Analysis of the results from the implementation of quality control program in mammography in Bulgaria

Size: px
Start display at page:

Download "Analysis of the results from the implementation of quality control program in mammography in Bulgaria"

Transcription

1 Analysis of the results from the implementation of quality control program in mammography in Bulgaria *, J. Vassileva, A. Dimov National Centre of Radiobiology and Radiation Protection, Sofia, Bulgaria *Presenting author:

2 Introduction The Bulgarian legislation adopted the requirements of the European protocol for the quality control of the physical and technical aspects of mammography screening in Since then quality control of the mammography units in the country is mandatory. Initial experience in this kind of measurements is gained.

3 Purpose The purpose of this work is to propose optimization measures based on the comprehensive analysis of the Quality control (QC) program implemented in several mammography rooms in Bulgaria.

4 Materials and methods The methods described in the national QC protocol for mammography adopting the European protocol were applied. Measurements were performed on 12 typical for the country film-screen mammography units: 3 Mammomat (Siemens) 2 Melody (Villa) 1 Flat E (Metaltronica) 1 Italray (Italray) 1 Giotto ST (IMS) 2 Affinity Lorad (Hologic) 1 Mammo Diagnost UC (Philips) 1 D-240/DG-40 (TuR)

5 Results % of units outside limits Remedial Suspension Tube voltage accuracy / 25% Central value 45% 18% of optical density control setting Difference per step 77% / of optical density control setting Object thickness and tube voltage 14% 43% compensation of AEC Short term reproducibility of AEC 8% 8%

6 Results % of units outside limits Remedial Suspension Light tightness of darkroom 60% / Extra fog from safelights 38% / Viewing box luminance / 100%! Viewing box homogeneity 13% 38% Ambient light level 25% /

7 Results AEC long term reproducibility for 3.5 years on unit Flat E Optical density Date Stable long term reproducibility, fluctuations are due to film processing

8 Results Film processing for 3.5 years on unit Flat E Speed index OD1 Date Big fluctuations of the sensitometry indexes mainly due to the slow processor replenishment rate

9 Conclusions The old mammography units type TuR D-240/DG-40 were suspended from clinical use due to the constructive limitations to cover the new requirements. Corrective actions were recommended to optimize the settings of some systems and to improve the practice (such as AEC optical density settings, film processor settings, decrease of ambient light level of the darkrooms, change of viewing boxes). Improvement was achieved in the performance of some of the mammography units.

10 Conclusions Implementation of the QC program is still ongoing process that needs good understanding and training of the clinical staff. This is particularly important because of the planed introduction of mammography screening in the country in near future.

Survey of Mammography Practice: Initial Results from Serbia

Survey of Mammography Practice: Initial Results from Serbia Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

1. Introduction. 2. Technical procedures

1. Introduction. 2. Technical procedures MAMMOGRAPHY QUALITY ASSURANCE IN MOROCCO i inn mill IT -7 v T i * ^ T i XA0101632 F. Zaoui, K. Talsmat, K. Lalaoui Departement des Applications Medicales et Biologiques, Centre National del'energie des

More information

Update on MQSA and Mammography Accreditation

Update on MQSA and Mammography Accreditation MQSA - Who s Who Update on MQSA and Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Priscilla F. Butler, M.S. Senior Director,

More information

Lorad Stereotactic Manual READ ONLINE

Lorad Stereotactic Manual READ ONLINE Lorad Stereotactic Manual READ ONLINE If you are searching for a ebook Lorad stereotactic manual in pdf form, in that case you come on to the right website. We furnish full variant of this book in DjVu,

More information

IAEA TECDOC 1517 Quality Control in Mammography Software

IAEA TECDOC 1517 Quality Control in Mammography Software IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Introduction 1. Executive Summary 5

Introduction 1. Executive Summary 5 Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local

More information

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose. Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives

More information

Digital Breast Tomosynthesis from a first idea to clinical routine

Digital Breast Tomosynthesis from a first idea to clinical routine International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen

More information

About me 10/9/2018. Digital Breast Tomosynthesis (DBT)

About me 10/9/2018. Digital Breast Tomosynthesis (DBT) About me Digital Breast Tomosynthesis (DBT) Deborah Thames, BSRS R.T.(R)(M)(QM) Advanced Health Education Center www.aheconline.com Work full time at a cancer hospital in Houston, Texas, born in South

More information

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan Quality Assurance in Mammography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Yi-Shuan Hwang 1 Hui-Yu Tsai 2 Chien-Chuan Chen 1 Pei-Kwei Tsay 3 Huay-Ben

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

Hong Kong College of Radiologists Mammography Statement

Hong Kong College of Radiologists Mammography Statement Hong Kong College of Radiologists Mammography Statement The Hong Kong College of Radiologists would like to give the following comments concerning mammography. Mammography screening: Breast cancer is the

