ULTRASOUND BLOOD FLOW IMAGING IN CAROTID ARTERIES BEFORE AND AFTER ENDARTERECTOMY

Size: px
Start display at page:

Download "ULTRASOUND BLOOD FLOW IMAGING IN CAROTID ARTERIES BEFORE AND AFTER ENDARTERECTOMY"

Transcription

1 ULTRASOUND BLOOD FLOW IMAGING IN CAROTID ARTERIES BEFORE AND AFTER ENDARTERECTOMY G. BAMBI 1, F. GUIDI 1, S. RICCI 1, P. TORTOLI 1 M.R. CIRELLI 2, L. PEDRINI 2 1 Electronics & Telecommunications Department,University of Florence, via Santa Marta 3, Firenze, Italy 2 Unità Operativa Complessa di Chirurgia Vascolare, Ospedale Maggiore, Azienda U.S.L. Città di Bologna, Italy 1. Introduction Echo-Doppler analysis is a fundamental medical tool for the evaluation of the presence of atherosclerotic plaques in the carotid artery (Carpenter et al, 1995), since it may show the dimensions of the stenosis and its haemodynamic significance. This exam is used both for pre-operative evaluations and for post-operative check ups in atherosclerotic patients subject to surgical treatment, for it has diagnostic capabilities of significant modifications to vessel morphology. Echo-Doppler investigation currently gives quantitative information on blood flow dynamics only for a small region inside the vessel. Using the colour- or power-doppler modes, a larger region can be scanned but mainly qualitative information is obtained in this way. The applications of a unique range-doppler instrument (called multigate ) implemented at the University of Florence showed the capability of extending the quantitative Doppler spectral analysis to multiple sample volumes along an entire ultrasound scan line. Such a system consists of a PC board that digitally processes the data received from ultrasound equipment. The results are shown on the Personal Computer (PC) monitor, which displays the distribution of velocity within the investigated vessel (velocity profiles). To facilitate the use of the multigate system in a clinical setting, commercial ultrasound equipment was modified to control the multigate capabilities and display both morphological (B-mode) and haemodynamic (velocity profile) information on the same screen. This allowed medical staff to conduct the analysis using only one instrument, the echograph, as in a standard investigation. This paper presents the results of the first clinical application of such system, addressed to investigate the relation between velocity profiles and morphology in carotid arteries (Perktold et al, 1991) of atherosclerotic patients subject to surgical treatment. Following a rigorous measure protocol, flow profiles were investigated at different sites both upstream and downstream the carotid bifurcation. Such measures were carried out in two different circumstances: pre-operative conditions in 18 patients

2 and post-operative conditions in 15 patients. The two analyses were performed on the same patients for 9 cases of the total set reported above. Both of the analyses, before and 8-days-after endarterectomy (EA), were carried out during the usual medical check-ups. The pre-operative evaluations showed that velocity profiles in correspondence to the stenosis were always notably distorted with respect to the shape visible in healthy conditions while in some cases the presence of the plaque could not be easily detected through a conventional sonogram related to a single sample volume. After the surgical treatment the laminar flow conditions were recovered in 5 cases while in 10 cases an irregular flow was still evident with a high variability both in time (during the cardiac cycle) and in space (along the depths). 2. Materials and methods The multigate system consists of a single card housed in a free slot of a PC. It can digitise, process and store 64 complex (range-gated) signals originating at different depths along the beam axis of an ultrasound transducer, fired in Pulsed Wave mode (PW). Powerful Digital Signal Processors (DSPs) elaborate in real-time the acquired data in such a way that the power spectral density of each of the 64 Doppler signals is produced. The results are directly transferred into the PC, where the monitor displays them in the so called spectral profile mode (Tortoli et al, 1996). A large memory buffer is also available for raw data storage. The PC runs a custom software, called G.A.S.P. (Global Acquisition and Signal Processing) that consists of an integrated shell for Microsoft Window (Microsoft Corporation, Redmond, Washington) family OS. It includes applications for post processing carried out in LabVIEW (National Instruments Corporation, Austin, Texas) and a Microsoft Visual C++ (Microsoft Corporation, Redmond, Washington) coded console for real-time data acquisition and visualisation. Processed data are displayed in real-time providing dynamic images of the distribution of Doppler frequencies along the investigated depth (spectral profiles). It is possible to extract the maximum and the mean frequencies from the spectral profiles for each depth with both being suitable for conversion to velocity. For this application, a commercial ultrasound equipment (AU3, Esaote, Florence, Italy) was expressly modified to control the multigate system from the echograph console and to display the spectral profile with the B-mode representation on the echograph monitor. An opportune command sequence allows the operator to easily switch from the normal use of the equipment to the multigate mode and back again. The controls of the multigate are set by means of the keyboard and the trackball of the AU3, and are sent to the board through a serial channel, while the information of the VGA signal are used to visualize the spectral profile on the AU3 monitor, in dual image mode (Figure 1).

