Lesson 4 ECG 2 Electrocardiography

Size: px
Start display at page:

Download "Lesson 4 ECG 2 Electrocardiography"

Transcription

1 Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 4 ECG 2 Electrocardiography PC running Windows XP or Mac OS X Lesson Revision BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA (805) , Fax (805) info@biopac.com

2 Page 2 Biopac Science Lab Lesson 4 ELECTROCARDIOGPAHY 2 I. SCIENTIFIC PRINCIPLES Recall from Lesson 3, ECG 1, that each cardiac cycle, or heartbeat, begins with a bioelectric impulse generated by the sinoatrial (SA) node, the pacemaker of the heart. The bioelectric impulse is conducted through the heart by the pacemaker system and travels as a wave of depolarization followed in a few thousandths of a second by a wave of repolarization. Ventricular muscle depolarization and repolarization are represented in the electrocardiogram (ECG) as the QRS complex and the T wave, respectively. The body contains fluids with ions that allow for electric conduction. This makes it possible to use electrodes on the surface of the skin to detect electrical activity in and around the heart and use an electrocardiograph to record the activity. Conveniently, the legs and arms act as simple extensions of points in the torso, allowing the recording and ground electrodes to be placed on the wrists and ankles (Fig. 4.1). - Lead I Lead II + Lead III + right arm left arm right leg (ground) left leg Fig. 4.1 Electrode placement options Fig. 4.2 Einthoven s triangle and vectorgraph The ECG is a record of the overall spread of electric current through the heart as a function of time in the cardiac cycle. The direction of polarity (+ or -) of the recorded waveforms depends upon the location of the recording electrodes on the surface of the body and whether the electrical activity during the cardiac cycle is coming toward or going away from the surface electrode. In general, as a wave of depolarization approaches a positive electrode, a positive voltage is seen by that electrode. If the wave of depolarization is traveling toward a negative electrode, a negative voltage will be seen. The term lead is defined as a spatial arrangement of two recording electrodes on the body. One lead is labeled positive (+) and the other negative (-). The electrode placements define the recording direction of the lead, which is called the lead axis or angle. The axis is determined by the direction when going from the negative to positive electrode. The electrocardiograph computes the voltage difference (magnitude) between the positive and negative electrodes and displays the changes in voltage difference with time. The standard bipolar limb leads are: Lead I = right arm (-), left arm (+) Lead II = right arm (-), left leg (+) Lead III = left arm (-), left leg (+) The three standard bipolar limb leads may be used to construct an equilateral triangle, called Einthoven s triangle, at the center of which lies the heart (Fig. 4.2, left). Each side of the triangle represents one of the bipolar limb leads, and each lead forms a 60 angle with the two opposite leads. By placing Lead I through the center of the heart and moving the axes of Lead II and Lead III horizontally until they intersect Lead I at the center of the heart, a vectorgraph can be constructed (Fig. 4.2, right). The vectorgraph is used to plot vectors and aid in visualizing the electrical axes of the heart. At any given moment during the cardiac cycle (during the QRS complex, for example), the net electrical activity seen by a lead may be represented by a vector. A vector is an entity that can be represented by an arrow of specific length pointing in a specific direction. A vector has both magnitude and direction. The length of the arrow, divided equally into 0.1 mv units, represents the vector s magnitude, and is expressed as a mean potential. The direction of the vector, in degrees, is represented by its position relative to the horizontal 0º to 180º line (Lead I) on the vectorgraph. The dominant waveform in all three bipolar limb leads is the QRS complex signifying ventricular depolarization. The mean axis of ventricular depolarization is called the mean QRS axis. The mean QRS axis is diagnostically important

3 Lesson 4: ECG 2 Page 3 because it represents the majority of the heart s electrical activity occurring during a cardiac cycle. When the bipolar limb leads are used to obtain data for plotting the vector, the mean QRS axis lies in the frontal plane because the electrical picture seen by the recording electrodes placed on the extremities lies in the frontal plane. In normal persons, the mean QRS axis is between -30 and +90 (Fig. 4.2). In the frontal plane, the mean QRS axis is often between 0 and +90. A mean QRS axis between +90 and +180 indicates right axis deviation (RAD). RAD occurs, for example, in persons with right ventricular hypertrophy (over-sized right ventricle). A mean QRS axis between -30 and -90 indicates left axis deviation (LAD). LAD is found in patients with left anterior blockage in the conduction system of the ventricle. To define the mean QRS axis precisely, the cardiologist defines it in three dimensions (X, Y and Z planes). This is done in practice by using a standardized set of 12 leads. In this lesson, you will sequentially record two of the bipolar limb leads (Lead I and Lead II) and extract the data from the ECG necessary to determine the mean QRS axis in the frontal plane. A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead II (Fig. 4.3). magnitude of R wave from leads 0,0 I Mean Electrical Axis III Mean Electrical Axis Fig. 4.3 Mean electrical axis In this lesson, you will record data from Lead I and Lead III connections with the Subject lying down and sitting up. II. EXPERIMENTAL OBJECTIVES 1) To record ECG from Leads I and III in the following conditions: lying down, sitting up, and breathing deeply while sitting. 2) To compare the direction of the QRS complex (+ or ) with the direction of the lead axis. 3) To estimate the mean ventricular potential. 4) To estimate the mean electrical axis of the QRS complex using two methods. III. MATERIALS Computer system (running Windows XP or Mac OS X) Biopac Science Lab system (MP40 and software) Electrode lead set (40EL lead set) Disposable vinyl electrodes (EL503), four electrodes per subject Cot or lab pillow Protractor Two different color pens or pencils

4 Page 4 Biopac Science Lab IV. EXPERIMENTAL METHODS A. Set Up EQUIPMENT SUBJECT - Lead I + right forearm WHITE lead left forearm RED lead Fig. 4.4 right leg BLACK lead (Ground) Fig. 4.5 Electrodes for Leads I & III/Lead I configuration FAST TRACK 1. Turn the computer ON. 2. Set the MP40 dial to OFF. 3. Plug the equipment in as follows: Electrode leads (40EL) MP Attach four electrodes to the Subject as shown in Fig Connect the electrode leads (40EL) for a Lead I configuration, matching lead color to electrode position as shown above. Details Attach four electrodes to the Subject to allow Lead I and Lead III configurations, as shown in Fig. 4.5: one electrode on the right leg, just above the ankle bone one electrode on the left leg, just above the ankle bone one electrode on the left wrist (same side of arm as the palm of hand) one electrode on the right wrist (same side of arm as the palm of hand). For optimal electrode response, place electrodes on the skin at least five minutes before starting the Check section. Lead I will be recorded first, so connect the leads for a Lead I configuration: white: right wrist; red: left wrist; black: right ankle. IMPORTANT Clip each electrode lead color to its specified electrode position. 6. Start the Biopac Science Lab software. 7. Choose lesson L04-ECG-2 and click OK. 8. Type in a unique file name. No two people can have the same file name, so use a unique identifier, such as the Subject s nickname or student ID#. 9. Click OK. This ends the Set Up procedure.

