1 Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz Freire, 1000, Recife, Pernambuco, Brazil. khoury@ufpe.br Abstract. A radiological screening or diagnostic unit has the primary aim of providing images with a maximum amount of diagnostic information while keeping breast radiation dose as low as reasonably achievable (ALARA). For this purpose it is essential that the whole imaging system be fully optimized. Optimization of image quality is particularly important in mammography. This paper presents results of a quality control (QC) program implemented in a mammography center of Recife, NE Brazil, with focus on patient dose and image quality. The QC program consisted of four different components: a) equipment testing, b) reject analysis, c) sensitometric variability analysis of the automatic processor, and d) dose evaluation to the standard phantom. Sources of radiological quality test protocols were the Brazilian Ministry of Health and the European Guidelines for Quality Assurance in Mammography Screening. Reject analysis was divided in periods between maintenances. Films used for diagnosis were also screened (artifacts, patient movement, film - too dark, too bright). The Entrance Surface Air Kerma and Mean Glandular Dose were determined with termoluminescent dosimeters for a 47mm PMMA phantom under automatic exposure mode and a loaded cassette in place for target film optical density measurements. Significant improvements were achieved with both simple corrective actions and some more technical interventions, such as tube voltage maintenance of the mammography unit by a local engineer. Reject analysis revealed a reduction from 16.1% to 0.6% of films rejected during the period July- December Results of phantom dose measurements were found to be mgy, higher than the current Brazilian recommended limits of 10 mgy, mainly because of AEC system inadequacies. 1. Introduction Quality assurance applied to the radio diagnostic practice is intrinsically related to medical ethics. It s primary goal is to guarantee the fulfillment of three basic criteria: a) Be necessary and appropriate to solve the clinical problem; b) Be able to produce images with enough information to solve the clinical problem; and c) Be optimized, in order that the screening or diagnostic examination results in the lowest possible radiation exposure, lower costs, and inconvenience to the patient. Brazilian National Ministry of Health has determined a series of quality criteria under Portaria 453 [1], to regulate and optimize public and private radio diagnostic installations in the country. It establishes that all medical X-ray equipment should be routinely verified to ensure adequate performance. Patient doses should also be monitored and reference values for incident air kerma at the skin position are provided for various radiological modalities. Assuming the importance of reducing patient doses and optimizing image quality, this paper describes the results of a quality control program (QC) implemented in a mammography center located in Recife, NE Brazil, with focus on patient dose and image quality. 2. Materials and Methods The medical center surveyed uses a mammography unit model Mk MCP (Emic Eletro Medicina Ind. Com. Ltda / Limex) equipped with a Varian M-113R tube, double focus (0.1 e 0.3 mm), molybdenum anode - molybdenum filter (Mo-Mo). Although this model has an Automatic Exposure Control System, it is not equipped with a post-exposure display of the tube load (mas). This facility, as many others in the region, uses a dedicated Macrotec MX-2 model film processor, not specifically designed for mammography film processing. The QC program consisted of four different components: a) Equipment testing, b) Reject analysis, c) Sensitometric variability analysis of the automatic processor; and d) dose evaluation for the standard phantom. In the first part, the Brazilian Ministry of Health protocols were used [1,2], along with the International Atomic Energy Agency (IAEA) ARCAL protocol for mammography [3] and the 1

