Breast CT and Dosimetry
|
|
- Joan Oliver
- 5 years ago
- Views:
Transcription
1 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of Biomedical Engineering University of California Davis Medical Center Sacramento, California
2 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary
3 Mammography: Standard of Care CC MLO CC MLO 3
4 Breast CT Mammography ~ mm mm 230 mm 250 mm ~0.50 mm 3 70 mm 70 mm 50,000 mm 4
5 Dedicated Breast CT
6 6
7 Computed Tomographic Mammography--CTM John J. Gisvold, Philip R. Karsell, David F. Reese Mayo Clinic 1975 Fall sec scan in water bath 116 kvp, 300 mas 10 mm slice thickness 1.56 x 1.56 mm pixels 4 to 6 10 mm slices per breast 256 shades of gray (8 bit)
8 CTM: 1.56 x 1.56 x 10 mm voxels Voxel volume: over 900 times smaller Special thanks to Cynthia McCollough and John Gisvold for providing this information
9 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary
10 Computer aided design / computer aided manufacture (CAD/CAM) Albion 2003 Bodega 2007 Cambria 2011 Doheny 2012
11 Albion 2004 Bodega 2007 Cambria 2012 Doheny 2013
12 UC Davis Medical Center Albion Bodega Cambria Doheny 12
13 13
14 Cambria 2011
15 Components (Albion and Bodega) Varian 4030CB 194 mm pixels 2x2 388 mm 1024 x 768 x 30 FPS Kollmorgen Servo Motor Propulsion Bearing Angle Encoder Comet 1 kw Tube 15
16 Components (Cambria and Doheny) Varian 4030CB Yaskawa Servo Motor Propulsion Bearing Angle Encoder Varian 11.3 kw Pulsed Tube M-1500 Dexela mm pixels 26 FPS / CMOS 70 2x2 16
17 Half Cone Beam CT Geometry
18 calibration, corrections, reconstruction Flat Field Correction I corr g I I raw grain I r offset I g offset Geometric Calibration
19 FDK Reconstruction Code >100x ~42 minutes ~35 minutes ~20 seconds
20 Spatial Resolution: Modeled & Measured MTF s g( t) f A a (1 exp( t / T )) a (1 exp( t / T )) a (1 exp( t / T )) Albion & Bodega Cambria (2 x 2) [388 mm pixels] Cambria (1 x 1) [194 mm pixels] Doheny 3X Spatial Resolution 20
21 Phantom Imaging
22 Phantom Imaging
23 Phantom Imaging
24 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary
25 Patient Imaging Radiologist Observers Over 600 patients imaged at two sites
26 296 second volunteer imaged: January
27 first breast cancer imaged: January
28 True 3D Display! 28
29 2008
30 bct (no injected contrast) 30
31 PRE CONTRAST POST CONTRAST bct (with contrast) PRE CONTRAST POST CONTRAST 31
32 2010 AUC DHU = 0.87
33 mass lesion microcalcifications
34 Prospective Clinical Trial 105 patients /103 lesions (BIRADS 4 or 5) imaged on VCO mammo / tomo / CE-bCT all biopsied malignant microcalcifications 31 masses 27 Shadi Shakeri, M.D. benign total Radiologists Rated Lesions using a 0 to 10 Conspicuity Score 0 = not seen 10 = excellent one-way ANOVA with correction for multiple comparisons
35 mammo tomo CE-bCT mammo tomo CE-bCT mammo tomo CE-bCT mammo tomo CE-bCT
36 Computer (Mathematical) Observer
37
38 38
39 # observations pre-whitened matched filter computer observer (f i I i ) = dꞌ lesion absent lesion present filter dꞌ inserted lesion inserted non-lesion breast CT image (actual images are used)
40 Breast CT 46% mammo mass lesions only / results do not reflect microcalcifications
41 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary
42
43 Radiation Dose from Breast CT? size matters
44 breast circumference diameter compressed breast thickness
45 25 20 N = 200 Frequency Diameter (cm)
46 Breast Diameter (cm) r = CB Thickness (cm)
47 Mean Glandular Dose in Mammography two-view mammography dose versus CB thickness Two-View Mean Dose (mgy) Average of 4 mammo systems Compressed Breast Thickness (cm)
48 Monte Carlo Assessment of Dose Deposition monoenergetic functions 50 breast modeled as a cylinder ugy per million photons cm dia 18 cm dia SIC=45 cm FOV=20 cm 50% glandular/50% adipose X-ray Energy (kev) JM Boone, N. Shah, and T.R. Nelson, A Comprehensive Analysis of DgN CT Coefficients for Pendant-Geometry Cone-Beam Breast Computed Tomography, Medical Physics 31: (2004)
49 X-ray spectral model for bct beams 1.2e+7 1e+7 1.2e+7 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e+6 60 kv 8e+6 80 kv Photon Fluence (mm -2 ) 6e+6 4e+6 2e+6 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e kv X-ray Energy (kev) X-ray Energy (kev) X-ray Energy (kev) Attenuation (%) model meas. data JM Boone and JA Seibert, An accurate method for computer-generating tungsten anode x-ray spectra from 30 kv to 140 kv, Medical Physics 24; , Thickness (mm Al)
50 Converting m-dgn to p-dgn using spectral weighting polyenergetic functions 1.1 ugy per million photons Photon Fluence (mm -2 ) 1e+7 8e+6 6e+6 4e+6 2e X-ray Energy (kev) Dose per Kerma (mgy/mgy) cm dia 18 cm dia Tube Voltage (kvp) X-ray Energy (kev) JM Boone, N. Shah, and T.R. Nelson, A Comprehensive Analysis of DgN CT Coefficients for Pendant-Geometry Cone-Beam Breast Computed Tomography, Medical Physics 31: (2004)
51 Probability (%) A B C D DD Bra Cup Size Probability (%) A B Bra Cup Size C D DD Breast Diameter (cm)
52 Comparison of breast dimensions X = 13.4 cm = 2.0 cm Median = 13.6 cm 2008 assessment on bct images (N = 137) 2001 tape measure results (N = 200) 52
53
