Dual Energy CT Attenuation Correction Methods for Quantitative Assessment of Response to Cancer Therapy with PET/CT Imaging

Size: px
Start display at page:

Download "Dual Energy CT Attenuation Correction Methods for Quantitative Assessment of Response to Cancer Therapy with PET/CT Imaging"

Transcription

1 Technology in Cancer Research and Treatment ISSN Volume 5, Number 4, August (2006) Adenine Press (2006) Dual Energy CT Attenuation Correction Methods for Quantitative Assessment of Response to Cancer Therapy with PET/CT Imaging We hypothesize that improved quantification for PET imaging of high atomic number materials can be achieved by combining low-dose x-ray imaging with dual energy CT for PET attenuation correction. Improved quantification of tracer uptake will lead to improved patient outcomes by providing more accurate information for therapeutic choices. Accurate PET/CT measurements of early response will be critical in determining the best cancer therapy option for each patient in a timely manner and in sparing patients the morbidity and cost of ineffective treatments. We first evaluate the potential errors in PET images arising from CT-based attenuation correction when iodine-based contrast is incorrectly classified as bone when forming the linear attenuation coefficient image. We then investigate two methods of reducing errors in the linear attenuation image: an approximate, but fast, hybrid classification/scaling algorithm and a model-based dual-energy CT method that incorporates the polyenergetic spectrum and a noise model in an iterative reconstruction method. Both methods are shown to reduce errors in the estimated linear attenuation coefficient image, but require further study to determine the effects of noise propagation if low-dose CT scans are used for the estimation of the linear attenuation image. Paul E. Kinahan, Ph.D. 1,* Adam M. Alessio, Ph.D. 1 Jeffrey A. Fessler, Ph.D. 2 1 Department of Radiology University of Washington 200 Old Fisheries Center Box Seattle, WA 98195, USA 2 Department of Electrical Engineering and Computer Science University of Michigan 4240 EECS Building 1301 Beal Ave. Ann Arbor, MI 48109, USA Introduction Positron emission tomography (PET) radiotracer imaging with the labeled glucose analog 18F-fluorodeoxyglucose (FDG) is being used increasingly in oncology imaging due to its usefulness in detecting and staging cancer and metastatic disease (1-4). A promising application of PET is tumor imaging with newly developed tracers for therapy monitoring with proliferation and apoptosis markers and definition of the tumor environment throughout therapy, constituting a strong basis for an individually tailored therapy for tumor patients (5-9). The advent of the dual modality PET/CT scanner (14) has significantly enhanced the physician s armamentarium for the diagnosis and staging of cancer as well as for therapy planning and monitoring response to therapy. PET/CT has become the most comprehensive diagnostic tool in oncology imaging by providing improved lesion identification and localization (15). Estimates of the numbers of FDG PET scans performed in the U.S. are shown in Figure 1, which also indicates the rapid saturation by PET/CT systems of the new PET scanner sales market. The primary purpose of combining CT and PET systems in a single scanner is the precise anatomical localization of regions identified on the PET tracer uptake images (16, 17). Although it is possible to use non-rigid image registration to align separately-acquired whole-body PET and CT images, challenges * Corresponding Author: Paul E. Kinahan, Ph.D. kinahan@u.washington.edu 319

2 320 Kinahan et al. Figure 1: Number of PET procedures per year. The dotted line is an exponential fit (r 2 = 0.99) indicating a doubling time of 19 months. Roughly 90% of PET procedures are oncology studies. Also shown is the percentage of new PET or PET/CT scanner sales that are PET/CT scanners (right axis). The total number of PET and PET/CT scanner sales has also been rising: 342 in 2001, 417 in 2002, and 450 in Data compiled from (10-13). remain in the practical implementation and validation of software-based methods (18, 19). Identifying Early Response to Cancer Therapy A widely used standard for monitoring the effect of cancer therapy on solid tumors is the evaluation of size changes measured from CT images. The Response Evaluation Criteria in Solid Tumors (RECIST) standard specifies a positive response as a 30% decrease in the maximum diameter (20). This approach has several problems, including the long time lag after therapy (six months or more) necessary for anatomical changes to become evident (21). By then if there is no response, time may have run out for other treatment options for the patient. In addition the patient may have undergone several months of ineffective but toxic therapy. Furthermore, cytostatic agents inhibit tumor growth when successful but may not lead to cell death and reduction in tumor size, in which case tumor size is not a good measure of response (22). While RECIST guidelines provide an important standard for evaluating response, alternative and robust methods of evaluating response by biochemical imaging are needed to improve patient management especially with increasing therapeutic choices (5). There has been increasing awareness of the value of identifying early response to cancer therapy. On January 10th, 2005, the National Cancer Institute (NCI) sponsored a Workshop on the Role of FDG-PET in the Evaluation of Therapeutic Response in Cancer. One of the conclusions from the workshop was that improved quantification is needed, including robust and accurate methods for PET/CT (23). Figure 2 illustrates the use of PET imaging to assess response to therapy. The images demonstrate an excellent response to therapy, which had been underestimated by conventional imaging modalities, including CT and bone scans. Some notable features are: The distribution of bony abnormalities is different between FDG PET and fluoride PET; and response to therapy is qualitatively and quantitatively apparent in the FDG PET image. The qualitative appearance of the fluoride image, however, is relatively stable over the course of treatment, and response to therapy is based largely on quantitative assessment. This emphasizes the need for accurate quantification. CT-based Attenuation Correction (CTAC) Figure 2: Change in glucose metabolism and fluoride incorporation in bone-dominant metastatic breast cancer. These images demonstrate glucose metabolism (FDG) and bone fluoride incorporation (F18) in a patient with widespread bony metastases from breast cancer before (left) and after hormonal therapy (right). Numbers shown are standardized uptake values (SUVs), a measure of relative tracer uptake. Data courtesy of Dr. David Mankoff, University of Washington. An important synergy of PET/CT scanners is the use of the CT images for attenuation correction of the PET emission data (24-26). All manufacturers of PET/CT scanners incorporate X-ray CT based attenuation correction (CTAC) algorithms in their systems, and for the majority of PET/CT scanners it is the only option offered (27). This approach offers the significant advantage that the CT data has much lower statistical noise and can be acquired in a shorter time than a standard PET transmission scan (24). CT transmission scans can also be acquired after the PET tracer is injected, allowing the ability to collect unbiased post-injection transmission scans. This reduces image bias from emission contamination while shortening the time spent by a patient on the scanner bed and providing more efficient use of scanner time. To be used for attenuation correction, the CT data must be transformed to an estimate of the attenuation coefficients at

3 Dual Energy PET/CT kev. However, there is no unique transformation from CT energies (~30 to 140 kev) to 511 kev due to the possibility of independent variations in density and atomic number (Z) (28). Two materials with different atomic numbers may have similar CT values but different attenuation coefficients at 511 kev. Conversely, it is possible for two distinct materials with the same value of attenuation coefficient at 511 kev to have different CT numbers. Biases in the CTAC image propagate to errors in the PET image in the same general location (29). Three methods have been considered for converting a CT image to attenuation coefficients at 511 kev: segmentation, scaling, and dual-energy X-ray scans. Segmentation: Methods can be used to separate the CT image into regions corresponding to different tissue types, which are then replaced with appropriate attenuation coefficients at 511 kev. However, some tissue and bone regions have continuously-varying densities that cannot be accurately represented by a discrete set of segmented values (30). The increased noise of low-dose CTAC scans can result in misclassification of voxels, with incorrect attenuation coefficients being assigned. For these reasons, segmentation methods have not been adopted for CTAC transforms, although they may have some application for the correction of focal accumulations of CT contrast agents (31, 32). Scaling: It is possible to estimate the attenuation map of the patient at 511 kev simply by multiplying the entire CT image by the ratio of attenuation coefficients of water (representing soft tissues) at the photon energies of CT and PET. For bone, however, linear scaling is a poor approximation, since photoelectric absorption dominates Compton scatter at the lower range of CT energies (33). Blankespoor et al. (34) used bilinear scaling to convert CT images to 140 kev for attenuation correction of SPECT data. In this method, different scaling factors (for water and air, and for water and bone, respectively) are used to calculate the attenuation values for CT numbers H for which < H < 0, and for H > 0. The bilinear scaling method has been shown to give reasonable results for low-z biological materials in practice (25, 35, 36). For high-z materials such as contrast agents there is some disagreement in the literature. Some investigators report no significant errors introduced by high-z materials when using scaling methods (37), while others have measured quantitative errors (32, 38-42). While in many cases any CTAC errors may not significantly affect diagnostic utility, they may affect decisions or therapies that depend on accurate estimation of tracer uptake in response to therapy. The errors introduced by high-z materials can be understood from Figure 3, which illustrates the standard bilinear method used for estimating the linear attenuation coefficient (μ(x,y) [cm -1 ]) image at 511 kev from the CT image. Each voxel in the CT image is scaled according to the bilinear transform, which has different slopes for air/water and water/bone mixtures, as described above (26). For iodine, the transformation that should be used is illustrated, but, unfortunately, there is no way to discriminate iodine from bone based on the CT voxel value alone. Thus, either voxels containing contrast are incorrectly scaled as bone, or if the iodine curve is used to produce correct results for contrast agent, then voxels containing bone are incorrectly scaled as contrast agent. Figure 3: Bilinear scaling factors used to convert CT numbers to linear attenuation coefficients at 511 kev. Also show is the transformation that should be used for iodine-based contrast agents. Dual Energy X-ray Imaging: A quantitatively accurate estimate of the linear attenuation coefficients at 511 kev can be obtained by collecting two CT scans using X-ray beams with different energy spectra and by estimating the energy dependence of attenuation coefficients in terms of a Compton scattering component and a photoelectric absorption component (43, 44), or by other material bases. The separate component images can then be combined to synthesize an accurate image of attenuation coefficients at any energy. Numerous medical applications of dual-energy imaging, such as bone mineral density measurements (45), as well as non-medical applications have been explored (46). Dual-energy CTAC would allow for accurate attenuation correction in PET/CT imaging that involves high-z materials, including bone, contrast, and metals. This approach was used to form a monoenergetic attenuation map at 140 kev by Hasegawa et al. for a prototype SPECT/CT detector block (47) and for separate SPECT and CT scans by Guy et al. (48). A drawback of the dual-energy CT method is that the inverse problem of estimating the component sinograms is poorly conditioned (49), leading to excessive noise amplification. To make dual-energy CTAC feasible for PET/CT imaging, additional steps are needed to reduce noise in the estimated attenuation image at 511 kev. Image reconstruction algorithms that model the acquisition physics can reduce statistical noise in

