1 Iranian Journal of Medical Physics Vol. 8, No. 1, Winter 212, 1-8 Received: January 15, 212; Accepted: February 1, 212 Original Article Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife Tayeb Allahverdi Pourfallah 1*, Mahmoud Allahverdi 2, Mohammad Hasan Zahmatkesh 3 Abstract Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with each other. Materials and Methods The head phantom was a custom-built 16 cm diameter plexiglas sphere. Inside the phantom, there were two cubic cutouts for inserting the gel vials and inhomogeneities. Following irradiation with the Gamma Knife unit, the polymer gel dosimeters were scanned with a 1.5 T MRI scanner. For the purpose of simulation the simplified channel of Co source of Gamma Knife BEAMnrc and for extracting the 3D dose distribution in the phantom, DOSXYZnrc codes were used. Results Within high isodose levels (>%), there are dose differences higher than 7%, especially between air inserted and PTFE inserted phantoms, which were obtained using both simulation and experiment. This means that these values exceed the acceptance criterion of conformal radiotherapy and stereotactic radiosurgery (i.e., within some isodose levels, less than 93% of prescription dose are delivered to the target). Conclusion The discrepancies observed between the results obtained from heterogeneous and homogeneous phantoms suggest that Leksell Gamma Knife planning system (LGP) predictions which assume the target as a homogeneous material must be corrected in order to take care of the air- and bone-tissue inhomogeneities. Keywords: Dose Profile, Gamma Knife, Inhomogeneity, Monte Carlo, Polymer Gel Dosimeter 1- Biochemistry and Biophysics Dept., Faculty of Medicine, Mazandaran University of Medical Sciences, Sari, Iran *Corresponding author: Tel: , ; tpourfallah@mazums.ac.ir 2- Medical Physics Dept., Faculty of Medicine, Tehran University of Medical Sciences, Tehran, Iran 3- Novin Medical Radiation Institute, Tehran, Iran Iran J Med Phys, Vol. 8, No. 1, Winter 212 1

2 Tayeb Allahverdi Pourfallah et al. 1. Introduction Stereotactic Gamma Knife radiosurgery plays an important role in managing small intracranial brain lesions. For patient treatment, a stereotactic frame is attached to the patient s head which establishes a threedimensional (3D) coordinate system for the determination of the precise target location with imaging techniques. All 21 Co beams intersecting at the unit center point (UCP) and four helmets with different sized collimators form four standard clinical beam sizes (18, 14, 8, and 4 nominal diameters) [1, 2]. The treatment planning system (TPS) of Gamma Knife assumes that the target is a homogeneous material and in dose calculations does not consider the presence of inhomogeneities (e.g., bony structures and sinuses) which may lead to considerable dose disturbances [3, 4]. In the investigation of dose perturbations produced by heterogeneities, Monte Carlo (MC) has proved to be a useful tool, because it adequately accounts for the lack of electron equilibrium close to interfaces. The degree of accuracy that can be attained by this method is determined mostly by the accuracy of the cross-section data, the radiation beams with respect to energy and angular distribution, the statistical accuracy of the MC calculation method, and how the phantom geometry and tissue properties are related to the radiation interactions that are modeled. The EGSnrc based BEAMnrc and DOSXYZnrc from the National Research Council of Canada (NRCC) group were used in this study [5]. The EGSnrc was developed from the EGS4 code [6]. The EGS (Electron Gamma Shower) system of computer codes is a general purpose package for the Monte Carlo simulation of the coupled transport of electrons and photons in an arbitrary geometry for particles with energies above a few kev up to several hundreds of GeV [7]. Moreover, polymer gel dosimetry is still the only dosimetry method for directly measuring threedimensional dose distributions. These dosimeters are tissue equivalent and can act as a phantom material [8, 9]. Polymer gel dosimetry is a technique that can map absorbed radiation dose distributions in three dimensions with a high spatial resolution and also offer a number of advantages over the traditional dosimeters such as ionization chambers, thermoluminescent dosimeters (TLDs), and radiographic films. The advantages include independence of radiation direction, radiological soft tissue equivalence, integration of the dose for a number of sequential treatment fields, and perhaps most importantly, evaluation of a complete volume at once. Several studies have been performed for investigating the effects of inhomogeneities on dose distribution using MC simulation and conventional dosimeters [4, 1-12]; however, studies related to polymer gel dosimetry along with simulation are rare [9, 13-15]. In this study, effects of inhomogeneities on those distributions have been investigated using both EGSnrc calculation and PAGAT (PAG (PolyAcrylamide Gel) And THPC antioxidant) polymer gel dosimeter and dose profiles in three-dimensional (3D) coordinate system have been investigated. 2. Materials and Methods 2.1. Gel fabrication PAGAT polymer gel dosimeter was fabricated according to composition proposed by Venning et al who noted that using MRI, the formulation to give the maximum change in the transverse relaxation rate R2 was determined to be 4.5% N,N methylene-bisacrylamide (bis), 4.5% acrylamide (AA), 5% gelatine, 5 mm THPC,.1 mm hydroquinone (HQ), and 86% H2O. For fabricating the gel dosimeter, the De Deene et al. [16] proposed method has been used in which for the npag gels containing crosslinker, the acrylamide and crosslinker N,N -methylene-bis-acrylamide were first dissolved in % total water volume by heating to 45 C and the gelatin solution was then cooled down to 35 C before it was mixed with the monomer solution. The antioxidant was added to the solution under heavy stirring just before filling the test tubes [17]. 2 Iran J Med Phys, Vol. 8, No. 1, Winter 212