5 Lesson 4: ECG 2 Page 5 B. Check MP40 Check FAST TRACK Details 1. Set the MP40 dial to ECG/EOG. 2. Press and hold the Check pad on the MP Click when the light is flashing. 4. Wait for the MP40 Check to stop. 5. Let go of the Check pad. 6. Click Continue. Signal Check 7. Click Check Signal. 8. Wait for the Signal Check to stop. 9. Review the data. If correct, go to the Record section. Continue to hold the pad down until prompted to let go. The MP40 check procedure will last five seconds. The light should stop flashing when you let go of the Check pad. When the light stops flashing, click Continue. Lead I configuration is used for the signal check. Subject should be supine, relaxed, and still. The electrocardiograph is very sensitive to small changes in voltage caused by contraction of skeletal muscles, and the Subject s arms and legs need to be relaxed so that the muscle (EMG) signal does not corrupt the ECG signal. The eight-second Signal Check recording should resemble Fig If incorrect, click. Fig. 4.6 If the recording does not show a recognizable ECG waveform with R-wave peaks and no large baseline drift, click Redo Signal Check.

6 Page 6 Biopac Science Lab C. Record FAST TRACK Details 1. Prepare for the recording and have the Subject lie down and relax. SEGMENT 1 Supine, Lead I Watch the Help menu videos to prepare for the recording. You will record two segments for each lead configuration, one with the Subject lying down, and another with the Subject sitting up. In order to work efficiently, read this entire section so you will know what to do before recording. Check the last line of the journal and note the total amount of time available for the recording. Stop each recording segment as soon as possible so you don t use an excessive amount of time (time is memory). Tips for obtaining optimal data To minimize muscle (EMG) corruption of the ECG and baseline drift: a) The Subject should not talk or laugh during any of the recording segments. b) When lying down or sitting up, the Subject should be completely relaxed. c) When sitting up, the Subject s arms should be supported on an armrest (if available). d) The recording should be suspended before the Subject prepares for the next recording segment. e) The Subject should breathe normally during the recording to minimize EMG from the chest area. f) Make sure electrodes do not peel up. 2. Click. When you click Record, the recording will begin and an append marker labeled Supine, Lead I will automatically be inserted. 3. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 0-30). 4. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 5. Review the data. Your data should resemble Fig If correct, go to Step 6. If incorrect, click Redo. Recording continues Fig. 4.7 Segment 1 (Supine, Lead I) A little baseline drift when the Subject breathes in and out does not indicate incorrect data. The data is incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If the data is incorrect, click Redo and repeat Steps 1-5; the last data segment you recorded will be erased.

7 Lesson 4: ECG 2 Page 7 SEGMENT 2 Supine, Lead III 6. Disconnect the leads from the Lead I configuration and reconnect to establish a Lead III configuration (Fig. 4.8). - Lead III + left forearm WHITE lead white: left wrist red: left ankle black: right ankle right leg BLACK lead (Ground) left leg RED lead Fig. 4.8 Lead III Configuration 7. Click Resume. When you click Resume, the recording will begin and an append marker labeled Lying down, Lead III will automatically be inserted. 8. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 30-60). 9. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 10. Review the data on the screen. If correct, go to Step 11. Your data should resemble Fig If incorrect, click Redo. Fig. 4.9 Segment 2 (Supine, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If data was incorrect, click Redo and repeat Steps 7-10; the last data segment you recorded will be erased. Recording continues

8 Page 8 Biopac Science Lab SEGMENT 3 Seated, Lead III 11. Have the Subject quickly get up and sit in a chair, with arms relaxed. 12. Click Resume as soon as possible after the Subject sits up. 13. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible Note Do not insert markers. Subject should sit with arms relaxed, preferably in a chair with armrests. In order to capture the heart rate variation, it is important that you resume recording as quickly as possible after the Subject sits. However, it is also important that you do not click Resume while the Subject is in the process of sitting or you will capture motion artifact. When you click Resume, the recording will continue and an append marker labeled Seated, Lead III will be automatically inserted. After 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds 60-90). 14. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 15. Review the data on the screen. If correct, go to Step 13. Your data should resemble Fig If incorrect, click Redo. Fig Segment 3 (Seated, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 11-15; the last data segment you recorded will be erased. Recording continues

9 Lesson 4: ECG 2 Page 9 SEGMENT 4 Seated, Lead I 16. Disconnect the leads from the Lead III configuration and reconnect to establish a Lead I configuration (Fig. 4.11) while Subject remains seated and relaxed. - Lead I + right forearm WHITE lead left forearm RED lead white: right wrist red: left wrist black: right ankle right leg BLACK lead (Ground) Fig Lead 1 configuration 17. Click Resume. When you click Resume, the recording will continue and an append marker labeled Seated, Lead I will be automatically inserted. 18. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible. Note Do not insert markers. After about 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds ). 19. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data. 20. Review the data. If correct, go to Step 21. Your data should resemble Fig If incorrect, click Redo. Recording continues Fig Segment 4 (Seated, Lead I) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 17-20; the last data segment you recorded will be erased.

10 Page Optional: Click Resume to record additional segments. 22. Click Done. 23. Click Yes. 24. Choose an option and click OK. 25. Remove the electrodes. END OF RECORDING Biopac Science Lab Optional: You can record additional segments by clicking Resume instead of Done. A time marker will be inserted at the start of each added segment. A pop-up window with options will appear. Click Yes (or No if you want to redo the last segment). When you click Yes, a dialog with options will be generated. Make your choice, and click OK. If you choose Analyze current data file, go to the Analyze section for directions. Unclip the electrode lead connectors and peel off the electrodes. Throw out the electrodes.