4 322 Kinahan et al. DECT. There have, however, been very few iterative image reconstruction algorithms proposed for DECT. Sukovic and Clinthorne have investigated iterative algorithms for dualenergy CT reconstruction based on a weighted least-squares approach assuming monoenergetic scans (50). We describe below a polyenergetic version of the Sukovic and Clinthorne weighted least-squares algorithm for CTAC estimation appropriate for PET/CT scanners (51, 52), and evaluate its usefulness for dual energy CTAC. We call this method the dualenergy weighted least-squares algorithm (DE-WLS). We also investigate an approximate dual-energy method (called hybrid DECTAC) that allows discrimination of contrast from bone (53, 54). This allows the selection of correct scaling factors for bone versus contrast, but not necessarily other materials. Recently Bacharach et al. described the same approach (55). Watson et al. described a related method (56) that used dual-energy CT to correctly estimate bone linear attenuation values. We first evaluate the potential errors in PET images arising from CT-based attenuation correction when contrast is incorrectly classified as bone in the estimation of the linear attenuation coefficient image. We then investigate the two methods described above for reducing errors in the linear attenuation image: the approximate, but fast, hybrid classification/scaling algorithm and the model-based dual-energy CT method that incorporates the polyenergetic spectrum, and a noise model in an iterative reconstruction method. Materials and Methods Impact of Errors from Contrast Agent Enhancement We first investigate the bias introduced by incorrect estimation of linear attenuation coefficients estimated by the standard bi-linear method. This used computer simulations of a abdomen-sized object with a set of six test objects (5 cm diameter regions) of differing standardized uptake values (SUVs) with two different local background SUVs as illustrated in Figure 4 (left). The two background regions had SUVs of 1.0 and 2.0, while the test objects had SUVs of 0.5, 1.0, 2.0, and 3.0 as shown. The corresponding attenuation values for the six objects should ideally have the linear attenuation coefficient as background. We simulated the effect of contrast agent being incorrectly scaled as bone, as illustrated in Figure 4 (right). The effect of contrast agent enhancement of 20, 100, and 500 Hounsfield unites (HU) were evaluated for each SUV of the test objects. Method 1: Hybrid Dual-Energy CTAC (DECTAC) The proposed dual-energy hybrid method classifies materials based on their changes in CT number from scans at two different kvp, taking advantage of the large differences in the photoelectric cross-sections of iodine and barium versus calcium (57) to determine how to scale each voxel in the CTAC image. Since bone and contrast values span a wide range of CT values, we have derived a classification scheme for the CT difference images (e.g., 80 and 140 kvp) by simply using the midpoint of the relative change in CT numbers, in other words the HU difference depends on the HU value to account for variable concentrations of contrast and variable bone density (Fig. 5) (53). When the correct material (bone or contrast) is determined, then the appropriate scaling factor from Figure 3 is applied to the corresponding voxel in the first CT image. To test this approach, we acquired CT measurements at 80, 100, 120, and 140 kvp of a 20 cm diameter test phantom containing water and 5 cm diameter cylinders of air, dilute iodine-based contrast agent and CaCl 2 in solution (bone equivalent atomic number). The resulting CT numbers were evaluated at each tube voltage. The standard bilinear CT scaling method and the hybrid method were applied to Figure 4: Test objects used for simulation study investigating the propagation of errors from biased attenuation correction factors (CTAC with errors) into the attenuation corrected PET image. Figure 5: DE-CTAC Classification scheme. Contrast versus bone discrimination is accomplished by determining if the change in CT number is greater than a threshold, which depends on the CT number for one of the kvp images.

5 Dual Energy PET/CT 323 the 140 kvp CT image. A standard PET transmission scan was also acquired for reference. Method 2: Dual-Energy Weighted Least-squares Algorithm (DE-WLS) The above Hybrid DE-CTAC hybrid scheme will help discriminate contrast from bone, but will not suffice if unknown or multiple high-z materials are present. We have implemented a new PWLS algorithm for dual energy CT similar to that of Sukovic and Clinthorne (50), but with a polyenergetic model of the spectrum, rather than their monoenergetic approach (51, 52). We term this the dual-energy weighted least-squares (DE-WLS) algorithm. To derive the algorithm we start with a parameterization that uses basis functions that are each separable in space and energy μ(x,e) = Σ l=1,l Σ j=1,np β l (E) b j (x )x lj, where β l (E) is the mass attenuation of the lth material type, b j (x ) are spatial basis functions, and x lj are the desired coefficients for material l and voxel j. Thus, the spatially-varying density of each material is given by r l (x ) = Σ j b j (x ) x lj and the system model is given by a ij = b L i j (x )dl, where L i denotes the ith line of response (LOR). The material basis sinograms are then given by To solve for x, we use the PWLS criterion: x = arg min Ψ(x), Δ N with Ψ(x) = Σ D (sˆ i s i (x))ʹw i (sˆ w h e r e i=1 i s i (x)) + R(x), N D is the number of lines of response, and s ˆi are the material basis sinograms estimated from the CT sinograms using standard dual-energy decomposition methods, and W i are the inverse variances of s ˆi. The function R(x) is a regularizer that controls noise. We tested the DE-WLS approach with simulated and measured data. For the simulations we used a 20 cm diameter phantom containing three smaller 5 cm diameter cylinders comprised of (i) air, (ii) half soft tissue and half bone with a total density of 1.5 g/cm 3 ; and (iii) a bone cylinder with a total density of 2 g/cm 3. The large background cylinder was all soft tissue component with a density of 1 g/cm 3. The attenuation coefficients at 511 kev were estimated by (i) standard bilinear scaling, (ii) dual energy CT with a standard dual-energy decomposition method followed by filtered backprojection (FBP) and the DE-WLS algorithm. Results Δ s il (x l ) = Li ρ l (x)dl = Σi=1,N p a ij x lj. Impact of Errors from Contrast Agent Enhancement Figure 6 illustrates the effects of biases from CT-based attenuation correction for noiseless PET simulations. The difference image between the true PET image (Fig. 4) and the reconstructed image is also shown. The errors, expressed as standardized uptake values (SUV), are plotted in Figure 7 as a function of the SUV of the test object for the different levels of contrast-induced CT number errors. Figure 6: Noiseless simulation study showing propagation of errors from biased attenuation correction factors into the attenuation corrected PET image (PET EM using CTAC). A difference image from the true PET image (Fig. 4) shows that both quantitative errors and complex artifacts can occur. Error [SUV] Method 1: Hybrid Dual-Energy CTAC (DECTAC) 500 HU error 100 HU error 20 HU error True SUV Figure 7: Results of simulation study showing propagation of CT error of different Hounsfield units (HU) into the error in the attenuation corrected PET standard uptake values (SUV). For the measured phantom data there was essentially no variation in CT number for air or water with tube voltage. For the CaCl 2 solution the CT numbers increased from 546 to 897 HU, while the dilute contrast agent values increased from 856 to 1721 HU as the tube voltage was dropped from 140 to 80 kvp. Thus, starting from a standard CT scan at 120 kvp, a second (low-dose) CT scan at 80 kvp would show a difference of ~400 HU between contrast agents and CaCl 2 (bone equivalent). This is similar to heuristic bone-imaging procedures for CT. One of the dual-energy difference images is shown in Figure 8, which shows that there is no variation in CT number with kvp for both air and water, while there are differing variations for bone and contrast agent. The difference image is sensitive to the photoelectric absorption, which is higher in iodine-based contrast agent than in bone.

