1 Radiography/Radiology Activity for 2017 Activity No: A1(17) Topic CT radiation Article CT radiation: key concepts for gentle and wise use Approved for (3) Clinical Continuing Educational Units (CEU s)

2 SAFETY 1706 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at CT Radiation: Key Concepts for Gentle and Wise Use 1 Mannudeep K. Kalra, MD Aaron D. Sodickson, MD, PhD William W. Mayo-Smith, MD Abbreviations: ACR = American College of Radiology, AEC = automatic exposure control, BEIR = Biologic Effects of Ionizing Radiation, CTDI vol = volume CT dose index, DLP = doselength product, DRL = diagnostic reference level, IR = iterative reconstruction, SSDE = sizespecific dose estimate RadioGraphics 2015; 35: Published online /rg Content Codes: 1 From the Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, Mass (M.K.K.); and Department of Radiology, Brigham and Women s Hospital, 75 Francis St, Boston, MA and Harvard Medical School, Boston, Mass (A.D.S., W.W.M.S.). Received April 17, 2015; revision requested June 3 and received June 16; accepted June 19. A.D.S. has provided disclosures (see p 1720); all other authors have disclosed no relevant relationships. Address correspondence to W.W.M.S. ( wmayo-smith@partners.org). RSNA, 2015 Use of computed tomography (CT) in medicine comes with the responsibility of its appropriate (wise) and safe (gentle) application to obtain required diagnostic information with the lowest possible dose of radiation. CT provides useful information that may not be available with other imaging modalities in many clinical situations in children and adults. Inappropriate or excessive use of CT should be avoided, especially if required information can be obtained in an accurate and time-efficient manner with other modalities that require a lower radiation dose, or non radiation-based imaging modalities such as ultrasonography and magnetic resonance imaging. In addition to appropriate use of CT, the radiology community also must monitor scanning practices and protocols. When appropriate, high-contrast regions and lesions should be scanned with reduced dose, but overly zealous dose reduction should be avoided for assessment of low-contrast lesions. Patients cross-sectional body size should be taken into account to deliver lower radiation dose to smaller patients and children. Wise use of CT scanning with gentle application of radiation dose can help maximize the diagnostic value of CT, as well as address concerns about potential risks of radiation. In this article, key concepts in CT radiation dose are reviewed, including CT dose descriptors; radiation doses from CT procedures; and factors and technologies that affect radiation dose and image quality, including their use in creating dose-saving protocols. Also discussed are the contributions of radiation awareness campaigns such as the Image Gently and Image Wisely campaigns and the American College of Radiology Dose Index Registry initiatives. RSNA, 2015 radiographics.rsna.org Introduction Technologic innovations in computed tomography (CT), including helical and multi detector row capabilities, have expanded the applications of CT in modern medicine (1 3). Use of CT has increased substantially in recent years, with approximately 70 million CT examinations performed annually in the United States (1). Abdominal, head, and chest CT examinations represent the majority of CT procedures. CT examinations represent about 17% of radiationbased medical procedures but contribute approximately 49% of the medical radiation dose to the population as a whole (2). Despite concerns about the potential risks of radiation at CT, in appropriate clinical settings CT provides information vital to accurate clinical diagnosis and patient care (1,3). Technologic advances have led to the introduction of many factors that can complicate scanning protocols but are crucial to optimize the technology and tailor examinations to clinical indications, body regions, and body sizes (4 7). Throughout this article, unless specified otherwise, when we mention size we are referring to the cross-sectional area

3 RG Volume 35 Number 6 Kalra et al 1707 TEACHING POINTS AEC techniques allow automatic modulation of tube current according to patient size and shape. There are several types of IR approaches that incorporate different x-ray photon noise statistics and system optics modeling. The most important step toward use of the appropriate radiation dose is ensuring its appropriateness on the basis of required clinical information. Compared with routine chest CT (eg, for cancer staging), lung nodule follow-up CT should have a much lower radiation dose or CTDI vol. Dose tracking or monitoring can help in CT radiation dose management by allowing audit and comparison of dose indexes for different CT examinations. of the patient in the scanned region. Automatic exposure control (AEC), automatic voltage selection, and iterative reconstruction (IR) techniques are examples of factors that affect image quality and associated radiation dose. In this article, we address key aspects of CT radiation dose as outlined in the American Board of Radiology Core Quality and Safety Study Guide (8). These include CT dose descriptors; radiation doses for CT procedures; and factors and technology that affect radiation dose and image quality, as well as the use of these factors in creating dose-saving protocols. Radiation awareness campaigns including Image Gently and Image Wisely and the American College of Radiology (ACR) Dose Index Registry initiative are discussed. Radiation Risks Radiation exposure has received much attention in both the medical literature and the lay media. This attention has focused, in part, on a number of highly publicized events in which radiation exposure from improperly performed CT perfusion examinations led to the deterministic effects of skin erythema and hair loss (9,10). It is important, however, to distinguish medical errors from properly performed examinations with dose levels that are well below these deterministic effect thresholds. Controversy over the shape of the doseresponse curve linking cancer risk to the radiation doses used at diagnostic imaging is ongoing (11,12). Cancer data from the survivors of the atomic bombs at Hiroshima and Nagasaki, Japan, show a statistically significant increase in cancer incidence for one-time acute exposure of msv or greater, yet most diagnostic imaging examinations (including CT) impart patient doses well below 20 msv (with the possible exception of multiphasic scanning of the torso) (13). There is debate about how best to incorporate the atomic bomb risk data into low-dose risk models. Some practitioners believe that there likely is a threshold dose below which there is no lasting effect (the linear threshold model) or even that low levels of radiation are protective (the hormesis model). In the absence of definitive data linking low-dose radiation to cancer induction, however, most organizations have adopted a linear no-threshold model such as that used in the seventh report of the Biologic Effects of Ionizing Radiation (BEIR) study (BEIR VII), in which doubling the radiation exposure is assumed to double the associated cancer risk (14). As shown in Figure 1, this model shows that for the same x-ray exposure, risks are expected to be higher for young patients, likely because cell division is more active in young people, and their life expectancy after the exposure is longer (14). These facts highlight the need to image gently in younger patients. Female patients are also thought to be at higher risk for a given exposure, due in part to risks of breast, ovarian, and uterine cancer, as well as to an increased risk of lung and thyroid cancer without a clearly understood biologic basis. A large amount of radiation exposure data linked to eventual cancer outcomes is needed to test the cancer risks of diagnostic imaging directly (13). The fundamental epidemiologic challenge is to detect a small incremental increase in the 42% baseline lifetime rate of cancer in the U.S. population. Large-scale epidemiologic studies to date have been focused primarily on pediatric cancer risks, and in these populations researchers generally have found cancer risks of a magnitude similar to those predicted by downward extrapolation of the atomic bomb data by using linear no-threshold models (15 18). However, fundamental questions remain as to whether the identified risk increases are real or related to issues such as reverse causation, indication bias, or predisposing patient factors (19 21). Although the absolute risk of performing CT is not definitively known, radiologists, physicists, technologists, and CT technology manufacturers must optimize available technology to reduce the potential risks from ionizing radiation while maintaining the ability to perform high-quality imaging that allows clinical questions to be answered. We believe that it is prudent to adhere to the as low as reasonably achievable, or ALARA, principle when performing medical imaging. Units of Radiation Dose and Metrics Used at CT Absorbed dose (also referred to as organ dose) is the energy absorbed by tissue divided by the tissue mass. It is measured in grays (grays =

4 1708 October Special Issue 2015 radiographics.rsna.org Figure 1. BEIR VII linear nothreshold risk model demonstrating lifetime attributable risk (LAR) of cancer incidence, an additive risk higher than the baseline rate of 42%. Graph shows that for patients who receive CT exposure of 10 msv, which is common at abdominal CT, excess risk is predicted to be 1/1000 (where the y-axis equals 1) for a 32-year-old woman or a 19-year-old man. Data for this graph were extracted from Table 12D of BEIR VII (14). joules per kilogram). CT radiation outputs are described by using the volume CT dose index (CTDI vol, expressed in milligrays) and doselength product (DLP, expressed in milligraycentimeters) (22,23). CTDI vol is calculated on the basis of measurements made in standard plastic (polymethyl methacrylate) 16-cm head or 32-cm body phantoms to quantify the x-ray tube output emitted by the scanner. CTDI vol is an estimate of average x-ray tube output from an acquired CT image series, and DLP is the total x-ray tube output integrated throughout the entire scan, which is obtained by multiplying CTDI vol by the scanned length (in centimeters) of the patient. Both CTDI vol and DLP can be used for comparison of different CT protocols and scanners. Anticipated values of these technique descriptors are displayed on the user interface of the scanner before acquisition of the CT data (with more accurate performed values reported after the scan when AEC is used), thus providing to the technologist an estimate of scanner output before the patient undergoes imaging. Recently, a CT dose alert feature, which allows users to set certain CTDI vol values to enable the scanner to alert the user if a specified limit is exceeded, has been introduced in scanners. The dose information page and structured Digital Imaging and Communications in Medicine, or DICOM, report available on most CT scanners tabulate CTDI vol and DLP for each acquired CT series, and can be used for quality assurance when reviewing scans on a picture archiving and communication system, or PACS, or for subsequent data capture for large-scale dose monitoring efforts (24). These metrics represent scanner x-ray tube output and not patient dose (25). For example, at any given CTDI vol value for the same diameter phantom, organ doses are substantially larger for small patients than for large patients, in whom many of the incident x-rays are blocked in the peripheral soft tissues and thus do not deposit energy in the centrally positioned, more radiation-sensitive organs (26). A recent report from the American Association of Physicists in Medicine (AAPM) defined a relatively new metric, the size-specific dose estimate (SSDE), which is calculated by adjusting (or normalizing) the CTDI vol to the size of the patient to better estimate patient-specific organ doses (27 29). To calculate SSDE, CTDI vol is multiplied by a correction factor on the basis of patient size by using tables in the AAPM report. Patient size may be estimated from the anteroposterior or lateral diameters, the sum of the anteroposterior and lateral diameters, or the effective diameter (square root of the product of anteroposterior and lateral diameters) of the patient. A subsequent AAPM report (30) identified the patient s water-equivalent diameter (diameter of a cylinder of water that would yield the same overall x-ray attenuation as the cross section through the patient) as the preferred size metric for automated SSDE calculation. SSDE is not used to estimate effective dose. Effective dose (measured in sieverts) is intended to represent the uniform whole-body equivalent dose that would be expected to cause the same overall cancer risk as the nonuniform or partial-body exposure delivered to the patient. It is calculated as a weighted sum of each organ s dose times that organ s relative risk of radiationinduced carcinogenesis; it is a single dose value that is commonly used to compare the relative biologic risks of different exposures (from the same or different imaging modalities), even if the exposure is to different body regions (23). Effective dose can be estimated by multiplying the DLP by a conversion factor on the basis of the body region scanned in a sex- and age-neutral reference patient, although these methods can be

