Transtelephone Pacemaker Monitoring

Size: px
Start display at page:

Download "Transtelephone Pacemaker Monitoring"

Transcription

1 Transtelephone Pacemaker Monitoring Five Years Later Seymour Furman, M.D., and Doris J. W. Escher, M.D. ABSTRACT Six hundred nineteen patients have been followed by remote monitoring of pacemaker function using ECG and rate or rate alone; 278 of 280 have had battery exhaustion or electronic failure demonstrated. Ten percent of exhausted pacemakers failed prior to the average longevity of the particular model, and 32% (89 of 280) exceeded 36 months longevity; of these, 13% (37 of 280) lasted more than 40 months and 4.6% (13 of 280) exceeded 50 months. The error rate is 0.7% (2 of 280). With pulse generator longevity increasing, monitoring is done less frequently during the first 2 years, then calls are made weekly after 24 months. T he durability of pacemaker systems has always been variable. Patient welfare has required observation following implantation to reduce the incidence of sudden and unpredicted failure of the implanted system as well as to detect early failure of some units and maximize the useful longevity of others. Pacemaker Failure The major reason for the necessity of follow-up over a prolonged period is that the time of failure is spread over a large portion of the total longevity of any pulse generator model. Elective removal of all generators during normal function, as when 10% of a series has failed, produces extremely short functional periods and accepts an undetected failure rate of 10%. A recent model of ventricular inhibited pacemaker (VVI)* had 10% failure at 15 months. Removal of the entire series at that point would have produced an average longevity of 13 months. When all the units were followed to exhaustion or electronic deterioration, half the series continued to function at 28 months, with an average longevity of 28.3 months and a standard deviation of 8.8 months (Fig. 1). A model contemporary with it had a 10% failure time of 16 months and would have had an average longevity of 15 months if all units had been removed at that time. Fifty percent had failed at 22 months, and the average longevity was 23 months with a standard deviation of 4.8 months [S]. Such wide standard deviations required that one group of patients be carefully monitored over 18 months (? 1 SD) and the other over 10 months (& 1 SD) in order to achieve a 66% detection rate for impending failure; these periods had to From the Cardiothoracic Service, Division of Surgery and the Cardiology Service, Division of Medicine, Montefiore Hospital and Medical Center, Bronx, N.Y. Supported in part by U.S. Public Health Service Grant no. HE Presented at the Eleventh Annual Meeting - of The Society of Thoracic Surgeons, - Montreal, Que., Canada, Jan , Address remint reauests to Dr. Furman. 111 E. 210th St., Bronx, N.Y *Terminology for pulse generators follows that suggested by Parsonnet, Furman, and Smyth [ THE ANNALS OF THORACIC SURGERY

2 Transtelephone Pacemaker Monitoring c % E - -..,, I 1 i s Number Of Months After Implant Number Of Failurem FIG. 1. This model, successful for its time, was implanted during Failure was approximately normal in distribution, with mean and median at 28 months and a standard deviation of 8.8 months. Ninety-five percent of failures occurred over a period of 35 months. Data on Cordis Model 143A7: total sample, 101; total calculated, 50; total dead or removed, 36; lost to follow-up, 5; remaining, 10 (average, 41.O mo); average, 28.3 mo; median, 28 mo; standard deviation, 8.8 mo. be doubled to detect 95% of failures. For neither group would elective replacement at 15 to 16 months have been consistent with patient safety or efficient use of the pulse generator. Were the spread, or standard deviation, of failure to narrow, monitoring could begin later and would be required for only a short period. A standard deviation of only 1 month for pulse generator failure would require monitoring for a 4-month period to detect 95% of all failures. Battery exhaustion has been and remains the single most common reason for pulse generator replacement, but electronic and packaging defects must also be considered. During , of 32 1 pulse generators (average longevity, 27 months) removed for some variety of failure, 123 of the failures were electronic in nature (average longevity, 19.1 months) and 198 were due to battery exhaustion (average longevity, months). A late postimplant rising threshold still occurs at a rate of about 3% each year. Electrode fracture remains a persistent event, causing 4 to 8% of the yearly secondary procedures [lo]. The need for monitoring will change if the average duration and reliability of pulse generators improve and if the initial failures occur later. The average useful functional period of generators removed for battery exhaustion during 1974 was 32 months (Fig. 2), during which 35% of the patients died. A reliable increase of the 10% pacemaker failure time from 16 to 32 months would substantially reduce the need for monitoring for a very large group and would shift the monitoring burden to failures occurring later. The mercury-zinc cell has established generator longevity to be between 18 and 36 months, with a sudden voltage decline and battery failure at end of life. An energy source that decays linearly and progressively can be monitored less frequently and removed when rate or another determinant reaches a point incom- VOL. 20, NO. 3, SEPTEMBER,