6 324 Kinahan et al. from 11% using bi-linear scaling to 9% with DE- FBP and to 7.5% with DE-WLS. Figure 8: Measured 20 cm diameter water cylinder with 5 cm cylindrical inserts containing air, bone-equivalent solution of CaCl 2 and dilute iodine-based contrast agent scanned at two different kvp. By classifying the voxels in the CT image (Fig. 8) based on the scheme illustrated in Figure 5, the correct scaling factors for bone and contrast agent (Fig. 3) can be used to convert the CT image to an estimate of the linear attenuation coefficients at 511 kev. This was evaluated as shown in Figure 9 for the same test phantom used in Figure 8. The bilinear method has a substantial error for the 5 cm diameter contrast agent filled region, when compared to the measured PET transmission scan at 511 kev. This error is reduced from -38% to + 6% with the use of the dual energy hybrid/bilinear method. The computation time of the DE-WLS (with ten iterations) and FBP methods are shown in Figure 11 for a standard workstation. There is a significant increase in computation time required for the DE-WLS, which would render it infeasible for clinical implementation with standard CT images. For attenuation correction of PET data, however, only (or perhaps even 64 64) images are needed, for relatively thick slices, which reduces the DE-WLS computation time down to clinically feasible levels. Figure 10: Comparison of CT-based attenuation images scaled to 511 kev by either bilinear scaling or dual energy CT. Also shown (bottom row) are the differences from the true values at 511 kev. Discussion Figure 9: Effect of bilinear and hybrid scaling methods on measured data, compared to measured PET transmission scan (TX). Currently, CT attenuation correction has the potential for significant bias in the attenuation correction factors. We present results showing the propagation of these errors into Method 2: Dual-Energy Weighted Least-squares Algorithm (DE-WLS) Results of the simulation studies are shown in Figure 10. The CT-based attenuation images were scaled to 511 kev by either bilinear scaling or dual energy CT. The dual-energy CT images were reconstructed with both FBP and the DE-WLS algorithms. Also shown (bottom) are the differences from the true values at 511 kev. There is a significant reduction of bias using DECT (from 20% to ~0% for the tissue/bone mixture) compared to bilinear scaling. The FBP image, however, has high levels of statistical noise, as expected. These noise levels are reduced from ~8% to ~3% by the use of our DE-WLS algorithm. Within the object RMS errors decrease Figure 11: Computation time for the DE-WLS and FBP algorithms as a function of image size when implemented on a standard workstation.

7 Dual Energy PET/CT 325 attenuation corrected PET images. Dual energy CT-based attenuation correction offers the potential of reducing bias in the CTAC image and, thus, can also reduce bias in the final PET emission image. Dual energy CT attenuation correction results in increased noise in reconstructions from the basis material decomposition process and from the potential use of low-dose CT. Noise from dual energy CT imaging can be reduced through the use of: (i) a statistically principled reconstruction method, (ii) a coarse reconstruction grid, and (iii) only the summed component image at 511keV. Both the simulated and measured phantom results demonstrate the feasibility of using dual-energy CTAC for accurate attenuation correction in PET/CT oncology imaging. We have shown that with the dual-energy hybrid method it is possible to discriminate contrast versus bone in the CTAC images and apply the correct scale factor for accurate calibration and attenuation correction. We note that this method may not be suitable for standard high resolution CT images due to patient motion or noise amplification. However, for low-resolution (or even 64 64) CTAC images reconstructed with statistically-principled algorithms we expect that the noise amplification will be acceptable. We also evaluated a polychromatic DE-PWLS iterative reconstruction method that is potentially more accurate than the dual-energy hybrid method. We showed that the reduction of the CTAC sinograms and image matrix sizes to or leads to iterative reconstructions of dual energy CT component images in a clinically feasible amount of time. In addition to errors from scaling the CTAC image to 511 kev described above, three other sources of error from CTbased attenuation have been recognized (14, 26): CT field of view truncation artifacts; CT beam hardening; and respiratory motion mismatch between the PET and CT acquisitions. An ameliorating factor for truncation artifacts is that missing attenuation correction factors can be closely approximated from the truncated projection data, leading to accurate CTAC images (58). We also note that beam hardening effects are minimized by the use of a dual energy CT method. Research in compensating for respiratory motion are ongoing. There is a clear consensus that PET/CT imaging has significant potential as biomarker for assessing response to therapy. Recent recommendations for such PET imaging protocols have been announced by the National Cancer Institute (23). In addition the Cancer Imaging Program at NCI is sponsoring the development of a database of CT and PET/CT images pre- and post-therapy as a testing ground for determining if there has been a response to therapy, by using anatomical and/or functional information (59). This RIDER (Reference Image Database to Evaluate Response) database is intended to enable industry and academia to develop, test, and compare semi-automated and automated software tools for change analysis in response to therapy. A key initial step towards this project is a consensus on procedures for obtaining quantitatively accurate PET images from PET/CT scanners. Conclusion We presented two successful DECT strategies for reducing quantitative errors from CT-based attenuation correction. The quantitative biases present in PET/CT imaging clearly do not prevent its effective use in the diagnosis and staging of cancer. However, a key component of the use of PET in the management of cancer will be for evaluating response to therapy and for making therapeutic choices. These applications require the use of quantitative PET/CT imaging to assure they have a positive impact on patient management. Accurate PET/CT measurements of early response will be critical in determining the best cancer therapy option for each patient in a timely manner and in sparing patients the morbidity and cost of ineffective treatments. Acknowledgements We would like to acknowledge helpful discussion with Drs. David Mankoff and Hubert Vesselle in the Department of Radiology at the University of Washington, and Drs Barbara Croft and Lawrence Clarke in the Cancer Imaging Program of the National Cancer Institute. This work was supported by NCI Grants R01-CA74135 and R01-CA References 1. Wahl, R. L. Clinical Oncology Update: The Emerging Role of Positron Emission Tomography: Part I. Principal and Practice of Oncology Updates 11, 1-18 (1997). 2. Weber, W. A., Avril, N., Schwaiger, M. Relevance of Positron Emission Tomography (PET) in Oncology. Strahlentherapie und Onkologie 175, (1999). 3. Bomanji, J. B., Costa, D. C., Ell, P. J. Clinical Role of Positron Emission Tomography in Oncology. Lancet Oncology 2, (2001). 4. Rohren, E. M., Turkington, T. G., Coleman, R. E. Clinical Applications of PET in Oncology. Radiology 231, (2004). 5. Krohn, K. A., Mankoff, D. A., Eary, J. F. Imaging Cellular Proliferation as a Measure of Response to Therapy. Journal of Clinical Pharmacology Suppl, 96S-103S (2001). 6. Mankoff, D. A., Muzi, M., Krohn, K. A. Quantitative Positron Emission Tomography Imaging to Measure Tumor Response to Therapy: What is the Best Method? Molecular Imaging & Biology 5, (2003). 7. Cohade, C., Wahl, R. L. PET Scanning and Measuring the Impact of Treatment. Cancer J 8, (2002). 8. Stahl, A. et al. Positron Emission Tomography as a Tool for Translational Research in Oncology. Molecular Imaging in Biology 6, (2004). 9. Juweid, M. E., Cheson, B. D. Positron-emission Tomography and Assessment of Cancer Therapy. N. Engl. J. Med. 354, (2006). 10. Burns, M. PET Market Appears Poised for Continued Strong Growth. Diagnostic Imaging (2003).