5 RG Volume 35 Number 6 Kalra et al 1709 inaccurate when applied to individual patients (31,32). Effective dose reflects an average population-based risk to an average individual of average size, not the actual risk to a specific patient. Benchmarks for CT Radiation Doses A primary purpose of surveys of radiation dose from CT is to establish diagnostic reference levels (DRLs) for different CT examinations, which help establish benchmark doses for various CT procedures to avoid excess radiation dose (2,33). Although DRLs typically represent radiation dose levels used by 75% of the surveyed CT facilities (recorded as CTDI vol ), users should use diagnostic reference ranges of 25% 75% levels as one common range. In 2013, the CT scanner accreditation process from the ACR set DRLs of 75 and 35 mgy (both with a 16-cm CTDI phantom size) for adult and pediatric head CT examinations, respectively (34). The DRLs for adult and pediatric (in a 5-year-old child) abdominal CT examinations were set at 25 mgy (32-cm CTDI phantom size) and 15 mgy (16-cm CTDI phantom), respectively. Although there is no chest CT reference dose evaluation in the ACR accreditation process, the American Association of Physicists in Medicine and ACR practice guideline has set an adult DRL of 21 mgy (32-cm phantom) (35). The National Council on Radiation Protection and Measurements report number 172 (36) states that, for institutions already operating below the DRLs, the next target should be achievable doses, which represent radiation dose levels below which 50% of the institutions are operating without compromising required diagnostic image quality (35 37). These DRLs and achievable doses refer to CTDI vol values from a single CT series for a standard-sized patient; smaller patients are expected to need lower doses, and larger patients may need higher radiation doses. Factors Affecting Radiation Dose and Quality Several CT parameters and patient factors affect CT radiation dose and image quality. The most important strategy for reducing radiation dose associated with CT is ensuring the appropriateness or justification for CT. These factors should be accounted for in the creation of CT protocols for different body regions and clinical indications. Patient Factors A few key concepts are important regarding the effect of patient cross-sectional area on CT radiation dose and image quality. If constant scanning factors such as tube current, tube potential, gantry rotation time, detector configuration, and image reconstruction parameters are maintained, image noise (a general inverse surrogate of image quality in CT) will increase with increasing patient size and will decrease with decreasing patient size. Thus, for the same clinical indication and body region, larger patients generally require higher x-ray tube output, while smaller patients can be scanned at a lower radiation dose to achieve the same image quality. Small children can and should be scanned with substantially lower radiation doses than those used with adults. Regions with lower x-ray attenuation (eg, the air-containing thorax) should be scanned at lower dose compared with regions with higher x-ray attenuation (eg, the abdomen). Both AEC and automatic tube potential selection techniques accommodate these attenuation differences by modifying scanning parameters appropriate to the size of the patient and the body region (1,38,39). The clinical indication for CT is an important patient factor, because conspicuity of different findings at CT is influenced differently by changes in image quality or applied radiation dose. Certain inherently high-tissue-contrast tasks such as lung nodule follow-up or kidney stone evaluation can be performed by using lower radiation doses, because image noise is less detrimental to detection of these high-contrast structures. CT examinations performed to detect low-contrast lesions such as mediastinal lymph nodes or liver tumors require less image noise and thus higher doses for accurate detection and characterization. Applied radiation dose affects some attributes of image quality, such as image noise and certain artifacts, which increase with decreasing radiation dose. These attributes can affect lesion conspicuity negatively, although other factors such as lesion size, location, contrast enhancement, and patient cross-sectional area also can have an effect on lesion conspicuity. This complex relationship demands attention to details in CT protocols, which should be designed with consideration of the patient cross-sectional area, body region scanned, and examination indication. Clinical indication also dictates the scanning coverage length and the number of scanning phases. Scanning Factors Many scanning factors affect image quality and radiation dose. Although selection of some factors such as tube current (in milliamperes) and tube voltage (in kilovolts) has been automated by some CT vendors, the CT scanners of several other vendors need manual adjustment, such as gantry rotation time, detector configuration, and pitch. Different CT vendors each have their own proprietary nomenclature for similar scanning factors, which can create confusion

6 1710 October Special Issue 2015 radiographics.rsna.org at institutions in which different scanners are used. A good resource to navigate the lexicon for all major CT vendors is now available on the American Association of Physicists in Medicine Web site (40). In addition to nomenclature variability, changes in certain scan factors produce different effects on image quality and radiation dose when equipment from different vendors is used. The terminology and effects have been reviewed in a recent article (1). It is important to work closely with technologists, medical physicists, and industry specialists to understand and optimize the unique and variable functionality of each type of scanner. In implementation of strategies for radiation dose reduction for CT, maintaining the required diagnostic quality and information for which the CT examination was requested is important. A lower-dose CT examination with insufficient visualization of required structures is just as suboptimal as a higher-dose CT examination that provides higher image quality than needed to establish a diagnosis. Tube Current and Tube Current Time Product. Some vendors use tube current (in milliamperes) as an independent factor, while others use tube current time product (in milliampere-seconds), which is milliamperes multiplied by the gantry rotation time and divided by the helical pitch factor. A change in tube current or tube current time product is linearly related to the radiation output from the x-ray tube, such that doubling of tube current or tube current time product doubles the radiation output. Adjustment of tube current or tube current time product remains the most practical and most commonly used method of adapting radiation output from CT scanners (4,5). Although most multi detector row CT scanners now use AEC techniques to modulate tube current as appropriate for patient attenuation, older scanners and certain low-dose scanning protocols still operate with fixed tube current or tube current time product. As a rule of thumb, smaller patients and body regions with lower x-ray attenuation (such as the chest) should be scanned at lower tube current or tube current time product settings, and larger patients or more higher attenuating body regions (such as the shoulders or abdomen) require higher tube current or tube current time product settings. Clinical indications for assessment of highcontrast regions or lesions (eg, lung nodules, kidney stones) are less affected by image noise and, therefore, enable a lower tube current or tube current time product technique compared with organs with lower inherent tissue contrast (eg, liver, pancreas). When AEC techniques are unavailable or not used, scan protocols should allow suitable adjustment of tube current time product to account for variations in patient size, region scanned, and clinical indications. Tube Voltage. Tube voltage represents the peak photon energy of an x-ray energy spectrum. A change in voltage has a much greater effect on radiation dose than does a change in fixed tube current or tube current time product. For example, a 15% reduction in tube current time product results in 15% dose reduction if all other scanning parameters are held constant, whereas a 15% reduction in voltage results in a 35% dose reduction if other scanning parameters are held constant. Smaller children and smaller adults should be scanned at lower voltage ( 100 kv) to reduce radiation dose. Unlike change in tube current, a change in voltage is associated with a change in CT numbers as well. This is particularly beneficial at contrast material enhanced examinations such as CT angiography (including CT pulmonary angiography and coronary CT angiography) and CT perfusion, because iodine enhancement increases substantially at lower voltage (41 44). A decrease in voltage reduces radiation dose and substantially improves image contrast, particularly for contrastenhanced CT examinations. Lower voltage can also help reduce the amount of contrast material required for CT scanning (41 44). Lower voltage is being used increasingly for contrast-enhanced CT in adults and children. Because a decrease in voltage is associated with an increase in image noise, since fewer of the x-rays in the lower energy spectrum make it through the patient to the detector array, a compensatory increase in tube current time product is often used to maintain similar image noise. As an alternative, IR techniques that reduce image noise may enable use of lower voltage in adult patients with reduced radiation dose (43,44). Regions with lower x-ray attenuation (eg, chest, extremities) also can be scanned at lower voltage ( 100 kv) compared with regions with higher x-ray attenuation (abdomen). On scanners with IR techniques, lower voltage is being used increasingly for contrast-enhanced CT studies in adult patients as well. Gantry Rotation Time and Pitch. Image quality is dependent on adequate x-ray flux passing through the patient and reaching the detector array. As we described previously, x-ray flux may be increased by increasing voltage (when relevant) or by increasing the tube current time product, which may be achieved by increasing either tube current, effective tube current time product, or gantry rotation time (expressed in seconds), which refers to the duration of a 360 revolution

