Doppler Echocardiography
|
|
- Brent Freeman
- 5 years ago
- Views:
Transcription
1 Doppler cardiography Use of a Graphical Display System GERALD LORCH, M.D., SIMEON RUENSTEIN, M.D., DONALD AKER, TERRYL DOOLEY, AND HAROLD DODGE, M.D. SUMMARY Pulsed Doppler echocardiography (PDE) is a technique for evaluating blood flow characteristics at specific locations within the heart and great vessels. ecause this method assesses blood flow rather than cardiac structures, PDE complements the findings of M-mode echocardiography. A new on-line graphical method for displaying pulsed Doppler information provides 1) a printed, permanent record of flow information, 2) precise timing of blood flow events, and 3) information on the direction of flow in the heart and great vessels. CARDIAC ULTRASOUND has greatly enhanced our ability to define structural disorders in the heart and great vessels. While certain cardiac diseases lend themselves to firm diagnosis by M-mode echocardiography, many valvular and congenital disorders result in normal or nonspecific echocardiographic findings. In these disorders structural abnormalities clearly exist which cannot be evaluated with present echocardiographic techniques. A range-gated pulsed Doppler device has therefore been developed which can identify blood flow patterns at well-defined intracardiac locations.1 4 y demonstrating whether intracardiac flow is normal or pathologic, direct inference can be made as to what structural abnormality exists. The use of the pulsed Doppler device for the evaluation of murmurs has been previously described using an audible signal as the primary output for interpretation.1 This system however has recently been further improved by the introduction of a graphical recording method which increases ease of interpretation, provides a printed record of pulsed Doppler findings, and allows for more precise timing of blood flow events. It is the purpose of this paper to illustrate the applications of this graphic technique in identifying valvular abnormalities and septal defects. Methods In accordance with the Doppler theory, the frequencies of sound reflected from an object will be altered if that object is moving. This alteration is known as Doppler shift. lood will reflect high frequency sound because it contains particulate matter - namely the blood cells. The pulsed Doppler device uses repetitive bursts, or pulses, of sound in the megahertz range, much like traditional M-mode echocardiographic systems. While most ultrasonic waves are reflected from the various cardiac structures, a small amount is backscattered by the blood cells within the chambers themselves. These waves can be analyzed using a special Doppler shift detector system. If an ultrasonic beam is directed at a nonturbulent moving blood sample, the reflected sound waves will be of fairly uniform Doppler shift. y contrast sound waves reflected from a turbulent blood sample will show wide fluctuations in Doppler shift. The pulsed Doppler functions by detecting the Doppler From the Division of Cardiology and ioengineering, University of Washington, Seattle, Washington. Supported by NHLI grants HL and HL Address for reprints: Donald aker, Center for ioengineering, FT-15, University of Washington, Seattle, Washington Received July 12, 1976; revision accepted May shift from moving blood within the heart or great vessels and provides an output by which blood flow characteristics can be recognized. The primary advantage of the pulsed, as opposed to the continuous wave Doppler is that blood flow velocity may be evaluated from discrete intracardiac positions. This flow velocity is detected in a teardrop-shaped region called the sample volume which may be chosen from any depth from between 2 and 15 centimeters along the ultrasonic beam through the use of a movable range-gating system. The sample volume measures approximately 2 mm in diameter and 4 mm in length. The device employs a 3 MHz piezoelectric transducer with a 1/2 inch diameter crystal focused at 5 cm. The transmitter circuit emits 1 msec ultrasonic bursts at a pulse repetition frequency that may be varied between 3,500 and 10,000 bursts per second depending on the depth of penetration desired. The returned signal is divided such that one portion is used to create a traditional A-mode display for orientation; however the quality of this A-mode Doppler display is not yet comparable to conventional devices due to inherent resolution limitations. The other portion of the returned signal is used for Doppler flow velocity information. The magnitude and direction of the Doppler shift is determined in a special phase detector circuit where the returned frequency is compared to the transmitted frequency. Positioning the transducer in a manner similar to that used in traditional echocardiography and using the A-mode display for orientation, the sample volume is placed at well defined intracardiac locations simply by adjusting the range gate knob (fig. 1). There are two types of output available from a pulsed Doppler device. The first provides an audible signal in the range of 400 Hz to 5 khz which corresponds to the spectrum of Doppler shifts produced by normal or disturbed blood flow. The interpretation is based on evaluating the tonal quality of the signal. Smooth blood flow characteristically has a narrow bandwidth pattern which produces a tonal "musical-like" sound whereas disturbed or turbulent flow has a wide bandwidth pattern which produces a harsh, scratchy sound. This method has been described in detail in previous publications.1-3 The second type of output available is an on-line graphical display of the Doppler shift frequencies which is printed on a (Honeywell #1856) strip chart recorder. A circuit has been built into the pulsed Doppler detector for performing analysis of flow and creating a display pattern. This circuit creates what is called a time interval histogram (TIH) (fig. 2), which is a dot pattern by which flow velocity characteris-
2 DOPPLER CARDIOGRAPHY/Lorch et al. 577 DOPPLER TRANSDUCE )UND EAM AXI S A DOPPLER SHIFT SIGNAL ZERO CROSSING PULSE I', 'b T1 T2 T3 T4 T5 T6 T7 TS <., ~~~~- Zero c rossi ng VJkJV zero line & e b d * 6 6- * lntp-ncltv LLn LJL LJ LLL C * 0 TIME INTERVAL H ISTOGRAM 0. GATE VOLUME FIGURE 1. The sample volume is a teardrop-shaped region which may be selected from anywhere along the ultrasonic beam axis, by adjusting the depth gate. An A -mode display is usedfor orientation. tics can be easily identified. The pattern is derived by measuring the time between successive zero crossings of the audible signal waveform. A zero crossing is defined as the instant in time when the Doppler shift signal passes through its zero intensity level (fig. 2). A dot is printed on the strip chart paper for each zero crossing event such that the distance of a particular dot above or below the zero flow velocity baseline is directly proportional to the Doppler shift frequency. For exjmple, if the Doppler shift frequency is relatively low, a dot will be printed near the zero velocity baseline. However, if the Doppler shift frequency is relatively high, a dot will be printed proportionately farther away from the baseline. Any dot printed above the baseline (positive) represents relative blood velocity toward the transducer and conversely any dot printed below the baseline (negative) represents flow velocity away from the transducer. A dot pattern is thus produced which shows the Doppler shift spectrum, and this pattern is used to identify blood flow characteristics within the sample volume. When laminar flow exists, blood cells within the sample volume have fairly uniform direction and velocity. Therefore backscattered sound waves will have a fairly uniform Doppler shift. The time interval histogram produced by this flow has a narrowly clustered dot pattern (narrow band pattern) (fig. 3A). When "disturbed" flow exists, the blood cells have varied direction and velocity which produce a time interval histogram with a widely dispersed dot pattern (wide band pattern) (fig. 3). Strip chart records contain simultaneous recordings of 1) the M-mode display of the received Doppler signal (in the form of a traditional M-mode echo display) with a superimposed line indicating the location of the sample volume at any particular time, 2) the time interval histogram of the 0 zero flow Ti me line FIGURE 2. Production of a time interval histogram. A) The Doppler shift frequency signal. Each time the zero intensity line is crossed by the Doppler sound wave signal, the time period to the previous zero crossing is measured. This period is the reciprocal of the Doppler shift frequency. T1, T2, T8, etc., = successive periods between zero crossings. ) Graph illustrating when zero crossings occur. C) The time interval histogram. For each zero crossing event of the Doppler shift frequency signal, the time period (T) to the previous zero crossing is measured and a dot is printed. Its distance from the zero flow line is inversely proportional to that time period (T). For longer periods (as in T1), the dot is printed near the zero flow line. For shorter periods (as in T.), the dot is printed proportionately farther from the zero flow line. The distance ofa dotfrom the zero flow line is therefore directly proportional to the Doppler shift frequency. Doppler signal (also called a spectral display), and 3) a standard electrocardiogram. It is also possible to record an analog waveform of flow which corresponds to the mean value of the Doppler frequency shift; this can be shown as a separate channel on the recording but has proved of limited value and has been deleted from most later recordings. Patients for the study were selected from the routine referrals to our ultrasound laboratory for clinical evaluation. Most, but not all, with abnormal findings have had documentation of their pathology by cardiac catheterization and angiography. Each patient seen in our ultrasound laboratory routinely undergoes systematic cardiac ultrasonic examination, beginning with a complete M-mode echocardiographic assessment. Following this a pulsed Doppler examination is performed. eginning from the suprasternal notch, sampling is obtained from the proximal aorta and the right pulmonary artery., The transducer is then placed on the precordium and sampling performed on the inflow and outflow side of each valve and within each valve orifice itself to determine whether forward flow across the respective valve has a laminar or turbulent pattern, and whether regurgitant flow exists. When a septal defect is suspected, sampling is performed along the septum on the outflow side of the suspected defect, following which attempts are made to pass the sample volume directly through the defect.
