1 Electronic Analysis for Pacemaker Failure Seymour Furman, M.D., Doris J. W. Escher, M.D., Bryan Parker, and Norman Solomon, M.D. T he physician or surgeon who implants a cardiac pacemaker incurs a continuing responsibility for the patient s welfare. The prosthetic device may be erratic in its behavior, and because of the consumable batteries it undergoes progressive decay and eventual cessation of function. No accurate or reliable figures exist that will be a guide to removal of all pacemaker units before clinical failure. One approach to reduction of sudden failure is the removal of all functioning pacemakers when the average failure interval for 10% has been reached [2]. This approach cannot be recommended because of the unnecessary premature surgery and inefficiency of equipment usage and because of the 10% sudden failure rate, which is still allowed to occur. Removal at an average failure time is inadequate, because such an interval is composed of a wide scatter of figures about the average and would still allow a large number of sudden failures, while some units would be removed when they are still capable of significant function. To limit the number of sudden failures with possible disastrous consequences and to achieve maximum use of equipment, both the patient and his implanted prosthesis must be monitored frequently enough to reduce unpredicted failures to a minimum level. PRINCIPALS OF TESTING Just as a patient is periodically observed for proper cardiac function, so the pacemaker should be observed, during scheduled patient visits, for proper operation of its various functions. Analysis of a standard electrocardiogram, analysis of triggered function in such a unit, and careful measurement of the parameters of pacemaker output with an interval timer and by display of the pacemaker artifact on an oscilloscope screen allow determination of the state of function of the pacemaker circuit [l]. The important considerations are: (1) the battery From the Division of Surgery and Medicine, Montefiore Hospital and Medical Center, Bronx, N.Y. Supported in part by US. Public Health Service Grant No. HE Presented at the Fifth Annual Meeting of The Society of Thoracic Surgeons, San Diego, Calif., Jan , 1969.

2 FURMAN ET AL. voltage must be maintained; (2) the other electrical characteristics of the pacer output, slope of the pulse, pulse duration and shape, must be constant [5]; (3) the mode of pacemaker function must remain constant at all times, and the triggered function must remain appropriate; and (4) the pacemaker rate, or for a triggered pacemaker the rate produced by a magnetic switch or external stimulation, must be constant. EQUIPMENT A pacemaker clinic must include equipment for electrocardiography, for testing of pacemaker function, and for accurate recording and retrieval of data on each patient and pacemaker. OSCILLOSCOPE An oscilloscope appropriate for such use should have a storage screen with a differential vertical amplifier and a variable time base, with a range to encompass pacemaker impulses (0.5 msec. per centimeter) and ECG s (0.1 sec. per centimeter). Such a scope might be a Tektronix 564 with a 2A61 differential amplifier and a Type 2B67 time base. A Polaroid camera should be available for recording the pacemaker output wave form as it appears on the screen, so that the illustration can be used for future reference and for recording data on paper or for computer analysis. Unipolar electrodes provide large artifacts at the limbs (50 to 300 mv.) and are more readily interpreted than bipolar artifacts, which are smaller (1 to 30 mv.). Both can be measured and compared to previously determined standards, and studies can be performed on the same patients. INTERVAL TIMER The interval timer (such as the Hewlett-Packard Electronic Counter No. 5223L) should be capable of reading the interval between pacemaker impulses to within one tenth of a millisecond. The interval is then converted to a rate accurate to one one-hundredth of a beat per minute. A variation in asynchronous rate by one-half beat per minute during successive beat intervals confirms instability of circuit function. A general slowing or speeding of the rate is indicative of circuit changes of a gradual nature, whereas variation of interval from beat to beat is an indication of impending pacemaker failure. Almost all pacemakers have a fixed rate that is a function of the pacemaker circuit and can be measured. Many triggered units, such as the American Optical Co. Cardiocare and the Cordis Ectocor, have a fixed rate mode activated by a magnetic switch for measurement of rate. 58 THE ANNALS OF THORACIC SURGERY