7 RG Volume 35 Number 6 Kalra et al 1711 Figure 2. Coronal CT images in a 5-year-old boy (weight, 35 kg) show nodules in the right lower (a, b) and left upper (b) lobes that were related to fungal infection (Histoplasma infection). Images were acquired with high nonoverlapping pitch (3:1) and scanning duration of 0.28 second to eliminate the need for sedation and general anesthesia. The CTDI vol and DLP were 2.6 mgy and 62 mgy cm, respectively. of the x-ray tube around the patient. Gantry rotation time and pitch are the two main parameters that determine the scanning speed. In general, gantry rotation time should be kept low to avoid motion artifacts. This is particularly important in small children or in imaging of mobile structures (eg, the lungs during breath-hold scanning), where motion artifacts may necessitate repeat scanning. However, in large patients in whom the maximum tube current limits of the x-ray tube are readily reached, further increases in x-ray flux may require scanning to be slowed down by using a slower gantry rotation time or a reduced pitch. The pitch of a helical CT acquisition is defined as the table travel (CT table movement in millimeters per gantry rotation) divided by the collimated beam width (in millimeters). If the table travel in millimeters for one 360 gantry rotation equals the x-ray collimated beam in millimeters, the pitch is 1. If table travel is less than the x-ray collimated beam width, there is overlap at each view angle from one gantry rotation to the next (overlapping pitch < 1). A pitch greater than 1 (nonoverlapping) implies that table travel is greater than the x-ray beam width, and some view angles are not recorded at certain table positions. Generally speaking, pitch selection should be based on the requirement of scanning speed. If all other scanning factors are held constant, higher pitch results in faster scanning (Fig 2). For coverage of longer regions of the body, a nonoverlapping pitch is often used, whereas regions with rapidly changing anatomy (eg, skull base) are scanned with overlapping pitch. Likewise, to minimize scanning times, as in patients with shortness of breath who are undergoing CT pulmonary angiography, a nonoverlapping pitch with fast gantry rotation time is preferred to minimize motion artifacts. The effect of changes in pitch varies according to CT scanner vendor. On most Siemens Healthcare (Forchheim, Germany) and Philips Healthcare (Andover, Mass) CT scanners, a change in pitch is associated with a compensatory change in the tube current, such that the tube current time product and radiation dose remain essentially constant. On most scanners from GE Healthcare (Milwaukee, Wis) and Toshiba Medical Systems (Tokyo, Japan), tube current is not automatically changed with pitch; thus, radiation dose decreases as pitch is increased and vice versa. Detector Configuration. Detector configuration refers to the number of detector rows and the width of each detector row. The detector configuration determines x-ray beam collimation or width. Most multi detector row CT scanners offer two or more detector configurations. Some scanners (mostly those with 64 or fewer detector rows) contain detector elements of variable thickness, such that the detector configuration is tied to the desired minimum section thickness. For example, acquisition at mm can be accomplished only with a detector configuration of mm and not with mm. Thus if submillimeter sections are desired, a thinner x-ray beam is necessary, which can increase the radiation dose because of increased scanning time and decreased radiation dose efficiency. For other scanners (mostly those with 64 or more detector rows), a uniform array of detector configurations does not affect the

8 1712 October Special Issue 2015 radiographics.rsna.org Table 1: AEC Techniques with CT Scanners from Different Vendors AEC Type GE Healthcare Philips Healthcare Siemens Healthcare Toshiba Medical Systems Transverse SmartScan D-DOM Care Dose NA Longitudinal AutomA Z-DOM NA Sure Exposure Angular and longitudinal Smart ma Work in progress Care Dose 4D Sure Exposure 3D Image quality reference parameter for AEC Noise index Reference image Quality reference mas Standard deviation Minimum and maximum tube current control Note. NA = not available as separate item. Present NA Controlled with modification of strength of modulation Present Figure 3. Coronal CT image in a 41-year-old woman (body mass index, 38 kg/m 2 ) with left flank pain who underwent nonenhanced CT of the abdomen and pelvis with AEC (AutomA [GE Healthcare], 35 noise index) and 120 kv (CTDI vol = 4.9 mgy). A nonobstructive 4-mm calculus (arrow) can be seen in the lower pole of the left kidney. Appropriate modification in AEC technique based on clinical indication (in this patient, much higher noise index than that with routine abdominal CT protocol noise indexes) helped to reduce radiation dose. minimum section thickness. With such scanners, a wider detector configuration is preferred for scanning longer body regions (such as the chest, abdomen, and extremities), and a shorter configuration is used for scanning in a short scanning range (such as those used to image the head and joints). Dose-saving Technology AEC Techniques Most modern scanners employ AEC techniques that modulate the tube current (and therefore the radiation dose) to the patient s body habitus or size on the basis of regional x-ray attenuation measured from planning projection (scout) radiographs. For most CT protocols in the body, AEC techniques are preferred over fixed or manual selection of tube current (5). AEC techniques allow automatic modulation of tube current according to patient size and shape. Spatial tube current modulation techniques include angular (transverse or x-y), longitudinal (z), and combined (x-y-z) modulation (Table 1), with combined modulation techniques preferred for most indications when available (23,45). In addition, temporal tube current modulation is available to vary the tube current during different portions of the cardiac cycle at electrocardiographically gated cardiac CT examinations. To configure AEC techniques, users prescribe an image quality parameter on the basis of the body region to be scanned and the clinical indication (Fig 3). For example, a higher image quality requirement is appropriate for CT pulmonary angiography compared with that appropriate for follow-up CT for lung nodules (4). Substantial dose reduction has been described with these techniques in both adult and pediatric patients compared with fixed tube current techniques (23,45,46). With the use of AEC techniques, close attention should be paid to patient position in the gantry isocenter, because off-centered patient positioning can lead to miscalculation of patient attenuation and, therefore, also miscalculation of the applied tube current (45). Off centering of the patient in the gantry by the technologist before the helical scan can increase the radiation dose without a substantial gain in diagnostic image quality. This occurs because off centering the patient too close to the x-ray tube for the planning radiograph (scout image) results in an inflated estimate of

9 RG Volume 35 Number 6 Kalra et al 1713 Figure 4. Radiation dose reduction with reduction in voltage for CT pulmonary angiography. With use of automatic voltage selection technique (Care kv; Siemens Healthcare), scanners can adapt the voltage on the basis of patient size and protocol type. (a) Axial image in a 66-year-old man with morbid obesity (body mass index, 52 kg/m 2 ) scanned at 120 kv (CTDI vol, 19 mgy) shows segmental pulmonary embolism in right lower lobe (arrow). (b) Axial image in a 44-year-old man (body mass index, 34 kg/m 2 ) scanned at 100 kv (CTDI vol, 7 mgy) shows lobar pulmonary emboli bilaterally in lower lobes (arrows). (c) Axial CT image in a 64-year-old woman (body mass index, 24 kg/m 2 ) scanned at 80 kv (CTDI vol, 3.7 mgy) shows segmental pulmonary embolism in right lower lobe (arrow). patient width. CT scanners use the patient width from the scout image to determine the appropriate tube output for AEC. If the patient width from the projection image is artificially high, the scanner will be fooled to use a higher-than-necessary tube current to generate an image. Automatic Voltage Selection Recent automatic tube potential selection techniques enable some modern CT scanners to automatically select the voltage that achieves the desired image quality at the lowest CTDI vol appropriate for the patient s size and the diagnostic task (eg, nonenhanced CT vs CT angiography). Unlike tube current, voltage is still manually selected by the users of most CT scanners. To simplify and automate the selection of voltage, some vendors have introduced automatic voltage selection techniques (Care kv, Siemens Healthcare; kv Assist, GE Healthcare). These techniques select the most appropriate voltage for the scan, whereas the AEC techniques modulate the tube current throughout the patient anatomy. To configure the automatic voltage selection techniques, users identify the examination type (for example, nonenhanced CT, contrast-enhanced CT, or CT angiography) and select a reference voltage and tube current (Fig 4) (38,39). On the basis of the examination type and the patient size, the algorithm determines the voltage that will achieve the desired image quality at the lowest CTDI vol. Users can constrain the range of allowed voltage for each protocol (for example, by eliminating 140 kv as an option in CT angiographic examinations because of the loss of iodine enhancement at high voltage). Because lower voltage increases contrast enhancement, automatic voltageselection techniques select lower voltage more frequently for CT angiography and contrast-enhanced scanning than for nonenhanced scanning. Substantial dose reductions have been reported for both chest and abdominal CT examinations with the use of automatic voltage selection techniques (38,39,47). IR Techniques CT image reconstruction techniques have a profound effect on image quality and artifacts. Traditional filtered backprojection techniques make a number of simplifying assumptions about photon noise statistics and scanner hardware details (presuming that the x-ray focal spot and detector

10 1714 October Special Issue 2015 radiographics.rsna.org Table 2: IR Techniques Available from Different CT Vendors Technique and Details GE Healthcare Philips Healthcare Siemens Healthcare Main IR technique Details of main technique Other available IR techniques Toshiba Medical Systems ASiR idose IRIS Adaptive iterative dose reduction ASiR level is selected idose level can be These techniques Three settings at 10% increments set on a scale of work on a scale of (mild, standard, (10% 100%), with 1 7, with in- 1 5, with increas- strong) represent increasing percent- creasing number ing number asage increasing associated with suggesting lower sociated with lower strength of noise lower noise noise noise reduction Veo, ASiR-V IMR Safire, Admire Adaptive iterative dose reduction 3D elements are infinitely small) that allow rapid image reconstruction but can produce high image noise and artifacts, particularly at lower radiation dose levels. The CT industry has introduced more computationally intense IR techniques to improve CT image quality and enable use of lower radiation dose compared with filtered backprojection techniques (48,49). There are several types of IR approaches (Table 2) that incorporate different x-ray photon noise statistics and system optics modeling to generate better image quality compared with that of filtered backprojection images. Many IR techniques allow the user to specify the strength of the IR techniques, which allows control of the degree of noise reduction (Figs 5, 6). When high-strength IR techniques are used, a distinct or altered image appearance (blotchy appearance on images) can be seen. IR techniques do not reduce radiation dose, but by decreasing image noise, they allow users to adjust scanning parameters (most commonly lowering tube voltage and/ or kilovolts) to reduce tube output and, hence, dose compared with that achieved with traditional filtered backprojection techniques (48,49). IR techniques also have been used in children to enable radiation dose reduction. Wallihan et al (50) recently reported substantial reduction in SSDE for children undergoing CT enterography with an adaptive iterative dose reduction IR technique (6 mgy ± 2.1) compared with that with filtered backprojection (17 mgy ± 5). IR techniques also have been found to help reduce radiation dose in children undergoing CT hematuria protocols and head CT (51 53). Authors of a recent study (54) reported degradation of low-contrast spatial resolution with at least two commercial IR techniques at a 25% 50% reduction in dose relative to full-dose CT and filtered backprojection techniques. Therefore, radiation dose should be reduced in small steps when IR techniques are implemented. Hardware-driven Technology for CT Dose Reduction Vendors have introduced several types of hardware-driven technologies to enable efficient use of radiation with CT (55 57). This type of technology tends to improve the radiation dose efficiency of CT scanners by reducing unused portions of the x-ray beam, which are outside the detector assembly relative to the useful portion of the x-ray beam. For example, beam-shaping or bow-tie filters help reduce radiation dose to the thinner peripheral portions of the body while allowing most x-ray photons to pass through the thicker central parts of the body. Most modern CT scanners are equipped with these filters, which also help to maintain a constant image quality on the central and peripheral parts of the images. These filters function correctly only when the patient is centered appropriately in the CT gantry. Off centering of the patient can result in asymmetric image noise and excessive radiation dose to the thinner peripheral portions of the body and to radiosensitive tissue such as that of the breasts and thyroid (55). Some multi detector row CT scanners with wide area detectors (typically 4 cm) use adaptive shielding to improve radiation dose efficiency, reducing overranging by blocking the portions of the wide x-ray beam that would otherwise fall outside the desired anatomic regions of the scan at the start and the end of a helical acquisition (56). Overranging can contribute to a substantial portion of the associated radiation dose, particularly for shorter scanning lengths. The detector array is one of the most expensive and important parts of a CT scanner. Recent development of more efficient detector elements has helped to improve radiation dose efficiency, as has introduction of more sophisticated and integrated data assembly systems that produce less electronic noise and thus allow scanning with lower tube outputs (46).

