Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants

Size: px
Start display at page:

Download "Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants"

Transcription

1 E13195: Slightly better series/treatment with PRONE combination with vertical field. - This field is only possible in case the tumor is not too caudal (problem with energy). - It will increase the integral dose as it can be seen from the DVHs: To promote excellency in patient care and innovative proton treatment Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants Dietlicher I, Casiraghi M, Ares C, Bolsi A, Weber DC, Lomax AJ, Albertini F PSI,

2 Rationale 40 patients with Spinal Axis-Associated Chordomas/Chondrosarcomas ( ) 21 patients (50%) treated with titanium implants [12/21 have a local failure in 5y] No implant With titanium implant The presence of metal has been shown to be a risk factor for patient outcome Investigation of the artefacts effects on dose distribution Staab et al IJROBP 2011

3 Rationale: Dealing with metal implants - in practice Titanium stopping power is well known (<1%) BUT artefacts introduce range uncertainties Set high HU values (>=3095) to stopping power of titanium in CT calibration curve Correct artifacts by delineating artifact areas and setting HU values to average values for fat and soft tissues Nominal kv-ct Uncorrected Corrected kv-ct Is this policy effective?

4 Material and Methods: Use of an anthropomorphic phantom Phantom with titanium implants Correct for artifacts by delineating artifact areas and setting HU values to average values for fat and soft tissues. Nominal kv-ct CT not corrected corrected for artefacts for artefacts Calculate (and deliver) clinically relevant IMPT and SFUD proton plans

5 Material and Methods: GafChromic EBT2 1) Measure the 2D dose distribution with radiochromic films positioned along sagital planes, close to the implant (note: slice 2 is <1mm to the metal) 2) Digitization: - 24h after irradiation - transmission - average of 3 scans - red channel extraction 3) Linearization of the Response calibration curve 138 MeV SOBP 5 cm Dose (Gy) LG

6 Results: without artifacts correction WITHOUT artifacts correction, SFUD composite plan % of points satisfying the gamma criteria of 3%/3mm 94% 89% 97% Dietlicher et al PMB (submitted)

7 Results: with artifacts correction WITH artifacts correction, SFUD composite plan % of points satisfying the gamma criteria of 3%/3mm 100% 97% 99% Dietlicher et al PMB (submitted)

8 Results: Effect of the artifacts on the phantom Gamma analysis (3%, 3mm) reported on the phantom Biggest area of failure criteria close to the metal Slice 1 Slice 2 Slice 3 Area of failure γ- criteria 1cm 2 5cm 2 2.6cm 2 Dietlicher et al PMB (submitted)

9 Summary Our results indicate that when artifacts are corrected in the planning CT, the accuracy of the dose calculation is quite good. With some limitations (e.g. use multiple fields, avoid passage through the metal) patients with implants can therefore be treated with protons. The poor clinical outcome of these patients has to (probably) be related to some other factors (tumour more aggressive/more advanced stage )

10 Thank you

Specifics of treatment planning for active scanning and IMPT

Specifics of treatment planning for active scanning and IMPT Specifics of treatment planning for active scanning and IMPT SFUD IMPT Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment planning for scanning 1. Single Field,

More information

State-of-the-art proton therapy: The physicist s perspective

State-of-the-art proton therapy: The physicist s perspective State-of-the-art proton therapy: Tony Lomax, Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Overview of presentation 1. State-of-the-art proton delivery 2. Current challenges 3. New

More information

Center for Proton Therapy :: Paul Scherrer Institut :: #2_3/2014

Center for Proton Therapy :: Paul Scherrer Institut :: #2_3/2014 poton+ Center for Proton Therapy :: Paul Scherrer Institut :: #2_3/2014 Dear Colleagues Welcome again to this second edition of this newsletter. It is a pleasure to inform you that our first patient was

More information

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical

More information

Active Scanning Beam 3 Checking Delivery/Dosimetry

Active Scanning Beam 3 Checking Delivery/Dosimetry Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute

More information

Future upcoming technologies and what audit needs to address

Future upcoming technologies and what audit needs to address Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple

More information

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS Heng Li, PhD Assistant Professor, Department of Radiation Physics, UT MD Anderson Cancer Center, Houston, TX, 773

More information

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad: General Aspects of RCF (TG-235), AAPM July 2015 1 Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad, PhD, FAAPM,

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients IDEAL DOSIMETRY FOR IMRT/SRS Absolute dose measurement Precise

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Application of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint

Application of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint Application of Implanted Markers in Proton Therapy Sung Yong Park, Ph.D. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint AAPM 2016, SAM Therapy Educational Course, 2016.08.04. Course