2 European quality assurance protocol (CEC, 2001). The following tests were performed on the mammography unit: a. Collimation of the X-ray beam (edge of film-edge of bucky distance) b. Reproducibility and exactness of the tube voltage (KVp) c. Reproducibility of the exposure time d. Half-value layer measurement e. Reproducibility and linearity of the incident air kerma rate f. Automatic Exposure Control (AEC) System g. Maximum compression force h. Image quality evaluation using a mammography phantom Image quality evaluation was performed using the Phantom Mama mammography phantom (Centro de Diagnóstico Mamário, Santa Casa da Misericórdia, Rio de Janeiro/RJ), which is similar to the ACR accreditation phantom and is the most commonly available in the region. This test artifact simulates a 45 mm thick breast, with 50%-50% adipose-glandular tissue content. It is fitted with a low contrast scale, metal grids, high contrast microcalcifications, and low contrast structures which simulate masses and fibers. Automatic processor evaluation was done using a 21 step sensitometer and a Clamshell densitometer (Victoreen Nuclear Associates). Tolerance values were taken from the CEC protocol (2001), ± 0.30 DO (± 0.20 DO recommended) for contrast and speed indexes and ± 0.03 DO (± 0.02 DO recommended) for base+fog. Entrance surface air kerma (ESAK) was measured using TLD-100 termoluminescent dosimeters placed on the test phantom (equivalent to a 45 mm, 50:50, breast) and exposure made using AEC. Incident air kerma (INAK) was derived from ESAK using the following equation: Kae, Kai, = (1) B where B backscatter factor (KRAMER et al., [4]) K a,e entrance surface air kerma (including backscatter) incident air kerma (free in air) K a,i 3. Results Results for equipment testing, presented in Table I, shows correspondence with adopted tolerances for most quality control tests, except for irradiation geometry and AEC tests. Table I. Results of the quality control tests of the mammography unit. Description Results Tolerance Adequate Geometry of irradiation (X-ray field to film border distance) 5.8 mm 3 mm No Reproducibility of the tube voltage (CV 1 ) 0.25 % 2% Yes Exactness of the tube voltage (kv) -1.1% 5% Yes Reproducibility of exposure time (CV 1 ) 0.047% 10% Yes Half-value layer (at 28 kv) 0.33 mm Al mm Al Yes Reproducibility of the incident air kerma (CV 1 ) 1.98% 10% Yes Linearity of the incident air kerma (CL 2 ) 0.81% 10% Yes Reprodutibilidade of the AEC system (CV 1 ) 5.9% 5% No Film OD difference, varying PMMA thickness, with AEC 2.34 DO 0.30 DO No Maximum force of compression 120 N N Yes 1 Variability Coefficient (MS, 2003 [2]; CEC, 2001 [4]). 2 Linearity Coefficient (MS, 2003 [2[). A mammographic phantom image was used do evaluate image quality, which was scored by an expert member of the medical staff. Results, in Table II, show minimum detectable structures 2

3 below tolerance limits established by the Ministry of Health and by phantom manufacturer. Based on these results it was concluded that the system has adequate image quality. Table II. Minimum size of the objects correctly visualized in the phantom image and tolerance limits. Results Tolerance Adequate Fibre diameter (mm): a Yes Microcalcification, diameter (mm): a Yes Masses, diameter / height (mm) / (mm): 4 / 2 4 / 2 b Yes Low contrast scale (%): b Yes a Portaria 453 (MS, 1998).[1] b Phantom manual. Reject analysis and investigation of films for defects (artifacts, stains, too dark, poor contrast, etc) were performed on a sample of 5078 films, during the 6 months of the QC programme. Results showed 169 films (3.33%) were rejected by radiographer and 28 films (0.57%) were rejected by the radiologist. The reasons for rejection were primarily due to inadequacies of irradiation parameters (tube voltage, mas) and inadequate film processor temperature. Figure 1 shows a diagram of the rejected films by the radiographer, radiologist and of the quality if the images used for diagnosis. Compared to the total number of films used, a very small amount of films have been rejected. On the other hand, our investigation of the quality of the films used shows that only 8.95% of the films used for diagnosis are actually defect-free. Primary causes are artifacts (75.13%), stains (39.99%), and image of the grid (34.75%). Other defects found were poor contrast films, light in the cassette, and scratches. FIG 1. Reasons for film rejection and number of defective films used for diagnosis. The survey was conducted during intervals between equipment maintenances and corrective actions taken, such as kvp calibration, installation of water filtering, and change of the viewing box light bulbs. The success of the quality program can be seen in Figure 2, showing the relative number 3