11 RG Volume 35 Number 6 Kalra et al 1715 Figure 5. (a) Axial image in a 52-year-old man (body mass index, 33 kg/m 2 ) who underwent standard-of-care CT, which was done with standard CT protocol (CTDI vol, 13 mgy). (b d) With informed consent, the patient was rescanned at CTDI vol of 0.9 (b), 0.4 (c), and 0.2 mgy (d). Reconstructed sections with sinogram-affirmed IR (Safire; Siemens Healthcare) technique at different strengths (S3, S4, and S5, respectively) showed that despite deterioration in image quality with decreasing doses, both nodules (arrow; arrowhead) in the right upper lobe are seen as well as in a with standard-of-care CT at much higher radiation dose. With decreasing radiation dose, the image noise can be reduced with increasing strength of IR. Use of extreme dose reduction (eg, 0.4 and 0.2 mgy) is not suitable for evaluation of nonpulmonary structures and should be reserved for patients undergoing lung nodule follow-up or detection only. Alternatives to CT The most important step toward use of the appropriate radiation dose is ensuring its appropriateness on the basis of required clinical information. Several radiologic organizations have developed algorithms to help guide appropriate use of imaging tests for specific clinical indications (1). The ACR appropriateness criteria suggest appropriate modalities for imaging on the basis of the patient s clinical presentation. Radiology order-entry decision support software has been developed by using these guidelines and other relevant evidence to improve appropriate justification for use of imaging examinations (58). Ten Steps to Wise and Gentle CT Protocols 1. Establish the appropriateness or justification guidelines for CT scanning. Decision support software can help set this most elementary rule for managing radiation dose. 2. Create clinical indication specific CT protocols for different anatomic areas of the body. Such protocols help tailor CT dose and image quality on the basis of specific clinical indications. 3. Ensure that the patient is positioned appropriately in the gantry isocenter. Off centering leads to incorrect estimation of patient attenuation, which results in incorrect tube current modulation with AEC techniques and disproportionate distribution of image noise. 4. Use AEC techniques to adapt tube current. AEC techniques help modulate tube current on the basis of required image quality and patient size. Some techniques allow users to set minimum and maximum tube current limits to prevent radiation levels that are too low or too high. Most body CT examinations must be performed with AEC to reduce radiation dose. 5. Use voltage adjustment on the basis of clinical indication and patient size. Automatic voltage selection techniques help to modify

12 1716 October Special Issue 2015 radiographics.rsna.org Figure 6. (a) Axial image in a 71-year-old man (body mass index, 34 kg/m 2 ) who underwent standard-of-care CT (CTDI vol, 13 mgy) shows a right upper lobe lung nodule (arrow). (b d) With informed consent, the patient was rescanned at CTDI vol of 0.8 (b), 0.4 (c), and 0.2 mgy (d). Reconstructed sections with adaptive statistical IR (ASiR; GE Healthcare) technique at different strengths (50%, 70%, and 90%, respectively) showed that despite the deterioration in image quality with decreasing doses, the right upper lobe lung nodule (arrow) could be seen almost as well as on images from the standard-of-care CT at much higher radiation dose. The strength of IR technique may require fine tuning, depending on associated radiation dose. voltage on the basis of the examination type and patient size. CT without such techniques should use low voltage for specific clinical indications (CT angiography) and patient size (children and smaller adults). 6. Select appropriate detector and section collimation on the basis of the clinical indication. Section collimation has an effect on image noise, with thinner collimation needing a higher dose to decrease image noise. At advanced CT, an acquire thin and reconstruct thick strategy helps decrease image noise and thus dose requirement. 7. Avoid overscanning beyond the anatomic area of interest: Extending the scanning range to regions beyond the required area increases radiation dose. For example, limit the scanning range to lung bases for lung nodule follow-up CT. Kidney stone protocol CT can be initiated from the top of the kidneys to the pubic symphysis instead of the dome of the diaphragm. 8. Limit the use of multiphasic CT protocols, and employ clever strategies for dose reduction. Repetitive scanning increases radiation dose; multiphasic CT should be performed only when it is expected to help in lesion detection or characterization. For example, routine acquisition of nonenhanced images before routine chest or abdominal CT does not add substantial information in most patients and, therefore, should be discouraged. Split-bolus contrast material injection techniques can help reduce radiation dose by decreasing the number of required phases for some protocols such as those for CT urography. 9. Use IR techniques to improve image quality and allow dose reduction. These techniques help to reduce image noise and artifacts seen at low radiation-dose CT. 10. Assign a specific personnel team to be responsible for protocol review and dose monitoring. This team should perform periodic and frequent review of CT protocols and also should review the outliers from the dose-monitoring techniques (including the ACR Dose Index Registry initiative).

13 RG Volume 35 Number 6 Kalra et al 1717 CT Protocols for Proper Dose Use Creation of optimally tailored CT protocols is a collaborative team effort among radiologists, CT technologists, medical physicists, and CT vendor specialists. Once the appropriateness of CT has been determined, protocols should be tailored on the basis of patient age, body region, and clinical indication. Timely review of scanning protocols and associated radiation doses is essential to ensure that protocols are correctly configured and updated on a regular basis. Separate protocols often are needed for children and adults. Head Although magnetic resonance (MR) imaging is ideal for many neurologic indications, head CT remains a commonly performed examination in patients suspected of having head trauma and acute stroke. Many routine nonenhanced head CT examinations are performed at 120 kv by using a step-and-shoot or nonhelical mode of acquisition with a 1-second gantry rotation time and an angulated gantry to avoid radiation to the eyes. However, many institutions are increasingly using helical acquisitions for head CT to increase scanning speed and to enable high-quality multiplanar reformations. To decrease beam-hardening artifacts, an overlapping pitch (<1) is typically used in helical head CT. Head CT can be performed at a fixed tube current or with AEC techniques. IR techniques reduce image noise and thus permit radiation dose reductions. Users should ensure that most examinations are performed at less than the DRL value of a CTDI vol of 75 mgy in adults and 35 mgy in children (37). CT protocols for evaluation of paranasal sinuses can be performed at a substantially lower dose than that used at routine head CT, because the primary structures of interest are of inherently high tissue contrast between the bones and air in the sinuses (59). CT angiography of the head and neck is another common examination that preferably should be performed at low voltage ( kv) to reduce radiation dose while improving contrast enhancement. As a rule, CT perfusion of the head should be performed at 80 kv to reduce radiation dose associated with repetitive scanning while improving detection of contrast enhancement (60). Chest The presence of high inherent tissue contrast and low x-ray attenuation in the predominantly airfilled lungs allow most chest abnormalities to be assessed at a much lower radiation dose (at least 30% 50% lower) than the dose used in abdominal scanning. Routine chest CT examinations with intravenous contrast material in small- to average-sized patients can be performed at kv to reduce radiation dose. To minimize motion artifacts, a fast gantry rotation time and a nonoverlapping pitch with wide beam collimation are preferred. AEC techniques should be used to tailor radiation dose according to patient size and large regional variations in tissue attenuation (ie, dose is typically much higher for imaging the shoulders and upper abdomen than for imaging the midthorax). Special attention should be paid to limit the inferior extent of the scanning range, particularly in young patients without a history of cancer. As in other body regions, IR is useful to reduce noise and to permit further dose reductions by lowering tube current, voltage, or both (49). Separate CT protocols are recommended for follow-up of lung nodules, because nodules can be observed easily in images obtained at a fraction of the radiation dose used in routine chest CT examinations such as those for chronic cough. AEC techniques or weight-based adjustment of fixed tube current can be used for reducing radiation dose. Thus, compared with routine chest CT (eg, for cancer staging), lung nodule follow-up CT should have much lower radiation dose or CTDI vol. The scanning range for this protocol should be restricted to the lungs only. Authors of several studies have reported that radiation doses for CT follow-up of lung nodules or lung cancer screening at CT can be performed at submillisievert dose levels, especially with use of IR techniques (Figs 5, 6) (61 63). Abdomen and Pelvis Abdominal and pelvic abnormalities (eg, lesions in liver, spleen, and pancreas) often show low contrast on images, compared with high-contrast chest CT images. Excessive radiation dose reduction in the abdomen and pelvis can result in high image noise and artifacts that can affect conspicuity of some low-contrast lesions. Therefore, abdominal and pelvic protocols should be stratified carefully on the basis of clinical indications. IR techniques should be used to reduce image noise and enable lower tube current and/or voltage to reduce dose. Abdominal CT should be performed with AEC techniques to adapt radiation dose to patient size. Careful attention should be paid to the scanning range to minimize inclusion of the chest and thighs. Because the contrast of kidney stones is high relative to that of background soft tissue, abdominal CT performed exclusively for evaluation of urinary calculi can be performed at a lower radiation dose by using a dedicated protocol (if there is not a differential diagnosis encompassing other lower-contrast abdominal structures) (Fig 3). Although most kidney stones can be seen on