More information

IMPT with Carbon Ions

IMPT with Carbon Ions IMPT with Carbon Ions PTCOG 48, Heidelberg, 28.09.-03.10.2009 Malte Ellerbrock Medical Physics Expert Heidelberg Ion-Beam Therapy Center HIT Betriebs GmbH am Universitätsklinikum Heidelberg http://www.hit-centrum.de

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Image Guided Proton Therapy and Treatment Adaptation

Image Guided Proton Therapy and Treatment Adaptation Image Guided Proton Therapy and Treatment Adaptation www.hollandptc.nl d.r.schaart@tudelft.nl Cancer in The Netherlands About 1 in 3 people get cancer in some stage of their life 86.800 new cancer patients

More information

Dose perturbation effect of metallic spinal implants in proton beam therapy

Dose perturbation effect of metallic spinal implants in proton beam therapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Dose perturbation effect of metallic spinal implants in proton beam therapy Yingcui Jia, 1a Li Zhao, 2 Chee-Wai Cheng, 3 Mark W. McDonald,

More information

Estimating Iodine Concentration from CT Number Enhancement

Estimating Iodine Concentration from CT Number Enhancement Estimating Iodine Concentration from CT Number Enhancement Rosemary Eaton, Andrew Shah, Jane Shekhdar Medical Physics, Mount Vernon Hospital CT Users Group 4 th October 212, Edinburgh Summary Background

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Discuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons

Discuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons Mark Pankuch, PhD Discuss the general planning concepts used in proton planning Review the unique handling of CTV / ITV / PTV when treating with protons Pencil Beam distributions and PBS optimization Cover

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Susannah Lazar David Followill, Ph.D. John Gibbons, Ph.D. * Anita Mahajan, M.D. Mohammad Salehpour, Ph.D. Marilyn Stovall, Ph.D. R. Allen White,

More information

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical sites. The final aim will be to be able to make out these

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

Proton and helium beams: the present and the future of light ion beam therapy

Proton and helium beams: the present and the future of light ion beam therapy Proton and helium beams: the present and the future of light ion beam therapy Dr. Andrea Mairani Group Leader Biophysics in Particle Therapy Heidelberg Ion Beam Therapy Center HIT Department of Radiation

More information

Implementation of an Anthropomorphic Phantom for the Evaluation of Proton Therapy Treatment Procedures

Implementation of an Anthropomorphic Phantom for the Evaluation of Proton Therapy Treatment Procedures Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 5-2010 Implementation of an Anthropomorphic Phantom for the Evaluation

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

Potential benefits of intensity-modulated proton therapy in head and neck cancer van de Water, Tara Arpana

Potential benefits of intensity-modulated proton therapy in head and neck cancer van de Water, Tara Arpana University of Groningen Potential benefits of intensity-modulated proton therapy in head and neck cancer van de Water, Tara Arpana IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's

More information

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms 1 st Workshop: Radiotherapy Modelling - Luz Saint Sauveur, Sep 2016 A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms A R Barbeiro 1, A Ureba 2, J A Baeza 3,

More information

Development and Implementation of an Anthropomorphic Head & Neck Phantom for the Assessment of Proton Therapy Treatment Procedures

Development and Implementation of an Anthropomorphic Head & Neck Phantom for the Assessment of Proton Therapy Treatment Procedures Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2016 Development and Implementation of an Anthropomorphic Head &

More information

Y FILMS DOSIMETR Nederland België / Belgique

Y FILMS DOSIMETR Nederland België / Belgique DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous

More information

Follow this and additional works at: Part of the Medicine and Health Sciences Commons

Follow this and additional works at:   Part of the Medicine and Health Sciences Commons Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2014 Development and Implementation of an Anthropomorphic Pediatric

More information

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy Guidelines for Planning and Irradiating the RPC Liver Phantom. Revised Dec 2005 Credentialing for this protocol requires four steps:

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

3D titanium interbody fusion cages sharx. White Paper

3D titanium interbody fusion cages sharx. White Paper 3D titanium interbody fusion cages sharx (SLM selective laser melted) Goal of the study: Does the sharx intervertebral cage due to innovative material, new design, and lordotic shape solve some problems

More information

Advanced film dosimetry for a UK brachytherapy audit

Advanced film dosimetry for a UK brachytherapy audit Advanced film dosimetry for a UK brachytherapy audit Tony Palmer Lead Scientist Medical Physics, Portsmouth Hospitals NHS Trust, UK University of Surrey, UK Institute of Physics and Engineering in Medicine

More information

Dosimetry and QA of proton and heavier ion beams

Dosimetry and QA of proton and heavier ion beams Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference

More information

8/1/2016. Motion Management for Proton Lung SBRT. Outline. Protons and motion. Protons and Motion. Proton lung SBRT Future directions

8/1/2016. Motion Management for Proton Lung SBRT. Outline. Protons and motion. Protons and Motion. Proton lung SBRT Future directions Motion Management for Proton Lung SBRT AAPM 2016 Outline Protons and Motion Dosimetric effects Remedies and mitigation techniques Proton lung SBRT Future directions Protons and motion Dosimetric perturbation