11 Lesson 4: ECG 2 Page 11 V. ANALYZE In this section, you will examine ECG components of cardiac cycles and measure amplitudes (mv) and durations (msecs) of the ECG components. Note: Interpreting an ECG is a skill that requires practice to distinguish between normal variation and those arising from medical conditions. Do not be alarmed if your ECG is different than the examples shown or from the tables and figures in Section I (Scientific Principles). Because ECGs are widely used, basic elements have been standardized to simplify reading ECGs. ECGs have standardized grids of lighter, smaller squares and, superimposed on the first grid, a second grid of darker and larger squares (Fig. 4.13). The smaller grid always has time units of 0.04 seconds on the x-axis, and the darker vertical lines are spaced 0.2 seconds apart. The horizontal lines represent amplitude in mv. The lighter horizontal lines are 0.1 mv apart, and the darker grid lines represent 0.5 mv. Fig Standard ECG grid FAST TRACK Details 1. Enter the Review Saved Data mode and choose the correct data file. Note Channel Number (CH) designations: To review saved data, choose Analyze current data file from the Done dialog after recording data, or choose Review Saved Data from the Lessons menu and browse to the required file. Channel Displays CH1 CH2 CH3 CH4 Supine, Lead I Supine, Lead III Seated, Lead III Seated, Lead I Important For this lesson analysis, segments that were recorded sequentially have been split into channels for relative comparison and do not share a time base. 2. Set up measurement boxes for: Channel SC Measurement Max Fig IMPORTANT: Channels do not share a time base The measurement boxes are above the marker region in the data window. Each measurement has three sections: channel number, measurement type, and result. The first two sections are pull-down menus that are activated when you click them. SC SC Delta Mean Max: finds the maximum amplitude value within the selected area (including the endpoints) by the I-Beam tool. Delta: computes the difference in amplitude between the last point and the first point of the selected area. It is particularly useful for taking ECG measurements because the baseline does not have to be at zero to obtain accurate, quick measurements. Note The polarity of Delta is based on (first point-last point). First point is where you begin your selection; last point is where you release the mouse to end your selection. 3. For each of the leads, note whether the R wave is positive or negative. A This is just a visual check as to whether the R wave goes up (positive) or down (negative).

12 Page 12 Biopac Science Lab Analysis continues 4. Set up the display window for optimal viewing of one cardiac cycle in the supine channels. 5. Using the I-Beam tool, select and measure Max from a cycle in the center of each channel. B Fig The following tools help you adjust the data window: Autoscale Horizontal Autoscale Waveforms Zoom Your window should resemble Fig Horizontal (Time) Scroll Bar Vertical (Amplitude) Scroll Bar Zoom Previous/Back You ll use the recorded measurements to construct the graphical estimate of the Mean QRS Potential in the Data Report. 6. Place an event marker to indicate where the QRS measurement was taken you will add four markers. Insert the event marker directly above the selected R wave. To place an event marker after the data has been recorded, click the cursor in the event marker region under the marker label bar (above the top data channel). An inverted triangle should appear, which, when selected, will be yellow. Type marker labels: Max Lead I down, Max Lead I up, Max Lead III down, and Max Lead III up. 7. Repeat Steps 5 and 6 for a cardiac cycle in the Seated channel. B 8. Select the breathe in area of a cardiac cycle in the CH3 and repeat Steps 5 and 6. B 9. Select the breathe in area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe in was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as an upward spike in the data. 10. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe out was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as a downward spike in the data. 11. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B 12. Go back to the Max Lead I down marker created in Step 6. This is the same QRS region you measured and marked in Steps 5 and 6. It represents a QRS cycle in the supine, relaxed

13 Lesson 4: ECG 2 Page Measure and record the amplitudes of the Q, R, and S waves individually for both Lead I and Lead III for each segment. C state. You will be taking a total of six measurements. You may wish to look at one channel at a time by hiding the other channel(s) using the channel boxes. To measure a peak, go from the baseline (isoelectric line) to the peak of the wave. Fig Sample measurement of the R-wave peak 14. Save or print the data file. You may save the data, save notes that are in the journal, or print the data file. 15. Exit the program. 16. Set the MP40 dial to Off. END OF LESSON 4 Complete the Lesson 4 Data Report that follows.

14 Page 14 Biopac Science Lab The Data Report starts on the next page.

15 Biopac Science Lab Page 15 Lesson 4 ELECTROCARDIOGRAPHY 2 DATA REPORT Student s Name: Lab Section: Date: These are sample questions. You should amend, add, or delete questions to support your curriculum objectives. I. Data and Calculations Subject Profile Name Gender: Male / Female Height Weight Age Direction of R-Waves for Different Leads A. Put a check in the column to indicate whether the R-wave is positive (+) or negative ( ) for each lead configuration: Table 4.1 Lead Lead I Lead III R-wave + Mean QRS Potential and Axis Graphical Estimate Using R-wave Amplitude B. Use Table 4.2 to record measurements from the Data Analysis section: Table 4.2 QRS Potential and QRS Axis Condition Potential QRS Supine Seated Breathe In Breathe Out Lead I Max Lead III Max Mean QRS Potential Mean QRS Axis

16 Page 16 Biopac Science Lab A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead III (Fig. 4.3, repeated below from Section 1). magnitude of R wave from leads 0,0 I Mean Electrical Axis III Mean Electrical Axis Fig ) Determine the R-wave amplitude (mv) in Lead I and Lead III. 2) Plot the Lead I R-wave amplitude value on the Lead I axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 3) Plot the Lead III R-wave amplitude value on the Lead III axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 4) Draw a line perpendicular to the tip of each arrowhead 5) Determine the point of intersection of these two perpendicular lines. 6) Draw a new vector from point 0,0 at the center of the vectorgraph to the point where the perpendicular lines intersect with one another. The direction (in degrees) of this resulting vector approximates the mean QRS axis. The length (mv) of this vector approximates the mean QRS potential. To determine Mean QRS Potential from the plotted vector, subdivide it into 0.1 mv increments exactly as the leads are in the vectorgraph. Alternatively, the length of the vector can be measured (with a compass, string, notepaper, etc.) and then transferred to a lead line to determine the mv value. One way to approximate the mean electrical axis in the frontal plane is to plot the magnitude of the R wave from Lead I and Lead III, as shown in Fig 4.2 in Section I (Scientific Principles). 1. Draw a perpendicular line from the ends of the vectors (right angles to the axis of the Lead) using a protractor or right angle guide. 2. Determine the point of intersection of these two perpendicular lines. 3. Draw a new vector from point 0,0 to the point of intersection. The direction of this resulting vector approximates the mean electrical axis of the heart. The length of this vector approximates the mean potential of the heart.

