IAEA-TECDOC-796. Radiation doses in diagnostic radiology
|
|
- Sharlene Jordan
- 6 years ago
- Views:
Transcription
1
2 IAEA-TECDOC-796 Radiation doses in diagnostic radiology
3 The IAEA does
4 The originating Section
5 FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose from all man-made sources of radiation. Large differences in radiation doses from the same procedures among different X ray rooms have led to the conclusion that there is a potential for dose reduction. An exercise in dose reduction, while maintaining the quality of the diagnostic information in the image, is a genuine process of optimization associated with an improved use of the X ray equipment.
6 In preparing this publication EDITORIAL NOTE
7 CONTENTS 1. INTRODUCTION...
8 Radiation doses from diagnostic radiology 1. INTRODUCTION
9 reference levels
10 TABLE l. GENERAL INFORMATION DESCRIBING THE CRP PILOT PROGRAMME Country Argentina Brazil Czech Republic Ethiopia Ghana Iran, Islam. Rep. Romania Number of hospitals Number of X ray rooms To ensure a close interaction and co-operation with the radiology staff, a small number of hospitals was selected (see Table I and Appendix). Comparisons are only possible if some examinations are common to all participants, and preferably belonging
11 3.1. INDICATOR FOR DOSE ASSESSMENT In order to be able to assess patient doses, to compare them with the reference dose values and
12 TABLE
13 Initially, after annealing procedures, TLDs of each participating country were sent to the X ray Dosimetry Laboratory
14 TABLE V. TECHNICAL CHARACTERISTICS OF TLD SYSTEMS Country Argentina Brazil Czech Republic Ethiopia Iran, Islam.Rep. Ghana Romania TL reader Harshaw 2000 Harshaw Victoreen Vinten Solaro Harshaw Harshaw 2000 Harshaw 2000 TL material LiF-TLD 100 LiF-TLD 100 LiF-TLD 100 LiF-TLD 100 LiF-TLD 100 LiF-TLD 100 LiF-TLD 100 SD of batch (*) < 4 7 < 6 < 4.5 < 5 Annealing procedure 400 C/lh 100 C/3h 400 C/lh 100 C/2h 400 C/lh 80 C/16h 300 C/l,6h 80 C/17h 400 /lh 100 C/2h 400 C/lh 100 C/2h 400 C/lh 80 C/25h Annealing after each reading (Y/N) Y Y Y Y Y Y Y Reading process: ~TMAX -time -N 2 (Y/N) 300 C 15s Y Y 256 C 23s Y 300 C 12s Y 280 C 25s Y 300 C 20s Y 240 C 30s N Reading period after exposures (days) ~1 Calibration after each annealing procedure (Y/N) Source used for calibration Y ^Co Y «'Co Y 137 Cs Y»Sr Y «Co^'Cs Y Xray (48) kev Y 137 Cs Cleaning procedure (Y/N) Y Y Y Y Y Y N After each reading, the annealing procedures were performed according to manufacturer's recommendations and routine calibration using ß, y or X radiation was carried out by all the countries PLAN OF WORK
15 8. To organize the checking of the radiographie images using to the CEC quality criteria document. 9. To collect the occupational exposures data from the personnel dosimetry, and classify them in different ranges SCOPE
16 Filtering Additional filtration 4 is needed to avoid unnecessary skin exposure due to low energy components of the X ray spectrum. The latter do not actually contribute to image formation but are absorbed
17 Film processing
18 TABLE VI. CHARACTERISTICS OF TLD SYSTEMS EVALUATED DURING THE CALIBRATION OF THE PILOT PROGRAMME Energy response relative
19 TABLE VIL INTERCOMPARISON OF THE PILOT PROGRAMME OF TLD SYSTEMS Country Czech Rep. Ethiopia Ghana Iran, Islam. Rep. Romania TL group Air kerma (mgy) Effective energy (kev) Evaluated air kerma (mgy) (VC) (%) Bl (2.2) (3.7) 20.6 B (3.1) (8.9) B (6.6) (14.6)
20 Range of dose values The overall statistics of the dosimetry measurements are given in Table VIII for each examination type. One observes the large variations in dose, particularly for chest and abdomen examinations, as found
21
22 TABLE
23 TABLE X. (cont.) Brazil 3/2 Country Hospital/Room Iran, Islam. Rep. 1/1 Iran, Islam. Rep. 1/2 Iran, Islam. Rep. 1/3 Iran, Islam. Rep. 2/4 Romania 1/4 Argentina 3/5 Argentina 3/6 Iran, Islam. Rep. 2/4 Abdomen Abdomen Abdomen Abdomen Abdomen Abdomen Breast Breast Breast Examination and Projection Method Increased filtration Increased kv Increased
24 TABLE XI. COMPLIANCE WITH THE CEC REFERENCE DOSE VALUES Examination type Lumbar spine (PA) Pelvis Chest (PA) Abdomen Breast (4.5
25 In
26 TABLE Xm. REJECT RATES AND FILM USAGE BY X RAY ROOM AND BY COUNTRY Country Hospital/Room Argentina 1/1-1/2 Brazil 1/D Brazil 2/D Czech