3 Figure 1. Visualization obtainable on the echograph with the dual image (B-mode + multigate) modality. The right side of the figure shows the spectral profile, with depth in the vertical axis, frequency in horizontal axis and power spectral density in colour code. 24 patient with atherosclerotic problems at the carotid artery level submitted to surgical treatment were monitored using the multigate system in two different phases: in pre-operative conditions, when the evaluation of the stenosis level was performed, and in post-operative conditions, 8-days after the surgical treatment, when the results of the operation were evaluated. A specific measurement protocol was defined to obtain comparable images in different sites for all patients: starting from the level of the carotid bifurcation, the acquisitions were performed in three points in the common carotid artery (1-2-3 cm upstream the bulb) and in three points in the internal carotid artery (1-2-3 cm downstream the bulb), where the plaque is usually located. This protocol allowed to analyze a long region of the carotid, to follow the development of the velocity profile outside the stenotic region, too. 3. Experimental results The profiles acquired in correspondence to the stenosis, which is typically located in the internal carotid artery 1 cm downstream the bifurcation, had the same long, narrow and filled in shape in all pre-operative cases (Figure 2). In some cases, a double peaked profile was detected. In the same cases, the sonogram related to the central region of the vessel showed a normal waveform, compatible with a non-pathological condition (Figure 3) When the plaque was calcified, it was difficult to get the data, since the received signal was low and in many cases totally immersed in noise.

4 Figure 2. Typical spectral profile detected in the internal carotid artery of patients before EA. This profile was taken at the time indicated by the white line in the sonogram (upper image) corresponding to the centre of the vessel. Here aliasing is evident, due to the high velocity through the narrow vessel (diameter less than 3 mm). A B C D Figure 3. Pre-operative profiles detected at different registration sites. In the fourth image, obtained in correspondence to the more stenotic region, a narrow profile with a double horn is visible. This profile was frozen at the instant indicated on the sonogram. The examinations carried out 8 days after the EA were usually easier, due to the plaque removal. Three different surgical treatments were used for the 15 patients examined: 10 cases had an EA with a direct suture of the arteriotomy, in 4 cases the arteriotomy was closed with a polyester patch; in the last a 6 mm polytetrafluoroethylene (PTFE) graft was inserted at the level the carotid bulb.

5 In 5 cases, spectral profiles looked regular and larger than in the pre-operative condition (Figure 4), while in 10 cases some turbulences were still detected (Figure 5). (a) (b) (c) Figura 4. Post-operative check-up after TEA with direct suture and a cylinder-like carotid bulb. (a) Delineated and symmetrical profile in the common carotid artery. (b) Large and M shaped profile in deceleration 1 cm downstream the bifurcation. (c) Large and symmetrical profile in the internal carotid artery, 2 cm downstream the bifurcation. Figure 5. Post-operative check-up after TEA with patch and the carotid bulb a little elliptic. Delineated and symmetrical profile in the common carotid artery (upper images), large profile with some turbulences 1 cm downstream the bifurcation (bottom images).

6 Figure 6. Post-operative check-up after TEA with patch. Spectral profiles for the patch region, corresponding to the subsequent instants indicated in the sonogram. The main flow is directed toward left (bottom-left image), and a strong reverse flow is evident in the bottom-right image. Figure 7. Post-operative check-up after TEA with PTFE graft. Spectral profiles for the graft region, corresponding to the two peaks indicated in the sonogram. The maximum frequency is shifted toward the lower side of the profile (posterior- medial wall, bottom-left image), and toward the upper side (anteriorlateral wall, bottom-right image), respectively.

7 In 3 of the 4 cases of patch insertion back flow in the internal carotid artery was evident (Figure 6). For the patient treated with a PTFE graft, in correspondence to the initial portion of the graft a double acceleration flow was revealed. The spectral profile was skew, with the peak shifted toward the lower side of the profile (posterior- medial wall) in the first phase of acceleration, and toward the upper side (anterior-lateral wall) in the second phase (Figure 7). 4. Discussion The capability of visualizing both the spectral profile and the B-mode image on the same echograph display, and of controlling all the system by the echograph console allowed medical staff to perform the analysis using the same procedure used for conventional echo-doppler examination. The spectral profiles provided by the multigate system were highly informative. By examining them for the pre-operative condition, the geometric structure of the vessel could be estimated. For example, by looking at the spectral profiles of Figure 3, the asymmetrical shape of profile (A) and (B) suggested the presence of a flow obstruction in the lower part of the vessel, while the long and narrow profile (D) could be related with the stenosis. Comparing the angiography with the spectral results, these hypotheses were confirmed. The pre-operative analyses showed a high variability of the spectral profiles, due to the presence of different morphologies of the plaques. A common characteristic for the spectral profiles corresponding to the region of the stenosis was the narrow and extended shape. In some cases (Figure 3) a particular M shape was recognized. For some patients, an analysis using only the usual sonogram obtained with the PWmode of the echograph did not show the presence of the stenosis and revealed a quasinormal shape. The possibility of having the complete spectral profile simplified the stenosis identification. In most cases the spectral profiles of the post-operative evaluations were larger than those obtained in pre-operative conditions, due to the stenosis removal. In 10 cases turbulences were again detected, while only in 5 cases the shape was regular, corresponding to a laminar flow. The data obtained in the examinations in post-operative conditions seem to be dependent on the type of surgical treatment, but at present there is not a large enough sample of analyses to carry out statistical results. For the direct EA operation the carotid bulb maintains a regular shape, even though dimensions are reduced. For the patch insertion the bulb is enlarged and extended, due to the fact that the patch walls are more rigid than the vessel walls (Archie, 1997). The interposition of a graft also produces a rectilinear conformation of the vessel. The differences in the velocity profiles introduced by these different surgical treatments are to be investigated in the next step of this research.