3 578 CIRCULATION r rw yn VOL 56, No 4, OCTOER 1977 w Ak r: 14-0 t I,I i I FIGIURE 3. Flow patterns. A ) Noonturbulent flow. The dot pattern is narrowly dispersed (narrow band pattern) a result of uniform Doppler frequency shift. This reflects uniform direction and velocity of blood cells in the sample volume. A cluster of dots above the zero line reflects flow towards the transducer. ) Turbulent flow (X). During X the dot pattern is widely dispersed (broad band pattern) as a result of variations of Doppler shift within the sample volume. This reflects marked variation in blood cell velocities and direction, with some blood cells moving toward and others away from the transducer. as There are inherent technical limitations in the zero crossing detection system of the graphical display which make it slightly less sensitive and less immune to extraneous noise interference than the audible detection system. Therefore representative recordings of Doppler abnormalities are obtained both on magnetic tape and the graphical display system. Results General Normal and Turbulent Flow lood flow within a vessel or chamber is usually laminar and produces a time interval histogram with a narrowly clustered dot pattern (fig. 3A). Turbulent or disturbed flow, by contrast, produces a time interval histogram with a widely dispersed dot pattern (fig. 3). With disturbed blood flow the presence of dots which are widely dispersed both above and below the zero line illustrate the marked variation in blood cell direction and velocity within the sample, with some cells moving toward and others away from the transducer. Valvular Stenosis and Regurgitation Within and on the outflow side of a stenotic valvular orifice a jet of disturbed flow exists which produces a broad band pattern on Doppler examination. Valvular regurgitant lesions produce a broad band Doppler pattern, which is detected on the inflow side of the valve being examined. Septal Defects Much like valvular abnormalities, these lesions are detected by the turbulence which they produce. Large gradients, as seen for example in small ventricular septal defects, cause severe turbulence and produce markedly broad band patterns on Doppler examination. y contrast small gradient defects, as seen in atrial septal defects and large ventricular septal defects, cause less severe turbulence and produce mildly broad band patterns on Doppler examination. Aortic Valve Flow With the transducer positioned over the precordium, the systolic flow in or just above the aortic valve is normally laminar or slightly disturbed (fig. 4A). In diastole no flow is appreciated. As a valve leaflet traverses the sample volume this produces a fine mark on the recording which has been arbitrarily labeled a valve click because of its audible Doppler characteristics. Clicks during opening and closure of the aortic leaflets are usually recorded when the sample volume is placed within the valve orifice. Flow in the left ventricular outflow tract is essentially identical to that in the
4 ~~~~~~AA DOPPLER CARDIOGRAPHY/Lorch et al. A N a~(k~aa RVA 579 LA L LA I MA- -Aft.. AR604 N.- f R.}, ' S,. V1.,.L_ FIGURE 4. Aortic valve flow (from the precordium). A) Normal. There is a narrow band pattern (S) during systole representing nonturbulent flow. The upper portion of the figure is a schematic drawing of the actual M-mode recording displayed below. ) Aortic stenosis. In contrast to A), a widely dispersed dot pattern (X) representing turbulence is detected during systole. This patient was found at catheterization to have a mean aortic valve gradient of 70 mm Hg. AA = anterior wall of the aorta; AP = posterior wall of the aorta; LA = posterior wall of the left atrium; R VA = anterior wall of the right ventricular outflow tract; SV = sample volume; = time interval histogram of pulsed Dopplerflow; VC - valve click. This very fine line ofdots, created as a leaflet passes through the sample volume, can be seen on the original tracing but is lost in reproduction. proximal aorta except that valve clicks are not detected. In aortic stenosis a jet of systolic turbulence is detected within and above the aortic valve and is represented by a broad band Doppler pattern (fig. 4). In diastole a narrow band dot pattern is recorded at or near the zero flow baseline because no flow is present and the Doppler shift is therefore zero. In aortic insufficiency a regurgitant jet may be detected during diastole within the valve orifice, in the left ventricular outflow tract, and frequently well down into the left ventricular chamber (fig. SA). When searching for this lesion in the presence of mitral stenosis, diastolic turbulence must be documented within the aortic valve or high left ventricular outflow tract to make a firm diagnosis of coexisting aortic insufficiency. If attempts are made to follow the disturbed flow into each valve orifice, mitral stenosis and aortic insufficiency can be easily differentiated from one another. With idiopathic hypertrophic subaortic stenosis or subaortic rings, a broad band systolic pattern is detected within the left ventricular outflow tract in most patients (fig. S). In some of these patients disturbed flow may also be detected to a lesser extent above the aortic valve. Mitral Valve Flow If the sample volume is placed in the left ventricular inflow region, nonturbulent, anteriorly directed flow is found to commence with opening of the mitral valve (fig. 6A). This flow diminishes but is sustained until late diastole when a small flow wave occurs with atrial contraction. The mitral valve then closes and flow recorded from this position ceases. In some patients the early and late diastolic flow waves are separated in mid-diastole by what appears to be a period of virtually no detectable flow by Doppler. In mitral stenosis a broad band diastolic pattern is found within the mitral valve orifice, in the left ventricular inflow region, and frequently just posterior to the mitral valve in the left atrial outflow region (fig. 6). Normal flow patterns in the left atrial outflow region are quite similar to those detected in the left ventricular inflow (fig. 7A). Mitral regurgitation is represented by a broad band pattern detected in the left
