Assessment of the Use of a Saline Chaser to Reduce the Volume of Contrast Medium in Abdominal CT

Size: px
Start display at page:

Download "Assessment of the Use of a Saline Chaser to Reduce the Volume of Contrast Medium in Abdominal CT"

Transcription

1 Saline Chaser in bdominal CT bdominal Imaging Original Research C D E M N E U T R Y L I M C I G O F I N G Florence Orlandini Stéphanie Boini Sandrine Iochum-Duchamps Toufik Batch Xavier Zhu lain Blum Orlandini F, Boini S, Iochum-Duchamps S, Batch T, Zhu X, Blum Keywords: abdominal CT, aorta, contrast media, liver, saline chaser DOI:0.4/JR Received June 5, 004; accepted after revision pril6,005. Service d Imagerie Guilloz, University Hospital of Nancy, 9, ave. Marechal de Lattre de Tassigny, Nancy 54 05, France. ddress correspondence to. Blum (a.blum@chu-nancy.fr). Service d Epidémiologie et évaluation cliniques, Centre d Epidémiologie Clinique, Hôpital Marin, University Hospital of Nancy, Nancy 54 05, France. JR 006; 87: X/06/87 5 merican Roentgen Ray Society ssessment of the Use of a Saline Chaser to Reduce the Volume of Contrast Medium in bdominal CT OBJECTIVE. The purpose of this study was to investigate, first, the effect on enhancement of the liver and aorta during abdominal CT of the use of saline solution as a partial substitute for or in addition to contrast medium when the dose of medium is determined by the patient s body weight and, second, whether use of a saline chaser allows a decrease in the dose of contrast medium to less than.5 ml/kg. SUBJECTS ND METHODS. We enrolled 407 patients undergoing abdominal exploration on MDCT, including an early arterial phase and a portal phase. received contrast medium at a dose of.5 ml/kg. received contrast medium at a dose of.5 ml/kg less 0 ml of iodine followed by 0 ml of saline solution. received the full.5 ml/kg dose of medium followed by a 0-mL saline chaser. ttenuation values were obtained from the aorta in the arterial phase and from the liver in the portal phase. RESULTS. The groups were comparable in mean body weight and heart rate. None of the differences between them in aortic enhancement in the early arterial phase were statistically significant (group, 06 ± H; group, 04 ± H; group, 09 ± 4 H). There was a negative correlation between weight and aortic enhancement (r = 0.4, p < 0.000) and a positive correlation between weight and hepatic enhancement (r = 0.9, p < 0.000). significant reduction (p = 0.000) in hepatic enhancement was observed in group (group, 5 H; group, 46 H; group, 54 H). Hepatic enhancement greater than 45 H was observed in all groups except for the group subgroup of patients weighing less than 55 kg. CONCLUSION. In MDCT, reducing the amount of contrast medium does not affect aortic enhancement in the early arterial phase but decreases hepatic enhancement at the portal phase. The saline chaser technique had no influence on the phases studied. Greater than.5 ml/kg reduction in the dose of contrast medium followed by saline flushing may not be advisable for liver CT in low- and medium-weight patients. bdominal CT is crucial in many clinical situations, such as detection of metastatic lesions of the liver and hypervascular liver neoplasms. This technique, however, has the disadvantages of exposing the patient to radiation and requiring administration of iodinated contrast medium. Many studies have addressed the question of how best to improve the image quality and diagnostic accuracy of CT. Imaging overkill may be tempting under certain circumstances but cannot be considered a serious option because of the risk of contrast-induced nephropathy, particularly among elderly patients. In addition to increasing safety, use of low doses of contrast medium may result in considerable cost savings. In a 00 study, Dorio et al. [] found that use of a saline chaser (also known as saline flushing) may allow a reduction in the amount of contrast medium needed in abdominal imaging examinations. Those investigators compared two fixed-dose injection protocols, one with 50 ml of contrast material and one with 00 ml of contrast material followed by a 50-mL saline chaser. Despite the reduced volume of contrast medium, there were no clinically significant differences in hepatic enhancement between the two protocols. dministration of a fixed amount of contrast medium can have limitations. volume insufficient for enhancement of the liver in a large patient may cause contrast medium induced nephropathy in someone thin and frail. It has therefore become usual practice to consider the patient s weight in decisions about the amount of medium to use [ 4]. The objective of this study was to explore the effects of using a saline JR:87, ugust 006 5