3 Effects of Inhomogeneities in Gamma Knife Figure 1. Schematic view of the phantom in three spatial coordinates. The location of the cubic cutouts for the gel vial and inserting the air and bone inhomogeneities are depicted Design and irradiation of the phantom Phantom in this study was a 16 cm spherical plexiglas in which there was a cubic cutout for inserting the gel cubes (4 4 4 cm 3 ) and another cutout (4 4 3 cm 3 ) for inserting the air and/or a bone equivalent material (Polytetra-fluoro-ethylene (PTFE)), with density of 2.2 gr/cm 3 ). Figure 1 shows a schematic view of the phantom in three spatial coordinates. In this view, the location of the cubic cutouts for the gel vial and inserting the air and bone inhomogeneities are also depicted. Figure 2 shows the phantom placed in a Gamma Knife unit (model C) for singleshot irradiations with 18 mm collimator to the maximum dose of Gy. The calibration tubes were irradiated using the Theratron Co machine using especial water filled container ( cm 3 ) in which the calibration vials could be placed horizontally at a depth of 5 cm. The calibration vials were irradiated from to 45 Gy with steps of 5 Gy. Post manufacture irradiation time was 24 h Evaluation of gel dosimeter The evaluation of the polymerized dosimeter was performed on a 1.5 T clinical Siemens scanner in the transmitter/receiver head coil. A multi-echo sequence with 32 echoes was used for the evaluation of irradiated polymer-gel dosimeters. The parameters used for the sequence were as follows: TR=3 ms, TE=22 6 ms, slice thickness=1 mm, FOV=128 mm, matrix size= , pixel size=.5.5 mm 2, and one acquisition. The R2 (spin-lattice relaxation rate, 1/T) was computed using modified radiotherapy gel dosimetry image processing software coded in MATLAB. The R2 matrix was subsequently converted into a relative dose matrix normalized to the maximum prescribed dose of Gy. Figure 2. The Leksell Gamma Knife stereotactic frame attached to the spherical plexiglas head phantom. Iran J Med Phys, Vol. 8, No. 1, Winter 212 3