17 Lesson 4: ECG 2 Page 17 Create two plots on each of the following graphs, using data from Table 4.2. Use a different color pencil or pen for each plot. Graph 1: Lead I Lying Down and Sitting Up 0,0 mv Lead I mv Lead III Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference (if any) in Mean QRS Potential and Axis under the two conditions:

18 Page 18 Biopac Science Lab Graph 2: Breathing In /Breathing Out or Lead III Lying Down/Sitting Up 0,0 mv Lead I mv Lead III Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference in Mean QRS Potential and Axis for the Lying Down data in this plot (Graph 2) and the first plot (Graph 1). C. Mean QRS Potential and Magnitude More Accurate Approximation Add the Q, R, and S potentials to obtain net potentials. Lying down: Lead I Q R S Lead III Q R S QRS Net 1 QRS Net 2

19 Lesson 4: ECG 2 Page 19 II. Data Summary and Questions D. Define ECG. E. Define Einthoven s Triangle. F. What factors affect the orientation of the mean electrical axis? G. Refer to Table 4.2: How did the amplitudes of Lead I and Lead III change between inhalation and exhalation? Did the heart s axis and magnitude change? H. What factors affect the amplitude of the R wave recorded on the different leads? I. Compare the mean electrical axis and magnitude obtained when i. using just the amplitude measurement of the R wave vs. net potentials ii. lying down vs. sitting up End of Biopac Science Lab Lesson 4 Data Report

20 Biopac Science Lab Page 20 VI. ACTIVE LEARNING LAB Design a new experiment to test or verify the scientific principle(s) you learned in the Biopac Science Lab recording and analysis segments. For this lesson, you might examine how gender, duration of exercise, and changing body position influence the ECG and/or time to return to baseline. Design Your Experiment Use a separate sheet to detail your experiment design, and be sure to address these main points: A. Hypothesis Describe the scientific principle to be tested or verified. B. Materials List the materials will you use to complete your investigation. C. Method Describe the experimental procedure be sure to number each step to make it easy to follow during recording. See the Set Up section or Help > About Electrodes for electrode placement guidelines. Run Your Experiment D. Setup Set up the equipment and prepare the subject for your experiment. E. Recording Use the Record, Resume, and Suspend buttons in the Biopac Science Lab program to record as many segments as necessary for your experiment. Click Done when you have completed all of the segments required for your experiment. Analyze Your Experiment F. Set measurements relevant to your experiment and record the results in a Data Report.

Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Analysis Procedure. Rev

Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Analysis Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Analysis Procedure Rev. 12292017 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University

More information

Lesson 2 EMG 2 Electromyography: Mechanical Work

Lesson 2 EMG 2 Electromyography: Mechanical Work Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 2 EMG 2 Electromyography: Mechanical Work PC running Windows XP or Mac OS X 10.3-10.4 Lesson Revision 5.23.2006 BIOPAC Systems, Inc. 42

More information

Physiology Lessons for use with the Biopac Student Lab

Physiology Lessons for use with the Biopac Student Lab Physiology Lessons for use with the Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY II Bipolar Leads (Leads I, II, III) Einthoven s Law Mean Electrical Axis on the Frontal Plane Richard Pflanzer, Ph.D.

More information

Lesson 10 Aerobic Exercise Physiology

Lesson 10 Aerobic Exercise Physiology Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 10 Aerobic Exercise Physiology PC running Windows XP or Mac OS X 10.3-10.4 Lesson Revision 12.19.2005 BIOPAC Systems, Inc. 42 Aero Camino,

More information

Biopac Student Lab Lesson 5 ELECTROCARDIOGRAPHY (ECG) I Procedure. Rev

Biopac Student Lab Lesson 5 ELECTROCARDIOGRAPHY (ECG) I Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 5 ELECTROCARDIOGRAPHY (ECG) I Procedure Rev. 07112013 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University School

More information

Lesson 5 BIOPAC Systems, Inc. Manual Revision PL3.7.5

Lesson 5 BIOPAC Systems, Inc. Manual Revision PL3.7.5 Physiology Lessons for use with the Biopac Student Lab Lesson 5 ELECTROCARDIOGRAPHY I Components of the ECG Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University School of Medicine Purdue

More information

Lesson 5 EEG 1 Electroencephalography: Brain Rhythms

Lesson 5 EEG 1 Electroencephalography: Brain Rhythms Physiology Lessons for use with the Biopac Science Lab MP40 PC running Windows XP or Mac OS X 10.3-10.4 Lesson 5 EEG 1 Electroencephalography: Brain Rhythms Lesson Revision 2.23.2006 BIOPAC Systems, Inc.

More information

Lesson 2 ELECTROMYOGRAPHY II Motor unit recruitment Fatigue

Lesson 2 ELECTROMYOGRAPHY II Motor unit recruitment Fatigue Physiology Lessons for use with the PC running Windows 98SE, Me, 2000 Pro, XP Pro/Home/Media Lesson 2 ELECTROMYOGRAPHY II Motor unit recruitment Fatigue Manual Revision PL3.7.0 M3.0.6 080706 Richard Pflanzer,

More information

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Lab #10 Table of Contents: Expected Learning Outcomes...... 83 Introduction....... 84 Activity 1: Collecting ECG Data..... 85 Activity

More information

Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG

Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG Physiology Lessons for use with the PC running Windows 98SE, Me, 2000 Pro, XP Pro/Home/Media Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG Manual Revision PL3.7.0 080706 Richard Pflanzer, Ph.D.

More information

Physiology Lessons for use with the Biopac Student Lab. Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG

Physiology Lessons for use with the Biopac Student Lab. Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG Physiology Lessons for use with the Biopac Student Lab Lesson 1 ELECTROMYOGRAPHY I Standard and Integrated EMG Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University School of Medicine

More information

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram Experiment HH-4: The Six-Lead Electrocardiogram Background The cardiac cycle involves a sequential contraction of the atria and the ventricles. These contractions are triggered by the coordinated electrical

More information

Biopac Student Lab Lesson 3 ELECTROENCEPHALOGRAPHY (EEG) I Analysis Procedure. Rev

Biopac Student Lab Lesson 3 ELECTROENCEPHALOGRAPHY (EEG) I Analysis Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 3 ELECTROENCEPHALOGRAPHY (EEG) I Analysis Procedure Rev. 12112014 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana

More information

INTRODUCTION TO ECG. Dr. Tamara Alqudah

INTRODUCTION TO ECG. Dr. Tamara Alqudah INTRODUCTION TO ECG Dr. Tamara Alqudah Excitatory & conductive system of the heart + - The ECG The electrocardiogram, or ECG, is a simple & noninvasive diagnostic test which records the electrical

More information

Biopac Student Lab Lesson 20 SPINAL CORD REFLEXES Analysis Procedure. Rev

Biopac Student Lab Lesson 20 SPINAL CORD REFLEXES Analysis Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 20 SPINAL CORD REFLEXES Analysis Procedure Rev. 12292017 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University

More information

Electrocardiography I Laboratory

Electrocardiography I Laboratory Introduction The body relies on the heart to circulate blood throughout the body. The heart is responsible for pumping oxygenated blood from the lungs out to the body through the arteries and also circulating

More information

Richard Pflanzer, Ph.D. J.C. Uyehara, Ph.D. William McMullen David Pittman, Ph.D. Manual Revision PL3.7.0 M3.0.6 Modified by David Pittman

Richard Pflanzer, Ph.D. J.C. Uyehara, Ph.D. William McMullen David Pittman, Ph.D. Manual Revision PL3.7.0 M3.0.6 Modified by David Pittman ELECTROMYOGRAPHY (EMG) Laboratory Motor unit recruitment & Fatigue Comparison of dominant versus non-dominant arm using a within subject repeated measures statistical analysis Written by: Richard Pflanzer,