2 chaser in abdominal CT when the dose of contrast medium is based on the patient s weight. nother objective was to evaluate the use of saline solution when doses of contrast medium are reduced to less than.5 ml/kg. Subjects and Methods Patient Population and Protocol Four hundred seven patients (08 men and 99 women) consecutively referred for abdominal CT to include liver and vascular studies were enrolled to receive IV contrast medium according to one of the following three protocols (all injected agents were delivered at a rate of ml/s): group (n = 5), iobitridol (Xenetix 50, Guerbet).5 ml/kg; group (n = 5), iobitridol.5 ml/kg with 0 ml (7 g) of iodine replaced by 0 ml of saline chaser; group (n = ), iobitridol.5 ml/kg followed by 0 ml of saline chaser. The study was conducted in three - week periods during which each of the test protocols was used. Subjects in each group were subdivided into three categories by body weight: less than 55 kg, kg, and 80 kg or more. Twenty patients who for medical reasons underwent a repeat examination or weeks after the first were included in two of the groups. Exclusion criteria were age less than 8 years, known or suspected hypersensitivity to iodinated contrast medium, severe renal impairment (serum creatinine value > 0 µmol/l), liver cirrhosis, and compromised venous access. Image cquisition Contrast-enhanced MDCT (Sensation 6, Siemens Medical Solutions) was performed. Each examination included one acquisition of unenhanced images from the hepatic dome to the pelvis with the following parameters: detector row configuration, 6.5 mm; pitch, (table feed per gantry rotation, 4 mm/s); slice thickness, 8 mm at 0-mm intervals; 0 kv; and product of tube current and exposure time in milliampere-seconds (ms), equal to the patient s body weight. Early arterial phase acquisition with an automated bolus-tracking device (Carebolus, Siemens) was initialized when enhancement in the aorta exceeded 00 H. cquisition lasted approximately 6 seconds and was performed with a 6.5 mm detector row configuration and a pitch of (table feed per gantry rotation, 4 mm/s; gantry rotation time, 0.5 seconds). Images were reconstructed with a 5- mm slice thickness at 7-mm intervals. Forty seconds after the end of the first acquisition, data were acquired for the entire abdomen in the portal phase mm or 6.5 mm detector row configuration was chosen with a pitch of. Slice thickness was 5 mm at 7-mm intervals. Other slices were obtained for examination and interpretation elsewhere. In all cases, 0 kv was used; the product of tube current and exposure time in milliampere-seconds was equal to two to three times the body weight. Contrast medium was injected through an 8- gauge IV catheter into an antecubital vein. t the beginning of the study, the saline chaser technique required two interconnected power injectors, each holding a 00-mL syringe as described by Haage and colleagues [5]. The two long tubes on the injectors were connected in line with a nonreturn valve and then with a Y-adapter leading to the IV catheter. To save time, dual-head injectors (Stellant, Medrad) were used when they became available toward the end of the study. Reports in the literature indicate the volume of chaser saline solution used varies from 0 to 50 ml, and there is no consensus on the optimal amount. Bae et al. [6] estimated that the dead space caused by the brachial vein is approximately 40 ml in a 70-kg patient. Because we believe dead space is minimally influenced by the patient s weight, a fixed dose of 0 ml was adopted to simplify the procedure and make it reproducible. Data cquisition For each of the 407 scans, one radiologist measured arterial and hepatic enhancement after identification of consistent regions of interest with areas ranging from 5 to 0 mm. Measurements first were obtained from unenhanced images of the aorta and liver. In the arterial phase, circular regions of interest were identified at the level of the celiac trunk, the superior mesenteric artery, and the iliac bifurcation. In the portal phase, measurements free of artifacts and enhanced vessels were obtained in the hepatic parenchyma and the aorta. Statistical nalysis Comparability between groups was assessed with tests based on the chi-square test and Student s t test for categorical and continuous variables, respectively (level of type error, 5%). Quantitative data obtained with the various protocols were compared through one-way analysis of variance. Bonferroni adjustment was used for multiple comparisons. difference was considered statistically significant at p < Pearson s correlation coefficient was calculated to test the hypothesis that aortic enhancement and hepatic enhancement are related to patient weight. nalysis of variance was used to determine the effect of weight (< 55 kg, kg, and 80 kg) on aortic and hepatic enhancement. Statistical analysis was performed with SS system software (release 8., SS Institute [7]). Results Characteristics of the patient population are shown in Table. lthough they had different numbers of patients, the groups were TBLE : Patient Population Characteristic No. of patients 5 5 ge (y) Mean a Range % CI Body weight (kg) Mean Range % CI Heart rate (beats/min) Mean Range % CI Sex Male a Female a Note No significant differences in weight or heart rate were observed between groups. CI = confidence interval. a Compared with the other two groups, group patients were predominantly female (p = 0.0) and older (p = 0.0). comparable in terms of mean weight and heart rate. included a greater proportion of women than did groups and (p = 0.0), and the mean age was greater (p = 0.0). The findings for age and sex were adjusted to account for these differences. Contrast enhancement data for the three groups are summarized in Tables 5 and illustrated in Figures and. The magnitudes of observed enhancement were comparable in the aorta during the early arterial phase: group, 06 H; group, 04 H; and group, 09 H (Table ). Differences between groups did not reach statistical significance. negative correlation was observed between patient weight and aortic enhancement (r = 0.4, p < 0.000) (Table ). verage hepatic enhancement in the three groups was 5 H, 46 H, and 54 H. statistically significant reduction in hepatic enhancement was observed in group compared with groups and (6.8 H, p = 0.00; 7.8 H, p < 0.00). When doses of contrast medium were similar (groups and ), hepatic enhancement was unaffected by the saline chaser (Table 4). Patient weight and hepatic enhancement were positively correlated (r = 0.9, p < 0.000). Hepatic enhancement was greater than 45 H in all patient categories 5 JR:87, ugust 006