8 326 Kinahan et al. 11. Ridley, E. L. PET/CT Growth Sparks Gains in U.S. PET Market. AuntMinnie.com (2003). 12. Staff. PET Usage Surges in AuntMinnie.com (2003). 13. Staff. PET Procedure Volume to Surge. AuntMinnie.com (2004). 14. Beyer, T. et al. A Combined PET/CT Scanner for Clinical Oncology. Journal of Nuclear Medicine 41, (2000). 15. Czernin, J. Summary of Selected PET/CT Abstracts from the 2003 Society of Nuclear Medicine Annual Meeting. Journal of Nuclear Medicine 45, 102S-103S (2004). 16. Wahl, R. L. et al. Anatometabolic Tumor Imaging: Fusion of FDG PET with CT or MRI to Localize Foci of Increased Activity. J. Nucl. Med. 34, (1993). 17. Townsend, D. W., Beyer, T., Blodgett, T. M. PET/CT Scanners: A Hardware Approach to Image Fusion. Semin. Nucl. Med. 33, (2003). 18. Slomka, P. J. Software Approach to Merging Molecular with Anatomic Information. J. Nucl. Med. 45 Suppl. 1, 36S-45S (2004). 19. Wahl, R. L. Why Nearly All PET of Abdominal and Pelvic Cancers will be Performed as PET/CT. Journal of Nuclear Medicine 45, 82S- 95S (2004). 20. Therasse, P. et al. New Guidelines to Evaluate the Response to Treatment in Solid Tumors. European Organization for Research and Treatment of Cancer, National Cancer Institute of the United States, National Cancer Institute of Canada. J. Natl. Cancer Inst. 92, (2000). 21. Husband, J. E. et al. Evaluation of the Response to Treatment of Solid Tumours A Consensus Statement of the International Cancer Imaging Society. Br. J. Cancer 90, (2004). 22. Michaelis, L. C., Ratain, M. J. Measuring Response in a Post-RE- CIST World: From Black and White to Shades of Grey. Nat. Rev. Cancer 6, (2006). 23. Shankar, L. K. et al. Consensus Recommendations for the Use of 18F-FDG PET as an Indicator of Therapeutic Response in Patients in National Cancer Institute Trials. J. Nucl. Med. 47, (2006). 24. Kinahan, P. E., Townsend, D. W., Beyer, T., Sashin, D. Attenuation Correction for a Combined 3D PET/CT Scanner. Med. Phys. 25, (1998). 25. Burger, C. et al. PET Attenuation Coefficients from CT Images: Experimental Evaluation of the Transformation of CT into PET 511- kev Attenuation Coefficients. Eur. J. Nucl. Med. Mol. Imaging 29, (2002). 26. Kinahan, P. E., Hasegawa, B. H., Beyer, T. X-ray Based Attenuation Correction for PET/CT Scanners. Seminars in Nuclear Medicine 33, (2003). 27. Alessio, A., Kinahan, P. E., Cheng, P., Vesselle, H., Karp, J. S. PET/ CT Scanner Instrumentation, Challenges, and Solutions. Radiologic Clinics of North America 42, (2004). 28. Schneider, W., Bortfeld, T., Schlegel, W. Correlation Between CT Numbers and Tissue Parameters Needed for Monte Carlo Simulations of Clinical Dose Distributions. Physics in Medicine and Biology 45, (2000). 29. Bai, C. Y. et al. An Analytic Study of the Effects of Attenuation on Tumor Detection in Whole-body PET Oncology Imaging. Journal of Nuclear Medicine 44, (2003). 30. Robinson, P. J., Kreel, L. Pulmonary Tissue Attenuation with Computed Tomography: Comparison of Inspiration and Expiration Scans. Journal of Computer Assisted Tomography 3, (1979). 31. Carney, J., Beyer, T., Brasse, D., Yap, J. T., Townsend, D. W. CT- Based Attenuation Correction for PET/CT Scanners in the Presence of Contrast Agent. IEEE Nuclear Science Symposium and Medical Imaging Conference (2002). 32. Nehmeh, S. A. et al. Correction for Oral Contrast Artifacts in CT Attenuation-corrected PET Images Obtained by Combined PET/CT. Journal of Nuclear Medicine 44, (2003) LaCroix, K. J., Tsui, B. M. W., Hasegawa, B. H., Brown, J. K. Investigation of the Use of X-ray CT Images for Attenuation Compensation in SPECT. IEEE Transactions on Nuclear Science NS-41, (1994). Blankespoor, S. C. et al. Attenuation Correction of SPECT Using X-Ray CT on an Emission-Transmission CT System: Myocardial Perfusion Assessment. IEEE Transactions on Nuclear Science 43, (1996). Kamel, E. et al. CT vs 68Ge Attenuation Correction in a Combined PET/CT System: Evaluation of the Effect of Lowering the CT Tube Current. Eur. J. Nucl. Med. Mol. Imaging 29, (2002). Nakamoto, Y. et al. PET/CT: Comparison of Quantitative Tracer Uptake Between Germanium and CT Transmission Attenuation-corrected Images. J. Nucl. Med. 43, (2002). Dizendorf, E., Hany, T. F., Buck, A., von Schulthess, G. K., Burger, C. Cause and Magnitude of the Error Induced by Oral CT Contrast Agent in CT-based Attenuation Correction of PET Emission Studies. Journal of Nuclear Medicine 44, (2003). Antoch, G. et al. Effect of Oral Contrast Agents on Computed Tomography-based Positron Emission Tomography Attenuation Correction in Dual-modality Positron Emission Tomography/Computed Tomography Imaging. Investigative Radiology 38, (2003). Bujenovic, S., Mannting, F., Chakrabarti, R., Ladnier, D. Artifactual 2-deoxy-2-[(18)F]fluoro-D-glucose Localization Surrounding Metallic Objects in a PET/CT Scanner using CT-based Attenuation Correction. Mol. Imaging Biol. 5, (2003). Goerres, G. W., Ziegler, S. I., Burger, C., von Schulthess, T., Buck, A. Artifacts at PET and PET/CT Caused by Metallic Hip Prosthetic Material. Radiology 226, (2003). Nakamoto, Y. et al. Effects of Nonionic Intravenous Contrast Agents at PET/CT Imaging: Phantom and Canine Studies. Radiology 227, (2003). Visvikis, D. et al. CT-based Attenuation Correction in the Calculation of Semi-quantitative Indices of F-18 FDG Uptake in PET. European Journal of Nuclear Medicine and Molecular Imaging 30, (2003). Alvarez, R., Seppi, E. Comparison of Noise and Dose in Conventional and Energy Selective Computed-Tomography. IEEE Transactions on Nuclear Science 26, (1979). Alvarez, R. E., Macovski, A. Energy-Selective Reconstructions in X- Ray Computerized Tomography. Physics in Medicine and Biology 21, (1976). Greenfield, M. A. Current Status of Physical Measurements of the Skeleton. Medical Physics 19, (1992). Engler, P., Friedman, W. D. Review of Dual-Energy Computed-Tomography Techniques. Materials Evaluation 48, (1990). Hasegawa, B. H. et al. Object Specific Attenuation Correction of SPECT with Correlated Dual-energy X-ray CT. IEEE Transactions on Nuclear Science NS-40, (1993). Guy, M. J. et al. DETECT Dual Energy Transmission Estimation CT For Improved Attenuation Correction in SPECT and PET. IEEE Transactions on Nuclear Science 45, (1998). Kalender, W. A., Klotz, E., Kostaridou, L. An Algorithm for Noise Suppression in Dual Energy CT Material Density Images. IEEE Transactions on Medical Imaging 7, (1988). Sukovic, P., Clinthorne, N. H. Penalized Weighted Least-squares Image Reconstruction for Dual Energy X-ray Transmission Tomography. IEEE Transactions on Medical Imaging 19, (2000). Fessler, J. A., Elbakri, I. A., Sukovic, P., Clinthorne, N. H. Maximumlikelihood Dual-energy Tomographic Image Reconstruction. Medical Imaging 2002: Image Processing 4684, (2002). Kinahan, P. E., Fessler, J. A., Alessio, A. M., Lewellen, T. K. Quantitative Attenuation Correction for PET/CT Using Iterative Reconstruction of Low-dose Dual-energy CT. Nuclear Science Symposium Conference Record, 2004 IEEE Nuclear Science Symposium Conference Record, 2004 IEEE 5, Vol. 5 (2004).

