Interleaved Water and Fat Imaging and Applications to Lipid Quantitation Using the Gradient Reversal Technique

Size: px
Start display at page:

Download "Interleaved Water and Fat Imaging and Applications to Lipid Quantitation Using the Gradient Reversal Technique"

Transcription

1 JOURNAL OF MAGNETIC RESONANCE IMAGING 26: (2007) Original Research Interleaved Water and Fat Imaging and Applications to Lipid Quantitation Using the Gradient Reversal Technique Haiying Tang, PhD, 1 * Ed X. Wu, PhD, 2 Richard Kennan, PhD, 1 Haiying Liu, PhD, 1 and Donald S. Williams, PhD 1 Purpose: To implement and evaluate the gradient reversalbased chemical shift imaging technique to obtain qualitative and quantitative spatially-registered fat and water images with high imaging efficiency at very high field. Materials and Methods: A multiecho gradient reversalbased sequence allowing interleaved water fat imaging during a single acquisition and quantitation of fat/water content is presented. The sequence was optimized and implemented at 11.7T. The quantitation was verified with water fat phantoms and applied to lipid measurement in an in vivo mouse model. Results: Results from phantoms, in vivo lipid measurement in mouse liver and hind limb muscle, and ex vivo rat knee imaging experiments demonstrated the robustness and high selectivity of this technique for interleaved and quantitative water and fat imaging at very high field. Conclusion: The proposed MRI technique permits interleaved water and fat imaging, with which spectrally wellseparated water and fat images at the identical slice locations could be obtained in a single acquisition without increasing scan time. The technique could be used for in vivo quantitative mapping of lipid content and applied to investigations using small animal experiment models. Key Words: MRI; water and fat imaging; chemical-shift; gradient reversal; liver; muscle J. Magn. Reson. Imaging 2007;26: Wiley-Liss, Inc. 1 Department of Imaging, Merck Research Laboratories, Rahway, New Jersey, USA. 2 Department of Electrical and Electronic Engineering, The University of Hong Kong, Hong Kong SAR, China. *Address reprint requests to: H.T., PhD, Department of Imaging, Merck Research Laboratories, RY80B-100, Rahway, NJ haiying_tang@merck.com Received October 13, 2006; Accepted June 1, DOI /jmri Publishedonline00Month2007inWileyInterScience( wiley.com). MR IMAGING of water and fat is important in a wide range of biomedical studies including obesity and diabetes research, cartilage delineation in joint imaging, body composition measurement, and evaluation of effect of drugs on lipids. A rapidly evolving and extremely useful area is evaluation of lipid content in genetically altered mice. At very high field, the commonly used spectral selection or saturation techniques are often hampered by strong susceptibility effects at tissue air interfaces, as well as other sources of field inhomogeneity (1,2). The spatial-spectral selective excitation techniques (3,4) are limited in practice by their requirement for high gradient amplitude and switching rate during the short radio frequency (RF) pulse train. The alternative phase sensitive techniques such as the Dixonmethod (5 7) may not be desirable for applications at very high field because the chemical shift during sliceselective excitation causes a misregistration of water and fat slices. Furthermore, phase sensitive methods require multiple data acquisitions for separation of water and fat signal. The recent three-point Dixon (6)- based iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) technique (8,9) maximized the noise performance of the water fat decomposition and thus improved the scan efficiency by combining the acquisition of asymmetric echoes with respect to the spin echo (SE) and an iterative least-squares separation algorithm. However the implementation of IDEAL at very high field and the related data acquisition issues require further investigation. In 1987, Park et al (10) proposed a gradient reversal technique for single-slice two-dimensional (2D) or single-slab 3D chemical shift selective imaging. It was applied to obtain water and fat images by utilizing the fat water chemical shift separation in the slice direction. This technique is applicable to spin echo 2D or 3D sequence that involves both excitation and refocusing slice-selective RF pulses. Because a single spectral resonance is selected in an imaging scan, two acquisitions are required to obtain both water and fat images. In the same year, Volk et al (11) proposed the chemical shift specific slice-selection technique and demonstrated the use of the gradient reversal technique for interleaved water and fat imaging during a single acquisition by interleaving water and fat excitations at the same locations. Because such an approach relies on the strong chemical shift effect along the slice direction for com Wiley-Liss, Inc. 1064

2 An Interweaved Water and Fat Imaging Technique 1065 plete separation of water and fat, the method is best suited for high field imaging in which larger chemical shifts enable the use of slice-selective RF pulses of reasonable durations. Since 1987, no further studies using the gradient reversal technique for fat/water imaging have been reported. On the other hand, besides proton MR spectroscopy based methods that provide limited spatial information (12), MR imaging methods that allowed accurate fat water quantitation have been scarce. Imaging lipid content remains an area of high priority for the pharmaceutical industry in diabetes and obesity research as well as in safety assessment of drugs (7). The technical issues of existing techniques include susceptibility gradients, effective spectral separation of fat and water, imaging the same slice location for fat and water, and imaging efficiency in terms of number of acquisitions and scan time required. Given the increasing availability of high-field scanners, and the need for small animal imaging, the gradient reversal based chemical shift imaging techniques (10,11) were reevaluated in this study with a view toward obtaining quantitative, spatially-registered fat and water images with high imaging efficiency. An interleaved water- and fatselective imaging sequence based on the gradient reversal technique designed for qualitative and quantitative water and fat imaging purposes was implemented at 11.7T to assess the robustness of this approach for in vivo imaging at very high field. The proposed imaging sequence is based on spin echo and is capable of multiecho and multislice acquisitions, allowing well-separated water and fat images at the identical slice locations to be acquired in a single acquisition without increasing acquisition time. The multiecho train is used to generate true proton density (PD) images for better quantitation of fat and water. Image quantitation was verified with a water/fat phantom, and its application to in vivo high-field small animal imaging was demonstrated in liver and in hind limb muscle fat quantitation, in obese vs. lean mice. Finally, ex vivo high-field musculoskeletal imaging was demonstrated with a formalin-fixed rat knee sample. MATERIALS AND METHODS Gradient Reversal Technique for Water and Fat Imaging Within a Single Acquisition In conventional proton MR imaging, the chemical-shift effect of water and fat occurs not only in the frequencyencoding direction, but also in the slice-selection direction. Provided the slice selection profile is approximately rectangular, water and fat slices could be completely separated in the slice direction when rf off, where rf is the bandwidth of slice-selective RF excitation or refocusing pulse, and off is the frequency offset between water and fat resonances. Under such circumstances, the gradient reversal method (10,11), a spin echo sequence with reversal of slice gradient polarity during the refocusing pulse, allows pure water and fat images from the same slice location to be obtained as long as a slice at the center of the gradients is selected. If the RF pulses are applied on Figure 1. A schematic of exciting both water and fat slices for a two-slice imaging protocol using the gradient reversal technique. a: RF is on-resonance with water; (b) RF is on-resonance with fat. Diagonal lines show the variation of resonances for water (thick line, W) and fat (thin line, F) for positive (solid line) and negative (dashed line) polarity gradients. W1 and W2 in (a), and F1 and F2 in (b) represent excited water and fat slices experiencing both 90 and 180 pulses, while F1 and F2 in (a) and W1 and W2 in (b) experience only 90 or the 180 pulse. water resonance, only the water slice will be excited and refocused to produce a spin echo because it experiences both 90 and 180 pulses. Repeating this with RF frequency on fat resonance will allow fat from the same slice location to be excited and refocused. Since fat (or water) spins in the slice are not excited when RF offset is set on water (or fat), the scheme can easily be implemented in a multislice acquisition mode to obtain water and fat images in the same scan without increasing acquisition time. Extending the scheme to multislice acquisition requires the reversal of frequency offset for the refocusing pulse and also places restrictions on slice gap. To help illustrate this, a schematic of the location of excited slices for a two-slice imaging protocol using the gradient reversal sequence is shown in Fig. 1 when RF applied is on resonance with water (Fig. 1a), and with fat (Fig. 1b) of the same slice location. Diagonal lines show the variation of resonances for water (thick line, W) and fat (thin line, F) for positive (solid line) and negative (dashed line) polarity gradients. Note that an echo will be formed from a slice location only if both 90 and 180 pulses excite the same slice location. W1 and W2 in (Fig. 1a), and F1 and F2 in (Fig. 1b) are excited water and fat slices with a solid-line frame, representing slices that experience both 90 and 180 pulses; while F1 and F2 in (Fig. 1a) and W1 and W2 in (Fig. 1b) represent slices that experience only the 90 (solidline frame) or the 180 (dotted-line frame) pulses. With the scheme, water and fat slice excitations can be interleaved within a TR period to image both water (Fig. 1a) and fat (Fig. 1b) species at the identical slice locations during a single acquisition without increasing overall scan time. However, as seen on Fig. 1, the multislice acquisition requires the slice interval (center-to-