27 J y 2 - CHEST ABDOMEN L. SPINE PELVIS MAMMOG. FIG.
28 20 15-' HSxl 10-' xt~tx1 o Too Dark Too Light Posit. Cent. Procès. Pat. Mot. Others FIG. 6. Rejected films (relative percentages) averaged over 22 rooms in six countries.
29 TABLE XIV. REJECT RATES BY CAUSE AND BY COUNTRY Country Hospital/Room Argentina (chest) Argentina (abdomen) Argentina (breast) Brazil Czech 1/1 (chest) Czech
30 Among the most important causes of reject, one observes inadequate setting of physical parameters which lead
31
32 REFERENCES
33 [20] SULEIMAN, O.H.,
34 Appendix SUMMARY OF REPORTS BY PARTICIPANTS This appendix deals with individual patient dose data and quality control measurement results obtained by each country participating in the CRP. It provides
35
36 Chief investigator ARGENTINA
37 TABLE I. SUMMARY OF THE MEASUREMENTS OF PATIENT ENTRANCE SURFACE DOSE (ARGENTINA) HOSPITAL
38 TABLE
39 U) oo TABLE III. COMPLIANCE WITH THE CEC IMAGE CRITERIA. CHEST (PA) (ARGENTINA) IMAGE CRITERIA Performed
40 Chief investigator BRAZIL
41 -po TABLE
42 TABLE IL QC PARAMETERS CONTROLLED (BRAZIL) PARAMETERS Film processing conditions kv Accuracy (10%)* kv Reproducibility (4%)* kv Consistency (10%)* HVL, mm of Al (measured at 80 kv) Timer accuracy (10%)* Timer reproducibility (5%)* ma- s linearity, consistency Output (mgy/ma- s) (5%)* Consistency Light/radiation beam alignment (deviation
43 to TABLE HI. COMPLIANCE WITH THE CEC IMAGE CRITERIA. CHEST (PA) (BRAZIL) IMAGE CRITERIA Performed
44 TABLE
45 TABLE VI. COMPLIANCE WITH THE CEC IMAGE CRITERIA. ABDOMEN (AP) (BRAZIL) IMAGE CRITERIA Symmetrical reproduction
46 CZECH REPUBLIC Chief investigator and research team: Chief investigator: Research team: O. Kodl I. Zachariasova V. Klener
47 TABLE
48 TABLE II. SUMMARY OF THE MEASUREMENTS OF PATIENT ENTRANCE SURFACE DOSE (CZECH REPUBLIC) HOSPITAL X RAY ROOM 1
49 TABLE III. QC PARAMETERS CONTROLLED (CZECH REPUBLIC) PARAMETERS Film Processing Conditions
50 TABLE
51 TABLE
52 TABLE VI. COMPLIANCE WITH THE CEC IMAGE CRITERIA. PELVIS (AP) (CZECH REPUBLIC) IMAGE CRITERIA Symmetrical reproduction of the pelvis Visualization of the sacrum and its intervertébral foramina Visualization
53 ETHIOPIA Chief investigator and research team: Chief investigator: Research team: Wondwosen Mengesha Hailu Wolde Tsegaye Tolossa Dereje Anbessei Institution(s): Radiation Protection Unit Institute
54 TABLE
55 TABLE II. QC PARAMETERS CONTROLLED (ETHIOPIA) PARAMETERS Film Processing Conditions kv Accuracy (10%)* kv Reproducibility (4%)* kv Consistency (10%)* HVL, mm of AI (measured