8 References Archie, J.P.Jr. (1997) Geometric dimension changes with carotid endarterectomy reconstruction, J. Vasc. Surg. 25, Carpenter, J.P., Lexa, F.J., Davis, J.T. (1995) Determination of sixty percent or greater carotid artery stenosis by duplex Doppler ultrasonography, J. Vasc. Surg. 22(6), Perktold, K., Resch, M., Peter, RO (1991) Three-dimensional numerical analysis of pulsatile flow and wall shear stress in the carotid artery bifurcation, J.Biomechanics 24(6), Tortoli, P., Guidi, F., Guidi, G., Atzeni, C. (1996) Spectral velocity profiles for detailed ultrasound flow analysis, IEEE Trans. UFFC 43,

Policies and Statements D16. Intracranial Cerebrovascular Ultrasound

Policies and Statements D16. Intracranial Cerebrovascular Ultrasound Policies and Statements D16 Intracranial Cerebrovascular Ultrasound SECTION 1: INSTRUMENTATION Policies and Statements D16 Intracranial Cerebrovascular Ultrasound May 2006 (Reaffirmed July 2007) Essential

More information

Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012

Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012 Carotid Abnormalities Coils, Kinks and Tortuosity David Lorelli M.D., RVT, FACS Michigan Vascular Association Conference Saturday, October 20, 2012 Page 1 Table of Contents Carotid Anatomy Carotid Duplex

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

VFI Technology to Change the Way You Work

VFI Technology to Change the Way You Work Analogic Ultrasound VFI Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler

More information

Doppler Basic & Hemodynamic Calculations

Doppler Basic & Hemodynamic Calculations Doppler Basic & Hemodynamic Calculations August 19, 2017 Smonporn Boonyaratavej MD Division of Cardiology, Department of Medicine Chulalongkorn University Cardiac Center, King Chulalongkorn Memorial Hospital

More information

GUNDERSEN/LUTHERAN ULTRASOUND DEPARTMENT POLICY AND PROCEDURE MANUAL

GUNDERSEN/LUTHERAN ULTRASOUND DEPARTMENT POLICY AND PROCEDURE MANUAL GUNDERSEN/LUTHERAN ULTRASOUND DEPARTMENT POLICY AND PROCEDURE MANUAL SUBJECT: Carotid Duplex Ultrasound SECTION: Vascular Ultrasound ORIGINATOR: Deborah L. Richert, BSVT, RDMS, RVT DATE: October 15, 2015

More information

Deb Coghlan AMS (Vascular and General ) Brisbane, Australia

Deb Coghlan AMS (Vascular and General ) Brisbane, Australia Deb Coghlan AMS (Vascular and General ) Brisbane, Australia ANEURYSMAL DIISEASE The infrarenal aorta enlarges with age, and is the commonest site for arterial aneurysms. An aneurysm is a permanent focal

More information

Diagnosis of Middle Cerebral Artery Occlusion with Transcranial Color-Coded Real-Time Sonography

Diagnosis of Middle Cerebral Artery Occlusion with Transcranial Color-Coded Real-Time Sonography Diagnosis of Middle Cerebral Artery Occlusion with Transcranial Color-Coded Real-Time Sonography Kazumi Kimura, Yoichiro Hashimoto, Teruyuki Hirano, Makoto Uchino, and Masayuki Ando PURPOSE: To determine

More information

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals Optimising your Doppler settings for an accurate PI Alison McGuinness Mid Yorks Hospitals Applications Both maternal uterine and fetal circulations can be studied with doppler sonography Uterine arteries

More information

No financial or commercial relationships to disclose

No financial or commercial relationships to disclose Deanna New, RVT No financial or commercial relationships to disclose IAC REQUIREMENTS: The main duty of a sonographer is to make the physician or radiologists job easier by capturing images and doing

More information

Disclosure Statement:

Disclosure Statement: Marsha M. Neumyer, BS, RVT, FSVU, FSDMS, FAIUM International Director Vascular Diagnostic Educational Services Vascular Resource Associates Harrisburg, PA Disclosure Statement: CME Calendar QR Code Marsha

More information

What effects will proximal or distal disease have on a waveform?

What effects will proximal or distal disease have on a waveform? Spectral Doppler Interpretation Director of Ultrasound Education & Quality Assurance Baylor College of Medicine Division of Maternal-Fetal Medicine Maternal Fetal Center Imaging Manager Texas Children

More information

ASDIN 7th Annual Scientific Meeting DISCLOSURES TECHNICAL CONSIDERATIONS TECHNICAL CONSIDERATIONS UTILITY OF ULTRASOUND IN EVALUATING ACCESS

ASDIN 7th Annual Scientific Meeting DISCLOSURES TECHNICAL CONSIDERATIONS TECHNICAL CONSIDERATIONS UTILITY OF ULTRASOUND IN EVALUATING ACCESS DISCLOSURES UTILITY OF ULTRASOUND IN EVALUATING ACCESS DYSFUNCTION None Vandana Dua Niyyar, MD Assistant Professor of Medicine, Division of Nephrology, Emory University UTILITY OF ULTRASOUND IN ACCESS

More information

Duplex Ultrasound of the Renal Arteries. Duplex Ultrasound. In the Beginning

Duplex Ultrasound of the Renal Arteries. Duplex Ultrasound. In the Beginning Duplex Ultrasound of the Renal Arteries DIMENSIONS IN HEART AND VASCULAR CARE 2013 PENN STATE HEART AND VASCULAR INSTITUTE ROBERT G. ATNIP MD PROFESSOR OF SURGERY AND RADIOLOGY Duplex Ultrasound Developed