4 of films rejected in each period. This Figure shows a reduction in the number of films rejected from 16.1% to 0.6%. The number of films with defects was also reduced as a consequence of the corrective actions taken during the QC program. Figure 3 shows this reduction along the time for films too dark, too light, and with stains. 18% 16% 14% % rejected 12% 10% 8% 6% 4% 2% 0% Períod FIG 2. Percentage of rejected films by period between equipment maintenances. % rejected 7% 6% 5% 4% 3% 2% 1% Too dark Too Light Stains 0% Period FIG 3. Relative number of films rejected due to: film too dark, too light, and stains. Sensitometry was performed daily, before the beginning of the service. Sensitometry indexes kept between acceptable limits, except for 3 days for Speed and 7 days for Contrast index, after equipment maintenances. Incident air kerma at the entrance of the breast position was estimated over a 40 mm PMMA phantom. Results showed INAK to be 17.1 mgy for this institution, which is 70% higher than the 10 4

5 mgy reference value established by the Ministry of Health [1]. Film optical density of 2.43 OD was also found to be higher than the suggested optimal range of 1.40 to These results reflect the AEC system test failure and also the low kvp (26 kv) used for 45 mm patient breasts. 4. Conclusion Equipment testing found inadequate the performance of the AEC system and collimation of the X-ray field. All other quality control tests performed were found to be adequate. Image quality, evaluated using phantom images, was found adequate. Daily sensitometric control of the automatic processor showed large variations after maintenances. Further investigations will be necessary to establish the adequacy of this type of processor for mammography, since it was not specifically designed for this purpose. Reject analysis revealed a reduction from 16.1% to 0.6% in the number of films rejected, consequence of corrective actions, such as correct film storage, film processor maintenances, adjustment of the mammography unit tube voltage, installation of water filtering, and change of the viewing box light bulbs. The number of films used for diagnosis but containing defects was also reduced. Incident air kerma at the phantom surface position was found to be 70% higher than the established reference value. The results indicate that the quality control programme needs to be extended in the studied centre, in order to reduce breast doses during mammographic screenings. 5. References 1. Ministério da Saúde (MS). Diretrizes de Proteção Radiológica em Radiodiagnóstico Médico e Odontológico - Portaria 453 de 01 Junho de Diário Oficial da União, Brasil (1998) 2. Ministério da Saúde (Ms). Agência Nacional de Vigilância Sanitária (Anvisa). Annex of the Resolution Re N. 64 de 04 de Abril de 2003, Brasil (2003). 3 Acuerdo Regional de Cooperacion para la Promocion de la Ciencia y la Tecnologia Nucleares en America Latina y el Caribe (ARCAL) / International Atomic Energy Agency (IAEA). Protocolo de Garantía de Calidad en Mamografia. Aspectos Físicos. Havana (1999). 4. Commission of the European Communities (CEC). European Guidelines for Quality Assurance in Mammography Screening. 3. Ed. Luxemburg: Office for Official Publications of the European Communities (2001). 5. Kramer, R.; Drexler, G.; Pentoussi-Henss, N.; Zankl, M.; Regulla, D.; Panzer, W. Backscatter Factors for Mammography Calculated with Monte Carlo Methods. Physics in Medicine and Biology, v. 46, p (2001). 5

IAEA TECDOC 1517 Quality Control in Mammography Software

IAEA TECDOC 1517 Quality Control in Mammography Software IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Breast phantom with silicone implant for evaluation in conventional mammography

Breast phantom with silicone implant for evaluation in conventional mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, winter 2011 Breast phantom with silicone implant for evaluation in conventional mammography Fábio A. R. Silva, 1a Luíza F Souza, 1 Carlos

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Development of techniques and methods for evaluation of quality of scanned image in mammography