14 1718 October Special Issue 2015 radiographics.rsna.org Figure 7. (a) Coronal CT image in a 9-yearold girl (weight, 43 kg) with soft-tissue sarcoma of the lower extremities was acquired at 100 kv and 25 mas (AEC) for evaluation of intrathoracic metastatic disease. (b) Coronal multiplanar reformatted CT image shows acceptable image quality at low radiation dose (CTDI vol, 0.4 mgy; DLP, 39 mgy cm) without evidence of intrathoracic metastatic disease. reduced-dose CT images, subtle changes of pancreatitis or ovarian abnormalities can be missed in some patients; therefore, if there is any suspicion of other confounding abnormalities in the abdomen, reduced-dose CT protocols should be used in moderation. Such dose reduction is generally achieved with a reduction in tube current through suitable modification of AEC techniques (lower image quality requirement compared with that at routine abdominal CT) (64,65). Multiphasic examinations of the abdomen should be restricted to appropriate clinical situations (1). For example, acquisition of nonenhanced images before contrast-enhanced routine abdominal and/or chest CT should be avoided. Delayed images should be acquired only when they may help in evaluation of an abnormality. Reduction of scanning range for one or more phases to the specific region of interest can help reduce radiation dose substantially. For example, at multiphasic liver CT, the arterial and delayed phases can be limited to the liver, with the entire abdomen imaged only during the portal venous phase, if necessary. In addition, for the arterial phase, the voltage can be reduced to improve contrast enhancement while reducing radiation dose. Key Features of Pediatric CT Protocols Special consideration should be given to CT protocols for children because radiation-induced cancer risks are believed to be greater in the pediatric population. Adult CT protocols should not be used in small children, and size-based adjustment of radiation dose must be followed rigidly, with appropriate adjustment of both voltage and tube current (Fig 7). Care must be taken to minimize motion artifacts in small or anxious children with appropriate calming methods and fast scanning speeds accomplished by fast gantry rotation times and nonoverlapping pitch. As they are with adult patients, CT radiation doses for children must be adjusted to the clinical indication (Fig 8). Good scan timing is crucial to avoid missed contrast boluses. Routine use of multiphasic CT protocols must be minimized in children. When possible and appropriate to the clinical indication, ultrasonography or MR imaging is preferred to eliminate ionizing radiation. Agencies and Organizations Involved with Medical Radiation Many organizations including the ACR, the American Association of Physicists in Medicine, the International Commission on Radiological Protection, the International Atomic Energy Agency, the National Council on Radiation Protection and Measurements, and the U.S. Food and Drug Administration have issued recommendations for management of CT radiation dose (6,35 37,66,67). The Image Gently campaign was created to raise awareness of radiation in the pediatric population (68). The campaign Web site provides teaching materials and optimized scanning protocols to promote safe use of ionizing radiation based modalities in children. The Image Wisely campaign ( is a joint effort of the ACR, the Radiological Society of North America, the American Association of Physicists in Medicine, and the American Society of Radiologic Technologists that advocates for appropriate use of radiation in adult patients (69). RadiologyInfo.org ( is a resource on radiation dose at CT and radiation safety for patients (70).

15 RG Volume 35 Number 6 Kalra et al 1719 Figure 8. (a, b) Images in a 13-year-old boy (weight, 45 kg) who underwent reduced-dose chest CT for evaluation of pectus excavatum deformity. Radiation dose was reduced to as low as achievable on the CT scanner with 80 kv, 4 mas, nonoverlapping pitch of 1.375:1, and detector configuration of mm (CTDI vol, 0.08 mgy; DLP, 2.5 mgy cm). Sagittal (a) and axial (b) images in bone windows reconstructed with ASiR (GE Healthcare) showed pectus excavatum deformity (Haller index, 3.2). (c) Axial CT image obtained with lung window setting demonstrates sufficient image quality for evaluation of lung abnormalities. According to the prepublication requirements of the Joint Commission issued on January 9, 2015, applicable to ambulatory care centers effective July 1, 2015, organizations must archive CTDI vol, DLP, or SSDE for every CT examination, summarized by series or anatomic area imaged (71). The prepublication document mandates annual CT review of CTDI vol, and also recommends that organizations review and analyze incidents where the CT x-ray tube output metric exceeded expected dose index ranges identified for that particular protocol. Dose Monitoring and Registries Several commercial dose-tracking software packages are available for monitoring CT dose (72). These software programs automatically extract CT dose metrics from the CT examinations and compile them in a searchable and analyzable database (24). In some, alerts can be configured for doses falling outside the expected range for the particular CT protocol. Dose tracking or monitoring can help in CT radiation dose management by allowing audit and comparison of dose indexes for different CT examinations. Practices are encouraged to participate in the ACR Dose Index Registry, which has captured CT data from more than 5 million CT examinations submitted by more than 1250 imaging facilities in North America (73,74). The ACR Dose Index Registry issues biannual reports to each contributing imaging facility in which values for CTDI vol, DLP, and SSDE (when available from centers that transfer planning radiographs along with the dose data) are compared among regional, national, and practice types on a per-protocol basis. Although the ACR did not have published protocol-specific dose reference levels at the time of preparation of this article, the dose report from the ACR Dose Index Registry allows practices to compare their doses by protocol to 1250 other sites in North America. These dose-tracking options have encouraged the radiologic community to become increasingly aware of CT radiation doses and image quality. The use of these dosetracking options is expected to increase after the Joint Commission recommendations highlighting

16 1720 October Special Issue 2015 radiographics.rsna.org the importance of tracking doses for every CT examination are published (71). Conclusion CT provides vital diagnostic information that affects patient care. Once the appropriateness of CT is established, scanning protocols should be optimized for the specific clinical indication, body habitus, anatomic area, and patient age. There are several technologic and scanning strategies available to optimize radiation dose associated with CT while maintaining diagnostic image quality. Periodic monitoring of scanning protocols and associated radiation doses with dose-tracking programs should be a part of any radiation dose management initiative. Wise and gentle use of CT with the strategies outlined in this review can help preserve expected diagnostic information while ensuring that radiation doses are as low as reasonably achievable. Disclosures of Conflicts of Interest. A.D.S. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: grant from Siemens Healthcare. Other activities: disclosed no relevant relationships. References 1. Mayo-Smith WW, Hara AK, Mahesh M, Sahani DV, Pavlicek W. How I do it: managing radiation dose in CT. Radiology 2014;273(3): Mettler FA Jr, Bhargavan M, Faulkner K, et al. Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources Radiology 2009;253(2): Fletcher JG, Kofler JM, Coburn JA, Bruining DH, McCollough CH. Perspective on radiation risk in CT imaging. Abdom Imaging 2013;38(1): Sodickson A. Strategies for reducing radiation exposure in multi-detector row CT. Radiol Clin North Am 2012;50 (1): Kalra MK, Maher MM, Toth TL, et al. Strategies for CT radiation dose optimization. Radiology 2004;230(3): McCollough CH, Chen GH, Kalender W, et al. Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT. Radiology 2012;264(2): Frush D, Denham CR, Goske MJ, et al. Radiation protection and dose monitoring in medical imaging: a journey from awareness, through accountability, ability and action but where will we arrive? J Patient Saf 2013;9(4): American Board of Radiology. Diagnostic radiology: core quality and safety study guide. /themes/abr-media/pdf/core_exam_quality_and_safety_ Syllabus_1st_ed_FINAL.pdf. Updated February 20, Accessed July 16, Bogdanich W. After stroke scans, patients face serious health risks. New York Times. July 31, Accessed October 20, Joint Commission. The Joint Commission Sentinel Event Alert: radiation risks of diagnostic imaging. Published August 24, Little MP, Wakeford R, Tawn EJ, Bouffler SD, Berrington de Gonzalez A. Risks associated with low doses and low dose rates of ionizing radiation: why linearity may be (almost) the best we can do. Radiology 2009;251(1): Tubiana M, Feinendegen LE, Yang C, Kaminski JM. The linear no-threshold relationship is inconsistent with radiation biologic and experimental data. Radiology 2009;251 (1): Brenner DJ, Doll R, Goodhead DT, et al. Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci U S A 2003;100(24): National Research Council. Health risks from exposure to low levels of ionizing radiation: BEIR VII, phase 2. Washington, DC: National Academies Press, Pearce MS, Salotti JA, Little MP, et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 2012;380(9840): Mathews JD, Forsythe AV, Brady Z, et al. Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 2013;346:f Huang WY, Muo CH, Lin CY, et al. Paediatric head CT scan and subsequent risk of malignancy and benign brain tumour: a nation-wide population-based cohort study. Br J Cancer 2014;110(9): Sodickson A. CT radiation risks coming into clearer focus. BMJ 2013;346:f Journy N, Rehel JL, Ducou Le Pointe H, et al. Are the studies on cancer risk from CT scans biased by indication? Elements of answer from a large-scale cohort study in France. Br J Cancer 2015;112(1): Krille L, Dreger S, Schindel R, et al. Risk of cancer incidence before the age of 15 years after exposure to ionising radiation from computed tomography: results from a German cohort study. Radiat Environ Biophys 2015;54(1): Boice JD Jr. Radiation epidemiology and recent paediatric computed tomography studies. Ann ICRP 2015;44(1 suppl): Huda W, Rowlett WT, Schoepf UJ. Radiation dose at cardiac computed tomography: facts and fiction. J Thorac Imaging 2010;25(3): McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J. Strategies for reducing radiation dose in CT. Radiol Clin North Am 2009;47(1): Sodickson A, Warden GI, Farkas CE, et al. Exposing exposure: automated anatomy-specific CT radiation exposure extraction for quality assurance and radiation monitoring. Radiology 2012;264(2): McCollough CH, Leng S, Yu L, Cody DD, Boone JM, McNitt-Gray MF. CT dose index and patient dose: they are not the same thing. Radiology 2011;259(2): Turner AC, Zhang D, Khatonabadi M, et al. The feasibility of patient size-corrected, scanner-independent organ dose estimates for abdominal CT exams. Med Phys 2011;38(2): AAPM task group 204. AAPM Report 204: size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. College Park, Md: American Association of Physicists in Medicine, reports/rpt_204.pdf. 28. Noferini L, Fulcheri C, Taddeucci A, Bartolini M, Gori C. Considerations on the practical application of the size-specific dose estimation (SSDE) method of AAPM Report 204. Radiol Phys Technol 2014;7(2): Brady SL, Kaufman RA. Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation. Radiology 2012;265(3): AAPM Report 220: use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT. College Park, Md: American Association of Physicists in Medicine, reports/rpt_220.pdf. 31. AAPM Report No. 96: The measurement, reporting, and management of radiation dose in CT. College Park, Md: American Association of Physicists in Medicine, Huda W, Mettler FA. Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 2011;258(1):