More information

EXPLORING ULTRASOUND IMAGES OF THE CAROTID ARTERIES USING NEURAL NETWORK TOOLS

EXPLORING ULTRASOUND IMAGES OF THE CAROTID ARTERIES USING NEURAL NETWORK TOOLS Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5430 EXPLORING ULTRASOUND IMAGES

More information

ScienceDirect. Ergonomics and workflow evaluation of automatic doppler angle technology implemented in a diagnostic ultrasound system

ScienceDirect. Ergonomics and workflow evaluation of automatic doppler angle technology implemented in a diagnostic ultrasound system Available online at www.sciencedirect.com ScienceDirect Procedia Manufacturing 3 (2015 ) 120 127 6th International Conference on Applied Human Factors and Ergonomics (AHFE 2015) and the Affiliated Conferences,

More information

Ultrasound Physics & Terminology

Ultrasound Physics & Terminology Ultrasound Physics & Terminology This module includes the following: Basic physics terms Basic principles of ultrasound Ultrasound terminology and terms Common artifacts seen Doppler principles Terms for

More information

Radial Artery Assessment for Coronary Artery Bypass

Radial Artery Assessment for Coronary Artery Bypass VASCULAR TECHNOLOGY PROFESSIONAL PERFORMANCE GUIDELINES Radial Artery Assessment for Coronary Artery Bypass This Guideline was prepared by the Professional Guidelines Subcommittee of the Society for Vascular

More information

Pulsed Doppler techniques are commonly used

Pulsed Doppler techniques are commonly used 336 Accurate Noninvasive Method to Diagnose Minor Atherosclerotic Lesions in Carotid Artery Bulb Tiny van Merode, MD, Jan Lodder, MD, Frans A.M. Smeets, Arnold P.G. Hoeks, PhD, and Robert S. Reneman, MD,

More information

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA 99. MRA Principles and Carotid MRA As described in Chapter 12, time of flight (TOF) magnetic resonance angiography (MRA) is commonly utilized in the evaluation of the circle of Willis. TOF MRA allows depiction

More information

PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING

PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING STANDARD - Primary Instrumentation 1.1 Cardiac Ultrasound Systems SECTION 1 Instrumentation Ultrasound instruments

More information

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus Page 1 of 7 11/18 Candidates are expected to cover all of the content of this syllabus when preparing

More information

Doppler ultrasound as noninvasive diagnosis of peripheral arterial disease

Doppler ultrasound as noninvasive diagnosis of peripheral arterial disease Doppler ultrasound as noninvasive diagnosis of peripheral arterial disease Poster No.: C-0246 Congress: ECR 2012 Type: Scientific Exhibit Authors: C. Ballester Valles, F. Aparici-Robles; Valencia/ES Keywords:

More information

Beyond Stenosis Severity: Top 5 Important Duplex Characteristics to Identify in a Patient with Carotid Disease

Beyond Stenosis Severity: Top 5 Important Duplex Characteristics to Identify in a Patient with Carotid Disease Beyond Stenosis Severity: Top 5 Important Duplex Characteristics to Identify in a Patient with Carotid Disease Jan M. Sloves RVT, RCS, FASE Technical Director New York Cardiovascular Associates Disclosures

More information

5 Working With Measurements

5 Working With Measurements 5 Working With Measurements Measurement Overview Measurements accompanying ultrasound images supplement other clinical procedures available to the attending physician. Accuracy of the measurements is determined

More information

Vascular Modeling & Hemodynamics Research OUTLINE. CLINICAL MOTIVATION: Stroke. BACKGROUND: Diagnosis

Vascular Modeling & Hemodynamics Research OUTLINE. CLINICAL MOTIVATION: Stroke. BACKGROUND: Diagnosis Biomedical Ultrasonics Research Lab Vascular Modeling & Hemodynamics Research Tamie Poepping Biomedical Ultrasonics Research Laboratory Dept. of Physics & Astronomy The University of Western Ontario UWO:

More information

Physician s Vascular Interpretation Examination Content Outline

Physician s Vascular Interpretation Examination Content Outline Physician s Vascular Interpretation Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 6 Cerebrovascular Abdominal Peripheral Arterial - Duplex Imaging Peripheral Arterial

More information

Categorical Course: Update of Doppler US 8 : 00 8 : 20

Categorical Course: Update of Doppler US 8 : 00 8 : 20 159 Categorical Course: Update of Doppler US 8 : 00 8 : 20 160 161 Table 1.Comparison of Recommended Values from Data in the Published Literature* S t u d y Lesion PSV E D V VICA/VCCA S e v e r i t y (

More information

Vascular Ultrasound: Current state, current needs, future directions

Vascular Ultrasound: Current state, current needs, future directions Vascular Ultrasound: Current state, current needs, future directions Laurence Needleman, MD Thomas Jefferson University Hospitals Sidney Kimmel Medical College of Thomas Jefferson University Disclosures

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

Ultrasound Applied Physics

Ultrasound Applied Physics Ultrasound Applied Physics University of Toronto Department of Medical Imaging Applied Physics Mini-Course #3 2016 Ultrasound Laboratory Manual and Examination Booklet 1/21/2016 Ultrasound Applied Physics

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Duplex Criteria for Determination of 50% or Greater Carotid Stenosis

Duplex Criteria for Determination of 50% or Greater Carotid Stenosis Article Duplex Criteria for Determination of 50% or Greater Carotid Stenosis David G. Neschis, MD, Frank J. Lexa, MD, Julia T. Davis, RN, RVT, Jeffrey P. Carpenter, MD, RVT Recently the North American