Development of techniques and methods for evaluation of quality of scanned image in mammography Development of techniques and methods for evaluation of quality of scanned image in mammography Priscila do Carmo Santana, Maria S. Nogueira * Development Centre for Nuclear Technology (CDTN / CNEN - MG),

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Development of Object Simulator for Radiation Field of Dental X-Rays

Development of Object Simulator for Radiation Field of Dental X-Rays Journal of Physics: Conference Series OPEN ACCESS Development of Object Simulator for Radiation Field of Dental X-Rays To cite this article: L F Silva et al 2013 J. Phys.: Conf. Ser. 477 012011 View the

More information

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography

Method for determination of the mean fraction of glandular tissue in individual female breasts using mammography INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Hong Kong College of Radiologists Mammography Statement

Hong Kong College of Radiologists Mammography Statement Hong Kong College of Radiologists Mammography Statement The Hong Kong College of Radiologists would like to give the following comments concerning mammography. Mammography screening: Breast cancer is the

More information

Survey of Mammography Practice: Initial Results from Serbia

Survey of Mammography Practice: Initial Results from Serbia Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information

1. Introduction. 2. Technical procedures

1. Introduction. 2. Technical procedures MAMMOGRAPHY QUALITY ASSURANCE IN MOROCCO i inn mill IT -7 v T i * ^ T i XA0101632 F. Zaoui, K. Talsmat, K. Lalaoui Departement des Applications Medicales et Biologiques, Centre National del'energie des

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

ENTRANCE SKIN DOSE MEASUREMENTS FOR PAEDIATRIC CHEST X-RAYS EXAMINATIONS IN BRAZIL

ENTRANCE SKIN DOSE MEASUREMENTS FOR PAEDIATRIC CHEST X-RAYS EXAMINATIONS IN BRAZIL ENTRANCE SKIN DOSE MEASUREMENTS FOR PAEDIATRIC CHEST X-RAYS EXAMINATIONS IN BRAZIL Mohamadain, K.E.M. 1 ; Azevedo, A.C. P. 2 ; da Rosa L.A.R. 3 ; Goncalves O.D. 4 ; Guebel, M.R.N. 5 and Mota, H. C. 3 1

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA Facility Certification Extension Requirements, Quality Assurance and Medical Physicists role for Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT) System Kish chakrabarti, Ph.D. Senior Physicist

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

ESTIMATION OF EFFECTIVE DOSE FROM PEDIATRIC RADIOGRAPHY OF THE CHEST IN MEDELLÍN-COLOMBIA a

ESTIMATION OF EFFECTIVE DOSE FROM PEDIATRIC RADIOGRAPHY OF THE CHEST IN MEDELLÍN-COLOMBIA a Revista Facultad de Ciencias Universidad Nacional de Colombia, Sede Medellín V 4 N 1 enero-junio 2015 ISSN 0121-747X / ISSN-e 2357-5749 Research Paper Páginas 11 a 16 DOI: https://doi.org/10.15446/rev.fac.cienc.v4n1.50759

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom

Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom Equivalent dose to organs and tissues in Hysterosalpingography calculated with the FAX (Female Adult voxel) phantom 1 R Kramer, PhD, 1 H J Khoury, PhD, 1 C. Lopes and 2 J W Vieira, PhD 1 Departamento de

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

Doses Received by Patients during Thorax X-Ray Examinations

Doses Received by Patients during Thorax X-Ray Examinations Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan

More information

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein

More information

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 CHARACTERIZATION OF TLD-100

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

RADIOGRAPHIC IMAGE QUALITY AND DOSE AT THÓRAX, ABDOMEN AND SKULL OF PATIENTS AT HC-FMB UNESP

RADIOGRAPHIC IMAGE QUALITY AND DOSE AT THÓRAX, ABDOMEN AND SKULL OF PATIENTS AT HC-FMB UNESP IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR RADIOGRAPHIC IMAGE