3 FURMAN AND ESCHER a.w 3a 37 M j; - 30! 20 I7 2 I6 rae c.- 18 pr, f 12 u mil 12 I a I011 I FIG. 2. Normal battery exhaustion is occurring later with improved pulse generator design. Pacemakers removed for exhaustion only, tabulated on a monthly basis, have increased in longevity from 21.3 months in 1971 to 31.8 months in patible with normal function. In the next few years, significant changes in time of onset as well as frequency of monitoring can be anticipated as new energy sources are used. A high incidence of electronic malfunction reduces the accuracy of failure detection, as comprehension of a pulse generator failure pattern is the basis of any follow-up system. Sudden changes in this pattern, or a high incidence of random failures, allow a high number of failures to go undetected until the monitoring system is adjusted to a-new schedule. From 1970 to 1973 the overall accuracy of failure prediction was about 80%. Emergencies during 1974 occurred because of electrode fracture (4%) and electronic failure (9.8%). Battery exhaustion is a process that usually falls within anticipated time limits and is almost always detected. Of the 19 pulse generator replacements done in 1974 for undetected failure, 16 (84%) were necessitated by failure of electronic components that occurred during the first year after implanting. Seven other impending electronic failures (3.7%) were detected by telephone and the generators replaced before failure occurred. This period previously had been one of light observation. The problems of failure detection were eventually resolved by early frequent telephone monitoring and by elective removal of a large number of suspect pulse generators (32 of 194, or 16.5% of the total for 1974). Methods of Monitoring The electronic clinic demonstrated that measurement of electronic function allows pulse generator failure to be predicted. One month, the shortest practical 328 THE ANNALS OF THORACIC SURGERY

4 Transtelephone PacemaRer Monitoring interval between patient visits, is not frequent enough to achieve the necessary accuracy for detecting battery depletion, which requires weekly data determination [2,7]. Electronic monitoring served as the mainstay of pacemaker follow-up for many years [l 1, 141 and is now complementary to the transtelephone system. Use of both telephonic and electronic follow-up has shown that visits at intervals of 4 months during the first year after implantation, every 3 months during the second year, and telephonic follow-up weekly thereafter are most efficient and productive, but the schedule must be modified as experience with specific pacemaker models dictates. If direct electronic follow-up is not practiced, telephone monitoring should be begun initially after implantation. The accuracy of telephone monitoring in detection of battery depletion is unmatched by that of direct electronic monitoring because of the ease and frequency of the transmissions [4]. The single area of failure has been the inability to detect impending lead fracture, which is, unfortunately, undetectable by most means. The purposes of clinic monitoring by electronic wave-form analysis are: 1. Evaluation of return of symptoms despite apparently normal pacing. This problem may require maximum operator and electronic skill to detect intermittent malfunction. 2. Adjustment and control of generators capable of variability in impulse duration (Medtronic and 5961 *). 3. Adjustment and optimum control of noninvasive programmable pacers (Cordis Omnicort). 4. Initial observation of pulse generators before telephone monitoring is required. 5. Evaluation of pulse generators suspected of having electronic defects that are not manifested by a change in rate or a clear-cut failure to pace or sense spontaneous cardiac activity. Electronic clinic monitoring continued on a monthly basis for prolonged periods is burdensome for staff and patient and should be replaced by telephone monitoring. Telephone monitoring alone, with rate determination and an electrocardiogram, can be used at frequent intervals following pacer implantation to ascertain electrode stability, then at longer intervals, and as the unit ages, at ever shorter intervals. The purposes of telephone monitoring are: 1. Careful pacemaker follow-up to achieve maximum safety and maximum longevity. 2. Careful and frequent observation of pacemakers in a series known to have a high incidence of electronic failure or of a unit that is no longer operating normally but for which clear-cut malfunction has not been demonstrated. *Medtronic, Inc., Minneapolis, Minn. Wordis Corp., Miami, Fla. VOL. 20, NO. 3, SEPTEMBER,

5 FURMAN AND ESCHER 3. Remote follow-up for patients who would otherwise receive little or no aftercare directed at the pacemaker. The capacity of the system to be used as a laboratory resource in which the primary physician receives adequate data about the patient to make a judgment, but does not relinquish responsibility for patient care, is especially important in the United States and has encouraged extension of pacemaker follow-up to several thousand patients from whom it otherwise might have been withheld [ The extension of monitoring to remote areas where travel is so great an impediment that follow-up would not be done. TRA NSTELEPHON IC MONITORING Rate Only. Rate-only monitoring was the initial approach. It is suitable for any variety of pacemaker and requires only that the unit have a magnetic switch to convert to asynchronous operation or that pacing be frequent enough that intervals between pacemaker stimuli can be determined. Battery failure is detected when the first cell is exhausted and pacer rate declines sharply [6, 131. Sensing function is difficult to determine except when tracking pulses can be used to distinguish between paced and sensed beats.* This method is still useful and is frequently employed. Rate and Digital Plethysmography. This early approach was adapted from the rate-only technique because of the occasional need to determine whether ventricular capture has occurred. The lack of positive evidence of capture is the major defect of the rate-only technique and is only partially solved by rate and digital plethysmography, as the incidence of false-positive and false-negative results may be substantial, though statistical data have never been made available [6, 121. Rate and Electrocardiogram. This provides rate determination and an ECG, usually but not necessarily lead I, to ascertain the adequacy of capture and pacer sensing. The ECG is superior to digital plethysmography since it is the universally recognized indicator of cardiac activity, and it is preferred when remote follow-up starts shortly after pacer implantation because failure to capture or sense ventricular activity is readily recognized. If follow-up is begun late, perhaps 1 to 2 years after pacer implantation, the ECG is less necessary since cessation of ventricular capture rarely occurs in the presence of a normal pacemaker rate and presumed normal output. Two general varieties of rate and ECG monitoring exist. Both have a carrier frequency modulated by the ECG, producing a wavering tone acoustically coupled to the telephone. In one,? a highly accurate ECG results from an 1,800 Hz carrier signal that transmits the pacer spike and ECG to the receiver. Both are demodulated and converted to a signal capable of operating an ECG and an interval timer that maintains both artifact amplitude and polarity. The other4 is a carrier-interrupt system with a 2,000 cycle carrier signal interrupted by the pacemaker impulse. The impulse is not transmitted, and a gap is left in the carrier signal where the pacer stimulus had occurred. An artifact is then added to the ECG at the receiver that distorts the low-frequency spectrum of the ECG. The *Starr-Edwards Cardiac Pacemaker System, Edwards Laboratories, Santa Ana, Calif. tmedalert Gorp., New York, N.Y. Spacer-Check, ESBIMedcor, Yardley, Pa. 330 THE ANNALS OF THORACIC SURGERY