17 RG Volume 35 Number 6 Kalra et al National Council on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States: NCRP report no Bethesda, Md: National Council on Radiation Protection and Measurements, McCollough C, Branham T, Herlihy V, et al. Diagnostic reference levels from the ACR CT Accreditation Program. J Am Coll Radiol 2011;8(11): American College of Radiology. Computed tomography accreditation. American College of Radiology Web site. Updated November 18, Accessed March 1, National Council on Radiation Protection and Measurements. Reference levels and achievable doses in medical and dental imaging: recommendations for the United States NCRP report no Bethesda, Md: National Council on Radiation Protection and Measurements, ACR-AAPM practice guideline for diagnostic reference levels and achievable doses in medical x-ray imaging, rev Practice-Guidelines-by-Modality/~/media/ACR/Documents/ PGTS/guidelines/Reference_Levels.pdf. 38. Lee KH, Lee JM, Moon SK, et al. Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 2012;265(2): Yu L, Li H, Fletcher JG, McCollough CH. Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys 2010;37(1): American Association of Physicists in Medicine. AAPM CT lexicon version documents/ctterminologylexicon.pdf. Published April 20, Accessed April 2, Lell MM, Jost G, Korporaal JG, et al. Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography. Invest Radiol 2015;50(3): Noda Y, Kanematsu M, Goshima S, et al. Reduction of iodine load in CT imaging of pancreas acquired with low tube voltage and an adaptive statistical iterative reconstruction technique. J Comput Assist Tomogr 2014;38(5): Lira D, Padole A, Kalra MK, Singh S. Tube potential and CT radiation dose optimization. AJR Am J Roentgenol 2015;204(1):W4 W Sodickson A, Weiss M. Effects of patient size on radiation dose reduction and image quality in low-kvp CT pulmonary angiography performed with reduced IV contrast dose. Emerg Radiol 2012;19(5): Kalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology 2004;233(3): Singh S, Kalra MK, Shenoy-Bhangle AS, et al. Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 2012;263(2): Ghoshhajra BB, Engel LC, Károlyi M, et al. Cardiac computed tomography angiography with automatic tube potential selection: effects on radiation dose and image quality. J Thorac Imaging 2013;28(1): Khawaja RD, Singh S, Otrakji A, et al. Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol 2015;45(7): Singh S, Khawaja RD, Pourjabbar S, Padole A, Lira D, Kalra MK. Iterative image reconstruction and its role in cardiothoracic computed tomography. J Thorac Imaging 2013;28(6): Wallihan DB, Podberesky DJ, Sullivan J, et al. Diagnostic performance and dose comparison of filtered back projection and adaptive iterative dose reduction three-dimensional CT enterography in children and young adults. Radiology 2015;276(1): Bombiński P, Warchoł S, Brzewski M, et al. Lower-dose CT urography (CTU) with iterative reconstruction technique in children: initial experience and examination protocol. Pol J Radiol 2014;79: McKnight CD, Watcharotone K, Ibrahim M, Christodoulou E, Baer AH, Parmar HA. Adaptive statistical iterative reconstruction: reducing dose while preserving image quality in the pediatric head CT examination. Pediatr Radiol 2014;44(8): Brady SL, Moore BM, Yee BS, Kaufman RA. Pediatric CT: implementation of ASIR for substantial radiation dose reduction while maintaining pre-asir image noise. Radiology 2014;270(1): McCollough CH, Yu L, Kofler JM, et al. Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology 2015 Mar 26: [Epub ahead of print] 55. Li J, Udayasankar UK, Toth TL, Seamans J, Small WC, Kalra MK. Automatic patient centering for MDCT: effect on radiation dose. AJR Am J Roentgenol 2007;188(2): Deak PD, Langner O, Lell M, Kalender WA. Effects of adaptive section collimation on patient radiation dose in multisection spiral CT. Radiology 2009;252(1): Duan X, Wang J, Leng S, et al. Electronic noise in CT detectors: impact on image noise and artifacts. AJR Am J Roentgenol 2013;201(4):W626 W Brink JA. Dose tracking and rational examination selection for the medically-exposed population. Health Phys 2014;106(2): Singh S, Kalra MK, Thrall JH, Mahesh M. Pointers for optimizing radiation dose in head CT protocols. J Am Coll Radiol 2011;8(8): Wintermark M, Lev MH. FDA investigates the safety of brain perfusion CT. AJNR Am J Neuroradiol 2010;31(1): Khawaja RD, Singh S, Gilman M, et al. Computed tomography (CT) of the chest at less than 1 msv: an ongoing prospective clinical trial of chest CT at submillisievert radiation doses with iterative model image reconstruction and idose4 technique. J Comput Assist Tomogr 2014;38 (4): Gordic S, Morsbach F, Schmidt B, et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping. Invest Radiol 2014;49(7): Murugan VA, Kalra MK, Rehani M, Digumarthy SR. Lung cancer screening: computed tomography radiation and protocols. J Thorac Imaging 2015 Apr 4. [Epub ahead of print] 64. Moore CL, Daniels B, Ghita M, et al. Accuracy of reduceddose computed tomography for ureteral stones in emergency department patients. Ann Emerg Med 2015;65(2): e Jin DH, Lamberton GR, Broome DR, et al. Effect of reduced radiation CT protocols on the detection of renal calculi. Radiology 2010;255(1): Rehani MM. ICRP and IAEA actions on radiation protection in computed tomography. Ann ICRP 2012;41(3-4): Rumack CM ACR Presidential Address: patientfocused radiology taking charge of radiation dose. J Am Coll Radiol 2010;7(11): Goske MJ, Applegate KE, Bulas D, et al. Image Gently: progress and challenges in CT education and advocacy. Pediatr Radiol 2011;41(suppl 2): Brink JA, Amis ES Jr. Image Wisely: a campaign to increase awareness about adult radiation protection. Radiology 2010;257(3): Ellenbogen PH, Tashjian JH. RadiologyInfo: reaching out to touch patients. J Am Coll Radiol 2007;4(11): Joint Commission. The Joint Commission prepublication requirements: revised requirements for diagnostic imaging services. CAH_DiagImag_Prepub_July2015release_ pdf. Published January 9, Accessed March 30, Duong PA, Little BP. Dose tracking and dose auditing in a comprehensive computed tomography dose-reduction program. Semin Ultrasound CT MR 2014;35(4): Morin RL, Coombs LP, Chatfield MB. ACR Dose Index Registry. J Am Coll Radiol 2011;8(4): Bhargavan-Chatfield M, Morin RL. The ACR Computed Tomography Dose Index Registry: the 5 million examination update. J Am Coll Radiol 2013;10(12):

18 QUESTIONNAIRE A1(17) CT radiation INSTRUCTIONS Read through the article and answer the multiple-choice questions provided below Some questions may have more than one answer; in which case you must please mark all the correct answers CT radiation: key concepts for gentle and wise use Question 1: Would it be correct to state that whilst CT examinations represent only about 17% of radiation-based medical procedures, they contribute nearly 50% of the medical radiation dose of the population as a whole? A: YES B: NO Question 2: The BEIR VII study found that for the same x-ray exposure: A: Risks are expected to be higher for young patients and female patients B: There is a threshold dose below which there is no lasting effect C: Low levels of radiation are protective D: Doubling the radiation exposure is assumed to double the associated cancer risk E: A and C Question 3: Which CT radiation output is expressed in milligray centimeters? A: Absorbed or organ dose B: CTDI C: DLP Question 4: Patient size may be estimated from? A: The anteroposterior or lateral diameters B: The sum of the anteroposterior and lateral diameters C: The effective diameter of the patient Question 5: Which of the following are TRUE regarding effective dose? A: Effective dose is measured in sieverts B: Effective dose is a single dose value that is commonly used to compare the relative biological risks of different exposures C: Effective dose reflects the actual risk to a specific patient D: Effective dose is intended to represent the uniform whole-body equivalent dose that would be expected to cause the same overall cancer risk as the non-uniform or partial-body exposure delivered to the patient Question 6: In 2013, the CT scanner accreditation process from the ACR set the DRLs for adult and pediatric (5-year-old) abdominal CT examinations, at... and..., respectively: A: 75 mgy (16-cm CTDI phantom size) and 35 mgy (16- cm CTDI phantom size) B: 25 mgy (32-cm CTDI phantom size) and 15 mgy (16- cm CTDI phantom) Question 7: Is it TRUE that for the same clinical indication and body region, smaller patients generally require higher x-ray tube output, while larger patients can be scanned at a lower radiation dose to achieve the same image quality? A: YES B: NO Question 8: CT examinations performed to detect... require... image noise and thus... doses for accurate detection and characterization: A: High-tissue-contrast tasks / less / higher B: Low-contrast lesions / more / lower C: Low-contrast lesions / less / higher Question 9: Adjustment of which of the following products remains the most practical and most commonly used method of adapting radiation output from CT scanners? A: Fixed tube current B: Tube current C: Tube current-time D: B or C Question 10: Which of the following require, as a rule of thumb, higher tube current or tube current-time product settings? A: Smaller patients B: Larger patients C: Body regions with higher x-ray attenuation D: Body regions with lower x-ray attenuation Question 11: Tube voltage represents the peak photon energy of an x-ray energy spectrum. A change in voltage does not have as great an effect on radiation dose as a change in fixed tube current or tube current-time product. Is this statement TRUE or FALSE? A: TRUE B: FALSE