5 580 CIRCULATION A RVA ~~ ~ RIVS LIVS J Sv MVA LA VOL 56, No 4, OCTOER RIVS ~~~~~~~ ~~~".LIVS 7MVA LVPW 9t IN T.- I ", I III~ II 4 z1a -,I II - q'it I... :0, w r, &A..1D.,,71 'f 1- FIGURE 5. Left ventricular outflow tract (L VOT). A) Aortic insufficiency. Turbulence (X) is detected in the L VOT during diastole. ) Idiopathic hypertrophic subaortic stenosis. Systolic turbulence (X) is detected in the L VOT. This patient had a resting gradient across the left ventricular outflow tract of 75 mm Hg. LIVS = left side ofthe interventricular septum; MVA = anterior leaflet of the mitral valve; LVPW = posterior wall (endocardium) of the left ventricle. atrial outflow region during systole (fig. 7). This turbulence is frequently confined to the latter half of systole in the presence of mitral valve prolapse. Tricuspid Valve Flow In both the normal and pathologic state, flow proximal and distal to this valve is quite similar to that detected at the mitral valve with the Doppler detector (fig. 8A, 8). Pulmonic Valve Flow The normal right ventricular outflow tract has a rather characteristic narrow band, negatively deflected pattern in systole since flow is directed posteriorly, away from the transducer. No significant flow occurs in diastole (fig. 9A). A similar pattern occurs in the proximal pulmonary artery; however mild systolic turbulence may be detected in the absence of a pathologic pulmonic valve. In pulmonic stenosis a broad band pattern is detected during systole in the proximal pulmonary artery. Pulmonic regurgitation produces a broad band pattern in the right ventricular outflow tract during diastole (fig. 9). Atrial Septal Defect When an atrial septal defect is suspected, the transducer may be directed in such a manner that the sample volume is passed from the left atrial chamber to the right atrial chamber while flow is being assessed (ASD maneuver). If this maneuver is performed several times with the range-gate set at slightly different distances, a sample can usually be obtained from directly within, and on the outflow side of the atrial septal defect. Intra-atrial shunting generally produces a turbulent pattern with the Doppler detector, but the timing has been variable. Recent intracardiac Doppler studies performed elsewhere"' suggest that shunting generally occurs at three separate points during each cycle: during 1) late systole, 2) early diastole, and 3) late diastole corresponding to atrial contraction. While we have frequently observed a similar timing pattern with the Doppler detector, in two
6 DOPPLER CARDIOGRAPHY/Lorch et al RIVS LIVS SV MVA PWLV Per. -j L. LJ I I r I RIVS MSV 6 P:PW I I i I 0 IV lbin WI', I-- I V t, -, 11, I 0. -T- % -t-- I 0 a til l 1- Analog I8Pt 6 FIGURE 6. Left ventricular inflow region (LVIR). A) Normal. With the sample volume (SV) placed anterior to the mitral valve, nonturbulent, anteriorly (positively) directedflow denotes the onset of diastole. This flow diminishes but is sustained until late systole when a smallflow wave occurs coincident with atrial contraction. ) Mitral stenosis. Diastolic turbulence (X) is detected in the L VIR and within the mitral valve. At catheterization this patient was found to have a mean mitral valve gradient of 7 mm Hg and a mitral valve area of 0.9 cm2. PER = pericardial/epicardial interface. Analog = wave form which corresponds to mean valve of Doppler frequency shift. patients shunting was observed to begin in late diastole and extend through early systole (fig. 10). Johnson et al. have shown that atrial septal defects with left-to-right shunt can be detected with the pulsed Doppler device in two-thirds of patients.'2 Ventricular Septal Defects Membranous ventricular septal defects are characterized by the presence of a broad band systolic jet in the right ventricular outflow tract (fig. 11). This turbulent jet can frequently be followed through the interventricular septal defect itself. The localization of systolic turbulence to the right ventricular outflow tract and through a portion of the interventricular septum differentiates this lesion from infundibular pulmonic stenosis. In the latter the systolic turbulence is confined to the right ventricular outflow tract. In the muscular form of ventricular septal defect, a systolic jet exists in the body of the right ventricle. For reasons that are not entirely clear, the jet is frequently difficult to identify and can only occasionally be followed through the septal defect itself. Discussion The range-gated pulsed Doppler instrument can greatly extend the amount of useful information obtained in a cardiac ultrasound examination. ecause it deals with cardiac blood flow rather than cardiac structure, this device tends to complement information obtained from M-mode echocardiography. The information generated is of four basic types. Identification of the origin of murmurs. y sampling at precise intracardiac locations, the pulsed Doppler device can reliably identify the location and timing of turbulent blood flow events. The timing corresponds to that heard with the stethoscope or recorded on phonocardiogram. From this Doppler information the abnormality causing a particular murmur may be directly inferred. This is especially useful in those lesions such as mitral regurgitation, tricuspid regurgitation, pulmonic regurgitation, and ventricular septal defects which produce either minimal or nonspecific abnormalities by traditional M-mode echocardiography. For example, the development of a new systolic murmur heard
7 582 CIRCULATION VOL 56, No 4, OCTOER 1977 RIVS LIVS MVA Sv LA f~~~~~~~~~~~~~~~~~~~~..... Ṉ :-.-, -,". I -,.1 -,.O'. N..,-1.,.1.,I.I.,-1, ". -,.,-'1--,..- RIVS t LI v/c V D AMV sv LA I. _~~~a t I I 11 1 I T r 11 XIC r_ 1 T EC G FIGURE 7. Left atrial outflow region (LA OR). A) Normal. ) Mitral regurgitation. A moderately turbulent pattern (X) is detected in the LA OR during systole. While the patient did not undergo catheterization, this 21-year-oldfemale had a strong family history of Marfan's disease and nonrheumatic mitral regurgitation. The patient had a loud mitral regurgitant murmur and diagnostic mitral valve prolapse on echocardiogram. e = early diastolic flow; a = late diastolic flow commencing with atrial contraction; VC = valve click. from the left lower sternal border to the apex in a patient several days after an acute myocardial infarction could represent an acquired ventricular septal defect or mitral regurgitation from severe papillary muscle dysfunction. Traditional M-mode echocardiography would not distinguish between these two entities. However by localizing the site of the flow disturbance, the pulsed Doppler detector can be used to provide a firm diagnosis. Similarly, pulmonic can be separated from aortic regurgitation, aortic stenosis from subvalvular obstructive disease, and mitral from tricuspid regurgitation. In light of recent reports of normal posterior leaflet motion in some cases of mitral stenosis,6 7 the Doppler detector should prove a useful adjunct in deciding whether this disease is present or not. The obtaining of information not recognized on physical examination. Loud murmurs on auscultation may sometimes obliterate the presence of other softer but potentially significant murmurs. For example, mitral or tricuspid regurgitation might be overlooked in the face of a harsh aortic flow murmur, but the presence of multiple areas of disturbed flow can be easily recognized on pulsed Doppler examination. As another example, the diagnosis of an atrial septal defect can sometimes be difficult to make with confidence using our usual clinical modalities. Paradoxical septal motion seen on echocardiogram is neither specific, nor a constant feature,8' 9 and provides only indirect support for the diagnosis.10 While the flow across an ASD is almost always inaudible by auscultation, it can usually be detected on Doppler examination. Estimation of the quantitative significance of murmurs. This is an application that is just beginning to be explored. The Doppler detector alone has not proved useful thus far in quantitating the severity of stenotic valvular lesions. y contrast, it appears that regurgitant lesions are detected over an increasingly larger area and for a longer period of time as they become increasingly severe. For example insignificant mitral regurgitation may be detected during just a portion