3 Electronic Analysis for Pacemaker Failure ELECTROCARDIOGRAPHIC ANALYSIS Interpretation of the electrocardiogram will indicate whether pacemaker artifacts produce a ventricular response and whether the pacemaker is responding appropriately to spontaneous cardiac function when it exists. Each triggered pacemaker should be tested for appropriateness of its triggered function. This is readily accomplished by stimulating the implanted pacemaker through the intact skin with an external pacemaker. The ventricular synchronous pacemaker will follow the imposed rate without delay between the imposed pacemaker artifact and the implanted artifact [3]. The atrial synchronous pacemaker responds similarly, but with the A-V delay, which is part of the circuit (Fig. 1). The demand unit will be disabled by the external pacemaker artifacts and will not emit its own stimuli during external stimulation (Fig. 2). ELECTRONIC ANALYSIS OF PULSE GENERATOR IMPULSES Several parameters should be checked. Stimulus amplitude is the maximum amplitude achieved by an impulse, although a measurement at 0.5 msec. following the beginning of the impulse may be used. A I I t I FIG. 1. Ventricular synchronous unit triggered for test purposes. Left: pacing at the escape rate. Right: triggered pacing by transcutaneous stimulation of the pulse generator. The upward-directed line is the external impulse; the downward line is the implanted pacemaker impulse. VOL. 8, NO. 1, JULY,

4 FURMAN ET AL. M,Nrl!D #I( ".*.A FIG. 2. Demand pacing with application of a 5 ma. external impulse over the pacer.. The low-energy artifacts produce no cardiac response and suppress the implanted pacemaker impulses. decrease is caused either by a reduction of output current (usually because of reduction in battery voltage) or by a change in the dipole vector of the impulse. Respiratory movement yields more pronounced vector changes for bipolar than for unipolar electrodes, but large vector changes suggest either electrode break or displacement. Consistent measurement in a single lead, usually lead I for a unipolar electrode and lead I1 for a bipolar electrode, is adequate to yield reproducible results (Fig. 3). Stimulus shape is a valuable indicator of the pulse generator and electrode system. It is linearly related to the current flowing between PACE MAKER MFG : Cordis TYPE: 1298 ELECTRODE MFG: Cordis TYPE: Unipolar 4mm Final pacemaker current into 500 ohms = 6mA Final pacemaker battery voltage =4.7 volts rnsec Amplitude (at t=0.5msec) Pulse duration (msec) I.7 2.I Months after generator implant Rate (beats per minute) Plateau time (msec) to 65.6 (erratic) 0.7 FIG. 3. Two groups of data deriued six months apart. The unipolar impulse shows a decline of 40%, lengthening of the pulse duration, and slowing of the pacemaker rate with development of an erratic pulse-to-pulse interval. Pulse generator replacement was indicated. 60 THE ANNALS OF THORACIC SURGERY

5 Electronic Analysis for Pacemaker Failure PACEMAKER MFG: Cordis TYPE: Ventrlcor 1118 ELECTRODE MFG: Cord16 TYPE: Unipolar 4mm Final pacemaker current into 500 ohms -2.5 ma Flnal pacemaker battery voltage = 23 volts 0 -m M mlk I Amplitude (at 1=0.5msec) Pulse duration (msec) Months after generator implant Rate (beats per minute) Plateau time (msee) FIG. 4. Data derived at two-month intervals. The rate has declined by approximately two beats, and both the pulse duration and the plateau time have become proionged. Pulse generator replacement was indicated. the electrodes, and it changes with electrode defects and deterioration of the pulse generator output. Each pulse generator and its electrode has a characteristic impulse shape. Three main categories exist: voltage limited, current limited, and current and voltage limited. In the voltage-limited shape, current flow increases sharply to its maximum and immediately decreases continuously and exponentially throughout the impulse. The Cordis Atricor 102F, Medtronic, and American Optical Co. Cardiocare units have different varieties of this output shape. In the current-limited shape, current flow increases sharply and remains constant throughout the impulse, producing a relatively rectangular shape. The Cordis 1 11E and the Ector 129F are of this variety. In the current- and voltage-limited shape, the impulse shape initially resembles the current-limited output, with a short plateau followed by a voltage-limited exponential decay. All Cordis Ventricor, Atricor, and Ectocor models not listed above are of this variety. As this type of unit fails, the plateau is prolonged and the exponential decay shortens, tending toward a current-limited appearance [6] (Fig. 4). PR 0 CED URE Each patient is seen at three- to six-month intervals, and more frequently as the second anniversary after implantation is reached. During each visit the patient s cardiac status and pacemaker status are reviewed. A 12-lead electrocardiogram, a long rhythm record of lead 11, and the electronic analysis of the pacemaker output are performed. Each triggered pacemaker is tested by triggering to observe whether its response is constant. VOL. 8, NO. 1, JULY,