More information

Lab 7. Physiology of Electrocardiography

Lab 7. Physiology of Electrocardiography 7.1 Lab 7. Physiology of Electrocardiography The heart is a muscular pump that circulates blood throughout the body. To efficiently pump the blood, cardiac contractions must be coordinated and are regulated

More information

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram

iworx Sample Lab Experiment HH-4: The Six-Lead Electrocardiogram Experiment HH-4: The Six-Lead Electrocardiogram Exercise 1: Six Lead ECG from Resting Subject Aim: To record a Six Lead ECG from a resting subject and determine the QRS axis of the subject s heart. Procedure

More information

Biopac Student Lab Lesson 8 RESPIRATORY CYCLE I Analysis Procedure. Rev

Biopac Student Lab Lesson 8 RESPIRATORY CYCLE I Analysis Procedure. Rev 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 8 RESPIRATORY CYCLE I Analysis Procedure Rev. 12292017 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University

More information

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Background The arterial system functions as a pressure reservoir. Blood enters via the heart and exits through the capillaries.

More information

LABORATORY INVESTIGATION

LABORATORY INVESTIGATION LABORATORY INVESTIGATION Recording Electrocardiograms The taking of an electrocardiogram is an almost universal part of any complete physical examination. From the ECG record of the electrical activity

More information

BSL PRO Lesson H03: Nerve Conduction Velocity: Along the Ulnar Nerve of a Human Subject

BSL PRO Lesson H03: Nerve Conduction Velocity: Along the Ulnar Nerve of a Human Subject Updated 12-22-03 BSL PRO Lesson H03: Nerve Conduction Velocity: Along the Ulnar Nerve of a Human Subject This PRO lesson describes hardware and software setup of the BSL PRO System to record and measure

More information

#4 Cardiovascular I The Heart & EKG

#4 Cardiovascular I The Heart & EKG #4 Cardiovascular I Objectives: Identify a list of human heart structures using a virtual human dissection Dissect a sheep heart to identify external and internal structures Identify a list of human heart

More information

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds

Science in Sport. 204a ECG demonstration (Graph) Read. The Electrocardiogram. ECG Any 12 bit EASYSENSE. Sensors: Loggers: Logging time: 10 seconds Sensors: Loggers: ECG Any 12 bit EASYSENSE Science in Sport Logging time: 10 seconds 204a ECG demonstration (Graph) Read Regular medical check ups are essential part of the life of a professional sports

More information

Overview. Page 1 of 9. Impedance Cardiography

Overview.  Page 1 of 9. Impedance Cardiography Updated 05.14.10 BSL PRO Lesson H21: Impedance Cardiography Data collected from a subject using the referenced set-up procedure. Note that dz/dt maximum is determined on a cycle by cycle basis from the

More information

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation Exercise 1: The ECG and the Pulse in a Resting Subject Aim: To measure and correlate the ECG and the pulse in a resting individual.

More information

Exercise 5: Muscle Physiology I - Electromyography

Exercise 5: Muscle Physiology I - Electromyography Exercise 5: Muscle Physiology I - Electromyography Readings: Silverthorn, 6 th ed. pg. 410 420 Your brain communicates with your muscles through action potentials on the motor neurons, which are then transmitted

More information

Experiment HH-2: The Electrocardiogram and Heart Sounds

Experiment HH-2: The Electrocardiogram and Heart Sounds Experiment HH-2: The Electrocardiogram and Heart Sounds Exercise 1: The ECG in a Resting Subject Aim: To measure the ECG in a resting individual. Procedure 1. Click on the Record button, located on the

More information

Cardiovascular Effects of Exercise. Background Cardiac function

Cardiovascular Effects of Exercise. Background Cardiac function Cardiovascular Effects of Exercise In this experiment, you will record an electrocardiogram, or ECG, and finger pulse from a healthy volunteer. You will then compare the ECG and pulse recordings when the

More information

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded.

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded. EXPERIMENT 1 ELECTROCARDIOGRAPHY The purpose of this experiment is to introduce you to the techniques of electrocardiography and the interpretation of electrocardiograms. In Part A of the experiment, you

More information

Sample. Analyzing the Heart with EKG. Computer

Sample. Analyzing the Heart with EKG. Computer Analyzing the Heart with EKG Computer An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring within the heart. In a healthy heart there is a natural pacemaker in

More information

Lect.6 Electrical axis and cardiac vector Cardiac vector: net result Vector that occurs during depolarization of the ventricles Figure:

Lect.6 Electrical axis and cardiac vector Cardiac vector: net result Vector that occurs during depolarization of the ventricles  Figure: Lect.6 Electrical axis and cardiac vector Objectives: 1. State the relationship between the direction of cardiac vector with the direction (-ve, +ve) and amplitude of an ECG waves. 2. Draw diagram indicting

More information

Warning: Avoid putting the tip of the syringe needle too deeply into the hole to avoid damage to the heart tissue.

Warning: Avoid putting the tip of the syringe needle too deeply into the hole to avoid damage to the heart tissue. Experiment AM-12: Crayfish Electrocardiogram (ECG) Exercise 1: The Crayfish ECG and Heart Rate Aim: To record the electrical trace (ECG) produced by the contraction of a resting heart, and to determine

More information

The Normal Electrocardiogram

The Normal Electrocardiogram C H A P T E R 1 1 The Normal Electrocardiogram When the cardiac impulse passes through the heart, electrical current also spreads from the heart into the adjacent tissues surrounding the heart. A small

More information

Figure 1 The respiratory cycle

Figure 1 The respiratory cycle RESPIRATORY CYCLE Introduction Three primary functions of the respiratory system are to provide oxygen for the body's energy needs, provide an outlet for C0 2 and help maintain the ph of the blood plasma.

More information

5- The normal electrocardiogram (ECG)

5- The normal electrocardiogram (ECG) 5- The (ECG) Introduction Electrocardiography is a process of recording electrical activities of heart muscle at skin surface. The electrical current spreads into the tissues surrounding the heart, a small

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology The mammalian heart is a pump that pushes blood around the body and is made of four chambers: right and left atria and right and left ventricles. The two atria act as collecting

More information

Figure 1 muscle tissue to its resting state. By looking at several beats you can also calculate the rate for each component.

Figure 1 muscle tissue to its resting state. By looking at several beats you can also calculate the rate for each component. ANALYZING THE HEART WITH EKG WITH LABQUEST LAB From Human Physiology with Vernier Westminster College INTRODUCTION An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring

More information

Lesson 9. Measuring Emotional Responses to the Acoustic Startle Reflex: Stimulus Habituation & Sensitization

Lesson 9. Measuring Emotional Responses to the Acoustic Startle Reflex: Stimulus Habituation & Sensitization Measuring Emotional Responses to the Acoustic Startle Reflex: Stimulus Habituation & Sensitization Revised 10 / 29 / 2012 Lesson 9 METHOD: GALVANIC SKIN RESPONSE & THE POLYGRAPH David W. Pittman, Ph.D.