3 1066 Tang et al. center) frequency to be greater than off rf, to avoid interference between adjacent slices. Finally, if during signal detection, the frequency is set to the Larmor frequency of the species excited, the fat images will no longer be shifted along the readout direction, and water and fat images will be spatially registered with one another. This technique is generally applicable to the 2D and 3D sequences that involve at least two slice-selective RF pulses, such as spin echo related sequences. It is especially suited for high-field applications. At low field, the requirement of rf off leads to unacceptably long RF pulse duration and restricts the practical applications. Quantitative Water and Fat Imaging Sequence Development For all phantom and in vivo experiments, optimized Gaussian-shaped RF pulses were used for both excitation and refocusing because they are short and produce slice profiles that contain no lasting ripples. The durations of the 90 and 180 RF pulses were chosen such that the slice excitation bandwidth of the pulses were matched, and the spin echo slice profile bandwidth was approximately 1750 Hz full-width at 20% of maximum, corresponding to the water fat chemical shift (3.5 ppm) at 11.7T. This is necessary to reduce the fat contamination during water selection and vice versa. Effective bandwidth for the given RF pulse shape and durations was experimentally verified by examining the actual slice profile obtained by Fourier transform of the echo formed by slice gradient reversal. For quantitative water/fat imaging, a temporally interleaved water- and fat-selective sequence capable of multiecho and multislice acquisition was implemented using the gradient reversal technique illustrated in Fig. 1. In a single acquisition, the sequence provided a series of multiecho water- and fat-selective images. Multiecho data were then fitted to monoexponential decay to generate fat and water PD images for accessing lipid quantitation and distribution. Fat/water content was calculated according to Ref. 13. Imaging Experiments All imaging experiments were performed on a Bruker Biospin 500WB spectrometer (Bruker Biospin Corporation, Billerica, MA, USA) with an 89-mm 11.7T vertical bore magnet (Oxford Instruments Ltd., UK) and a shielded gradient system capable of up to 100 G/cm gradients. In the first phantom study to demonstrate the effectiveness of fat and water separation, a tube of 29 mm inner diameter (ID) was filled with olive oil, and contained two smaller tubes of water doped with 1:100 dilution gadolinium diethylenetriamine pentaacetic acid (Gd-DTPA) (Magnevist). The phantom was imaged using a 30-mm ID RF coil. Interleaved fat and water images were obtained during the same scan using the proposed chemical shift selective and gradient reversal based sequence, with slice thickness 2 mm, and TR/TE 1000 msec/4.5 msec. To demonstrate the quantitative aspects of the interleaved water/fat imaging, another phantom study was conducted, in which water and fat concentrations were varied at a subvoxel scale. The phantom was based on a water/lipid emulsion (mayonnaise) that contained approximately 75% fat (14), and was diluted with water from 20% to 100% mayonnaise to vary lipid concentration. Separated water- and fat-selective images of single slices were acquired with TR/TE 10 seconds/4.5 msec, number of echoes 8, and slice thickness 1.2 mm. To demonstrate quantitative water and fat imaging in vivo, the interleaved water- and fat-selective multiecho and multislice imaging sequence was applied to a mouse. For in vivo lipid quantitation, water and fat proton densities were obtained by fitting the multiecho images pixel by pixel to a monoexponential, from which relative lipid percentage maps were generated. Lipid measurements in mouse liver and hind limb muscle were performed in wild-type lean mice and obese mice that were on a high fat diet for five months. All in vivo experiments were approved by the Institutional Animal Care and Use Committee. Mice were anesthetized with 1.5% isoflurane/o 2 gas mixture delivered through a nose cone during imaging. For in vivo liver imaging, a 30-mm ID RF coil was used. A total of six water and fat images at three slice locations were acquired with respiratory gating (per slice triggering) in a single acquisition, resulting in a TR 6 seconds. Strengths of the spoilers at both sides of the reversal gradient were reduced, and flow saturation in the descending aorta was applied to minimize the flow artifacts. Other parameters were TE 4.5 msec, number of echoes 8, field of view (FOV) 30 mm, acquisition matrix , slice thickness 1.5 mm, slices spacing 3.2 mm, and number of averages 2. The total acquisition time was approximately 40 minutes. Unlike the mouse liver, in which a wide range of lipid levels could be detected, the in vivo MRI lipid detection and quantification in skeletal muscle is challenging because of the low lipid content and heterogeneous spatial distribution. To increase sensitivity, muscle lipid studies were carried out with a 10-mm ID RF coil. Highresolution water and fat images at a single slice location were acquired from the left hind limbs, with TR/TE 2000 msec/4.5 msec, number of echoes 8, FOV 12.8 mm, acquisition matrix size , slice thickness 1 mm, and number of averages 32. The acquisition time was 137 minutes. As a further demonstration of this technique to highfield musculoskeletal imaging, water and fat imaging of an ex vivo formalin-fixed rat knee was also performed. Water and fat images were acquired at three slice locations, with TR/TE 2000 msec/5.3 msec, FOV 16.5 mm, acquisition matrix size , slice thickness 0.6 mm, slices spacing 1.3 mm, and number of averages 8. The acquisition time was 68 minutes. RESULTS Figure 2 shows fat (Fig. 2a) and water (Fig. 2b) images of the first phantom acquired at the same slice location in a single acquisition using the gradient reversal tech-

