1 Cardiopulmonary Imaging Original Research Christe et al. CT of Lung Nodules Cardiopulmonary Imaging Original Research CT Screening and Follow-Up of Lung Nodules: Effects of Tube Current Time Setting and Nodule Size and Density on Detectability and of Tube Current Time Setting on Apparent Size Andreas Christe 1,2 Jaled Charimo Torrente 2 Margaret Lin 1 Andrew Yen 1 Rich Hallett 1 Kingshuk Roychoudhury 3 Florian Schmitzberger 1 Peter Vock 2 Justus Roos 1 Christe A, Torrente JC, Lin M, et al. Keywords: CT, density, dose, lung nodules, sensitivity, volume measurement error DOI: /AJR Received July 8, 2010; accepted after revision January 17, Supported by the Swiss National Science Foundation (SSMBS, Patronage of Swiss Academy of Medical Sciences), the Swiss Gottfried und Julia Bangerter- Rhyner Foundation and the Swiss Huggenberg- Bischoff Foundation. 1 Department of Radiology, Stanford University School of Medicine, Stanford, CA. 2 Department of Diagnostic, Interventional and Pediatric Radiology, Inselspital, University Hospital, University of Bern, Freiburgstrasse 10, CH-3010 Bern, Switzerland. Address correspondence to A. Christe (andreas.christe@insel.ch). 3 Statistics Department, University College Cork, Cork, Ireland. AJR 2011; 197: X/11/ American Roentgen Ray Society OBJECTIVE. The purpose of the study was to quantify and compare the effect of CT dose and of size and density of nodules on the detectability of lung nodules and to quantify the influence of CT dose on the size of the nodules. MATERIALS AND METHODS. From 50 patients a total of 125 cuboidal regions of interest ( cm volumes) showing a single nodule ( 8 mm) and 27 normal cuboids were selected. Image sets were reconstructed with the software from raw data simulating different dose levels: 300 (original dose), 220, 180, 140, 100, 80, 60, 50, 40, 30, 20, 10, and 5 reference mas. A logistic regression model was used to analyze detectability for three blinded readers. Odds ratios were calculated for nodule size smaller than 3 mm versus 3 mm and larger and for nodule attenuation of 300 HU and greater versus less than 300 HU. RESULTS. Tube current time settings of 10 mas and greater were not associated with a significant difference in individual reader sensitivity compared with the standard setting of 300 mas. At 5 mas only one reader had a significant decrease in sensitivity, from 82% to 77% (p = ). According to the odds ratios and logistic regression results, the strongest negative effect on sensitivity can be assumed for low nodule density followed by small nodule size and dose level. The mean nodule volume measurement error between 5 and 300 mas was 2.2% ± 18% (SD) and much lower than the interobserver volume measurement error rate of 38% ± 45%. CONCLUSION. The results show the feasibility of a low-dose CT protocol at 10 mas for follow-up of lung nodules. Computer-aided volume measurement in follow-up of lung nodules decreases interobserver variability. C T screening has increased the rate of detection of small nodules, including those of early peripheral lung cancer [1 3]. Despite the increasingly higher spatial and contrast resolution of CT, nodular lesions are missed at chest CT. The sensitivity of chest CT for lung nodules depends on nodule characteristics and image acquisition parameters. In a study by Ko et al. [4], a small nodule size of 5 mm or less in diameter (detection sensitivity, 74%; > 5 mm, 82%), ground-glass opacity nodules (sensitivity, 65%; solid nodules, 83%), and lesion location (sensitivity for central nodules, 61%; peripheral nodules, 80%) were found to be major factors contributing to the difficulty of detecting nodules. In several studies, investigators have defined the lowest acceptable tube current time settings without significant loss of sensitivity compared with standard-dose CT ( mas). Hetmaniak et al. [5] found 30 mas; Weng et al. [6], 43 mas; Gergely et al. [7], 5 mas; and Das et al. [8], 10 mas. Nodule detection can be improved by advances in computer-aided detection systems that are being developed and evaluated to provide a second perspective on nodule detection [9, 10]. However, Lee et al. [11] found that the sensitivity of a computer-aided detection system (81%) did not differ significantly from that of radiologists (85%). Radiologists were more sensitive at detecting nodules attached to other structures, whereas computer-aided detection was better for finding isolated nodules and those 5 mm in diameter or smaller. The size of the nodule is important because volume-doubling time is a predictor of malignancy [12, 13]. There is considerable interreader and intrareader variability in assessing the size of pulmonary nodules [14 16]. The measurement is affected by measuring method (determination of longest axis, segmentation of computer-aided measure- AJR:197, September

2 Christe et al. ment), FOV, window width and level settings [17], and changes in lung volume during breathing [18]. Even though computerized methods seem to be immune to some factors that affect radiologists, challenges persist owing to image acquisition parameters and nodule characteristics. Tube current time setting, slice thickness, reconstruction algorithm, nodule attenuation, and nodule size influence volume error in studies of computer-aided detection [19, 20]. The purposes of this study were to quantify and compare the effect of CT tube current time setting, nodule size, and nodule density on the detectability of lung nodules and to investigate the effect of CT tube current time setting on apparent nodule size. The major differences between this investigation and previous studies include the larger study sample, greater continuous range of nodule attenuation and tube current time settings, comparison between radiologist and computer methods, and investigation of lung nodules in real patients. We also examined the influence of both nodule characteristics and image acquisition parameters on sensitivity and the influence of tube current time setting on volume measurement. Materials and Methods Patients Raw data from thoracic CT scans of 50 consecutively registered patients referred for cancer staging, lung nodule screening, or follow-up were prospectively collected between January and March The images were read by a consensus panel of two thoracic radiologists with 9 and 10 years of experience. Lung nodules were found in 41 patients (22 men, 19 women; median age, 55.5 years; range, years; median weight, 73 kg; range, kg), and a total of 125 nodules were selected. The nine patients with normal (true-negative) thoracic CT findings served as a control group. The study protocol was approved by the local institutional review board, and informed consent was waived because of the retrospective data processing. Image Acquisition All image data were obtained with a 64-MDCT scanner (Somatom Sensation, Siemens Healthcare) with collimation of mm, pitch of 2, 120 kvp, and reference tube current time setting of 300 mas. Automated tube current time modulation (CARE Dose, Siemens Healthcare) was engaged to ensure a similar noise index for each patient. Acquisition parameters were identical to the parameters used in our standard clinical protocol. All patients were scanned in the supine position from the lung apex to the base of the chest at breath-hold full inspiration. No IV contrast material was administered. Image Reconstruction After reconstruction of the original DICOM images with an image matrix of and a B45 kernel, 12 additional image sets were reconstructed from the same raw data simulating different dose levels. Computer-calculated noise was superimposed on the original raw dataset by use of the validated CT software (Somatom Noise+ V5.Obeta, 2006, Siemens Healthcare) [21], and reference tube current time settings of 220, 180, 140, 100, 80, 60, 50, 40, 30, 20, 10, and 5 mas were simulated. This process resulted in 13 datasets per patient available for image analysis. The rather low kernel was a compromise for improving the 3D reformation and the accuracy of semiautomated volume measurement. Expert-Derived Standard of Reference A consensus panel of two thoracic radiologists (9 and 10 years of experience) identified small nodules ( 8 mm), making sure that approximately one half of the selected nodules were solid (maximum attenuation, 300 HU) and one half ground-glass nodules (< 300 HU). Cuboidal regions of interest ( cm volumes) containing only one nodule each were selected by transferring the center coordinates to the lung nodule evaluation platform (Internet readout workstation described later). To compensate for potential enhanced nodule detection for limited FOV and z-axis, only smaller ( 8 mm) nodules were selected. A total of 125 cuboids showing a single nodule were selected for subsequent image analysis. A total of 27 cuboids of the nine patients without disease were randomly collected and added for image analysis. The greatest axial dimension of each CT finding was measured with digital calipers, and its maximum attenuation was noted in Hounsfield units. Image Analysis Detectability of nodules Interpretation was performed with a lung nodule evaluation platform that was not commercially available. The system featured a split screen for automated display of two orthogonal cross sections (axial and coronal planes combined for multiplanar reconstruction) in stacked cine mode of a cuboidal region of interest. For our study the screen was zoomed to the axial cross sections, and the coronal plane was hidden. Before appearing on the screen, the cuboids were randomly axially rotated to positions 0, 90, 180, 270. The cuboids then could be browsed slice by slice. In a separate integrated window, readers recorded their answers by mouse click. Data then were automatically transmitted online to a statistician. The lung nodule evaluation platform was installed on a PC workstation (Dell E520, 2.1 GHz, Intel Core 2 Duo, 4 GB RAM) with a 20-inch (51 cm) liquid crystal display (Dell Ultrasharp, pixels). Ambient light conditions, which were similar to our clinical reading conditions, and the display window settings (window center, 600 HU; width, 1500 HU) were kept constant for all reading sessions and readers. Size of nodules Digitized measurements of nodule size were performed with a Mac Book Pro 15-inch (38 cm) screen (Apple Computer) running OsiriX Imaging Software (OsiriX). Center coordinates were used to find the nodules in the full CT datasets. Zoom of 2000% was used for all nodules. Computer-aided volume measurement (LMS Lung/ Track, Median Technologies) was performed. This semiautomatic lesion measure ment solution seg ments pulmonary nodules identi fied by radiologists (radiologist chooses nodule by mouse click). The segmentation is the result of a regiongrowing algorithm followed by thresholding on attenuation (Hounsfield units) and concluded by mathematic morphologic operations (erosion and dilation) to separate the suspicious lesion from surrounding anatomic structures. The method published by Bland and Altman for assessing agree ment between two methods of clinical measure ment [22] was used to determine intraobserver and inter observer variability and agreement of volume measurements at different CT tube current time levels. This method was used to calculate limits of agreement. Evaluation Detectability of nodules Three readers (board-certified radiologists who had 5 years [reader 1], 6 years [reader 2], and 12 years [reader 3] of experience in interpretation of chest CT scans) were instructed to identify lung nodules. A pilot study with a different reader (9 years of chest CT experience) of 36 cuboids from 20 patients at five dose levels (300, 100, 50, 20, and 5 mas) was performed to decrease the number of readout cases. Only at dose levels of 20 and 5 mas were different answers given compared with 300 mas. Therefore, we limited the evaluation to seven relevant tube current time levels: 300, 100, 40, 30, 20, 10, and 5 mas. The readers answers were recorded by use of the lung nodule evaluation platform. Image interpretation took place in multiple reading sessions with blocks not exceeding 1 hour of successive reading to minimize reader fatigue. Presentation of the cases to the readers was randomized, and the cuboids were randomly rotated over 360 in the axial plane to minimize recognition bias. Size of nodules Comparison of measurements on two consecutive scans requires knowledge of observer agreement because all observed differences 624 AJR:197, September 2011

