Breast CT and Dosimetry

Size: px
Start display at page:

Download "Breast CT and Dosimetry"

Transcription

1 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of Biomedical Engineering University of California Davis Medical Center Sacramento, California

2 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary

3 Mammography: Standard of Care CC MLO CC MLO 3

4 Breast CT Mammography ~ mm mm 230 mm 250 mm ~0.50 mm 3 70 mm 70 mm 50,000 mm 4

5 Dedicated Breast CT

6 6

7 Computed Tomographic Mammography--CTM John J. Gisvold, Philip R. Karsell, David F. Reese Mayo Clinic 1975 Fall sec scan in water bath 116 kvp, 300 mas 10 mm slice thickness 1.56 x 1.56 mm pixels 4 to 6 10 mm slices per breast 256 shades of gray (8 bit)

8 CTM: 1.56 x 1.56 x 10 mm voxels Voxel volume: over 900 times smaller Special thanks to Cynthia McCollough and John Gisvold for providing this information

9 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary

10 Computer aided design / computer aided manufacture (CAD/CAM) Albion 2003 Bodega 2007 Cambria 2011 Doheny 2012

11 Albion 2004 Bodega 2007 Cambria 2012 Doheny 2013

12 UC Davis Medical Center Albion Bodega Cambria Doheny 12

13 13

14 Cambria 2011

15 Components (Albion and Bodega) Varian 4030CB 194 mm pixels 2x2 388 mm 1024 x 768 x 30 FPS Kollmorgen Servo Motor Propulsion Bearing Angle Encoder Comet 1 kw Tube 15

16 Components (Cambria and Doheny) Varian 4030CB Yaskawa Servo Motor Propulsion Bearing Angle Encoder Varian 11.3 kw Pulsed Tube M-1500 Dexela mm pixels 26 FPS / CMOS 70 2x2 16

17 Half Cone Beam CT Geometry

18 calibration, corrections, reconstruction Flat Field Correction I corr g I I raw grain I r offset I g offset Geometric Calibration

19 FDK Reconstruction Code >100x ~42 minutes ~35 minutes ~20 seconds

20 Spatial Resolution: Modeled & Measured MTF s g( t) f A a (1 exp( t / T )) a (1 exp( t / T )) a (1 exp( t / T )) Albion & Bodega Cambria (2 x 2) [388 mm pixels] Cambria (1 x 1) [194 mm pixels] Doheny 3X Spatial Resolution 20

21 Phantom Imaging

22 Phantom Imaging

23 Phantom Imaging

24 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary

25 Patient Imaging Radiologist Observers Over 600 patients imaged at two sites

26 296 second volunteer imaged: January

27 first breast cancer imaged: January

28 True 3D Display! 28

29 2008

30 bct (no injected contrast) 30

31 PRE CONTRAST POST CONTRAST bct (with contrast) PRE CONTRAST POST CONTRAST 31

32 2010 AUC DHU = 0.87

33 mass lesion microcalcifications

34 Prospective Clinical Trial 105 patients /103 lesions (BIRADS 4 or 5) imaged on VCO mammo / tomo / CE-bCT all biopsied malignant microcalcifications 31 masses 27 Shadi Shakeri, M.D. benign total Radiologists Rated Lesions using a 0 to 10 Conspicuity Score 0 = not seen 10 = excellent one-way ANOVA with correction for multiple comparisons

35 mammo tomo CE-bCT mammo tomo CE-bCT mammo tomo CE-bCT mammo tomo CE-bCT

36 Computer (Mathematical) Observer

37

38 38

39 # observations pre-whitened matched filter computer observer (f i I i ) = dꞌ lesion absent lesion present filter dꞌ inserted lesion inserted non-lesion breast CT image (actual images are used)

40 Breast CT 46% mammo mass lesions only / results do not reflect microcalcifications

41 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary

42

43 Radiation Dose from Breast CT? size matters

44 breast circumference diameter compressed breast thickness

45 25 20 N = 200 Frequency Diameter (cm)

46 Breast Diameter (cm) r = CB Thickness (cm)

47 Mean Glandular Dose in Mammography two-view mammography dose versus CB thickness Two-View Mean Dose (mgy) Average of 4 mammo systems Compressed Breast Thickness (cm)

48 Monte Carlo Assessment of Dose Deposition monoenergetic functions 50 breast modeled as a cylinder ugy per million photons cm dia 18 cm dia SIC=45 cm FOV=20 cm 50% glandular/50% adipose X-ray Energy (kev) JM Boone, N. Shah, and T.R. Nelson, A Comprehensive Analysis of DgN CT Coefficients for Pendant-Geometry Cone-Beam Breast Computed Tomography, Medical Physics 31: (2004)