4 An Interweaved Water and Fat Imaging Technique 1067 Figure 2. Water and fat images of a phantom acquired at the same slice location in a single acquisition. a: Fat image (olive oil); (b) water image; (c) sum of fat and water images, illustrating that chemical shift artifact along the read direction has been eliminated by setting resonance frequency on the species excited during acquisition; (d) conventional spin echo image, containing both in-plane and through-plane chemical shift artifacts; (e) conventional spin echo image with presaturation of fat. nique. The chemical shift effect along the frequency encoding direction was eliminated by setting the resonance frequency on either water or fat during acquisition. Figure 2c is the chemical shift artifact free image obtained by the summation of Fig. 2a and b. Figure 2d is the conventional spin echo image, containing both in-plane and through-plane chemical shift artifacts. In comparison to the conventional spin echo image, a slight signal inhomogeneity is noted at the edge of the coil in Fig. 2c, indicating that the gradient reversal technique is sensitive to B 0 field inhomogeneity, thus requiring good shimming over the imaging region. The image in Fig. 2e was acquired with the same shimming, using the conventional spin echo sequence with presaturation of fat. It is seen that poor fat suppression is achieved because the spectral saturation technique is sensitive to both B 0 and B 1 field inhomogeneities. Moreover, spectral saturation or selection techniques do not lend themselves to interleaved water and fat imaging in a single acquisition. Finally, it should be noted that a small signal in the oil region in Fig. 2b and 2e was caused by a small fraction of olefinic fatty acids having resonance frequency very close to that of water, thus it could not be suppressed (15) with any frequency-sensitive imaging technique including the gradient reversal based water and fat imaging technique. Figure 3 shows the fat (Fig. 3a) and water (Fig. 3b) images (PD weighted, echo train spacing 4.5 msec) of the water/lipid emulsion phantom acquired using the proposed interleaved and quantitative water and fat imaging sequence. As expected, the signal intensity varied with the lipid concentration. It should be noted that the center tube of the phantom contains only water, and there is no water contamination found in the center tube location in the fat image. Figure 3c is the chemical shift artifact free image obtained by summation of the fat and water images. In comparison, Figure 3d shows the phantom image acquired using the conventional spin echo sequence, in which the in-plane and through-plane chemical shift artifacts (blurring) were clearly seen at the edges. Figure 3e is the lipid percentage map generated from the PD images of water and fat calculated by extrapolating the multiecho images. Figure 3f shows that the MRI lipid content measurement correlated linearly with the water/lipid emulsion concentrations. Figure 4 demonstrates the in vivo water (Fig. 4a) and fat (Fig. 4b) transaxial images (PD weighted, echo train spacing 4.5 msec) in the liver region of an obese mouse acquired in an interleaved acquisition. Figure 4c shows the chemical shift artifact free water and fat image combined. In Fig. 4d, a lipid percentage map was generated from the PD images of water and fat. With the multiecho imaging capability, the transverse relaxation rate (R2) maps of the water and fat could also be obtained separately and simultaneously in a single acquisition. Figure 4e and f are the calculated R2 maps of water and fat at the identical slice location, and the R2 appears much larger for the water component ( l/second) than the lipid component ( l/second) in liver. The increase relaxation rate of liver water is likely due to the ferritin that is mainly existed in the water component. To better visualize the spatial distribution of lipid concentration in mouse liver, the lipid percentage was color-coded in liver and the surrounding tissue. Figure 4g shows a color-coded liver lipid percentage map of the obese mouse. For comparison, a color-coded liver lipid percentage map of the lean mouse is shown in Fig. 4h. About 30% lipid was de- Figure 3. Quantitative water and fat imaging using a diluted water/lipid emulsion (mayonnaise) phantom. a: Fat image; (b) water image; (c) sum of fat and water images; (d) conventional spin echo image; (e) lipid percentage map; and (f) relationship between MRI measured lipid % and mayonnaise content in phantoms.

5 1068 Tang et al. tected in the obese liver, compared to about 8% lipid in the lean liver. These images demonstrate that the gradient reversal based water and fat imaging method can provide high-resolution lipid distribution in mouse liver. Figure 5 shows the high-resolution cross-sectional water and fat images and the lipid percentage map in hind limb muscle of the obese and lean mice. Figure 5a and c are water images of the obese and lean mice, respectively; Fig. 5b and d are fat images. Several regions of interest marked in Fig. 5a are the tibialis anterior (TA); soleus (S); and medial head of the gastrocnemius (MHG). The color-coded lipid percentage maps in the muscle of the left hind limbs are shown in Fig. 5e Figure 4. In vivo water and fat images (PD weighted, echo train spacing 4.5 msec) and lipid quantification in mouse liver. a: Water image of an obese mouse; (b) fat image of the obese mouse; (c) fat plus water image; (d) lipid percentage map; (e) R2 map of water; (f) R2 map of fat; (g) color-coded liver lipid percentage map of the obese mouse; and (h) color-coded liver lipid percentage map of the lean mouse. Figure 5. In vivo water and fat images (PD weighted, echo train spacing 4.5 msec) and lipid quantification in mouse hind limb muscle. a: Water image of an obese mouse, with regions of interest marked: TA, tibialis anterior; S, soleus; MHG, medial head of the gastrocnemius; (b) fat image of the obese mouse; (c) water image of a lean mouse; (d) fat image of the lean mouse; (e) color-coded lipid percentage map in left hind limb muscle of the obese mouse; (f) color-coded lipid percentage map in left hind limb muscle of the lean mouse; and (g) normalized histogram of lipid percentage in MHG. [Color figure can be viewed in the online issue, which is available at

6 An Interweaved Water and Fat Imaging Technique 1069 Figure 6. Ex vivo interleaved water and fat images of a rat knee sample. a: Three-slice water images of the rat knee; and (b) three-slice fat images of the rat knee. and f. The subcutaneous and intermuscular fat regions were not included in the muscle lipid percentage map. Analysis of regional lipid percentage maps showed that the lipid content of the MHG was higher than that of the TA and soleus. The normalized histogram of lipid percentage in MHG of the two mice is shown in Fig. 5g. About 50% more lipid was detected in the MHG of the obese mouse (mean, 2.9%) than that of the lean mouse (1.9%). No significant difference was found in the histograms of lipid percentage in TA (obese: 1.7%, lean: 1.3%) and soleus (obese: 1.8%, lean: 1.6%) between the two mice. The regional muscle lipid distribution and quantification was consistent with that observed in a human imaging study (16). These results indicate that the lipid content in mouse limb muscle is detectable using the water and fat imaging protocol performed with a high-field magnet. Figure 6a shows the three-slice water images of a formalin-fixed rat knee, giving a good depiction of bone structure and cartilage. The corresponding fat images are shown in Fig. 6b. The central slice shows the separation of water and fat images and allows a much clearer delineation of cartilage on the femur from the adjacent fat pad. Such accurate water and fat imaging could be useful for quantifying the spatial water and fat content in bone for osteoarthritis as well as other joint disease/injury applications. DISCUSSION The utility of the gradient reversal technique for interleaved and quantitative water/fat imaging was evaluated at 11.7T in several applications requiring separation of fat and water. The technique has its pros and cons. The main advantage of this technique is that the highly effective separation of water and fat, and chemical shift artifact free water and fat imaging could be achieved in a single acquisition. It employs a modified spin echo sequence without requiring presaturation or spatial-spectral selective RF pulses. Its implementation does not demand a high gradient amplitude or switching rate. At very high field, the requirement of this technique, rf off, can be easily satisfied with RF pulses of reasonable durations. On the other hand, the sequence requires the presence of at least two sliceselective RF pulses. In our study, durations of the 90 and 180 RF pulses were chosen such that the spin echo slice excitation bandwidth was approximately the water fat chemical shift separation, and was applicable to very thin slices as long as the slice gradient strength was sufficient. Since both water and fat images were acquired at the same slice locations during a multislice 2D acquisition, the maximum number of slices for either water or fat was actually reduced by half of that of the conventional spin echo acquisition. Although a minimum slice gap is required during a multislice 2D acquisition to avoid adjacent slice interference, contiguous slices could be obtained in the same acquisition using the 3D gradient reversal based sequence. As with all other fat-/water-selective imaging techniques, gradient reversal method is sensitive to the B 0 inhomogeneity. In this study, for example, it was necessary to shim the slice region to well below the water fat chemical shift difference of 3.5 ppm to separate water and fat completely for accurate quantitation. The other requirement is optimization of the RF pulse to minimize the water/fat contamination due to imperfect slice-selection profiles. Several pulse shapes were examined, including Sinc, Hermite, and Gaussian. For the results presented, the Gaussian-shaped RF pulses generating the slice-selective bandwidth of 1750 Hz (full-width at 20% maximum) were employed for the excitation and refocusing because they are power efficient and produce no lasting ripples in the slice profile. During the fat-selective imaging, for example, the water contamination arising from the overlapping of 90 and 180 water slice-selection profiles was found to be below 1% based on the phantom assessments. Potentially, the power-efficient RF pulses could be further optimized using numerical methods, e.g., the Shinnar-Le Roux (SLR) RF pulse shape design (17 19), to provide a good rectangular slice profile for more accurate quantitative measurements. In conclusion, the gradient reversal technique for chemical shift selective imaging was evaluated for applications at very high field. An interleaved water and fat imaging sequence with in-plane chemical-shift correction was successfully implemented at 11.7T, by which complete separation of water and fat was achieved in a single acquisition without increasing acquisition time. The ability for quantitative lipid measurements was verified in phantoms and demonstrated in mouse liver and hind limb muscle. Compared to the phase-related method such as the Dixon method, the gradient reversal based technique can image both water and fat at the identical slice locations and does not require multiple acquisitions. At very high field, this water and fat imaging approach is simple to implement, robust, and quantitative. It is applicable to the study of lipid-modifying therapies in animal models as well as lipid quantification in humans. ACKNOWLEDGMENTS We thank Xiaodong Yang for animal assistance, Corin Miller for support with the NMR imaging facilities, and