More information

Treatment Planning & IGRT Credentialing for NRG SBRT Trials

Treatment Planning & IGRT Credentialing for NRG SBRT Trials Treatment Planning & IGRT Credentialing for NRG SBRT Trials Hania Al Hallaq, Ph.D. Department of Radiation & Cellular Oncology The University of Chicago Learning Objectives Explain rationale behind credentialing

More information

Multi-Channel Film Dosimetry and Gamma Map Verification

Multi-Channel Film Dosimetry and Gamma Map Verification Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending Single Channel Film Dosimetry Calibration Curve X=R R ave

More information

Image Fusion, Contouring, and Margins in SRS

Image Fusion, Contouring, and Margins in SRS Image Fusion, Contouring, and Margins in SRS Sarah Geneser, Ph.D. Department of Radiation Oncology University of California, San Francisco Overview Review SRS uncertainties due to: image registration contouring

More information

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas 1,3, Natasa Milickovic 1, Nikolaos Zamboglou 2 1 Dept. of Medical Physics & Engineering, 2 Strahlenklinik,

More information

Establishing the impact of temporary tissue expanders on

Establishing the impact of temporary tissue expanders on Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions A. Asena 1,*, T. Kairn 1,2, S.B. Crowe 1 and J.V Trapp 1 1 School of Chemistry, Physics and Mechanical

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

Verification of micro-beam irradiation

Verification of micro-beam irradiation Journal of Physics: Conference Series OPEN ACCESS Verification of micro-beam irradiation To cite this article: Qiongge Li et al 2015 J. Phys.: Conf. Ser. 573 012047 View the article online for updates

More information

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas Dept. of Medical Physics & Engineering, Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Avoid exposure of film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities Dual-Energy Imaging of Bone Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities W Zbijewski, 1 A Sisniega, 1 JW Stayman, 1 N Packard, 2 J Yorkston, 2 G Thawait, 3 S Demehri, 3 J Fritz,

More information

The Physics of Oesophageal Cancer Radiotherapy

The Physics of Oesophageal Cancer Radiotherapy The Physics of Oesophageal Cancer Radiotherapy Dr. Philip Wai Radiotherapy Physics Royal Marsden Hospital 1 Contents Brief clinical introduction Imaging and Target definition Dose prescription & patient

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the PMMA ring holder with 7 containers are also shown. Relative

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

R. Harding, P. Trnková, S. J. Weston, J. Lilley, C. M. Thompson, S. C. Short, C. Loughrey, V. P. Cosgrove, A. J. Lomax, and D. I.

R. Harding, P. Trnková, S. J. Weston, J. Lilley, C. M. Thompson, S. C. Short, C. Loughrey, V. P. Cosgrove, A. J. Lomax, and D. I. Benchmarking of a treatment planning system for spot scanning proton therapy: Comparison and analysis of robustness to setup errors of photon IMRT and proton SFUD treatment plans of base of skull meningioma

More information

Dosimetric Analysis Report

Dosimetric Analysis Report RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

Clinical Results of Carbon Ion Radiotherapy: The Heidelberg Experience

Clinical Results of Carbon Ion Radiotherapy: The Heidelberg Experience Clinical Results of Carbon Ion Radiotherapy: The Heidelberg Experience Stephanie E. Combs, MD Department of Radiation Oncology University of Heidelberg, Germany Carbon ion RT at GSI Active beam delivery

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

Correction of Prompt Gamma Distribution for Improving Accuracy of Beam Range Determination in Inhomogeneous Phantom

Correction of Prompt Gamma Distribution for Improving Accuracy of Beam Range Determination in Inhomogeneous Phantom Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.27 pissn 258-4445, eissn 258-4453 Correction of Prompt Gamma Distribution for Improving Accuracy

More information

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital

More information

CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING

CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING Julian Liu a, Keisha Robinson a, DhanaJayan Kothandan a and Joshua Luis b (a) Cancer Centre London (b) University College London

More information

Treatment Planning (Protons vs. Photons)

Treatment Planning (Protons vs. Photons) Treatment Planning Treatment Planning (Protons vs. Photons) Acquisition of imaging data Delineation of regions of interest Selection of beam directions Dose calculation Optimization of the plan Hounsfield

More information

Proton and heavy ion radiotherapy: Effect of LET

Proton and heavy ion radiotherapy: Effect of LET Proton and heavy ion radiotherapy: Effect of LET As a low LET particle traverses a DNA molecule, ionizations are far apart and double strand breaks are rare With high LET particles, ionizations are closer

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Evaluation of a single-scan protocol for radiochromic film dosimetry

Evaluation of a single-scan protocol for radiochromic film dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Evaluation of a single-scan protocol for radiochromic film dosimetry Yoshinobu Shimohigashi, 1a Fujio Araki, 2 Masato Maruyama, 1

More information

Metal Artefact Reduction in CT

Metal Artefact Reduction in CT Metal Artefact Reduction in CT DANIEL MARRINER Metal Artefact Reduction in CT Metal Artefact Clinical Indications for MAR SEMAR and How It Works Technical Considerations Case Studies utilising SEMAR Metal

More information

Uncertainties in proton therapy: Analysis of the effects of density changes, calibration curve errors and setup errors in proton dose distributions.