3 CT of Lung Nodules between measurements of the same nodule on different scans can be attributed to intraobserver variability or to interscan variability. Two readers (other than the aforementioned board-certified radiologists who had 3 years [reader 4] and 6 years [reader 5] of experience in interpretation of chest CT) measured the size and volume of the 59 solid nodules at all 13 tube current time levels manually and with the computer-aided lesion measurement solution software. The readers measured volume by multiplying the longest axial axis (according to the Response Evaluation Criteria in Solid Tumors [RECIST]), the perpendicular short axial axis, the longest coronal z-axis, and π/6. Reader 4 performed the same evaluation 3 months later to determine intrareader agreement. The computer measurements were obtained with segmentation and threshold methods for automated volume and longest axial diameter measurement (RECIST). This semiautomatic process was performed by both readers. Reader 5 repeated the measurement with a delay of 1 day for intrareader agreement. This technique was chosen because it was our hypothesis that repeated measurement with different starting points might give different volumes. Statistical Analysis Detectability of nodules Individual results from the three readers were compared with the standard of reference. Sensitivity and specificity for each reader in detection of pulmonary nodules were calculated at each dose level. Paired comparisons of the sensitivity of detection at each lower dose level against the corresponding detection at 300 ma were performed. The null hypothesis of equality of sensitivity for this dose pair was then tested by McNemar test for each reader [23]. For comparison of the effect of tube current time level and nodule size and density on sensitivity for lung nodules the odds ratio for each variable and reader was calculated. For practical reasons, data on each variable were dichotomized: size 3 mm and larger versus less than 3 mm, attenuation 300 HU versus less than 300 HU (ground-glass opacity), and standard dose (300 mas) versus low dose (5 mas). Odds ratios and sensitivities were calculated, and p values for each variable were analyzed by chi-square test. If an absolute cell frequency was lower than 5, the Fisher exact probability test was applied. A value of p < 0.05 was considered to indicate a statistically significant difference. To rule out a bias of dichotomizing level setting and to obtain a more general statement, the following logistic regression model was fitted to diagnosis [24]: log[p / (1 p)] = c + b 0 LD st + b 1 size st + b 2 HU st + error where p is the probability of a reader s giving a positive diagnosis (sensitivity); c is the intercept; b 0, b 1, and b 2 are regression coefficients (log odds ratios); LD is log(dose) in milliampere-seconds (tube current time level); st indicates standardized, size is the maximum axial diameter of a nodule (average of readers 4 and 5 at standard tube current time level); HU is attenuation; and error is the specification error. The variables were standardized, and the mean was made 0 with 1 SD. The advantage of using this calculation is that the regression coefficients b 0, b 1, and b 2 become comparable on the same scale. Because a logarithmic relation between dose and sensitivity can be assumed, the logarithmic dose was used. Size of nodules The method described by Bland and Altman [22] for assessing agreement between two methods of clinical measurement was used to determine intraobserver and interobserver variability and agreement of volume measurements between the standard tube current time setting of 300 mas and all lower tube current time levels (220 5 mas). This method of calculation of limits of agreement results in the range of 95% of all volume measurement errors. Because the percentage of changes is more important than absolute volume values, the volume measurement error (VME) is given as a percentage of absolute volume measurement difference (V 1 V 2 ) in relation to nodule volume, where V 1 is the volume of the first reading. The true nodule volume is not known; therefore, the mean nodule volume of both measurements (V 1 + V 2 ) / 2 is the best estimate obtainable [22]: (V 1 V 2 ) VME = 100% (V 1 + V 2 ) / 2 0% CI 2sE LA +2sE +CI 2sd CI VME% VME: Volume measurement error, differences in % LA: Limits of agreement CI: Confidence interval sd: Measured standard deviation of VME se: Standard error of VME = (sd 2 /n) se: Standard error of LA = (3*sd 2 /n) In addition, the 95% CI of the mean volume measurement error of the upper and lower limits of agreement was calculated. The limit of agreement was calculated from the SD of the measurement differences (VME), and the 95% CI was derived from the standard error for normally distributed data, which can be assumed for measurement differences [22] (Fig. 1). To find significant differences in volume measurements, the 95% CI of the volume measurement error should not include 0 for significant systematic bias. The corrected SD of measurement differences was applied for repeated measurements [22]. Results Standard of Reference Diameter and volume are indicated as the average of readers at a tube current time level of 300 mas. The range of largest axial diameters of the 125 nodules was mm (mean, 3.11 ± 1.66 [SD] mm; median, 2.65 mm). The range of nodule volumes was mm 3 (mean, 58.1 ± 83.5 mm 3 ; median, 24.9 mm 3 ). Attenuation fluctuated from 800 to 212 HU (mean, ± 317 HU; median, 352 HU). Forty-nine of the 125 nodules (39%) were small (< 3 mm), and 66 (53%) were ground-glass nodules (< 300 HU). Dose-Dependent Sensitivity and Specificity The tube current times levels of 10, 20, 30, 40, and 100 mas did not exhibit a significant difference in individual reader sensitivity compared with the standard 300 mas (Table 1). At the lowest tube current level of 5 mas, only reader 1 had a significant decrease in sensitivity from 82% to 77% (p = ). Nodule detection by readers 2 and 3 decreased from 98% and 94% at 300 mas to 95% and 92% at 5 mas, but the difference was not statistically significant (p = 0.39 and p = 0.45, McNemar test). Individual specificities were high and unaffected by tube current 2se +2se 2sE +2sE +CI Limits of Agreement (corrected) CI 100% Fig. 1 Chart shows Bland-Altman method. Limits of agreement (LA) are volume measurement error (VME) ± 2 SD, indicating limits, where 95% of errors lie. LA are only estimates of values that apply to whole population. Second sample would give different limits. Therefore, CI of VME and LA better describe real limits for whole population (corrected LA). CI can be calculated with standard error of VME, which follows approximately normal distribution. +2sd +LA +CI AJR:197, September