7 1070 Tang et al. Richard Hargreaves for critical reading of the manuscript and helpful comments. All MRI experiments were performed at the Department of Imaging, Merck Research Laboratories, Merck & Co., Inc. REFERENCES 1. Haase A, Frahm J, Hänicke W, Matthaei D. 1 H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985;30: Laurent WM, Bonny JM, Renou JP. Imaging of water and fat fractions in high-field MRI with multiple slice chemical shift-selective inversion recovery. J Magn Reson Imaging 2000;12: Schick F. Simultaneous highly selective MR water and fat imaging using a simple new type of spectral-spatial excitation. Magn Reson Med 1998;40: Kwok WE, Totterman SM, Zhong J. 3D interleaved water and fat image acquisition with chemical-shift correction. Magn Reson Med 2000;44: Dixon WT. Simple proton spectroscopic imaging. Radiology 1984; 153: Glover G, Schneider E. Three-point Dixon technique for true fat/ water decomposition with B0 inhomogeneity correction. Magn Reson Med 1991;18: Zhang X, Tengowski M, Fasulo L, Botts S, Suddarth BS, Johnson GA. Measurement of fat/water ratios in rat liver using 3D threepoint Dixon MRI. Magn Reson Med 2004;51: Reeder SB, Pineda AR, Wen Z, et al. Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med 2005;54: Reeder SB, Wen Z, Yu H, et al. Multicoil Dixon chemical species separation with an iterative least squares estimation method. Magn Reson Med 2004;51: Park HW, Kim DJ, Cho ZH. Gradient reversal technique and its applications to chemical-shift-related NMR imaging. Magn Reson Med 1987;4: Volk A, Tiffon B, Mispelter J, Lhoste JM. Chemical shift-specific slice selection a new method for chemical shift imaging at high magnetic field. J Magn Reson 1987;71: Garbow JR, Lin X, Sakata N, Chen Z, Koh D, Schonfeld G. In vivo MRS measurement of liver lipid levels in mice. J Lipid Res 2004; 45: Poon CS, Szumowski J, Plewes DB, Ashby P, Henkelman RM. Fat/water quantitation and differential relaxation time measurement using chemical shift imaging technique. Magn Reson Imaging 1989;7: Corran JW. Emulsion technology. Brooklyn, NY: Chem Publ Co; Emmanuelle MD, Javier B, Glyn J, Jean R, Xavier M, Anne C. Fat suppression in MR imaging: techniques and pitfalls. Radiographics 1999;19: Goodpaster BH, Stenger VA, Boada F, et al. Skeletal muscle lipid concentration quantified by magnetic resonance imaging. Am J Clin Nutr 2004;79: Shinnar M, Bolinger L, Leigh JS. The synthesis of soft pulses with a specified frequency response. Magn Reson Med 1989;12: Shinnar M, Bolinger L, Leigh JS. The use of finite impulse response filters in pulse design. Magn Reson Med 1989;12: Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm. IEEE Trans Med Imaging 1991;10:53 65.

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE In Practice RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE By Atsuya Watanabe, MD, PhD, Director, Advanced Diagnostic Imaging Center and Associate Professor, Department of Orthopedic Surgery, Teikyo

More information

Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases

Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases Removal of Nuisance Signal from Sparsely Sampled 1 H-MRSI Data Using Physics-based Spectral Bases Qiang Ning, Chao Ma, Fan Lam, Bryan Clifford, Zhi-Pei Liang November 11, 2015 1 Synopsis A novel nuisance

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/35124 holds various files of this Leiden University dissertation. Author: Wokke, Beatrijs Henriette Aleid Title: Muscle MRI in Duchenne and Becker muscular

More information

Assessment of Adipose Tissue from Whole Body 3T MRI Scans

Assessment of Adipose Tissue from Whole Body 3T MRI Scans Assessment of Adipose Tissue from Whole Body 3T MRI Scans Ting Song 1, Jing An 2, Qun Chen 2, Vivian Lee 2, Andrew Laine 1 1 Department of Biomedical Engineering, Columbia University, New York, NY, USA

More information

Magnetic Resonance Angiography

Magnetic Resonance Angiography Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous

More information

Liver Fat Quantification

Liver Fat Quantification Liver Fat Quantification Jie Deng, PhD, DABMP Department of Medical Imaging May 18 th, 2017 Disclosure Research agreement with Siemens Medical Solutions 2 Background Non-alcoholic fatty liver diseases

More information

High-Sensitivity Coil Array for Head and Neck Imaging: Technical Note

High-Sensitivity Coil Array for Head and Neck Imaging: Technical Note AJNR Am J Neuroradiol 22:1881 1886, November/December 2001 Technical Note High-Sensitivity Coil Array for Head and Neck Imaging: Technical Note Roland G. Henry, Nancy J. Fischbein, William P. Dillon, Daniel

More information

High Field MR of the Spine

High Field MR of the Spine Department of Radiology University of California San Diego 3T for MR Applications Advantages High Field MR of the Spine Increased signal-to-noise Better fat suppression Increased enhancement with gadolinium

More information

MR Advance Techniques. Vascular Imaging. Class II

MR Advance Techniques. Vascular Imaging. Class II MR Advance Techniques Vascular Imaging Class II 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology

More information

Real-Time Black-Blood MRI Using Spatial Presaturation

Real-Time Black-Blood MRI Using Spatial Presaturation JOURNAL OF MAGNETIC RESONANCE IMAGING 13:807 812 (2001) Technical Note Real-Time Black-Blood MRI Using Spatial Presaturation Krishna S. Nayak, Ph.D., 1 * Pedro A. Rivas, M.D., 2 John M. Pauly, Ph.D., 1

More information

Fat Suppression in the Abdomen

Fat Suppression in the Abdomen Clinical How I do it? Fat Suppression in the Abdomen Wilhelm Horger Siemens Medical Solutions, Erlangen, Germany Introduction Due to the different chemical environment, hydrogen nuclei in - and in -tissue

More information

Comparison of Muscle Image Acquisition & Analysis Techniques

Comparison of Muscle Image Acquisition & Analysis Techniques Comparison of Muscle Image Acquisition & Analysis Techniques Toronto General Research Institute Andy Kin On Wong PhD Assistant UHN Scientist Scanco User Meeting, September 20, 2016 2:15 pm Muscle & Bone

More information

Extraneous Lipid Contamination in Single-Volume Proton MR Spectroscopy: Phantom and Human Studies

Extraneous Lipid Contamination in Single-Volume Proton MR Spectroscopy: Phantom and Human Studies Extraneous Lipid Contamination in Single-Volume Proton MR Spectroscopy: Phantom and Human Studies Lester Kwock, Mark A. Brown, and Mauricio Castillo PURPOSE: To determine the degree of extraneous lipid

More information

FieldStrength. Achieva 3.0T enables cutting-edge applications, best-in-class MSK images

FieldStrength. Achieva 3.0T enables cutting-edge applications, best-in-class MSK images FieldStrength Publication for the Philips MRI Community Issue 33 December 2007 Achieva 3.0T enables cutting-edge applications, best-in-class MSK images Palo Alto Medical Clinic Sports Medicine Center employs