6 FURMAN ET AL. Based on our observations of pacemaker longevity, we have established a routine pulse generator replacement interval. This interval is the time beyond which a pulse generator is not allowed to remain, even if all its electronic parameters seem satisfactory. In 1967, only 12% of pulse generator replacements reached this period; in 1968, 14% did. The various types of units performed as follows: asynchronous, 30 months; demand, 30 months; atrial synchronous, 24 months; and ventricular synchronous, 24 months, although few units have yet reached this period. PULSE GENERATOR REPLACEMENT When all data are available, a pulse generator is replaced if any one of the following exists: (1) a rate change of 10% or more has occurred; (2) the electrocardiogram and triggered studies indicate that the pulse generator no longer functions in the mode for which it was designed; (3) the heart is no longer consistently stimulated; (4) the stimulus artifact has fallen by 40%; (5) stimulus duration has changed by 25% or has become grossly distorted; or (6) the routine replacement interval has passed [4]. RESULTS PACEMAKER LONGEVITY Accumulated data have demonstrated two major characteristics of pacemaker longevity: (1) 80% of asynchronous pulse generator failures occur over a one-year interval during the second year of pacemaker function; and (2) monthly analysis of pulse generator failure indicates a wide scatter, rather than a grouping (Fig. 5). These characteristics, therefore, do not allow a statistical prediction of individual pulse generator failure and do not allow the choice of a single period that will adequately avoid the bulk of failures while not substantially shortening the functional life of the pacemaker. CLINICAL EFFECTIVENESS The final tests of the adequacy of any follow-up technique will be based on the following characteristics: (1) as low as possible a death rate because of sudden pacemaker failure; (2) as low as possible an emergency pulse generator replacement rate because of unpredicted pacemaker failure; (3) maximum utilization of each pulse generator; (4) maximum number of pulse generator replacements done on a predicted basis while they approach exhaustion but are still functioning; and (5) as few pulse generators as possible removed while still capable of a significant duration of function. 62 THE ANNALS OF THORACIC SURGERY

7 Electronic Analysis for Pacemaker Failure CORDIS 127A Total sample - 33 Total calculated-22 Dead or pacemaker removed while functioning- 5 Remainina-6 Average -longevity months c E ; : 14 O n 10 E Number of months after implant E 10 : 12 L o 14 E E 20 L Number of failures FIG. 5. The behavior of a single model of cardiac pacemaker with units remaining operatiue (left) and individual failures (right) as a function of time. In this series, 44 replacements were done during 1967, a period in which pacemakers were closely observed but neither electronic analysis nor interval timing was used. During this period, 45% of all pulse generator replacements were emergencies and 39% were elective replacements. Sixteen percent were emergency pulse generator replacements done pari passu with repair of an electrode malfunction. In 1968, a transitional period, 89 replacements were done, and the technique described was used in the last seven months of the year. Pulse TABLE 1. PULSE GENERATOR REPLACEMENT Reason for Replacement " - Emergency 27 (61%) 25 (28%) Unpredicted pulse generator failure 20 (45%) 18 (20%) Cessation of pacing 12 (27%) 15 (17%) Gross change (continued pacing) 8 (18%) 3( 3%) Electrode failure 7 (16%) 7( 8%) Elective 17 (39%) 64 (72%) Predicted pulse generator failure 11 (25%) 47 (53%) Mode change 1 ( 2%) 6( 7%) Routine replacement 5 (12%) 11 (12%) Total "Percentages are of total replacements for the year. VOL. 8, NO. 1, JULY,