4 Christe et al. TABLE 1: Sensitivity for Lung Nodules (n = 125) Calculated With McNemar Test Tube Current Time Reader 1 Reader 2 Reader 3 Setting (mas) Sensitivity p Sensitivity p Sensitivity p a 0.82 NA 0.98 NA 0.94 NA Note Bold type indicates lower sensitivities. NA = not available. a Reference standard. time level. Specificity at the standard dose was 96%, 89%, and 92% for readers 1, 2, and 3 and did not decrease at lower dose levels (p > 0.31). Influence of Nodule Size, Density, and Tube Current Time Setting on Sensitivity Sensitivity as a function of nodule characteristic is shown in Table 2. There were large individual differences in ability to detect individual lung nodules. Reader 1 had significantly lower sensitivity for smaller and groundglass nodules. The difference in sensitivity for ground-glass versus solid nodules even increased, from 22% at standard dose to 39% at the lowest dose level. The sensitivity of reader 1 for smaller nodules was 12 20% lower than that reader s sensitivity for larger nodules. Interestingly, the sensitivity of reader 2 was not statistically affected by nodule size, and nodule density influenced sensitivity only at dose levels lower than 30 mas. Reader 3 was influenced by nodule size but not by nodule density. The sensitivity for smaller nodules was 8 14% lower than the sensitivity for larger nodules. The readers had one thing in common: Sensitivity for all 125 nodules was not significantly influenced by dose level. The tube current time products exhibited different individual sensitivities (Table 1), which did not decrease significantly from 300 to 5 mas, according to results of the chi-square test (p > 0.25). The probability of a true-positive finding if a condition is present (size 3 mm) compared with the probability of a true-positive finding if the condition is not present (size < 3 mm) resulted in individual reader odds ratios of Only the odds ratio of reader 1 was significant (p = ). For variable density, the odds ratios were , only reader 1 having a statistically significant result (p = ). The odds ratios for variable tube current were a nonsignificant (p > 0.25). Therefore, the strongest negative effect on sensitivity can be assumed for attenuation less than 300 HU, followed by size less than 3 mm. Examples of small ground-glass nodules at different dose levels are shown in Figure 2. Logistic Regression The most significant predictor of sensitivity was nodule density (p < for all three readers). Nodule size significantly affected the sensitivity of reader 1 only (p = 0.02). Tube current time level did significantly influence the sensitivity of reader 1 alone (p = 0.007), but for readers 2 and 3, it was not a significant predictor of sensitivity (p = 0.20, p = 0.19) in the logistic regression model. Measurement of Nodule Size by Radiologists Agreement of nodule size for lower tube current time levels compared with reference standard Agreement between the 5 and 300 mas levels is presented; the relative volume differences are shown in Figure 3. The results for the other tube current time levels are listed in Table 3. The mean volume measurement error of two readers combined was 2.2% with a 95% CI of 7.8% to 3.4%; hence a systematic bias between the measurements cannot be assumed (zero is inside the 95% CI). The negative sign means that volume 2 (5 mas) was measured larger than volume 1 (300 mas). The lower limit of agreement was 39% (95% CI, 47% to 31%), and the upper limit was 35% (95% CI, 27 43%). An increase in nodule volume of more than 43% from the standard tube current time level (300 mas) to a hypothetical low-dose follow-up examination (5 mas) is more likely to indicate real nodule growth than interscan variability. Often the nodules were distorted at lower tube current time levels because of superimposed noise, and their diameter might have increased (Fig. 4) or decreased but would have had no statistical effect on the overall volume measurement. However, the longest axial diameter measurement exhibited significant TABLE 2: Sensitivity of Nodule Detection in Relation to Nodule Size and Attenuation Reader 5 mas 10 mas 20 mas 30 mas 40 mas 100 mas 300 mas Characteristic Sensitivity, 3 mm Sensitivity, < 3 mm Change in size sensitivity Sensitivity 300 HU Sensitivity < 300 HU Change in attenuation sensitivity Note Bold type indicates p < AJR:197, September 2011

5 CT of Lung Nodules Fig. 2 CT scans show ground-glass nodules at different tube current time levels. A, 54-year-old man with 2-mm ground-glass nodule (arrows) with attenuation of 584 HU detectable at all levels. B, 48-year-old man with 1.5-mm ground-glass nodule with attenuation of 647 HU clearly visible at standard and higher tube current time levels (white arrows) that disappears at lower levels (black arrows). Relative Volume Differences Relative Volume Differences differences for lower tube current time levels. The diameters were measured longer for levels of 100 mas and less (Fig. 4). The mean measurement error for 5 mas was 5.3% (95% CI, 7.5% to 3%) with corrected limits of agreement of 28% and 17%. Interreader and intrareader agreement of volume and diameter measurement With application of the same method for the two Mean Volume (mm 3 ) Mean Volume (mm 3 ) VME Mean VME Upper limit of agreement Lower limit of agreement VME Mean VME Upper limit of agreement Lower limit of agreement readers (reader 1 volume reader 2 volume / mean volume), the mean interobserver variability of volume measurements at the standard tube current time level of 300 mas was 38% (95% CI, 27 49%), meaning a significant difference between the two readers. The mean volume measurement of reader 1 was substantially larger. A lower limit of agreement of 51% (95% CI, 71% to 32%) and Fig. 3 Graphs show agreement of manual volume measurement of all nodules for standard and lowest tube current time levels and for different readers (interreader agreement). Mean difference of two readers (bottom) is much higher than mean difference between two currents (top), and scattering of volume differences is much larger for different readers than for different currents. Therefore, upper limit of agreement is greater than 100% for two readers and less than 50% for different tube current time settings. VME = volume measurement error. an upper limit of agreement of 127% (95% CI, %) were found. In absolute figures, these results meant a variability among the readers of 9.3 ± 20.7 mm 3 with limits of 32.1 mm 3 (95% CI, 41 to 23 mm 3 ) to 51 mm 3 (95% CI, mm 3 ) for an average absolute nodule volume of 58.1 mm 3 (mean of readers 1 and 2). Interobserver variability did not change substantially for lower tube current time levels (Table 3). Intraobserver agreement of volume measurement by reader 1 at the standard tube current time level was 3.9% (95% CI, 2.4% to 10%) with a lower limit of agreement of 35% (95% CI, 43% to 27%) and an upper limit of agreement of 43% (95% CI, 35 51%). The intraobserver measurement error for the longest diameter was 2.8% (95% CI, 0.1% to 5.8% and therefore not significant) and had a smaller limit of agreement according to the single dimension (compared with the volume) of 20% to 26%. The measurement errors for neither volume nor diameter differed substantially for lower tube current time levels (Table 3). Computer-Aided Volume Measurement Agreement on nodule size for lower tube current time levels compared with standard level Less agreement of volume and diameter measurement for the computer-aided le- AJR:197, September