9 Dual Energy PET/CT Kinahan, P. E. et al. Dual-Energy CT For Quantitative Attenuation Correction in PET/CT Imaging RSNA (abstract) (2004). Kinahan, P. E., Alessio, A. M., Fessler, J. A., Lewellen, T. K., Vesselle, H. A Simple Dual Energy Method for CT-Based Attenuation Correction that Accounts for Contrast Agents in PET/CT Imaging. Journal of Nuclear Medicine 46, 113P (abstract) (2005). Bacharach, S. L., Freedman, N. M. T., Maass-Moreno, R., Le Meunier, L. Dual Energy CT for Detecting Contrast Media in PET/CT. Journal of Nuclear Medicine 46, 168P (abstract) (2005). Watson, C. C., Rappoport, V., Faul, D., Townsend, D. W., Carney, J. P. A Method for Calibrating the CT-based Attenuation Correction of PET in Human Tissue. Nuclear Science, IEEE Transactions on Nuclear Science, IEEE Transactions on 53, (2006) Hubbell, J. H., Seltzer, S. M. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients. JOURNAL?! January, 2003, (1997). Hsieh, J. et al. Algorithm to Extend Reconstruction Field-of-View IEEE International Symposium on Biomedical Imaging (2004). Clarke, L. P. et al. Building of Large Public Databases For Validation of Software Tools. Sixth National Forum on Biomedical Imaging in Oncology (2005). Received: April 21, 2006; Revised: July 2, 2006; Accepted: July 5, 2006

1 Department of Radiology, University of Washington 2 Department of Electrical Engineering and Computer Science, University of Michigan

1 Department of Radiology, University of Washington 2 Department of Electrical Engineering and Computer Science, University of Michigan Dual Energy CT Attenuation Correction Methods for Quantitative Assessment of Response to Cancer Therapy with PET/CT Imaging. Paul E. Kinahan 1, Adam M. Alessio 1, Jeffrey A. Fessler 2 1 Department of Radiology,

More information

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? Typical PET Image Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located? PET/CT Oncology Imaging Anatometabolic fusion images are useful in the management

More information

Photon Attenuation Correction in Misregistered Cardiac PET/CT

Photon Attenuation Correction in Misregistered Cardiac PET/CT Photon Attenuation Correction in Misregistered Cardiac PET/CT A. Martinez-Möller 1,2, N. Navab 2, M. Schwaiger 1, S. G. Nekolla 1 1 Nuklearmedizinische Klinik der TU München 2 Computer Assisted Medical

More information

Biases affecting tumor uptake measurements in FDG-PET

Biases affecting tumor uptake measurements in FDG-PET Biases affecting tumor uptake measurements in FDG-PET M. Soret, C. Riddell, S. Hapdey, and I. Buvat Abstract-- The influence of tumor diameter, tumor-tobackground activity ratio, attenuation, spatial resolution,

More information

Reproducibility of Uptake Estimates in FDG PET: a Monte Carlo study

Reproducibility of Uptake Estimates in FDG PET: a Monte Carlo study Reproducibility of Uptake Estimates in FDG PET: a Monte Carlo study Juliette Feuardent, Marine Soret, Irène Buvat 1 Abstract Tumor glucose metabolism measurements from Fluoro-deoxyglucose (FDG) Positron

More information

Identifying Image Artifacts, Their Causes and How to Fix Them: PET. Brad J Kemp, PhD Mayo Clinic, Rochester, MN

Identifying Image Artifacts, Their Causes and How to Fix Them: PET. Brad J Kemp, PhD Mayo Clinic, Rochester, MN Identifying Image Artifacts, Their Causes and How to Fix Them: PET Brad J Kemp, PhD Mayo Clinic, Rochester, MN Case 1 Can we scan with a defective block detector? Daily Quality Assurance Results Singles

More information

Dual-Energy CT: The Technological Approaches

Dual-Energy CT: The Technological Approaches Dual-Energy CT: The Technological Approaches Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure

More information

Biases Affecting the Measurements of Tumor-to-Background Activity Ratio in PET

Biases Affecting the Measurements of Tumor-to-Background Activity Ratio in PET 2112 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002 Biases Affecting the Measurements of Tumor-to-Background Activity Ratio in PET M. Soret, C. Riddell, S. Hapdey, and I. Buvat Abstract

More information

Quantitative Molecular Imaging Using PET/CT to Assess Response to Therapy

Quantitative Molecular Imaging Using PET/CT to Assess Response to Therapy Quantitative Molecular Imaging Using PET/CT to Assess Response to Therapy Paul Kinahan, PhD Director of PET/CT Physics Imaging Research Laboratory, Department of Radiology University of Washington, Seattle,

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

Calculation methods in Hermes Medical Solutions dosimetry software

Calculation methods in Hermes Medical Solutions dosimetry software Calculation methods in Hermes Medical Solutions dosimetry software Helena McMeekin MSc. Clinical Applications Scientist, Hermes Medical Solutions MRTDosimetry Scientific Workshop The Principals and Clinical

More information

Clinical PET/CT imaging. PET/CT - Optimization of torso imaging. CT-based attenuation correction. PET/CT torso imaging

Clinical PET/CT imaging. PET/CT - Optimization of torso imaging. CT-based attenuation correction. PET/CT torso imaging / - Optimization of torso imaging Clinical / imaging Torso FDG, 1976 Torso FDG-/, 2007 Thomas Beyer thomas.beyer@cmi-experts.com CE 30, SNM Annual Meeting, Toronto, June 2009 Software Hardware Neck Thorax

More information

GE Healthcare. Rad Rx. White Paper

GE Healthcare. Rad Rx. White Paper GE Healthcare Rad Rx White Paper Introduction This publication is part of a series of white papers aimed at communicating the importance of each component in the image chain of a PET/CT study. From data

More information

PET in Radiation Therapy. Outline. Tumor Segmentation in PET and in Multimodality Images for Radiation Therapy. 1. Tumor segmentation in PET

PET in Radiation Therapy. Outline. Tumor Segmentation in PET and in Multimodality Images for Radiation Therapy. 1. Tumor segmentation in PET Tumor Segmentation in PET and in Multimodality Images for Radiation Therapy Wei Lu, Ph.D. Department of Radiation Oncology Mallinckrodt Institute of Radiology Washington University in St. Louis Outline

More information

The Proper Use of PET/CT in Tumoring Imaging

The Proper Use of PET/CT in Tumoring Imaging The Proper Use of PET/CT in Tumoring Imaging Mijin Yun, M.D. Jong Doo Lee, M.D. Department of Radiology / Division of Nuclear Medicine Yonsei University College of Medicine, Severance Hospital E mail :

More information

The combination of functional and anatomic images, Optimized Intravenous Contrast Administration for Diagnostic Whole-Body 18 F-FDG PET/CT

The combination of functional and anatomic images, Optimized Intravenous Contrast Administration for Diagnostic Whole-Body 18 F-FDG PET/CT Optimized Intravenous Contrast Administration for Diagnostic Whole-Body 18 F-FDG PET/CT Thomas Beyer, PhD 1 ; Gerald Antoch, MD 2 ; Andreas Bockisch, MD, PhD 1 ; and Joerg Stattaus, MD 2 1 Department of

More information

Patterns and Prevalence of Misregistration in 18 F-FDG PET/CT

Patterns and Prevalence of Misregistration in 18 F-FDG PET/CT Egyptian J. Nucl. Med., Vol 2, No. 2, Dec. 2009 55 ONCOLOGY, Original Article Patterns and Prevalence of Misregistration in 18 F-FDG PET/CT Farghaly HR,* Muzaffar R,** Bohle RJ,** Nguyen NC,** Osman MM,**

More information

To Shield or Not to Shield? Lincoln L. Berland, M.D.

To Shield or Not to Shield? Lincoln L. Berland, M.D. To Shield or Not to Shield? Lincoln L. Berland, M.D. Disclosures Consultant to: Nuance, Inc. Page 2 Breast Radiation on CT Use of chest CT has increased in women vulnerable to cancer induction by radiation.

More information

Evaluation of Lung Cancer Response: Current Practice and Advances

Evaluation of Lung Cancer Response: Current Practice and Advances Evaluation of Lung Cancer Response: Current Practice and Advances Jeremy J. Erasmus I have no financial relationships, arrangements or affiliations and this presentation will not include discussion of

More information

PET is currently the accepted gold standard in noninvasive

PET is currently the accepted gold standard in noninvasive CT Attenuation Correction for Myocardial Perfusion Quantification Using a PET/CT Hybrid Scanner Pascal Koepfli, MD 1 ; Thomas F. Hany, MD 2 ; Christophe A. Wyss, MD 1 ; Mehdi Namdar, MD 1 ; Cyrill Burger,

More information

PET-CT for radiotherapy planning in lung cancer: current recommendations and future directions

PET-CT for radiotherapy planning in lung cancer: current recommendations and future directions PET-CT for radiotherapy planning in lung cancer: current recommendations and future directions Gerry Hanna Centre for Cancer Research and Cell Biology Queen s University of Belfast @gerryhanna Talk Outline

More information

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information

Regular PET Scanning Without CT. Christian Cohade and Richard L. Wahl

Regular PET Scanning Without CT. Christian Cohade and Richard L. Wahl Applications of Positron Emission Tomography/Computed Tomography Image Fusion in Clinical Positron Emission Tomography Clinical Use, Interpretation Methods, Diagnostic Improvements Christian Cohade and