8 FURMAN ET AL. generator replacement for unpredicted failure was done in 20%; 8% of pulse generator replacements were done during repair of an electrode malfunction. Seventy-two percent of all replacements were elective (Table 1). Of these pulse generators, one that was replaced after 30 months showed significant residual life. One pacemaker was removed at 27 months on a false determination of battery exhaustion. All other pulse generators that were removed were confirmed in their actual or impending failure upon return to the manufacturers. One patient died of a clearly malfunctioning pacemaker, although she had been checked and the unit had been satisfactory one month earlier. SUMMARY Careful serial observations of implanted cardiac pacemakers are needed to reduce unforeseen pacemaker failures and their consequences. Analysis of triggered pacemaker function, constancy of pacemaker rate, and electronic analysis of pacemaker output allow accurate prediction of impending pacemaker failure or exhaustion before they are clinically manifest. REFERENCES 1. Knuckey, L., McDonald, R., and Sloman, G. A method of testing implanted cardiac pacemakers. Brit. Heart J. 27:483, Pacemaker Follow-up Reliability Report. Medtronic, Inc., Oct. 16, Smyth, N. P. D., Bacas, J. M., and Keller, J. W. Experimental and clinical use of a variable parameter cardiac pacer. Dis. Chest 53:93, Sowton, E. Detection of impending pacemaker failure. Israel J. Med. Sci. 3:260, Sprawls, P., Miller, W. B., and Logan, W. D. Observation of electromagnetic signals from implantable pacemakers. J. Thorac. Cardiouasc. Surg. 49:748, Van Den Berg, J. W., Rodrigo, F. A., Thalen, H. J., and Koops, J. Photoanalysis of the condition of implanted pacemakers and electrode circuits. Proc. Kon. Nederl. Akad. Wet. [Biol. Med.1 70:419, DISCUSSION DR. HAROLD STERN (New Haven, Conn.): I have had the privilege of reviewing Dr. Furman s excellent paper and was delighted to learn about his scientific approach to the latest plague of the chest surgeon, acute pacemaker failure. Analysis of the stimulus wave form and his very accurate measurement of stimulus intervals will serve to teach us about the failure characteristics of the devices we are using. His work, as you know, was done with the Cordis product, and I would like to see him carry it over to other products, perhaps in other institutions. For example, a rate change of 10% with the unit I use may be much more ominous than a rate change of 10% with a Cordis. Perhaps he can answer that. 64 THE ANNALS OF THORACIC SURGERY

9 Electronic Analysis for Pacemaker Failure What we really want to know is the time during which the pacemaker is safe, even if it means taking out some units prematurely. Electronic component failure should approach zero in this space age, when we are told that devices circling the globe are good for four, five, or six years. We have been changing our units routinely at 18 months whether they need it or not, and the hospitalization has been very brief. The patients are more confident and worry less about battery failure with this method. Our experience is much more limited than Dr. Furman s, but it seems to me that 25 emergency replacements in one year is excessive. What we need is a regular, detailed analysis by the manufacturers of these products which would include not only the longevity curve but also what happens when the pacemaker fails: slowing, speeding up, sudden standstill, battery failure, component failure. In that way we could learn more about the apparatus we use, and perhaps even change to a better product. DR. SAMUEL W. HUNTER (St. Paul, Minn.): It has been ten years now since we first put a pacemaker in a human being. I would like to remind the Society that there are approximately 15,000 new cases of total heart block per year in the United States. Moreover, pacemakers for arrhythmias other than total heart block have been developed to the point that if we multiply that 15,000 by 4 or 5, we would come closer to the annual number of pacemakers that will be used in this country within the next several years. The problem arises when you get a call at three o clock in the morning from a patient whose pacemaker has failed. Perhaps this is not going to happen after the sophisticated methods described by Dr. Furman have been introduced, but I am not so sure that it won t. Pacemakers fail for a multiplicity of reasons, such as the electrode eroding through the right ventricle, as just described by Dr. Peter Allen of Vancouver. I am exploring with Medtronic the idea of establishing a computerized record system. If such a system were established, we could dial into a number at the computer center and obtain the necessary information quickly and accurately. Some such system is necessary because of the increasing number of pacemaker patients. Not only do we need this information ourselves, but we must be prepared to send it quickly to surgeons at other centers. DR. FURMAN: Dr. Stern, this method of analysis is suitable for any of the pacemakers presently available. It is somewhat more difficult to get a satisfactory analysis with a bipolar unit because of the smaller size of the electrocardiographic wave form; nevertheless, it is perfectly possible to do. I completely agree that one should have immediate access to all the pacemaker data on any particular patient. It is incumbent upon us to follow our patients very carefully. Despite the best efforts of the manufacturers, and they are improving their efforts, it is really up to us, once we know what the pacemaker will do, to see the patients sufficiently frequently so we will pick up failures. Of course, it is up to the manufacturers to let us know what the pacemaker will do and, very honestly, when they can anticipate that it may fail. VOL. 8, NO. 1, JULY,

