MRI evaluation of metal acupuncture needles

Size: px
Start display at page:

Download "MRI evaluation of metal acupuncture needles"

Transcription

1 1 Paul C Lauterbur Research Centers for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China 2 School of Electronic and Information Engineering, South China University of Technology, Guangzhou, China 3 Shenzhen International TCM Training Center, Shenzhen Hospital of Traditional Chinese Medicine, Shenzhen, Guangdong, China 4 Department of Rehabilitation, Shenzhen Nanshan Hospital, Shenzhen, Guangdong, China 5 Bastyr University Research Institute, Bastyr University, Kenmore, Washington, USA 6 Department of Radiology, University of Washington School of Medicine, Seattle, Washington, USA Correspondence to Professor Bensheng Qiu, and Dr Xiaojing Long, Paul C Lauterbur Research Centers for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, 1068 Xueyuan Boulevard, University Town of Shenzhen, Xili Nanshan, Shenzhen, Guangdong , China; bqiu@ustc.edu.cn and xj.long@ siat.ac.cn Received 6 May 2013 Accepted 1 August 2013 To cite: Mei L, Long X, Diao Y, et al. Acupunct Med Published Online First: [ please include Day Month Year] doi: /acupmed MRI evaluation of metal acupuncture needles Ling Mei, 1,2 Xiaojing Long, 1 Yanjun Diao, 1,3 Haibo Yu, 3 Wanzhang Yang, 4 Leanna J Standish, 5,6 Bensheng Qiu 1,5 ABSTRACT Objective To evaluate the MR compatibility of three metal acupuncture needles (a standard stainless steel needle, a gold needle and an austenitic stainless steel needle) by comparing their imaging artefacts, radiofrequency heating effects and ease of operation. Methods The MRI artefacts of the three metal needles were first evaluated by placing them in an agar gel phantom and performing MRI of the phantom. The increase in temperature during MRI was recorded using an MR-compatible fibreoptic thermometer. MRI of acupuncture at SP6 was performed using the MR-compatible gold needle and the austenitic stainless steel needle. Results The standard stainless steel acupuncture needle produced large imaging artefacts on MRI. The gold needle was superior for MRI but not rigid enough for some clinical applications such as scalp acupuncture. The austenitic stainless steel needle is non-ferromagnetic and compatible with MRI. None of these acupuncture needles introduced radiofrequency heating during MRI. Conclusions The evaluation of MR compatibility showed that gold and austenitic stainless steel needles are MR-compatible and therefore can be used for MRI of acupuncture. INTRODUCTION Acupuncture, an important component of Traditional Chinese Medicine, has gradually been recognised and accepted by the Western medical community. However, the underlying mechanisms of acupuncture have not yet been fully established. With the development of functional MRI (fmri), investigators have begun to explore the correlation between acupuncture points and specific functional areas of the brain. 1 7 Traditional metal needles made of ferromagnetic materials 89 are not MR-compatible and generate large artefacts on MRI. Because of the lack of MR-compatible acupuncture needles, researchers have been unable to study the effects of stimulation acupuncture points on the upper torso or the head, such as GV24 Shenting, LI20Yingxiang and GV14 Dazhui, or scalp acupuncture points used for stroke treatment using MRI. Ferromagnetic needles experience magnetic force within the MRI scanner which may lead to needles being pulled out of tissues and pose a risk to the subject, especially to the head or eyes. In addition, metal acupuncture needles may act as electrical conductors and induce a temperature increase during radiofrequency (RF) transmission in MRI. 10 In order for MRI technology to be applied to acupuncture research, needles must be non-magnetic and non-susceptible to RF-related heating that could damage tissues adjacent to the inserted needle. Beissner et al 11 recently evaluated several acupuncture needles and concluded that conventional acupuncture needles should not be used in the MRI environment and suggested that non-ferromagnetic or plastic needles may be used as alternatives. However, the plastic needle they assessed has some disadvantages in comparison with metal needles for example, it cannot be used for electroacupuncture. In this study we performed experiments to evaluate MR compatibility of three metal acupuncture needles, including a standard stainless steel needle, a gold needle and an austenitic stainless steel needle, by comparing their imaging artefacts, RF heating effects, cost and ease of operation. In vivo MRI of acupuncture at the lower leg acupuncture point SP6 Sanyinjiao was performed using MR-compatible gold and austenitic stainless steel needles. MATERIALS AND METHODS Metal acupuncture needles The metal acupuncture needles used in this study were a traditional stainless steel acupuncture needle, a 75% pure gold Mei L, et al. Acupunct Med 2013;0:1 5. doi: /acupmed Copyright 2013 by British Medical Journal Publishing Group.