56 TABLE III. COMPLIANCE WITH THE CEC IMAGE CRITERIA. CHEST (PA) (ETHIOPIA) IMAGE CRITERIA Performed
57 ON TABLE
58 Chief investigator and research team: GHANA
59 TABLE I. SUMMARY OF THE MEASUREMENTS OF PATIENTS ENTRANCE SURFACE DOSE (GHANA) HOSPITAL
60 TABLE
61 TABLE III. COMPLIANCE WITH
62 TABLE IV. COMPLIANCE WITH THE CEC IMAGE CRITERIA. PELVIS (AP) (GHANA) IMAGE CRITERIA Symmetrical reproduction
63 O\ (O TABLE V. COMPLIANCE WITH THE CEC IMAGE CRITERIA. LUMBAR SPINE (AP) (GHANA) IMAGE CRITERIA Linear reproduction of the upper
64 Chief investigator and research team: Chief investigator: ISLAMIC REPUBLIC OF BRAN
65 TABLE I. SUMMARY OF THE MEASUREMENTS OF PATIENT ENTRANCE SURFACE DOSE (IRAN, ISLAM. REP.) HOSPITAL X RAY ROOM
66 TABLE II. QC PARAMETERS CONTROLLED (IRAN, ISLAM. REP.) PARAMETERS Film Processing Conditions kv Accuracy (10%)* kv Reproducibility (4%)* kv Consistency (10%)* HVL, mmofal (measured at 80 kv) Timer Accuracy (10%)* Timer Reproducibility (5%)* ma s Linearity, Consistency Output (mgy/ma s) (5%)* Consistency Light/Radiation Beam Alignment (deviation
67 o\ OS TABLE III. COMPLIANCE WITH THE CEC IMAGE CRITERIA. MAMMOGRAPHY (IRAN, ISLAM. REP.) IMAGE CRITERIA Visually sharp reproduction of
68 TABLE
69 GO TABLE V. COMPLIANCE WITH THE CEC IMAGE CRITERIA. ABDOMEN (AP) (IRAN, ISLAM. REP.) IMAGE CRITERIA* Symmetrical reproduction
70 ROMANIA Chief investigator and research team: Chief investigator: Research team: C. Milu
71 TABLE
72 TABLE II.
73 (O TABLE III. COMPLIANCE WITH THE CEC IMAGE CRITERIA. CHEST (PA) (ROMANIA) IMAGE CRITERIA Performed at deep inspiration and with suspended respiration Symmetrical reproduction
74 Annex
75
76 ABDOMEN AP/PA, FTRST ROUND OF MEASUREMENTS. RESULTS BY COUNTRY, HOSPITAL AND X RAY ROOM Countries Hospital Argentina Brazil Brazil Brazil Iran, Islam. Rep. Iran, Islam. Rep. Iran, Islam. Rep. Romania
77 CHEST AP/PA, FIRST ROUND
78 IVU, FIRST ROUND
79 oo MAMMOGRAPHY, FIRST ROUND OF MEASUREMENTS Countries Hospital Argentina Argentina Iran, Islam Rep
80 PELVIS
81 ABDOMEN AP/PA, SECOND ROUND OF MEASUREMENTS Countries Hospital Argentina Brazil Brazil Brazil Iran, Islam. Rep. Iran, Islam. Rep. Iran, Islam. Rep. Romania
82 CHEST PA, SECOND ROUND OF MEASUREMENTS Countries Hospital X ray Room Argentina Argentina Brazil Brazil Czech Republic C 2 D 2 2 Ethiopia Ethiopia Ghana Ghana Iran, Islam. Rep. Iran, Islam. Rep. Iran, Islam. Rep. Romania Romania