4 Tayeb Allahverdi Pourfallah et al. Figure 3 shows the R2-dose response curve obtained from PAGAT polymer gel dosimeter in which the R2-dose sensitivity from -2 Gy and 2- Gy were.9±.1 s-1 Gy -1 and.8±.1 s-1 Gy -1, respectively. Figure 3. R2-dose curve obtained from PAGAT polymer gel dosimeter Monte Carlo modeling The EGSnrc-based BEAMnrc code [18, 19] was used to simulate the geometry of the simplified Gamma Knife source channel, and outputs phase-space data (phase-space files), which include all the particle information (i.e., the charge, position, direction, energy, and history tag for each particle). Previous studies have shown that this simplified geometry allows a significant reduction of the simulation time, without loss of accuracy in the doses delivered to the phantom [2]. Another general-purpose MC EGSnrs user code DOSXYZnrc [19, 21] was employed to model the 21 Co sources of Gamma Knife unit. For performing this modeling, the Phase-space source from multiple directions, source type of DOSXYZnrc code was used. This code was also employed to obtain the 3D dose distributions in the phantom which considers the phantom divided in a large number of small volume elements, or voxels. PAGAT polymer gel dosimeter and MC simulation. Figure 5 compares simulation with measurements in three spatial coordinates. Table 1 shows dose differences between homogeneous and inhomogeneous phantoms and Table 2 shows comparison between measurements and the simulation. Because of the especial phantom geometry and also isocentre within gel cubes, the correct dose distribution using gel dosimeter could only be obtained at the Y direction. Therefore, in order to compare dose profiles only the negative direction of Y axis was considered. Table 1. Dose differences (DDs) for MC simulation and PAGAT polymer gel dosimeter. Homogeneous vs. air inserted phantom (DD(H-A)%), Homogeneous vs. PTFE inserted phantom (DD(H-P)%), and air inserted vs. PTFE inserted phantom (DD(A-P)%), in three coordinate axes. MC simulation Axes DD (H-A)% DD (H-P)% DD (A-P)% X 4.8± ±1.1 7.±1. Y 4.94± ± ±1.84 Z 3.64± ± ±2.1 PAGAT gel dosimeter Axes DD (H-A)% DD (H-P)% DD (A-P)% X 3.23± ± ±3.37 Y 3.1± ± ±2.5 Z 4.27± ± ±3.5 Table 2. Dose differences (DDs) for MC simulation vs. PAGAT polymer gel dosimeter in air inserted, homogeneous and PTFE inserted phantoms along the three coordinate axes. Axes Air inserted phantom Homogeneous phantom PTFE inserted phantom X 4.16± ± ±.79 Y 3.39± ± ±1.16 Z 2.32± ± ± Results Figure 4 (a-f) compares the relative dose profiles in three spatial coordinates for 4 Iran J Med Phys, Vol. 8, No. 1, Winter 212

5 Effects of Inhomogeneities in Gamma Knife (a) Homogen ph. (b) 12 Air inserted ph. 12 Relative dose (%). 2 PTFE inserted ph. Relative dose (%) Z axis (mm) Z axis (mm) (c) (d) Relative dose (%)... Relative dose (%) (e) (f) Relative dose (%)... Relative dose (%) Y axis (mm) Y axis (mm) Figure 4. Comparing relative dose profiles along three coordinate axes (X, Y, and Z) for MC simulation (b, d, f) and PAGAT polymer gel dosimeter (a, c, e) in irradiation with 18 mm collimator. Relative dose (%)... (a) Gel dosimetry 12 MC simulation Relative dose (%)... (b) Relative dose (%)... (c) Figure 5. Comparing relative dose profiles along three coordinate axes (X, Y, and Z). MC simulation vs. PAGAT polymer gel dosimeter in a) air inserted, b) homogeneous, and c) PTFE inserted phantoms, in irradiation with 18 mm collimator. Iran J Med Phys, Vol. 8, No. 1, Winter 212 5