19 Question 12: A decrease in which scanning factor reduces radiation dose and substantially improves image contrast, particularly for contrast-enhanced CT examinations? A: Tube current B: Gantry rotation time C: Tube current-time product D: Tube voltage Question 13: Which type of gantry pitch and time should be used to minimize scanning times, as for patients with shortness of breath who are undergoing CT pulmonary angiography? A: Overlapping pitch with fast gantry rotation time B: Non-overlapping pitch with fast gantry rotation time C: Overlapping pitch with slow gantry rotation time D: Non-overlapping pitch with slow gantry rotation time Question 14: Which of the following statements are TRUE regarding detector configuration? A: It determines x-ray beam collimation or width B: Most multi-detector row CT scanners offer two or more detector configurations C: All scanners contain detector elements of variable thickness, such that detector configuration is tied to the desired minimum section thickness D: All the above Question 15: Which of the following is the result of off-centering a patient too close to the x-ray tube for the planning radiograph? A: Miscalculation of patient attenuation B: Increase in radiation dose without substantial gain in diagnostic image quality C: Inflated estimate of patient width Question 16: For which procedure, would you, for instance, constrain the range of allowed voltage by eliminating 140Kv as an option because of the loss of iodine enhancement at high voltage? A: Non-enhanced CT B: Contrast-enhanced CT C: CT angiography Question 17: By using an adaptive iterative dose reduction IR technique, Wallihan et al recently reported a substantial reduction in SSDE for children undergoing CT enterography. Is this statement TRUE? A: YES B: NO Question 18: Some hardware-driven technology for CT dose reduction, such as some multi-detector row CT scanners with wide area detectors, use... to improve radiation dose efficiency: A: Beam-shaping B: Bow-tie filters C: Adaptive shielding Question 19: Which one of the following, in the opinion of the authors, is the most important step toward use of the appropriate radiation dose during CT scanning? A: Performing CT scans in accordance with set protocols B: Ensuring the appropriateness of the treatment based on required clinical information C: Regularly updating, sharpening and checking the skills of the operator Question 20: Which technique can help to reduce radiation dose by decreasing the number of required phases for some CT protocols? A: Acquire thin and reconstruct thick B: Voltage adjustment C: Split-bolus contrast material injection D: Section collimation Question 21: Once the appropriateness of CT has been confirmed, protocols should be tailored based on...? A: Body region B: Age C: Clinical indication D: Gender E: Weight Question 22: Regarding head CT, users should ensure that most examinations are performed at less than the DRL value of a CTDI of mgy in adults and. mgy in children? A: 75 and 35 B: 60 and 40 C: 85 and 45 D: 90 and 55 Question 23: As a rule, CT perfusion of the head should be performed at... to reduce radiation dose associated with repetitive scanning: A: 90 kv B: 120 kv C: 100 kv D: 80 kv Question 24: Authors of several studies have reported that radiation doses for CT follow-up of lung nodules or lung cancer screening at CT can be performed at supra millisieverts dose levels, especially with use of IR techniques. Is this statement TRUE or FALSE? A: TRUE B: FALSE

20 Question 25: Regarding scans of the abdomen and pelvis, which of the following statements are TRUE? A: Excessive dose reduction can result in high image noise and artifacts that can affect conspicuity of some lowcontrast lesions B: Protocols should be stratified carefully based on clinical indications C: Abdominal scans should be performed with AEC techniques to adapt radiation dose to patient size D: Abdominal CT performed exclusively for evaluation of urinary calculi, must be performed at a higher radiation dose by using a dedicated protocol E: All of the above Question 26: Which of the following are TRUE regarding multiphasic examinations of the abdomen? A: It should be restricted to appropriate clinical situations B: Delayed images should always be acquired C: Acquisition of nonenhanced images before contrastenhanced routine abdominal and/or chest CT is advised D: Reduction of scanning range for one or more phases to the specific region of interest can help reduce radiation dose substantially E: For the arterial phase, the voltage can be reduced to improve contrast enhancement while reducing radiation dose Question 27: Radiation-induced cancer risks are believed to be greater in the pediatric population. Which techniques are preferred to eliminate ionising radiation? A: Multiphasic CT B: Ultrasonography C: MR imaging D: All the above Ten steps to wise and gentle CT protocols Question 28: Is it TRUE that most body CT examinations must be performed with AEC to reduce radiation dose? A: YES B: NO Question 29: Routine acquisition of nonenhanced images before routine chest or abdominal CT adds substantial information in most patients. Is this statement TRUE or FALSE? A: TRUE B: FALSE Question 30: Appropriate detector and section collimation should be selected based on: A: Clinical indication B: Age C: Weight D: Gender E: All of the above End

21 400 Theuns van Niekerk Street Wierda Park 0157 PO Box 71 Wierda Park Cell: Tel: /2373 /2873 Mon-Fri: 07:30-16:30 *ID / HPCSA No PERSONAL INFO (If your personal details have not changed, only complete the sections marked with an asterisk *) *FOH Number Initials &Surname *Cell Number needed for confirmation sms Employer address *Time spent on activity Hour Min Any additional comments / change of personal details ANSWER SHEET A1(17) CT radiation: key concepts for gentle and wise use Some questions may have more than one answer; in this case please select every correct answer. A B C D E A B C D E A B C D E This article is accredited for THREE CLINICAL (3) CEU s I hereby declare that the completion of this document is my own effort without any assistance. Signed: Date: SEND ANSWER SHEET TO: FAX: / OR WHATSAPP: OR (YOU WILL RECEIVE A CONFIRMATION OF RECEIPT SMS WITHIN HOURS, IF NOT RECEIVED PLEASE SEND AGAIN) PLEASE NOTE No answer sheets will be marked if your account is in arrears Please rate the article: POOR 1 FAIR 2 AVERAGE 3 GOOD 4 EXCELLENT 5 /30 = % MARKED: FAILED PASSED (70% PASS RATE) (R50 to resubmit) (IAR will be sent) MODERATED BY: CAPTURED: DATE:

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School OUTLINE 1 st Presentation: Dose optimization strategies

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center Mannudeep K. Kalra, MD Webster Center for Quality and Safety Massachusetts General Hospital Harvard Medical School

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean Doses from Cervical Spine Computed Tomography (CT) examinations in the UK John Holroyd and Sue Edyvean Why a new dose survey? Number of enquires received concerning the current NDRL Concern that could

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

CT SCAN PROTOCOL. Shoulder

CT SCAN PROTOCOL. Shoulder CT SCAN PROTOCOL Shoulder Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

CT Dose Reduction in Pediatric Patients

CT Dose Reduction in Pediatric Patients CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Radiation Dose Reduction Strategies in Coronary CT Angiography

Radiation Dose Reduction Strategies in Coronary CT Angiography Radiation Dose Reduction Strategies in Coronary CT Angiography Noor Diyana Osman, PhD noordiyana@usm.my Contents: Introduction Radiation dosimetry in CT Radiation risk associated with coronary CT angiography

More information

Patient / Organ Dose in CT

Patient / Organ Dose in CT Patient / Organ Dose in CT Patient specific and organ dose estimation H.D. Nagel Dr. HD Nagel, Science & Technology for Radiology Buchholz / Germany www.sascrad.com 1 Topics CTDI & patient dose SSDE Organ

More information

CT Radiation Risks and Dose Reduction

CT Radiation Risks and Dose Reduction CT Radiation Risks and Dose Reduction Walter L. Robinson, M.S. D.A.B.S.N.M., D.A.B.M.P., D.A.B.R. Consultant Certified Medical Radiation Health & Diagnostic Imaging Physicist Medical Radiation and Children

More information

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program CT Low Dose Lung Cancer Screening Part I Journey to LDCT LCS Program Paul Johnson, M.S., DABHP, DABR Cleveland Clinic September 26, 2015 Lung Caner is No. 1 In Cancer Related Death In The United States

More information

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy What is radiation? Radiation is energy emitted from a source, that travels through space and can penetrate matter. Listed below are two types that we are exposed to and contribute to our overall radiation

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Chief Radiographer TEI Clinical Associate 2016

Chief Radiographer TEI Clinical Associate 2016 MDCT Principles i and Applications Ε ΑGADAKOS MSc Ε. ΑGADAKOS MSc Chief Radiographer TEI Clinical Associate 2016 Aim To understand d recent technological advances in MSCT and how they can be effectively

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Aquilion ONE: Pediatric Imaging. Richard Mather, PhD. Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc.