The Influence of Electromagnetic

The Influence of Electromagnetic The nfluence of Electromagnetic Environment on the Performance of Artificial Cardiac Pacemakers Seymour Furman, M.D., Bryan Parker, Martin Krauthamer, M.D., and Doris J. W. Escher, M.D. T he danger of

More information

A Mechanism for "False" Inhibition of Demand Pacemakers

A Mechanism for False Inhibition of Demand Pacemakers A Mechanism for "False" Inhibition of Demand Pacemakers By KENNETH C. LASSETER, M.D., JACK W. BUCHANAN, JR., M.S.E.E., AND KARL F. YOSHONIS, M.D. SUMMARY Certain variations in discharge rate of demand

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

Transtelephone Pacemaker Monitoring

Transtelephone Pacemaker Monitoring Transtelephone Pacemaker Monitoring Five Years Later Seymour Furman, M.D., and Doris J. W. Escher, M.D. ABSTRACT Six hundred nineteen patients have been followed by remote monitoring of pacemaker function

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

Radioelectrocardiography in Diagnosis of Rhythm

Radioelectrocardiography in Diagnosis of Rhythm Brit. Heart J., 1969, 31. 480. Radioelectrocardiography in Diagnosis of Rhythm Disturbances in Patients with Implanted Cardiac Pacemakers MARIUSZ JAN STOPCZYK, ZYTA KRASZEWSKA, AND RYSZARD JACEK ZOCHOWSKI

More information

Pacemaker Concepts and Terminology*

Pacemaker Concepts and Terminology* Pacemaker Concepts and Terminology* BAROUH V BERKOVITS, E.E., Ing. Associate in Surgery, Harvard Medical School, Boston, Mass., Associate in Electrophysiology, Miami University, Miami, Fla., Senior Research

More information

The Normal Electrocardiogram

The Normal Electrocardiogram C H A P T E R 1 1 The Normal Electrocardiogram When the cardiac impulse passes through the heart, electrical current also spreads from the heart into the adjacent tissues surrounding the heart. A small

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

Interpreting Electrocardiograms (ECG) Physiology Name: Per: Interpreting Electrocardiograms (ECG) Physiology Name: Per: Introduction The heart has its own system in place to create nerve impulses and does not actually require the brain to make it beat. This electrical

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

5- The normal electrocardiogram (ECG)

5- The normal electrocardiogram (ECG) 5- The (ECG) Introduction Electrocardiography is a process of recording electrical activities of heart muscle at skin surface. The electrical current spreads into the tissues surrounding the heart, a small

More information

Peri-operative management of pacemakers and implantable cardiac defibrillators

Peri-operative management of pacemakers and implantable cardiac defibrillators Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV. Contents I. Introduction II. Website and Relevant Documents III. Background Information IV. Pacemaker System V. i. Components ii. Operating Modes iii. Parameters Example: Pacemaker in DDD Mode Pacemaker

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Sheet 5 physiology Electrocardiography-

Sheet 5 physiology Electrocardiography- *questions asked by some students Sheet 5 physiology Electrocardiography- -why the ventricles lacking parasympathetic supply? if you cut both sympathetic and parasympathetic supply of the heart the heart

More information

Low-Cost Daily Pacemaker Monitoring System

Low-Cost Daily Pacemaker Monitoring System LowCost Daily Pacemaker Monitoring System George Benzing, 111, M.D., Rumult Iltis, M.S.E.E., Samuel Kaplan, M.D., J. Tracy Schreiber, M.D., and James Helmsworth, M.D. ABSTRACT To reduce the cost of pacemaker

More information

Interpretation of complex demand pacemaker arrhythmias

Interpretation of complex demand pacemaker arrhythmias British Heart Journal, I972, 34, 312-3I7. Interpretation of complex demand pacemaker arrhythmias S. Serge Barold, John J. Gaidula, Richard L. Banner, George I. Litman, and Sidney Goldstein From the Department

More information

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current

Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current Factors Determining Vulnerability to Ventricular Fibrillation Induced by 60-CPS Alternating Current By Tsuneoki Sugimoto, M.D., Stephen F. School, M.D., and Andrew G. Wallace, M.D. ABSTRACT Very weak,

More information

Diploma in Electrocardiography

Diploma in Electrocardiography The Society for Cardiological Science and Technology Diploma in Electrocardiography The Society makes this award to candidates who can demonstrate the ability to accurately record a resting 12-lead electrocardiogram