More information

Personal use only. MRI Metal Artifact Reduction: Shoulder Implants and Arthroplasty. Reto Sutter, MD

Personal use only. MRI Metal Artifact Reduction: Shoulder Implants and Arthroplasty. Reto Sutter, MD MRI Metal Artifact Reduction: Shoulder Implants and Arthroplasty Reto Sutter, MD University Hospital Balgrist Zurich University of Zurich Cor PD fat sat 56-year old male patient with positive lift-off

More information

Echelon Oval provides a robust suite of leading musculoskeletal imaging capabilities for detailed assessment of all anatomy for your most challenging

Echelon Oval provides a robust suite of leading musculoskeletal imaging capabilities for detailed assessment of all anatomy for your most challenging Echelon Oval provides a robust suite of leading musculoskeletal imaging capabilities for detailed assessment of all anatomy for your most challenging cases. Hitachi Medical Systems America, Inc. 1959 Summit

More information

Rapid Quantitation of High-Speed Flow Jets

Rapid Quantitation of High-Speed Flow Jets Rapid Quantitation of High-Speed Flow Jets Krishna S. Nayak, 1 * Bob S. Hu, 1,2 and Dwight G. Nishimura 1 Magnetic Resonance in Medicine 50:366 372 (2003) Flow jets containing velocities up to 5 7 m/s

More information

High-Resolution 3D Cartilage Imaging with IDEAL SPGR at 3 T

High-Resolution 3D Cartilage Imaging with IDEAL SPGR at 3 T Siepmann et al. Knee MRI with IDEL SPGR Musculoskeletal Imaging Technical Innovation David. Siepmann 1 Jeff McGovern 2 Jean H. rittain 3 Scott. Reeder 1,4 Siepmann D, McGovern J, rittain JH, Reeder S Keywords:

More information

Quantification of liver steatosis in MRI: available techniques and use of transverse magnetization decay curve in patients with iron overload

Quantification of liver steatosis in MRI: available techniques and use of transverse magnetization decay curve in patients with iron overload Quantification of liver steatosis in MRI: available techniques and use of transverse magnetization decay curve in patients with iron overload Poster No.: C-1302 Congress: ECR 2013 Type: Educational Exhibit

More information

8/4/2016. MRI for Radiotherapy: MRI Basics. Nuclear Magnetic Resonance. Nuclear Magnetic Resonance. Wilson Miller

8/4/2016. MRI for Radiotherapy: MRI Basics. Nuclear Magnetic Resonance. Nuclear Magnetic Resonance. Wilson Miller MRI for Radiotherap: MRI asics Wilson Miller Universit of Virginia Department of Radiolog & Medical Imaging AAPM 2016 August 4, 2016 Nuclear Magnetic Resonance Magnetic resonance images are created using

More information

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new

More information

High-field MR imaging systems such as 3T strength aid in

High-field MR imaging systems such as 3T strength aid in TECHNICAL NOTE H.-S. Liu H.-W. Chung C.-J. Juan S.-Y. Tsai C.-Y. Wang C.-C. Chan G.-S. Huang M.-C. Chou C.-S. Lee C.-W. Ko N.-Y. Cho C.-Y. Chen Anomalous J-Modulation Effects on Amino Acids in Clinical

More information

Problem Set #8 Rad 226

Problem Set #8 Rad 226 Problem Set #8 1. J-editing. In class, we discussed J-editing for the doublet resonance of lactate. Other in vivo peaks (e.g. GABA) are more complicated (triplets, quartets, etc.). In this problem, we

More information

ACR MRI Accreditation: Medical Physicist Role in the Application Process

ACR MRI Accreditation: Medical Physicist Role in the Application Process ACR MRI Accreditation: Medical Physicist Role in the Application Process Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives

More information

Clinical Applications

Clinical Applications C H A P T E R 16 Clinical Applications In selecting pulse sequences and measurement parameters for a specific application, MRI allows the user tremendous flexibility to produce variations in contrast between

More information

Correction of CSF Motion Artifact on MR Images of the Brain and Spine by Pulse Sequence Modification: Clinical Evaluation

Correction of CSF Motion Artifact on MR Images of the Brain and Spine by Pulse Sequence Modification: Clinical Evaluation 1069 Correction of CSF Motion Artifact on MR Images of the Brain and Spine by Pulse Sequence Modification: Clinical Evaluation Nikolaus M. Szeverenyi 1 Stephen A. Kieffer Edwin D. Cacayorin A modification

More information

Optimal Timing for In Vivo 1 H-MR Spectroscopic Imaging of the Human Prostate at 3T

Optimal Timing for In Vivo 1 H-MR Spectroscopic Imaging of the Human Prostate at 3T Magnetic Resonance in Medicine 53:1268 1274 (2005) Optimal Timing for In Vivo 1 H-MR Spectroscopic Imaging of the Human Prostate at 3T Tom W.J. Scheenen, 1 * Giulio Gambarota, 1 Elisabeth Weiland, 2,3

More information

Basics of MRI Part I

Basics of MRI Part I Basics of MRI Part I Mathew J. Dixon, D.O. Chairman Department of Radiology Memorial Health University Medical Center Savannah, GA Objectives Brief History Concept of MRI Creation of a Magnetic Field Concepts

More information

How Much Tesla Is Too Much?

How Much Tesla Is Too Much? How Much Tesla Is Too Much? Johnny U. V. Monu, MB, BS; MSc Professor of Radiology and Orthopedics University of Rochester School of Medicine Rochester, New York Historical Timeline Clinical Imaging 1970

More information

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine

Non Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast

More information

T2, T2*, ute. Yeo Ju Kim. Radiology, Inha University Hospital, Incheon, Korea

T2, T2*, ute. Yeo Ju Kim. Radiology, Inha University Hospital, Incheon, Korea SY28-1 T2, T2*, ute Yeo Ju Kim Radiology, Inha University Hospital, Incheon, Korea T2 relaxation times relate to the rate of transverse magnetization decay, caused by the loss of phase coherence induced

More information

Why Talk About Technique? MRI of the Knee:

Why Talk About Technique? MRI of the Knee: Why Talk About Technique? MRI of the Knee: Part 1 - Imaging Techniques Mark Anderson, M.D. University of Virginia Health Sciences Center Charlottesville, Virginia Always had an interest teach our fellows

More information

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA

Essentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA 99. MRA Principles and Carotid MRA As described in Chapter 12, time of flight (TOF) magnetic resonance angiography (MRA) is commonly utilized in the evaluation of the circle of Willis. TOF MRA allows depiction

More information

Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping

Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping TR_002 Technical Reports Knee Articular Cartilage in an Asymptomatic Population : Comparison of T1rho and T2 Mapping Min A Yoon 1,*, Suk-Joo Hong 1, Chang Ho Kang 2, Baek Hyun Kim 3 1 Korea University

More information

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation

Previous talks. Clinical applications for spiral flow imaging. Clinical applications. Clinical applications. Coronary flow: Motivation for spiral flow imaging Joao L. A. Carvalho Previous talks Non-Cartesian reconstruction (2005) Spiral FVE (Spring 2006) Aortic flow Carotid flow Accelerated spiral FVE (Fall 2006) 2007? Department of Electrical

More information

Cardiovascular MR Imaging at 3 T: Opportunities, Challenges, and Solutions 1

Cardiovascular MR Imaging at 3 T: Opportunities, Challenges, and Solutions 1 TECHNICAL ADVANCEMENTS IN CARDIAC MR IMAGING 1612 Cardiovascular MR Imaging at 3 T: Opportunities, Challenges, and Solutions 1 Prabhakar Rajiah, MD, FRCR Michael A. Bolen, MD Abbreviations: BOLD = blood

More information

In vivo quantitative lipidic map of brown adipose tissue by chemical shift imaging at 4.7 tesla