More information

Lab 5: Electromyograms (EMGs)

Lab 5: Electromyograms (EMGs) Lab 5: Electromyograms (EMGs) Overview A motorneuron and all the muscle fibers that it innervates is known as a motor unit. Under normal circumstances, a neuronal action potential activates all of the

More information

iworx Sample Lab Experiment HC-3: Pulse Wave Velocity

iworx Sample Lab Experiment HC-3: Pulse Wave Velocity Experiment HC-3: Pulse Wave Velocity Exercise 1: The Pulse Wave Velocity at Rest Aim: To measure the pulse wave velocity in a resting individual. Procedure 1. Click on the Record button, located on the

More information

Analyzing the Heart with EKG

Analyzing the Heart with EKG Analyzing the Heart with EKG LabQuest An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring within the heart. In a healthy heart there is a natural pacemaker in

More information

Exercise 6: Muscle Physiology II Twitch & Summation

Exercise 6: Muscle Physiology II Twitch & Summation Exercise 6: Muscle Physiology II Twitch & Summation Text Reading: Silverthorn, 5 th ed. 412 419, 425 427; 6 th ed. pg. 410 420 In this exercise, we will investigate the physiology of contraction in the

More information

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

Interpreting Electrocardiograms (ECG) Physiology Name: Per: Interpreting Electrocardiograms (ECG) Physiology Name: Per: Introduction The heart has its own system in place to create nerve impulses and does not actually require the brain to make it beat. This electrical

More information

ELECTROCARDIOGRAPHY (ECG)

ELECTROCARDIOGRAPHY (ECG) ELECTROCARDIOGRAPHY (ECG) The heart is a muscular organ, which pumps blood through the blood vessels of the circulatory system. Blood provides the body with oxygen and nutrients, as well as assists in

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Axis. B.G. Petty, Basic Electrocardiography, DOI / _2, Springer Science+Business Media New York 2016

Axis. B.G. Petty, Basic Electrocardiography, DOI / _2, Springer Science+Business Media New York 2016 Axis 2 The electrical axis of any electrocardiogram (EKG) waveform is the average direction of electrical activity. It is not a vector, because by definition a vector has both direction and amplitude,

More information

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent?

CASE 10. What would the ST segment of this ECG look like? On which leads would you see this ST segment change? What does the T wave represent? CASE 10 A 57-year-old man presents to the emergency center with complaints of chest pain with radiation to the left arm and jaw. He reports feeling anxious, diaphoretic, and short of breath. His past history

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

EKG. Danil Hammoudi.MD

EKG. Danil Hammoudi.MD EKG Danil Hammoudi.MD What is an EKG? The electrocardiogram (EKG) is a representation of the electrical events of the cardiac cycle. Each event has a distinctive waveform, the study of which can lead to

More information

Lab 4: Introduction to Physiological Measurements - Cardiovascular

Lab 4: Introduction to Physiological Measurements - Cardiovascular Lab 4: Introduction to Physiological Measurements - Cardiovascular INTRODUCTION: This lab will demonstrate cardiovascular measurements by creating an ECG with instruments used in previous labs. Students

More information

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals ECG SIGNAL ACQUISITION HARDWARE DESIGN Origin of Bioelectric Signals 1 Cell membrane, channel proteins Electrical and chemical gradients at the semi-permeable cell membrane As a result, we get a membrane

More information

HUMAN ANATOMY AND PHYSIOLOGY

HUMAN ANATOMY AND PHYSIOLOGY HUMAN ANATOMY AND PHYSIOLOGY NAME Detection of heart sounds. Clean the ear pieces of the stethoscope before using. The ear pieces should be pointing slightly forward when inserted into the ears because

More information

BIO 360: Vertebrate Physiology Performing and analyzing an EKG Lab 11: Performing and analyzing an EKG Lab report due April 17 th

BIO 360: Vertebrate Physiology Performing and analyzing an EKG Lab 11: Performing and analyzing an EKG Lab report due April 17 th BIO 60: Vertebrate Physiology Lab : Lab report due April 7 th All muscles produce an electrical current when they contract. The heart is no exception. An electrocardiogram (ECG or EKG) is a graphical recording

More information

Sheet 5 physiology Electrocardiography-

Sheet 5 physiology Electrocardiography- *questions asked by some students Sheet 5 physiology Electrocardiography- -why the ventricles lacking parasympathetic supply? if you cut both sympathetic and parasympathetic supply of the heart the heart

More information

= BIOPAC. J ~ ~ t. Lesson 6 ELECTROCARDIOGRAPHY Bipolar Leads (Leads 1, II, III) Einthoven's Law Mean Electrical Axis on the Frontal Plane

= BIOPAC. J ~ ~ t. Lesson 6 ELECTROCARDIOGRAPHY Bipolar Leads (Leads 1, II, III) Einthoven's Law Mean Electrical Axis on the Frontal Plane = BIOPAC Systems, Inc. Physiology Lessons for use with the Biopac Student Lab PC under Windows 98SE, Me, 2000 Pro or Macintosh" 8.6-9.1 Lesson 6 ELECTROCARDIOGRAPHY II Bipolar Leads (Leads 1, II, III)

More information

BASIC CONCEPT OF ECG

BASIC CONCEPT OF ECG BASIC CONCEPT OF ECG Electrocardiogram The electrocardiogram (ECG) is a recording of cardiac electrical activity. The electrical activity is readily detected by electrodes attached to the skin. After the

More information

ELECTROCARDIOGRAPHY, ECG

ELECTROCARDIOGRAPHY, ECG ELECTROCARDIOGRAPHY, ECG Introduction: Electrocardiography is a method to register the heart electric activity, e.g. electric potentials from outside of body. It is possible to record the changes of these

More information

Lesson 1: Types of ECG s

Lesson 1: Types of ECG s Electrocardiography Lesson 1: Types of ECG s Objectives: Distinguish between a single-channel ECG machine and a multi-channel machine. Distinguish between a manual ECG machine and an automatic ECG machine.

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13

ECG. Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13 ECG Prepared by: Dr.Fatima Daoud Reference: Guyton and Hall Textbook of Medical Physiology,12 th edition Chapters: 11,12,13 The Concept When the cardiac impulse passes through the heart, electrical current

More information

ECG INTERPRETATION MANUAL

ECG INTERPRETATION MANUAL Lancashire & South Cumbria Cardiac Network ECG INTERPRETATION MANUAL THE NORMAL ECG Lancashire And South Cumbria Cardiac Physiologist Training Manual THE NORMAL ECG E.C.G CHECKLIST 1) Name, Paper Speed,

More information

Introduction to ECG Gary Martin, M.D.