6 Transtelephone Pacemaker Monitoring pacemaker spike as seen on the final ECG is a total artifact and bears only a time relation to the stimulus generated by the pacemaker. Neither amplitude nor polarity can be determined. The carrier-interrupt system is easier for the patient to work and allows an unskilled operator to derive useful data. The more discriminating system provides a superior ECG but requires an operator of greater skill and is most readily performed in a central laboratory facility. In each instance the transmission is performed from the patient s home. Rate and ECG ahd Electronic Analysis. Potentially the most sophisticated system for remote information transmission, it requires that the patient visit a peripheral facility with a relatively elaborate transmitter from which data are transmitted to a central station for analysis.* The need for the patient to travel to the transmitter makes the technique similar to a conventional electronic clinic operation, removing the ease of home transmission. A device of similar capability to transmit from the patient s home is under development [3]. ELECTROCARDIOGRAPHY IN TRANSTELEPHONE MONITORING The two varieties of ECG provide interpretation and artifacts which are similar to conventional electrocardiography, and both clearly indicate failure to capture, sense, or both. A few artifacts are unique to transtelephone monitoring. The most serious artifact is that caused by the use of a single-lead rhythm strip, usually lead I, in which the QRS or the pacemaker artifact may be isoelectric or for some reason insufficiently diagnostic. Another lead should be selected and transmission repeated. The diagnostic lead can be used thereafter. Telephone noise is frequently picked up from continuous interference and electrical transients. Usually both are detected readily. Patient reversal of the limb leads is readily detected and is of little consequence. Motion artifacts occur frequently. There can be troublesome 60 Hz interference with all systems. The stimulus rate and the ECG can usually be discriminated. Because the magnetic determination of rate requires a magnet to be held over the pacemaker, its movement (as in a patient with Parkinson s disease) can cause a variety of artifacts, including cessation of pacing and inhibition of the pacer [51. Respiratory movement is recognized, as on a conventional ECG, by a slow, rhythmic oscillation of the baseline. The degradation of the ECG signal on the carrier-interrupt monitor may cause difficulty with interpretation of the QRS complex. A wide variety of changes may be detected with ECG monitoring. These include the range of pacemaker malfunction: rate change and loss of pacing or sensing or both (Fig. 3). Relatively common intercurrent findings are multifocal premature ventricular contractions, tachycardia, atrial fibrillation, and other arrhythmias (Fig. 4). In more than 99.9% of transmissions the induction of a competitive, magnet-produced rhythm to detect pacemaker rate is without meaningful effect. Pacemaker-produced premature ventricular contractions occur but are not sustained, and only very rarely do multiple PVCs require cessation of use of the *Pacemaker Diagnostic Clinic of America, Gainesville, Fla. VOL. 20, NO. 3, SEPTEMBER,

7 FURMAN AND ESCHER FIG. 3, Normal pacing with intermittent unsensed complexes (carrier-interrupt transmission). magnet (see Fig. 4). No episode of ventricular fibrillation or sustained tachycardia, competitively induced, has occurred. Monitoring Technique The schedule of follow-up visits or telephone calls should be based on the best available knowledge of the characteristics of pacemaker operation and should be adhered to as closely as possible. Careful records will provide quick diagnostic FIG. 4. In addition to findings related to pacemaker function, a variety of other events can be documented: (A) Competitive pemature ventricular contractions and an asynchronous pacemaker. (B) Competition between a noncompetitive pacemaker in the magnetic asynchronous mode and spontaneous contractions poduces short, self-limited runs of ventricular tachycardia (the only such instance in this series). (C) Premature ventricular contractions, compensatory pause, and a pseudofusion beat. (0) Spontaneous multijocal ventricular tachycardia, pacemaker normal. 332 THE ANNALS OF THORACIC SURGERY