More information

Radiographic techniques in screen-film mammography

Radiographic techniques in screen-film mammography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 3, SUMMER 2002 Radiographic techniques in screen-film mammography Thomas R. LaVoy* 4314 Kelsey Drive, Syracuse, New York 13215 Walter Huda

More information

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system

Mean glandular dose values used for the mammography screening program in Poland according to the type of image registration system Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

Quality control in mammography: an initiative in France

Quality control in mammography: an initiative in France f 1994, The British Journal of Radiology, 67, 371-383 Quality control in mammography: an initiative in France 1 C MACCIA, PhD, 2 R RENAUD, MD, 1 S CASTELLANO, PhD, 2 P SCHAFFER, MD, PhD, 2 R WAHL, MD,

More information

Mdiagnosing breast by using x-ray to both breasts. It

Mdiagnosing breast by using x-ray to both breasts. It Asian Journal of Medical and Clinical Sciences Estimation radiation risk during mammography in Sudan 1 1 1, Hifaa.Mohamed Khair, Hiba.Osman, Abdelmoneim Sulieman* 1 Al Neelain University, Faculty of Science

More information

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Nadia Alatta 1, Abdelrahaman Elnour 2, Ikhlas Abdelaziz 3, Tugua Tageldeen 4, Duha Abdu 5, Asma Elamin

More information

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis

Effect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of a New Tomosynthesis Program: A Physicists Perspective Implementation of a New Tomosynthesis Program: A Physicists Perspective Bill Geiser, MS DABR Senior Medical Physicist wgeiser@mdanderson.org 1 Conflict of Interest None of the authors nor their immediate

More information

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom 2001 AAPM Summer School Seattle, Washington ACR R/F Phantom Charles R. Wilson, Ph.D., FACR Medical College of Wisconsin Milwaukee, Wisconsin MCMC Circa 1975 ACR R/F Phantom This lecture has been approved

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

CALDose_X a software tool for absorbed dose calculations in diagnostic radiology

CALDose_X a software tool for absorbed dose calculations in diagnostic radiology CALDose_X a software tool for absorbed dose calculations in diagnostic radiology Richard Kramer *a, Helen Jamil Khoury a and José Wilson Vieira b, c a Departamento de Energia Nuclear, Universidade Federal

More information

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

IAEA-TECDOC-796. Radiation doses in diagnostic radiology IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose

More information

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis J. Antunes 1,2, J. Machado 1,2, L. Peralta 1,2, N. Matela 1,3 1 Departamento de Física da Faculdade de Ciências da Universidade

More information

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Establishment of local diagnostic reference levels for quality control in intraoral radiography Oral Radiol (27) 33:38 44 DOI.7/s282-6-24-9 ORIGINAL ARTICLE Establishment of local diagnostic reference levels for quality control in intraoral radiography Maki Izawa Yasuo Harata Noriyoshi Shiba Nobuhide

More information

CT Quality Control Manual FAQs

CT Quality Control Manual FAQs CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME

EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME S. Ferrer (), M Ramos (), J.I. Villaescusa (), G Verdú (), M.D. Salas (3), M.D. Cuevas (4) () Departamento

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, Mail Code 25-01

More information

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan Quality Assurance in Mammography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Yi-Shuan Hwang 1 Hui-Yu Tsai 2 Chien-Chuan Chen 1 Pei-Kwei Tsay 3 Huay-Ben

More information

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

More information

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability.