In vivo quantitative lipidic map of brown adipose tissue by chemical shift imaging at 4.7 tesla In vivo quantitative lipidic map of brown adipose tissue by chemical shift imaging at 4.7 tesla Ernesto Lunati,*, Pasquina Marzola,* Elena Nicolato,* Mauro Fedrigo, Marco Villa, and Andrea Sbarbati 1,

More information

ACR Accreditation Update in MRI

ACR Accreditation Update in MRI ACR Accreditation Update in MRI Whole Body Systems Extremity (MSK) Ron Price Vanderbilt University Medical Center Nashville, TN Dedicated Breast MRI Accreditation Update 1. ACR MRI Accreditation: Overview,

More information

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use

Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use International Congress Series 1281 (2005) 793 797 www.ics-elsevier.com Visualization strategies for major white matter tracts identified by diffusion tensor imaging for intraoperative use Ch. Nimsky a,b,

More information

Spiral Coronary Angiography Using a Blood Pool Agent

Spiral Coronary Angiography Using a Blood Pool Agent JOURNAL OF MAGNETIC RESONANCE IMAGING 22:213 218 (2005) Original Research Spiral Coronary Angiography Using a Blood Pool Agent Steffen Ringgaard, PhD, 1 * Michael Pedersen, PhD, 1 Jonas Rickers, MD, 1,2

More information

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord Mark D. Keiper, Robert I. Grossman, John C. Brunson, and Mitchell D. Schnall PURPOSE:

More information

ASL BASICS II. Learning Objectives. Outline. Acquisition. M. A. Fernández-Seara, Ph. D. Arterial spin labeled perfusion MRI: basic theory

ASL BASICS II. Learning Objectives. Outline. Acquisition. M. A. Fernández-Seara, Ph. D. Arterial spin labeled perfusion MRI: basic theory Acquisition ASL BASICS II M. A. Fernández-Seara, Ph. D. Neuroimaging Laboratory Center for Applied Medical Research University of Navarra Pamplona, Spain Outline Arterial spin labeled perfusion MRI: basic

More information

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration Richard Kijowski, Michael Fazio, Benjamin Beduhn, and Fang Liu Department of Radiology University of Wisconsin School of Medicine

More information

Magnetic Resonance Imaging. Basics of MRI in practice. Generation of MR signal. Generation of MR signal. Spin echo imaging. Generation of MR signal

Magnetic Resonance Imaging. Basics of MRI in practice. Generation of MR signal. Generation of MR signal. Spin echo imaging. Generation of MR signal Magnetic Resonance Imaging Protons aligned with B0 magnetic filed Longitudinal magnetization - T1 relaxation Transverse magnetization - T2 relaxation Signal measured in the transverse plane Basics of MRI

More information

Master Thesis in Radiation Physics 09/10 Annie Olsson. Supervisors: Frank Risse Lars E. Olsson. Imaging centre AstraZeneca R&D Mölndal

Master Thesis in Radiation Physics 09/10 Annie Olsson. Supervisors: Frank Risse Lars E. Olsson. Imaging centre AstraZeneca R&D Mölndal Determination of a preclinical protocol for quantitative measurements of perfusion and permeability in the rat lung using dynamic contrast enhanced-mri Master Thesis in Radiation Physics 9/1 Annie Olsson

More information

MR Imaging of the Spine at 3.0T with T2-Weighted IDEAL Fast Recovery Fast Spin-Echo Technique

MR Imaging of the Spine at 3.0T with T2-Weighted IDEAL Fast Recovery Fast Spin-Echo Technique Original rticle http://dx.doi.org/10.3348/kjr.2012.13.1.44 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2012;13(1):44-52 MR Imaging of the Spine at 3.0T with T2-Weighted IDEL Fast Recovery Fast Spin-Echo

More information

Ultrashort TE Spectroscopic Imaging (UTESI): Application to the Imaging of Short T2 Relaxation Tissues in the Musculoskeletal System

Ultrashort TE Spectroscopic Imaging (UTESI): Application to the Imaging of Short T2 Relaxation Tissues in the Musculoskeletal System JOURNAL OF MAGNETIC RESONANCE IMAGING 29:412 421 (2009) Original Research Ultrashort TE Spectroscopic Imaging (UTESI): Application to the Imaging of Short T2 Relaxation Tissues in the Musculoskeletal System

More information

2D-Sigmoid Enhancement Prior to Segment MRI Glioma Tumour

2D-Sigmoid Enhancement Prior to Segment MRI Glioma Tumour 2D-Sigmoid Enhancement Prior to Segment MRI Glioma Tumour Pre Image-Processing Setyawan Widyarto, Siti Rafidah Binti Kassim 2,2 Department of Computing, Faculty of Communication, Visual Art and Computing,

More information

1Pulse sequences for non CE MRA

1Pulse sequences for non CE MRA MRI: Principles and Applications, Friday, 8.30 9.20 am Pulse sequences for non CE MRA S. I. Gonçalves, PhD Radiology Department University Hospital Coimbra Autumn Semester, 2011 1 Magnetic resonance angiography

More information

Repeatability of 2D FISP MR Fingerprinting in the Brain at 1.5T and 3.0T

Repeatability of 2D FISP MR Fingerprinting in the Brain at 1.5T and 3.0T Repeatability of 2D FISP MR Fingerprinting in the Brain at 1.5T and 3.0T Guido Buonincontri 1,2, Laura Biagi 1,3, Alessandra Retico 2, Michela Tosetti 1,3, Paolo Cecchi 4, Mirco Cosottini 1,4,5, Pedro

More information

Fat-Corrected T2 Measurement as a Marker of Active Muscle Disease in Inflammatory Myopathy

Fat-Corrected T2 Measurement as a Marker of Active Muscle Disease in Inflammatory Myopathy Musculoskeletal Imaging Original Research Yao and Gai MRI of Inflammatory Myopathy Musculoskeletal Imaging Original Research Lawrence Yao 1 Neville Gai Yao L, Gai N Keywords: MRI, muscle, myopathy, myositis,

More information

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution

Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution COMMUNICATIONS Magnetic Resonance in Medicine 44:511 515 (2000) Turbo ASL: Arterial Spin Labeling With Higher SNR and Temporal Resolution Eric C. Wong,* Wen-Ming Luh, and Thomas T. Liu A modified pulsed

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants Knee Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Knee Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All knee implants of Smith

More information

MAJOR PAPER. Introduction. Young Han Lee 1, Seok Hahn 1, Eunju Kim 2, and Jin-Suck Suh 1*

MAJOR PAPER. Introduction. Young Han Lee 1, Seok Hahn 1, Eunju Kim 2, and Jin-Suck Suh 1* Magn Reson Med Sci, Vol. 15, No. 4, pp. 371 378, 2016 doi:10.2463/mrms.mp.2015-0055 MAJOR PAPER Introduction Fat-suppressed MR Imaging of the Spine for Metal Artifact Reduction at 3T: Comparison of STIR

More information

Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast

Musculoskeletal MRI at 3.0 T: Relaxation Times and Image Contrast Downloaded from www.ajronline.org by 46.3.25.67 on 12/5/17 from IP address 46.3.25.67. Copyright RRS. For personal use only; all rights reserved Garry E. Gold 1 Eric Han 2 Jeff Stainsby 3 Graham Wright

More information

An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles

An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles An fmri Phantom Based on Electric Field Alignment of Molecular Dipoles Innovative Graduate Student Proposal Yujie Qiu, Principal Investigator, Graduate Student Joseph Hornak, Faculty Sponsor, Thesis Advisor

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Gregg NM, Kim AE, Gurol ME, et al. Incidental cerebral microbleeds and cerebral blood flow in elderly individuals. JAMA Neurol. Published online July 13, 2015. doi:10.1001/jamaneurol.2015.1359.