3 Saline Chaser in bdominal CT TBLE : ortic Enhancement (H) in the Early rterial Phase Value Mean aortic enhancement 95% CI Note None of the differences observed was statistically significant. CI = confidence interval. TBLE : ortic Enhancement (H) in the Early rterial Phase by Weight Class Patient Weight (kg) ll s < ± 7 ± 9 40 ± 8 8 ± ± 4 09 ± 4 ± 5 ± ± 6 7 ± 7 84 ± 7 78 ± 4 Note Values are mean ± SD. nalysis of variance showed all groups data had significant differences between weight categories (p < 0.000). TBLE 4: Hepatic Enhancement (H) in the Portal Phase Value Mean hepatic enhancement 95% CI Note No difference was observed between groups receiving.5 ml of iodine, regardless of use of saline flush. CI = confidence interval. TBLE 5: Hepatic Enhancement (H) in the Portal Phase by Weight Class Patient Weight (kg) ll s < ± 4 8 ± 4 47 ± 44 ± ± 5 45 ± 8 5 ± 50 ± ± 5 ± 60 ± 55 ± Note Values are mean ± SD. nalysis of variance showed hepatic enhancement superior to 45 H was observed in all groups other than the low-weight subgroup of group. other than group subjects weighing less than 55 kg. The only significant difference in hepatic enhancement was an increase among group patients weighing kg compared with the same weight category in group (p = 0.0) (Table 5). Discussion The advent of MDCT allows increased z- axis coverage with shorter scanning duration. Improved understanding of the phases of enhancement necessitates refinement of injection acquisition techniques [8]. shorter duration of injection implies a need for lower doses of contrast medium and consequent reductions in cost and risk of nephropathy [9, 0]. The total amount of contrast medium, however, must be sufficient to produce acceptable aortic and hepatic enhancement throughout acquisition. Consensus has not been reached about the ideal contrast agent or the optimal dose and rate of injection. Issues concerning the choice of monophasic or biphasic injection and scan delay after injection also remain to be fully addressed. Some authors have advocated use of a saline flush to improve vessel and parenchymal enhancement achieved with low doses of contrast medium []. Hopper et al. [] looked at flushing techniques as a way of reducing the dose of contrast medium in thoracic CT. Their hypothesis was that medium remaining in the brachiocephalic vein and the right-heart cavities provided few data and that saline pushing would reduce the transit time of the column of medium and decrease the amount of iodine in these so-called dead spaces. 75-mL dose of contrast medium pushed with saline solution was shown to enhance the thoracic vasculature to a degree equivalent to that obtained with 5 ml of contrast material alone and to cause significantly fewer beam-hardening artifacts. To the best of our knowledge, there have been few studies of the value of saline flushing in abdominal CT. Dorio and colleagues [] evaluated 86 hypoattenuating metastatic lesions of the liver in 6 patients who underwent CT in two sessions separated by a mean of 85 days. t the first session, 50 ml of contrast material (iohexol; Omnipaque 00, Nycomed) was administered at a rate varying from.0 to.0 ml/s depending on venous access. t the second session, 00 ml of contrast medium was followed by a 50-mL saline chaser. Liver and tumor attenuation was slightly better with the 50-mL dose, but the difference was of questionable clinical significance (95 H vs 89 H for the liver). Schoellnast and colleagues [] conducted a study in which 4 patients (average weight, 65 kg) with malignant tumors or inflammatory disease underwent two successive sessions of abdominal MDCT. The subjects received 00 ml of nonionic contrast medium (concentration, 00 mg/ml) either alone or pushed with 0 ml of saline solution through a double-syringe power injector at a rate of.5 ml/s. Mean enhancement of the liver improved 9 ± 9 H when saline solution was used. Itoh et al. [] studied the effects of a 5% dextrose flush on aortic, portal, and hepatic enhancement in 80 patients given. ml/kg of contrast medium (concentration, 00 mg/ml) in four different protocols. No statistically significant difference in hepatic enhancement in the portal phase was observed between protocols with or without 5% dextrose. Our results contradict those of Dorio et al. [] and Schoellnast et al. [] but are in accord with those of Itoh et al. []. Our data do not show that bolus flushing improves hepatic enhancement in the portal phase or allows significant reduction in the amount of iodine administered. Numerous factors, such as dose of contrast medium, injection rate, venous site of injection, and the liver disease involved, may account for the discrepancies. Dorio et al. and Schoellnast et al. used a fixed dose of contrast medium, whereas we determined the dose according to the patient s body weight. Our results suggest that hepatic enhancement depends on the total dose of contrast medium and not on whether a saline chaser is administered. In the portal phase, all the contrast material is available to contribute to hepatic enhancement because the dead spaces of the right-heart cavities and the brachiocephalic veins have already been washed out. One of the aims of this study was to determine whether the use of saline solution allows reduction in doses of contrast medium. Some investigators have devised strategies to lower iodine doses, but their findings have been contradictory. Most authors consider maximal hepatic enhancement greater than 50 H to be adequate. Bree and colleagues [4] found that 5 ml of iohexol 00 (4.7 g iodine) resulted in quantitatively optimal hepatic enhancement but that qualitatively optimal enhancement required 50 ml (45 g iodine). Baker et al. [5] found no significant difference in hepatic enhancement in a 68-patient study in which one group received 50 ml of iopamidol 00 (45 g iodine) and another received 5 ml of ioversol 0 (40 g iodine). Blumeke et al. [6] extrapolated from data on rabbits and concluded that 49 mg/kg is the lowest dose with which to produce a 0-H increase in hepatic enhancement. This figure is close to the 457 mg/kg used by Brink et al. []. In a large, clinical multicenter trial, Megibow and colleagues [] established that a JR:87, ugust 006 5

4 dose of 450 mg/kg (.5 ml/kg of iopromide at a concentration of 00 mg/ml) produces acceptable scans in most patients. The effect of patient weight has also been noted in previous studies. Brink s group [] found that in heavy patients an iodine dose of 8 g was required for adequate enhancement, whereas in thin patients doses as low as 6 g might be sufficient. In a prospective randomized study by Yamashita et al. [7], hepatic parenchymal enhancement at visual analysis was graded as Fig. rterial acquisition for 65-year-old woman in groups and. No significant differences were observed between enhancement of celiac trunk and hepatic artery.,, injection of.5 ml/kg of contrast medium. B,, injection of.5 ml/kg of contrast medium followed by 0 ml of saline flush at ml/s. Fig. Portal acquisition for 65-year-old woman in groups and.,, injection of.5 ml/kg of contrast medium. B,, injection of.5 ml/kg of contrast medium followed by 0 ml of saline solution at ml/s. Hepatic enhancement is similar to that in. good or excellent in 64% of patients receiving.5 ml/kg, 85% of those receiving ml/kg, 94% of those receiving.5 ml/kg, and 65% of those receiving a fixed dose. We chose a basic dose of.5 ml/kg because it is close to the minimum recommended by Megibow and colleagues []. fixed dose reduction of 0 ml for all patients in group was adopted because it was simple and reproducible. Moreover, this amount was considered a reasonable approximation of the dead space in the peripheral vein that does not contribute to aortic or hepatic enhancement. Because it corresponds to the vascular compartment, we considered it unlikely this amount would be significantly affected by body weight. These hypotheses were not confirmed because hepatic enhancement apparently was unaffected by flushing. The dose used for thin patients therefore is clearly too low. That reductions in contrast medium were not made on the basis of weight is a poten- B B 54 JR:87, ugust 006