Aquilion ONE: Pediatric Imaging. Richard Mather, PhD. Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc. Aquilion ONE: Pediatric Imaging Richard Mather, PhD Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc. The use of CT in pediatric diagnostic procedures has increased significantly

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose Ask EuroSafe Imaging Tips & Tricks CT Working Group Optimization of scan length to reduce CT radiation dose Alban Gervaise (Centre Hospitalier Universitaire Nancy, FR) Mika Kortesniemi (HUS Medical Imaging

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Minimizing CT Radiation Dose CT examinations improve health care and are an essential part

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

The radiation dose in retrospective

The radiation dose in retrospective The radiation dose in retrospective gated tdcoronary computed td tomography (CCT) Saeed AL Ahmari, Ghormallah AL Zahrani, Sumiah AL Helali, Samir AL Dulikan, Abdullah Bafagih, HibaKhashojji Prince Sultan

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

CT Guided Procedures And Interesting Cases. Stephen Kim, MD Diagnostic and Interventional Radiology

CT Guided Procedures And Interesting Cases. Stephen Kim, MD Diagnostic and Interventional Radiology CT Guided Procedures And Interesting Cases Stephen Kim, MD Diagnostic and Interventional Radiology CT guided procedure benefits Precise lesion targeting Clear image guidance for needle placement Immediate

More information

Low Dose Era in Cardiac CT

Low Dose Era in Cardiac CT Low Dose Era in Cardiac CT DIANA E. LITMANOVICH, MD Department of Radiology Beth Israel Deaconess Medical Center Harvard Medical School Disclosures Neither I nor my immediate family members have a financial

More information

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

A more accurate method to estimate patient dose during body CT examinations with tube current modulation A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga

More information

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Alice B. Smith, MD

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

CT dose survey data acquisition form

CT dose survey data acquisition form CT dose survey data acquisition form CT Protocol page CT Head (acute stroke) C-spine (fracture) Chest (lung cancer) Chest High-Res. (interstitial lung disease) CTA (blood vessels) CTPA (PE) Abdomen (liver

More information

Managing Patient Dose in Computed Tomography (CT)

Managing Patient Dose in Computed Tomography (CT) Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Debra Pennington, MD Director of Imaging Dell Children s Medical Center Debra Pennington, MD Director of Imaging Dell Children s Medical Center 1 Gray (Gy) is 1 J of radiation energy/ 1 kg matter (physical quantity absorbed dose) Diagnostic imaging doses in mgy (.001 Gy)

More information

Radiation dose reduction in computed tomography: techniques and future perspective

Radiation dose reduction in computed tomography: techniques and future perspective REVIEW Radiation dose reduction in computed tomography: techniques and future perspective Despite universal consensus that computed tomography (CT) overwhelmingly benefits patients when used for appropriate

More information

Preparing for Medical Physics Components of the ABR Core Examination

Preparing for Medical Physics Components of the ABR Core Examination Preparing for Medical Physics Components of the ABR Core Examination The ABR core examination for radiologists contains material on medical physics. This content is based on the medical physics that is

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016 Computed Tomography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled with selected clinical experiences,

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Radiation Dose in Pediatric Imaging

Radiation Dose in Pediatric Imaging Radiation Dose in Pediatric Imaging A Brief History of Radiology Dose: Why Does It Matter? Measuring Exposure and Dose Deterministic Effects Stochastic Effects Common Exams: What is the Risk? Reducing

More information

Medical Diagnostic Imaging

Medical Diagnostic Imaging Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features

More information

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Re: PSM-044-10 of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report Dear members of the National Quality Forum Board: As Professors of Radiologic Physics (at Mayo Clinic

More information

The Computed Tomography Examination

The Computed Tomography Examination CONTENT SPECIFICATIONS The Computed Tomography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Computed Tomography Examination is to assess the knowledge and cognitive

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

Low-Dose CT: Clinical Studies & the Radiologist Perspective

Low-Dose CT: Clinical Studies & the Radiologist Perspective Low-Dose CT: Clinical Studies & the Radiologist Perspective RD-ASiR RD-MBIR SD-FBP RD=0.35 msv (80% dose reduction) Perry J. Pickhardt, MD UW School of Medicine & Public Health Low-Dose CT: Clinical Overview

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters

Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters Medical Physics and Informatics Review Maldjian and Goldman Reducing Radiation Dose in Body CT Medical Physics and Informatics Review FOCUS ON: Pierre D. Maldjian 1 Alice R. Goldman Maldjian PD, Goldman

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

Low-dose and High-resolution Cardiac Imaging with Revolution CT

Low-dose and High-resolution Cardiac Imaging with Revolution CT GE Healthcare Case study Low-dose and High-resolution Cardiac Imaging with Revolution CT Prof. Philipp A. Kaufmann, M.D. Ronny R. Buechel, M.D. Fran Mikulicic, M.D. Dominik C. Benz, M.D. University of

More information

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT protocols Tips Leukaemia / lymphoma: ~ 35% acute lymphoblastic

More information

Clinical Image Gallery Next Generation Volume 1

Clinical Image Gallery Next Generation Volume 1 Clinical Image Gallery Next Generation Volume 1 Dr. Russell Bull Royal Bournemouth Hospital, Bournemouth, United Kingdom After long experience with the first generation, a next generation Aquilion ONE

More information

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement 3 rd Annual Imaging & Physiology Summit November 20-21, 21, 2009 Seoul, Korea Wm. Guy Weigold, MD, FACC Cardiovascular

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC)

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC) EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC) BY KOFI OKYERE AKYEA-LARBI 10442894 THIS THESIS IS SUBMITTED TO THE MEDICAL

More information

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest (Non-EKG Gated) CTA Thoracic Aorta = CTA Chest Reviewed By: Dan Verdini, MD, Rachael Edwards, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the ALARA principle, TRA

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT

SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Journal of Cardiovascular Computed Tomography (2011) 5, 198 224 Guidelines SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT Sandra S. Halliburton, PhD a, *, Suhny

More information

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Radiation Exposure in Pregnancy John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Illustrative Clinical Scenario 32 year old female 34 weeks pregnant with recent onset shortness of breath and central chest pain

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

Dose-equivalent equivalent = absorbed

Dose-equivalent equivalent = absorbed UCSF General Surgery 2010 Radiation Risks of Diagnostic Radiology in Trauma Robert A. Izenberg, M.D., FACS University of California, San Francisco San Francisco General Hospital Context Increasingly liberal

More information

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Cardiac CT - Coronary Calcium Basics Workshop II (Basic) Cardiac CT - Coronary Calcium Basics Workshop II (Basic) J. Jeffrey Carr, MD, MSCE Dept. of Radiology & Public Health Sciences Wake Forest University School of Medicine Winston-Salem, NC USA No significant

More information

Managing the imaging dose during image-guided radiation therapy

Managing the imaging dose during image-guided radiation therapy Managing the imaging dose during image-guided radiation therapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Richmond VA Imaging during radiotherapy Radiographic

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Tracking Doses in the Pediatric Population

Tracking Doses in the Pediatric Population Tracking Doses in the Pediatric Population Frederic H. Fahey DSc Boston Children s Hospital Harvard Medical School frederic.fahey@childrens.harvard.edu Disclosures Sadly, none that pay me any money! SNMMI

More information

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear Sciences 2016 Seminar Jerrold Bushberg Ph.D., DABMP,

More information

Original Article Thoracic Imaging

Original Article Thoracic Imaging Original Article Thoracic Imaging https://doi.org/10.3348/kjr.2018.19.6.1179 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2018;19(6):1179-1186 Size-Specific Dose Estimation In the Korean Lung Cancer

More information

Breast Tomosynthesis. What is breast tomosynthesis?

Breast Tomosynthesis. What is breast tomosynthesis? Scan for mobile link. Breast Tomosynthesis Breast tomosynthesis is an advanced form of mammography, a specific type of breast imaging that uses low-dose x-rays to detect cancer early when it is most treatable.

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? Typical PET Image Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? PET/CT Oncology Imaging Anatometabolic fusion images are useful in the management

More information

How to Develop CT Protocols for Children

How to Develop CT Protocols for Children How to Develop CT Protocols for Children Introduction Prior to 2001 the vast majority of CT imaging of children was conducted using the same or similar techniques used for adult imaging. In 2001, several

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SUNDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SUNDAY CURRICULUM A. INJECTION TECHNIQUES a. Drug administration routes i. Oral ii. Buccal iii. Topical iv. Rectal v. Parenteral b. Injection techniques i. Hand ii. Mechanical injector c. Injection parameters i. Catheters

More information

Radiation related cancer risk & benefit/risk assessment for screening procedures

Radiation related cancer risk & benefit/risk assessment for screening procedures WHO Workshop on Justification of CT for IHA 15-17 Oct 2014 Radiation related cancer risk & benefit/risk assessment for screening procedures Elke A. Nekolla BfS Federal Office for Radiation Protection Radiation

More information

Liver Perfusion Analysis New Frontiers in Dynamic Volume Imaging. Case Study Brochure Chang Gung Memorial Hospital.

Liver Perfusion Analysis New Frontiers in Dynamic Volume Imaging. Case Study Brochure Chang Gung Memorial Hospital. New Frontiers in Dynamic Volume Imaging dynamic volume CT Case Study Brochure Chang Gung Memorial Hospital http://www.toshibamedicalsystems.com Toshiba Medical Systems Corporation 2010-2011. All rights

More information

With increasing use of computed tomography (CT) in modern medicine, concerns have arisen regarding increasing radiation dose to the community from med

With increasing use of computed tomography (CT) in modern medicine, concerns have arisen regarding increasing radiation dose to the community from med Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. ORIGINAL RESEARCH

More information

THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING

THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING White paper THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING This paper discusses why radiation exposure in medical imaging is such a hot topic, how a dose management

More information

The Increasing Use of CT and Its Risks

The Increasing Use of CT and Its Risks STUDENT SCOPE The Increasing Use of CT and Its Risks Matthew Voress is a radiography student at Owens Community College in Toledo, Ohio. This article was awarded first prize in the Ohio Society of Radiologic

More information

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate) Overview Ga-68 dotatate binds to somatostatin receptors, with highest affinity for subtype 2 receptors (sstr2). It binds to cells that express

More information

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017 EXAMINATION CONTENT SPECIFICATIONS Computed Tomography The purpose of the computed tomography examination is to assess the knowledge and cognitive skills underlying the intelligent performance of the tasks

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System Varian Edge Experience Jinkoo Kim, Ph.D Henry Ford Health System Disclosures I participate in research funded by Varian Medical Systems. Outline of Presentation Review advanced imaging in Varian Edge Linear

More information