2 needle and an austenitic stainless steel needle customised for MRI of acupuncture. Details of these three needles are summarised in table 1. The length and diameter of the standard stainless steel needle and the gold needle were 4 cm and 0.3 mm, respectively, whereas the austenitic stainless steel needle was slightly longer and thicker (5 cm long with a diameter of 0.32 mm). In vitro MRI evaluation of metal acupuncture needles Experiments were performed to compare MRI artefacts of the needles and to measure temperature increases in an agar gel phantom during MRI scanning. MRI scans were obtained on a whole body 3 T MRI scanner (MAGNETOM Trio Tim, Siemens, Erlangen, Germany) equipped with a body coil for RF transmission. The needle was inserted into the phantom and a gradient-echo fast low angle shot (FLASH) sequence was used to obtain T1-weighted MRI with TR 100 ms, TE 2.86 ms, field of view (FOV) 150 mm 117 mm, flip angle 70, matrix size , slice thickness 5 mm and in-plane resolution 0.6 mm 0.5 mm. Five slices around the needle were obtained to evaluate the imaging artefacts of the needles. To measure the temperature variation during MRI scanning, the needle was removed from the phantom and attached to an MR-compatible fibreoptic thermometer with a sensitivity of 0.01 C. The tip of the thermometer, where the sensor is situated, was attached to the needle and the attached needle and thermometer were inserted into the phantom. The temperature was recorded every 5 s during MRI scanning. The MRI pulse sequences included: (1) a T2*-weighted echo planar imaging (EPI) sequence with TR 2000 ms, TE 30 ms, FOV 150 mm 150 mm, flip angle 90 and slice thickness 3 mm; and (2) a T1-weighted gradient-echo FLASH sequence with TR 8 ms, TE 2.86 ms, FOV 150 mm 117 mm, flip angle 70, matrix size and slice thickness 3 mm, which can produce a high specific absorption rate (SAR) to induce a potential heating effect. In both MRI sequences the number of measurements was set to 700 and the total scanning time was 23 min and 32 s (700 TRs). The temperature records were processed using MATLAB 2010 (Matlab, Asheboro, North Table 1 Technical specifications of the three acupuncture needles used in the study Attribution Carolina, USA), in which the temperatures at every 12 consecutive time points were averaged; 23 average temperatures were obtained and presented as a curve of temperature versus imaging time. In vivo MRI of acupuncture using metal needles After MR evaluation of the compatibility of the three needles in the phantom, the gold needle and the austenitic stainless steel needle were used for in vivo MRI of human acupuncture; the standard stainless needle was not used because of the in vitro imaging artefact (see below). SP6, located on the medial side of the leg 10 cm above the tip of the medial malleolus and posterior to the medial border of the tibia, was chosen for needle insertion by a professional acupuncturist. Two orientations, the coronal and transversal planes, were scanned using a fast spin echo sequence with TR 700 ms, TE 27 ms, flip angle 120, matrix size and slice thickness 2 mm The FOV of the coronal plane was 300 mm 201 mm, while the FOV of the transversal plane was 300 mm 120 mm. After MRI of acupuncture, the needle was removed from the acupuncture point. RESULTS In vitro phantom experiment The T1-weighted MRI of acupuncture needles in the phantom are shown in figure 1. The MRI results show that the standard stainless steel needle produced a large artefact, where most signal generated by the needle and its surrounding was missed. The gold needle generated a relatively small artefact, and the austenitic stainless steel needle also induced relatively small artefact on the images, which provide clear data at most regions within the FOV. The temperature increases of the three needles during 700-TR MR scanning are shown in figures 2 and 3. For the EPI sequence, all of the temperature curves were almost horizontal, indicating no RF heating risk when using these metal needles (figure 2). For the FLASH sequence, the slope of the increase in temperature was slightly larger than for EPI scanning (figure 3), but the temperature variation was still small and should be safe for patients. The temperature evaluation of the three needles was repeated at least four times. Type Material Manufacturer Length (cm) Diameter (mm) Standard stainless steel Standard stainless steel Suzhou Acupuncture and Moxibustion Appliance Co needle Ltd, China Gold needle Gold needle with a silver handle Suzhou Medical Appliance factory, China Austenitic stainless steel Austenitic stainless steel with a copper Suzhou Acupuncture and Moxibustion Appliance Co needle handle Ltd, China 2 Mei L, et al. Acupunct Med 2013;0:1 5. doi: /acupmed

3 Figure 1 MRI of an agar gel medium with different acupuncture needles: (A) standard stainless steel needle produced a large imaging artefact; (B) gold needle generated a small artefact; and (C) austenitic stainless steel needle produced some artefact on the MRI. Comparing the temperature changes of the three metal needles, we found that all the temperatures tended to increase, but the maximum rise was less than 0.5 C and should not be perceptible to the patient nor influence the MRI. In vivo MRI of acupuncture The coronal and transversal planes of the in vivo MRI are shown in figure 4. The MRI showed that the gold needle (figure 4A) produced a smaller artefact than the austenitic stainless steel needle (figure 4B). However, it was difficult to determine the exact depth of insertion of the gold needle, especially in muscle, since the contrast between the needle and its surrounding is low. In addition, the gold needle is less rigid than the stainless steel needle and bends during insertion under the skin, particularly in areas of resistance such as the scalp. DISCUSSION Acupuncture needles evolved originally from bianstone, bamboo and bone to metals such as bronze, gold, silver and stainless steel. However, the same acupuncture manipulation with different kinds of needles may produce different effects. Among these materials, the stiffness and toughness of copper, gold and silver needles is less than that of stainless steel needles. As fmri is being increasingly used for acupuncture studies, MRI-compatible metal needles have attracted increased attention from investigators. Plastic needles can be used for MRI of acupuncture, 11 but they are not commonly used in the clinic because of their low elasticity and extremely low electrical conductivity. Standard stainless steel needles are widely used in the clinic; however, they generate serious artefacts and may experience a magnetic force inside the MRI scanner. Evidence from previous work 11 and our present study confirms that standard stainless steel acupuncture needles are unsafe and generate a significant artefact, so cannot be used in MRI scanners. Our experimental results show that the gold needle is superior because of the small imaging artefacts and the high level of safety in MRI. However, it is soft and tends to bend during acupuncture, which makes it difficult to control the insertion depth when being used for certain treatments. In addition, the higher cost makes it less practical for widespread use in the clinic. Acupunct Med: first published as /acupmed on 16 August Downloaded from Figure 2 Temperature changes of the acupuncture needles induced by an echo planar imaging (EPI) sequence. All three needles showed no obvious temperature increase during MRI using an EPI sequence. Figure 3 Temperature changes of acupuncture needles scanned using a fast low angle shot (FLASH) sequence. Although the FLASH sequence had a high specific absorption rate, the temperature changes of the three needles remained very small. on 22 July 2018 by guest. Protected by copyright. Mei L, et al. Acupunct Med 2013;0:1 5. doi: /acupmed