49 X-ray spectral model for bct beams 1.2e+7 1e+7 1.2e+7 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e+6 60 kv 8e+6 80 kv Photon Fluence (mm -2 ) 6e+6 4e+6 2e+6 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e kv X-ray Energy (kev) X-ray Energy (kev) X-ray Energy (kev) Attenuation (%) model meas. data JM Boone and JA Seibert, An accurate method for computer-generating tungsten anode x-ray spectra from 30 kv to 140 kv, Medical Physics 24; , Thickness (mm Al)

50 Converting m-dgn to p-dgn using spectral weighting polyenergetic functions 1.1 ugy per million photons Photon Fluence (mm -2 ) 1e+7 8e+6 6e+6 4e+6 2e X-ray Energy (kev) Dose per Kerma (mgy/mgy) cm dia 18 cm dia Tube Voltage (kvp) X-ray Energy (kev) JM Boone, N. Shah, and T.R. Nelson, A Comprehensive Analysis of DgN CT Coefficients for Pendant-Geometry Cone-Beam Breast Computed Tomography, Medical Physics 31: (2004)

51 Probability (%) A B C D DD Bra Cup Size Probability (%) A B Bra Cup Size C D DD Breast Diameter (cm)

52 Comparison of breast dimensions X = 13.4 cm = 2.0 cm Median = 13.6 cm 2008 assessment on bct images (N = 137) 2001 tape measure results (N = 200) 52

53

54 actual and modeled x-ray spectra were tuned measured values spectral model 1.2e+7 1e+7 1.2e+7 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e+6 Photon Fluence (mm -2 ) 8e+6 6e+6 4e+6 2e+6 Photon Fluence (mm -2 ) 1.0e+7 8.0e+6 6.0e+6 4.0e+6 2.0e X-ray Energy (kev) X-ray Energy (kev) X-ray Energy (kev)

55 X-ray system output carefully measured

56 Trans (%) 14 iso mr/mas HVL E ave (ent) E ave (exit) BH factor F/mGy

57 MGD(t) in two-view mammography Relationship between bct(d) and mammo(t)

58 MGD (mgy) as a function of air kerma at isocenter (mgy) 10 cm 0% 18 cm 0%

59 isocenter to yield same dose as 2V mammography 18 cm 16 cm 14 cm 12 cm 10 cm

60 isocenter to yield same dose as 2V mammography ~0.113 mgy / mas / = 39.3 mas

61 mas (on our scanner) to yield same dose as 2V mammography

62

63 Dose assessment repeated by Thacker/Glick Thacker S and Glick S, PMB 2004,

64 Other Considerations

65 Average Glandular Dose (mgy) kvp 100 kvp 120 kvp Compressed Breast Thickness (cm)

66 Mammography a a b = high dose area b = low dose area

67 Breast CT 26 kvp Glandular Dose (mgy) a b c a b c Position (cm)

68 Other Considerations (2) DgN s used in mammography probably underestimate dose: Skin Thickness Issue Glandular Density Spectrum Quality factor for E < 25 kev 1.3

69 Breast CT and Dosimetry Breast CT Background Technical Development Initial Clinical Results Dose Issues in Breast CT Summary

70 Breast CT and Dosimetry: Summary Breast CT has strong clinical potential MGD levels are comparable to mammography MC Methods modern x-ray dosimetry When dosimetry methods are not available, you have to develop them yourself!

71 Breast Tomography Project University of California Davis Acknowledgements: California BCRP 7EB-0075 California BCRP 11I-0114 R01 CA R01 EB (BRP) R01 CA (RDB) P30 CA (CCSG) Susan G. Komen Foundation University of Pittsburgh

72

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

for Mammography Dosimetry

for Mammography Dosimetry Changing Perceptions and Updated Methods for Mammography Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE Professor of Radiology & Biomedical Engineering University of California Davis Sacramento,

More information

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics

More information

Skin Model and its impact on Digital Mammography

Skin Model and its impact on Digital Mammography Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1 Outline Motivation Methodology Results

More information

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,

More information

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,

More information

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION Managing Pediatric CT Patient Doses Keith J. Strauss, MSc, FAAPM, FACR President X-Ray Computations, Inc. Boston, Massachusetts Acknowledgements Marilyn Goske, MD John Boone, PhD Cynthia McCollough, PhD

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine ESTABLISHING DRLs in PEDIATRIC CT Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine CT Dose Indices CTDI INTRODUCTION CTDI 100, CTDI w, CTDI vol