More information

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis 60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

Mammography Quality Control: A Refresher

Mammography Quality Control: A Refresher Mammography Quality Control: A Refresher MATT WAIT, MS, DABR Objectives Attendees will re-familiarize themselves with the purpose of quality assurance and quality control Attendees will re-familiarize

More information

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County

More information

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

Breast dosimetry system in screen/film mammography

Breast dosimetry system in screen/film mammography Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,

More information

A tissue-equivalent phantom series for mammography dosimetry

A tissue-equivalent phantom series for mammography dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Research Article. Corresponding author H. Osman

Research Article. Corresponding author H. Osman Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific

More information

Measurement of Absorbed Doses in Anatomical Phantoms

Measurement of Absorbed Doses in Anatomical Phantoms International Journal of Pure and Applied Physics. ISSN 0973-1776 Volume 8, Number 1 (2012), pp. 53-58 Research India Publications http://www.ripublication.com/ijpap.htm Measurement of Absorbed Doses in

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

CALDose_X online: Web-based, real time Monte Carlo calculations for patient dosimetry in X-ray diagnosis

CALDose_X online: Web-based, real time Monte Carlo calculations for patient dosimetry in X-ray diagnosis CALDose_X online: Web-based, real time Monte Carlo calculations for patient dosimetry in X-ray diagnosis Richard Kramer 1, Arthur Cavalcanti 2, Vagner Ferreira Cassola 1, Rafael Dueire Lins 2, Carlos Alberto

More information

Patient doses in general radiography examinations

Patient doses in general radiography examinations peer reviewed ORIGINAL ARTICLE Patient doses in general radiography examinations Wambani JS 1 MBChB, M. Med (Rad), Paediatric (Rad) Onditi EG 2 MBChB, M.Med (Rad) Korir GK 3 B.Sc, MSc (Nuclear Science),

More information

History of MQSA ACR creates testing procedures Oct Mammography Quality Standards Act Dec 1993 Interim Regulations Oct All mammog

History of MQSA ACR creates testing procedures Oct Mammography Quality Standards Act Dec 1993 Interim Regulations Oct All mammog 15 Years of MQSA, What works, what doesn t Jerry Soen, MS DABR History of MQSA 1990-9191 ACR creates testing procedures Oct. 1992 Mammography Quality Standards Act Dec 1993 Interim Regulations Oct. 1994

More information

Dental Intraoral X-ray Systems

Dental Intraoral X-ray Systems Dental Intraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 2.0 4732.#### DENTAL INTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Commented [JC(1]: Based on part 4732.0880. Subpart 1. Applicability.

More information

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE ACR Update on FFDM Accreditation Overview New ACR Activities Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO QC Today QC Tomorrow *No financial disclosures to report

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences 1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,

More information

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation? The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Revised: September 7, 2017; updated questions in red) Table of Contents Application - General...

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography Available online at http://www.biij.org/2006/3/e35 doi: 10.2349/biij.2.3.e35 biij Biomedical Imaging and Intervention Journal ORIGINAL PAPER Evaluation of radiation dose and image quality following changes

More information

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units

Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units Evaluation of the doses delivered to premature babies in the Belgian Neonatal Intensive Care Units PreDos Project Jérémie Dabin Lara Struelens Filip Vanhavere Table of Contents Executive summary... Introduction...

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Dose Evaluation for Common Digital Radiographic s in Selected Hospitals in Pahang Malaysia Soo-Foon Moey 1*, Zubir Ahmad Shazli 1, Inayatullah Shah Sayed

More information

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN

More information

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose

Comparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman

More information

Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography

Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography Characteristic curve measurement based on bootstrap method using a new calcium phosphate stepwedge in mammography Norimitsu SHINOHARA* a, Hiroshi FUJITA b, Takeshi HARA b, and Katsuhei HORITA c a Electronic

More information

Mammographic Accreditation Phantom

Mammographic Accreditation Phantom Quality Control in Mammography RMI 156 Mammographic Accreditation Phantom RMI 156 A User s Guide GAMMEX RMI Table of Contents Product Overview...................... 1 Product Description....................

More information

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography Diagn Interv Radiol 2007; 13:134 139 Turkish Society of Radiology 2007 BREAST IMAGING ORIGINAL ARTICLE Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY Journal of Bangladesh Academy of Sciences, Vol. 3, No. 1, 7-17, 211 A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY MAHFUZA BEGUM 1, A. S. MOLLAH

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,

More information