4 The austenitic stainless steel, which differs from the standard stainless steel in its composition and molecular structure,19 is a possible option as an MR-compatible material. The austenitic matrix changes the susceptibility of the normal stainless steel, resulting in a reduction in magnetic field distortion and imaging artefacts.20 In addition, it is stiffer than a gold needle, so it is easier to manufacture and to use for acupuncture studies. Previous studies have shown that long conductive wires can induce RF heating in the surrounding tissue during MRI Some investigations showed that the effect may lead to an increase in temperature of more than 60 C at the tip when the resonance conditions were met for an antenna.21 In our study, two commonly used sequences for fmri of acupuncture, the EPI and FLASH sequences, were used for temperature measurement. EPI is a fast data collection sequence with multiple echoes in an excitation, and the TR of 4 EPI should be set long enough to meet its acquisition scheme.13 The EPI sequence therefore has a lower SAR value, so the RF heating power could spread out during a long TR. To ensure the safety of the needles with regard to heating, we designed our second test at a critical condition using a FLASH pulse sequence with a short TR and high SAR. Although all the temperature measurements showed a tendency to increase, none of the three needles produced an increment larger than 0.5 C during 20 min MR scanning, indicating that none of these metal needles induced RF heating problems during MR scanning. However, it is necessary to conduct further tests to evaluate the safety of MRI when the needles are connected to a long cable for electrostimulation. In conclusion, we performed a series of in vitro and in vivo experiments to evaluate MR compatibility of three metal acupuncture needles made of standard Mei L, et al. Acupunct Med 2013;0:1 5. doi: /acupmed Figure 4 MRI of the needles inserted into SP6 (top row: coronal plane, bottom row: transversal plane). (A) The gold needle produced a small artefact; however, the contrast between the needle and its surrounding muscle was low. (B) The austenitic stainless steel needle generated some artefact but it was easy to identify the needle and measure the insertion depth.

5 stainless steel, gold and austenitic stainless steel, respectively. The experimental results showed that the traditional stainless steel needle is not MR-compatible and thus is not recommended for use in MRI. The gold needle and the austenitic stainless steel needle demonstrated higher MR compatibility and smaller imaging artefacts. None of the three needles showed RF heating risk during MRI of acupuncture. The clinical use of gold needles is not practical due to their high cost and softness. The austenitic needle is inexpensive, strong enough to penetrate thick skin such as the scalp and appropriate for electrical stimulation, so it is the most practical type of acupuncture needle for MRI of acupuncture. Summary points We compared the performance of standard, gold and austenitic stainless steel needles in MRI. The temperature increase was <0.5 C for all three needles. Standard needles gave an obscure image and so were unsuitable for MR; the other two types are MR-compatible and austenitic stainless has some advantages over gold. Contributors BQ and LM designed the experiment. LM performed the experiment, assisted by YD and HY who were licensed acupuncturists. The austenitic stainless steel needle was provided by WY. After the experiment, LM and XL processed the data and wrote the paper which was finally revised by XL, BQ and LJS. Funding This work was supported by National Basic Research Program of China (973 Program)-2010CB732604, Guangdong Innovation Research Team Fund for Low-Cost Healthcare Technologies and the Bairen Talent Project of CAS. Competing interests None. Patient consent Obtained. Ethics approval The experiment was approved by the Institutional Review Board of Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences. Provenance and peer review Not commissioned; internally peer reviewed. REFERENCES 1 Cho ZH, Chung SC, Jones JP, et al. New findings of the correlation between acupoints and corresponding brain cortices using functional MRI. Proc Natl Acad Sci USA 1998;95: Hui KK, Liu J, Makris N, et al. Acupuncture modulates the limbic system and subcortical gray structures of the human brain: evidence from fmri studies in normal subjects. Hum Brain Mapp 2000;9: Li G, Yang ES. An fmri study of acupuncture-induced brain activation of aphasia stroke patients. Complement Ther Med 2011;19(Suppl 1):S Na BJ, Jahng GH, Park SU, et al. An fmri study of neuronal specificity of an acupoint: electroacupuncture stimulation of Yanglingquan (GB34) and its sham point. Neurosci Lett 2009;464: Park SU, Shin AS, Jahng GH, et al. Effects of scalp acupuncture versus upper and lower limb acupuncture on signal activation of blood oxygen level dependent (BOLD) fmri of the brain and somatosensory cortex. J Altern Complement Med 2009;15: Wu MT, Sheen JM, Chuang KH, et al. Neuronal specificity of acupuncture response: a fmri study with electroacupuncture. Neuroimage 2002;16: Yan B, Li K, Xu J, et al. Acupoint-specific fmri patterns in human brain. Neurosci Lett 2005;383: Hui K, Liu J, Chen A, et al. Effects of acupuncture on human limbic system and basal ganglia measured by fmri. Neuroimage 1997;5:s Hui K, Liu J, Marina O, et al. The integrated response of the human cerebro-cerebellar and limbic systems to acupuncture stimulation at ST 36 as evidenced by fmri. Neuroimage 2005;27: Armenean C, Perrin E, Armenean M, et al. RF-induced temperature elevation along metallic wires in clinical magnetic resonance imaging: influence of diameter and length. Magn Reson Med 2004;52: Beissner F, Nöth U, Schockert T. The problem of metal needles in acupuncture-fmri studies. Evid-Based Complement Alternat Med 2011;2011: Frahm J, Haase A, Hanicke W, et al. Chemical shift selective MR imaging using a whole-body magnet. Radiology 1985;156: Matthaei D, Frahm J, Haase A, et al. Three-dimensional FLASH MR imaging of thorax and abdomen without triggering or gating. Magn Reson Imaging 1986;4: Mansfield P. Multi-planner image formation using NMR spin echoes. J Phys 1977;10: Mansfield P, Pykett I. Biological and medical imaging by NMR. J Magn Reson 1978;29: Rzedzian R, Pykett I. Instant images of the human heart using a new whole body MR imaging system. Am J Roentgenol 1987;149: Nghiem H, Herfkens R, Francis I. The pelvis: T2-weighted fast spin-echo MR imaging. Radiology 1992;185: Smith R, Reinhold C, Lange R, et al. Fast spin echo MR imaging of the female pelvis. I. Use of a whole volume coil. Radiology 1992;184: Xu J, Ju C, Ji G. Research progress of phase transformation of stainless steel. Hot Work Technol 2008;37: McGuire M. Austenitic stainless steels. In: Kaufman JG, ed. Stainless steels for design engineers. Materials Park, OH: ASM International, 2008: Dempsey M, Condon B, Hadley D. Investigation of the factors responsible for burns during MRI. J Magn Reson Imaging 2001;13: Mei L, et al. Acupunct Med 2013;0:1 5. doi: /acupmed