More information

Quality control in mammography: an initiative in France

Quality control in mammography: an initiative in France f 1994, The British Journal of Radiology, 67, 371-383 Quality control in mammography: an initiative in France 1 C MACCIA, PhD, 2 R RENAUD, MD, 1 S CASTELLANO, PhD, 2 P SCHAFFER, MD, PhD, 2 R WAHL, MD,

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, Mail Code 25-01

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

The relationship of the mean glandular dose with compressed breast thickness in mammography

The relationship of the mean glandular dose with compressed breast thickness in mammography Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS ABBY DUROJAYE,M.D CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS BOTH HAVE BEEN SHOWN TO BE EFFECTIVE TOOLS EARLY DETECTION OF BREAST

More information

Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition

Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition Sandra S. Rao, M.D. Northwestern Medicine I have nothing to disclose. Objectives: Review

More information

MQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR

MQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR 2027 North 36 th Street Saint Joseph, Mo 64506 (86) 390-90 (800) 306-4477 MQSA Physicist Qualifications Jon J. Erickson, Ph.D., DABR Contents Missouri Mammography Physicist Initial Qualifications Certificate

More information

Experience with Tomosynthesis (3-D) Mammography in a Mobile Setting and Current Controversies

Experience with Tomosynthesis (3-D) Mammography in a Mobile Setting and Current Controversies Experience with Tomosynthesis (3-D) Mammography in a Mobile Setting and Current Controversies Jennifer Lance, MHSA, MS Jerry McLarty, PhD LSU Feist-Weiller Cancer Center Shreveport LA Sept 15, 2008 A brief

More information

The New Jersey Radiographic Quality Assurance Program at 5 Years

The New Jersey Radiographic Quality Assurance Program at 5 Years The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided

More information

Proven clinical effectiveness at low radiation dose

Proven clinical effectiveness at low radiation dose MicroDose Mammography Solutions Proven clinical effectiveness at low radiation dose Several studies provide evidence that Philips MicroDose Mammography* can provide outstanding image quality at 18% to

More information

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD Update of Digital Breast Tomosynthesis Susan Orel Roth, MD NCI estimates that : Why DBT? Approximately 20% of breast cancers are missed at mammography screening Average recall rates approximately 10%

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

Straight to the point with your diagnosis.

Straight to the point with your diagnosis. CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS HELIODENT DS INTRAORAL X-RAY WITH EXCELLENT IMAGE QUALITY Straight to the point with your diagnosis. T h e D e n t a l C o

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of a New Tomosynthesis Program: A Physicists Perspective Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate

More information

Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics

Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics Volume 2 Number 1 Winter Issue, 2001 Evaluation of a Digital Measurement Tool to Estimate Working Length in Endodontics Abstract The purpose of the study was to compare the segmental measurement tool from

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

POLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA

POLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA POLICY FOR THE USE AND CONTROL OF IONIZING RADIATION FOR DIAGNOSTIC IMAGING AT THE SCHOOL OF DENTISTRY, UNIVERSITY OF MINNESOTA The radiation use policy has been in effect since November 1981. This revised

More information

Dental Intraoral X-ray Systems

Dental Intraoral X-ray Systems Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.

More information

Film processing procedures have a direct effect on the quality of a radiograph.

Film processing procedures have a direct effect on the quality of a radiograph. . Film processing procedures have a direct effect on the quality of a radiograph. Processing is a series of steps that changes the latent image on the exposed film into a radiograph by producing a visible

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Evaluation of Doses and Image Quality in Mammography with Screen-Film, CR, and DR Detectors Application of the ACR Phantom

Evaluation of Doses and Image Quality in Mammography with Screen-Film, CR, and DR Detectors Application of the ACR Phantom Signature: Pol J Radiol, 1; 1: -91 DOI: 1.159/PJR.97 ORIGINAL ARTICLE Received: 15.1. Accepted: 1.1.1 Published: 1..1 Authors Contribution: A Study Design B Data Collection C Statistical Analysis D Data

More information

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist

More information

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis 60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

More information

Chapter 10: Special Topics

Chapter 10: Special Topics Chapter 10: Special Topics Slide set of 71 slides based on the chapter authored by D. McLean and J. Shepherd of the publication (ISBN 978-92-0-131010-1): Diagnostic Radiology Physics: A Handbook for Teachers

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Medical Physics in the Breast Screening Programme Kath Schofield & Anne McCurrach NSS Medical Physics

Medical Physics in the Breast Screening Programme Kath Schofield & Anne McCurrach NSS Medical Physics Medical Physics in the Breast Screening Programme Kath Schofield & Anne McCurrach NSS Medical Physics COR103-012 V1 Who we are John Wright Assistant Director Facilities Services MPE Medical Physics Expert