5 Saline Chaser in bdominal CT tial limitation, because thin patients were disproportionately sensitive to dose lowering. The observation of less enhancement in thin patients in all groups is in accord with the finding by Megibow et al. [] that the larger percentage of unacceptable enhancement was found in patients in the lowweight group. We observed that the relation between a patient s weight and hepatic enhancement was not linear and that the highest levels of hepatic enhancement occurred in the highest weight category, even in groups and. The reason for this effect remains hypothetical. It may be that in large patients, the ratio between the diffusion space of contrast medium (i.e., interstitial and blood pools) and the intracellular pool is lower. It is interesting that hepatic enhancement greater than 45 H was seen in all subjects except for the thin and medium-sized patients in group (< 55 kg, hepatic enhancement, 8 H; kg, hepatic enhancement, 45 H). Even if it is agreed that maximal hepatic enhancement greater than 50 H produces 00% acceptable scans, important individual variability occurs in acceptability levels. Thus in group, scans were acceptable for large patients. Therefore, if hepatic enhancement alone is considered, one might use a lower dose of contrast medium for these patients. This practice, however, is not acceptable for vascular structures because aortic enhancement is lower in large patients. With regard to aortic enhancement, attenuation values remained high even when the dose of contrast medium was reduced. This finding is consistent with recent evidence that MDCT allows dramatic reduction in contrast medium in vascular studies [8]. In our study, there was no difference among the three groups. This finding is in agreement with the finding by Itoh et al. [] of no significant difference in aortic enhancement in the early arterial phase between four injection techniques. The lack of discrepancy between groups can probably be attributed to assessment of aortic enhancement before the arterial peak, which is known to occur (approximately) at the end of continuous injection after a continuous increase in arterial opacification due to recirculation effects []. With MDCT, acquisition time is very short. Because it lasted 6 seconds, the arterial phase in this study was completed before the end of injection in many patients. bsence of late arterial data was a limitation of the investigation. The effect on aortic enhancement of the volume injected and the chaser was probably greater in the late arterial phase, as suggested by the results of Itoh et al. []. In their study, a statistically significant increase in aortic enhancement was observed in the late arterial phase and corresponded to the aortic peak of enhancement, when saline flush was used. In this study, multiphasic acquisitions with early and late arterial phases were not clinically justified. In summary, reducing the volume of contrast medium in this context leads to a reduction in hepatic enhancement not compensated by saline flushing. In the portal phase, it is unlikely that contrast medium in the arm vein will be available for saline flushing. However, if most examinations are to be acceptable, greater than.5 ml/kg reduction in contrast bolus may not be advisable for scanning of the livers of thin and medium-sized patients. With regard to aortic enhancement in the early arterial phase, a low dose of contrast medium is sufficient to produce the same enhancement as.5 ml/kg with or without a bolus chaser. cknowledgments We thank Jean Bérard for his assistance and William Francis for reviewing the manuscript. References. Dorio P, Lee FT, Henseler K, et al. Using a saline chaser to decrease contrast media in abdominal CT. JR 00; 80: Megibow J, Jacob G, Heiken JP, et al. Quantitative and qualitative evaluation of volume of low osmolality contrast medium needed for routine helical abdominal CT. JR 00; 76: Brink J, Heiken J, Forman H, Sagel S, Molina P, Brown P. Hepatic spiral CT: reduction of dose of intravenous contrast material. Radiology 995; 97: Bonaldi VM, Garcia P, Bret PM. Parameters influencing hepatic parenchyma enhancement in X-ray computed tomography after intravenous injection of iodinated contrast medium [in French]. J Radiol 997; 78: Haage P, Schmitz-Rode T, Hubner D, Piroth W, Gunther RW. Reduction of contrast material dose and artifacts by a saline flush using a double power injector in helical CT of the thorax. JR 000; 74: Bae KT, Heiken JP, Brink J. ortic and hepatic contrast medium enhancement at CT. Part. Prediction with a computer model. Radiology 998; 07: SS Institute. SS/STT User s guide, release 8.. Cary, NC: SS Institute, Fleishman D. Use of high concentration contrast media: principles and rationale. Eur J Radiol 00; 45:S88 S9 9. Katzberg R. Urography into the st century: new contrast media, renal handling, imaging characteristics, and nephrotoxicity. Radiology 997; 04:97 0. Lautin E, Freeman NJ, Schoenfeld H, et al. Radiocontrast-associated renal dysfunction: a comparison of lower-osmolality and conventional high-osmolality contrast media. JR 99; 57: Hopper KD, Mosher TJ, Kasales CJ, TenHave TR, Tully D, Weaver JS. Thoracic spiral CT: delivery of contrast material pushed with injectable saline solution in power injector. Radiology 997; 05:69 7. Schoellnast H, Tillich M, Deutschmann H, Schaffler G, Schoellnast R, Uggowitzer M. Improvement of parenchymal and vascular enhancement using saline flush and power injection for multiple-detector-row abdominal CT. Eur Radiol 004; 4: Itoh S, Ikeda M, chiwa M, Ota T, Satake H, Ishigaki T. Multiphase contrast enhanced CT of the liver with a multislice CT scanner. Eur Radiol 00; : Bree RL, Parisky YR, Bernardino ME, Costello P, Leder R, Brown PC. Cost-effective use of low-osmolality contrast media for CT of the liver: evaluation of liver enhancement provided by various doses of iohexol. JR 994; 6: Baker ME, Beam C, Leder R, Gulliver D, Paine SS, Dunnick NR. Contrast material for combined abdominal and pelvic CT of the liver: can cost be reduced by increasing the concentration and decreasing the volume? JR 99; 60: Blumeke D, Fishman EK, nderson JH. Dose requirements for a nonionic contrast agent for spiral computed tomography of the liver in rabbits. Invest Radiol 994; 9: Yamashita Y, Komohara Y, Takahashi M, et al. bdominal helical CT: evaluation of optimal doses of intravenous contrast material a prospective randomized study. Radiology 000; 6: Rubin G. MDCT imaging of the aorta and peripheral vessels. Eur J Radiol 00; 45:S4 S49 JR:87, ugust

High Opacification of Hilar Pulmonary Vessels with a Small Amount of Nonionic Contrast Medium for General Thoracic CT: A Prospective Study

High Opacification of Hilar Pulmonary Vessels with a Small Amount of Nonionic Contrast Medium for General Thoracic CT: A Prospective Study Pierre Loubeyre 1 Isabelle Debard 2 Chantal Nemoz 3 Van André Tran Minh 2 Received April 5, 2001; accepted after revision December 28, 2001. 1 Service de Radiologie, l Hôpital Cantonal Universitaire de

More information

CT angiography techniques. Boot camp

CT angiography techniques. Boot camp CT angiography techniques Boot camp Overview Basic concepts Contrast administration arterial opacification Time scan acquisition during the arterial phase Protocol examples Helical non-gated CTA Pulmonary

More information

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

Customizing Contrast Injection for Body MDCT: Algorithmic Approach Customizing Contrast Injection for Body MDCT: Algorithmic Approach Lincoln L. Berland, M.D., F.A.C.R. University of Alabama at Birmingham Before Contrast Prep and Hydration Hydration single most important

More information

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760 Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

ORIGINAL ARTICLE INTRODUCTION. Byoung Goo Jo 1, Yun Gyu Song 1, Sang Goon Shim 2, and Young Wook Kim 3

ORIGINAL ARTICLE INTRODUCTION. Byoung Goo Jo 1, Yun Gyu Song 1, Sang Goon Shim 2, and Young Wook Kim 3 ORIGINAL ARTICLE Korean J Intern Med 2016;31:461-469 Comparison of enhancement and image quality: different iodine concentrations for liver on 128-slice multidetector computed tomography in the same chronic

More information

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography 16-MDCT Coronary Angiography Shim et al. 16-MDCT Coronary Angiography Sung Shine Shim 1 Yookyung Kim Soo Mee Lim Received December 1, 2003; accepted after revision June 1, 2004. 1 All authors: Department

More information

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method

Patient-Tailored Scan Delay for Multiphase Liver CT: Improved Scan Quality and Lesion Conspicuity With a Novel Timing Bolus Method Gastrointestinal Imaging Original Research Schneider et al. Multiphase Liver CT Gastrointestinal Imaging Original Research J. Gabriel Schneider 1 Zhen J. Wang Wilbur Wang Judy Yee Yanjun Fu Benjamin M.