8 Transtelephone Pacemaker Monitm'ng information when failure occurs and must be readily available for comparison with earlier data (Fig. 5). Over half of the patients, even in large metropolitan areas with good medical facilities, do not receive adequate pacemaker follow-up [ 13. Pacemaker systems tend to fail during three distinct periods after implantation and should be monitored correspondingly. At present the schedule for telephone calls is every 2 months for the first 6 months after implantation and at shorter intervals until the eighteenth or twentyfourth month (depending on pacemaker model), when weekly calls are begun. Weekly calls are started earlier if the behavior of a specific pacemaker model or individual unit is suspicious. Early Postimplant (months I -3). Electrode malfunction occurs during the first 2 weeks and should be detected during postimplant observation. At Montefiore Hospital and Medical Center, 3.4% of transvenous pacemaker implants require revision during the early period following implantation. Surgical problems of wound healing cannot be detected remotely, and conventional observation by the surgeon is required. Random electronic failure occurs infrequently. Failure to sense or pace properly happens often during this period in the general population of pacemaker implants. A follow-up service* accepting patients from a variety of physicians reports that 10% of all pacemaker system failures occur during the first 3 months after implantation, representing problems of electrode placement. Middle Period (4-24 months). Three varieties of failure occur, though all are relatively uncommon with a modern, well-designed pacemaker system: (1) Battery exhaustion, which should not occur in more than 5% of modern pulse generators before the twenty-fourth month - some models appear to have reached that goal; (2) Lead fracture (4% in 1974); and (3) Electronic failure. The method of follow-up should be matched to experience with the specific pulse generator. Late Period (beyond24 months). Battery exhaustion is the typical failure mode. Lead fracture continues at a constant rate (4% per year). Electronic failures increase as pulse generator components drift out of specification despite satisfactory battery function. Results Our follow-up program has made only 2 errors. One occurred with rate-only monitoring when ventricular capture was erroneously assumed to be present because the pacemaker rate was constant. The other happened with rate and ECG monitoring in which battery voltage was down: the rate was stable and pacing was consistent during each transmission, but pacing had become intermittent and syncope had returned. The error rate of transtelephone monitoring is thus about 0.7% of the failure determinations made on patients followed to generator removal, 2 patients of 280 (Table). *Medalert Corp., New York, N.Y. VOL. 20, NO. 3, SEPTEMBER,

9 FURMAN AND ESCHER hhjz-4.- I c , - I ; A FIG. 5. Record keeping is as important as data acquisition. (A) Face ofjle card used for telephone transmissions. Of the 6 19 patients followed since 1969,280 have had failure detected before cessation of pacing and return of symptoms. Two hundred two patients are now actively followed. The average number of calls has been 35 per patient, 1 being the least and 132 the most for a pacemaker suspected of premature battery failure but which lasted 56 months. The average longevity of pacemakers removed was 34 months. Ten percent of exhausted pacemakers failed prior to the average life for the particular model, and 32% (89 of 280) exceeded 36 months' longevity. Of 334 THE ANNALS OF THORACIC SURGERY

10 Transtelephone Pacemaker Monitoring FIG. 5. (Continued.) (B) Back of file card. Rate stability is clearly visible, as is a rate change during the twenty-jifth month, the significance of which was unknown. End of pacemaker lfe was heralded by a beginning decline and then a sharp decline of 6.8 beats per minute in one week. The pulse generator was replaced during the thirty-jifth month of operation. (Retouched to hide patient s identity.) these, 13% (37 of 280) lasted longer than 40 months and 4.6% (13 of 280) exceeded 50 months. One pulse generator was removed 63 months after implantation. These results have been derived from a carefully controlled hospital-based patient population. Pacemaker follow-up in the general population is of equal interest. Data provided by a non-hospital-based follow-up system* that accepts patients without any preselection provides an informative comparison. During the period March, 1972, to September, 1974, a total of 606 patients were followed for some period, with an average enrollment of 182. Fifty-four pacemakers were electively replaced despite follow-up. Failure or impending failure was detected in 104 pacemaker systems. This was 57% of the average number followed, and 10 failures were discovered at the very first (enrollment) transmission, so that function during the preceding period of no pacing is unknown. Forty-three (41%) of the failures were due to battery exhaustion, 3 (3%) to electrode displacement, 6 (6%) to wire breaks, 8 (8%) to sensing failures, 13 *Medalert Gorp., New York, N.Y VOL. 20, NO. 3, SEPTEMBER,

11 FURMAN AND ESCHER RESULTS OF TRANSTELEPHONE MONITORING IN 619 PATIENTS Result Failure detected Errors Pacer electronic failure (undetected) Lead fracture - pacer normal Patient died - pacer normal Elective cessation of monitoring Patients actively followed No. of Patients (12.5%) to loss of ventricular capture with pacemaker rate normal, 13'(12.5%) to rate increases of more than 10 beats per minute in pulse generators that should have decreased in rate with battery depletion, and a variety to sudden cessation of pacer function and other failures, the nature of which was never determined. The failure rate requiring surgical intervention was 23% per year of pacemaker operation.* References 1. Bilitch, M., Cassady, E. E., and Lloyd, J. S. Physician Follow-up of Patients with Permanent Cardiac Pacemakers. In H. J. T. Thalen (Ed), Cardiac Pacing: Proceedings of the 4th International Symposium. Assen, Netherlands: van Gorcum, P Escher, D. J. W., and Furman, S. Oscilloscopic and recent other methods of implantable pacemaker follow-up. Ann Cardiol Angiol 20:503, Escher, D. J. W., Hochberg, H., King, E., Chudoba, P., and Kaufman, P. Pacemaker electronic follow-up: Complete unique system. Circulation 49,50 (Suppl 111): 94, Furman, S. Transtelephone observation of implanted cardiac pacemakers. Med Znstrum 7:196, Furman, S., Escher, D. J. W., and Parker, B. Failure of triggered pacemakers. Am Heart J 82:28, Furman, S., Escher, D. J. W., and Parker, B. The pacemaker follow-up clinic. Prog Cardiovasc DW 14:515, Furman, S., Parker, B., and Escher, D. J. W. Transtelephone pacemaker clinic.] Thorac Cardiovarc Surg 61:827, Morse, D. P., Tesler, U. F., and Lemole, G. M. The actual lifespan of pacemakers. Chest 64:454, Parsonnet, V., Furman, S., and Smyth, N. Implantable cardiac pacemakers: Status report and resource guideline. (Pacemaker study group.) Circulation 50:A21, Parsonnet, V., Gilbert, L., and Zucker, I. R. Natural history of pacemaker wires. J Thorac Cardiovasc Surg 65:315, Parsonnet, V., Meyers, G. H., Gilbert, L., and Zucker, I. R. Prediction of impending pacemaker failure in a pacemaker clinic. Am J Cardiol 25:311, Pennock, R. S., Dreifus, L. S., Morse, D. P., and Watanabe, Y. Cardiac pacemaker function. JAMA 222:1379, Ruben, S. Sealed zinc-mercuric oxide cells for implantable cardiac pacemakers. Ann NY Acad Sci 167:627, *A. Auerbach, personal communication. 336 THE ANNALS OF THORACIC SURGERY