54 actual and modeled x-ray spectra were tuned measured values spectral model 1.2e+7 1e+7 1.2e+7 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e+6 Photon Fluence (mm -2 ) 8e+6 6e+6 4e+6 2e+6 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e X-ray Energy (kev) X-ray Energy (kev) X-ray Energy (kev)
55 X-ray system output carefully measured
56 Trans (%) 14 iso mr/mas HVL E ave (ent) E ave (exit) BH factor F/mGy
57 MGD(t) in two-view mammography Relationship between bct(d) and mammo(t)
58 MGD (mgy) as a function of air kerma at isocenter (mgy) 10 cm 0% 18 cm 0%
59 isocenter to yield same dose as 2V mammography 18 cm 16 cm 14 cm 12 cm 10 cm
60 isocenter to yield same dose as 2V mammography ~0.113 mgy / mas / = 39.3 mas
61 mas (on our scanner) to yield same dose as 2V mammography
62
63 Dose assessment repeated by Thacker/Glick Thacker S and Glick S, PMB 2004,
64 Other Considerations
65 Average Glandular Dose (mgy) kvp 100 kvp 120 kvp Compressed Breast Thickness (cm)
66 Mammography a a b = high dose area b = low dose area
67 Breast CT 26 kvp Glandular Dose (mgy) a b c a b c Position (cm)
68 Other Considerations (2) DgN s used in mammography probably underestimate dose: Skin Thickness Issue Glandular Density Spectrum Quality factor for E < 25 kev 1.3
69 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary
70 Breast CT and Dosimetry: Summary Breast CT has strong clinical potential MGD levels are comparable to mammography MC Methods modern x-ray dosimetry When dosimetry methods are not available, you have to develop them yourself!
71 Breast Tomography Project University of California Davis Acknowledgements: California BCRP 7EB-0075 California BCRP 11I-0114 R01 CA R01 EB (BRP) R01 CA (RDB) P30 CA (CCSG) Susan G. Komen Foundation University of Pittsburgh
72
Patient Dosimetry in Mammography and Tomosynthesis:
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor
More informationfor Mammography Dosimetry
Changing Perceptions and Updated Methods for Mammography Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE Professor of Radiology & Biomedical Engineering University of California Davis Sacramento,
More informationCT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center
CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics
More informationSkin Model and its impact on Digital Mammography
Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1 Outline Motivation Methodology Results
More informationCURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC
CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,
More informationPatient Dose Estimates. from CT Examinations. Patient Dose Estimates
Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,
More information8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION
Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD
More informationDosimetry in digital mammography
Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationMammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433
- 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationEffect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis
Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography
More informationToshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery
Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,
More informationESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine
ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol
More informationPilot study of patient and phantom breast dose measurements in Bulgaria
Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast
More informationHacia la imagenología tomográfica de mama
Hacia la imagenología tomográfica de mama Futuro y presente Ioannis Sechopoulos, Ph.D., DABR Advanced X ray Tomographic Imaging (AXTI) Lab Department of Radiology and Nuclear Medicine Radboud University
More informationTranslating Protocols Across Patient Size: Babies to Bariatric
Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
More informationRadiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist
Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry
More informationUsing Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography
Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,
More informationContrast Enhanced Spectral Mammography (CESM) Updates
Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da
More informationDigital Breast Tomosynthesis from a first idea to clinical routine
International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen
More informationDianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX
Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods
More information8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree
DBT Physics Basic to Advanced: Primer On Tomosynthesis Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology Acknowledgements of Support Research support from the Komen Foundation,
More informationComparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose
Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman
More informationCone-Beam CT for MSK Extremities
8/6/0 Diagnostic Image Quality Evaluation of an Extremity Cone-Beam CT Scanner: Pre-Clinical and First Clinical Results Abdullah Muhit Wojciech Zbijewski, J Webster Stayman John Yorkston, Nathan Packard,
More informationDirect half value layer measurements in mammography - is near enough good enough?
Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.
More informationIntroduction and Background
CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section
More informationState of the art and future development for standardized estimation of organ doses in CT
State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida
More informationMammography calibration qualities establishment in a Mo- Mo clinical system
Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733
More informationInvivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
More informationHead to new heights with your imaging SCANORA 3D
SCANORA 3D Head to new heights with your imaging Benefits at a glance The solution for dentomaxillofacial and ENT imaging Easy Patient seated for added stability during exposure. Clear, self-explinatory
More informationA tissue-equivalent phantom series for mammography dosimetry
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.
More informationBetter View. Getting a. Plus. enters new phase of clinical testing. Conducting Employee Surveys Radiology Master s Program Ground Transients
RT awardwinning magazine the weekly source for radiology professionals Getting a Better View Dedicated breast CT enters new phase of clinical testing Plus Conducting Employee Surveys Radiology Master s
More informationCT Dosimetry in the Clinical Environment: Methods and Analysis
CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida
More informationNote: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, conta
Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Karen K. Lindfors,
More informationGATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT
GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS
More informationCalculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer
Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity
More informationSince its introduction in 2000, digital mammography has become
Review Article Smith A, PhD email : Andrew.smith@hologic.com Since its introduction in 2000, digital mammography has become an accepted standard of care in breast cancer screening and has paved the way
More informationDual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities
Dual-Energy Imaging of Bone Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities W Zbijewski, 1 A Sisniega, 1 JW Stayman, 1 N Packard, 2 J Yorkston, 2 G Thawait, 3 S Demehri, 3 J Fritz,
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationSOMATOM Drive System Owner Manual Dosimetry and imaging performance report
www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General
More informationMeasurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method
Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.
More informationMethod for determination of the mean fraction of glandular tissue in individual female breasts using mammography
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in
More informationNational Cancer Institute
National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,
More informationEstimating Patient Radiation Dose from Computed Tomography
Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical
More informationContrast-Enhanced Digital Mammography
2015 ARRS Breast Symposium Contrast-Enhanced Digital Mammography John Lewin, M.D. Diversified Radiology of Colorado CEDM - Outline History Technique Literature Review / Cases Clinical Status Inexpensive,
More informationTo Shield or Not to Shield? Lincoln L. Berland, M.D.
To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.
More informationIMAGE GENTLY HOW CAN YOU HELP?
IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationEstablishment of the new IEC mammography qualities in a clinical system used for instruments calibration
Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas
More informationQuality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio
Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz
More informationPhD Title: Improving QA/QC in mammography screening and breast in Montenegro
3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:
More informationManaging Radiation Risk in Pediatric CT Imaging
Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist
More informationMonte Carlo Modelling: a reliable and efficient tool in radiation dosimetry
Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative
More informationLow Dose Molecular Breast Imaging
Low Dose Molecular Breast Imaging Dr. M.K. O Connor Conflict of Interest Royalties - Gamma Medica Research funding GE Healthcare Research support MTTI Michael O Connor, Ph.D Dept. of Radiology Mayo Clinic
More informationAn Introduction to Dual Energy Computed Tomography
An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the
More informationX-Ray & CT Physics / Clinical CT
Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED
More informationRadiation Dose Reduction: Should You Use a Bismuth Breast Shield?
Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT
More informationBICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives
More informationBreast tomosynthesis reduces radiologist performance variability compared to digital mammography
Breast tomosynthesis reduces radiologist performance variability compared to digital mammography Andrew Smith 1, Elizabeth Rafferty 2, Loren Niklason 1 1 Hologic, Inc., Bedford MA, USA 2 Massachusetts
More informationALARA in mammography screening. Hilde Bosmans et al.