6 Christe et al. TABLE 3: Manual Nodule Volume Measurement Errors and Corrected Limits of Agreement Tube Current Time Setting (mas) Volume Compared With 300 mas Intrareader Agreement Interreader Agreement Volume Measurement Error Limits of Agreement Volume Measurement Error Limits of Agreement Volume Measurement Error Limits of Agreement Mean Lower Upper Mean Lower Upper Mean Lower Upper Note Values are percentages. Bold type indicates significant differences in volume measurement. sion measurement solution resulted in higher variability and therefore wider limits of agreement. For example, at 5 mas, the mean volume measurement error was 6% (95% CI, 1.8% to 13.8%) with a lower limit of agreement of 58% (95% CI, 71% to 44%) and an upper limit of agreement of 70% (95% CI, 56 83%). Interreader and intrareader agreement on volume and diameter measurement Much better interreader agreement was found for the computer-aided lesion measurement solution than for manual measurements (Fig. 5). The volume measurement error at standard dose was 4.1%, and the corrected limits of agreement were 69% and 60%. In contrast to manual measurement, the lesion measurement solution exhibited no significant difference between the two readers. Good intrareader agreement with a similarly low rate of volume measurement error of 5.9% was found with corrected limits of agreement of 68% and 56%. The reader measurement error for longest diameter also exhibited better interreader and intrareader agreement with the computer-aided technique than without it. Human Versus Machine At each dose level, the readers measured a significantly larger volume than did the computer. At the standard dose level (300 mas) the readers measured 25% more, with corrected limits of agreement of 86% and 136%. The longest diameter was also 15.8% longer measured manually. Fig year-old man with metastases from bronchogenic carcinoma. CT scans show measurements of longest axial diameter (green line) at different tube current time levels in milliampereseconds (top left number in each image). Top left image shows 3D reconstruction for computer-aided volume measurement (LMS Lung/Track, Median Technologies). Increases in manually measured size are due to distortion of nodule from increasing superimposed noise. Computer-aided volume measurements (yellow dots) are shown at 300 and 5 mas. Although readers included larger parts of ground-glass transition zone in nodule diameter, computer-aided measurement had higher attenuation threshold level. Therefore, volumes measured with computer-aided technique were smaller than manually measured volumes. 628 AJR:197, September 2011

7 CT of Lung Nodules Volume 2 (mm 3 ) Volume 1 (mm 3 ) Discussion This study showed a significant difference between tube current time settings of 300 and 5 mas with respect to sensitivity of chest CT for lung nodules for one of the three readers and no significant change in sensitivity for doses of 100, 40, 30, 20, and 10 mas compared with 300 mas. These results suggest 10 mas is the threshold for diagnostic image quality in this study setting and confirm the results of other low-dose CT studies [7, 8]. Reducing the tube current time level from 300 to 10 mas automatically reduces the radiation dose received to an average of 3.3% of the standard dose. The risk of radiation-induced cancer death for one standard chest CT examination is estimated at approximately 1:4000 [25] and may hypothetically be reduced to 1:120,000 if a linear relation between dose and risk of cancer death is assumed [25]. A low-dose protocol would be of special interest to patients with lung nodules because the cumulative radiation dose of repetitive follow-up CT examinations might be reduced considerably. In our study, nodule characteristics had the most important negative effect on detectability of small pulmonary nodules, and the technical parameter tube current time product had less influence on sensitivity. Ko et al. [4] reported a similar odds ratio of 3.7 for nodule size (> 5 mm vs 5 mm) and an odds ratio of 3.2 for ground-glass opacity. One reader in our study had a higher odds ratio of 9.9 for nodule density, probably because of the selection of smaller nodules. Small ground-glass nodules are much more difficult to identify, leading to a larger gap of sensitivities. Volume 2 (mm 3 ) Volume 1 (mm 3 ) Fig. 5 Graphs show computer-aided (left) and manual (right) volumetric results for each nodule for reader 1 versus reader 2. Dashed lines show correlation between readers for same nodules. Solid line indicates y = x, that is, 100% correlation. Agreement between readers is much better with computer-aided than with manual technique. 500 The odds ratios of the tube current time levels exhibited little influence of dose on nodule sensitivity. In addition, in the reader-specific logistic regressions, the effect of tube current time level was significant only for reader 1. The McNemar test, which is the better test for comparing paired data alone, showed significantly lower sensitivity only at 5 mas for the same reader, indicating that the tube current time level may influence the detectability of lung nodules, but not as much as the nodule characteristics. The manual volume measurement in our study was much more influenced by interreader variability than by variability of tube current time level or intraobserver variability. The tube current time setting can be reduced from 300 to 5 mas with remarkable reproducibility of volume measurement: a volume measurement error of 2.2% compared with more than 32% for different readers at any tube current time level. The problem with volume-doubling rate calculations is not a change in technical parameters between scans but the change of readers. Gietema et al. [26] found narrower limits of agreement ( 21.2% to 23.8%) for interscan volume measurement; Wormanns et al. [27] and Hein et al. [28, 29] found narrower limits of agreement for automated interscan volume measurement ( 20.4% to 21.9% and 26.5% to 29.6%) and even narrower interobserver and intraobserver agreement. The narrower limits of agreement in these studies is probably due to the smaller tube current time gap (75 vs 5 mas) and the selection of larger nodules (mean diameter, 11 mm) whereby measurement error automatically leads to a smaller relative difference. We used corrected limits of agreement [22], which widened the range of agreement. The interobserver variability for manual measurement of 38% at the standard tube current time setting appears large for a volume but would be much reduced in evaluation of diameter. For a hypothetical spherical nodule with a diameter of 1 mm, a 38% decrease in volume would mean a reduction of the diameter to 0.85 mm. Taking large interreader variability into account, a clinically significant volume increase would be detected when the volume difference exceeded the 95% limit of agreement of approximately 146%. With an increase in volume measurement of greater than 146%, the probability of real growth, rather than interreader inaccuracy, would be 95% for small lung nodules. Measurement variability at different tube current time levels is much smaller. From a first examination at 300 mas to a hypothetical low-dose follow-up CT examination at 5 mas, a volume increase greater than 43% (95% limit of agreement) for the same reader would be 95% more likely to be real growth than variability between tube current time levels. The readers measured significantly larger volumes manually than with the computeraided lesion measurement solution. The readers probably tended to include large parts of the small ground-glass transition around the nodule into the measurement of the diameter. The computer algorithm, however, measured from fixed higher-attenuation threshold levels with the potential of being inappropriately small owing to elimination of partial volume at the lung interface. Over all, the lesion measurement solution was associated with slightly greater volume measurement error on average at the different dose levels but would be definitely superior in tumor follow-up CT with different readers, especially because the volume difference between two readers was significant for the manual measurement but not for the computer-aided lesion measurement solution. Although studies have shown imprecision in automated volume measurements and volume-doubling time estimation of pulmonary nodules [30], our study showed that volume-doubling time estimation would be more accurate if an automated measurement tool were used for both measurements than if individual manual measurements were made by two readers. A limitation of this study was that it did not capture depth of inspiration as a factor affecting nodule volume. In addition, the design whereby interscan variability is assessed by AJR:197, September