12 Transtelephone Pacemaker Monitoring 14. Van Den Berg, J. W., Rodrigo, F. A., Thalen, H. J. T., and Koops, J, Photoanalysis of the condition of implanted pacemaker electrode circuits. Proc Kon Ned Akad Wet (Biol Med) 70:419, Discussion DR. DAVID C. MACGREGOR (Toronto, Ont., Canada): I would like to congratulate Dr. Furman on his pioneer work in transtelephone monitoring and to thank him for his assistance in establishing our own follow-up program at the Toronto General Hospital. We have recently reviewed our experience with 224 pulse generators replaced over the past 28 months on the basis of telephone evaluation alone. ECG and rate determination were used in every case, and only units which had been in place for 18 months or more were considered. Our results differ considerably from those of Dr. Furman. Seventy-eight (35%) of the pulse generators were replaced electively, 99 (44%) were predicted failures on the basis of rate change alone, and 47 (21%) were unpredicted failures in that a rate change was not observed prior to loss of capture or failure of sensing function or both. The mean longevity of all pulse generators was 29.7 months. Ofthe 47 unpredicted pulse generator failures, 25, or 11.2% ofthe total group, lost the ability to sense or pace, or both, without an accompanyingchange in rate, and as such would not be detected by rate-only monitoring. Four failures resulted from unpredictable electrode fractures. Eight patients had clinical symptoms, but there were no deaths between the recognition of pulse generator failure and the time of replacement. Although it is often stated that, at present, the stimulation rate of virtually all pacemakers varies as battery exhaustion occurs, numerous patients still have pacemakers that do not incorporate this feature. For this reason we recommend strict adherence to manufacturers replacement recommendations. We also believe that transtelephone monitoring should include an electrocardiogram to identify premature battery exhaustion in this group of patients as well as to detect random component failures, which continue to occur in all types of pacemakers. Furthermore, the electrocardiogram may identify pacemaker-related or other underlying cardiac arrhythmias that often require treatment. DR. ALBERT STARR (Portland, Ore.): We certainly owe a great deal to Dr. Furman for his pioneering work with pacemakers, especially in pointing out the need for an organized program of follow-up and also in demonstrating that the telephone system has certain advantages. We regard this kind of follow-up system as a supplement to standard cardiological care, but not as a substitute for it, and that is the important concept. Our own transtelephone system began its clinical use in 1970, and we are currently following 350 patients. The system has been made possible by the development of a proved, truly rate stable pulse generator of a ventricular tracking type. Rate is thus an exquisite indicator of battery voltage, and hence rate alone provides adequate pacemaker analysis. Competitive pacing is avoided during telephone transmission, since at both the high and low rates the pulse generator is always in the demand mode. The tracking signal in this pacemaker facilitates phone monitoring by marking the patient s own QRS complex; thus it can be considered a kind of simplified digitized electrocardiogram. We have organized this so that the clinician receives a pacergram as a laboratory service, not as a professional service. This reduces the cost enormously and avoids the practice of cardiology on the telephone. DR. FURMAN: I would like to thank the discussants for their very kind and searching comments on the technology involved. I must disagree with Dr. Starr concerning the electrocardiogram, which is the universally recognized means of determining cardiac action and should be used when at all possible. Many of the intercurrent findings we have made that were not related to pacer function were detected because of the electrocardiogram, and neither rate-only monitoring nor tracking would have yielded that information. VOL. 90, NO. 3, SEPTEMBER,

13 FURMAN AND ESCHER My second point is that we probably will see a noticeable decline in the loss of cardiac capture without a significant rate change, as that was representative of only one major manufacturer and only one major model. All the units now in use seem to have a rate change or some other clear-cut indication of battery depletion. In any event, even in that rate-stable model we were able to determine that a very small rate change was the indicator of battery depletion, and except for 1 patient we had very little trouble even in that single model. NOTICE FROM THE AMERICAN BOARD OF THORACIC SURGERY A limited number of copies of the 1974 Self-Assessment Examination for Thoracic Surgery (SATSE), together with an answer sheet, are available for $25 per copy from the office of the American Board of Thoracic Surgery, E. Seven Mile Rd., Detroit, Mich THE ANNALS OF THORACIC SURGERY

