Analysis of the results from the implementation of quality control program in mammography in Bulgaria
|
|
- Giles Lyons
- 5 years ago
- Views:
Transcription
1 Analysis of the results from the implementation of quality control program in mammography in Bulgaria *, J. Vassileva, A. Dimov National Centre of Radiobiology and Radiation Protection, Sofia, Bulgaria *Presenting author:
2 Introduction The Bulgarian legislation adopted the requirements of the European protocol for the quality control of the physical and technical aspects of mammography screening in Since then quality control of the mammography units in the country is mandatory. Initial experience in this kind of measurements is gained.
3 Purpose The purpose of this work is to propose optimization measures based on the comprehensive analysis of the Quality control (QC) program implemented in several mammography rooms in Bulgaria.
4 Materials and methods The methods described in the national QC protocol for mammography adopting the European protocol were applied. Measurements were performed on 12 typical for the country film-screen mammography units: 3 Mammomat (Siemens) 2 Melody (Villa) 1 Flat E (Metaltronica) 1 Italray (Italray) 1 Giotto ST (IMS) 2 Affinity Lorad (Hologic) 1 Mammo Diagnost UC (Philips) 1 D-240/DG-40 (TuR)
5 Results % of units outside limits Remedial Suspension Tube voltage accuracy / 25% Central value 45% 18% of optical density control setting Difference per step 77% / of optical density control setting Object thickness and tube voltage 14% 43% compensation of AEC Short term reproducibility of AEC 8% 8%
6 Results % of units outside limits Remedial Suspension Light tightness of darkroom 60% / Extra fog from safelights 38% / Viewing box luminance / 100%! Viewing box homogeneity 13% 38% Ambient light level 25% /
7 Results AEC long term reproducibility for 3.5 years on unit Flat E Optical density Date Stable long term reproducibility, fluctuations are due to film processing
8 Results Film processing for 3.5 years on unit Flat E Speed index OD1 Date Big fluctuations of the sensitometry indexes mainly due to the slow processor replenishment rate
9 Conclusions The old mammography units type TuR D-240/DG-40 were suspended from clinical use due to the constructive limitations to cover the new requirements. Corrective actions were recommended to optimize the settings of some systems and to improve the practice (such as AEC optical density settings, film processor settings, decrease of ambient light level of the darkrooms, change of viewing boxes). Improvement was achieved in the performance of some of the mammography units.
10 Conclusions Implementation of the QC program is still ongoing process that needs good understanding and training of the clinical staff. This is particularly important because of the planed introduction of mammography screening in the country in near future.
Survey of Mammography Practice: Initial Results from Serbia
Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental
More informationPilot study of patient and phantom breast dose measurements in Bulgaria
Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast
More information1. Introduction. 2. Technical procedures
MAMMOGRAPHY QUALITY ASSURANCE IN MOROCCO i inn mill IT -7 v T i * ^ T i XA0101632 F. Zaoui, K. Talsmat, K. Lalaoui Departement des Applications Medicales et Biologiques, Centre National del'energie des
More informationUpdate on MQSA and Mammography Accreditation
MQSA - Who s Who Update on MQSA and Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Priscilla F. Butler, M.S. Senior Director,
More informationLorad Stereotactic Manual READ ONLINE
Lorad Stereotactic Manual READ ONLINE If you are searching for a ebook Lorad stereotactic manual in pdf form, in that case you come on to the right website. We furnish full variant of this book in DjVu,
More informationIAEA TECDOC 1517 Quality Control in Mammography Software
IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationIntroduction 1. Executive Summary 5
Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local
More informationDisclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.
Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives
More informationDigital Breast Tomosynthesis from a first idea to clinical routine
International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen
More informationAbout me 10/9/2018. Digital Breast Tomosynthesis (DBT)
About me Digital Breast Tomosynthesis (DBT) Deborah Thames, BSRS R.T.(R)(M)(QM) Advanced Health Education Center www.aheconline.com Work full time at a cancer hospital in Houston, Texas, born in South
More informationEffects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan
Quality Assurance in Mammography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Yi-Shuan Hwang 1 Hui-Yu Tsai 2 Chien-Chuan Chen 1 Pei-Kwei Tsay 3 Huay-Ben
More informationQuality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio
Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz
More informationHong Kong College of Radiologists Mammography Statement
Hong Kong College of Radiologists Mammography Statement The Hong Kong College of Radiologists would like to give the following comments concerning mammography. Mammography screening: Breast cancer is the
More informationQuality control in mammography: an initiative in France
f 1994, The British Journal of Radiology, 67, 371-383 Quality control in mammography: an initiative in France 1 C MACCIA, PhD, 2 R RENAUD, MD, 1 S CASTELLANO, PhD, 2 P SCHAFFER, MD, PhD, 2 R WAHL, MD,
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, Mail Code 25-01
More informationRadiographic techniques in screen-film mammography
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda
More informationThe relationship of the mean glandular dose with compressed breast thickness in mammography
Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu
More informationEstablishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology
Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics
More informationMammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433
- 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.
More informationCURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS
ABBY DUROJAYE,M.D CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS BOTH HAVE BEEN SHOWN TO BE EFFECTIVE TOOLS EARLY DETECTION OF BREAST
More informationMaintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition
Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition Sandra S. Rao, M.D. Northwestern Medicine I have nothing to disclose. Objectives: Review
More informationMQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR
2027 North 36 th Street Saint Joseph, Mo 64506 (86) 390-90 (800) 306-4477 MQSA Physicist Qualifications Jon J. Erickson, Ph.D., DABR Contents Missouri Mammography Physicist Initial Qualifications Certificate
More informationExperience with Tomosynthesis (3-D) Mammography in a Mobile Setting and Current Controversies
Experience with Tomosynthesis (3-D) Mammography in a Mobile Setting and Current Controversies Jennifer Lance, MHSA, MS Jerry McLarty, PhD LSU Feist-Weiller Cancer Center Shreveport LA Sept 15, 2008 A brief
More informationThe New Jersey Radiographic Quality Assurance Program at 5 Years
The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided
More informationProven clinical effectiveness at low radiation dose
MicroDose Mammography Solutions Proven clinical effectiveness at low radiation dose Several studies provide evidence that Philips MicroDose Mammography* can provide outstanding image quality at 18% to
More informationUpdate of Digital Breast Tomosynthesis. Susan Orel Roth, MD
Update of Digital Breast Tomosynthesis Susan Orel Roth, MD NCI estimates that : Why DBT? Approximately 20% of breast cancers are missed at mammography screening Average recall rates approximately 10%
More informationPATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME
Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY
More informationStraight to the point with your diagnosis.
CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS HELIODENT DS INTRAORAL X-RAY WITH EXCELLENT IMAGE QUALITY Straight to the point with your diagnosis. T h e D e n t a l C o
More informationImpact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy
More informationBREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0
BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2
More informationImplementation of a New Tomosynthesis Program: A Physicists Perspective
Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate
More informationEvaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics
Volume 2 Number 1 Winter Issue, 2001 Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics Abstract The purpose of the study was to compare the segmental measurement tool from
More informationPatient Dosimetry in Mammography and Tomosynthesis:
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor
More informationPOLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA
POLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA The radiation use policy has been in effect since November 1981. This revised
More informationDental Intraoral X-ray Systems
Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.
More informationFilm processing procedures have a direct effect on the quality of a radiograph.
. Film processing procedures have a direct effect on the quality of a radiograph. Processing is a series of steps that changes the latent image on the exposed film into a radiograph by producing a visible
More informationPhD Title: Improving QA/QC in mammography screening and breast in Montenegro
3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationEvaluation of Doses and Image Quality in Mammography with Screen-Film, CR, and DR Detectors Application of the ACR Phantom
Signature: Pol J Radiol, 1; 1: -91 DOI: 1.159/PJR.97 ORIGINAL ARTICLE Received: 15.1. Accepted: 1.1.1 Published: 1..1 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data
More informationKish chakrabarti, Ph.D. Senior Physicist CDRH/FDA
Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist
More informationQuality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis
60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis
More informationChapter 10: Special Topics
Chapter 10: Special Topics Slide set of 71 slides based on the chapter authored by D. McLean and J. Shepherd of the publication (ISBN 978-92-0-131010-1): Diagnostic Radiology Physics: A Handbook for Teachers
More informationBICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:
BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:
More informationMedical Physics in the Breast Screening Programme Kath Schofield & Anne McCurrach NSS Medical Physics
Medical Physics in the Breast Screening Programme Kath Schofield & Anne McCurrach NSS Medical Physics COR103-012 V1 Who we are John Wright Assistant Director Facilities Services MPE Medical Physics Expert
More informationUpdate on the ACR FFDM QC Manual
Update on the ACR FFDM QC Manual Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO AAPM 2011 Vancouver, Canada 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000
More informationBreast phantom with silicone implant for evaluation in conventional mammography
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, winter 2011 Breast phantom with silicone implant for evaluation in conventional mammography Fábio A. R. Silva, 1a Luíza F Souza, 1 Carlos