Introduction to ECG Gary Martin, M.D. Brief review of basic concepts Introduction to ECG Gary Martin, M.D. The electrical activity of the heart is caused by a sequence of rapid ionic movements across cell membranes resulting first in depolarization

More information

ECG and Cardiac Electrophysiology

ECG and Cardiac Electrophysiology ECG and Cardiac Electrophysiology Simon Some very basic electrophysiology Intracellular fluid: 10 mm Na, 140 mm K, etc. K Na-K ATPase Extracellular fluid: 140mM Na, 4mM K, etc. Na Ion gradient plus selective

More information

Overview Active Learning

Overview Active Learning : EMG Active Learning Updated 01-06-11 This lesson is based on a presentation given by Dr. Jennifer Lundmark and Ms. Andrea Salmi at the HAPS Conference, Maui, HI, June 2-7, 2001 Overview Active Learning

More information

Electrocardiography negative zero LA/VL RA/VR LL/VF recording electrode exploring electrode Wilson right arm right arm, left arm left arm

Electrocardiography negative zero LA/VL RA/VR LL/VF recording electrode exploring electrode Wilson right arm right arm, left arm left arm Electrocardiography In the previous lecture, we were talking about the unipolar limb leads. We said that to make the unipolar lead, you have to make the negative electrode as zero electrode, this is done

More information

- why the T wave is deflected upwards although it's a repolarization wave?

- why the T wave is deflected upwards although it's a repolarization wave? Cardiac Electrograph: - why the T wave is deflected upwards although it's a repolarization wave? After depolarization the ventricle contracts but since the heart is a volume conductor (3D not 2D), when

More information

How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body

How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body How Do We Sense, Think, and Move? -- Lab #11 Bioelectronics Measuring Electrical Properties of the Body Experiment #1 Your Body's Resistance Equipment: Digital multimeter, Banana leads Important Equipment

More information

Exercise 2: Effects of Cold Temperature Aim: To record changes in heart rate after the heart is bathed in cold Ringer s solution.

Exercise 2: Effects of Cold Temperature Aim: To record changes in heart rate after the heart is bathed in cold Ringer s solution. Experiment AM-3: Heart Muscle Warning: The heart preparation used in this experiment is functional for a limited period of time. If the muscle is bathed periodically in Ringer s solution, it will work

More information

Experiment HM-11: Electromyograms (EMG) for Paired Arm Wrestling

Experiment HM-11: Electromyograms (EMG) for Paired Arm Wrestling Experiment HM-11: Electromyograms (EMG) for Paired Arm Wrestling Background The movement of parts of the body is accomplished through a system of levers composed of skeletal muscles and bones. In a lever,

More information

Birmingham Regional Emergency Medical Services System

Birmingham Regional Emergency Medical Services System Birmingham Regional Emergency Medical Services System 2018 ALCTE Summer Conference EKG Basics Brian Gober, MAT, ATC, NRP, CSCS Education Services Manager ECC Training Center Coordinator Birmingham Regional

More information

iworx Sample Lab Experiment HP-7: Hypothesis-driven Biofeedback Lab/Research Study

iworx Sample Lab Experiment HP-7: Hypothesis-driven Biofeedback Lab/Research Study Experiment HP-7: Hypothesis-driven Biofeedback Lab/Research Study What is Biofeedback? Biofeedback is a technique that people can use to learn to control their body's physiological functions. The definition

More information

Experiment HM-7: Electromyogram (EMG) Activity in Antagonistic Muscles and Range of Motion

Experiment HM-7: Electromyogram (EMG) Activity in Antagonistic Muscles and Range of Motion Experiment HM-7: Electromyogram (EMG) Activity in Antagonistic Muscles and Range of Motion Exercise 1: Antagonistic Muscles in Forearm Aim: To study the EMG activity in muscles that work in opposition

More information

This presentation will deal with the basics of ECG description as well as the physiological basics of

This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 1 Electrocardiography basics This presentation will deal with the basics of ECG description as well as the physiological basics of Snímka 2 Lecture overview 1. Cardiac conduction system functional

More information

2The Concept of Periodic Functions

2The Concept of Periodic Functions The Concept of Periodic Functions EKG INTRODUCTION The heart is a fist-sized muscle that acts as an electrical generator lying in a conducting medium made of body tissue and fluids. The heart pumps oxygen-rich

More information

Teacher s Guide. Physiology Lessons for use with the Biopac Science Lab MP40

Teacher s Guide. Physiology Lessons for use with the Biopac Science Lab MP40 Physiology Lessons for use with the Biopac Science Lab MP40 Teacher s Guide Overview 2 About Electrodes 3 The Body Electric 6 Waveforms 7 National Science Standards 19 Lesson 1 Lesson 2 Lesson 3 Lesson

More information

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD The Electrocardiogram part II Dr. Adelina Vlad, MD PhD Basic Interpretation of the ECG 1) Evaluate calibration 2) Calculate rate 3) Determine rhythm 4) Determine QRS axis 5) Measure intervals 6) Analyze

More information

Educators have placed an emphasis on the development of laboratory materials that

Educators have placed an emphasis on the development of laboratory materials that VIRTUAL EXPERIMENT FOR UNDERSTANDING THE ELECTROCARDIOGRAM AND THE MEAN ELECTRICAL AXIS Jamie Anderson and Stephen E. DiCarlo Department of Physiology, Wayne State University, School of Medicine, Detroit,

More information

8/20/ Identify the functions of common ECG machines. 3.3 Explain how each ECG machine control is used. 3.4 Recognize common electrodes.

8/20/ Identify the functions of common ECG machines. 3.3 Explain how each ECG machine control is used. 3.4 Recognize common electrodes. 1 2 Electrocardiography for Healthcare Professionals Chapter 3: The Electrocardiograph Learning Outcomes 3.1 Identify three types of leads, and explain how each is recorded. 3.2 Identify the functions

More information

EKG Sensor Product Number: ENEKG189

EKG Sensor Product Number: ENEKG189 imagine explore learn EKG Sensor Product Number: ENEKG189 Overview An electrocardiogram abbreviated as EKG or ECG is a test that measures the electrical activity of the heartbeat. With each beat, an electrical

More information

ECE ECE PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM

ECE ECE PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM ECE0909.404.01 ECE 0909.504.03 PRINCIPLES OF BIOMEDICAL SYSTEMS & DEVICES LAB 1 - ELECTROCARDIOGRAM The purpose of this laboratory is to introduce you to electrocardiogram, its acquisition and interpretation.