6 Tayeb Allahverdi Pourfallah et al. 4. Discussion Regarding acceptance criteria for conformal radiotherapy, not more than 7% dose difference in delivered dose to the target is acceptable [22], and it is obvious that the Gamma Knife as a stereotactic radiosurgery unit must convey such a degree of accuracy. According to values of Table 1 within high isodose levels (>%), there are dose differences higher than 7%, especially between air inserted and PTFE inserted phantoms, which were obtained using both simulation and experiment. This means that these values exceed the acceptance criterion of conformal radiotherapy and stereotactic radiosurgery (i.e., within some isodose levels less than 93% of prescription dose are delivered to the target). It seems that different linear attenuation coefficient of inserted materials within the phantom can cause such dose disturbances. Several studies have been performed for investigating the effects of inhomogeneities on dose distribution using MC simulation and conventional dosimeters [4, 1-12]; however, in the case of polymer gel dosimetry along with simulation only a few studies exist which will be reviewed in the following. In one study [23], considerable differences were found between diameter of isodoses less than % between homogeneous and inhomogeneous phantoms using MAGIC gel dosimeter which is in contrast with our study. According to their study, the diameters of the 5% isodose curves differed 43% in the X axis and 32% in the Y axis. We found no reasonable answer for these differences between the two studies. In another study [4], using conventional dosimeters and simulation (PENELOPE), it was found that the dose delivered to the target area away from an air-tissue interface may be underestimated by up to 7% by GammaPlan due to overestimation of attenuation of photon beams passing through air cavities. Their findings are somewhat similar to our findings in air inserted phantom. Al-Dweri et al. [1] determined that dose distribution for heterogeneous phantoms including the bone- and/or air-tissue interfaces show non-negligible differences with respect to those calculated for a homogeneous one, mainly when the Gamma Knife isocentre approaches the separation surfaces. Their findings confirm an important underdosage (~1%) nearby the air-tissue interface. However, their study was in interfaces of bone-tissue and air-tissue and somewhat different from our goal of study. Allahverdi Pourfallah et al. [15] in one study through investigating the dose-volume histograms using simulation and polymer gel dosimetry showed that in irradiation with 18 mm collimator of the Gamma Knife unit, 23.24% difference in DVH within 9% % relative isodose level for homogeneous and inhomogeneous phantoms exist. Furthermore, authors revealed that a significant part of the target (28.56%) received relative doses higher than the maximum dose, which exceeds the acceptance criterion. In another study [14], the authors showed that the presence of inhomogeneities in head phantom could cause spatial uncertainty higher than ±2 mm and dose uncertainty higher than 7% when measurement and simulation in homogeneous and inhomogeneous phantoms are used. Regarding the results of comparison between simulation and measurement (Table 2), it is clear that the observed difference between them is within the acceptance criterion (<7%). However, some differences (at most 4.16±2.83% in air inserted phantom) were observed between the two which may be due to experimental uncertainty and parameters that can affect the accuracy of polymer gel dosimetry [24]. Imaging artifacts may affect the accuracy of gel dosimetry. Dosimetric imaging artifacts can be related to MRI machine or the dosimeter itself. Machine-related artifacts originate from imperfections in the scanning device while dosimeter-related artifacts are mainly attributed to a temperature drift during scanning or molecular self-diffusion. 6 Iran J Med Phys, Vol. 8, No. 1, Winter 211