Uncertainties in proton therapy: Analysis of the effects of density changes, calibration curve errors and setup errors in proton dose distributions. Uncertainties in proton therapy: Analysis of the effects of density changes, calibration curve errors and setup errors in proton dose distributions. A treatment planning study performed on water phantoms

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ Eighth Annual Chapman Phytosanitary Irradiation Forum Dosimetry for phytosanitary applications: dos and don ts Florent KUNTZ June 14 th, 2018 What does a dosimeter measure? Question A dosimeter measures

More information

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry Journal of Radiation Research, 2012, 53, 930 935 doi: 10.1093/jrr/rrs030 Advance Access Publication 21 August 2012 Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

More information

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During

More information

High-Dose Standardization Study For γ-rays radiation processing at NIM

High-Dose Standardization Study For γ-rays radiation processing at NIM High-Dose Standardization Study For γ-rays radiation processing at NIM ZHANG Yanli, GAO Junchen, YANG Yuandi. Division of Metrology in Ionizing Radiation and medicine, National Institute of Metrology,

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

25 Years of Proton Radiation Therapy at PSI an Overview

25 Years of Proton Radiation Therapy at PSI an Overview 25 Years of Proton Radiation Therapy at PSI an Overview Gudrun Goitein for the Team of the Center for Proton Therapy Center for Proton Therapy Paul Scherrer Institut (www.psi.ch) 5232 Villigen PSI Switzerland

More information

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy Antony L Palmer, 1,2a David Bradley,

More information

Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer

Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 1, WINTER 2009 Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer Annelise Giebeler, 1,2 Jonas Fontenot,

More information

Strengthening the weakest link: improving tumour definition

Strengthening the weakest link: improving tumour definition Strengthening the weakest link: improving tumour definition Marianne C Aznar marianne.aznar@manchester.ac.uk University of Manchester / The Christie NHS Clinical Trial Service Unit, University of Oxford

More information

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

A review of RTTQA audit. Karen Venables

A review of RTTQA audit. Karen Venables A review of RTTQA audit Karen Venables CHART 8 hours on the machine Chart cord CHART QA: Phantoms(1 st use of anatomical phantoms) Designed for treatments in Bronchus and Head and Neck (2D only) Outlines

More information

First, how does radiation work?

First, how does radiation work? Hello, I am Prajnan Das, Faculty Member in the Department of Radiation Oncology at The University of Texas MD Anderson Cancer Center. We are going to talk today about some of the basic principles regarding

More information

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Dosimetry Standards and Dissemination Systems for Radiation Processing in China Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria Organization Founded in 1955, National Institute

More information

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim

More information

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging Research Journal of Applied Sciences, Engineering and Technology (): 6-, ISSN: -767 Maxwell Scientific Organizational, Submitted: November 9, Accepted: November 9, Published: February, Comparison of Energy

More information

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Abdul Kazi, Ph.D., DABR Assistant Professor, Department of Radiation Oncology University of Maryland

More information

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania

Proton Treatment. Keith Brown, Ph.D., CHP. Associate Director, Radiation Safety University of Pennsylvania Proton Treatment Keith Brown, Ph.D., CHP Associate Director, Radiation Safety University of Pennsylvania Proton Dose vs. Depth Wilson,. R.R. Radiological use of fast protons. Radiology 47:487-491, 1946.

More information

A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy

A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy Poster No.: R-0008 Congress: 2014 CSM Type: Scientific Exhibit Authors: T. Tran, W. Ding, L. Melvern, M. Chao, B. Subramanian;

More information

Collateral Exposure: The Additional Dose From Radiation Treatment

Collateral Exposure: The Additional Dose From Radiation Treatment Department of Physics and Astronomy, University of Canterbury, Private Bag 4800, Christchurch, New Zealand Collateral Exposure: The Additional Dose From Radiation Treatment Katherine Fricker (BSc) A thesis

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

Applications in brachytherapy

Applications in brachytherapy SAM Practical Medical Physics Session: TU-F-201 Radiochromic Film Dosimetry Update Tuesday, July 14, 2015, 2:45 pm 3:45 pm Applications in brachytherapy Samuel Trichter, M.Sc., Department of Radiation

More information