Electronic Analysis for Pacemaker Failure

Electronic Analysis for Pacemaker Failure Electronic Analysis for Pacemaker Failure Seymour Furman, M.D., Doris J. W. Escher, M.D., Bryan Parker, and Norman Solomon, M.D. T he physician or surgeon who implants a cardiac pacemaker incurs a continuing

More information

Critical Analysis of Pulse Generator Replacement

Critical Analysis of Pulse Generator Replacement Critical Analysis of Pulse Generator Replacement B. S. Goldman, M.D., J. G. Heller, Ph.D., E. J. Noble, M.D., H. D. Covvey, M.Sc., D. C. MacGregor, M.D., and J. D. Morrow, M.D. ABSTRACT A review of 633

More information

A Mechanism for "False" Inhibition of Demand Pacemakers

A Mechanism for False Inhibition of Demand Pacemakers A Mechanism for "False" Inhibition of Demand Pacemakers By KENNETH C. LASSETER, M.D., JACK W. BUCHANAN, JR., M.S.E.E., AND KARL F. YOSHONIS, M.D. SUMMARY Certain variations in discharge rate of demand

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission

Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission Takeo SAKURAI, M.D., Shuichi TOKUTSU, M.D., Osamu NISHIMUR M.D., Yutaka TAGAMI, M.D., Motoyuki TAKIMOTO,

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

The Influence of Electromagnetic

The Influence of Electromagnetic The nfluence of Electromagnetic Environment on the Performance of Artificial Cardiac Pacemakers Seymour Furman, M.D., Bryan Parker, Martin Krauthamer, M.D., and Doris J. W. Escher, M.D. T he danger of

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Low-Cost Daily Pacemaker Monitoring System

Low-Cost Daily Pacemaker Monitoring System LowCost Daily Pacemaker Monitoring System George Benzing, 111, M.D., Rumult Iltis, M.S.E.E., Samuel Kaplan, M.D., J. Tracy Schreiber, M.D., and James Helmsworth, M.D. ABSTRACT To reduce the cost of pacemaker

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Pacemaker Concepts and Terminology*

Pacemaker Concepts and Terminology* Pacemaker Concepts and Terminology* BAROUH V BERKOVITS, E.E., Ing. Associate in Surgery, Harvard Medical School, Boston, Mass., Associate in Electrophysiology, Miami University, Miami, Fla., Senior Research

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.

Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing. Friedman, Rott, Wokhlu, Asirvatham, Hayes 201 Figure.7 Shortening of the AV interval during pacing. 202 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization Figure.8 is obtained from

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement 217 Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement Partha Prateem Choudhury, Vivek Chaturvedi, Saibal Mukhopadhyay, Jamal Yusuf Department

More information

BEDSIDE ECG INTERPRETATION

BEDSIDE ECG INTERPRETATION BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes

More information

Different indications for pacemaker implantation are the following:

Different indications for pacemaker implantation are the following: Patient Resources: ICD/Pacemaker Overview ICD/Pacemaker Overview What is a pacemaker? A pacemaker is a device that uses low energy electrical pulses to prompt the heart to beat whenever a pause in the

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017 1. 2017 BDKA Review 2. 3. 4. Interpretation 5. QT 6. 7. 8. 9. 10. QT 11. 12. 13. 14. 15. 16. 17. 18. QT 19. 20. QT 21. 22. QT 23. 24. Where are pacer spikes? Before the P wave or before the QRS complex?

More information

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Basic Dysrhythmia Interpretation

Basic Dysrhythmia Interpretation Basic Dysrhythmia Interpretation Objectives 2 To understand the Basic ECG To understand the meaning of Dysrhythmia To describe the normal heart conduction system. To describe the normal impulse pathways.

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

This leaflet is intended for patients with permanent pacemakers and their parents.

This leaflet is intended for patients with permanent pacemakers and their parents. Permanent PACEMAKER This leaflet is intended for patients with permanent pacemakers and their parents. It serves as an information tool and a guide that offers some advice in order to live safely. General

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 5: Rhythm Strip Interpretation and Sinus Rhythms Learning Outcomes 5.1 Explain the process of evaluating ECG tracings

More information

Panorama. Arrhythmia Analysis Frequently Asked Questions

Panorama. Arrhythmia Analysis Frequently Asked Questions Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Chapter 9: Ventricular Dysrhythmias 2012 The Companies, Inc. All rights reserved. Learning Outcomes 9.1 Describe the various ventricular dysrhythmias 9.2

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV. Contents I. Introduction II. Website and Relevant Documents III. Background Information IV. Pacemaker System V. i. Components ii. Operating Modes iii. Parameters Example: Pacemaker in DDD Mode Pacemaker

More information

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8)

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8) Page 1 of 12 Centers for Medicare & Medicaid Services National Coverage Determination (NCD) for Cardiac Pacemakers (20.8) Tracking Information Publication Number 100-3 Manual Section Number 20.8 Manual

More information

Radioelectrocardiography in Diagnosis of Rhythm

Radioelectrocardiography in Diagnosis of Rhythm Brit. Heart J., 1969, 31. 480. Radioelectrocardiography in Diagnosis of Rhythm Disturbances in Patients with Implanted Cardiac Pacemakers MARIUSZ JAN STOPCZYK, ZYTA KRASZEWSKA, AND RYSZARD JACEK ZOCHOWSKI

More information

Peri-operative management of pacemakers and implantable cardiac defibrillators

Peri-operative management of pacemakers and implantable cardiac defibrillators Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),