More information3980-DM Digital Mammography
Turnkey Laboratories Solutions 3980-DM Digital Mammography EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES Analytical Technologies Limited An ISO 9001 Certified Company www.analyticalgroup.net
More information-Ray Producing Equipment
Temple University R adiation Safety Guide For Diagnostic Imaging X -Ray Producing Equipment Environmental Health And Safety Radiation Safety Department Table of Content Page 1. Radiation 2 A. Radiation
More informationCharacteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography
Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography Norimitsu SHINOHARA* a, Hiroshi FUJITA b, Takeshi HARA b, and Katsuhei HORITA c a Electronic
More informationDental Extraoral X-ray Systems
Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental
More informationMR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015
MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute
More informationMammography calibration qualities establishment in a Mo- Mo clinical system
Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733
More informationA STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY
Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH
More informationDIAGNOSTIC IMAGING (DMI)
Diagnostic Imaging (DMI) 1 DIAGNOSTIC IMAGING (DMI) DMI 101 Introduction to Medical Imaging Introduction to Diagnostic Medical Imaging modalities with a special emphasis on Radiologic Technology and Diagnostic
More informationA Standardised Approach to Optimisation
The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained
More informationExamination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA
Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi
More informationOverview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS
ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements
More informationDÜRR X-RAY FILM PROCESSING. The development goes on and on DÜRR XR 24 PRO
DÜRR X-RAY FILM PROCESSING The development goes on and on From a professional picture to an exact diagnosis DÜRR X-RAY FILM PROCESSING Facts in black and white Causal dental medicine presupposes an exact
More informationMethod for determination of the mean fraction of glandular tissue in individual female breasts using mammography
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationLevel 3 Diploma in Dental Nursing
LearnerName CompanyName AssessorName Date Submitted Level 3 Diploma in Dental Nursing - 5234 Unit 314 Dental Radiography Below contains all areas that make up the 314 dental exam, please be aware of all
More informationRadiation Protection in Dental Intraoral Radiography Edition
Radiation Protection In Intraoral Dental Radiography is one of a series of training CDs designed for staff that use or are exposed to radiation in their daily work. It would be useful to take the Basic
More informationBasic tests in mammography as a tool in quality improvement
reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 145 152 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Technical note Basic tests in mammography as a tool in quality
More informationMammographic Accreditation Phantom
Quality Control in Mammography RMI 156 Mammographic Accreditation Phantom RMI 156 A User s Guide GAMMEX RMI Table of Contents Product Overview...................... 1 Product Description....................
More informationTHE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES
IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele
More informationACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program
ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new
More informationCT Quality Control Manual FAQs
CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues
More informationQ: Why is breast cancer a big deal?
I hate breast cancer. As a radiologist who specializes in breast imaging, my career is devoted to the detection and diagnosis of breast cancer. I am passionate about women s health and my goal is to find
More informationPatient dose measurement in common medical X-ray examinations in Iran
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Patient dose measurement in common medical X-ray examinations in Iran Behrouz Rasuli, 1a Ali Mahmoud-Pashazadeh, 2 Mohammad Ghorbani,
More informationRadiation Physics Consultants, Inc. Medical Physicist. Mammography Qualifications, Continuing Experience and Continuing Education
Radiation Physics Consultants, Inc. Mammography Qualifications, Continuing Experience and Continuing Education For: Steven T. Nicholas, M.S. DABMP MAP ID s. Each medical physicist who provides medical
More informationMedical Diagnostic Imaging
Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique
More informationAbsorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry
Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,
More informationReceptor dose and Patient Dose in Radiographic exposures - a 15 year review
Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County
More informationABSTRACT I. INTRODUCTION
2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography
More informationImpact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography
Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.
More informationNEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS GENERAL REQUIREMENTS
CHRIS CHRISTIE DEPARTMENT OF ENVIRONMENTAL PROTECTION BOB MARTIN Governor Division of Environmental Safety and Health Commissioner Bureau of X-Ray Compliance, Technologist Certification Section KIM GUADAGNO
More informationCHAPTER ONE: EXECUTIVE SUMMARY
CHAPTER ONE: EXECUTIVE SUMMARY X-Ray Analysis Digital X-Ray Evolution Boosting Efficiency Significant Uses Chest X-Rays Economics Cardiovascular Disease Market Analysis Traditional X-Ray Digital X-Ray
More informationVolume 14 - Issue 3, Matrix
Volume 14 - Issue 3, 2014 - Matrix Digital Breast Tomosynthesis for Screening and Diagnosis of Breast Cancer Author ECRI ECRI Institute 29 Broadwater Road Suite 104 Welwyn Garden City AL7 3BQ United Kingdom
More informationA Methodology for Developing Speed and Endurance of Teenagers
Journal of Sports Science 3 (2015) 223-227 doi: 10.17265/2332-7839/2015.05.004 D DAVID PUBLISHING A Methodology for Developing Speed and Endurance of Teenagers Velizar T. Pavlov 1 and Veneta Simeonova
More informationEffect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis
Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography
More informationThe American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)
The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red) Table of Contents General... 1 Applicability... 3 Transitioning
More informationPlastic scintillation detectors for dose monitoring in digital breast tomosynthesis
Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis J. Antunes 1,2, J. Machado 1,2, L. Peralta 1,2, N. Matela 1,3 1 Departamento de Física da Faculdade de Ciências da Universidade
More informationBreast CT and Dosimetry
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of
More informationFinancial Disclosures
Financial Disclosures 3D Mammography: The Latest Developments in the Breast Imaging Arena I have no financial disclosures Dr. Katharine Lampen-Sachar Breast and Body Radiologist Radiology Associates of
More informationMultilayer Gafchromic film detectors for breast skin dose determination in vivo
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination
More information#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis =
#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? January K. Lopez, MD Hoag Breast Care Center Newport Beach, CA Disclosures: None TERMS (AKA) Digital breast tomosynthesis = DBT Tomo 3D Full
More informationDental Radiography
Western Technical College 10508103 Dental Radiography Course Outcome Summary Course Information Description Career Cluster Instructional Level Total Credits 2.00 Prepares dental auxiliary students to operate
More informationDisclosure. Objectives 2/6/ Allina Health System. What Every Technologist Would Like Their Radiologists To Know
What Every Technologist Would Like Their Radiologists To Know Louise C. Miller, RTRM Director of Education Mammography Educators San Diego, CA February 6, 2016 Disclosure There are no conflicts of interest
More informationComparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose
Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman
More informationThe wonderful world of dental radiology
The wonderful world of dental radiology Dr Christine Hawke BSc(Vet)(Hons) BVSc(Hons) PhD MACVSc (Veterinary Dentistry) PO Box 3001 Willoughby North 2068 Phone: 0408 782 611 Email: christine@sydneypetdentistry.com.au
More informationQ. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?
The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Revised: September 7, 2017; updated questions in red) Table of Contents Application - General...
More informationRadiologic Technology Training Program
Radiologic Technology Training Program 418 Radiologic Technology Training Program Radiologic Technology Training Program Program Coordinator: Mansour, Zepure General Information The Radiologic Technology
More information3D Dosimetry with Polymer Gel
3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer
More informationScreening with Abbreviated Breast MRI (AB-MR)
Screening with Abbreviated Breast MRI (AB-MR) Christopher Comstock, MD, FACR, FSBI Department of Radiology Memorial Sloan-Kettering Cancer Center New York, NY Outline History of our approach to screening
More informationPitfalls and Remedies in PET/CT imaging for RT planning
Pitfalls and Remedies in PET/CT imaging for RT planning Tinsu Pan, Ph.D. M.D. Anderson Cancer Center The University of Texas Outlines Background Average CT (< 1 msv) to reduce mis-alignment of PET and
More informationTEMPLATE PORTFOLIO FOR THE REGIONAL CLINICAL TRAINING PROGRAMME IN RADIOTHERAPY MEDICAL PHYSICS
RAF/6/044 STRENGTHENING MEDICAL PHYSICS IN SUPPORT OF CANCER MANAGEMENT - PHASE II TEMPLATE PORTFOLIO FOR THE REGIONAL CLINICAL TRAINING PROGRAMME IN RADIOTHERAPY MEDICAL PHYSICS REPORT OF A TASK FORCE
More informationA comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV
A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical
More information