More information

Portal Venous-Phase CT of the Liver in Patients without Chronic Liver Damage: Does Portal-Inflow Tracking Improve Enhancement and Image Quality?

Portal Venous-Phase CT of the Liver in Patients without Chronic Liver Damage: Does Portal-Inflow Tracking Improve Enhancement and Image Quality? Open Journal of Radiology, 2013, 3, 112-116 http://dx.doi.org/10.4236/ojrad.2013.33018 Published Online September 2013 (http://www.scirp.org/journal/ojrad) Portal Venous-Phase CT of the Liver in Patients

More information

The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications

The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications The Impact of Warmed Intravenous Contrast Material on the Bolus Geometry of Coronary CT Angiography Applications Tuncay Hazirolan, MD 1 Baris Turkbey, MD 1 Erhan Akpinar, MD 1 Murat Canyigit, MD 1 Musturay

More information

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation

Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Computed tomographic pulmonary angiography procedures: Contrast media dilution from the venous to the systemic circulation Petter Bugge Askeland Project thesis at the Faculty of Medicine UNIVERSITETET

More information

Purpose. Methods and Materials

Purpose. Methods and Materials The dose of contrast medium should be determined on the basis of estimated lean body weight (LBW) instead of the body weight--excessive amounts are currently given to females. Poster No.: C-2641 Congress:

More information

Multidetector CT of the Liver and Hepatic Neoplasms: Effect of Multiphasic Imaging on Tumor Conspicuity and Vascular Enhancement

Multidetector CT of the Liver and Hepatic Neoplasms: Effect of Multiphasic Imaging on Tumor Conspicuity and Vascular Enhancement Isaac R. Francis 1 Richard H. Cohan 1 Nancy J. McNulty 1 Joel F. Platt 1 Melvyn Korobkin 1 Achamyeleh Gebremariam 2 Kartik Ragupathi 2 Received June 3, 2002; accepted after revision October 21, 2002. Presented

More information

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Poster No.: C-3331 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Head and Neck G.

More information

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group Ask EuroSafe Imaging Tips & Tricks CT Working Group The use of bi-phase injection protocols to reduce the number of acquisition phases and radiation dose Alban Gervaise (Medical Imaging Department, HIA

More information

Hepatobiliary Imaging Original Research. rior researchers have described. the usefulness of biphasic contrast-enhanced

Hepatobiliary Imaging Original Research. rior researchers have described. the usefulness of biphasic contrast-enhanced MDCT of the Liver and Hypervascular Hepatocellular Carcinomas Hepatobiliary Imaging Original Research A C D E M N E U T R Y L I A M C A I G O F I N G MDCT of the Liver and Hypervascular Hepatocellular

More information

Cardiac CT Techniques in Neonates (and infants)

Cardiac CT Techniques in Neonates (and infants) Cardiac CT Techniques in Neonates (and infants) Siddharth P. Jadhav, MD Director, Body CT and MRI Edward B. Singleton Department of Pediatric Radiology Texas Children s Hospital Disclosures None Objectives

More information

Gastrointestinal Imaging Original Research

Gastrointestinal Imaging Original Research Gastrointestinal Imaging Original Research oursey et al. Gastrointestinal Imaging Original Research ourtney. oursey 1 Rendon. Nelson 1 Paul W. Weber 2 Laurens E. Howle 2 Eli. Nichols 3 Daniele Marin 1

More information

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate

Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper

More information

Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors

Pseudoenhancement of Renal Cysts: Influence of Lesion Size, Lesion Location, Slice Thickness, and Number of MDCT Detectors Genitourinary Imaging Original Research Tappouni et al. MDCT of Renal Cysts Genitourinary Imaging Original Research Rafel Tappouni 1 Jennifer Kissane 2 Nabeel Sarwani 1 Erik B. Lehman 1 Tappouni R, Kissane

More information

Methods and materials

Methods and materials The British Journal of Radiology, 84 (2011), 698 708 Effect of varying contrast material iodine concentration and injection technique on the conspicuity of hepatocellular carcinoma during 64-section MDCT

More information

Abdomen and Pelvis CT (1) By the end of the lecture students should be able to:

Abdomen and Pelvis CT (1) By the end of the lecture students should be able to: RAD 451 Abdomen and Pelvis CT (1) By the end of the lecture students should be able to: State the common indications for Abdomen and pelvis CT exams Identify possible contra indications for Abdomen and

More information

Liver 4 Phase CT Abdomen WO W - NC.A.V.D

Liver 4 Phase CT Abdomen WO W - NC.A.V.D Liver 4 Phase CT Abdomen WO W - NC.A.V.D Reviewed By: Rachael Edwards, MD; Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the ALARA

More information

Analysis of Changes in Attenuation of Proven Renal Cysts on Different Scanning Phases of Triphasic MDCT

Analysis of Changes in Attenuation of Proven Renal Cysts on Different Scanning Phases of Triphasic MDCT Eugene P. Chung 1 Brian R. Herts 1,2 Grant Linnell 1 Andrew C. Novick 2 Nancy Obuchowski 1,3 Deirdre M. Coll 1,4 Mark E. Baker 1 Received June 24, 2003; accepted after revision August 28, 2003. Presented

More information

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest

(Non-EKG Gated) CTA Thoracic Aorta = CTA Chest (Non-EKG Gated) CTA Thoracic Aorta = CTA Chest Reviewed By: Dan Verdini, MD, Rachael Edwards, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the ALARA principle, TRA

More information

Dual Energy CT Aortography: Can We Reduce Iodine Dose??

Dual Energy CT Aortography: Can We Reduce Iodine Dose?? Dual Energy CT Aortography: Can We Reduce Iodine Dose?? William P. Shuman MD, FACR FSCBTMR Department of Radiology University of Washington SCBTMR Annual Course Boston, October 10, 2012 Conflict of Interest

More information

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review Mannudeep K. Kalra, MD, DNB Massachusetts General Hospital Harvard Medical School Financial Disclosure This presentation

More information

Liver Perfusion Analysis New Frontiers in Dynamic Volume Imaging. Case Study Brochure Chang Gung Memorial Hospital.