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

Hacia la imagenología tomográfica de mama

Hacia la imagenología tomográfica de mama Hacia la imagenología tomográfica de mama Futuro y presente Ioannis Sechopoulos, Ph.D., DABR Advanced X ray Tomographic Imaging (AXTI) Lab Department of Radiology and Nuclear Medicine Radboud University

More information

Translating Protocols Across Patient Size: Babies to Bariatric

Translating Protocols Across Patient Size: Babies to Bariatric Translating Protocols Across Patient Size: Babies to Bariatric Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

Contrast Enhanced Spectral Mammography (CESM) Updates

Contrast Enhanced Spectral Mammography (CESM) Updates Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da

More information

Digital Breast Tomosynthesis from a first idea to clinical routine

Digital Breast Tomosynthesis from a first idea to clinical routine International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen

More information

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX Learning Objectives: Limitations for estimating patient dose for CT Methods

More information

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree DBT Physics Basic to Advanced: Primer On Tomosynthesis Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology Acknowledgements of Support Research support from the Komen Foundation,

More information

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman

More information

Cone-Beam CT for MSK Extremities

Cone-Beam CT for MSK Extremities 8/6/0 Diagnostic Image Quality Evaluation of an Extremity Cone-Beam CT Scanner: Pre-Clinical and First Clinical Results Abdullah Muhit Wojciech Zbijewski, J Webster Stayman John Yorkston, Nathan Packard,

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

State of the art and future development for standardized estimation of organ doses in CT

State of the art and future development for standardized estimation of organ doses in CT State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

Head to new heights with your imaging SCANORA 3D

Head to new heights with your imaging SCANORA 3D SCANORA 3D Head to new heights with your imaging Benefits at a glance The solution for dentomaxillofacial and ENT imaging Easy Patient seated for added stability during exposure. Clear, self-explinatory

More information

A tissue-equivalent phantom series for mammography dosimetry

A tissue-equivalent phantom series for mammography dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.

More information

Better View. Getting a. Plus. enters new phase of clinical testing. Conducting Employee Surveys Radiology Master s Program Ground Transients

Better View. Getting a. Plus. enters new phase of clinical testing. Conducting Employee Surveys Radiology Master s Program Ground Transients RT awardwinning magazine the weekly source for radiology professionals Getting a Better View Dedicated breast CT enters new phase of clinical testing Plus Conducting Employee Surveys Radiology Master s

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, conta

Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, conta Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Karen K. Lindfors,

More information

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS

More information

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity

More information

Since its introduction in 2000, digital mammography has become

Since its introduction in 2000, digital mammography has become Review Article Smith A, PhD email : Andrew.smith@hologic.com Since its introduction in 2000, digital mammography has become an accepted standard of care in breast cancer screening and has paved the way

More information

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities Dual-Energy Imaging of Bone Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities W Zbijewski, 1 A Sisniega, 1 JW Stayman, 1 N Packard, 2 J Yorkston, 2 G Thawait, 3 S Demehri, 3 J Fritz,

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

National Cancer Institute

National Cancer Institute National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,

More information

Estimating Patient Radiation Dose from Computed Tomography

Estimating Patient Radiation Dose from Computed Tomography Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical

More information

Contrast-Enhanced Digital Mammography

Contrast-Enhanced Digital Mammography 2015 ARRS Breast Symposium Contrast-Enhanced Digital Mammography John Lewin, M.D. Diversified Radiology of Colorado CEDM - Outline History Technique Literature Review / Cases Clinical Status Inexpensive,

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

IMAGE GENTLY HOW CAN YOU HELP?

IMAGE GENTLY HOW CAN YOU HELP? IMAGE GENTLY HOW CAN YOU HELP? Keith J. Strauss, MSc, FAAPM, FACR Director, Radiology Physics & Engineering Children s s Hospital Boston Harvard Medical School Acknowledgment Marilyn J. Goske,, MD Robert

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

Managing Radiation Risk in Pediatric CT Imaging

Managing Radiation Risk in Pediatric CT Imaging Managing Radiation Risk in Pediatric CT Imaging Mahadevappa Mahesh, MS, PhD, FAAPM, FACR, FACMP, FSCCT. Professor of Radiology and Cardiology Johns Hopkins University School of Medicine Chief Physicist

More information

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative

More information

Low Dose Molecular Breast Imaging

Low Dose Molecular Breast Imaging Low Dose Molecular Breast Imaging Dr. M.K. O Connor Conflict of Interest Royalties - Gamma Medica Research funding GE Healthcare Research support MTTI Michael O Connor, Ph.D Dept. of Radiology Mayo Clinic

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography Breast tomosynthesis reduces radiologist performance variability compared to digital mammography Andrew Smith 1, Elizabeth Rafferty 2, Loren Niklason 1 1 Hologic, Inc., Bedford MA, USA 2 Massachusetts

More information

ALARA in mammography screening. Hilde Bosmans et al.