More information

45 Hr PET Registry Review Course

45 Hr PET Registry Review Course 45 HR PET/CT REGISTRY REVIEW COURSE Course Control Document Timothy K. Marshel, MBA, R.T. (R), (N)(CT)(MR)(NCT)(PET)(CNMT) The PET/CT Training Institute, Inc. SNMMI-TS 028600-028632 45hr CEH s Voice Credits

More information

Understanding Biological Activity to Inform Drug Development

Understanding Biological Activity to Inform Drug Development National Cancer Policy Forum Understanding Biological Activity to Inform Drug Development December 12, 2016 Wolfgang Weber Molecular Imaging and Therapy Service Department of Radiology RECIST Response

More information

An Introduction to Dual Energy Computed Tomography

An Introduction to Dual Energy Computed Tomography An Introduction to Dual Energy Computed Tomography Michael Riedel University of Texas Health Science Center at San Antonio Introduction The idea of computed tomography (CT) was first introduced in the

More information

X-Ray & CT Physics / Clinical CT

X-Ray & CT Physics / Clinical CT Computed Tomography-Basic Principles and Good Practice X-Ray & CT Physics / Clinical CT INSTRUCTORS: Dane Franklin, MBA, RT (R) (CT) Office hours will be Tuesdays from 5pm to 6pm CLASSROOM: TIME: REQUIRED

More information

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities

Dual-Energy Imaging of Bone Marrow Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities Dual-Energy Imaging of Bone Edema on a Dedicated Multi-Source Cone-Beam CT System for the Extremities W Zbijewski, 1 A Sisniega, 1 JW Stayman, 1 N Packard, 2 J Yorkston, 2 G Thawait, 3 S Demehri, 3 J Fritz,

More information

Integrated PET/CT systems State of the art and Clinical Applications

Integrated PET/CT systems State of the art and Clinical Applications Integrated PET/CT systems State of the art and Clinical Applications V. Bettinardi Nuclear Medicine Dep. Scientific Institute San Raffaele Hospital Milan Italy The announcement of a new Diagnostic Imaging

More information

Automated Tumour Delineation Using Joint PET/CT Information

Automated Tumour Delineation Using Joint PET/CT Information Automated Tumour Delineation Using Joint PET/CT Information Vaclav Potesil 1,2, Xiaolei Huang 1,3, Xiang Sean Zhou 1, 1 Siemens Medical Solutions USA, Computer Aided Diagnosis and Therapy Solutions; 2

More information

First Clinical Experience with

First Clinical Experience with First Clinical Experience with Discovery MI Digital PET/CT Martin Huellner Department of Nuclear Medicine University Hospital Zurich / University of Zurich Switzerland Agenda BelNuc GE Symposium 1. Digital

More information

Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field

Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field Measurement of Respiratory and Cardiac Motion Using a Multi Antenna Continuous Wave Radar Operating in the Near Field Florian Pfanner 1,2, Thomas Allmendinger 2, Thomas Flohr 2, and Marc Kachelrieß 1,3

More information

Use of molecular and functional imaging for treatment planning The Good, The Bad and The Ugly

Use of molecular and functional imaging for treatment planning The Good, The Bad and The Ugly Use of molecular and functional imaging for treatment planning The Good, The Bad and The Ugly Robert Jeraj, PhD Associate Professor of Medical Physics, Human Oncology, Radiology and Biomedical Engineering

More information

The SMART scanner: a combined PETET tomograph for clinical oncologys

The SMART scanner: a combined PETET tomograph for clinical oncologys The SMART scanner: a combined PETET tomograph for clinical oncologys David W. Townsend', Senior Member IEEE, Thomas Beyer', Student Member IEEE, Paul E. Kinahan', Member IEEE, Tony Brun2, Raymond Roddy2,

More information

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols Personalization of Pediatric Imaging in Terms of Needed Indication-Based Quality Per Dose Acknowledgments Duke University Medical Center Ehsan Samei, PhD Donald Frush, MD Xiang Li PhD DABR Cleveland Clinic

More information

Radiation Exposure to Staff Using PET/CT Facility

Radiation Exposure to Staff Using PET/CT Facility Egyptian J. Nucl. Med., Vol. 8, No. 2, December 2013 1 Editorial Radiation Exposure to Staff Using PET/CT Facility Taalab, Kh; and Mohsen, Z Department of Nuclear Medicine, International Medical Center;

More information

Dual-Energy CT Applications in Radiation Therapy

Dual-Energy CT Applications in Radiation Therapy THE UNIVERSITY OF WISCONSIN MADISON Dual-Energy CT Applications in Radiation Therapy - Jessica Miller 1 Disclosures Funding provided by Siemens Medical 2 Learning objectives General principles of dual

More information

IN VIVO IMAGING Proton Beam Range Verification With PET/CT

IN VIVO IMAGING Proton Beam Range Verification With PET/CT IN VIVO IMAGING Proton Beam Range Verification With PET/CT Antje-Christin Knopf 1/3 K Parodi 2, H Paganetti 1, T Bortfeld 1 Siemens Medical Solutions Supports This Project 1 Department of Radiation Oncology,

More information

Simulations of Preclinical andclinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy. Using GATE

Simulations of Preclinical andclinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy. Using GATE GATE Simulations of Preclinical andclinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy Using GATE Quick tour & Highlights! GATE Training, INSTN-Saclay, October 2015 Albertine

More information

Nuclear Medicine and PET. D. J. McMahon rev cewood

Nuclear Medicine and PET. D. J. McMahon rev cewood Nuclear Medicine and PET D. J. McMahon 150504 rev cewood 2018-02-15 Key Points Nuclear Medicine and PET: Imaging: Understand how Nuc Med & PET differ from Radiography & CT by the source of radiation. Be

More information

Imaging Decisions Start Here SM

Imaging Decisions Start Here SM Owing to its high resolution and wide anatomic coverage, dynamic first-pass perfusion 320-detector-row CT outperforms PET/CT for distinguishing benign from malignant lung nodules, researchers from Japan

More information

Molecular Imaging and Cancer

Molecular Imaging and Cancer Molecular Imaging and Cancer Cancer causes one in every four deaths in the United States, second only to heart disease. According to the U.S. Department of Health and Human Services, more than 512,000

More information

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens DEfinitely Siemens True Dual Energy Dr. Stefan Ulzheimer, Siemens Healthcare GmbH International version. Not for distribution in the US. Unrestricted Siemens AG 2015 All rights reserved. The products/features

More information

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT

Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT Introduction Pediatric malignancies Changing trends & Radiation burden Radiation exposure from PET/CT Image gently PET & CT modification - PET/CT protocols Tips Leukaemia / lymphoma: ~ 35% acute lymphoblastic

More information

The Role of PET / CT in Lung Cancer Staging

The Role of PET / CT in Lung Cancer Staging July 2004 The Role of PET / CT in Lung Cancer Staging Vlad Vinarsky, Harvard Medical School Year IV Patient AM HPI: 81 yo F p/w hemoptysis x 1 month LLL lesion on CXR, not responsive to Abx 35 pack-year

More information

Los Angeles Radiological Society 62 nd Annual Midwinter Radiology Conference January 31, 2010

Los Angeles Radiological Society 62 nd Annual Midwinter Radiology Conference January 31, 2010 Los Angeles Radiological Society 62 nd Annual Midwinter Radiology Conference January 31, 2010 Self Assessment Module on Nuclear Medicine and PET/CT Case Review FDG PET/CT IN LYMPHOMA AND MELANOMA Submitted

More information

J of Nuclear Medicine Technology, first published online January 24, 2014 as doi: /jnmt

J of Nuclear Medicine Technology, first published online January 24, 2014 as doi: /jnmt J of Nuclear Medicine Technology, first published online January 24, 2014 as doi:10.2967/jnmt.113.133736 Effects of Pacemaker, Implantable Cardioverter Defibrillator, and Left Ventricular Leads on CT-Based

More information

CLINICAL IMPLEMENTATION OF HYBRID IMAGING CLINICAL IMPLEMENTATION OF HYBRID IMAGING COMBINED IMAGING CLINICAL IMPLEMENTATION OF

CLINICAL IMPLEMENTATION OF HYBRID IMAGING CLINICAL IMPLEMENTATION OF HYBRID IMAGING COMBINED IMAGING CLINICAL IMPLEMENTATION OF HYBRID OBJECTIVES OF ADDING CT-SCANNING TO NM? CLINICAL IMPLEMENTATION OF François Jamar UCL, Brussels ABR-BVS Workshop, Antwerp, Feb. 18, 2011 HYBRID COMBINED COMBINED ( PET AND CT) In-WBC 111 In COMBINED

More information

OPTIMIZATION OF Rb-82 PET ACQUISITION AND RECONSTRUCTION PROTOCOLS FOR MYOCARDIAL PERFUSION DEFECT DETECTION

OPTIMIZATION OF Rb-82 PET ACQUISITION AND RECONSTRUCTION PROTOCOLS FOR MYOCARDIAL PERFUSION DEFECT DETECTION 2008 IEEE Nuclear Science Symposium Conference Record M10-44 OPTIMIZATION OF Rb-82 PET ACQUISITION AND RECONSTRUCTION PROTOCOLS FOR MYOCARDIAL PERFUSION DEFECT DETECTION ling Tang, Member, IEEE, Arman

More information

AAPM Initiatives in Quantitative Imaging

AAPM Initiatives in Quantitative Imaging AAPM Initiatives in Quantitative Imaging John M. Boone, Ph.D., FAAPM, FSBI, FACR Chair, AAPM Science Council Chair, Ad Hoc Committee on Quantitative Imaging Chair, TG on QI in CT Professor and Vice Chair

More information

Does the Use of IV Contrast Enhanced CT for Attenuation Correction Affect Clinical Interpretation of Head and Neck PET/CT?