Liver Perfusion Analysis New Frontiers in Dynamic Volume Imaging. Case Study Brochure Chang Gung Memorial Hospital. New Frontiers in Dynamic Volume Imaging dynamic volume CT Case Study Brochure Chang Gung Memorial Hospital http://www.toshibamedicalsystems.com Toshiba Medical Systems Corporation 2010-2011. All rights

More information

CT Contrast Protocols for Different Organ Imaging

CT Contrast Protocols for Different Organ Imaging CT Contrast Protocols for Different Organ Imaging g Paul Shreve, M.D. Advanced Radiology Services, P.C. & Spectrum Health Grand Rapids, MI, USA Correlative Imaging Council Society of Nuclear Medicine 56

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

Outcomes in the NLST. Health system infrastructure needs to implement screening

Outcomes in the NLST. Health system infrastructure needs to implement screening Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes Poster No.: C-0488 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E.-J. Kang,

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Apfaltrer et al. High-Pitch Versus Standard-Pitch CTA of the Aorta Vascular and Interventional Radiology Original Research Paul Apfaltrer 1,2 E.

More information

Combined Anatomical and Functional Imaging with Revolution * CT

Combined Anatomical and Functional Imaging with Revolution * CT GE Healthcare Case studies Combined Anatomical and Functional Imaging with Revolution * CT Jean-Louis Sablayrolles, M.D. Centre Cardiologique du Nord, Saint-Denis, France Case 1 Whole Brain Perfusion and

More information

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling Investigations and research Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling J. Wang Z. Ying V. Yao L. Ciancibello S. Premraj S. Pohlman

More information

Pictorial Essay. Multidetector CT of the Pancreas and Bile Duct System: Value of Curved Planar Reformations

Pictorial Essay. Multidetector CT of the Pancreas and Bile Duct System: Value of Curved Planar Reformations Downloaded from www.ajronline.org by 46.3.207.229 on 02/03/18 from IP address 46.3.207.229. Copyright RRS. For personal use only; all rights reserved n important feature distinguishing multidetector CT

More information

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology

Pulmonary Nodule Volumetric Measurement Variability as a Function of CT Slice Thickness and Nodule Morphology CT of Pulmonary Nodules Chest Imaging Original Research Myria Petrou 1 Leslie E. Quint 1 in Nan 2 Laurence H. aker 3 Petrou M, Quint LE, Nan, aker LH Keywords: chest, lung disease, MDCT, oncologic imaging,

More information

Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation

Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation Poster No.: C-0072 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. S. K. Teo, C. Ong, A. Ying,

More information

Right arm injection of contrast medium reduces venous artifacts in head and neck multislice spiral computed tomography angiography

Right arm injection of contrast medium reduces venous artifacts in head and neck multislice spiral computed tomography angiography European Review for Medical and Pharmacological Sciences 2015; 19: 4698-4702 Right arm injection of contrast medium reduces venous artifacts in head and neck multislice spiral computed tomography angiography

More information

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION Donald L. Renfrew, MD Radiology Associates of the Fox Valley, 333 N. Commercial Street, Suite 100, Neenah, WI 54956 6/23/2012 Radiology Quiz of the Week # 78 Page 1 CLINICAL PRESENTATION AND RADIOLOGY

More information

CT Pancreas 3 Phase CT Abdomen WO W - NC.A.V

CT Pancreas 3 Phase CT Abdomen WO W - NC.A.V CT Pancreas 3 Phase CT Abdomen WO W - NC.A.V Reviewed By: Rachael Edwards, MD; Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the

More information

Hepatocellular carcinoma (HCC) is a malignant liver neoplasm

Hepatocellular carcinoma (HCC) is a malignant liver neoplasm Diagn Interv Radiol 2011; 17:328 333 Turkish Society of Radiology 2011 ABDOMINAL IMAGING ORIGINAL ARTICLE Correlation of dynamic multidetector CT findings with pathological grades of hepatocellular carcinoma

More information

Computed tomography. Department of Radiology, University Medical School, Szeged

Computed tomography. Department of Radiology, University Medical School, Szeged Computed tomography Department of Radiology, University Medical School, Szeged voxel +1-4 +2 +5 +3 +1 0-2 pixel -2 0 +1-4 -6 +5 +2 +1 Department of Radiology, University Medical School, Szeged

More information

After the Chest X-Ray:

After the Chest X-Ray: After the Chest X-Ray: What To Do Next Alan S. Brody Professor of Radiology and Pediatrics Chief of Thoracic Imaging Cincinnati Children s Hospital Cincinnati, Ohio USA What Should We Do Next? CT scan?

More information

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis

Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis GE Healthcare Gemstone Spectral Imaging quantifies lesion characteristics for a confident diagnosis CT clinical case study lesion characterization Desiree Morgan, MD Vice Chair of Clinical Research Professor

More information

Vascular CT Protocols

Vascular CT Protocols Vascular CT Protocols V 1D: Chest and abdominal CT angiogram (aortic dissection protocol) V 1T: Chest CT angiogram (aortic trauma protocol) V 2: Abdominal and pelvis CT angiogram (aortic aneurysm protocol)

More information

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography Supalerk Pattanaprichakul, MD 1, Sutipong Jongjirasiri, MD 2, Sukit Yamwong, MD 1, Jiraporn Laothammatas,

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Reducing Radiation Dose in Body CT: A Practical Approach to Optimizing CT Protocols

Reducing Radiation Dose in Body CT: A Practical Approach to Optimizing CT Protocols Medical Physics and Informatics Review Goldman and Maldjian Reducing Radiation Dose in ody CT Medical Physics and Informatics Review Downloaded from www.ajronline.org by lice Goldman on 03/26/13 from IP

More information

Medical Radiology Contrast Medium Injection Technique

Medical Radiology Contrast Medium Injection Technique Medical Radiology Contrast Medium Injection Technique --Manuscript Draft-- Manuscript Number: Article Type: Corresponding Author: Multidetector-Row CT of the Thorax D. Fleischmann, MD UNITED STATES First

More information

Handzettel 1. CT Contrast Media. Agenda. Contrast Media Definition. Agenda. Why we need contrast media? Agenda

Handzettel 1. CT Contrast Media. Agenda. Contrast Media Definition. Agenda. Why we need contrast media? Agenda Agenda CT Contrast Media Weena Swatdiswanee Factorinvolvein contrast enchancement Senior Application Specialist, CT Regional Headquarter Asia Australia weena.swat@siemens.com Page 1 Page 2 Agenda Contrast