ALARA in mammography screening. Hilde Bosmans et al. ALARA in mammography screening Hilde Bosmans et al. ALARA in the mammography Overview 1. Screening 2. The balance: 1. Image quality 2. Radiation dose 3. In practice Screening leeftijd Mammography Screening

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose. Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

Mdiagnosing breast by using x-ray to both breasts. It

Mdiagnosing breast by using x-ray to both breasts. It Asian Journal of Medical and Clinical Sciences Estimation radiation risk during mammography in Sudan 1 1 1, Hifaa.Mohamed Khair, Hiba.Osman, Abdelmoneim Sulieman* 1 Al Neelain University, Faculty of Science

More information

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes BioMedical quantitative X-Ray Imaging Emmanuel Brun Researcher Inserm Université Grenoble Alpes 1 Outline Introduction K-Edge Imaging Patient imaging at the European synchrotron Medical Phase Contrast

More information

Fetal Dose Calculations and Impact on Patient Care

Fetal Dose Calculations and Impact on Patient Care Fetal Dose Calculations and Impact on Patient Care Matt Hough, MS, DABR, DABMP Florida Hospital Diagnostic Medical Physics and Radiation Safety Resource ACR-SPR Practice Parameter for Imaging Pregnant

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com with many thanks to Vince Polkus, Advanced Applications Product Mgr. 1 Content

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Accounting for Imaging Dose

Accounting for Imaging Dose Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some

More information

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist

More information

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences 1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

Imaging of Radiation Dose Using Cherenkov Light

Imaging of Radiation Dose Using Cherenkov Light Imaging of Radiation Dose Using Cherenkov Light Eric Brost 1, Yoichi Watanabe 1, Fadil Santosa 2, Adam Green 3 1 Department of Radiation Oncology, University of Minnesota 2 Institute for Mathematics and

More information

8/3/2016. CAMPEP Accredited, 2016!! CHERENKOV IMAGING IN RADIOTHERAPY: 2D, 3D, REALTIME 3D, Disclosure

8/3/2016. CAMPEP Accredited, 2016!! CHERENKOV IMAGING IN RADIOTHERAPY: 2D, 3D, REALTIME 3D, Disclosure CHERENKOV IMAGING IN RADIOTHERAPY: 2D, 3D, REALTIME 3D, Real time IMRT Real time MRI/Co Whole breast radiotherapy Total skin therapy Brian W. Pogue PhD Engineering, Physics, Surgery, Dartmouth College

More information

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography

Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.

More information

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski White Paper October 2012 SenoBright Contrast Enhanced Spectral Mammography Technology Ann-Katherine

More information

Radiology Review Course Hotel del Coronado Coronado, California

Radiology Review Course Hotel del Coronado Coronado, California 37 th Annual Radiology Review Course Hotel del Coronado Coronado, California Saturday, April 22, 2017 - AM TABLE OF CONTENTS Saturday, April 22, 2017 - AM 7:00 AM 7:30 AM Coffee and Pastries for Registrants

More information

Preparing for Medical Physics Components of the ABR Core Examination

Preparing for Medical Physics Components of the ABR Core Examination Preparing for Medical Physics Components of the ABR Core Examination The ABR core examination for radiologists contains material on medical physics. This content is based on the medical physics that is

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

3D Dosimetry with Polymer Gel

3D Dosimetry with Polymer Gel 3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

Accurate Accurate boost or Simply Accuboost

Accurate Accurate boost or Simply Accuboost Accurate Accurate boost or Simply Accuboost Zoubir Ouhib MS DABR Lynn Cancer Institute of Boca Raton Community Hospital Disclosure: Advisory board Items to be discussed Big picture on management of Breast

More information

Neutron Measurements for Intensity Modulated Radiation Therapy

Neutron Measurements for Intensity Modulated Radiation Therapy SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

Breast dosimetry system in screen/film mammography

Breast dosimetry system in screen/film mammography Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

Andrew Karellas, PhD

Andrew Karellas, PhD Advanced Imaging for Breast Cancer: Screening, Diagnosis, and Assessing Response to Therapy The Role of Tomosynthesis Andrew Karellas, PhD Department of Radiology University of Massachusetts Medical School

More information

Outcomes in the NLST. Health system infrastructure needs to implement screening

Outcomes in the NLST. Health system infrastructure needs to implement screening Outcomes in the NLST Health system infrastructure needs to implement screening Denise R. Aberle, MD Professor of Radiology and Bioengineering David Geffen School of Medicine at UCLA 1 Disclosures I have

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information