More information

Update on the ACR FFDM QC Manual

Update on the ACR FFDM QC Manual Update on the ACR FFDM QC Manual Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO AAPM 2011 Vancouver, Canada 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000

More information

Breast phantom with silicone implant for evaluation in conventional mammography

Breast phantom with silicone implant for evaluation in conventional mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, winter 2011 Breast phantom with silicone implant for evaluation in conventional mammography Fábio A. R. Silva, 1a Luíza F Souza, 1 Carlos

More information

3980-DM Digital Mammography

3980-DM Digital Mammography Turnkey Laboratories Solutions 3980-DM Digital Mammography EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES Analytical Technologies Limited An ISO 9001 Certified Company www.analyticalgroup.net

More information

-Ray Producing Equipment

-Ray Producing Equipment Temple University R adiation Safety Guide For Diagnostic Imaging X -Ray Producing Equipment Environmental Health And Safety Radiation Safety Department Table of Content Page 1. Radiation 2 A. Radiation

More information

Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography

Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography Norimitsu SHINOHARA* a, Hiroshi FUJITA b, Takeshi HARA b, and Katsuhei HORITA c a Electronic

More information

Dental Extraoral X-ray Systems

Dental Extraoral X-ray Systems Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental

More information

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015 MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH

More information

DIAGNOSTIC IMAGING (DMI)

DIAGNOSTIC IMAGING (DMI) Diagnostic Imaging (DMI) 1 DIAGNOSTIC IMAGING (DMI) DMI 101 Introduction to Medical Imaging Introduction to Diagnostic Medical Imaging modalities with a special emphasis on Radiologic Technology and Diagnostic

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements

More information

DÜRR X-RAY FILM PROCESSING. The development goes on and on DÜRR XR 24 PRO

DÜRR X-RAY FILM PROCESSING. The development goes on and on DÜRR XR 24 PRO DÜRR X-RAY FILM PROCESSING The development goes on and on From a professional picture to an exact diagnosis DÜRR X-RAY FILM PROCESSING Facts in black and white Causal dental medicine presupposes an exact

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Level 3 Diploma in Dental Nursing

Level 3 Diploma in Dental Nursing LearnerName CompanyName AssessorName Date Submitted Level 3 Diploma in Dental Nursing - 5234 Unit 314 Dental Radiography Below contains all areas that make up the 314 dental exam, please be aware of all

More information

Radiation Protection in Dental Intraoral Radiography Edition

Radiation Protection in Dental Intraoral Radiography Edition Radiation Protection In Intraoral Dental Radiography is one of a series of training CDs designed for staff that use or are exposed to radiation in their daily work. It would be useful to take the Basic

More information

Basic tests in mammography as a tool in quality improvement

Basic tests in mammography as a tool in quality improvement reports of practical oncology and radiotherapy 1 5 (2 0 1 0) 145 152 available at www.sciencedirect.com journal homepage: http://www.rpor.eu/ Technical note Basic tests in mammography as a tool in quality

More information

Mammographic Accreditation Phantom

Mammographic Accreditation Phantom Quality Control in Mammography RMI 156 Mammographic Accreditation Phantom RMI 156 A User s Guide GAMMEX RMI Table of Contents Product Overview...................... 1 Product Description....................

More information

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele

More information

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Q: Why is breast cancer a big deal?

Q: Why is breast cancer a big deal? I hate breast cancer. As a radiologist who specializes in breast imaging, my career is devoted to the detection and diagnosis of breast cancer. I am passionate about women s health and my goal is to find

More information

Patient dose measurement in common medical X-ray examinations in Iran

Patient dose measurement in common medical X-ray examinations in Iran JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Patient dose measurement in common medical X-ray examinations in Iran Behrouz Rasuli, 1a Ali Mahmoud-Pashazadeh, 2 Mohammad Ghorbani,

More information

Radiation Physics Consultants, Inc. Medical Physicist. Mammography Qualifications, Continuing Experience and Continuing Education

Radiation Physics Consultants, Inc. Medical Physicist. Mammography Qualifications, Continuing Experience and Continuing Education Radiation Physics Consultants, Inc. Mammography Qualifications, Continuing Experience and Continuing Education For: Steven T. Nicholas, M.S. DABMP MAP ID s. Each medical physicist who provides medical

More information

Medical Diagnostic Imaging

Medical Diagnostic Imaging Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.