More information

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation for spiral flow imaging Joao L. A. Carvalho Previous talks Non-Cartesian reconstruction (2005) Spiral FVE (Spring 2006) Aortic flow Carotid flow Accelerated spiral FVE (Fall 2006) 2007? Department of Electrical

More information

Carotid Stenosis (carotid artery disease)

Carotid Stenosis (carotid artery disease) 1 Carotid Stenosis (carotid artery disease) Overview Carotid stenosis is a narrowing of the carotid arteries, the two major arteries that carry oxygenrich blood from the heart to the brain. Also called

More information

Duplex ultrasound is first-line imaging for all

Duplex ultrasound is first-line imaging for all Our Protocol for Transabdominal Pelvic Vein Duplex Ultrasound A summary of s protocol for pelvic vein duplex ultrasonography, including equipment, patient positioning, ultrasound settings, and technique.

More information

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System

Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System Guide to Small Animal Vascular Imaging using the Vevo 770 Micro-Ultrasound System January 2007 Objectives: After completion of this module, the participant will be able to accomplish the following: Understand

More information

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial

More information

Diagnostic approach to heart disease

Diagnostic approach to heart disease Diagnostic approach to heart disease Initial work up History Physical exam Chest radiographs ECG Special studies Echocardiography Cardiac catheterization Echocardiography principles Technique of producing

More information

Goals. Access flow and renal artery stenosis evaluation by Doppler ultrasound. Reimbursement. WHY use of Doppler Ultrasound

Goals. Access flow and renal artery stenosis evaluation by Doppler ultrasound. Reimbursement. WHY use of Doppler Ultrasound Access flow and renal artery stenosis evaluation by Doppler ultrasound Adina Voiculescu, MD Interventional Nephrology Brigham and Women s Hospital Boston Instructor at Harvard Medical School Understand

More information

Congenital. Unicuspid Bicuspid Quadricuspid

Congenital. Unicuspid Bicuspid Quadricuspid David Letterman s Top 10 Aortic Stenosis The victim can be anyone: Echo is the question and the answer!!!! Hilton Head Island Echocardiography Conference 2012 Timothy E. Paterick, MD, JD, MBA Christopher

More information

Protokollanhang zur SPACE-2-Studie Neurology Quality Standards

Protokollanhang zur SPACE-2-Studie Neurology Quality Standards Protokollanhang zur SPACE-2-Studie Neurology Quality Standards 1. General remarks In contrast to SPACE-1, the neurological center participating in the SPACE-2 trial will also be involved in the treatment

More information

Recommendations for documentation of neurosonographic examinations

Recommendations for documentation of neurosonographic examinations Recommendations for documentation of neurosonographic examinations The documentation of ultrasound examinations is subject to a dynamic development particularly as regards newer applications. The present

More information

Doppler Color Flow Imaging #4

Doppler Color Flow Imaging #4 Doppler Color Flow Imaging #4 Joseph A. Kisslo, MD David B. Adams, RDCS INTRODUCTION Doppler color flow imaging is a method for noninvasively imaging blood flow through the heart by displaying flow data

More information

What effects will proximal or distal disease have on an waveform?

What effects will proximal or distal disease have on an waveform? Spectral Doppler Interpretation Director Director of of Ultrasound Ultrasound Education Education & & Quality Quality Assurance Assurance Baylor Baylor College College of of Medicine Medicine Division

More information

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),

More information

Evaluation of Carotid Vessels and Vertebral Artery in Stroke Patients with Color Doppler Ultrasound and MR Angiography

Evaluation of Carotid Vessels and Vertebral Artery in Stroke Patients with Color Doppler Ultrasound and MR Angiography Evaluation of Carotid Vessels and Vertebral Artery in Stroke Patients with Color Doppler Ultrasound and MR Angiography Dr. Pramod Shaha 1, Dr. Vinay Raj R 2, Dr. (Brig) K. Sahoo 3 Abstract: Aim & Objectives:

More information

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS ULTRASOUND QA SOLUTIONS Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS INNOVATORS IN ADVANCED ULTRASOUND TECHNIQUES Gammex is the

More information

Carotid Artery Doppler

Carotid Artery Doppler Carotid Artery Doppler Patient Position supine or semisupine head slightly hyper extended rotated 45 away from the side being examined. Higher frequency linear transducers (7 MHz) Vessels should be imaged

More information

RPVI Exam Review ecourse

RPVI Exam Review ecourse RPVI Exam Review ecourse The RPVI Exam Review ecourse consists of ten Vascular Physics Modules and fourteen Vascular Specialty Modules. Detailed descriptions of module content are listed below. During

More information

Neuro Quiz 29 Transcranial Doppler Monitoring

Neuro Quiz 29 Transcranial Doppler Monitoring Verghese Cherian, MD, FFARCSI Penn State Hershey Medical Center, Hershey Quiz Team Shobana Rajan, M.D Suneeta Gollapudy, M.D Angele Marie Theard, M.D Neuro Quiz 29 Transcranial Doppler Monitoring This

More information

8/20/18. The Doppler Effect. Objectives. What is the Doppler Effect. Doppler principles. Spectral Waveform. Image recognition. Vascular Ultrasound

8/20/18. The Doppler Effect. Objectives. What is the Doppler Effect. Doppler principles. Spectral Waveform. Image recognition. Vascular Ultrasound Vascular Ultrasound: Physics and Haemodynamics Objectives Doppler principles Spectral Waveform Key factors Haemodynamics: Stenosis Waveforms Image recognition Vascular Ultrasound: A flawed paradigm What