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs

Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Use of Conventional Dual Chamber Pacemakers With Custom Lead Adapters to Induce Atrial Fibrillation or Heart Failure in Dogs Faruk Cingoz, MD, George Yanulis, MS, Elizabeth Ching, RN, Kiyotaka Fukamachi,

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Critical Analysis of Pulse Generator Replacement

Critical Analysis of Pulse Generator Replacement Critical Analysis of Pulse Generator Replacement B. S. Goldman, M.D., J. G. Heller, Ph.D., E. J. Noble, M.D., H. D. Covvey, M.Sc., D. C. MacGregor, M.D., and J. D. Morrow, M.D. ABSTRACT A review of 633

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2)

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2) EKG Sign-in Book HTEC 91 Review of protocol Review of placement of chest leads (V1, V2) Medical Office Diagnostic Tests Week 2 http://www.cvphysiology.com/arrhythmias/a013c.htm The Cardiac Cycle Represents

More information

UNDERSTANDING YOUR ECG: A REVIEW

UNDERSTANDING YOUR ECG: A REVIEW UNDERSTANDING YOUR ECG: A REVIEW Health professionals use the electrocardiograph (ECG) rhythm strip to systematically analyse the cardiac rhythm. Before the systematic process of ECG analysis is described

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

EC6001 MEDICAL ELECTRONICS

EC6001 MEDICAL ELECTRONICS EC6001 MEDICAL ELECTRONICS UNIT I ELECTRO-PHYSIOLOGY AND BIO-POTENTIAL RECORDING The origin of Bio-potentials; bio-potential electrodes, biological amplifiers, ECG, EEG, EMG, PCG, lead systems and recording

More information

BASIC CONCEPT OF ECG

BASIC CONCEPT OF ECG BASIC CONCEPT OF ECG Electrocardiogram The electrocardiogram (ECG) is a recording of cardiac electrical activity. The electrical activity is readily detected by electrodes attached to the skin. After the

More information

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants EKG Pop Quiz Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants 1 Disclosures No financial relationships to disclose. EKG #1 75 y/o woman with a dual-chamber pacemaker

More information

V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7

V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7 V4, V5 and V6 follow the 5 th intercostal space and are NOT horizontal as indicated in the image. Page 1 of 7 "Is it correct to place the electrodes under the female breasts?" Placement of electrodes directly

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

This presentation will deal with the basics of ECG description as well as the physiological basics of

This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 1 Electrocardiography basics This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 2 Lecture overview 1. Cardiac conduction system functional

More information

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine

More information

BEDSIDE ECG INTERPRETATION

BEDSIDE ECG INTERPRETATION BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction

More information

also carried out with a portable high duty electric drill (Black and Decker), a short wave diathermy apparatus (Magatherm Junior 4), a 5 watt neon

also carried out with a portable high duty electric drill (Black and Decker), a short wave diathermy apparatus (Magatherm Junior 4), a 5 watt neon British Heart journal, I970, 32, 626-632. Electrical interference in non-competitive pacemakers E. Sowton, K. Gray, and T. Preston From The Institute of Cardiology, London W.I Patients with 4I implanted

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS BEFORE INTERPRETING THE ECG: Nora Goldschlager, M.D. MACP, FACC, FAHA, FHRS. Cardiology San Francisco General Hospital UCSF Disclosures: None 1 2 QUESTIONS

More information

Chapter 12: Cardiovascular Physiology System Overview

Chapter 12: Cardiovascular Physiology System Overview Chapter 12: Cardiovascular Physiology System Overview Components of the cardiovascular system: Heart Vascular system Blood Figure 12-1 Plasma includes water, ions, proteins, nutrients, hormones, wastes,

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Relation between QT interval and heart rate

Relation between QT interval and heart rate Br Heart J 1981; 45: 56-61 Relation between QT interval and heart rate New design of physiologically adaptive cardiac pacemaker A F RICKARDS, J NORMAN From The National Heart Hospital, London The relation

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Electrocardiography Normal 5. Faisal I. Mohammed, MD, PhD

Electrocardiography Normal 5. Faisal I. Mohammed, MD, PhD Electrocardiography Normal 5 Faisal I. Mohammed, MD, PhD 1 Objectives 2 1. Describe the different waves in a normal electrocardiogram. 2. Recall the normal P-R and Q-T interval time of the QRS wave. 3.