7 Effects of Inhomogeneities in Gamma Knife In addition, MRI protocol with more acquisition or larger voxel size can improve signal to noise ratio. Noisy results can cause deviations between polymer gel measurements and MC plus LGP calculations. More studies may be necessary to answer these questions. 5. Conclusions The discrepancies observed between the results obtained for heterogeneous and homogeneous phantoms suggest that LGP predictions which assume the target as a homogeneous material must be corrected in order to deal with the air-and bone-tissue inhomogeneities. These results suggest that algorithms considering the tissue differences in the head would calculate the delivered dose more accurately. The aim of treatment using Gamma Knife unit is to cover the tumor or the lesion with the high isodose levels. However, ignoring tissue inhomogeneities can mislead accurate localization of dose levels. Clinically, this can cause underdose irradiation of the tumor in some regions or overdose irradiation of the normal tissue in an unwanted region. It could also be concluded that the applied MC code, i.e., EGSnrc is a suitable tool for 3D evaluation of dose distribution in irradiation with Gamma Knife unit, which can be used as an important evaluation criterion for 3D dose distribution in clinical practice. Acknowledgment We would like to thank staff of the "Iran Gamma Knife Center" and "Novin Medical Radiation Institute" for their kind Cooperation. References 1. Leksell L. The stereotaxic method and radiosurgery of the brain. Acta Chir Scand. 1951;12: Mack A, Scheib S, Major J, Gianolini S, Pazmandi G, Feist H, Czempiel H, Kreiner H. Precision dosimetry for narrow photon beams used in radiosurgery-determination of Gamma Knife output factors. Med Phys. 22;29(9): Yu C, Shepard D. Treatment planning for stereotactic radiosurgery with photon beams. Technol Cancer Res Treat. 23;2: Moskvin V, Timmerman R, DesRosiers C, Randall M, DesRosiers P, Dittmer P, Papiez L. Monte Carlo simulation of the Leksell gamma Knife (R): II. Effects of heterogeneous versus homogeneous media for stereotactic radiosurgery. Phys Med Biol. 24;49(21): Rogers D, Faddegon B, Ding G, Ma C, Wei J, Mackie T. BEAM: A Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22: Nelson W, Hirayama H, Rogers D. The EGS4 Code System. Stanford Linear Accelerator Center, Stanford, California; 1985.Report SLAC Kawrakow I, Mainegra-Hing E, Rogers D, Tessier F, Walters B. The EGSnrc Code System: Monte Carlo Simulation of Electron and Photon Transport. Ottawa : National Research Council of Canada; 211.Report PIRS Fong PM, Keil DC, Does MD, Gore JC. Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere. Phys Med Biol. 21;46(12): Venning AJ, Nitschke KN, Keall PJ, Baldock C. Radiological properties of normoxic polymer gel dosimeters. Med Phys. 25;32(4): Al-Dweri FM, Rojas EL, Lallena AM. Effects of bone- and air-tissue inhomogeneities on the dose distributions of the Leksell Gamma Knife calculated with PENELOPE. Phys Med Biol. 25;5(23): Cheung JY, Yu KN. Dose distribution close to metal implants in gamma knife radiosurgery: A Monte Carlo study [letter]. Med Phys. 25;32(5): Lewis RD, Ryde SJ, Seaby AW, Hancock DA, Evans CJ. Use of Monte Carlo computation in benchmarking radiotherapy treatment planning system algorithms. Phys Med Biol. 2;45(7): Pourfallah TA, Allahverdi M, Alam NR, Ay MR, Zahmatkesh MH, Ibbott GS. Performance evaluation of MRI-based PAGAT polymer gel dosimeter in an inhomogeneous phantom using EGSnrc code on a Co- machine. Appl Radiat Isot. 29;67(1): Iran J Med Phys, Vol. 8, No. 1, Winter 212 7

8 Tayeb Allahverdi Pourfallah et al. 14. Allahverdi Pourfallah T, Allahverdi M, Riahi Alam N, Ay MR, Zahmatkesh MH. Verifying the accuracy of dose distribution in gamma knife unit in presence of inhomogeneities using PAGAT polymer gel dosimeter and MC simulation. Iran J Radiat Res. 29;7: Pourfallah TA, Allahverdi M, Alam NR, Ay MR, Zahmatkesh MH. Differential dose volume histograms of Gamma Knife in the presence of inhomogeneities using MRI polymer gel dosimetry and MC simulation. Med Phys. 29;36(7): De Deene Y, Vergote K, Claeys C, De Wagter C. The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels. Phys Med Biol. 26;51(3): Venning AJ, Hill B, Brindha S, Healy BJ, Baldock C. Investigation of the PAGAT polymer gel dosimeter using magnetic resonance imaging. Phys Med Biol. 25;5(16): Rogers DWO, Walters B, Kawrakow I. BEAMnrc Users Manual. Ottawa : National Research Council of Canada; 27 May. 268 p. Report PIRS-59(A)revL. 19. Treurniet JR, Walters BR, Kawrakow I, Rogers DWO. BEAMnrc, DOSXYZnrc and BEAMDP GUI User s Manual. Ottawa: National Research Council of Canada; 25 Jul. 17 p. Report PIRS-623(rev C). 2. Al-Dweri FM, Lallena AM, Vilches M. A simplified model of the source channel of the Leksell GammaKnife (R) tested with PENELOPE. Phys Med Biol. 24;49(12): Walters B, Kawrakow I, Rogers DWO. DOSXYZnrc Users Manual. Ottawa : National Research Council of Canada; 29 Jul. Report PIRS-794revB. 22. Perez CA, Brady LW, Halperin EC, Schmidt-Ullrich RK. Principles and Practice of Radiation Oncology.4nd ed. Philadelpia: LIPPINCOTT WILLIAMS & WILKINS; Isbakan F, Ulgen Y, Bilge H, Ozen Z, Agus O, Buyuksarac B. Gamma Knife 3-D dose distribution near the area of tissue inhomogeneities by normoxic gel dosimetry. Med Phys. 27;34(5): De Deene, Y. On the accuracy and precision of gel dosimetry. J Phys Conf Ser. 26;56: Iran J Med Phys, Vol. 8, No. 1, Winter 211