More information

SS11LA Pneumotach Transducer

SS11LA Pneumotach Transducer BIOPAC Systems, Inc. 42 Aero Camino Goleta, CA 93117 Ph (805) 685-0066 Fax (805) 685-0067 www.biopac.com info@biopac.com SS11LA Pneumotach Transducer The SS11LA can be used to measure respiratory flow

More information

iworx Sample Lab Experiment AN-5: Cockroach Leg Mechanoreceptors

iworx Sample Lab Experiment AN-5: Cockroach Leg Mechanoreceptors Experiment AN-5: Cockroach Leg Mechanoreceptors Exercise 1: Chordotonal Organs Aim: To explore the basic characteristics of the chordotonal organs, their response to direction and intensity of leg movement,

More information

Biology 212: Anatomy and Physiology II. Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures

Biology 212: Anatomy and Physiology II. Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures Biology 212: Anatomy and Physiology II Lab #5: Physiology of the Cardiovascular System For Labs Associated With Dr. Thompson s Lectures References: Saladin, KS: Anatomy and Physiology, The Unity of Form

More information

Electrocardiography Normal 5. Faisal I. Mohammed, MD, PhD

Electrocardiography Normal 5. Faisal I. Mohammed, MD, PhD Electrocardiography Normal 5 Faisal I. Mohammed, MD, PhD 1 Objectives 2 1. Describe the different waves in a normal electrocardiogram. 2. Recall the normal P-R and Q-T interval time of the QRS wave. 3.

More information

Determining Axis and Axis Deviation on an ECG

Determining Axis and Axis Deviation on an ECG Marquette University e-publications@marquette Physician Assistant Studies Faculty Research and Publications Health Sciences, College of 7-15-2010 Determining Axis and Axis Deviation on an ECG Patrick Loftis

More information

Lab Activity 24 EKG. Portland Community College BI 232

Lab Activity 24 EKG. Portland Community College BI 232 Lab Activity 24 EKG Reference: Dubin, Dale. Rapid Interpretation of EKG s. 6 th edition. Tampa: Cover Publishing Company, 2000. Portland Community College BI 232 Graph Paper 1 second equals 25 little boxes

More information

ECG-PS ECG SIGNAL MEASUREMENT MODULE. Operation Manual. March 2005

ECG-PS ECG SIGNAL MEASUREMENT MODULE. Operation Manual. March 2005 ECG-PS ECG SIGNAL MEASUREMENT MODULE Operation Manual March 2005 2555 Collins Avenue, Suite C-5 - Miami Beach FL - 33140 - U.S.A. Phone #: (305) 534-5905 Fax: (305) 534-8222 e-mail: info@galix-gbi.com

More information

ELECTROMYOGRAM ANALYSIS OF MUSCLE FUNCTION INTRODUCTION

ELECTROMYOGRAM ANALYSIS OF MUSCLE FUNCTION INTRODUCTION ELECTROMYOGRAM ANALYSIS OF MUSCLE FUNCTION STANDARDS: 3.3.10.B - Explain cell functions and processes in terms of chemical reactions and energy changes. 3.3.12.B - Evaluate relationships between structure

More information

PROBLEM SET 3. Assigned: February 19, 2004 Due: February 26, 2004

PROBLEM SET 3. Assigned: February 19, 2004 Due: February 26, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

CARDIOVASCULAR PHYSIOLOGY ECG. Dr. Ana-Maria Zagrean

CARDIOVASCULAR PHYSIOLOGY ECG. Dr. Ana-Maria Zagrean CARDIOVASCULAR PHYSIOLOGY ECG Dr. Ana-Maria Zagrean Electrocardiogram (ECG) ECG is a non-invasive method to record at the body surface the electrical activity of the heart. - the rate and regularity of

More information

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2)

3/26/15 HTEC 91. EKG Sign-in Book. The Cardiac Cycle. Parts of the ECG. Waves. Waves. Review of protocol Review of placement of chest leads (V1, V2) EKG Sign-in Book HTEC 91 Review of protocol Review of placement of chest leads (V1, V2) Medical Office Diagnostic Tests Week 2 http://www.cvphysiology.com/arrhythmias/a013c.htm The Cardiac Cycle Represents

More information

1724 Lab: Frog Skeletal Muscle Physiology (Marieb Exercise 16A) Marieb/iWorx / Ziser, 2002

1724 Lab: Frog Skeletal Muscle Physiology (Marieb Exercise 16A) Marieb/iWorx / Ziser, 2002 1724 Lab: Frog Skeletal Muscle Physiology (Marieb Exercise 16A) Marieb/iWorx / Ziser, 2002 I. Introduction. Read the introductory material in your lab manual Marieb Ex 16A: Skeletal Muscle Physiology Frog

More information

Relax and Learn At the Farm 2012

Relax and Learn At the Farm 2012 Relax and Learn At the Farm 2012 Session 2: 12 Lead ECG Fundamentals 101 Cynthia Webner DNP, RN, CCNS, CCRN-CMC, CHFN Though for Today Mastery is not something that strikes in an instant, like a thunderbolt,

More information

CARDIOCOMP Electrocardiography Primer

CARDIOCOMP Electrocardiography Primer TM CARDIOCOMP Electrocardiography Primer Richmond, VA, USA 23230 CARDIOCOMP 1 Version 8 Electrocardiography Primer Revised: June 2015 Copyright 2015. Phipps & Bird, Inc. All Rights Reserved. This is a

More information

Humans make voluntary decisions to talk, walk, stand up, or sit down. The

Humans make voluntary decisions to talk, walk, stand up, or sit down. The 2 E X E R C I S E Skeletal Muscle Physiology O B J E C T I V E S 1. To define motor unit, twitch, latent period, contraction phase, relaxation phase, threshold, summation, tetanus, fatigue, isometric contraction,

More information

By the end of this lecture, you will be able to: Understand the 12 lead ECG in relation to the coronary circulation and myocardium Perform an ECG

By the end of this lecture, you will be able to: Understand the 12 lead ECG in relation to the coronary circulation and myocardium Perform an ECG By the end of this lecture, you will be able to: Understand the 12 lead ECG in relation to the coronary circulation and myocardium Perform an ECG recording Identify the ECG changes that occur in the presence

More information

Farah Khreisat. Raghad Abu Jebbeh. Faisal Mohammad. 1 P a g e

Farah Khreisat. Raghad Abu Jebbeh. Faisal Mohammad. 1 P a g e 5 Farah Khreisat Raghad Abu Jebbeh Faisal Mohammad 1 P a g e بسم هللا الرحمن الرحيم Hello guys, hope you're doing well, as you've seen in the previous lecture, the Dr started with an extremely important

More information

12 Lead ECG Interpretation: Color Coding for MI s

12 Lead ECG Interpretation: Color Coding for MI s 12 Lead ECG Interpretation: Color Coding for MI s Anna E. Story, RN, MS Director, Continuing Professional Education Critical Care Nurse Online Instructional Designer 2004 Anna Story 1 Objectives review

More information