More information

In vivo diffusion tensor imaging (DTI) of articular cartilage as a biomarker for osteoarthritis

In vivo diffusion tensor imaging (DTI) of articular cartilage as a biomarker for osteoarthritis In vivo diffusion tensor imaging (DTI) of articular cartilage as a biomarker for osteoarthritis Jose G. Raya 1, Annie Horng 2, Olaf Dietrich 2, Svetlana Krasnokutsky 3, Luis S. Beltran 1, Maximilian F.

More information

Table 1. Summary of PET and fmri Methods. What is imaged PET fmri BOLD (T2*) Regional brain activation. Blood flow ( 15 O) Arterial spin tagging (AST)

Table 1. Summary of PET and fmri Methods. What is imaged PET fmri BOLD (T2*) Regional brain activation. Blood flow ( 15 O) Arterial spin tagging (AST) Table 1 Summary of PET and fmri Methods What is imaged PET fmri Brain structure Regional brain activation Anatomical connectivity Receptor binding and regional chemical distribution Blood flow ( 15 O)

More information

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging

Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR Imaging Provläsningsexemplar / Preview TECHNICAL REPORT ISO/TR 16379 First edition 2014-03-01 Tissue-engineered medical products Evaluation of anisotropic structure of articular cartilage using DT (Diffusion Tensor)-MR

More information

Focal fatty sparing and focal fatty infiltration can be observed in the

Focal fatty sparing and focal fatty infiltration can be observed in the Diagn Interv Radiol 2011; 17:323 327 Turkish Society of Radiology 2011 ABDOMINAL IMAGING ORIGINAL ARTICLE Focal sparing of iron and fat in liver tissue in patients with hemosiderosis: diagnosis with combination

More information

Effects of Magnetic Susceptibility Artifacts and Motion in Evaluating the Cervical Neural Foramina on 30FT Gradient-Echo MR Imaging

Effects of Magnetic Susceptibility Artifacts and Motion in Evaluating the Cervical Neural Foramina on 30FT Gradient-Echo MR Imaging 237 Effects of Magnetic Susceptibility Artifacts and Motion in Evaluating the Cervical Neural Foramina on 30FT Gradient-Echo MR Imaging JayS. Tsuruda 1 Kent Remley The purpose of this study was to evaluate

More information

6/23/2009. Inversion Recovery (IR) Techniques and Applications. Variations of IR Technique. STIR, FLAIR, TI and TI Null. Applications of IR

6/23/2009. Inversion Recovery (IR) Techniques and Applications. Variations of IR Technique. STIR, FLAIR, TI and TI Null. Applications of IR The Anatomy of Basic R Pulse Sequences Inversion Recovery () Techniques and Applications Chen Lin, PhD Indiana University School of edicine & Clarian Health Partners agnetization Preparation Section Chemical

More information

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol Sample referral sticker Referral Sticker Insert here Corin 17 Bridge Street Pymble NSW Australia 2073 P: +61 (0)2 9497 7400 F: +61 (0)2 9497 7498 E: KAS.customerservice@coringroup.com

More information

40 th Annual Meeting of the SCBT/MR

40 th Annual Meeting of the SCBT/MR Liver Fat and Iron Quantification 40 th Annual Meeting of the SCBT/MR Nashville, TN September 11, 2017 Scott B. Reeder, MD, PhD 100% 0% Department of Radiology University of Wisconsin Madison, WI Disclosures

More information

Perfusion-Based fmri. Thomas T. Liu Center for Functional MRI University of California San Diego May 7, Goal

Perfusion-Based fmri. Thomas T. Liu Center for Functional MRI University of California San Diego May 7, Goal Perfusion-Based fmri Thomas T. Liu Center for Functional MRI University of California San Diego May 7, 2006 Goal To provide a basic understanding of the theory and application of arterial spin labeling

More information

Raja Muthupillai, PhD. Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital. Research Support: Philips Healthcare

Raja Muthupillai, PhD. Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital. Research Support: Philips Healthcare 3D Cardiac Imaging Raja Muthupillai, PhD Department of Diagnostic and Interventional Radiology St. Luke s Episcopal Hospital Houston, TX Disclosures Research Support: Philips Healthcare This presentation

More information

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils Clinical Orthopedic Imaging Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils Elena Ferrer 1 ; Rafael Coronado Santos 2 1 Radiology Department, Clínica Creu Blanca, Barcelona,

More information

Standardized, Reproducible, High Resolution Global Measurements of T1 Relaxation Metrics in Cases of Multiple Sclerosis

Standardized, Reproducible, High Resolution Global Measurements of T1 Relaxation Metrics in Cases of Multiple Sclerosis AJNR Am J Neuroradiol 24:58 67, January 2003 Standardized, Reproducible, High Resolution Global Measurements of T1 Relaxation Metrics in Cases of Multiple Sclerosis Radhika Srinivasan, Roland Henry, Daniel

More information

Real-Time MRI of Joint Movement With TrueFISP

Real-Time MRI of Joint Movement With TrueFISP JOURNAL OF MAGNETIC RESONANCE IMAGING 15:710 715 (2002) Technical Note Real-Time MRI of Joint Movement With TrueFISP Harald H. Quick, MSc, 1 * Mark E. Ladd, PhD, 1 Matthias Hoevel, MD, 2 Silke Bosk, RT,

More information

The Effects of Music intervention on Functional connectivity. Supplemental Information

The Effects of Music intervention on Functional connectivity. Supplemental Information Yang et al. 0 The Effects of Music intervention on Functional connectivity strength of Brain in Schizophrenia Supplemental Information Mi Yang,#, Hui He #, Mingjun Duan,, Xi Chen, Xin Chang, Yongxiu Lai,

More information

Intrinsic Signal Optical Imaging

Intrinsic Signal Optical Imaging Intrinsic Signal Optical Imaging Introduction Intrinsic signal optical imaging (ISOI) is a technique used to map dynamics in single cells, brain slices and even and most importantly entire mammalian brains.

More information

Clinical application of 3.0 T proton MR spectroscopy in evaluation of pancreatic diseases

Clinical application of 3.0 T proton MR spectroscopy in evaluation of pancreatic diseases Clinical application of 3.0 T proton MR spectroscopy in evaluation of pancreatic diseases Award: Cum Laude Poster No.: C-1762 Congress: ECR 2012 Type: Scientific Paper Authors: T. Su, E. Jin; Beijing/CN

More information

Sung Hong Park. M.S. in Electrical Engineering, KAIST, South Korea, Submitted to the Graduate Faculty of

Sung Hong Park. M.S. in Electrical Engineering, KAIST, South Korea, Submitted to the Graduate Faculty of NONINVASIVE IMAGING OF BRAIN VASCULATURE WITH HIGH RESOLUTION BLOOD OXYGENATION LEVEL DEPENDENT VENOGRAPHY IN MAGNETIC RESONANCE IMAGING: APPLICATIONS TO FUNCTIONAL AND CLINICAL STUDIES by Sung Hong Park

More information

In vivo MRI using positive-contrast techniques in detection of cells labeled with superparamagnetic iron oxide nanoparticles y

In vivo MRI using positive-contrast techniques in detection of cells labeled with superparamagnetic iron oxide nanoparticles y NMR IN BIOMEDICINE NMR Biomed. 2008; 21: 242 250 Published online 13 June 2007 in Wiley InterScience (www.interscience.wiley.com) DOI:10.1002/nbm.1187 In vivo MRI using positive-contrast techniques in

More information

BioMatrix Tuners: CoilShim

BioMatrix Tuners: CoilShim MAGNETOM Vida special issue Head and Neck Imaging Clinical 11 BioMatrix Tuners: CoilShim Miriam R. Keil, Ph.D.; Jörg Rothard; Carmel Hayes, Ph.D. Siemens Healthineers, Erlangen, Germany A cervical spine

More information

On the feasibility of speckle reduction in echocardiography using strain compounding

On the feasibility of speckle reduction in echocardiography using strain compounding Title On the feasibility of speckle reduction in echocardiography using strain compounding Author(s) Guo, Y; Lee, W Citation The 2014 IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL.,

More information

NIH Public Access Author Manuscript Obesity (Silver Spring). Author manuscript; available in PMC 2010 April 1.