More information

noninvasive, nonionizing, portable, inexpensive, safe for serial or prolonged studies

noninvasive, nonionizing, portable, inexpensive, safe for serial or prolonged studies TRANS CRANIAL DOPPLER Presented by : Anil Garg Transcranial Doppler 1982, Aaslid and colleagues introduced TCD as a non-invasive technique for monitoring blood flow velocity in basal cerebral arteries

More information

Imaging of the Basal Cerebral Arteries and Measurement of Blood Velocity in Adults by Using Transcranial Real-Time Color Flow Doppler Sonography

Imaging of the Basal Cerebral Arteries and Measurement of Blood Velocity in Adults by Using Transcranial Real-Time Color Flow Doppler Sonography 497 Imaging of the Basal Cerebral Arteries and Measurement of Blood Velocity in Adults by Using Transcranial Real-Time Color Flow Doppler Sonography Takashi Tsuchiya 1 Masahiro Yasaka Takenori Yamaguchi

More information

Human Systems. Technology - Ultrasounds

Human Systems. Technology - Ultrasounds Human Systems Technology - Ultrasounds What is General Ultrasound Imaging? Ultrasound imaging, also called ultrasound scanning or sonography, involves exposing part of the body to high-frequency sound

More information

Image Formation (10) 2 Evaluation and Selection of Representative Images (10)

Image Formation (10) 2 Evaluation and Selection of Representative Images (10) STRUCTURED SELF ASSESSMENT CONTENT SPECIFICATIONS SSA LAUNCH DATE: JANUARY 1, 2018 Vascular Sonography The purpose of continuing qualifications requirements (CQR) is to assist registered technologists

More information

Vascular. Intraoperative Surgical Guidance and Quality Assessment. Immediate Feedback

Vascular. Intraoperative Surgical Guidance and Quality Assessment. Immediate Feedback Vascular Intraoperative Surgical Guidance and Quality Assessment Immediate Feedback Improve surgical outcomes, demonstrate quality, and increase cost efficency. The Medistim MiraQ Vascular System combines

More information

Vascular Sonography Examination

Vascular Sonography Examination Vascular Sonography Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Vascular Sonography Examination is to assess the knowledge and cognitive skills underlying the intelligent

More information

Thyroid and Parathyroid Ultrasound Protocol

Thyroid and Parathyroid Ultrasound Protocol Thyroid and Parathyroid Ultrasound Protocol Reviewed By: Anna Ellermeier, MD Last Reviewed: December 2017 Contact: (866) 761-4200, Option 1 **NOTE for all examinations: 1. If documenting possible flow

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

DOW-RAD, DOW DIAGNOSTIC COMPLEX, DUHS TRAINING PROGRAM HANDBOOK 2013

DOW-RAD, DOW DIAGNOSTIC COMPLEX, DUHS TRAINING PROGRAM HANDBOOK 2013 DOW-RAD, DOW DIAGNOSTIC COMPLEX, DUHS TRAINING PROGRAM HANDBOOK 2013 CERTIFICATE COURSE INVASCULAR/DOPPLER ULTRASOUND: Introduction: Ultrasound is an evolving technology with wide spectrum application

More information

Pattern of Doppler Flow Indices at the Carotid Bifurcation

Pattern of Doppler Flow Indices at the Carotid Bifurcation Article Pattern of Doppler Flow Indices at the Carotid Bifurcation Evaluation by Hemodynamic Color Doppler Imaging G. Sharat Lin, PhD, Veena S. Naval, MA Objective. To visualize the spatial variation of

More information

Case 8038 Renal allograft complicated with renal artery stenosis

Case 8038 Renal allograft complicated with renal artery stenosis Case 8038 Renal allograft complicated with renal artery stenosis Santiago I, Canelas A, Pinto AP Section: Cardiovascular Published: 2009, Nov. 30 Patient: 61 year(s), male Clinical History A 61-year-old

More information

Antegrade and retrograde flow of carotid

Antegrade and retrograde flow of carotid Antegrade and retrograde flow of carotid The ECA waveform is high resistance and may have retrograde flow in diastole.. They should always demonstrate antegrade flow (toward the brain) and be. external

More information

Ultrasound imaging is a noninvasive medical test that helps physicians diagnose and treat medical conditions.

Ultrasound imaging is a noninvasive medical test that helps physicians diagnose and treat medical conditions. CAROTID ULTRASOUND What is Carotid Ultrasound Imaging? Ultrasound imaging, also called ultrasound scanning or sonography, involves exposing part of the body to highfrequency sound waves to produce pictures

More information

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2016

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2016 STRUCTURED EDUCATION REQUIREMENTS Vascular Sonography The purpose of structured education is to provide the opportunity for individuals to develop mastery of discipline-specific knowledge that, when coupled

More information

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS ULTRASOUND QA SOLUTIONS Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS INNOVATORS IN ADVANCED ULTRASOUND TECHNIQUES Gammex is the

More information

Pre-and Post Procedure Non-Invasive Evaluation of the Patient with Carotid Disease

Pre-and Post Procedure Non-Invasive Evaluation of the Patient with Carotid Disease Pre-and Post Procedure Non-Invasive Evaluation of the Patient with Carotid Disease Michael R. Jaff, D.O., F.A.C.P., F.A.C.C. Assistant Professor of Medicine Harvard Medical School Director, Vascular Medicine