ALARA in mammography screening Hilde Bosmans et al. ALARA in the mammography Overview 1. Screening 2. The balance: 1. Image quality 2. Radiation dose 3. In practice Screening leeftijd Mammography Screening
More informationCone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood
Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex
More informationDisclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.
Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives
More informationSkyscan 1076 in vivo scanning: X-ray dosimetry
Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More informationMdiagnosing breast by using x-ray to both breasts. It
Asian Journal of Medical and Clinical Sciences Estimation radiation risk during mammography in Sudan 1 1 1, Hifaa.Mohamed Khair, Hiba.Osman, Abdelmoneim Sulieman* 1 Al Neelain University, Faculty of Science
More informationBioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes
BioMedical quantitative X-Ray Imaging Emmanuel Brun Researcher Inserm Université Grenoble Alpes 1 Outline Introduction K-Edge Imaging Patient imaging at the European synchrotron Medical Phase Contrast
More informationFetal Dose Calculations and Impact on Patient Care
Fetal Dose Calculations and Impact on Patient Care Matt Hough, MS, DABR, DABMP Florida Hospital Diagnostic Medical Physics and Radiation Safety Resource ACR-SPR Practice Parameter for Imaging Pregnant
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationOpportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI
Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com with many thanks to Vince Polkus, Advanced Applications Product Mgr. 1 Content
More informationTesting of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution
Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi
More informationAccounting for Imaging Dose
Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some
More informationKish chakrabarti, Ph.D. Senior Physicist CDRH/FDA
Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist
More informationMammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences
1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,
More informationImplementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL
Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau
More informationProf. Dr. Doğan BOR Ankara University Institute of Nuclear Science
PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural
More informationImaging of Radiation Dose Using Cherenkov Light
Imaging of Radiation Dose Using Cherenkov Light Eric Brost 1, Yoichi Watanabe 1, Fadil Santosa 2, Adam Green 3 1 Department of Radiation Oncology, University of Minnesota 2 Institute for Mathematics and
More information8/3/2016. CAMPEP Accredited, 2016!! CHERENKOV IMAGING IN RADIOTHERAPY: 2D, 3D, REALTIME 3D, Disclosure
CHERENKOV IMAGING IN RADIOTHERAPY: 2D, 3D, REALTIME 3D, Real time IMRT Real time MRI/Co Whole breast radiotherapy Total skin therapy Brian W. Pogue PhD Engineering, Physics, Surgery, Dartmouth College
More informationImpact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography
Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.
More informationA comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:
A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research
More informationWhy is CT Dose of Interest?
Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase
More informationBREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0
BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2
More informationSenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski
SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski White Paper October 2012 SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine
More informationRadiology Review Course Hotel del Coronado Coronado, California
37 th Annual Radiology Review Course Hotel del Coronado Coronado, California Saturday, April 22, 2017 - AM TABLE OF CONTENTS Saturday, April 22, 2017 - AM 7:00 AM 7:30 AM Coffee and Pastries for Registrants
More informationPreparing for Medical Physics Components of the ABR Core Examination
Preparing for Medical Physics Components of the ABR Core Examination The ABR core examination for radiologists contains material on medical physics. This content is based on the medical physics that is
More informationPGY-1. Resident Review Session Schedule
1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT
More informationDual-Energy CT: The Technological Approaches
Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure
More information3D Dosimetry with Polymer Gel
3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer
More informationIGRT1 technologies. Paweł Kukołowicz Warsaw, Poland
IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained
More informationAccurate Accurate boost or Simply Accuboost
Accurate Accurate boost or Simply Accuboost Zoubir Ouhib MS DABR Lynn Cancer Institute of Boca Raton Community Hospital Disclosure: Advisory board Items to be discussed Big picture on management of Breast
More informationNeutron Measurements for Intensity Modulated Radiation Therapy
SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,
More informationIntroduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy
X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction
More informationImaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT
Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted
More informationBreast dosimetry system in screen/film mammography
Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationBICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:
BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:
More informationPatient effective dose evaluation for chest X-ray examination in three digital radiography centers
Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical
More informationAndrew Karellas, PhD
Advanced Imaging for Breast Cancer: Screening, Diagnosis, and Assessing Response to Therapy The Role of Tomosynthesis Andrew Karellas, PhD Department of Radiology University of Massachusetts Medical School
More informationOutcomes in the NLST. Health system infrastructure needs to implement screening
Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have
More informationTHE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM
1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter
More information