8 Christe et al. repeating scanning after very short time intervals and patients are required to leave the imaging table can strongly influence volume measurement, and this aspect was neglected in our analysis. In this study, low-dose images were simulated, and using a setting less than 50 mas with artificial noise might not have been an accurate representation of an actual 50-mAs scan because other issues beyond quantum mottle can affect image quality. Furthermore, to compensate for potential enhanced nodule detection for limited FOV and z-axis, only smaller nodules ( 8 mm) were selected. Whether our study design represents circumstances in daily clinical routine has yet to be investigated. Results of ongoing low-dose studies without simulation should clarify the answer to this question. Acknowledgment We thank the Department of Radiology, Stanford University, for providing infrastructure and staffing. References 1. Henschke CI, Naidich DP, Yankelevitz DF, et al. Early Lung Cancer Action Project: initial findings on repeat screenings. Cancer 2001; 92: Henschke CI, Yankelevitz DF, Mirtcheva R, et al. CT screening for lung cancer: frequency and significance of part-solid and nonsolid nodules. AJR 2002; 178: Lee IJ, Gamsu G, Czum J, Wu N, Johnson R, Chakrapani S. Lung nodule detection on chest CT: evaluation of a computer-aided detection (CAD) system. Korean J Radiol 2005; 6: Ko JP, Rusinek H, Naidich DP, et al. Wavelet compression of low-dose chest CT data: effect on lung nodule detection. Radiology 2003; 228: Hetmaniak Y, Bard JJ, Albuisson E, et al. Pulmonary nodules: dosimetric and clinical studies at low dose multidetector CT [in French]. J Radiol 2003; 84: Weng MJ, Wu MT, Pan HB, Kan YY, Yang CF. The feasibility of low-dose CT for pulmonary metastasis in patients with primary gynecologic malignancy. Clin Imaging 2004; 28: Gergely I, Neumann C, Reiger F, Dorffner R. Lung nodule detection with ultra-low-dose CT in routine follow-up of cancer patients [in German]. Rofo 2005; 177: Das M, Mühlenbruch G, Heinen S, et al. Performance evaluation of a computer-aided detection algorithm for solid pulmonary nodules in lowdose and standard-dose MDCT chest examinations and its influence on radiologists. Br J Radiol 2008; 81: Li F, Arimura H, Suzuki K, et al. Computer-aided detection of peripheral lung cancers missed at CT: ROC analyses without and with localization. Radiology 2005; 237: Armato SG, Li F, Giger ML, MacMahon H, Sone S, Doi K. Lung cancer: performance of automated lung nodule detection applied to cancers missed in a CT screening program. Radiology 2002; 225: Lee JW, Goo JM, Lee HJ, Kim JH, Kim S, Kim YT. The potential contribution of a computeraided detection system for lung nodule detection in multidetector row computed tomography. Invest Radiol 2004; 39: Lillington GA, Caskey CI. Evaluation and management of solitary and multiple pulmonary nodules. Clin Chest Med 1993; 14: Usuda K, Saito Y, Sagawa M, et al. Tumor doubling time and prognostic assessment of patients with primary lung cancer. Cancer 1994; 74: Bogot NR, Kazerooni EA, Kelly AM, Quint LE, Desjardins B, Nan B. Interobserver and intraobserver variability in the assessment of pulmonary nodule size on CT using film and computer display methods. Acad Radiol 2005; 12: Erasmus JJ, Gladish GW, Broemeling L, et al. Interobserver and intraobserver variability in measurement of non-small-cell carcinoma lung lesions: implications for assessment of tumor response. J Clin Oncol 2003; 21: Meyer CR, Johnson TD, McLennan G, et al. Evaluation of lung MDCT nodule annotation across radiologists and methods. Acad Radiol 2006; 13: Harris KM, Adams H, Lloyd DC, Harvey DJ. The effect on apparent size of simulated pulmonary nodules of using three standard CT window settings. Clin Radiol 1993; 47: Petkovska I, Brown MS, Goldin JG, et al. The effect of lung volume on nodule size on CT. Acad Radiol 2007; 14: Ko JP, Rusinek H, Jacobs EL, et al. Small pulmonary nodules: volume measurement at chest CT phantom study. Radiology 2003; 228: Way TW, Chan HP, Sahiner B, et al. Effect of CT scanning parameters on volumetric measurements of pulmonary nodules by 3D active contour segmentation: a phantom study. Phys Med Biol 2008; 53: Mayo JR, Whittall KP, Leung AN, et al. Simulated dose reduction in conventional chest CT: validation study. Radiology 1997; 202: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: Zar JH. Biostatistical analysis, 5th ed. Upper Saddle River, NJ: Pearson Prentice-Hall, 2010: Hosmer DW, Lemeshow S. Applied logistic regression, 2nd ed. New York, NY: Wiley, National Council on Radiation Protection and Measurements. Risk estimations for radiation protection. Bethesda, MD: NCRP Publications, 1993: NCRP report Gietema HA, Wang Y, Xu D, et al. Pulmonary nodules detected by lung cancer screening: interobserver variability of semiautomated volume measurements. Radiology 2006; 241: Wormanns D, Kohl G, Klotz E, et al. Volumetric measurements of pulmonary nodules at multi-row detector CT: in vivo reproducibility. Eur Radiol 2004; 14: Hein PA, Romano VC, Rogalla P, et al. Variability of semiautomated lung nodule volumetry on ultralow-dose CT: comparison with nodule volumetry on standard-dose CT. J Digit Imaging 2010; 23: Hein PA, Romano VC, Rogalla P, et al. Linear and volume measurements of pulmonary nodules at different CT dose levels: intrascan and interscan analysis. Rofo 2009; 181: Nietert PJ, Ravenel JG, Leue WM, et al. Imprecision in automated volume measurements of pulmonary nodules and its effect on the level of uncertainty in volume doubling time estimation. Chest 2009; 135: AJR:197, September 2011

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology

Chapter 6. Hester Gietema Cornelia Schaefer-Prokop Willem Mali Gerard Groenewegen Mathias Prokop. Accepted for publication in Radiology Chapter 6 Interscan variability of semiautomated volume measurements in intraparenchymal pulmonary nodules using multidetector-row computed tomography: Influence of inspirational level, nodule size and

More information

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays

Pulmonary Nodules: Contrast- Enhanced Volumetric Variation at Different CT Scan Delays Cardiopulmonary Imaging Original Research Rampinelli et al. Contrast-Enhanced CT of Pulmonary Nodules Cardiopulmonary Imaging Original Research Cristiano Rampinelli 1 Sara Raimondi 2 Mauro Padrenostro

More information

Computer-Aided Volumetry of Pulmonary Nodules Exhibiting Ground-Glass Opacity at MDCT

Computer-Aided Volumetry of Pulmonary Nodules Exhibiting Ground-Glass Opacity at MDCT Cardiopulmonary Imaging Original Research Oda et al. MDCT and Volumetry of Pulmonary Nodules Cardiopulmonary Imaging Original Research Computer-Aided Volumetry of Pulmonary Nodules Exhibiting Ground-Glass

More information

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology CT of Pulmonary Nodules Chest Imaging Original Research Myria Petrou 1 Leslie E. Quint 1 in Nan 2 Laurence H. aker 3 Petrou M, Quint LE, Nan, aker LH Keywords: chest, lung disease, MDCT, oncologic imaging,

More information

Identification of Missed Pulmonary Nodules on Low Dose CT Lung Cancer Screening Studies Using an Automatic Detection System

Identification of Missed Pulmonary Nodules on Low Dose CT Lung Cancer Screening Studies Using an Automatic Detection System Identification of Missed Pulmonary Nodules on Low Dose CT Lung Cancer Screening Studies Using an Automatic Detection System Carol L. Novak *a, Li Fan a, Jianzhong Qian a, Guo-Qing Wei a, David P. Naidich

More information

Chapter 5. Pulmonary nodules detected at lung cancer screening: Interobserver variability of semiautomated volume measurements

Chapter 5. Pulmonary nodules detected at lung cancer screening: Interobserver variability of semiautomated volume measurements Chapter 5 Pulmonary nodules detected at lung cancer screening: Interobserver variability of semiautomated volume measurements Hester Gietema Ying Wang Dongming Xu Rob van Klaveren Harry de Koning Ernst

More information

Malignant solitary pulmonary nodules: assessment of mass growth rate and doubling time at follow-up CT

Malignant solitary pulmonary nodules: assessment of mass growth rate and doubling time at follow-up CT Original Article Malignant solitary pulmonary nodules: assessment of mass growth rate and doubling time at follow-up CT Jingxu Li*, Tingting Xia*, Xinguan Yang, Xiao Dong, Jiamin Liang, Nanshan Zhong,