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE In Practice RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE By Atsuya Watanabe, MD, PhD, Director, Advanced Diagnostic Imaging Center and Associate Professor, Department of Orthopedic Surgery, Teikyo

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants Knee Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Knee Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All knee implants of Smith

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants Shoulder Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Shoulder Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All shoulder implants

More information

Assessment of Adipose Tissue from Whole Body 3T MRI Scans

Assessment of Adipose Tissue from Whole Body 3T MRI Scans Assessment of Adipose Tissue from Whole Body 3T MRI Scans Ting Song 1, Jing An 2, Qun Chen 2, Vivian Lee 2, Andrew Laine 1 1 Department of Biomedical Engineering, Columbia University, New York, NY, USA

More information

Combining tdcs and fmri. OHMB Teaching Course, Hamburg June 8, Andrea Antal

Combining tdcs and fmri. OHMB Teaching Course, Hamburg June 8, Andrea Antal Andrea Antal Department of Clinical Neurophysiology Georg-August University Goettingen Combining tdcs and fmri OHMB Teaching Course, Hamburg June 8, 2014 Classical Biomarkers for measuring human neuroplasticity

More information

Visual cortical activations on fmri upon stimulation of the vision-implicated acupoints

Visual cortical activations on fmri upon stimulation of the vision-implicated acupoints BRAIN IMAGING Visual cortical activations on fmri upon stimulation of the vision-implicated acupoints Geng Li, 1,2 Raymond T. F. Cheung, 2,CA Qi-Yuan Ma 1 and Edward S. Yang 1 1 The Jockey Club MRI Engineering

More information

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use International Congress Series 1281 (2005) 793 797 www.ics-elsevier.com Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use Ch. Nimsky a,b,

More information

Effects of Acupuncture on Chinese Adult Patients with Psoriatic Arthritis: A Prospective Cohort Study

Effects of Acupuncture on Chinese Adult Patients with Psoriatic Arthritis: A Prospective Cohort Study Effects of Acupuncture on Chinese Adult Patients with Psoriatic Arthritis: A Prospective Cohort Study Peiyi Chen 1, Tiantian Xin 2* and Yingchun Zeng 3,4 1 School of Nursing, Guangzhou University of Chinese

More information

Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius.

Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius. Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius. Technique Guide Part of the External Fixation System Low-Profile Wrist Fixator Indications Intended for stabilization of

More information

An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles

An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles Innovative Graduate Student Proposal Yujie Qiu, Principal Investigator, Graduate Student Joseph Hornak, Faculty Sponsor, Thesis Advisor

More information

Q-Sense fmri in the MRI environment

Q-Sense fmri in the MRI environment Q-Sense fmri in the MRI environment An evaluation at the Erwin L. Hahn Institute for Magnetic Resonance Imaging, and at the Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen

More information

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015 MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute

More information

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015

UK Biobank. Imaging modality Cardiovascular Magnetic Resonance (CMR) Version th Oct 2015 Imaging modality Cardiovascular Magnetic Resonance (CMR) Version 1.0 http://www.ukbiobank.ac.uk/ 30 th Oct 2015 This document details the procedure for the CMR scan performed at an Imaging assessment centre

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants Hip Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Hip Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All hip implants of Smith &

More information

BioMatrix Tuners: CoilShim

BioMatrix Tuners: CoilShim MAGNETOM Vida special issue Head and Neck Imaging Clinical 11 BioMatrix Tuners: CoilShim Miriam R. Keil, Ph.D.; Jörg Rothard; Carmel Hayes, Ph.D. Siemens Healthineers, Erlangen, Germany A cervical spine

More information

Supplementary Information Methods Subjects The study was comprised of 84 chronic pain patients with either chronic back pain (CBP) or osteoarthritis

Supplementary Information Methods Subjects The study was comprised of 84 chronic pain patients with either chronic back pain (CBP) or osteoarthritis Supplementary Information Methods Subjects The study was comprised of 84 chronic pain patients with either chronic back pain (CBP) or osteoarthritis (OA). All subjects provided informed consent to procedures

More information

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview

CARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview CARDIAC MRI Dr Yang Faridah A. Aziz Department of Biomedical Imaging University of Malaya Medical Centre Cardiovascular Disease Diseases of the circulatory system, also called cardiovascular disease (CVD),

More information

An fmri Study Showing the Effect of Acupuncture in Chronic Stage Stroke Patients With Aphasia

An fmri Study Showing the Effect of Acupuncture in Chronic Stage Stroke Patients With Aphasia J Acupunct Meridian Stud 2010;3(1):53 57 BRIEF REPORT An fmri Study Showing the Effect of Acupuncture in Chronic Stage Stroke Patients With Aphasia Anson C.M. Chau 1, Raymond Tak Fai Cheung 2 *, Xianyong

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging Provläsningsexemplar / Preview TECHNICAL REPORT ISO/TR 16379 First edition 2014-03-01 Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR

More information

The Effects of Music intervention on Functional connectivity. Supplemental Information

The Effects of Music intervention on Functional connectivity. Supplemental Information Yang et al. 0 The Effects of Music intervention on Functional connectivity strength of Brain in Schizophrenia Supplemental Information Mi Yang,#, Hui He #, Mingjun Duan,, Xi Chen, Xin Chang, Yongxiu Lai,

More information

Methods of Visualizing the Living Human Brain

Methods of Visualizing the Living Human Brain Methods of Visualizing the Living Human Brain! Contrast X-rays! Computerized Tomography (CT)! Magnetic Resonance Imaging (MRI)! Positron Emission Tomography (PET)! Functional MRI! Magnetoencephalography

More information

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system External Distal Radius Fixator. Supplement to the 8 mm rod fixator system Surgical technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation

More information

Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases

Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases Qiang Ning, Chao Ma, Fan Lam, Bryan Clifford, Zhi-Pei Liang November 11, 2015 1 Synopsis A novel nuisance

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

How Much Tesla Is Too Much?