More information

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS GENERAL REQUIREMENTS

NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) DENTAL RADIOGRAPHY CURRICULUM REQUIREMENTS GENERAL REQUIREMENTS CHRIS CHRISTIE DEPARTMENT OF ENVIRONMENTAL PROTECTION BOB MARTIN Governor Division of Environmental Safety and Health Commissioner Bureau of X-Ray Compliance, Technologist Certification Section KIM GUADAGNO

More information

CHAPTER ONE: EXECUTIVE SUMMARY

CHAPTER ONE: EXECUTIVE SUMMARY CHAPTER ONE: EXECUTIVE SUMMARY X-Ray Analysis Digital X-Ray Evolution Boosting Efficiency Significant Uses Chest X-Rays Economics Cardiovascular Disease Market Analysis Traditional X-Ray Digital X-Ray

More information

Volume 14 - Issue 3, Matrix

Volume 14 - Issue 3, Matrix Volume 14 - Issue 3, 2014 - Matrix Digital Breast Tomosynthesis for Screening and Diagnosis of Breast Cancer Author ECRI ECRI Institute 29 Broadwater Road Suite 104 Welwyn Garden City AL7 3BQ United Kingdom

More information

A Methodology for Developing Speed and Endurance of Teenagers

A Methodology for Developing Speed and Endurance of Teenagers Journal of Sports Science 3 (2015) 223-227 doi: 10.17265/2332-7839/2015.05.004 D DAVID PUBLISHING A Methodology for Developing Speed and Endurance of Teenagers Velizar T. Pavlov 1 and Veneta Simeonova

More information

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography

More information

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red) The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red) Table of Contents General... 1 Applicability... 3 Transitioning

More information

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis J. Antunes 1,2, J. Machado 1,2, L. Peralta 1,2, N. Matela 1,3 1 Departamento de Física da Faculdade de Ciências da Universidade

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Financial Disclosures

Financial Disclosures Financial Disclosures 3D Mammography: The Latest Developments in the Breast Imaging Arena I have no financial disclosures Dr. Katharine Lampen-Sachar Breast and Body Radiologist Radiology Associates of

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis =

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis = #46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? January K. Lopez, MD Hoag Breast Care Center Newport Beach, CA Disclosures: None TERMS (AKA) Digital breast tomosynthesis = DBT Tomo 3D Full

More information

Dental Radiography

Dental Radiography Western Technical College 10508103 Dental Radiography Course Outcome Summary Course Information Description Career Cluster Instructional Level Total Credits 2.00 Prepares dental auxiliary students to operate

More information

Disclosure. Objectives 2/6/ Allina Health System. What Every Technologist Would Like Their Radiologists To Know

Disclosure. Objectives 2/6/ Allina Health System. What Every Technologist Would Like Their Radiologists To Know What Every Technologist Would Like Their Radiologists To Know Louise C. Miller, RTRM Director of Education Mammography Educators San Diego, CA February 6, 2016 Disclosure There are no conflicts of interest

More information

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman

More information

The wonderful world of dental radiology

The wonderful world of dental radiology The wonderful world of dental radiology Dr Christine Hawke BSc(Vet)(Hons) BVSc(Hons) PhD MACVSc (Veterinary Dentistry) PO Box 3001 Willoughby North 2068 Phone: 0408 782 611 Email: christine@sydneypetdentistry.com.au

More information

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation? The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Revised: September 7, 2017; updated questions in red) Table of Contents Application - General...

More information

Radiologic Technology Training Program

Radiologic Technology Training Program Radiologic Technology Training Program 418 Radiologic Technology Training Program Radiologic Technology Training Program Program Coordinator: Mansour, Zepure General Information The Radiologic Technology

More information

3D Dosimetry with Polymer Gel

3D Dosimetry with Polymer Gel 3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer

More information

Screening with Abbreviated Breast MRI (AB-MR)

Screening with Abbreviated Breast MRI (AB-MR) Screening with Abbreviated Breast MRI (AB-MR) Christopher Comstock, MD, FACR, FSBI Department of Radiology Memorial Sloan-Kettering Cancer Center New York, NY Outline History of our approach to screening

More information

Pitfalls and Remedies in PET/CT imaging for RT planning

Pitfalls and Remedies in PET/CT imaging for RT planning Pitfalls and Remedies in PET/CT imaging for RT planning Tinsu Pan, Ph.D. M.D. Anderson Cancer Center The University of Texas Outlines Background Average CT (< 1 msv) to reduce mis-alignment of PET and

More information

TEMPLATE PORTFOLIO FOR THE REGIONAL CLINICAL TRAINING PROGRAMME IN RADIOTHERAPY MEDICAL PHYSICS

TEMPLATE PORTFOLIO FOR THE REGIONAL CLINICAL TRAINING PROGRAMME IN RADIOTHERAPY MEDICAL PHYSICS RAF/6/044 STRENGTHENING MEDICAL PHYSICS IN SUPPORT OF CANCER MANAGEMENT - PHASE II TEMPLATE PORTFOLIO FOR THE REGIONAL CLINICAL TRAINING PROGRAMME IN RADIOTHERAPY MEDICAL PHYSICS REPORT OF A TASK FORCE

More information

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical

More information