More information

11 TH ANNUAL VASCULAR NONINVASIVE TESTING SYMPOSIUM NOVEMBER 10, 2018

11 TH ANNUAL VASCULAR NONINVASIVE TESTING SYMPOSIUM NOVEMBER 10, 2018 11 TH ANNUAL VASCULAR NONINVASIVE TESTING SYMPOSIUM NOVEMBER 10, 2018 RENAL ARTERY DISEASE AND RENOVASCULAR HYPERTENSION 1 WHAT IS RENOVASCULAR HYPERTENSION? https://my.clevelandclinic.org/health/diseases/16459-renovascular-hypertension

More information

Vivid S6. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Vivid S6. Cardiovascular ultrasound system. GE Healthcare. Davis Medical GE Healthcare Vivid S6 Cardiovascular ultrasound system Taking care of everyone. Strong performance and innovative miniaturization make the Vivid S6 a highly desirable system for a variety of clinical

More information

Hemodynamic Analysis System

Hemodynamic Analysis System Hemodynamic Analysis System Advisor: Professor Mitch Tyler Client: Professor Naomi Chesler Team members: Sarah Czaplewski, Megan Jones, Sara Schmitz, and William Zuleger Overview Pulmonary Hypertension

More information

Introduction History Preceded by Arterial Doppler and ABI Indications

Introduction History Preceded by Arterial Doppler and ABI Indications Elise Brady, RVT, RDMS Introduction History Preceded by Arterial Doppler and ABI Indications 1) Abnormal ABI (within 2weeks of duplex) 2) Abnormal Doppler waveforms 3) Claudication 4) History of PVD 5)

More information

J. Laustsen ~, W. P. Paaske ~, S. Oyre 2 and E. Morre Pedersen ~'2

J. Laustsen ~, W. P. Paaske ~, S. Oyre 2 and E. Morre Pedersen ~'2 Eur J Vasc Endovasc Surg 9, 383-388 (1995) Dynamic Quantification, Visualisation and Animation of Blood Velocities and Flows in Infrarenal Aortic Aneurysms in vivo by Three-Dimensional MR Phase Velocity

More information

Ultrasound in Medicine

Ultrasound in Medicine Ultrasound in Medicine Experimental Equipment for Medical Education Universities Colleges Medical Schools Medical and Med-Technical Training Education can befun! WELCOME TO GAMPT Devices and accessories

More information

GLOBAL INNOVATION BY DESIGN

GLOBAL INNOVATION BY DESIGN GLOBAL INNOVATION BY DESIGN For over 130 years Toshiba s research and development has improved the health and welfare of people around the world. Today, Toshiba Medical Systems offers a full range of diagnostic

More information

Vascular Portfolio: Carotid Reflection. Paige Fabre

Vascular Portfolio: Carotid Reflection. Paige Fabre Vascular Portfolio: Carotid Reflection Paige Fabre 13654584 14 Carotid Reflection For this portfolio I produced three pieces of work; a case study, a PowerPoint of study protocol and a poster of stenosis

More information

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS Jeffrey P. Milligan, Daniel T. Peters, Structural Integrity Associates, Inc., USA Many advances in Non-Destructive Examination

More information

Basic Training Programme. 16 Februrary 2018, ROTTERDAM. Pre and Post-Course Test Answers

Basic Training Programme. 16 Februrary 2018, ROTTERDAM. Pre and Post-Course Test Answers Basic Training Programme 16 Februrary 2018, ROTTERDAM Pre and Post-Course Test Answers Your details: Name: Conference registration number/ BT delegate number: Email address: Are you already performing

More information

Brain Atrophy. Brain Atrophy

Brain Atrophy. Brain Atrophy Aging Central Nervous System Processes Age related brain atrophy Non-age related brain atrophy Cerebrovascular disease Cerebral infarction Hypertensive hemorrhage Carotid artery stenosis and occlusion

More information

Non-invasive examination

Non-invasive examination Non-invasive examination Segmental pressure and Ankle-Brachial Index (ABI) The segmental blood pressure (SBP) examination is a simple, noninvasive method for diagnosing and localizing arterial disease.

More information

Appendix II: ECHOCARDIOGRAPHY ANALYSIS

Appendix II: ECHOCARDIOGRAPHY ANALYSIS Appendix II: ECHOCARDIOGRAPHY ANALYSIS Two-Dimensional (2D) imaging was performed using the Vivid 7 Advantage cardiovascular ultrasound system (GE Medical Systems, Milwaukee) with a frame rate of 400 frames

More information

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography

B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with 64-MD-CT Angiography Med. J. Cairo Univ., Vol. 85, No. 2, March: 805-809, 2017 www.medicaljournalofcairouniversity.net B-Flow, Power Doppler and Color Doppler Ultrasound in the Assessment of Carotid Stenosis: Comparison with

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-449 Log Summary: Type of Modification Correct Value Name of Standard PS 3.16-2004 Rationale for Correction: Clinical applications use arterial and venous branches

More information

BEDSIDE ULTRASOUND BEDSIDE ULTRASOUND. Deep Vein Thrombosis. Probe used

BEDSIDE ULTRASOUND BEDSIDE ULTRASOUND. Deep Vein Thrombosis. Probe used BEDSIDE ULTRASOUND Part 2 Diagnosis of deep vein thrombosis Kishore Kumar Pichamuthu, Professor, Department of Critical Care, CMC, Vellore Summary: Deep vein thrombosis (DVT) is a problem encountered in