More information

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms JUNCTIONAL RHYTHMS The AV Junction (Bundle of His and surrounding cells) only acts as pacemaker of the heart when the SA Node is not firing normally

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

UNDERSTANDING YOUR ECG: A REVIEW

UNDERSTANDING YOUR ECG: A REVIEW UNDERSTANDING YOUR ECG: A REVIEW Health professionals use the electrocardiograph (ECG) rhythm strip to systematically analyse the cardiac rhythm. Before the systematic process of ECG analysis is described

More information

TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT

TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT Link download full: http://testbankair.com/download/test-bank-for-ecgs-made-easy-5thedition-by-aehlert/ TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT Chapter 5 TRUE/FALSE 1. The AV junction consists

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Management of RT patients with implanted cardiac devices: From recommendation to implementation Management of RT patients with implanted cardiac devices: From recommendation to implementation Coen Hurkmans Clinical Physicist Catharina Hospital Eindhoven The Netherlands 1/22 Outline CIED relocation.

More information

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece

Leadless pacing. P.E. Vardas Professor of Cardiology University of Crete, Greece Leadless pacing P.E. Vardas Professor of Cardiology University of Crete, Greece DISCLOSURES Small teaching fees from: Bayer/J&J, Boeringer, Menarini Small consulting fees from: Medtronic 1957: First External

More information

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent? CASE 10 A 57-year-old man presents to the emergency center with complaints of chest pain with radiation to the left arm and jaw. He reports feeling anxious, diaphoretic, and short of breath. His past history

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

The protection you need - without touching your heart

The protection you need - without touching your heart The S ICD System The protection you need - without touching your heart listen to your heart Patient information You have options An implantable defibrillator, commonly known as an ICD, is a device designed

More information

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D.

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. NOTES An Implantable Demand Pacemaker Lawrence I. Zaroff, M.D.,* Barouh V. Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. E lectrical stimulation in the treatment of heart

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital

Leadless pacing: Technical challenges and concerns. Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Leadless pacing: Technical challenges and concerns Panos Vardas President Elect of the ESC, Prof. of Cardiology, Heraklion University Hospital Modest consultancy fees/honoraria from Bayer Boehringer Ingelheim

More information

OHTAC Recommendation: Internet- Based Device-Assisted Remote Monitoring of Cardiovascular Implantable Electronic Devices

OHTAC Recommendation: Internet- Based Device-Assisted Remote Monitoring of Cardiovascular Implantable Electronic Devices OHTAC Recommendation: Internet- Based Device-Assisted Remote Monitoring of Cardiovascular Implantable Electronic Devices Ontario Health Technology Advisory Committee January 2012 Issue Background The Ontario

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network Pacing troubleshooting NASPE training Lancashire & South Cumbria Cardiac Network Pacing stimulus present failure to capture Lead dislodgement Early unstable position Late twiddlers syndrome Abnormal ECG

More information

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Programming of Bradycardic Parameters C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Programming of ICD Brady Parameters Conflict of Interest Biotronik

More information

Ambulatory ECG Monitoring. Gill Bouncir Blackpool Victoria Hospital

Ambulatory ECG Monitoring. Gill Bouncir Blackpool Victoria Hospital Ambulatory ECG Monitoring Gill Bouncir Blackpool Victoria Hospital History Developed in late 1950 s as a result of research of Dr Norman Holter TYPES Holter - continuous recording. Event recorder- patient

More information

MARS Ambulatory ECG Analysis The power to assess and predict

MARS Ambulatory ECG Analysis The power to assess and predict GE Healthcare MARS Ambulatory ECG Analysis The power to assess and predict Connecting hearts and minds Prevention starts with knowledge Around the world, heart disease is one of our fastest-growing health

More information

MEDICAL POLICY Cardiac Event Monitors/ Cardiac Event Detection

MEDICAL POLICY Cardiac Event Monitors/ Cardiac Event Detection POLICY: PG0039 ORIGINAL EFFECTIVE: 10/01/11 LAST REVIEW: 12/12/17 MEDICAL POLICY Cardiac Event Monitors/ Cardiac Event Detection GUIDELINES This policy does not certify benefits or authorization of benefits,

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

EVALUATION OF CONVENTIONAL AND COMPUTERISED MONITORING OF CARDIAC FUNCTION IN THE CORONARY CARE UNIT

EVALUATION OF CONVENTIONAL AND COMPUTERISED MONITORING OF CARDIAC FUNCTION IN THE CORONARY CARE UNIT Med. J. Malaysia Vol. 38 No. 4 December 1983 EVALUATION OF CONVENTIONAL AND COMPUTERISED MONITORING OF CARDIAC FUNCTION IN THE CORONARY CARE UNIT NG KW AN HOONG SUMMARY One ofthe important functions ofthe

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Epicardial Pacemaker Implantation for Complete Heart Block

Epicardial Pacemaker Implantation for Complete Heart Block for Complete Heart Block Donald G. Mulder, M.D., and C. Gordon Frank, M.D. S udden death or progressive cardiac failure threatens life in the patient with complete heart block. The slow heart rate and

More information

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology)

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) ECG Interpretation Cat Williams, DVM DACVIM (Cardiology) Providing the best quality care and service for the patient, the client, and the referring veterinarian. GOAL: Reduce Anxiety about ECGs Back to

More information

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand: Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical

More information

2009 CPT Codes for Cardiac Device Monitoring

2009 CPT Codes for Cardiac Device Monitoring 2009 CPT Codes for Cardiac Device Monitoring December 2008 Notices Current Procedural Terminology (CPT ) is copyright 2008 American Medical Association. All Rights reserved. No fee schedules, basic units,

More information

CURRICULUM GOALS AND OBJECTIVES CLINICAL CARDIOVASCULAR ELECTROPHYSIOLOGY TRAINING PROGRAM. University of Florida Gainesville, Florida

CURRICULUM GOALS AND OBJECTIVES CLINICAL CARDIOVASCULAR ELECTROPHYSIOLOGY TRAINING PROGRAM. University of Florida Gainesville, Florida CURRICULUM GOALS AND OBJECTIVES CLINICAL CARDIOVASCULAR ELECTROPHYSIOLOGY TRAINING PROGRAM University of Florida Gainesville, Florida 1. Mission Statement To achieve excellence in the training of fourth

More information

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Faruk Cingoz, MD, George Yanulis, MS, Elizabeth Ching, RN, Kiyotaka Fukamachi,

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

Intermittent episodes of paced tachycardia: what is the cause?

Intermittent episodes of paced tachycardia: what is the cause? Intermittent episodes of paced tachycardia: what is the cause? Lindsey Parkinson and Lucy Broadhurst, Rotherham NHS Foundation Trust Contact: Lindsey.Parkinson@rothgen.nhs.uk Introduction A hospital in-patient

More information

Cardiac Arrhythmias in Sleep

Cardiac Arrhythmias in Sleep Cardiac Arrhythmias in Sleep Only 53 Slides! Gauresh H Kashyap, MD, FACP, FCCP, FAASM 1 2 Cardiac Arrhythmias in Sleep Out of 400 Patients with OSA, 48% had some Arrhythmias 20% - 2 PVCs/min 7% - Bradycardia

More information

ECG MACHINE KEC- Series

ECG MACHINE KEC- Series KIZLON ECG MACHINE KEC- Series 404 Crescent Royale, Off New Link Road Andheri West, Mumbai 40053, Phones: 022-6708747 Email: info@kizlon.com Website: www.kizlon.com ECG Machine KEC-A100 This ECG machine

More information

Changing Indications for Pacemakers in Children

Changing Indications for Pacemakers in Children Changing Indications for Pacemakers in Children William H. Fleming, M.D., Lynne B. Sarafian, M.N., John D. Kugler, M.D., Philip J. Hofschire, M.D., and Edward B. Clark, M.D. ABSTRACT Permanent pacemakers

More information

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution Francesco Zanon, MD, FESC, FHRS Arrhythmia and Electrophysiology Unit, Cardiology Dept. Santa Maria della Misericordia General

More information

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs). Management of Children with Cardiac Devices Guideline originally developed by Leann Miles, APRN; Lindsey Pumphrey, RN; Srikant Das, MD, and the ANGELS Team. Last reviewed by Lindsey Pumphrey, RN, Srikant

More information

To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device.

To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device. 043 Lauren Evans To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device. 4061.40 Summary With an ageing UK population, the number of patients with

More information

Waterproof quality IP X4

Waterproof quality IP X4 Actual size Thanks to its high level of miniaturisation, the R.Test Evolution 3 is compact and lightweight (45 g. batteries included), for total discretion. Its exceptional design ensures high patient

More information

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Need to Know: Implantable Devices Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia Disclosure Statement I have no relationships to disclose. Objectives Discuss the most

More information

Patient Resources: Arrhythmias and Congenital Heart Disease

Patient Resources: Arrhythmias and Congenital Heart Disease Patient Resources: Arrhythmias and Congenital Heart Disease Overview Arrhythmias (abnormal heart rhythms) can develop in patients with congenital heart disease (CHD) due to thickening/weakening of their

More information

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg Electrocardiography 31650 Biomedical Engineering Kaj-Åge Henneberg Electrocardiography Plan Function of cardiovascular system Electrical activation of the heart Recording the ECG Arrhythmia Heart Rate

More information

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children

Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children Concurrent Failure of Active and Redundant Ventricular Epicardial Electrodes in Children ERALD A. SERWER, M.D., MACDONALD DICK 11, M.D., KAREN UZARK, R.N., Ph.D., WILLIAM A. SCOTT, M.D., and EDWARD L.

More information

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2 Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division:

More information

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?

More information

The Function of an ECG in Diagnosing Heart Conditions. A useful guide to the function of the heart s electrical system for patients receiving an ECG

The Function of an ECG in Diagnosing Heart Conditions. A useful guide to the function of the heart s electrical system for patients receiving an ECG The Function of an ECG in Diagnosing Heart Conditions A useful guide to the function of the heart s electrical system for patients receiving an ECG Written by Erhan Selvi July 28, 2014 Audience and Scope

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

Ventricular Parasystole

Ventricular Parasystole Ventricular Parasystole 1 WHAT IS IT? In addition to the sinus node, there are many accessory pacemakers throughout the conducting system of the atria, junction and ventricles that are ready to assume

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters

More information

Chapter 16: Arrhythmias and Conduction Disturbances

Chapter 16: Arrhythmias and Conduction Disturbances Complete the following. Chapter 16: Arrhythmias and Conduction Disturbances 1. Cardiac arrhythmias result from abnormal impulse, abnormal impulse, or both mechanisms together. 2. is the ability of certain

More information