8 DOPPLER CARDIOGRAPHY/Lorch et al. 583 A J _} iz TV RA TV sv RA Analog FIGURE 8. Right atrial outflow. A) Normal. ) Tricuspid regurgitation (X). This patient had severe mitral stenosis at catheterization with mildfunctional tricuspid regurgitation. Exact quantitation of the tricuspid regurgitation was not performed. of systole and the turbulence is localized to the inflow side of the mitral valve. However with severe mitral regurgitation turbulence is generally pansystolic and detected through much of the left atrium. This correlation awaits further validation studies by catheterization and angiography. Separation of innocent from pathologic flow murmurs. The significance of systolic flow murmurs in young people is sometimes difficult to assess with the usual noninvasive methods. y Doppler examination innocent murmurs produce minimal or no turbulence, whereas pathologic flow murmurs tend to be quite turbulent. The reason for this is not entirely clear, as presumably some turbulence is being produced in the presence of functional murmurs. It may well be, however, that this turbulence is so mild that it is not detected by the pulsed Doppler device. This area of investigation also requires further evaluation and validation. The use of a graphic display for the Doppler system has several advantages when compared to sole reliance upon an 771! r 4E audible output. First, it provides a printed, permanent record which simultaneously shows 1) the M-mode display of the Doppler signal with a superimposed line indicating the locations of the sample volume at any particular time, 2) the Doppler spectral display (time interval histogram), and 3) the. With the audible signal, storage must be done on electromagnetic tape, a somewhat cumbersome recording method which has made retrieval and transferring of information difficult. Secondly, the graphic display improves ease of interpretation. A fair degree of expertise is necessary to interpret reliably the audible Doppler signal, and this becomes less critical with use of the graphic display. Thirdly, a graphic display provides the opportunity to precisely time the onset and completion of turbulent blood flow events. This is useful in understanding flow characteristics in such disorders as atrial septal defects. Finally, information generated from the phase shift detector system can now be applied to determine the direction of
9 584 PV CIRCULATION VOL 56, No 4, OCTOER 1977 % sv PV Analog FIGURE 9. Right ventricular outflow (R VO T). A) Normal. A characteristic, negatively deflectedpattern is detected during systole because flow is posteriorly directed in the RVOT. ) Pulmonic regurgitation. A moderately turbulent, positively deflected pattern (X) is detected in the R VOT during diastole. PV = pulmonic valve. > w~~~~~~~~~~~~~qf di.. w ~ ~ ~ ~ I u e. i E _z ti r I. TV Sv RA FIGURE 10. Atrial septal defect. A scan is initiated with the sample volume in the left atrium. The transducer is then directed to the patient's right so that the sample volume passes through the defect into the right atrium. A bnormalflow is detected within and on the outflow side of the ASD. In this case the abnormalflow begins in late diastole and ends in mid-systole. Catheterization confirmed a left-to-right shunt in this patient at the atrial level with a pulmonic/systemic flow ratio of 2.2 to 1.
10 DOPPLER CARDIOGRAPHY/Lorch et al5 sv LIVS MVA _.t LVPW Acknowledgement References ti<ptitt -.A t k: w Thl 1! 5Ei5r' 2.h =iai - 11 FIGURE 1 1. Ventricular septal defect (VSD). Marked turbulence (X) isfound in the right ventricle during systole. A VSD with left-toright shunt was documented in this nine-year-oldfemale by left ventricular angiography and oxygen saturations. A pulmonic/systemic flow ratio of 1.8 to 1 (4 liter shunt) was found. 585 flow by pulsed Doppler. This is not possible with the audible signal. The graphic system in its present form requires sensitive adjustment to obtain a proper recording and therefore has not replaced the audible system at this time. However the ability to display information graphically about intracardiac flow events has added a new dimension to the ultrasonic examination of patients. The authors wish to acknowledge the contributions of Vernon Simmons, Robert Olson, and Ronald Daigle from the Ultrasound Program in the Center for ioengineering for instrumentation and technical assistance; to Steve L. Johnson, M.D., for his contributions to the clinical analysis and to arbara Eyer for typing the manuscript. 1. Johnson SL, aker DW, Lute RA, Dodge HT: Doppler echocardiography: The localization of cardiac murmurs. Circulation 48: 810, Johnson SL: Pulse Doppler cardiography. Seattle, Celcom Press, aker DW, Johnson SL: Doppler echocardiography. In Cardiac Ultrasound, edited by Gramiak R, St. Louis, Mosby Co, aker DW: Pulsed ultrasonic doppler blood-flow sensing. IEEE Transactions on Sonics and Ultrasonics, Vol SU-17, No 3, Goldberg : Suprasternal ultrasonography. JAMA 215: 245, Ticzon AR, Damato AN, Caracta AR, Lau SH, Gross L: cardiographic manifestations of "false" mitral stenosis that was. Ann Intern Med 83: 503, Flaherty JT, Livergood S, Fortuin NJ: Atypical posterior leaflet motion in echocardiogram in mitral stenosis. Am J Cardiol 35: 675, Hagan AD, Francis GS, Sahn DJ, Karliner JS, Friedman WF, O'Rourke RA: Ultrasound evaluation of systolic anterior septal motion in patients with and without right ventricular volume overload. Circulation 50: 248, Radtke WE, Tajik AJ, Gau GT, Schaltenberg TT, Guiliani ER, Tancredi RG: Atrial septal defect: cardiographic observations. Ann Intern Med 84: 246, Solinger RE: Congenital heart disease. In Cardiac Ultrasound, edited by Gramiak R. St Louis, Mosby Co, Kalmanson D, Veyrat C, et al: Non-invasive technique for diagnosing atrial septal defect and assessing shunt volume using directional Doppler ultrasound. r Heart J 34: 981, Johnson SL, Rubenstein SA, Kawabori I, Dooley T, aker OW: The detection of atrial septal defect by pulsed doppler flowmeter. Circulation 54 (suppl II): , 1976
Assessment of pulsed Doppler echocardiography in detection and quantification of aortic and mitral
Br Heart J 198; 44: 612-2 Assessment of pulsed Doppler echocardiography in detection and quantification of aortic and mitral regurgitation* MIGUEL A QUINONES, JAMES B YOUNG, ALAN D WAGGONER, MIODRAG C
More informationHISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.
