinical Densitometry The International Society for Clinical Densitometry (ISCD) is a not-for-profit multidisciplinary

More information

Note that metric units are used in the calculation of BMI. The following imperial-metric conversions are required:

Note that metric units are used in the calculation of BMI. The following imperial-metric conversions are required: Body Composition Body Composition: Assessment and Interpretation Body composition has great practical and functional significance for many of us: scientists, clinicians and the general population. It can

More information

Effect of hand positioning on idxa precision error for total and regional bone and body

Effect of hand positioning on idxa precision error for total and regional bone and body Effect of hand positioning on idxa precision error for total and regional bone and body composition parameters. Abstract Dual energy X-ray absorptiometry (DXA) body composition measurements are performed

More information

Fitness Concepts + Principles Packet. Name: Period:

Fitness Concepts + Principles Packet. Name: Period: Fitness Concepts + Principles Packet Name: Period: Content Pages + Directions 5 Components of Fitness.................................................... Slides 3-12 FITT Principle.........................................................

More information

Body Composition Profiles of Elite American Heptathletes

Body Composition Profiles of Elite American Heptathletes 162 / Houtkooper et al. International Journal of Sport Nutrition and Exercise Metabolism, 2001, 11, 162-173 2001 Human Kinetics Publishers, Inc. Body Composition Profiles of Elite American Heptathletes

More information

Clinical Study Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA

Clinical Study Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA ISRN Obesity Volume 2013, Article ID 462394, 5 pages http://dx.doi.org/10.1155/2013/462394 Clinical Study Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model

More information

Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67

Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67 WRN94060.qxd 4/11/05 4:57 PM Page 60 Simopoulos AP (ed): Nutrition and Fitness: Obesity, the Metabolic Syndrome, Cardiovascular Disease and Cancer. Basel, Karger, 2005, vol 94, pp 60 67 Physical Activity

More information

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT

Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT O S T E O P O R O S I S A S S E S S M E N T Advanced Point-of-Care Bone Health Assessment HOLOGIC OSTEOPOROSIS ASSESSMENT Identify Patients at Risk The evaluation of bone health has become an essential

More information

Dual-energy Vertebral Assessment

Dual-energy Vertebral Assessment Dual-energy Vertebral Assessment gehealthcare.com Dual-energy Vertebral Assessment More than 40% of women with normal or osteopenic BMD had a moderate or severe vertebral deformation seen with DVA. Patrick

More information

TOTAL BODY COMPOSITION

TOTAL BODY COMPOSITION Specialists in Joint Replacement, Spinal Surgery, Orthopaedics and Sport Injuries DEXA TOTAL BODY COMPOSITION www.sportsurgeryclinic.com WHY CHOOSE DEXA TOTAL BODY COMPOSITION? DEXA, short for Dual Energy

More information