83 00 to IVUR, SECOND ROUND
84 MAMMOGRAPHY, SECOND ROUND OF MEASUREMENTS Countries Hospital Argentina Iran, Islam. Rep.
85 CRP RESULTS BY COUNTRY, BY HOSPITAL AND BY X RAY ROOM. (SKULL) - SECOND ROUND OF MEASUREMENTS Countries Hospital Brazil Brazil Brazil Brazil
86 Annex n EXAMPLES OF IMAGE QUALITY CRITERIA AND OF GOOD RADIOGRAPHIC TECHNIQUE, TAKEN FROM
87
88 LUNGS CHEST
89 AP/PA PROJECTIONS LUMBAR SPINE 1. DIAGNOSTIC REQUIREMENTS Image criteria 1.1 Linear reproduction of the upper and tower-plate surfaces in the centred beam area and visualisation of tite intervertébral spaces 1.2 Visually sharp reproduction of the pedicles 1.3 Visualisation of the intervertébral joints 1.4 Reproduction of the spinous and transverse processes 1.5 Visually sharp reproduction
90 Alm Carlsson, G. Benini,
Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology
Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationImpact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy
More informationTHE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS
THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS Larysa Stadnyk, Pylypenko M., Shalepa O., Nosyk O. Grigorev Institute for Medical Radiology,
More informationPATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS
Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,
More informationOPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY
March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD
More informationTesting of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution
Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi
More informationDosimetry in digital mammography
Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol
More informationP T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3
Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,
More informationPatient doses in general radiography examinations
peer reviewed ORIGINAL ARTICLE Patient doses in general radiography examinations Wambani JS 1 MBChB, M. Med (Rad), Paediatric (Rad) Onditi EG 2 MBChB, M.Med (Rad) Korir GK 3 B.Sc, MSc (Nuclear Science),
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationA Standardised Approach to Optimisation
The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained
More informationABSTRACT I. INTRODUCTION
2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography
More informationSurvey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations
Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein
More informationX-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS
X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology
More informationQuality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio
Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz
More informationReceptor dose and Patient Dose in Radiographic exposures - a 15 year review
Receptor dose and Patient Dose in Radiographic exposures - a 15 year review D J Peet*, N Tyler, M Pryor, P Hollaway, C Strudley and L Leavesley Regional Radiation Protection Service, Royal Surrey County
More informationQuality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis
60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis
More informationSPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS
SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)
More informationResearch Article. Corresponding author H. Osman
Scholars Journal of Applied Medical Sciences (SJAMS) ISSN 2320-6691 Sch. J. App. Med. Sci., 2013; 1(5):511-515 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific
More informationTrends in Patients Exposure Doses during Radiographic Examination in Japan
Trends in Patients Exposure Doses during Radiographic Examination in Japan S. Suzuki 1*, Y. Asada 1, M. Kobayashi 2, K. Kobayashi 2, Y. Matsunaga 3, T. Otsuka 4 1 School of Health Sciences, Fujita Health
More informationEnergy dependent response of Al 2 O 3 and its potential application in personal monitoring.
Energy dependent response of Al 2 O 3 and its potential application in personal monitoring. Nelson, V. K. a, b a Sydney University, Faculty of Health Sciences, Discipline of Medical Imaging and Radiation
More informationRadiation Doses to Patients Undergoing Selected Radiographic Examinations in Three Hospitals South-West, Nigeria
International Journal of Chemical and Biomedical Science 2017; 3(3): 26-31 http://www.aascit.org/journal/ijcbs ISSN: 2472-9574 (Print); ISSN: 2472-9590 (Online) Radiation Doses to Patients Undergoing Selected
More informationApplication of international standards to diagnostic radiology dosimetry
Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,
More informationTHE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES
IJRRAS 9 (1) October 11 www.arpapress.com/volumes/vol9issue1/ijrras_9_1_16.pdf THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES Bamidele
More informationMeasurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran
Global Journal of Health Science; Vol. 7, No. 5; 2015 ISSN 1916-9736 E-ISSN 1916-9744 Published by Canadian Center of Science and Education Measurement of Entrance Skin Dose and Calculation of Effective
More informationPatti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety
Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February 2008. Radiation Safety Sub -headings Background Radiation Effects of Radiation Safe Levels Effective Doses ALARA Principle Radiation
More informationRadiation Protection Services in R.Macedonia
Radiation Protection Services in R.Macedonia Lidija Nikolovska Institute of Public Health Ionizing Radiation Laboratory Skopje, Republic of Macedonia Background - no NPP or research reactors, - limited
More informationApplication of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology
Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni
More informationRadiologic Units: What You Need to Know
Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report
More informationIAEA TECDOC 1517 Quality Control in Mammography Software
IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.