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging Research Journal of Applied Sciences, Engineering and Technology (): 6-, ISSN: -767 Maxwell Scientific Organizational, Submitted: November 9, Accepted: November 9, Published: February, Comparison of Energy

More information

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique Australian Journal of Basic and Applied Sciences, 5(11): 879-886, 2011 ISSN 1991-8178 Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and

More information

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry Journal of Physics: Conference Series Optimization of the T parametric image map calculation in MRI polymer gel dosimetry To cite this article: Fotini Zacharopoulou et al J. Phys.: Conf. Ser. 5 9 Related

More information

Australian Journal of Basic and Applied Sciences. Evaluation of CT based MAGAT gel dosimetry system for measuring 3D dose distribution

Australian Journal of Basic and Applied Sciences. Evaluation of CT based MAGAT gel dosimetry system for measuring 3D dose distribution AENSI Journals Australian Journal of Basic and Applied Sciences Journal home page: www.ajbasweb.com Evaluation of CT based MAGAT gel dosimetry system for measuring 3D dose distribution 1 M. Aljamal and

More information

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,

More information

Dose differences between the three dose calculation algorithms in Leksell GammaPlan

Dose differences between the three dose calculation algorithms in Leksell GammaPlan JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Dose differences between the three dose calculation algorithms in Leksell GammaPlan Andy (Yuanguang) Xu, 1a Jagdish Bhatnagar 1, Greg

More information

MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery

MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 2, spring 2011 MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery N. Gopishankar,

More information

3D Dosimetry with Polymer Gel

3D Dosimetry with Polymer Gel 3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer

More information

X-ray computed tomography imaging of polymer gel dosimeters

X-ray computed tomography imaging of polymer gel dosimeters Journal of Physics: Conference Series X-ray computed tomography imaging of polymer gel dosimeters To cite this article: M Hilts 2006 J. Phys.: Conf. Ser. 56 95 View the article online for updates and enhancements.

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara 1, Z Kuncic 1, J Adamovics 2 and C Baldock 1,3 1 Institute of Medical Physics, School of Physics, University

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

RadioTherapy. 80 wt% 90 wt% 3wt% 5wt% 3 wt% 10 wt% 1

RadioTherapy. 80 wt% 90 wt% 3wt% 5wt% 3 wt% 10 wt% 1 With rapid advances being made in radiotherapy treatment, three dimensional (3D) dose measurement techniques with great precision are required more than ever before. It is expected that polymer gel dosimeters

More information

Study of a spherical phantom for Gamma knife dosimetry

Study of a spherical phantom for Gamma knife dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Study of a spherical phantom for Gamma knife dosimetry Dengsong Zhu, 1a Carlos Austerlitz, 2 Sidi Benhabib, 2 Helvecio Mota,

More information

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague Radiosurgery by Leksell gamma knife Josef Novotny, Na Homolce Hospital, Prague Radiosurgery - Definition Professor Lars Leksell The tools used by the surgeon must be adapted to the task and where the human

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Evaluation of the basic properties of the BANGkit gel dosimeter

Evaluation of the basic properties of the BANGkit gel dosimeter Title page a) Title of article Evaluation of the basic properties of the BANGkit gel dosimeter b) Authors and addresses Y Murakami 1, T Nakashima 2, Y Watanabe 3, T Akimitsu 4, K Matsuura 1, M Kenjo 1,

More information

The Development of Normoxic Polymer Gel Dosimetry using High Resolution MRI

The Development of Normoxic Polymer Gel Dosimetry using High Resolution MRI Queensland University of Technology School of Physical and Chemical Sciences The Development of Normoxic Polymer Gel Dosimetry using High Resolution MRI Christopher Hurley M.App.Sci.(Med Phys), M.Ed.Admin.,

More information

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method Muhammad B. Kakakhel, 1 Tanya Kairn,

More information

NMR spectroscopy of saccharide-doped PAGAT dosimeters

NMR spectroscopy of saccharide-doped PAGAT dosimeters University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2015 NMR spectroscopy of saccharide-doped PAGAT

More information

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,

More information

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy Journal of Physics: Conference Series OPEN ACCESS Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy To cite this article: Shin-ichiro Hayashi et al 2015 J. Phys.: Conf.