More information

Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening

Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening Guidelines for the Management of Pulmonary Nodules Detected by Low-dose CT Lung Cancer Screening 1. Introduction In January 2005, the Committee for Preparation of Clinical Practice Guidelines for the Management

More information

Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times

Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times Small Pulmonary Nodules: Our Preliminary Experience in Volumetric Analysis of Doubling Times Andrea Borghesi, MD Davide Farina, MD Roberto Maroldi, MD Department of Radiology University of Brescia Brescia,

More information

Methods and validation of nodule measurement in a lung cancer screening Ying, Wang

Methods and validation of nodule measurement in a lung cancer screening Ying, Wang University of Groningen Methods and validation of nodule measurement in a lung cancer screening Ying, Wang IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish

More information

Measurement error of spiral CT Volumetry:

Measurement error of spiral CT Volumetry: Measurement error of spiral CT Volumetry: Influence of Low Dose CT Technique 1 Tae Gyu Lee, M.D. 2, Myung Jin Chung, M.D., Sung Bum Cho, M.D. 2, Jae Min Cho, M.D., Seog Joon Kim, M.D. 2, Sang Hyun Baik,

More information

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008:

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008: Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008: Computer Aided Diagnosis and is made available as an

More information

Ryutaro Kakinuma 1,2,3, Yukio Muramatsu 1,4, Junta Yamamichi 1,5, Shiho Gomi 1,6, Estanislao Oubel 7, Noriyuki Moriyama 1,8.

Ryutaro Kakinuma 1,2,3, Yukio Muramatsu 1,4, Junta Yamamichi 1,5, Shiho Gomi 1,6, Estanislao Oubel 7, Noriyuki Moriyama 1,8. Original Article Evaluation of the 95% limits of agreement of the volumes of 5-year clinically stable solid nodules for the development of a follow-up system for indeterminate solid nodules in CT lung

More information

Copyright 2007 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, volume 6514, Medical Imaging

Copyright 2007 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, volume 6514, Medical Imaging Copyright 2007 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, volume 6514, Medical Imaging 2007: Computer Aided Diagnosis and is made available as

More information

Copyright 2009 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 7260, Medical Imaging 2009:

Copyright 2009 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 7260, Medical Imaging 2009: Copyright 2009 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 7260, Medical Imaging 2009: Computer Aided Diagnosis and is made available as an

More information

Outcomes in the NLST. Health system infrastructure needs to implement screening

Outcomes in the NLST. Health system infrastructure needs to implement screening Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have

More information

Lung nodule volumetry: segmentation algorithms within the same software package cannot be used interchangeably

Lung nodule volumetry: segmentation algorithms within the same software package cannot be used interchangeably Eur Radiol (2010) 20: 1878 1885 DOI 10.1007/s00330-010-1749-z COMPUTED TOMOGRAPHY H. Ashraf B. de Hoop S. B. Shaker A. Dirksen K. S. Bach H. Hansen M. Prokop J. H. Pedersen Lung nodule volumetry: segmentation

More information

Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2

Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2 Low-dose CT Lung Cancer Screening Guidelines for Pulmonary Nodules Management Version 2 The Committee for Management of CT-screening-detected Pulmonary Nodules 2009-2011 The Japanese Society of CT Screening

More information

Copyright 2007 IEEE. Reprinted from 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro, April 2007.

Copyright 2007 IEEE. Reprinted from 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro, April 2007. Copyright 27 IEEE. Reprinted from 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro, April 27. This material is posted here with permission of the IEEE. Such permission of the

More information

Computer Assisted Radiology and Surgery

Computer Assisted Radiology and Surgery Computer Assisted Radiology and Surgery How Can a Massive Training Artificial Neural Network (MTANN) Be Trained With a Small Number of Cases in the Distinction Between Nodules and Vessels in Thoracic CT?

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

New Horizons in the Imaging of the Lung

New Horizons in the Imaging of the Lung New Horizons in the Imaging of the Lung Postprocessing. How to do it and when do we need it? Peter M.A. van Ooijen, MSc, PhD Principal Investigator, Radiology, UMCG Discipline Leader Medical Imaging Informatics

More information

Computer-Aided Detection in Screening CT for

Computer-Aided Detection in Screening CT for Yuan et al. Screening CT for Pulmonary Chest Imaging Original Research C M E D E N T U R I C L I M G I N G JR 2006; 186:1280 1287 0361 803X/06/1865 1280 merican Roentgen Ray Society Y O Ren Yuan 1 Patrick

More information

Pulmonary Nodule Size Evaluation with Chest Tomosynthesis 1

Pulmonary Nodule Size Evaluation with Chest Tomosynthesis 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Åse A. Johnsson,

More information

Cardiopulmonary Imaging Original Research

Cardiopulmonary Imaging Original Research Cardiopulmonary Imaging Original Research Ebner et al. MIP and CAD Algorithms in CT for Lung Cancer Screening Cardiopulmonary Imaging Original Research Lukas Ebner 1 Justus E. Roos 1 Jared D. Christensen

More information

Volume versus diameter assessment of small pulmonary nodules in CT lung cancer screening

Volume versus diameter assessment of small pulmonary nodules in CT lung cancer screening Review Article Volume versus diameter assessment of small pulmonary nodules in CT lung cancer screening Daiwei Han, Marjolein A. Heuvelmans, Matthijs Oudkerk University of Groningen, University Medical

More information

Xiaohuan Pan 1,2 *, Xinguan Yang 1,2 *, Jingxu Li 1,2, Xiao Dong 1,2, Jianxing He 2,3, Yubao Guan 1,2. Original Article

Xiaohuan Pan 1,2 *, Xinguan Yang 1,2 *, Jingxu Li 1,2, Xiao Dong 1,2, Jianxing He 2,3, Yubao Guan 1,2. Original Article Original Article Is a 5-mm diameter an appropriate cut-off value for the diagnosis of atypical adenomatous hyperplasia and adenocarcinoma in situ on chest computed tomography and pathological examination?

More information

LUNG CANCER continues to rank as the leading cause

LUNG CANCER continues to rank as the leading cause 1138 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 24, NO. 9, SEPTEMBER 2005 Computer-Aided Diagnostic Scheme for Distinction Between Benign and Malignant Nodules in Thoracic Low-Dose CT by Use of Massive

More information

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008:

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008: Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008: Computer Aided Diagnosis and is made available as an

More information

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? Journal of Physics: Conference Series PAPER OPEN ACCESS Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose? To cite this article: M Smarda et

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Volume and Mass Doubling Times of Persistent Pulmonary Subsolid Nodules Detected in Patients without Known Malignancy 1

Volume and Mass Doubling Times of Persistent Pulmonary Subsolid Nodules Detected in Patients without Known Malignancy 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Original Research

More information

Small solid noncalcified pulmonary nodules detected by screening chest computed tomography

Small solid noncalcified pulmonary nodules detected by screening chest computed tomography Respiratory Medicine (2007) 101, 1880 1884 Small solid noncalcified pulmonary nodules detected by screening chest computed tomography Sang-Man Jin a,b, Seung-Ho Choi c, Chul-Gyu Yoo a,b, Young-Whan Kim

More information

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software*

Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 3, Summer 2010 Improved image quality of low-dose thoracic CT examinations with a new postprocessing software* Anne Catrine Traegde Martinsen,

More information

Estimating Iodine Concentration from CT Number Enhancement

Estimating Iodine Concentration from CT Number Enhancement Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Prevent Cancer Foundation Quantitative Imaging Workshop XIII

Prevent Cancer Foundation Quantitative Imaging Workshop XIII Status of the Quantitative Imaging Profile Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening Prevent Cancer Foundation Quantitative Imaging Workshop XIII June 13-14, 2016 David S. Gierada,