More information

UROGENITAL. Abstract Objective To assess the effect of oral hydration and contrastmedium

UROGENITAL. Abstract Objective To assess the effect of oral hydration and contrastmedium Eur Radiol DOI 10.1007/s00330-010-1785-8 UROGENITAL Dieter H. Szolar Manfred Tillich Klaus W. Preidler Multi-detector CT urography: effect of oral hydration and contrast medium volume on renal parenchymal

More information

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Automatic Patient Centering for MDCT: Effect on Radiation Dose Patient Centering for MDCT CT Imaging Original Research Jianhai Li 1 Unni K. Udayasankar 1 Thomas L. Toth 2 John Seamans 2 William C. Small 1 Mannudeep K. Kalra 1,3 Li J, Udayasankar UK, Toth TL, Seamans

More information

Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma

Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma Poster No.: C-0245 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Yata,

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries

Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Utility of Variable Helical Pitch CT Scanning Technique for CT Angiography of Aortoiliac and Lower Extremity Arteries Poster No.: C-0863 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. Nakamoto,

More information

Newcastle HPB MDM updated radiology imaging protocol recommendations. Author Dr John Scott. Consultant Radiologist Freeman Hospital

Newcastle HPB MDM updated radiology imaging protocol recommendations. Author Dr John Scott. Consultant Radiologist Freeman Hospital Newcastle HPB MDM updated radiology imaging protocol recommendations Author Dr John Scott. Consultant Radiologist Freeman Hospital This document is intended as a guide to aid radiologists and clinicians

More information

CT Versus MR for the Runoff

CT Versus MR for the Runoff CT Versus MR for the Runoff Robert R. Edelman, M.D. Dept. of Radiology NorthShore University HealthSystem Feinberg School of Medicine, Northwestern University Magnetic Resonance Computed Tomography Radio

More information

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients

128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients 128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,

More information

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention Poster No.: C-0088 Congress: ECR 2014 Type: Scientific Exhibit Authors: C. Y. Lee, C. W.

More information

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention

Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention Imaging Features of Acute Pyelonephritis in Contrast Computed Tomography as Predictors of Need for Intervention Poster No.: C-0088 Congress: ECR 2014 Type: Scientific Exhibit Authors: C. Y. Lee, C. W.

More information

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing

Improving Diagnostic Accuracy of MDCT Coronary Angiography in Patients with Mild Heart Rhythm Irregularities Using ECG Editing Cademartiri et al. Heart Rhythm Irregularities on MDCT Angiography Cardiac Imaging Original Research A C M E D E N T U R I C A L I M A G I N G AJR 2006; 186:634 638 0361 803X/06/1863 634 American Roentgen

More information

Ct angiography, icd-10

Ct angiography, icd-10 Ct angiography, icd-10 The Borg System is 100 % Ct angiography, icd-10 Head and Neck. BW29 Head and Neck. BW290 High Osmolar. BW2900 Unenhanced and Enhanced. BW2900Z Computerized Tomography (CT Scan) of

More information

Contrast material enhanced computed tomography (CT) is the most commonly used imaging modality for the detection and characterization of liver metasta

Contrast material enhanced computed tomography (CT) is the most commonly used imaging modality for the detection and characterization of liver metasta Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Hiroshi Kondo, MD

More information

MDCT PROTOCOLS FOR POLYTRAUMA PATIENTS

MDCT PROTOCOLS FOR POLYTRAUMA PATIENTS András Palkó Department of Radiology, University of Szeged, Hungary MDCT PROTOCOLS FOR POLYTRAUMA PATIENTS Agenda Definition and significance Clinical implications Roleof imaging Examination protocols

More information

CT: Common Protocols. Michael Steigner, M.D. Director, Vascular CT/MR Assistant Professor of Radiology

CT: Common Protocols. Michael Steigner, M.D. Director, Vascular CT/MR Assistant Professor of Radiology CT: Common Protocols BRIGHAM AND WOMEN S HOSPITAL Heart & Vascular Center Harvard Medical School Teaching Hospital Michael Steigner, M.D. Director, Vascular CT/MR Assistant Professor of Radiology msteigner@bwh.harvard.edu

More information

Paediatric Dose Reduction and Image Quality

Paediatric Dose Reduction and Image Quality Paediatric Dose Reduction and Image Quality Alan Whiteside The majority of this work was undertaken as part of MSc Thesis of Helen Dixon. Introduction Paediatric CT protocols result in a higher effective

More information

CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous)

CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous) CTA Chest Pulmonary Embolism & Routine CT Abdomen + Pelvis W CTA Chest W (arterial) & CT Abdomen + Pelvis W (venous) Reviewed By: Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: August 2018 Contact:

More information

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography.

Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. ISPUB.COM The Internet Journal of Radiology Volume 11 Number 2 Detectability of subsegmental pulmonary vessels in 64 MDCT-pulmonary angiography. T Niemann, G Bongartz Citation T Niemann, G Bongartz. Detectability

More information

Three-dimensional CT angiography of the canine hepatic vasculature

Three-dimensional CT angiography of the canine hepatic vasculature J. Vet. Sci. (2008), 9(4), 407 413 JOURNAL OF Veterinary Science Three-dimensional CT angiography of the canine hepatic vasculature Yucheol Jeong, Changyun Lim, Sunkyoung Oh, Joohyun Jung, Jinhwa Chang,

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Lee et al. CT of Chronic Total and Subtotal Occlusive rterial Disease Vascular and Interventional Radiology Original Research Jung Eun Lee 1 Hee

More information

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA

Radiation Exposure in Pregnancy. John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Radiation Exposure in Pregnancy John R. Mayo UNIVERSITY OF BRITISH COLUMBIA Illustrative Clinical Scenario 32 year old female 34 weeks pregnant with recent onset shortness of breath and central chest pain

More information

CTA Pulmonary Embolism CTA Chest W (arterial)

CTA Pulmonary Embolism CTA Chest W (arterial) CTA Pulmonary Embolism CTA Chest W (arterial) Reviewed By: Rachael Edwards, MD; Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: January 2019 Contact: (866) 761-4200, Option 1 In accordance with the

More information

Role of three-phase contrast-enhanced multislice helical CT in evaluating resectability of pancreatic carcinoma