HISTORY 18-year-old man. CHIEF COMPLAINT: Heart murmur present since early infancy. PRESENT ILLNESS: Although normal at birth, a heart murmur was heard at the six week check-up and has persisted since
More informationPART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING
PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING STANDARD - Primary Instrumentation 1.1 Cardiac Ultrasound Systems SECTION 1 Instrumentation Ultrasound instruments
More informationHISTORY. Question: What category of heart disease is suggested by the fact that a murmur was heard at birth?
HISTORY 23-year-old man. CHIEF COMPLAINT: Decreasing exercise tolerance of several years duration. PRESENT ILLNESS: The patient is the product of an uncomplicated term pregnancy. A heart murmur was discovered
More informationPulmonary valve echo motion in pulmonary
British HeartJournal, I975, 37, ii84-ii90. Pulmonary valve echo motion in pulmonary regurgitation' Arthur E. Weyman, James C. Dillon, Harvey Feigenbaum, and Sonia Chang From the Department of Medicine,
More informationPHONOCARDIOGRAPHY (PCG)
PHONOCARDIOGRAPHY (PCG) The technique of listening to sounds produced by the organs and vessels of the body is called auscultation. The areas at which the heart sounds are heard better are called auscultation
More informationGiovanni Di Salvo MD, PhD, FESC Second University of Naples Monaldi Hospital
Giovanni Di Salvo MD, PhD, FESC Second University of Naples Monaldi Hospital VSD is one of the most common congenital cardiac abnormalities in the newborn. It can occur as an isolated finding or in combination
More informationAdult Echocardiography Examination Content Outline
Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,
More informationHemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics
Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial
More informationAppendix II: ECHOCARDIOGRAPHY ANALYSIS
Appendix II: ECHOCARDIOGRAPHY ANALYSIS Two-Dimensional (2D) imaging was performed using the Vivid 7 Advantage cardiovascular ultrasound system (GE Medical Systems, Milwaukee) with a frame rate of 400 frames
More informationHISTORY. Question: What type of heart disease is suggested by this history? CHIEF COMPLAINT: Decreasing exercise tolerance.
HISTORY 15-year-old male. CHIEF COMPLAINT: Decreasing exercise tolerance. PRESENT ILLNESS: A heart murmur was noted in childhood, but subsequent medical care was sporadic. Easy fatigability and slight
More informationCOMBINED CONGENITAL SUBAORTIC STENOSIS AND INFUNDIBULAR PULMONARY STENOSIS*
COMBINED CONGENITAL SUBAORTIC STENOSIS AND INFUNDIBULAR PULMONARY STENOSIS* BY HENRY N. NEUFELD,t PATRICK A. ONGLEY, AND JESSE E. EDWARDS From the Sections of Pa?diatrics and Pathological Anatomy, Mayo
More informationCardiac Ausculation in the Elderly
Cardiac Ausculation in the Elderly 박성하 신촌세브란스병원심장혈관병원심장내과 Anatomy Surface projection of the Heart and Great Vessels Evaluating pulsation Superior vena cava Rt. pulmonary artery Right atrium Right ventricle
More informationPROSTHETIC VALVE BOARD REVIEW
PROSTHETIC VALVE BOARD REVIEW The correct answer D This two chamber view shows a porcine mitral prosthesis with the typical appearance of the struts although the leaflets are not well seen. The valve
More informationUncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency
Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency Rahul R. Jhaveri, MD, Muhamed Saric, MD, PhD, FASE, and Itzhak Kronzon, MD, FASE, New York, New York Background: Two-dimensional
More informationAortic root and left atrial wall motion
British Heart Journal, 1977, 39, 1082-1087 Aortic root and left atrial wall motion An echocardiographic study GONES AKGtYN AND CLIVE LAYTON1 From the Cardiac Department, The London Hospital, London The
More informationfound that some patients without stenotic lesions had blood velocity or pressure measurement across the
Br Heart J 1985; 53: 640-4 Increased blood velocities in the heart and great vessels of patients with congenital heart disease An assessment of their significance in the absence of valvar stenosis STANLEY
More informationASCeXAM / ReASCE. Practice Board Exam Questions Monday Morning
ASCeXAM / ReASCE Practice Board Exam Questions Monday Morning Ultrasound Physics Artifacts Doppler Physics Imaging, Knobology, and Artifacts Echocardiographic Evaluation of the RV Tricuspid and Pulmonary
More informationMethods Forty patients representing all consecutive patients referred to. our laboratory on randomly selected days were chosen for
DIAGNOSTIC METHODS DOPPLER ECHOCARDIOGRAPHY Comparison of multiple views for the evaluation of pulmonary arterial blood flow by Doppler echocardiography GEORGE W. LIGHTY, JR., M.D., PH.D., ANTHONY GARGIULO,
More informationAdel Hasanin Ahmed 1
Adel Hasanin Ahmed 1 PERICARDIAL DISEASE The pericardial effusion ends anteriorly to the descending aorta and is best visualised in the PLAX. PSAX is actually very useful sometimes for looking at posterior
More informationCardiac ultrasound protocols
Cardiac ultrasound protocols IDEXX Telemedicine Consultants Two-dimensional and M-mode imaging planes Right parasternal long axis four chamber Obtained from the right side Displays the relative proportions
More informationDiagnostic approach to heart disease
Diagnostic approach to heart disease Initial work up History Physical exam Chest radiographs ECG Special studies Echocardiography Cardiac catheterization Echocardiography principles Technique of producing
More informationClinical significance of cardiac murmurs: Get the sound and rhythm!
Clinical significance of cardiac murmurs: Get the sound and rhythm! Prof. dr. Gunther van Loon, DVM, PhD, Ass Member ECVDI, Dip ECEIM Dept. of Large Animal Internal Medicine Ghent University, Belgium Murmurs
More informationECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT
ECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT Karen Stout, MD, FACC Divisions of Cardiology University of Washington Medical Center Seattle Children s Hospital NO DISCLOSURES
More informationThe Doppler Examination. Katie Twomley, MD Wake Forest Baptist Health - Lexington
The Doppler Examination Katie Twomley, MD Wake Forest Baptist Health - Lexington OUTLINE Principles/Physics Use in valvular assessment Aortic stenosis (continuity equation) Aortic regurgitation (pressure
More informationDoppler Basic & Hemodynamic Calculations
Doppler Basic & Hemodynamic Calculations August 19, 2017 Smonporn Boonyaratavej MD Division of Cardiology, Department of Medicine Chulalongkorn University Cardiac Center, King Chulalongkorn Memorial Hospital
More informationDoppler Color Flow Imaging #4
Doppler Color Flow Imaging #4 Joseph A. Kisslo, MD David B. Adams, RDCS INTRODUCTION Doppler color flow imaging is a method for noninvasively imaging blood flow through the heart by displaying flow data
More informationDoppler-echocardiographic findings in a patient with persisting right ventricular sinusoids
Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 1990 Doppler-echocardiographic findings in a patient with persisting right
More informationHeart sounds and murmurs. Dr. Szathmári Miklós Semmelweis University First Department of Medicine 15. Oct
Heart sounds and murmurs Dr. Szathmári Miklós Semmelweis University First Department of Medicine 15. Oct. 2013. Conditions for auscultation of the heart Quiet room Patient comfortable Chest fully exposed
More informationICE: Echo Core Lab-CRF
APPENDIX 1 ICE: Echo Core Lab-CRF Study #: - Pt Initials: 1. Date of study: / / D D M M M Y Y Y Y 2. Type of Study: TTE TEE 3. Quality of Study: Poor Moderate Excellent Ejection Fraction 4. Ejection Fraction
More information2. The heart sounds are produced by a summed series of mechanical events, as follows:
Heart Sounds. Phonocardiography 1 Objectives 1. Phonocardiography - Definition 2. What produces the heart sounds 3. Where to listen for the heart sounds 4. How to record a phonocardiogram 5. Normal heart
More informationEchocardiography of Cardiac Valves in
Echocardiography of Cardiac Valves in Pericardial Effusion NAVN C. NANDA, M.D., RAYMOND GRAMAK, M.D., AND CHARLES M. GROSS, M.D. SUMMARY Nine adult patients with large pericardial effusions (PE) demonstrated
More informationCase # 1. Page: 8. DUKE: Adams
Case # 1 Page: 8 1. The cardiac output in this patient is reduced because of: O a) tamponade physiology O b) restrictive physiology O c) coronary artery disease O d) left bundle branch block Page: 8 1.