How Much Tesla Is Too Much? How Much Tesla Is Too Much? Johnny U. V. Monu, MB, BS; MSc Professor of Radiology and Orthopedics University of Rochester School of Medicine Rochester, New York Historical Timeline Clinical Imaging 1970

More information

Supplementary information Detailed Materials and Methods

Supplementary information Detailed Materials and Methods Supplementary information Detailed Materials and Methods Subjects The experiment included twelve subjects: ten sighted subjects and two blind. Five of the ten sighted subjects were expert users of a visual-to-auditory

More information

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol Sample referral sticker Referral Sticker Insert here Corin 17 Bridge Street Pymble NSW Australia 2073 P: +61 (0)2 9497 7400 F: +61 (0)2 9497 7498 E: KAS.customerservice@coringroup.com

More information

University of Groningen. Neuro-imaging of visual field defects Boucard, Christine

University of Groningen. Neuro-imaging of visual field defects Boucard, Christine University of Groningen Neuro-imaging of visual field defects Boucard, Christine IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please

More information

fmri and Voxel-based Morphometry in Detection of Early Stages of Alzheimer's Disease

fmri and Voxel-based Morphometry in Detection of Early Stages of Alzheimer's Disease fmri and Voxel-based Morphometry in Detection of Early Stages of Alzheimer's Disease Andrey V. Sokolov 1,3, Sergey V. Vorobyev 2, Aleksandr Yu. Efimtcev 1,3, Viacheslav S. Dekan 1,3, Gennadiy E. Trufanov

More information

High Field MR of the Spine

High Field MR of the Spine Department of Radiology University of California San Diego 3T for MR Applications Advantages High Field MR of the Spine Increased signal-to-noise Better fat suppression Increased enhancement with gadolinium

More information

Table 1. Summary of PET and fmri Methods. What is imaged PET fmri BOLD (T2*) Regional brain activation. Blood flow ( 15 O) Arterial spin tagging (AST)

Table 1. Summary of PET and fmri Methods. What is imaged PET fmri BOLD (T2*) Regional brain activation. Blood flow ( 15 O) Arterial spin tagging (AST) Table 1 Summary of PET and fmri Methods What is imaged PET fmri Brain structure Regional brain activation Anatomical connectivity Receptor binding and regional chemical distribution Blood flow ( 15 O)

More information

3D high-resolution MR imaging can provide reliable information

3D high-resolution MR imaging can provide reliable information Published April 11, 2013 as 10.3174/ajnr.A3472 ORIGINAL RESEARCH HEAD & NECK High-Resolution MRI of the Intraparotid Facial Nerve Based on a Microsurface Coil and a 3D Reversed Fast Imaging with Steady-State

More information

Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T

Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T International Congress Series 1273 (2004) 394 398 www.ics-elsevier.com Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T F. Risi*, A. Saldanha, R. Leigh, P. Gibson Cochlear Limited,

More information

P2 Visual - Perception

P2 Visual - Perception P2 Visual - Perception 2014 SOSE Neuroimaging of high-level visual functions gyula.kovacs@uni-jena.de 11/09/06 Functional magnetic resonance imaging (fmri) The very basics What is fmri? What is MRI? The

More information

Fat Suppression in the Abdomen

Fat Suppression in the Abdomen Clinical How I do it? Fat Suppression in the Abdomen Wilhelm Horger Siemens Medical Solutions, Erlangen, Germany Introduction Due to the different chemical environment, hydrogen nuclei in - and in -tissue

More information

Acupuncture and Drugs Found Equally Effective For Mild Depression

Acupuncture and Drugs Found Equally Effective For Mild Depression Acupuncture and Drugs Found Equally Effective For Mild Depression Published by HealthCMi on 25 May 2018 Guangdong Province Traditional Chinese Medicine Hospital researchers find acupuncture with moxibustion

More information

2014 AAPM 56 th Annual Meeting

2014 AAPM 56 th Annual Meeting 2014 AAPM 56 th Annual Meeting SAM Diagnostic Radiology MR Safety - Deep Brain Stimulator and Other Neurostimulators Yunhong Shu, Ph.D. Mayo Clinic, Rochester, MN Outline Background of Neurostimulator

More information

Orthopedic Hardware Imaging Part II: MRI v. Metal

Orthopedic Hardware Imaging Part II: MRI v. Metal Orthopedic Hardware Imaging Trent Roth, MD And Lauren Ladd, MD Indiana University School of Medicine IU Health Physicians-Radiology Recap: Imaging Techniques Radiography Standard for initial and surveillance

More information

Procedia - Social and Behavioral Sciences 159 ( 2014 ) WCPCG 2014

Procedia - Social and Behavioral Sciences 159 ( 2014 ) WCPCG 2014 Available online at www.sciencedirect.com ScienceDirect Procedia - Social and Behavioral Sciences 159 ( 2014 ) 743 748 WCPCG 2014 Differences in Visuospatial Cognition Performance and Regional Brain Activation

More information

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils Clinical Orthopedic Imaging Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils Elena Ferrer 1 ; Rafael Coronado Santos 2 1 Radiology Department, Clínica Creu Blanca, Barcelona,

More information

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry Journal of Physics: Conference Series Optimization of the T parametric image map calculation in MRI polymer gel dosimetry To cite this article: Fotini Zacharopoulou et al J. Phys.: Conf. Ser. 5 9 Related

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/35124 holds various files of this Leiden University dissertation. Author: Wokke, Beatrijs Henriette Aleid Title: Muscle MRI in Duchenne and Becker muscular

More information

Post-Stroke Depression Relief With Acupuncture

Post-Stroke Depression Relief With Acupuncture Post-Stroke Depression Relief With Acupuncture Published by HealthCMi on August 2017 Acupuncture is more effective than the antidepressant paroxetine for the treatment of depression following a stroke.