More information

B. CT protocols for the spine

B. CT protocols for the spine B. CT protocols for the spine Poster No.: A-003 Congress: ECR 2010 Type: Invited Speaker Topic: Neuro Authors: B. Tins; Oswestry/UK Keywords: CT, spine, diagnostic imaging protocol DOI: 10.1594/ecr2010/A-003

More information

The latest developments - Automated Breast Volume Scanning. Dr. med. M. Golatta

The latest developments - Automated Breast Volume Scanning. Dr. med. M. Golatta The latest developments - Automated Breast Volume Scanning Dr. med. M. Golatta Automated Breast Volume US: Why? o Mammography is limited in dense breasts: high false negative rate o Many of these tumors

More information

Chapter 11. Summary and general discussion

Chapter 11. Summary and general discussion Chapter 11 Summary and general discussion Low Dose Computed Tomography of the Chest: Applications and Limitations INTRODUCTION The introduction of spiral, multidetector-row computed tomography (CT) has

More information

CT Screening for Lung Cancer: Frequency and Significance of Part-Solid and Nonsolid Nodules

CT Screening for Lung Cancer: Frequency and Significance of Part-Solid and Nonsolid Nodules Claudia I. Henschke 1 David F. Yankelevitz 1 Rosna Mirtcheva 1 Georgeann McGuinness 2 Dorothy McCauley 1 0lli S. Miettinen 3 for the ELCAP Group Received June 19, 2001; accepted after revision November

More information

doi: /

doi: / Yiting Xie ; Matthew D. Cham ; Claudia Henschke ; David Yankelevitz ; Anthony P. Reeves; Automated coronary artery calcification detection on low-dose chest CT images. Proc. SPIE 9035, Medical Imaging

More information

Effect of CT Image Compression on Computer-assisted Lung Nodule Volume Measurement 1

Effect of CT Image Compression on Computer-assisted Lung Nodule Volume Measurement 1 Computer Applications Radiology Jane P. Ko, MD Jeffrey Chang, MD Elan Bomsztyk, BS James S. Babb, PhD David P. Naidich, MD Henry Rusinek, PhD Published online before print 10.1148/radiol.2371041079 Radiology

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Institutional review board approval was obtained prior to the start of this study.

Institutional review board approval was obtained prior to the start of this study. Lower Limb Alignment and Length Measurements - Comparison of Computed Tomography, Upright Full-Length Conventional Radiography and Upright Biplanar Linear-Low Dose X-ray Scanner Poster No.: C-1382 Congress:

More information

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study

Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Poster No.: C-1207 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Shimomiya, M. Kondo, M.

More information

Segmentation of nodules on chest computed tomography for growth assessment

Segmentation of nodules on chest computed tomography for growth assessment Segmentation of nodules on chest computed tomography for growth assessment William Mullally, a) Margrit Betke, and Jingbin Wang Computer Science Department, Boston University, Boston, Massachusetts 02215

More information

Radiologic assessment of response of tumors to treatment. Copyright 2008 TIMC, Matthew A. Barish M.D. All rights reserved. 1

Radiologic assessment of response of tumors to treatment. Copyright 2008 TIMC, Matthew A. Barish M.D. All rights reserved. 1 Radiologic assessment of response of tumors to treatment Copyright 2008 TIMC, Matthew A. Barish M.D. All rights reserved. 1 Objective response assessment is important to describe the treatment effect of

More information

MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES

MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES Although response is not the primary endpoint of this trial, subjects with measurable disease will be assessed by standard criteria. For the purposes of this

More information

AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS

AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS Isaac N. Bankman', William A. Christens-Barryl, Irving N. Weinberg2, Dong W. Kim3, Ralph D. Semmell, and William R. Brody2 The Johns Hopkins

More information

Recommendations for Measuring Pulmonary Nodules at CT: A Statement from the Fleischner Society 1

Recommendations for Measuring Pulmonary Nodules at CT: A Statement from the Fleischner Society 1 This copy is for personal use only. To order printed copies, contact reprints@rsna.org Alexander A. Bankier, MD, PhD Heber MacMahon, MB, BCh Jin Mo Goo, MD, PhD Geoffrey D. Rubin, MD Cornelia M. Schaefer-Prokop,

More information

Evaluation of Lung Cancer Response: Current Practice and Advances

Evaluation of Lung Cancer Response: Current Practice and Advances Evaluation of Lung Cancer Response: Current Practice and Advances Jeremy J. Erasmus I have no financial relationships, arrangements or affiliations and this presentation will not include discussion of

More information

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 4, APRIL

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 4, APRIL IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 25, NO. 4, APRIL 2006 435 On Measuring the Change in Size of Pulmonary Nodules Anthony P. Reeves*, Senior Member, IEEE, Antoni B. Chan, David F. Yankelevitz,

More information

Performance of computer-aided detection of pulmonary nodules in low-dose CT: comparison with double reading by nodule volume

Performance of computer-aided detection of pulmonary nodules in low-dose CT: comparison with double reading by nodule volume Eur Radiol (2012) 22:2076 2084 DOI 10.1007/s00330-012-2437-y CHEST Performance of computer-aided detection of pulmonary nodules in low-dose CT: comparison with double reading by nodule volume Yingru Zhao

More information

Comparison of CT findings between MDR-TB and XDR-TB

Comparison of CT findings between MDR-TB and XDR-TB Comparison of CT findings between MDR-TB and XDR-TB Poster No.: C-0757 Congress: ECR 2017 Type: Authors: Keywords: DOI: Scientific Exhibit K. Yoon, H. Soohee; Changwon-si/KR Thorax, Lung, Respiratory system,

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Small Irregular Pulmonary Nodules in Low-Dose CT: Observer Detection Sensitivity and Volumetry Accuracy

Small Irregular Pulmonary Nodules in Low-Dose CT: Observer Detection Sensitivity and Volumetry Accuracy Cardiopulmonary Imaging Original Research Xie et al. CT of Phantom Model of Small Irregular Pulmonary Nodules Cardiopulmonary Imaging Original Research FOCUS ON: Xueqian Xie 1,2 Martin J. Willemink 3 Pim

More information

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies Research plan submitted for approval as a PhD thesis Submitted by: Refael Vivanti Supervisor: Professor Leo Joskowicz School of Engineering and Computer Science, The Hebrew University of Jerusalem Computer

More information

Comparison of RECIST version 1.0 and 1.1 in assessment of tumor response by computed tomography in advanced gastric cancer

Comparison of RECIST version 1.0 and 1.1 in assessment of tumor response by computed tomography in advanced gastric cancer Original Article Comparison of RECIST version 1.0 and 1.1 in assessment of tumor response by computed tomography in advanced gastric cancer Gil-Su Jang 1 *, Min-Jeong Kim 2 *, Hong-Il Ha 2, Jung Han Kim

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/183711

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Small Pulmonary Nodules: Volume Measurement at Chest CT Phantom Study 1

Small Pulmonary Nodules: Volume Measurement at Chest CT Phantom Study 1 Jane P. Ko, MD Henry Rusinek, PhD Erika L. Jacobs, MD James S. Babb, PhD Margrit Betke, PhD Georgeann McGuinness, MD David P. Naidich, MD Index terms: Computed tomography (CT), image processing, 60.12117

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Lung nodules: size still matters

Lung nodules: size still matters SERIES RADIOLOGY Lung nodules: size still matters Anna Rita Larici, Alessandra Farchione, Paola Franchi, Mario Ciliberto, Giuseppe Cicchetti, Lucio Calandriello, Annemilia del Ciello and Lorenzo Bonomo

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND

PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT OF VASCULAR OCCLUSION EXTENT AND LOCALIZATION OF EMBOLI 1. BACKGROUND JOURNAL OF MEDICAL INFORMATICS & TECHNOLOGIES Vol. 11/2007, ISSN 1642-6037 Damian PTAK * pulmonary embolism, AngioCT, postprocessing techniques, Mastora score PULMONARY EMBOLISM ANGIOCT (CTA) ASSESSMENT