More information

Dosimetric Analysis Report

Dosimetric Analysis Report RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------

More information

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter Applications of Nuclear Techniques (CRETE15) International Journal of Modern Physics: Conference Series Vol. 44 (2016) 1660221 (9 pages) The Author(s) DOI: 10.1142/S2010194516602210 Dosimetric characteristics

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

Gamma Knife radiosurgery with CT image-based dose calculation

Gamma Knife radiosurgery with CT image-based dose calculation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Gamma Knife radiosurgery with CT image-based dose calculation Andy (Yuanguang) Xu, 1a Jagdish Bhatnagar, 1 Greg Bednarz, 1 Ajay Niranjan,

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Dosimetry of the Gamma Knife TM

Dosimetry of the Gamma Knife TM Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife

More information

FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system

FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system Collaboration Project: Θ Θ Θ INFN, Milan; Medical Physics Department, Niguarda Ca'Granda Hospital, Milan; Elekta

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

3D MRI gel dosimetry based on image intensity (A new approach)

3D MRI gel dosimetry based on image intensity (A new approach) Iran. J. Radiat. Res., 2003; 1 (1): 45-50 3D MRI gel dosimetry based on image intensity (A new approach) T Sh. Akhlaghpoor *, M.H. Zahmatkesh, H. Pourbeigi Novin Medical Radiation Institute, Tehran, Iran

More information

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA M. Pimpinella, A. S. Guerra, S. La Civita and R. F. Laitano Istituto

More information

Applications of Monte Carlo simulations to radiation dosimetry

Applications of Monte Carlo simulations to radiation dosimetry Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry N. Gopishankar, 1a

More information

Formaldehyde increases MAGIC gel dosimeter melting point and sensitivity

Formaldehyde increases MAGIC gel dosimeter melting point and sensitivity Journal of Physics: Conference Series Formaldehyde increases MAGIC gel dosimeter melting point and sensitivity To cite this article: Juliana P Fernandes et al 2009 J. Phys.: Conf. Ser. 164 012004 Recent

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife

Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife Conflicts of Interest Strategies and Technologies for Cranial Radiosurgery Planning: Gamma Knife David Schlesinger, Ph.D. Research support: Elekta Instruments, AB Lars Leksell Gamma Knife Center University

More information

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD

DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies

Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies J. Y. C. Cheung and K. N. Yu a Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong,

More information

How to scan polymer gels with MRI?

How to scan polymer gels with MRI? Home Search Collections Journals About Contact us My IOPscience How to scan polymer gels with MRI? This content has been downloaded from IOPscience. Please scroll down to see the full text. 2013 J. Phys.:

More information

VIP. MRI magnetic resonance imaging. Keywords: PHITS, gel dosimeter, three dimensional dose distribution, LET dependence, dose verification MRI 2)

VIP. MRI magnetic resonance imaging. Keywords: PHITS, gel dosimeter, three dimensional dose distribution, LET dependence, dose verification MRI 2) ,,,,,,, We study the radiological characteristics of VIP polymer gel dosimeters under 135 AMeV and 290 AMeV carbon beam irradiation. We examine the transverse (or spin-spin) relaxation rate R 2 of the

More information

High Resolution Polymer Gel Dosimetry for Small and Micro Field Dosimetry, and Development of Innovative Polymer Gel Dosimeters

High Resolution Polymer Gel Dosimetry for Small and Micro Field Dosimetry, and Development of Innovative Polymer Gel Dosimeters High Resolution Polymer Gel Dosimetry for Small and Micro Field Dosimetry, and Development of Innovative Polymer Gel Dosimeters A thesis submitted in fulfilment of the requirements for the degree of Doctor

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

IMRT point dose measurements with a diamond detector

IMRT point dose measurements with a diamond detector IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,

More information

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy Sofie Ceberg, PhD Medical Physicist and Lund University Lund, Sweden Learning objectives 1: 3D Dosimetry in the Clinic: Background

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy Guidelines for Planning and Irradiating the RPC Liver Phantom. Revised Dec 2005 Credentialing for this protocol requires four steps:

More information

Effect of skull contours on dose calculations in Gamma Knife Perfexion stereotactic radiosurgery

Effect of skull contours on dose calculations in Gamma Knife Perfexion stereotactic radiosurgery JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 Effect of skull contours on dose calculations in Gamma Knife Perfexion stereotactic radiosurgery Hisato Nakazawa, 1,2a Masataka Komori,

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Y FILMS DOSIMETR Nederland België / Belgique

Y FILMS DOSIMETR Nederland België / Belgique DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

Verification of micro-beam irradiation

Verification of micro-beam irradiation Journal of Physics: Conference Series OPEN ACCESS Verification of micro-beam irradiation To cite this article: Qiongge Li et al 2015 J. Phys.: Conf. Ser. 573 012047 View the article online for updates

More information

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

Research Note CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY * Iranian Journal of Science & Technology, Transaction A, Vol. 32, No. A3 Printed in the Islamic Republic of Iran, 2008 Shiraz University "Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 :

More information

Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry

Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry Original Article 3 Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry Timothy Olding 1, Oliver Holmes 1, Paul DeJean 1, Kim B. McAuley 2, Ken

More information

Gel dosimetry: The effect of gold nanoparticles on the dose enhancement in the external radiation therapy

Gel dosimetry: The effect of gold nanoparticles on the dose enhancement in the external radiation therapy RESEARCH ARTICLE Gel dosimetry: The effect of gold nanoparticles on the dose enhancement in the external radiation therapy Hossein Khosravi 1 *, Bijan Hashemi 1, Faezeh Rahmani 1 Department of Medical

More information

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

X-Ray Guided Robotic Radiosurgery for Solid Tumors

X-Ray Guided Robotic Radiosurgery for Solid Tumors X-Ray Guided Robotic Radiosurgery for Solid Tumors Mohan Bodduluri Accuray Incorporated 570 Del Rey Avenue Sunnyvale, CA 94085 USA and J. M. McCarthy Department of Mechanical and Aerospace Engineering

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Comparing the results of 3D treatment planning and practical dosimetry in craniospinal radiotherapy using Rando phantom

Comparing the results of 3D treatment planning and practical dosimetry in craniospinal radiotherapy using Rando phantom Iran. J. Radiat. Res., 2011; 9(3): 151-158 Comparing the results of 3D treatment planning and practical dosimetry in craniospinal radiotherapy using Rando phantom H.R. Baghani 1*, S.M.R. Aghamiri 2, H.

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Chapters from Clinical Oncology

Chapters from Clinical Oncology Chapters from Clinical Oncology Lecture notes University of Szeged Faculty of Medicine Department of Oncotherapy 2012. 1 RADIOTHERAPY Technical aspects Dr. Elemér Szil Introduction There are three possibilities

More information

Review of TG-186 recommendations

Review of TG-186 recommendations Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP Iran. J. Radiat. Res., 2006; 4 (1): 15-19 Dose distribution and dosimetry parameters calculation of MED3633 palladium- source in water phantom using MCNP A.A. Mowlavi 1*,A. Binesh 2, H. Moslehitabar 3

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

Hampton University Proton Therapy Institute

Hampton University Proton Therapy Institute Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

The use of the BANG-3 polymer gel to quantify the three-dimensional dose distribution of IMRT

The use of the BANG-3 polymer gel to quantify the three-dimensional dose distribution of IMRT Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2003 The use of the BANG-3 polymer gel to quantify the three-dimensional dose distribution of IMRT Paul Anthony Bruce

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Transition to Heterogeneity Corrections. Why have accurate dose algorithms? Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY Translated English of Chinese Standard: YY0831-2011 www.chinesestandard.net Email: Sales@ChineseStandard.net ICS 11.040.50 YY C 43 PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA YY

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

Transmission portal in vivo dosimetry by means of the Monte Carlo method and the mathematical programming language MATLAB

Transmission portal in vivo dosimetry by means of the Monte Carlo method and the mathematical programming language MATLAB Radioprotection 49(3), 205-211 (2014) c EDP Sciences 2014 DOI: 10.1051/radiopro/2014008 Available online at: www.radioprotection.org Article Transmission portal in vivo dosimetry by means of the Monte

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information