More information

How I do it: Non Contrast-Enhanced MR Angiography (syngo NATIVE)

How I do it: Non Contrast-Enhanced MR Angiography (syngo NATIVE) Clinical How-I-do-it Cardiovascular How I do it: Non Contrast-Enhanced MR Angiography (syngo NATIVE) Manuela Rick, Nina Kaarmann, Peter Weale, Peter Schmitt Siemens Healthcare, Erlangen, Germany Introduction

More information

MR Advance Techniques. Vascular Imaging. Class II

MR Advance Techniques. Vascular Imaging. Class II MR Advance Techniques Vascular Imaging Class II 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

3T functional magnetic resonance in healthy subjects: effects evaluation of MS6 scalp acupuncture

3T functional magnetic resonance in healthy subjects: effects evaluation of MS6 scalp acupuncture 3T functional magnetic resonance in healthy subjects: effects evaluation of MS6 scalp acupuncture Poster No.: C-0897 Congress: ECR 2013 Type: Scientific Exhibit Authors: R. Zanardi, M. Maieron, C. M. Giovanardi,

More information

PHYSICS OF MRI ACQUISITION. Alternatives to BOLD for fmri

PHYSICS OF MRI ACQUISITION. Alternatives to BOLD for fmri PHYSICS OF MRI ACQUISITION Quick Review for fmri HST-583, Fall 2002 HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology

More information

Daniel Bulte. Centre for Functional Magnetic Resonance Imaging of the Brain. University of Oxford

Daniel Bulte. Centre for Functional Magnetic Resonance Imaging of the Brain. University of Oxford Daniel Bulte Centre for Functional Magnetic Resonance Imaging of the Brain University of Oxford Overview Signal Sources BOLD Contrast Mechanism of MR signal change FMRI Modelling Scan design details Factors

More information

Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.

Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use. Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices

More information

Functional Magnetic Resonance Imaging with Arterial Spin Labeling: Techniques and Potential Clinical and Research Applications

Functional Magnetic Resonance Imaging with Arterial Spin Labeling: Techniques and Potential Clinical and Research Applications pissn 2384-1095 eissn 2384-1109 imri 2017;21:91-96 https://doi.org/10.13104/imri.2017.21.2.91 Functional Magnetic Resonance Imaging with Arterial Spin Labeling: Techniques and Potential Clinical and Research

More information

In-vivo MRI tcs compatibility

In-vivo MRI tcs compatibility In-vivo MRI tcs compatibility Neuroelectrics White Paper WP201505 Author(s): B. Otal (PhD), G. Mitjà, G. Ruffini (PhD), M. Mendes Released on: 24 th November 2015 www.neuroelectrics.com 1 In-vivo MRI tcs

More information

The physiology of the BOLD signal What do we measure with fmri?

The physiology of the BOLD signal What do we measure with fmri? The physiology of the BOLD signal What do we measure with fmri? Methods and Models in fmri, 10.11.2012 Jakob Heinzle Translational Neuromodeling Unit (TNU) Institute for Biomedical Engineering (IBT) University

More information

Hoffmann II MRI External Fixation Systems. Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis

Hoffmann II MRI External Fixation Systems. Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis Hoffmann II MRI External Fixation Systems Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis Introduction In the ever changing world of medicine, new technology

More information

ACR MRI Accreditation: Medical Physicist Role in the Application Process

ACR MRI Accreditation: Medical Physicist Role in the Application Process ACR MRI Accreditation: Medical Physicist Role in the Application Process Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives

More information

Study of the CNS. Bent O. Kjos' Richard L. Ehman Michael Brant-Zawadzki William M. Kelly David Norman Thomas H. Newton

Study of the CNS. Bent O. Kjos' Richard L. Ehman Michael Brant-Zawadzki William M. Kelly David Norman Thomas H. Newton 271 Reproducibility of Relaxation Times and Spin Density Calculated from Routine MR Imaging Sequences: Clinical Study of the CNS Bent O. Kjos' Richard L. Ehman Michael Brant-Zawadzki William M. Kelly David

More information

Cardiac MRI at 7T Syllabus contribution: Matthew Robson

Cardiac MRI at 7T Syllabus contribution: Matthew Robson Cardiac MRI at 7T Syllabus contribution: Matthew Robson Field strength escalation has occurred over the last 10 years. Whilst 1.5T remains the most prevalent field strength the loss of 1T and the popularity

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Gregg NM, Kim AE, Gurol ME, et al. Incidental cerebral microbleeds and cerebral blood flow in elderly individuals. JAMA Neurol. Published online July 13, 2015. doi:10.1001/jamaneurol.2015.1359.

More information

Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping

Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping TR_002 Technical Reports Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping Min A Yoon 1,*, Suk-Joo Hong 1, Chang Ho Kang 2, Baek Hyun Kim 3 1 Korea University

More information

A quality control program for MR-guided focused ultrasound ablation therapy

A quality control program for MR-guided focused ultrasound ablation therapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A quality control program for MR-guided focused ultrasound ablation therapy Tao Wu* and Joel P. Felmlee Department of Radiology,

More information

WHAT DOES THE BRAIN TELL US ABOUT TRUST AND DISTRUST? EVIDENCE FROM A FUNCTIONAL NEUROIMAGING STUDY 1

WHAT DOES THE BRAIN TELL US ABOUT TRUST AND DISTRUST? EVIDENCE FROM A FUNCTIONAL NEUROIMAGING STUDY 1 SPECIAL ISSUE WHAT DOES THE BRAIN TE US ABOUT AND DIS? EVIDENCE FROM A FUNCTIONAL NEUROIMAGING STUDY 1 By: Angelika Dimoka Fox School of Business Temple University 1801 Liacouras Walk Philadelphia, PA