More information

Combined pulmonary fibrosis and emphysema; prevalence and follow up among health-care personnel

Combined pulmonary fibrosis and emphysema; prevalence and follow up among health-care personnel Combined pulmonary fibrosis and emphysema; prevalence and follow up among health-care personnel Poster No.: C-0698 Congress: ECR 2013 Type: Scientific Exhibit Authors: K. Chae, G. Jin, S. Chon, Y. Lee;

More information

Automatic Ascending Aorta Detection in CTA Datasets

Automatic Ascending Aorta Detection in CTA Datasets Automatic Ascending Aorta Detection in CTA Datasets Stefan C. Saur 1, Caroline Kühnel 2, Tobias Boskamp 2, Gábor Székely 1, Philippe Cattin 1,3 1 Computer Vision Laboratory, ETH Zurich, 8092 Zurich, Switzerland

More information

Cancer Cells Detection using OTSU Threshold Algorithm

Cancer Cells Detection using OTSU Threshold Algorithm Cancer Cells Detection using OTSU Threshold Algorithm Nalluri Sunny 1 Velagapudi Ramakrishna Siddhartha Engineering College Mithinti Srikanth 2 Velagapudi Ramakrishna Siddhartha Engineering College Kodali

More information

A comparative study of machine learning methods for lung diseases diagnosis by computerized digital imaging'"

A comparative study of machine learning methods for lung diseases diagnosis by computerized digital imaging' A comparative study of machine learning methods for lung diseases diagnosis by computerized digital imaging'" Suk Ho Kang**. Youngjoo Lee*** Aostract I\.' New Work to be 1 Introduction Presented U Mater~al

More information

Radiographic Assessment of Response An Overview of RECIST v1.1

Radiographic Assessment of Response An Overview of RECIST v1.1 Radiographic Assessment of Response An Overview of RECIST v1.1 Stephen Liu, MD Georgetown University May 15 th, 2015 Presentation Objectives To understand the purpose of RECIST guidelines To describe the

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

National Cancer Institute

National Cancer Institute National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,

More information

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD Update of Digital Breast Tomosynthesis Susan Orel Roth, MD NCI estimates that : Why DBT? Approximately 20% of breast cancers are missed at mammography screening Average recall rates approximately 10%

More information

Subsolid pulmonary nodule morphology and associated patient characteristics in a routine clinical population

Subsolid pulmonary nodule morphology and associated patient characteristics in a routine clinical population Eur Radiol (2017) 27:689 696 DOI 10.1007/s00330-016-4429-9 CHEST Subsolid pulmonary nodule morphology and associated patient characteristics in a routine clinical population Onno M. Mets 1 & Pim A. de

More information

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008:

Copyright 2008 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 6915, Medical Imaging 2008: Copyright 28 Society of Photo Optical Instrumentation Engineers. This paper was published in Proceedings of SPIE, vol. 695, Medical Imaging 28: Computer Aided Diagnosis and is made available as an electronic

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Regan EA, Lynch DA, Curran-Everett D, et al; Genetic Epidemiology of COPD (COPDGene) Investigators. Clinical and radiologic disease in smokers with normal spirometry. Published

More information

Early Detection of Lung Cancer

Early Detection of Lung Cancer Early Detection of Lung Cancer Aswathy N Iyer Dept Of Electronics And Communication Engineering Lymie Jose Dept Of Electronics And Communication Engineering Anumol Thomas Dept Of Electronics And Communication

More information

Subsolid lung nodules, also termed ground-glass nodules

Subsolid lung nodules, also termed ground-glass nodules ORIGINAL ARTICLE Long-Term Surveillance of Ground-Glass Nodules Evidence from the MILD Trial Silva Mario, MD,* Sverzellati Nicola, MD, PhD,* Manna Carmelinda, MD,* Negrini Giulio, MD,* Marchianò Alfonso,

More information

With the introduction of multidetector computed tomography (CT) and the ongoing advances in multidetector CT technology, the detection of small pulmon

With the introduction of multidetector computed tomography (CT) and the ongoing advances in multidetector CT technology, the detection of small pulmon Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Effect of Nodule

More information

By choosing to view this document, you agree to all provisions of the copyright laws protecting it.

By choosing to view this document, you agree to all provisions of the copyright laws protecting it. Copyright 2009 IEEE. Reprinted from 31 st Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2009. EMBC 2009. Sept. 2009. This material is posted here with permission

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

Doubling Time of Thymic Epithelial Tumors Correlates With World Health Organization Histopathologic Classification

Doubling Time of Thymic Epithelial Tumors Correlates With World Health Organization Histopathologic Classification Cardiopulmonary Imaging Original Research Jeong et al. Thymic Epithelial Tumors Cardiopulmonary Imaging Original Research JOURNAL CLUB Dong Young Jeong 1 Kyung Soo Lee 1 Myung Jin Chung 1 Jae Ill Zo 2

More information

NIH Public Access Author Manuscript AJR Am J Roentgenol. Author manuscript; available in PMC 2011 July 6.

NIH Public Access Author Manuscript AJR Am J Roentgenol. Author manuscript; available in PMC 2011 July 6. NIH Public Access Author Manuscript Published in final edited form as: AJR Am J Roentgenol. 2010 September ; 195(3): W221 W228. doi:10.2214/ajr.09.3928. New Response Evaluation Criteria in Solid Tumors

More information

Precision of manual two-dimensional segmentations of lung and liver metastases and its impact on tumour response assessment using RECIST 1.

Precision of manual two-dimensional segmentations of lung and liver metastases and its impact on tumour response assessment using RECIST 1. Cornelis et al. European Radiology Experimental (2017) 1:16 DOI 10.1186/s41747-017-0015-4 European Radiology Experimental ORIGINAL ARTICLE Precision of manual two-dimensional segmentations of lung and

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Loren Ketai, MD; Mathurn Malby, BS; Kirk Jordan, MD; Andrew Meholic, MD; and Julie Locken, MD

Loren Ketai, MD; Mathurn Malby, BS; Kirk Jordan, MD; Andrew Meholic, MD; and Julie Locken, MD Small Nodules Detected on Chest Radiography* Does Size Predict Calcification? Loren Ketai, MD; Mathurn Malby, BS; Kirk Jordan, MD; Andrew Meholic, MD; and Julie Locken, MD Study objectives: To determine

More information

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects

Method Comparison for Interrater Reliability of an Image Processing Technique in Epilepsy Subjects 22nd International Congress on Modelling and Simulation, Hobart, Tasmania, Australia, 3 to 8 December 2017 mssanz.org.au/modsim2017 Method Comparison for Interrater Reliability of an Image Processing Technique

More information

doi: /

doi: / Yiting Xie ; Yu Maw Htwe ; Jennifer Padgett ; Claudia Henschke ; David Yankelevitz ; Anthony P. Reeves; Automated aortic calcification detection in low-dose chest CT images. Proc. SPIE 9035, Medical Imaging

More information

Welcome to the RECIST 1.1 Quick Reference

Welcome to the RECIST 1.1 Quick Reference Welcome to the RECIST 1.1 Quick Reference *Eisenhauer, E. A., et al. New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1). Eur J Cancer 2009;45:228-47. Subject Eligibility

More information

Low-Dose CT: Clinical Studies & the Radiologist Perspective

Low-Dose CT: Clinical Studies & the Radiologist Perspective Low-Dose CT: Clinical Studies & the Radiologist Perspective RD-ASiR RD-MBIR SD-FBP RD=0.35 msv (80% dose reduction) Perry J. Pickhardt, MD UW School of Medicine & Public Health Low-Dose CT: Clinical Overview

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Original Article Thoracic Imaging

Original Article Thoracic Imaging Original Article Thoracic Imaging https://doi.org/10.3348/kjr.2018.19.6.1179 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2018;19(6):1179-1186 Size-Specific Dose Estimation In the Korean Lung Cancer

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information