More information

The Electrocardiogram

The Electrocardiogram The Electrocardiogram Chapters 11 and 13 AUTUMN WEDAN AND NATASHA MCDOUGAL The Normal Electrocardiogram P-wave Generated when the atria depolarizes QRS-Complex Ventricles depolarizing before a contraction

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Chapter 3 Biological measurement 3.1 Nerve conduction

Chapter 3 Biological measurement 3.1 Nerve conduction Chapter 3 Biological measurement 3.1 Nerve conduction Learning objectives: What is in a nerve fibre? How does a nerve fibre transmit an electrical impulse? What do we mean by action potential? Nerve cells

More information

Introduction to Lesson 2 - Heartbeat

Introduction to Lesson 2 - Heartbeat Introduction to Lesson 2 - Heartbeat Activity: Locate your pulse at rest. Count how many times it beats in 15 seconds (look at a clock), then multiply this number by 4. This is your pulse rate Approximately

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG Introduction An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG data can give important information about the health of the heart and can help physicians to diagnose

More information

Defibrillator. BiomedGuy

Defibrillator. BiomedGuy Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

Course Objectives. Proper Lead Placements. Review the ECG print paper. Review the mechanics of the Myocardium. Review basics of ECG Rhythms

Course Objectives. Proper Lead Placements. Review the ECG print paper. Review the mechanics of the Myocardium. Review basics of ECG Rhythms ECG Interpretations Course Objectives Proper Lead Placements Review the ECG print paper Review the mechanics of the Myocardium Review basics of ECG Rhythms How Leads Work The ECG Leads we use are Bipolar

More information

09/05/2018. True/False: All pacemakers have defibrillation capabilities.

09/05/2018. True/False: All pacemakers have defibrillation capabilities. Paul G. Tarasi M.D. Staff Anesthesiologist Allegheny Health Network Adjunct Clinical Assistant Professor of Anesthesiology Temple University School of Medicine Describe different types of implanted cardiac

More information

INTRODUCTION TO ECG. Dr. Tamara Alqudah

INTRODUCTION TO ECG. Dr. Tamara Alqudah INTRODUCTION TO ECG Dr. Tamara Alqudah Excitatory & conductive system of the heart + - The ECG The electrocardiogram, or ECG, is a simple & noninvasive diagnostic test which records the electrical

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

The Ventricular Evoked Response as Monitor for Adrenergic Stimulation

The Ventricular Evoked Response as Monitor for Adrenergic Stimulation 216 June 1999 The Ventricular Evoked Response as Monitor for Adrenergic Stimulation I. DJAJADISASTRA, G. WEYERS, P. SCHWEIZER Evangelisches Krankenhaus, Bergisch Gladbach, Germany Summary We investigated

More information

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand: Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Is This Thing Working?

Is This Thing Working? Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu Disclosures I currently receive research support from Medtronic, Inc. I have been compensated

More information

Birmingham Regional Emergency Medical Services System

Birmingham Regional Emergency Medical Services System Birmingham Regional Emergency Medical Services System 2018 ALCTE Summer Conference EKG Basics Brian Gober, MAT, ATC, NRP, CSCS Education Services Manager ECC Training Center Coordinator Birmingham Regional

More information

This leaflet is intended for patients with permanent pacemakers and their parents.

This leaflet is intended for patients with permanent pacemakers and their parents. Permanent PACEMAKER This leaflet is intended for patients with permanent pacemakers and their parents. It serves as an information tool and a guide that offers some advice in order to live safely. General

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

Biology 212: Anatomy and Physiology II. Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures

Biology 212: Anatomy and Physiology II. Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures Biology 212: Anatomy and Physiology II Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures References: Saladin, KS: Anatomy and Physiology, The Unity of Form

More information

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

ELECTROCARDIOGRAPHY (ECG)

ELECTROCARDIOGRAPHY (ECG) ELECTROCARDIOGRAPHY (ECG) The heart is a muscular organ, which pumps blood through the blood vessels of the circulatory system. Blood provides the body with oxygen and nutrients, as well as assists in

More information

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Pacing and Device Jargon Made Simple Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital Disclosure Dr Jonathan Timperley Honorarium from Bayer Arne Larson (1915

More information

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital Treatment Bradyarrhythmia Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital 1 What is the rhythm? Sinus Bradycardia What s rhythm Sinus Bradycardia Treatment Asymptomatic No treatment