More information

Resting-State Functional Connectivity in Stroke Patients After Upper Limb Robot-Assisted Therapy: A Pilot Study

Resting-State Functional Connectivity in Stroke Patients After Upper Limb Robot-Assisted Therapy: A Pilot Study Resting-State Functional Connectivity in Stroke Patients After Upper Limb Robot-Assisted Therapy: A Pilot Study N. Kinany 1,3,4(&), C. Pierella 1, E. Pirondini 3,4, M. Coscia 2, J. Miehlbradt 1, C. Magnin

More information

Magnetic Resonance Imaging on Soft Tissue. Jiten K. Mistry Calvin Gan

Magnetic Resonance Imaging on Soft Tissue. Jiten K. Mistry Calvin Gan Magnetic Resonance Imaging on Soft Tissue 1 Jiten K. Mistry Calvin Gan Outline Background of Medical Imaging Introduction to MRI How MRI works MRI of Soft Tissue Benefits & Risks Recent Advances 2 The

More information

Quantitative Analysis of Vascular Canals in Vertebral Endplate

Quantitative Analysis of Vascular Canals in Vertebral Endplate Quantitative Analysis of Vascular Canals in Vertebral Endplate Kristine Tan 1, Won C. Bae, PhD 1, Tomonori Yamaguchi, MS 1,2, Kelli Xu, BS 1, Iris Shieh, BS 1, Jade He, BS 1, Robert L. Sah, MD, ScD 1,

More information

ACR Accreditation Update in MRI

ACR Accreditation Update in MRI ACR Accreditation Update in MRI Whole Body Systems Extremity (MSK) Ron Price Vanderbilt University Medical Center Nashville, TN Dedicated Breast MRI Accreditation Update 1. ACR MRI Accreditation: Overview,

More information

Warning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed.

Warning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed. Prometra and Prometra II Programmable Pumps Magnetic Resonance Imaging (MRI) Safety Information GENERAL MR Conditional WARNING: FAILURE TO EMPTY THE PUMP PRIOR TO EXPOSURE TO MRI ENVIRONMENT COULD RESULT

More information

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 MIT OpenCourseWare http://ocw.mit.edu HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms.

More information

User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017

User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017 User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017 Ives EEG Solutions; 25 Storey Ave., #118, Newburyport, MA, 01950; ph: 978-358-8006; fax: 978-358-7825; www.iveseegsolutions.com

More information

Comparison of 1.5 T Tesla and 3.0 T Tesla Magnetic Resonance Imaging for Evaluating Local Extension of Endometrial Cancer

Comparison of 1.5 T Tesla and 3.0 T Tesla Magnetic Resonance Imaging for Evaluating Local Extension of Endometrial Cancer Showa Univ J Med Sci 27 1, 21 28, March 2015 Original Comparison of 1.5 T Tesla and 3.0 T Tesla Magnetic Resonance Imaging for Evaluating Local Extension of Endometrial Cancer Naomi YAGI, Masanori HIROSE,

More information

Journal of Foot and Ankle Research

Journal of Foot and Ankle Research Journal of Foot and Ankle Research BioMed Central Research Mechanical characteristics of three staples commonly used in foot surgery Ulfin Rethnam* 1, Jan Kuiper 2 and Nilesh Makwana 3 Open Access Address:

More information

T 2 * -weighted imaging: distinction between the representation of the forepaw and hindpaw in the somatosensory cortex

T 2 * -weighted imaging: distinction between the representation of the forepaw and hindpaw in the somatosensory cortex NMR IN BIOMEDICINE NMR Biomed 11, 115 119 (1998) Brainmapping of a-chloralose anesthetized rats with T 2 * -weighted imaging: distinction between the representation of the forepaw and hindpaw in the somatosensory

More information

ASL BASICS II. Learning Objectives. Outline. Acquisition. M. A. Fernández-Seara, Ph. D. Arterial spin labeled perfusion MRI: basic theory

ASL BASICS II. Learning Objectives. Outline. Acquisition. M. A. Fernández-Seara, Ph. D. Arterial spin labeled perfusion MRI: basic theory Acquisition ASL BASICS II M. A. Fernández-Seara, Ph. D. Neuroimaging Laboratory Center for Applied Medical Research University of Navarra Pamplona, Spain Outline Arterial spin labeled perfusion MRI: basic

More information

Case Reports: Tumor Detection by Diffusion-Weighted MRI and ADC-Mapping with Correlation to PET/CT Results

Case Reports: Tumor Detection by Diffusion-Weighted MRI and ADC-Mapping with Correlation to PET/CT Results Case Reports: Tumor Detection by Diffusion-Weighted MRI and ADC-Mapping with Correlation to PET/CT Results Matthias Philipp Lichy, M.D.; Philip Aschoff, M.D.; Christina Pfannenberg, M.D.; Schlemmer Heinz-Peter,

More information

Meridians in acupuncture and infrared imaging

Meridians in acupuncture and infrared imaging Medical Hypotheses (2002) 58(1), 72±76 1 doi: 10.1054/mehy.2001.1453, available online at http://www.idealibrary.com on Meridians in acupuncture and infrared imaging Shui-yin Lo Infrared Health Company,

More information

Twelve right-handed subjects between the ages of 22 and 30 were recruited from the

Twelve right-handed subjects between the ages of 22 and 30 were recruited from the Supplementary Methods Materials & Methods Subjects Twelve right-handed subjects between the ages of 22 and 30 were recruited from the Dartmouth community. All subjects were native speakers of English,

More information

MRI Guidelines for the Axonics Sacral Neuromodulation System

MRI Guidelines for the Axonics Sacral Neuromodulation System MRI Guidelines for the Axonics Sacral Neuromodulation System For Healthcare Professionals Rx only Note: This document is a supplement to the physician manuals provided with the Axonics SNM System. This

More information

Investigations in Resting State Connectivity. Overview

Investigations in Resting State Connectivity. Overview Investigations in Resting State Connectivity Scott FMRI Laboratory Overview Introduction Functional connectivity explorations Dynamic change (motor fatigue) Neurological change (Asperger s Disorder, depression)

More information

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration Richard Kijowski, Michael Fazio, Benjamin Beduhn, and Fang Liu Department of Radiology University of Wisconsin School of Medicine

More information

Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps.

Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps. Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices

More information

Acupuncture And Herbs Proven Effective For PID Treatment

Acupuncture And Herbs Proven Effective For PID Treatment Acupuncture And Herbs Proven Effective For PID Treatment Published by HealthCMi on 29 May 2018 Researchers find acupuncture combined with herbal medicine effective for the treatment of chronic pelvic inflammatory

More information

EXTERNAL FIXATION SYSTEM

EXTERNAL FIXATION SYSTEM EXTERNAL FIXATION SYSTEM The Apex Pin Fixation System is a one step procedure reducing insertion time and reducing insertion temperature. There are three types of fixation pins: Self-drilling / Self-tapping

More information

Transcranial Focused Ultrasound for BOLD fmri Signal Modulation in Humans

Transcranial Focused Ultrasound for BOLD fmri Signal Modulation in Humans Transcranial Focused Ultrasound for BOLD fmri Signal Modulation in Humans Leo Ai, Jerel K. Mueller, Andrea Grant, Yigitcan Eryaman, and Wynn Legon Abstract Transcranial focused ultrasound (tfus) is an

More information

A characteristic feature of acute haematomas in the brain on echo-planar diffusion-weighted imaging

A characteristic feature of acute haematomas in the brain on echo-planar diffusion-weighted imaging Neuroradiology (2002) 44: 907 911 DOI 10.1007/s00234-002-0860-5 DIAGNOSTIC NEURORADIOLOGY N. Morita M. Harada K. Yoneda H. Nishitani M. Uno A characteristic feature of acute haematomas in the brain on

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500). ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro

More information

Brain gray matter volume changes associated with motor symptoms in patients with Parkinson s disease

Brain gray matter volume changes associated with motor symptoms in patients with Parkinson s disease Kang et al. Chinese Neurosurgical Journal (2015) 1:9 DOI 10.1186/s41016-015-0003-6 RESEARCH Open Access Brain gray matter volume changes associated with motor symptoms in patients with Parkinson s disease

More information

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging

Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging Med Biol Eng Comput (2008) 46:729 733 DOI 10.1007/s11517-008-0307-1 TECHNICAL NOTE Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging

More information

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution COMMUNICATIONS Magnetic Resonance in Medicine 44:511 515 (2000) Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution Eric C. Wong,* Wen-Ming Luh, and Thomas T. Liu A modified pulsed

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Green SA, Hernandez L, Tottenham N, Krasileva K, Bookheimer SY, Dapretto M. The neurobiology of sensory overresponsivity in youth with autism spectrum disorders. Published

More information

Chronic Fatigue Recovery With Acupuncture

Chronic Fatigue Recovery With Acupuncture Chronic Fatigue Recovery With Acupuncture Published by HealthCMi on August 2017 Shenzhen Hospital of Chinese Medicine researchers determined that acupuncture is safe and effective for the alleviation of

More information

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new

More information

Motohiro Inoue, 1 Miwa Nakajima, 1 Tatsuya Hojo, 2 Hiroshi Kitakoji, 1 Megumi Itoi 3. Original paper

Motohiro Inoue, 1 Miwa Nakajima, 1 Tatsuya Hojo, 2 Hiroshi Kitakoji, 1 Megumi Itoi 3. Original paper 1 Department of Clinical Acupuncture and Moxibustion, Meiji University of Integrative Medicine, Kyoto, Japan 2 Department of Health and Sports Science, Doshisha University, Kyoto, Japan 3 Department of

More information

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 201x Algovita

More information

Titelmaster The physics of functional magnetic resonance imaging (fmri)

Titelmaster The physics of functional magnetic resonance imaging (fmri) Titelmaster The physics of functional magnetic resonance imaging (fmri) Outline 1.Introduction 2.The fmri experiment 2 3.The physics basis of fmri 4.Application Outline 3 1.Introduction Introduction Phrenology

More information

High-resolution diffusion-weighted MRI of the breast using readout-segmented EPI and single-shot EPI

High-resolution diffusion-weighted MRI of the breast using readout-segmented EPI and single-shot EPI High-resolution diffusion-weighted MRI of the breast using readout-segmented EPI and single-shot EPI Objective: Compared to dynamic contrast enhanced MRI (DCE-MRI), image quality in diffusion-weighted

More information

KneeAlign System Surgical Technique Guide

KneeAlign System Surgical Technique Guide KneeAlign System Surgical Technique Guide Table of Contents Step 1 System Assembly... 1 Step 2 System Assembly... 2 Step 3 System Assembly... 2 Step 4 System Assembly... 2 Step 5 Sensor Pairing... 2 Step

More information

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS Jeffrey P. Milligan, Daniel T. Peters, Structural Integrity Associates, Inc., USA Many advances in Non-Destructive Examination

More information

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH MRI for Patients with MRI-Conditional Pacing System: Radiographers Role Lawrance Yip DM, DR, QMH MRI & Pacemaker MRI is a crucial and growing imaging modality Plays an important role in clinical management

More information

Neural Networks: Tracing Cellular Pathways. Lauren Berryman Sunfest 2000

Neural Networks: Tracing Cellular Pathways. Lauren Berryman Sunfest 2000 Neural Networks: Tracing Cellular Pathways Lauren Berryman Sunfest 000 Neural Networks: Tracing Cellular Pathways Research Objective Background Methodology and Experimental Approach Results and Conclusions

More information