More information

Intraoperative Surgical Guidance and Quality Assessment

Intraoperative Surgical Guidance and Quality Assessment Intraoperative Surgical Guidance and Quality Assessment For best surgical outcomes Improve surgical outcomes, demonstrate quality, and increase cost efficiency. The Medistim MiraQ Cardiac combines ultrasound

More information

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care 2017 OPTIONS FOR INDIVIDUAL MEASURES: CLAIMS ONLY MEASURE TYPE:

More information

2015 ARDMS Physicians Vascular Interpretation Job Task Analysis Summary Report

2015 ARDMS Physicians Vascular Interpretation Job Task Analysis Summary Report P a g e 1 2015 ARDMS Physicians Vascular Interpretation Job Task Analysis Summary Report American Registry for Diagnostic Medical Sonography (ARDMS) P a g e 2 Table of Contents ABOUT THE REPORT... 3 METHODOLOGY...

More information

Duplex Carotid Sonography Peak Systolic Velocity in Quantifying Internal Carotid Artery Stenosis

Duplex Carotid Sonography Peak Systolic Velocity in Quantifying Internal Carotid Artery Stenosis Duplex Carotid Sonography Peak Systolic Velocity in Quantifying Internal Carotid Artery Stenosis Cynthia E Withers, MD", Barbara B Gosink, MD", Alison M Keightley, MD", Giovanna Casola, MD", Arthur A Lee,

More information

Terminology Tissue Appearance

Terminology Tissue Appearance By Marc Nielsen, MD Advantages/Disadvantages Generation of Image Ultrasound Machine/Transducer selection Modes of Ultrasound Terminology Tissue Appearance Scanning Technique Real-time Portable No ionizing

More information

Transducer Selection. Renal Artery Duplex Exam. Renal Scan. Renal Scan Echogenicity. How to Perform an Optimal Renal Artery Doppler Examination

Transducer Selection. Renal Artery Duplex Exam. Renal Scan. Renal Scan Echogenicity. How to Perform an Optimal Renal Artery Doppler Examination How to Perform an Optimal Renal Artery Doppler Examination Director of Ultrasound Education & Quality Assurance Baylor College of Medicine Division of Maternal-Fetal Medicine Maternal Fetal Center Imaging

More information

Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first

Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first choice examination for many disorders. If the subtlest

More information

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2 Page: 1 of 6 TABLE OF CONTENTS SECTION PAGE NO. 1.0 Introduction 2 2.0 Reference Documents 2 3.0 Summary of Practice 2 4.0 Significance & Use 3 5.0 Apparatus 3 6.0 Calibration & Adjustment of Apparatus

More information

Carotid Imaging. Dr Andrew Farrall. Consultant Neuroradiologist

Carotid Imaging. Dr Andrew Farrall. Consultant Neuroradiologist 20121123 SSCA http://www.neuroimage.co.uk/network Andrew Farrall Carotid Imaging Dr Andrew Farrall Consultant Neuroradiologist SFC Brain Imaging Research Centre (www.sbirc.ed.ac.uk), SINAPSE Collaboration

More information

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector ECNDT 2006 - Poster 198 The Evolution and Benefits of Phased Array Technology for the Every Day Inspector Dan KASS, Tom NELLIGAN, and Erich HENJES Olympus NDT, Waltham, USA Abstract. Phased arrays were

More information

DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS

DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS H. Ishida 1, Y. Kurozumi 1, and Y. Kaneshima 2 1 Institute of Nuclear Safety System,

More information

Half a century ago: when D. Eugene Strandness

Half a century ago: when D. Eugene Strandness Kirk W. Beach, Ph.D, M.D. VAPSHCS/GENERAL SURGERY Arterial Vibrometry Tissue Pulsatility Imaging of Brain Tissue Pulsatility Imaging of Placenta Multi-dimensional Doppler and Computational Hemodynamics

More information

Pipeline Technology Conference 2007

Pipeline Technology Conference 2007 The Complete Solution: Combined Crack and Metal Loss Detection Tool using Phased Array Technology By A. Hugger, J. Franz, H. Charbon, R. Bauernschmitt, M. Tschuch, K.-H. Käshammer, I. Lachtchouk, J. Ehrhardt.

More information

CT and MR Imaging in Young Stroke Patients

CT and MR Imaging in Young Stroke Patients CT and MR Imaging in Young Stroke Patients Ashfaq A. Razzaq,Behram A. Khan,Shahid Baig ( Department of Neurology, Aga Khan University Hospital, Karachi. ) Abstract Pages with reference to book, From 66

More information

It is generally accepted that with most of the

It is generally accepted that with most of the 846 Pulsed Multigated Doppler Ultrasonography in the Diagnosis of Carotid Artery Disease Henrik Sillesen, MD, Karen R. Bitsch, MD, Torben Schroeder, MD, Hans Jrgen Steenberg, MD, MSc, Linda Hansen, VT,

More information

ACUSON P500. Ultrasound Anytime, Anywhere. ACUSON P500. siemens.com/acusonp500. siemens.com/acusonp500 1

ACUSON P500. Ultrasound Anytime, Anywhere. ACUSON P500. siemens.com/acusonp500. siemens.com/acusonp500 1 Ultrasound Anytime, Anywhere. ACUSON P500 1 Enabling ultrasound imaging anytime, anywhere. 2 Ultrasound Anytime, Anywhere Siemens Healthineers engineered the compact and powerful ACUSON P500, a portable

More information