More informationRadiation Protection Dosimetry Advance Access published April 22, 2008
Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE
More informationSUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES
A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith
More informationPatient dose in routine X-ray examinations in Yazd state
Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:
More informationShielding Calculation: Radiographic Room. Jerry Williams
Shielding Calculation: Radiographic Room Jerry Williams What you need to know Room use and layout DAP workload DAP averaged kv Distance to barrier Construction details Walls Ceilings/ floors Surrounding
More informationDETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA
DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,
More informationPhD Title: Improving QA/QC in mammography screening and breast in Montenegro
3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:
More informationInternational Journal of Radiation Research, January 2017
Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing
More informationModule Rhodes
Module 6 10-526-197 Rhodes Health Physicist Concerned with providing occupation radiation protection and minimizing radiation dose to the public. Diagnostic Imaging has changed our world Live longer Work
More informationEntrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals
Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and
More informationPNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY
PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN
More informationProf. Dr. Doğan BOR Ankara University Institute of Nuclear Science
PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural
More informationPatient Dosimetry in Mammography and Tomosynthesis:
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor
More informationPatient dose measurement in common medical X-ray examinations in Iran
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Patient dose measurement in common medical X-ray examinations in Iran Behrouz Rasuli, 1a Ali Mahmoud-Pashazadeh, 2 Mohammad Ghorbani,
More informationA Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria
International Journal of Medical Imaging 2015; 3(4): 69-74 Published online June 6, 2015 (http://www.sciencepublishinggroup.com/j/ijmi) doi: 10.11648/j.ijmi.20150304.11 ISSN: 2330-8303 (Print); ISSN: 2330-832X
More information2001 AAPM Summer School Seattle, Washington ACR R/F Phantom
2001 AAPM Summer School Seattle, Washington ACR R/F Phantom Charles R. Wilson, Ph.D., FACR Medical College of Wisconsin Milwaukee, Wisconsin MCMC Circa 1975 ACR R/F Phantom This lecture has been approved
More informationIntroduction and Background
CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section
More informationManaging the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University
Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm
More informationIAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*
THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service
More informationMeasurement of Absorbed Doses in Anatomical Phantoms
International Journal of Pure and Applied Physics. ISSN 0973-1776 Volume 8, Number 1 (2012), pp. 53-58 Research India Publications http://www.ripublication.com/ijpap.htm Measurement of Absorbed Doses in
More informationPatient effective dose evaluation for chest X-ray examination in three digital radiography centers
Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationPatient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran
Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Soraya Emamgholizadeh Minaei 1, Fariba Firouzi 1 2, 3,*, Hamid Reza Khosravi 1 Department of radiology, Shahid
More informationRadiation exposure of the Yazd population from medical conventional X-ray examinations
Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical
More informationThe need for standardization of dosimetry in experimental radiation biology
The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall
More informationEVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2
Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 7 doi:10.1093/rpd/ncn143 EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY
More informationBureau of Laboratory Quality Standards
1. Condom -Lubricant ISO 4074:2002/Cor.1:2003 Annex C -Thickness ISO 4074:2002/Cor.1:2003 Annex F - Length ISO 4074:2002 / Cor1 : 2003 - Width ISO 4074 : 2002/Cor 1 : 2003 - Bursting pressure and Volume
More informationDevelopment and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.
Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,
More informationEvaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography
Available online at http://www.biij.org/2006/3/e35 doi: 10.2349/biij.2.3.e35 biij Biomedical Imaging and Intervention Journal ORIGINAL PAPER Evaluation of radiation dose and image quality following changes
More informationDoses Received by Patients during Thorax X-Ray Examinations
Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,
More informationAccreditation of Dosimetry Services
Accreditation of Dosimetry Services J. Van Cauteren, E. De Geest, M. Bricoult AV-Controlatom Content History Regulatory requirements Discussion of reference documents/standards Practical implementation
More informationHong Kong College of Radiologists Mammography Statement
Hong Kong College of Radiologists Mammography Statement The Hong Kong College of Radiologists would like to give the following comments concerning mammography. Mammography screening: Breast cancer is the
More informationMammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433
- 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.
More informationReport 85 ISSN March 1997 ISRN ULI-RAD-R--85--SE
Report 85 ISSN 1102-1799 March 1997 ISRN ULI-RAD-R--85--SE Collection and analysis of patient and image data for calibration of a voxelphantom based Monte Carlo code and for the modelling of important
More informationFETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS
FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção
More informationEntrance Surface Air Kerma for Skull, Pelvis and Abdomen X-ray Examinations in Some Diagnostic Radiology Facilities in Akwa Ibom State, Nigeria
Radiation Science and Technology 201; 2(1): -12 http://www.sciencepublishinggroup.com/j/rst doi: 10.1148/j.rst.2010201.12 Entrance Surface Air Kerma for Skull, Pelvis and Abdomen X-ray Examinations in
More informationMEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS
RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU
More informationEvaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray
Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray Poster No.: C-1492 Congress: ECR 2012 Type: Scientific Exhibit Authors: R. M. Lopes, L. H. Domingos ; Mortágua/PT, Azoia de
More informationUsing Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography
Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,
More informationPractice and Risk at Medical Facilities in Agency Operations
Practice and Risk at Medical Facilities in Agency Operations Igor Gusev Radiation Protection Unit IAEA International Atomic Energy Agency Outline What is medical radiation exposure? Radiation sources and
More informationIranian Journal of Medical Physics
Iranian Journal of Medical Physics ijmp.mums.ac.ir Dose Evaluation for Common Digital Radiographic s in Selected Hospitals in Pahang Malaysia Soo-Foon Moey 1*, Zubir Ahmad Shazli 1, Inayatullah Shah Sayed
More informationStefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.
Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn
More informationInvestigation of the clinical performance of a novel solid-state diagnostic dosimeter
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and
More informationQuality Control Assessment of Radiology Devices in Kerman Province, Iran
Iranian Journal of Medical Physics Vol. 13, No. 1, March 16, 25-35 Received: January 19, 16; Accepted: March 11, 16 Original Article Quality Control Assessment of Radiology Devices in Province, Iran Abstract
More informationBICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:
BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, Mail Code 25-01
More informationRADIOLOGIC TECHNOLOGY PROGRAM COURSE DESCRIPTIONS
RADIOLOGIC TECHNOLOGY PROGRAM COURSE DESCRIPTIONS Term I: Radiographic Procedures I (3 credits) This course is the first in a series of courses dealing with principals, techniques and radiographic procedures
More informationMeasurements of Air Kerma Index in Computed Tomography: A comparison among methodologies
Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques
More informationMedical Diagnostic Imaging
Medical Diagnostic Imaging Laboratories Medical Diagnostic Imaging Lab Name Location Person in Charge Programs Served Courses Served Patient Care and Management (2) Introduction to MDI Radiographic Technique
More informationSurvey of Mammography Practice: Initial Results from Serbia
Survey of Mammography Practice: Initial Results from Serbia Olivera Ciraj Bjelac *, Dusko Kosutic, Danijela Arandjic, Milojko Kovacevic Vinca Institute of Nuclear Sciences, Radiation and Environmental
More informationBREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0
BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2
More informationTLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience
IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson
More information"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *
Iranian Journal of Science & Technology, Transaction A, Vol. 32, No. A3 Printed in the Islamic Republic of Iran, 2008 Shiraz University "Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 :
More informationSurvey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria Bidemi I.
More informationU.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS
U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS F. M. Cummings ^ R. D. Carlson^ R. M. Loesch INTRODUCTION Accreditation of personnel dosimetry systems
More informationThe New Jersey Radiographic Quality Assurance Program at 5 Years
The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided
More informationTo Shield or Not to Shield? Lincoln L. Berland, M.D.
To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.
More informationIntroduction 1. Executive Summary 5
Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationDental Extraoral X-ray Systems
Dental Extraoral X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL EXTRAORAL X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental
More informationIntroduction to RP154 methology for population dose estimates
Introduction to RP154 methology for population dose estimates Hilde M. Olerud (NRPA, NO) DDM 1 & DDM2 WP 3 Contract ENER/2010/NUCL/SI2.581237 Study on European Population Doses from Medical Exposure (Dose
More informationA Method for Statistical Process Control of Radiation Sterilization Facilities
A Method for Statistical Process Control of Radiation Sterilization Facilities 2006 A Method for Statistical Process Control (SPC) of Radiation Sterilization Facilities Scope This document describes a
More informationMedical dosimetry using Optically Stimulated Luminescence dots and microstar readers
Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers Christopher A Perks *a, Cliff Yahnke b, Marc Million c a LANDAUER EUROPE, Unit 28, Bankside, Station Approach, Kidlington,
More informationBone Densitometry Equipment Operator
Bone Densitometry Equipment Operator The purpose of the Bone Densitometry Equipment Operator Examination, which is made available to state licensing agencies, is to assess the knowledge and cognitive skills
More information2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals
2017 Course of the Nordic Association for Clinical Physics on occupational dosimetry in hospitals 27 September 2017 Dr. Pedro Ortiz López Retired member of the ICRP and the IAEA 2 Pedro Ortiz López (Chair)
More informationThe determination of timer error and its role in the administration of specified doses
JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School
More informationPilot study of patient and phantom breast dose measurements in Bulgaria
Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast
More informationAbsorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry
Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
233-3 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 29 Dosimetry Framework, Formalism including Uncertainties Costas Hourdakis Greek Atomic
More information