NIH Public Access Author Manuscript Obesity (Silver Spring). Author manuscript; available in PMC 2010 April 1. NIH Public Access Author Manuscript Published in final edited form as: Obesity (Silver Spring). 2010 April ; 18(4): 841 847. doi:10.1038/oby.2009.352. Comparison of Fat Water MRI and Single-voxel MRS in

More information

Functional Chest MRI in Children Hyun Woo Goo

Functional Chest MRI in Children Hyun Woo Goo Functional Chest MRI in Children Hyun Woo Goo Department of Radiology and Research Institute of Radiology Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea No ionizing radiation

More information

Case Report: Knee MR Imaging of Haemarthrosis in a Case of Haemophilia A

Case Report: Knee MR Imaging of Haemarthrosis in a Case of Haemophilia A Clinical > Pediatric Imaging Case Report: Knee MR Imaging of Haemarthrosis in a Case of Haemophilia A M. A. Weber, J. K. Kloth University Hospital Heidelberg, Department of Diagnostic and Interventional

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle   holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/35124 holds various files of this Leiden University dissertation. Author: Wokke, Beatrijs Henriette Aleid Title: Muscle MRI in Duchenne and Becker muscular

More information

Using a Phantom to Compare MR Techniques for Determining the Ratio of Intraabdominal to Subcutaneous Adipose Tissue

Using a Phantom to Compare MR Techniques for Determining the Ratio of Intraabdominal to Subcutaneous Adipose Tissue Lane F. Donnelly 1,2 Kendall J. O Brien 1 Bernard J. Dardzinski 1,2 Stacy A. Poe 2 Judy A. Bean 2 Scott K. Holland 1,2 Stephen R. Daniels 2 Received April 29, 2002; accepted after revision August 27, 2002.

More information

Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (HASTE) MR: Comparison with Fast Spin-Echo MR in Diseases of the Brain

Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (HASTE) MR: Comparison with Fast Spin-Echo MR in Diseases of the Brain Half-Fourier Acquisition Single-Shot Turbo Spin-Echo (HASTE) MR: Comparison with Fast Spin-Echo MR in Diseases of the Brain Mahesh R. Patel, Roman A. Klufas, Ronald A. Alberico, and Robert R. Edelman PURPOSE:

More information

PERFUSION MRI CONTRAST BASED TECHNIQUES

PERFUSION MRI CONTRAST BASED TECHNIQUES PERFUSION MRI CONTRAST BASED TECHNIQUES by Kenny K Israni Mar 28, 2006 PERFUSION - MRI Dynamic Susceptibility contrast Dynamic Relaxivity contrast STEADY-STATE STATE TECHNIQUES Steady-state Susceptibility

More information

Methods of MR Fat Quantification and their Pros and Cons

Methods of MR Fat Quantification and their Pros and Cons Methods of MR Fat Quantification and their Pros and Cons Michael Middleton, MD PhD UCSD Department of Radiology International Workshop on NASH Biomarkers Washington, DC April 29-30, 2016 1 Disclosures

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants Shoulder Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Shoulder Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All shoulder implants

More information

Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, cont

Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, cont Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Peter Beddy, FFRRCSI,

More information

Comparison of Fat Water MRI and Single-voxel MRS in the Assessment of Hepatic and Pancreatic Fat Fractions in Humans

Comparison of Fat Water MRI and Single-voxel MRS in the Assessment of Hepatic and Pancreatic Fat Fractions in Humans nature publishing group articles Comparison of Fat Water MRI and Single-voxel MRS in the Assessment of Hepatic and Pancreatic Fat Fractions in Humans Houchun H. Hu, Hee-Won Kim, Krishna S. Nayak and Michael

More information

Abdominal applications of DWI

Abdominal applications of DWI Postgraduate course, SPR San Antonio (Texas), May 14-15, 2013 Abdominal applications of DWI Rutger A.J. Nievelstein Wilhelmina Children s s Hospital, Utrecht (NL) Outline What is DWI? How to perform? Challenges

More information

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants Hip Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Hip Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All hip implants of Smith &

More information

NEW SUBTRACTION ALGORITHMS FOR EVALUATION OF BREAST LESIONS ON DYNAMIC CONTRAST ENHANCED MR MAMMOGRAPHY

NEW SUBTRACTION ALGORITHMS FOR EVALUATION OF BREAST LESIONS ON DYNAMIC CONTRAST ENHANCED MR MAMMOGRAPHY A-056 NEW SUBTRACTION ALGORITHMS FOR EVALUATION OF BREAST LESIONS ON DYNAMIC CONTRAST ENHANCED MR MAMMOGRAPHY So Hee Cho, M.D., Byung Gil Choi, M.D., Hak Hee Kim, M.D., Euy Neyng Kim, M.D., Bum-soo Kim,

More information

ACR Breast MRI Accreditation Program - DRAFT

ACR Breast MRI Accreditation Program - DRAFT ACR Breast MRI Accreditation Program - DRAFT Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Provide an overview of the ACR Breast MRI Accreditation Program (BMRAP)

More information

Pictorial Essay. Jang Gyu Cha, MD, Hyun Sook Hong, MD, Jai Soung Park, MD, Sang Hyun Paik, MD, Hae Kyung Lee, MD INTRODUCTION

Pictorial Essay. Jang Gyu Cha, MD, Hyun Sook Hong, MD, Jai Soung Park, MD, Sang Hyun Paik, MD, Hae Kyung Lee, MD INTRODUCTION Pictorial Essay http://dx.doi.org/10.3348/kjr.2012.13.3.332 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2012;13(3):332-341 Practical pplication of Iterative Decomposition of Water and Fat with Echo

More information

Initial Clinical Experience of TOSHIBA 3T MRI

Initial Clinical Experience of TOSHIBA 3T MRI The 21st Conference of the Japanese Society of Cardiovascular Imaging & Dynamics Sponsored Seminar The Leading Edge of CT/MRI Diagnosis for the Cardiovascular System Initial Clinical Experience of TOSHIBA

More information

Basics of Quantification of ASL

Basics of Quantification of ASL Basics of Matthias van Osch Contents Thoughts about quantification Some quantification models Bloch equations Buxton model Parameters that need to be estimated Labeling efficiency M Bolus arrival time

More information

Phase-Sensitive Inversion Recovery for Detecting Myocardial Infarction Using Gadolinium-Delayed Hyperenhancement

Phase-Sensitive Inversion Recovery for Detecting Myocardial Infarction Using Gadolinium-Delayed Hyperenhancement Magnetic Resonance in Medicine 47:372 383 (2002) Phase-Sensitive Inversion Recovery for Detecting Myocardial Infarction Using Gadolinium-Delayed Hyperenhancement Peter Kellman,* Andrew E. Arai, Elliot

More information

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015 MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute

More information

Personalized Solutions. MRI Protocol for PSI and Signature Guides

Personalized Solutions. MRI Protocol for PSI and Signature Guides Personalized Solutions MRI Protocol for PSI and Signature Guides 2 Personalized Solutions MRI Protocol for PSI and Signature Guides Purpose and Summary This protocol is applicable for the Zimmer Biomet

More information

Time-Of-Flight MRA. Faculty Disclosures Vincent B. Ho, M.D. Presentation Objectives. MRA Techniques. Pros and Cons of MRA

Time-Of-Flight MRA. Faculty Disclosures Vincent B. Ho, M.D. Presentation Objectives. MRA Techniques. Pros and Cons of MRA Faculty Disclosures Vincent B. Ho, M.D. MR Angiography Techniques and Pitfalls Financial Disclosure Grant/Research Support General Electric Medical Systems Off-Label/Investigational Drug Use Dr. Ho will

More information