More information

ECG and Cardiac Electrophysiology

ECG and Cardiac Electrophysiology ECG and Cardiac Electrophysiology Simon Some very basic electrophysiology Intracellular fluid: 10 mm Na, 140 mm K, etc. K Na-K ATPase Extracellular fluid: 140mM Na, 4mM K, etc. Na Ion gradient plus selective

More information

Intraoperative Pacemaker Electrical Testing

Intraoperative Pacemaker Electrical Testing Intraoperative Pacemaker Electrical Testing James W. Calvin, M.D. ABSTRACT To determine the clinical value of intraoperative pacemaker system electrical testing, the results of 200 consecutive pacemaker

More information

An Official Journal of the American Heart

An Official Journal of the American Heart Circulation Research An Official Journal of the American Heart SEPTEMBER VOL. XXIII Association 8 NO. Incidence of Arrhythmias in the Dog Following Transthoracic Ventricular Defibrillation with Unidirectional

More information

Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission

Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission Surveillance and Follow-up of Patients with Implanted Cardiac Pacemaker by Telephone Transmission Takeo SAKURAI, M.D., Shuichi TOKUTSU, M.D., Osamu NISHIMUR M.D., Yutaka TAGAMI, M.D., Motoyuki TAKIMOTO,

More information

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals ECG SIGNAL ACQUISITION HARDWARE DESIGN Origin of Bioelectric Signals 1 Cell membrane, channel proteins Electrical and chemical gradients at the semi-permeable cell membrane As a result, we get a membrane

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D.

Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. NOTES An Implantable Demand Pacemaker Lawrence I. Zaroff, M.D.,* Barouh V. Berkovits, E.E. (Ing.),? Walter Zuckerman, M.D., and Dwight E. Harken, M.D. E lectrical stimulation in the treatment of heart

More information

Introduction to ECG Gary Martin, M.D.

Introduction to ECG Gary Martin, M.D. Brief review of basic concepts Introduction to ECG Gary Martin, M.D. The electrical activity of the heart is caused by a sequence of rapid ionic movements across cell membranes resulting first in depolarization

More information

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13 ECG Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13 The Concept When the cardiac impulse passes through the heart, electrical current

More information

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded.

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded. EXPERIMENT 1 ELECTROCARDIOGRAPHY The purpose of this experiment is to introduce you to the techniques of electrocardiography and the interpretation of electrocardiograms. In Part A of the experiment, you

More information

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER MODEL 53401 EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER In-service Guide 53401 TRAINING LEARNING OBJECTIVES Identify the control knobs and be able to explain how to use them to program device. Distinguish

More information

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II). Cardiac Pacemaker Premature Beats When one of ectopic foci becomes irritable, it may spontaneously fire, leading to one or more premature beats. Atrial and junctional foci may become irritable from excess

More information

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds Sensors: Loggers: ECG Any 12 bit EASYSENSE Science in Sport Logging time: 10 seconds 204a ECG demonstration (Graph) Read Regular medical check ups are essential part of the life of a professional sports

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Sample. Analyzing the Heart with EKG. Computer

Sample. Analyzing the Heart with EKG. Computer Analyzing the Heart with EKG Computer An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring within the heart. In a healthy heart there is a natural pacemaker in

More information

ECG INTERPRETATION MANUAL

ECG INTERPRETATION MANUAL Lancashire & South Cumbria Cardiac Network ECG INTERPRETATION MANUAL THE NORMAL ECG Lancashire And South Cumbria Cardiac Physiologist Training Manual THE NORMAL ECG E.C.G CHECKLIST 1) Name, Paper Speed,

More information

EEG, ECG, EMG. Mitesh Shrestha

EEG, ECG, EMG. Mitesh Shrestha EEG, ECG, EMG Mitesh Shrestha What is Signal? A signal is defined as a fluctuating quantity or impulse whose variations represent information. The amplitude or frequency of voltage, current, electric field

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device.

To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device. 043 Lauren Evans To investigate the safe levels of radiotherapy administered to patients who have an implanted cardiac device. 4061.40 Summary With an ageing UK population, the number of patients with

More information

ECG (MCQs) In the fundamental rules of the ECG all the following are right EXCEP:

ECG (MCQs) In the fundamental rules of the ECG all the following are right EXCEP: ECG (MCQs) 2010 1- In the fundamental rules of the ECG all the following are right EXCEP: a- It is a biphasic record of myocardial action potential fluctuations. b- Deflection record occurs only during

More information