More informationThe Cardiac Cycle Clive M. Baumgarten, Ph.D.
The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,
More informationEchocardiographic Patterns of Pulmonic Valve Motion with Pulmonary Hypertension
Echocardiographic Patterns of Pulmonic Valve Motion with Pulmonary Hypertension By ARTHUR E. WEYMAN, M.D., JAMES C. DILLON, M.D., HARVEY FEIGENBAUM, M.D., AND SONIA CHANG, B.A. SUMMARY Echocardiographic
More informationCARDIOVASCULAR PHYSIOLOGY
CARDIOVASCULAR PHYSIOLOGY LECTURE 4 Cardiac cycle Polygram - analysis of cardiac activity Ana-Maria Zagrean MD, PhD The Cardiac Cycle - definitions: the sequence of electrical and mechanical events that
More informationCARDIAC EXAMINATION MINI-QUIZ
CARDIAC EXAMINATION MINI-QUIZ 1. Sitting bolt upright, your dyspneic (short of breath) patient has visible jugular venous pulsations to the angle of his jaw, which is 12 cm above his sternal angle. What
More informationHeart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall.
Heart and Lungs Normal Sonographic Anatomy THORAX Axial and coronal sections demonstrate integrity of thorax, fetal breathing movements, and overall size and shape. LUNG Coronal section demonstrates relationship
More informationP = 4V 2. IVC Dimensions 10/20/2014. Comprehensive Hemodynamic Evaluation by Doppler Echocardiography. The Simplified Bernoulli Equation
Comprehensive Hemodynamic Evaluation by Doppler Echocardiography Itzhak Kronzon, MD North Shore LIJ/ Lenox Hill Hospital New York, NY Disclosure: Philips Healthcare St. Jude Medical The Simplified Bernoulli
More informationHEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE
Vet Times The website for the veterinary profession https://www.vettimes.co.uk HEART MURMURS: DECIPHERING THEIR CAUSE AND SIGNIFICANCE Author : Pedro Oliveira Categories : Vets Date : May 27, 2013 PEDRO
More informationHISTORY. Question: How do you interpret the patient s history? CHIEF COMPLAINT: Dyspnea of two days duration. PRESENT ILLNESS: 45-year-old man.
HISTORY 45-year-old man. CHIEF COMPLAINT: Dyspnea of two days duration. PRESENT ILLNESS: His dyspnea began suddenly and has been associated with orthopnea, but no chest pain. For two months he has felt
More informationCongenital heart disease. By Dr Saima Ali Professor of pediatrics
Congenital heart disease By Dr Saima Ali Professor of pediatrics What is the most striking clinical finding in this child? Learning objectives By the end of this lecture, final year student should be able
More informationAdult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016
1 Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016 DISCLOSURES I have no disclosures relevant to today s talk 2 Why should all echocardiographers
More informationMETHODS WILLIAM J. STEWART, MD,* KATHLEEN A. GALVIN, LINDA D. GILLAM, MD, FACC, DAVID E. GUYER, MD, FACC, ARTHUR E.
lacc Vol. 6, No.3 September 1985:565 71 565 METHODS Comparison of High Pulse Repetition Frequency and Continuous Wave Doppler Echocardiography in the Assessment of High Flow Velocity in Patients With Valvular
More information2) VSD & PDA - Dr. Aso
2) VSD & PDA - Dr. Aso Ventricular Septal Defect (VSD) Most common cardiac malformation 25-30 % Types of VSD: According to position perimembranous, inlet, muscular. According to size small, medium, large.
More informationM-Mode Echocardiography Is it still Alive? Itzhak Kronzon, MD,FASE. Sampling Rate M-Mode: 1800 / sec 2D: 30 / sec
M-Mode Echocardiography Is it still Alive? Itzhak Kronzon, MD,FASE Honoraria: Philips Classical M-mode Echocardiography M-Mode offers better time and image resolution. Sampling Rate M-Mode: 1800 / sec
More informationleft atrial myxoma causes paradoxical motion of the catheter; posterior
Am JRoentgenolla6:II55-II58, 1976 ABNORMAL LEFT VENTRICULAR CATHETER MOTION: AN ANCILLARY ANGIOGRAPHIC SIGN OF LEFT ATRIAL MYXOMA ABsTRACT: J. M. RAU5CH, R. T. REINKE, K. L. PETERSON,2 AND C. B. HIGGINs
More informationLittle is known about the degree and time course of
Differential Changes in Regional Right Ventricular Function Before and After a Bilateral Lung Transplantation: An Ultrasonic Strain and Strain Rate Study Virginija Dambrauskaite, MD, Lieven Herbots, MD,
More informationTissue Doppler Imaging in Congenital Heart Disease
Tissue Doppler Imaging in Congenital Heart Disease L. Youngmin Eun, M.D. Department of Pediatrics, Division of Pediatric Cardiology, Kwandong University College of Medicine The potential advantage of ultrasound
More informationBritish Society of Echocardiography
British Society of Echocardiography Affiliated to the British Cardiac Society A Minimum Dataset for a Standard Adult Transthoracic Echocardiogram From the British Society of Echocardiography Education
More information5 Working With Measurements
5 Working With Measurements Measurement Overview Measurements accompanying ultrasound images supplement other clinical procedures available to the attending physician. Accuracy of the measurements is determined
More informationHEMODYNAMIC ASSESSMENT
HEMODYNAMIC ASSESSMENT INTRODUCTION Conventionally hemodynamics were obtained by cardiac catheterization. It is possible to determine the same by echocardiography. Methods M-mode & 2D echo alone can provide
More informationUltrasound 10/1/2014. Basic Echocardiography for the Internist. Mechanical (sector) transducer Piezoelectric crystal moved through a sector sweep
Ultrasound Basic Echocardiography for the Internist Carol Gruver, MD, FACC UT Erlanger Cardiology Mechanical wave of compression and rarefaction Requires a medium for transmission Ultrasound frequency
More informationA Magnetic Resonance Imaging Method for
Journal of Cardiovascular Magnetic Resonance, 1(1), 59-64 (1999) INVITED PAPER Use of MRI in ASD Asessment A Magnetic Resonance Imaging Method for Evaluating Atrial Septa1 Defects Godtfred Holmvang Cardiac
More informationCOLIC AND MURMURS: AN OVERVIEW
COLIC AND MURMURS: AN OVERVIEW Gunther van Loon, DVM, PhD, Department of Large Animal Internal Medicine, Ghent University, Merelbeke, Belgium Introduction Many horses with colic present with a cardiac
More informationEcho Doppler Assessment of Right and Left Ventricular Hemodynamics.