Role of three-phase contrast-enhanced multislice helical CT in evaluating resectability of pancreatic carcinoma ISPUB.COM The Internet Journal of Radiology Volume 8 Number 1 Role of three-phase contrast-enhanced multislice helical CT in evaluating resectability of pancreatic E Ergul, E Gozetlik Citation E Ergul,

More information

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience

Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Original Article Fused monochromatic imaging acquired by single source dual energy CT in hepatocellular carcinoma during arterial phase: an initial experience Shun-Yu Gao, Xiao-Peng Zhang, Yong Cui, Ying-Shi

More information

A multicentric study on patient dose in multislice CT

A multicentric study on patient dose in multislice CT A multicentric study on patient dose in multislice CT A.Stratis 1, M.Molfetas 1, S.Kottou 2, A.Louizi 2 1. Medical Physics department, Evangelismos General hospital of Athens, Athens, Greece 2. Medical

More information

Prospective Study of Access Site Complications of Automated Contrast Injection With Peripheral Venous Access in MDCT

Prospective Study of Access Site Complications of Automated Contrast Injection With Peripheral Venous Access in MDCT Special Article Original Research Wienbeck et al. Access Site Complications in MDCT Special Article Original Research Susanne Wienbeck 1 Roman Fischbach 1,2 Stephan P. Kloska 1 Peter Seidensticker 3 Noriaki

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information

A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT

A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT Poster No.: C-0623 Congress: ECR 2017 Type: Authors: Keywords: DOI:

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

CT & MRI of Benign Liver Neoplasms Srinivasa R Prasad

CT & MRI of Benign Liver Neoplasms Srinivasa R Prasad CT & MRI of Benign Liver Neoplasms Srinivasa R Prasad No financial disclosures Acknowledgements Many thanks to Drs. Heiken, Narra & Menias (MIR) Dr. Sahani (MGH) for sharing images Benign Liver Tumors:

More information

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT YAO ET AL ORIGINAL ARTICLE Radiation Medicine: Vol. 21 No. 6, 252 257 p.p., 2003 Communicating Vein between the Left Renal Vein and Left Ascending Lumber Vein: Incidence and Significance on Abdominal CT

More information

Visualization of Normal Pulmonary Fissures on Sagittal Multiplanar Reconstruction MDCT

Visualization of Normal Pulmonary Fissures on Sagittal Multiplanar Reconstruction MDCT Normal Pulmonary Fissures on MDCT Chest Imaging Original Research C D E M N E U T R Y L I M C I G O F I N G Koji Takahashi 1,2 rad Thompson 2 William Stanford 2 Yutaka Sato 2 Kenichi Nagasawa 1 Hiroaki

More information

International Journal of Radiology

International Journal of Radiology International Journal of Radiology Online Submissions: http://www.ghrnet.org/index./ijr/ doi:10.17554/j.issn.2313-3406.2016.03.31 Int. J. of Radiology 2016 March 3(1): 90-94 ISSN 2313-3406(print) ORIGINAL

More information

Breast Cancer PET/CT Imaging Protocol

Breast Cancer PET/CT Imaging Protocol Breast Cancer PET/CT Imaging Protocol Scanning Protocol: Patients are scanned from the top of the neck through the pelvis. Arms-up position is used to avoid beam-hardening artifact in the chest and abdomen.

More information

Measurement error of spiral CT Volumetry:

Measurement error of spiral CT Volumetry: Measurement error of spiral CT Volumetry: Influence of Low Dose CT Technique 1 Tae Gyu Lee, M.D. 2, Myung Jin Chung, M.D., Sung Bum Cho, M.D. 2, Jae Min Cho, M.D., Seog Joon Kim, M.D. 2, Sang Hyun Baik,

More information

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,

More information

CT Renal 3 Phase + Pelvis CT Abdomen Pelvis WO W - NC.A.V, Pelvis during V

CT Renal 3 Phase + Pelvis CT Abdomen Pelvis WO W - NC.A.V, Pelvis during V CT Renal 3 Phase + Pelvis CT Abdomen Pelvis WO W - NC.A.V, Pelvis during V Reviewed By: Anna Ellermeier, MD; Brett Mollard, MD Last Reviewed: August 2018 Contact: (866) 761-4200, Option 1 In accordance

More information

Ultralow Dose Chest CT with MBIR

Ultralow Dose Chest CT with MBIR Ultralow Dose Chest CT with MBIR Ella A. Kazerooni, M.D. Professor & Director Cardiothoracic Radiology Associate Chair for Clinical Affairs University of Michigan Disclosures Consultant: GE Healthcare

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Detectability of unruptured intracranial aneurysms on thinslice non-contrast-enhanced CT

Detectability of unruptured intracranial aneurysms on thinslice non-contrast-enhanced CT Detectability of unruptured intracranial aneurysms on thinslice non-contrast-enhanced CT Poster No.: C-9 Congress: ECR 5 Type: Scientific Exhibit Authors: M. Nakadate, Y. Iwasa, M. Kishino, U. Tateishi;

More information

With the rapid advancement of CT

With the rapid advancement of CT Cardiopulmonary Imaging Perspective Weininger et al. Contrast Medium Delivery in Cardiothoracic CT Angiography Cardiopulmonary Imaging Perspective FOCUS ON: Markus Weininger 1 J. Michael Barraza 1 Corey

More information

Perfusion Measurement in Acute Pancreatitis Using Dynamic Perfusion MDCT

Perfusion Measurement in Acute Pancreatitis Using Dynamic Perfusion MDCT Perfusion Measuremen t in cute Pancreatitis with MDT Gastrointestinal Imaging Technical Innovation M E D E N T U R I L I M G I N G Pierre E. Bize 1 lexandra Platon hristoph D. Becker Pierre-lexandre Poletti

More information

JVS. Original Article. Introduction. Soo-Young Choi 1, In Lee 1, Ji-Won Seo 2, Hyun-Young Park 2, Ho-Jung Choi 2, Young-Won Lee 2, *

JVS. Original Article. Introduction. Soo-Young Choi 1, In Lee 1, Ji-Won Seo 2, Hyun-Young Park 2, Ho-Jung Choi 2, Young-Won Lee 2, * Original Article J Vet Sci 2016, 17(4), 555-561 ㆍ https://doi.org/10.4142/jvs.2016.17.4.555 JVS Optimal scan delay depending on contrast material injection duration in abdominal multi-phase computed tomography

More information

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION Managing Patient Dose in Computed Tomography (CT) International Commission on Radiological Protection Information abstracted from ICRP Publication 87 Available at www.icrp.org Task Group: M.M. Rehani,

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information