Does the Use of IV Contrast Enhanced CT for Attenuation Correction Affect Clinical Interpretation of Head and Neck PET/CT? 12 The Open Nuclear Medicine Journal, 2011, 3, 12-18 Open Access Does the Use of IV Contrast Enhanced CT for Attenuation Correction Affect Clinical Interpretation of Head and Neck PET/CT? Lance T. Hall,

More information

4D PET: promises and limitations

4D PET: promises and limitations 4D PET: promises and limitations Tinsu Pan, Ph.D. M.D. Anderson Cancer Center The University of Texas Background Outlines Gating techniques: Deep inspiration breath hold 4D PET/CT Non-gating techniques

More information

Uptake of 18 F-FDG in malignant tumors is subject to

Uptake of 18 F-FDG in malignant tumors is subject to Time Course of Tumor SUV in 18 F-FDG PET of Breast Cancer: Presentation of a Simple Model Using a Single Reference Point for Time Corrections of Tumor SUVs Alexander R. Stahl 1, Till A. Heusner 2, Verena

More information

8/10/2016. PET/CT Radiomics for Tumor. Anatomic Tumor Response Assessment in CT or MRI. Metabolic Tumor Response Assessment in FDG-PET

8/10/2016. PET/CT Radiomics for Tumor. Anatomic Tumor Response Assessment in CT or MRI. Metabolic Tumor Response Assessment in FDG-PET PET/CT Radiomics for Tumor Response Evaluation August 1, 2016 Wei Lu, PhD Department of Medical Physics www.mskcc.org Department of Radiation Oncology www.umaryland.edu Anatomic Tumor Response Assessment

More information

Breast CT and Dosimetry

Breast CT and Dosimetry 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of

More information

Metal Artefact Reduction in CT

Metal Artefact Reduction in CT Metal Artefact Reduction in CT DANIEL MARRINER Metal Artefact Reduction in CT Metal Artefact Clinical Indications for MAR SEMAR and How It Works Technical Considerations Case Studies utilising SEMAR Metal

More information

Repeatability and reproducibility of 18 F-NaF PET quantitative imaging biomarkers

Repeatability and reproducibility of 18 F-NaF PET quantitative imaging biomarkers Repeatability and reproducibility of 18 F-NaF PET quantitative imaging biomarkers Christie Lin, Tyler Bradshaw, Timothy Perk, Stephanie Harmon, Glenn Liu, Robert Jeraj University of Wisconsin Madison,

More information

Quantitative Theranostics in Nuclear Medicine

Quantitative Theranostics in Nuclear Medicine Quantitative Theranostics in Nuclear Medicine M. Lassmann Klinik und Poliklinik für Nuklearmedizin Direktor: Prof. Dr. A. Buck Contents What is Theranostics? Potential Targets Basic Principles of Quantitative

More information

Segmenting CT images of bronchogenic carcinoma with bone metastases using PET intensity markers approach

Segmenting CT images of bronchogenic carcinoma with bone metastases using PET intensity markers approach doi:10.2478/v10019-009-0023-y research article Segmenting CT images of bronchogenic carcinoma with bone metastases using PET intensity markers approach Iman Avazpour 1, Ros Ernida Roslan 1, Peyman Bayat

More information

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Radiation Dose Reduction: Should You Use a Bismuth Breast Shield? Lincoln L. Berland, M.D., F.A.C.R. Michael V. Yester, Ph.D. University of Alabama at Birmingham Breast Radiation on CT Use of chest CT

More information

High Tech Imaging Quick Reference Guide

High Tech Imaging Quick Reference Guide High Tech Imaging Quick Reference Guide 1 High Tech Imaging Authorizations may now be requested through our secure provider portal, BlueAccess. Getting Started Step 1: Log into BlueAccess from www.bcbst.com

More information

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes BioMedical quantitative X-Ray Imaging Emmanuel Brun Researcher Inserm Université Grenoble Alpes 1 Outline Introduction K-Edge Imaging Patient imaging at the European synchrotron Medical Phase Contrast

More information

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy.

8/3/2016. Consultant for / research support from: Astellas Bayer Bracco GE Healthcare Guerbet Medrad Siemens Healthcare. Single Energy. U. Joseph Schoepf, MD Prof. (h.c.), FAHA, FSCBT-MR, FNASCI, FSCCT Professor of Radiology, Medicine, and Pediatrics Director, Division of Cardiovascular Imaging Consultant for / research support from: Astellas

More information

Clinical evaluation of respiration-induced attenuation uncertainties in pulmonary 3D PET/CT

Clinical evaluation of respiration-induced attenuation uncertainties in pulmonary 3D PET/CT Kruis et al. EJNMMI Physics (2015) 2:4 DOI 10.1186/s40658-014-0107-7 ORIGINAL RESEARCH Open Access Clinical evaluation of respiration-induced attenuation uncertainties in pulmonary 3D PET/CT Matthijs F

More information

ROLE OF PET-CT IN BREAST CANCER, GUIDELINES AND BEYOND. Prof Jamshed B. Bomanji Institute of Nuclear Medicine UCL Hospitals London

ROLE OF PET-CT IN BREAST CANCER, GUIDELINES AND BEYOND. Prof Jamshed B. Bomanji Institute of Nuclear Medicine UCL Hospitals London ROLE OF PET-CT IN BREAST CANCER, GUIDELINES AND BEYOND Prof Jamshed B. Bomanji Institute of Nuclear Medicine UCL Hospitals London CANCER Key facts Estimated 15.2 million new cases per year in 2015 worldwide

More information

Bone PET/MRI : Diagnostic yield in bone metastases and malignant primitive bone tumors

Bone PET/MRI : Diagnostic yield in bone metastases and malignant primitive bone tumors Bone PET/MRI : Diagnostic yield in bone metastases and malignant primitive bone tumors Lars Stegger, Benjamin Noto Department of Nuclear Medicine University Hospital Münster, Germany Content From PET to

More information

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies

Computer based delineation and follow-up multisite abdominal tumors in longitudinal CT studies Research plan submitted for approval as a PhD thesis Submitted by: Refael Vivanti Supervisor: Professor Leo Joskowicz School of Engineering and Computer Science, The Hebrew University of Jerusalem Computer

More information

Quantitative outcome of registration methods for correcting cardiac drift in cardiac PET/CT imaging

Quantitative outcome of registration methods for correcting cardiac drift in cardiac PET/CT imaging JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Quantitative outcome of registration methods for correcting cardiac drift in cardiac PET/CT imaging Jonathon A. Nye, a Dana Tudorascu,

More information

Cigna - Prior Authorization Procedure List: Radiology & Cardiology

Cigna - Prior Authorization Procedure List: Radiology & Cardiology Cigna - Prior Authorization Procedure List: Radiology & Cardiology Product Category CPT Code CPT Code Description Radiology MR 70336 MRI Temporomandibular Joint(s), (TMJ) Radiology CT 70450 CT Head or

More information

An Introduction to PET Imaging in Oncology

An Introduction to PET Imaging in Oncology January 2002 An Introduction to PET Imaging in Oncology Janet McLaren, Harvard Medical School Year III Basics of PET Principle of Physiologic Imaging: Allows in vivo visualization of structures by their

More information

State-of-the-Art SPECT/CT: Cardiac Imaging

State-of-the-Art SPECT/CT: Cardiac Imaging State-of-the-Art SPECT/CT: Cardiac Imaging Ernest V Garcia*, PhD Endowed Professor in Cardiac Imaging Director, Nuclear Cardiology R&D Laboratory Disclosure: Dr. Garcia receives royalties from the sale

More information

Dual-Energy 101: Principles, Methods and Dose

Dual-Energy 101: Principles, Methods and Dose Dual-Energy 101: Principles, Methods and Dose Juan Carlos Ramirez-Giraldo, Ph.D Staff Scien2st, Collabora2ons Manager SE Region ISCT San Francisco, 2017 Siemens Medical Solu2ons USA, Inc., 2017 Page 1