Echo Doppler Assessment of Right and Left Ventricular Hemodynamics. Itzhak Kronzon, MD, FASE, FACC, FESC, FAHA, FACP, FCCP Northwell, Lenox Hill Hospital, New York Professor of Cardiology Hofstra University
More informationDepartment of Pediatrics and Child Health, Kurume University School of Medicine, Kurume, 830 Japan. Received for publication October 26, 1992
THE KURUME MEDICAL JOURNAL Vol.39, p.291-296, 1992 Jon-Invasive Evaluation of Pulmonary Arterial and Right Ventricular Pressures with Contrast Enhanced Doppler Signals of Tricuspid Regurgitation Flow Using
More informationComprehensive Hemodynamics By Doppler Echocardiography. The Echocardiographic Swan-Ganz Catheter.
Comprehensive Hemodynamics By Doppler Echocardiography. The Echocardiographic Swan-Ganz Catheter. Itzhak Kronzon, MD, FASE, FACC, FESC, FAHA, FACP, FCCP North Shore HS, LIJ/Lenox Hill Hospital, New York
More informationTGA atrial vs arterial switch what do we need to look for and how to react
TGA atrial vs arterial switch what do we need to look for and how to react Folkert Meijboom, MD, PhD, FES Dept ardiology University Medical entre Utrecht The Netherlands TGA + atrial switch: Follow-up
More informationBrief View of Calculation and Measurement of Cardiac Hemodynamics
Cronicon OPEN ACCESS EC CARDIOLOGY Review Article Brief View of Calculation and Measurement of Cardiac Hemodynamics Samah Alasrawi* Pediatric Cardiologist, Al Jalila Children Heart Center, Dubai, UAE *
More informationCardiac Cycle MCQ. Professor of Cardiovascular Physiology. Cairo University 2007
Cardiac Cycle MCQ Abdel Moniem Ibrahim Ahmed, MD Professor of Cardiovascular Physiology Cairo University 2007 1- Regarding the length of systole and diastole: a- At heart rate 75 b/min, the duration of
More informationPulsed Wave Doppler and Color Flow Doppler Evaluation in Healthy Dogs and Dogs with Cardiac Disease
Cloud Publications International Journal of Advanced Veterinary Science and Technology 2016, Volume 5, Issue 2, pp. 256-265, Article ID Sci-446 ISSN 2320-3595 Research Article Open Access Pulsed Wave Doppler
More informationThe production of murmurs is due to 3 main factors:
Heart murmurs The production of murmurs is due to 3 main factors: high blood flow rate through normal or abnormal orifices forward flow through a narrowed or irregular orifice into a dilated vessel or
More informationAortic Stenosis: LVOT Obstruction
Aortic Stenosis: LVOT Obstruction Raymond Stainback, MD 7 th annual Houston Echo Review 2016: Boot Camp for the Echo Board Murmur: Additional heart or vascular sound due to normal or abnormal turbulent
More informationInspiratory Right Ventricular Outflow Obstruction in a Patient with Hypertrophic Cardiomyopathy
Case Reports Inspiratory Right Ventricular Outflow Obstruction in a Patient with Hypertrophic Cardiomyopathy Kazufumi TSUCHIHASHI, M.D., Akihito TSUCHIDA, M.D., Nobuichi HIKITA, M.D., Shuji YONEKURA, M.D.,
More informationPRACTICAL ECHOCARDIOGRAPHY IN THE ADULT with Doppler and color-doppler flow imaging
PRACTICAL ECHOCARDIOGRAPHY IN THE ADULT with Doppler and color-doppler flow imaging PRACTICAL ECHOCARDIOGRAPHY IN THE ADULT with Doppler and color-doppler flow imaging by J.P.M. HAMER Thoraxcentre, Department
More informationIntroduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03
Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount
More informationCases in Adult Congenital Heart Disease
Cases in Adult Congenital Heart Disease Sabrina Phillips, MD FACC FASE Associate Professor of Medicine The University of Oklahoma Health Sciences Center No Disclosures I Have Palpitations 18 Year old Man
More informationMuscle dynamics in hypertrophic cardiomyopathy
Postgraduate Medical Journal (1986) 62, 545-551 Muscle dynamics in hypertrophic cardiomyopathy Roando F. Alvares and John F. Goodwin Division ofcardiovascular Diseases, Hammersmith Hospital, London W12,
More informationThe background of the Cardiac Sonographer Network News masthead is a diagnostic image:
Number 5 Welcome Number 5 Welcome to the newsletter created just for you: sonographers who perform pediatric echocardiograms in primarily adult echo labs. Each issue features tips on echocardiography of
More informationECHOCARDIOGRAPHY DATA REPORT FORM
Patient ID Patient Study ID AVM - - Date of form completion / / 20 Initials of person completing the form mm dd yyyy Study period Preoperative Postoperative Operative 6-month f/u 1-year f/u 2-year f/u
More informationThe production of murmurs is due to 3 main factors:
Heart murmurs The production of murmurs is due to 3 main factors: high blood flow rate through normal or abnormal orifices forward flow through a narrowed or irregular orifice into a dilated vessel or
More informationPhysical Exam Part II
Physical Exam Part II University of Michigan Cardiovascular Center Kim A. Eagle, MD Albion Walter Hewlett Professor Director Physical Exam: Part II Heart Sounds Heart Murmurs HEART SOUNDS S1 MITRAL + TRICUSPID
More informationAssessment of LV systolic function
Tutorial 5 - Assessment of LV systolic function Assessment of LV systolic function A knowledge of the LV systolic function is crucial in the undertanding of and management of unstable hemodynamics or a
More informationPediatric Echocardiography Examination Content Outline
Pediatric Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 Anatomy and Physiology Normal Anatomy and Physiology 10% 2 Abnormal Pathology and Pathophysiology
More informationCONGENITAL HEART DISEASE (CHD)
CONGENITAL HEART DISEASE (CHD) DEFINITION It is the result of a structural or functional abnormality of the cardiovascular system at birth GENERAL FEATURES OF CHD Structural defects due to specific disturbance
More informationPRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan
PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan Fetal Cardiology Unit, Harris Birthright Research Centre for Fetal Medicine, King's College Hospital, London, UK IMPORTANCE OF PRENATAL
More informationAtrial Septal Defects
Supplementary ACHD Echo Acquisition Protocol for Atrial Septal Defects The following protocol for echo in adult patients with atrial septal defects (ASDs) is a guide for performing a comprehensive assessment
More informationCertificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE)
Certificate in Clinician Performed Ultrasound (CCPU) Syllabus Rapid Cardiac Echo (RCE) Purpose: Rapid Cardiac Echocardiography (RCE) This unit is designed to cover the theoretical and practical curriculum
More informationIdiopathic Hypertrophic Subaortic Stenosis and Mitral Stenosis
CASE REPORTS Idiopathic Hypertrophic Subaortic Stenosis and Mitral Stenosis Martin J. Nathan, M.D., Roman W. DeSanctis, M.D., Mortimer J. Buckley, M.D., Charles A. Sanders, M.D., and W. Gerald Austen,
More informationvolume allows characterization of areas of normal and abnormal blood flow within the cardiac chambers and vessels. If an abnormal highvelocity
Subject Review Use of Doppler Techniques (Continuous-Wave, Pulsed-Wave, and Color Flow Imaging) in the Noninvasive Hemodynamic Assessment of Congenital Heart Disease GUY S. REEDER, M.D., PHILIP J. CURRIE,
More informationpulmonary valve on, 107 pulmonary valve vegetations on, 113
INDEX Adriamycin-induced cardiomyopathy, 176 Amyloidosis, 160-161 echocardiographic abnormalities in, 160 intra-mural tumors similar to, 294 myocardial involvement in, 160-161 two-dimensional echocardiography
More informationMITRAL STENOSIS. Joanne Cusack
MITRAL STENOSIS Joanne Cusack BSE Breakdown Recognition of rheumatic mitral stenosis Qualitative description of valve and sub-valve calcification and fibrosis Measurement of orifice area by planimetry
More informationTEE Outside of the Cardiac OR
TEE Outside of the Cardiac OR STEVE GIBSON MD PHD OU DEPARTMENT OF ANESTHEIOLOGY I have no financial relationships or conflicts of interest to disclose TRANSESOPHAGEAL ECHOCARDIOGRAPHY Basic principles
More informationB-Mode measurements protocols:
Application Note How to Perform the Most Commonly Used Measurements from the Cardiac Measurements Package associated with Calculations of Cardiac Function using the Vevo Lab Objective The Vevo LAB offline
More informationTricuspid and Pulmonic Valve Disease
Chapter 31 Tricuspid and Pulmonic Valve Disease David A. Tate Acquired disease of the right-sided cardiac valves is much less common than disease of the leftsided counterparts, possibly because of the
More information8/31/2016. Mitraclip in Matthew Johnson, MD
Mitraclip in 2016 Matthew Johnson, MD 1 Abnormal Valve Function Valve Stenosis Obstruction to valve flow during that phase of the cardiac cycle when the valve is normally open. Hemodynamic hallmark - pressure
More informationPULMONARY VALVE ECHOCARDIOiGRAM IN THE EVALUATION OF PULMONARY ARTERIAL HYPERTENSION IN THE PRESENCE OF INTRACARDIAC SHUNTS
Med. J. Malaysia Vol. 41 No. 3 September 1986 PULMONARY VALVE ECHOCARDIOiGRAM IN THE EVALUATION OF PULMONARY ARTERIAL HYPERTENSION IN THE PRESENCE OF INTRACARDIAC SHUNTS FONG CHEE YEE K. T. SINGHAM SUMMARY
More informationMurmur Sounds made by turbulence in the heart or blood stream. 1. Timing. 5. Intensity 2. Shape. 6. Pitch 3. Location of maximum intensity
Definition Items in description of Timing Shape Location of maximum intensity Murmur Sounds made by turbulence in the heart or blood stream. 1. Timing 5. Intensity 2. Shape 6. Pitch 3. Location of maximum
More informationNegative Contrast Echocardiography: A New Method for Detecting Left-to-Right Shunts
Negative Contrast Echocardiography: A New Method for Detecting Left-to-Right Shunts ARTHUR E. WEYMAN, M.D., L. SAM WANN, M.D., RANDALL L. CALDWELL, M.D., ROGER A. HURWITZ, M.D., JAMES C. DILLON, M.D.,
More informationIndex of subjects. effect on ventricular tachycardia 30 treatment with 101, 116 boosterpump 80 Brockenbrough phenomenon 55, 125
145 Index of subjects A accessory pathways 3 amiodarone 4, 5, 6, 23, 30, 97, 102 angina pectoris 4, 24, 1l0, 137, 139, 140 angulation, of cavity 73, 74 aorta aortic flow velocity 2 aortic insufficiency
More informationULTRASOUND OF THE FETAL HEART
ULTRASOUND OF THE FETAL HEART Cameron A. Manbeian, MD Disclosure Statement Today s faculty: Cameron Manbeian, MD does not have any relevant financial relationships with commercial interests or affiliations
More informationNon-invasive assessment of aortic stenosis by
Br Heart3' 198; 43: 284-292 Non-invasive assessment of aortic stenosis by Doppler ultrasound L HTLE, B NGELSEN, TROMSDL From the Section of Cardiology, Medical Department, University Hospital, Trondheim;
More information5.8 Congenital Heart Disease
5.8 Congenital Heart Disease Congenital heart diseases (CHD) refer to structural or functional heart diseases, which are present at birth. Some of these lesions may be discovered later. prevalence of Chd
More informationRight Heart Catheterization. Franz R. Eberli MD Chief of Cardiology Stadtspital Triemli, Zurich
Right Heart Catheterization Franz R. Eberli MD Chief of Cardiology Stadtspital Triemli, Zurich Right Heart Catheterization Pressure measurements Oxygen saturation measurements Cardiac output, Vascular
More information1. how a careful cardiovascular evaluation can accurately assess pathology and physiology at the bedside, and
This program will demonstrate: 1. how a careful cardiovascular evaluation can accurately assess pathology and physiology at the bedside, and 2. the importance of integrating this information with selected
More informationTricuspid and Pulmonary Valve Disease
Tricuspid and Pulmonary Valve Disease Lawrence Rudski MD FRCPC FACC FASE Professor of Medicine Director, Division of Cardiology Jewish General Hospital McGill University Question 1 All of the following
More informationAssessment of Aortic Blood Flow Velocities With Continuous Wave Doppler Ultrasound in the Neonate and Young Child
JCC Vol. 5, No. I January 1985:113S-9S ll3s ssessment of ortic Blood Flow Velocities With Continuous Wave Doppler Ultrasound in the Neonate and Young Child LIV, MD Trondheim, Norway ortic flow velocities
More informationMost common fetal cardiac anomalies
Most common fetal cardiac anomalies Common congenital heart defects CHD % of cardiac defects Chromosomal Infants Fetuses anomaly (%) 22q11 deletion (%) VSD 30 5~10 20~40 10 PS 9 5 (PA w/ VSD) HLHS 7~9
More information