More information

PET-MRI in malignant bone tumours. Lars Stegger Department of Nuclear Medicine University Hospital Münster, Germany

PET-MRI in malignant bone tumours. Lars Stegger Department of Nuclear Medicine University Hospital Münster, Germany PET-MRI in malignant bone tumours Lars Stegger Department of Nuclear Medicine University Hospital Münster, Germany Content From PET to PET/MRI General considerations Bone metastases Primary bone tumours

More information

Quantitation of Cerebral Glucose Utilization using the Arterial Input Function or the Standardized Uptake Value (SUV)

Quantitation of Cerebral Glucose Utilization using the Arterial Input Function or the Standardized Uptake Value (SUV) Quantitation of Cerebral Glucose Utilization using the Arterial Input Function or the Standardized Uptake Value (SUV) Brian R. Moyer BRMoyer & Associates, LLC, Amherst, NH 03031 http://www.brmassocllc-org.com/

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement Range of charged particles (e.g.,!: µm; ": mm) Range of high energy photons (cm) Two main types of interactions of high energy photons Compton scatter Photoelectric

More information

Optimized PET/CT protocols: how much CT is needed? Increasing use of PET-CT. Imaging in Lymphoma

Optimized PET/CT protocols: how much CT is needed? Increasing use of PET-CT. Imaging in Lymphoma BVS/ABR Workshop 2011 on dose related to multimodality imaging Optimized PET/CT protocols: how much CT is needed? 2008 Mean medical radiation exposure/head Belgium= 2.42 msv The Netherlands= 0.8 msv Belgium

More information

Breast Cancer PET/CT Imaging Protocol

Breast Cancer PET/CT Imaging Protocol Breast Cancer PET/CT Imaging Protocol Scanning Protocol: Patients are scanned from the top of the neck through the pelvis. Arms-up position is used to avoid beam-hardening artifact in the chest and abdomen.

More information

Hybrid systems in Medical Imaging

Hybrid systems in Medical Imaging Hybrid systems in Medical Imaging from PET/CT to PET/MR Osman Ratib, MD, PhD, FAHA Professor and chair Department of Medical Imaging and Information Sciences Head of division of Nuclear Medicine University

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Role of DE-CT in Oncology

Role of DE-CT in Oncology Role of DE-CT in Oncology Dushyant Sahani, M.D Director of CT Associate Professor of Radiology Massachusetts General Hospital Harvard Medical School Email-dsahani@partners.org Disclosure Research Grant

More information

Precision of pre-sirt predictive dosimetry

Precision of pre-sirt predictive dosimetry International Course on THERANOSTICS AND MOLECULAR RADIOTHERAPY Precision of pre-sirt predictive dosimetry Hugo Levillain Department of Nuclear Medicine Medical Physics Jules Bordet Institute, Université

More information

BlueAdvantage SM. & BlueChoice SM Radiology Prior Authorization Program Code List CPT /HCPS

BlueAdvantage SM. & BlueChoice SM Radiology Prior Authorization Program Code List CPT /HCPS BlueAdvantage SM & BlueChoice SM Radiology Prior Authorization Program Code List CPT /HCPS 70336 MRI TMJ 70450 CT Head Without Contrast 70460 CT Head With Contrast 70470 CT Head Without & With Contrast

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,

More information

On the feasibility of speckle reduction in echocardiography using strain compounding

On the feasibility of speckle reduction in echocardiography using strain compounding Title On the feasibility of speckle reduction in echocardiography using strain compounding Author(s) Guo, Y; Lee, W Citation The 2014 IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL.,

More information

The development of combined PET/CT scanners has

The development of combined PET/CT scanners has Decreased Perfusion in the Lateral Wall of the Left Ventricle in PET/CT Studies with 13 N-Ammonia: Evaluation in Healthy Adults William C. Klingensmith III 1, Carolyn Noonan 2, Jack H. Goldberg 2, Dedra

More information

POSITRON EMISSION TOMOGRAPHY PHANTOM STUDIES FOR RADIATION THERAPY TARGET DELINEATION MICHAEL VINTSON LAWRENCE

POSITRON EMISSION TOMOGRAPHY PHANTOM STUDIES FOR RADIATION THERAPY TARGET DELINEATION MICHAEL VINTSON LAWRENCE POSITRON EMISSION TOMOGRAPHY PHANTOM STUDIES FOR RADIATION THERAPY TARGET DELINEATION BY MICHAEL VINTSON LAWRENCE A Dissertation Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL

More information

Animals. Male C57Bl/6 mice (n=27) were obtained from Charles River (Sulzfeld, Germany) and

Animals. Male C57Bl/6 mice (n=27) were obtained from Charles River (Sulzfeld, Germany) and Supplemental Methods Animals. Male C57Bl/6 mice (n=27) were obtained from Charles River (Sulzfeld, Germany) and housed in groups of 5-10 in individually ventilated BCU cages within a temperature controlled

More information

C ancer cells require a great deal of sugar (glucose) for. PET/CT imaging: detection of choroidal melanoma SCIENTIFIC REPORT

C ancer cells require a great deal of sugar (glucose) for. PET/CT imaging: detection of choroidal melanoma SCIENTIFIC REPORT 265 SCIENTIFIC REPORT PET/CT imaging: detection of choroidal melanoma S Reddy, M Kurli, L B Tena, P T Finger... Aim: To determine the size of untreated choroidal melanomas resolved by whole body positron

More information

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell CT Optimisation for Paediatric SPECT/CT Examinations Sarah Bell Sarah.bell14@nhs.net Outline 1. Introduction 2. Aims and Objectives 3. Methods 4. Results 5. Discussion 6. Conclusions 7. References Introduction

More information

Outline. Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) SPECT/CT and PET/CT Dosimetry

Outline. Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) SPECT/CT and PET/CT Dosimetry SPECT/CT and PET/CT Dosimetry Lifetime Attributable Risk 10 mgy in 100,000 exposed persons (BEIR VII 2006) All Solid Tumors Leukemia Frederic H. Fahey DSc Male Female Male Female Children s Hospital Boston

More information

Positron Emission Tomography Computed Tomography (PET/CT)

Positron Emission Tomography Computed Tomography (PET/CT) Positron Emission Tomography Computed Tomography (PET/CT) What is Positron Emission Tomography Computed Tomography (PET/CT) Scanning? What are some common uses of the procedure? How should I prepare for

More information

Accurate attenuation correction (AC) is a major clinical

Accurate attenuation correction (AC) is a major clinical Single-Phase CT Aligned to Gated PET for Respiratory Motion Correction in Cardiac PET/CT R. Glenn Wells 1, Terrence D. Ruddy 1,2, Rob A. DeKemp 1,2, Jean N. DaSilva 1, and Rob S. Beanlands 1,2 1 Division

More information

Effects on the FDG distribution by a high uptake of brown adipose tissue at PET examination

Effects on the FDG distribution by a high uptake of brown adipose tissue at PET examination Lindholm et al. EJNMMI Research 2014, 4:72 ORIGINAL RESEARCH Open Access Effects on the FDG distribution by a high uptake of brown adipose tissue at PET examination Henry Lindholm 1*, Fredrik Brolin 2,

More information

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging Ultrasound Computed tomography Case studies Utility of IQon Spectral CT in cardiac imaging Cardiac imaging is a challenging procedure where it is necessary to image a motion-free heart. This requires a

More information

8/10/2016. PET/CT for Tumor Response. Staging and restaging Early treatment response evaluation Guiding biopsy

8/10/2016. PET/CT for Tumor Response. Staging and restaging Early treatment response evaluation Guiding biopsy PET/CT for Tumor Response Evaluation August 4, 2016 Wei Lu, PhD Department of Medical Physics www.mskcc.org Department of Radiation Oncology www.umaryland.edu FDG PET/CT for Cancer Imaging Staging and

More information

Dual Energy CT of the Liver

Dual Energy CT of the Liver 34th Annual Course October 2011 Washington, DC Dual Energy CT of the Liver Vassilios Raptopoulos, MD Beth Israel Deaconess Medical Center Harvard Medical School Dual Energy CT (DECT) Different materials

More information

CARDIAC PET PERFUSION IMAGING with RUBIDIUM-82

CARDIAC PET PERFUSION IMAGING with RUBIDIUM-82 CARDIAC PET PERFUSION IMAGING with RUBIDIUM-82 Pr Denis AGOSTINI Président du Groupe de Cardiologie Nucléaire et IRM CHU Caen Bordeaux 2006 Cardiac Perfusion-Metabolism Mismatch with PET Cumulative Survival

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information