High ultrasonic power standard by calorimetric method in NMIJ. Takeyoshi Uchida Masahiro Yoshioka
|
|
- Myra McKinney
- 5 years ago
- Views:
Transcription
1 High ultrasonic power standard by calorimetric method in NMIJ Takeyoshi Uchida Masahiro Yoshioka 1
2 Typical method for measuring ultrasonic power Radiation force balance (RFB) method A method measuring a force generated at ultrasound exposure time Calorimetric method A method measuring temperature change generated at ultrasound exposure time 2
3 NMIJ s ultrasonic power standard Calibration method Calibration range Relative expanded uncertainty Radiation force balance (RFB) 1 mw-500 mw (15 MHz-20 MHz) 1 mw-15 W (0.5 MHz-15 MHz) 5%-12% Calorimetry 15 W-100 W (1 MHz-3 MHz) 9% 3
4 Ultrasonic power standard by RFB method Ultrasonic power: 1 mw~15 W Frequency: 0.5 MHz~20 MHz Relative expanded uncertainty: 5%~12% 4
5 Radiation force balance method A force is generated when ultrasound is irradiated to target. Force is measurd as a change in the weight of targets by using electric balance. Electric balance Receiving target Ultrasound wave Ultrasonic transducer 5
6 天秤本体 Electric balance Radiation force balance device Hanging hook Water vessel Transducer Temperature sensor 6
7 Water vessel Ultrasonic transducer Attachment 7
8 Receiving target Reflecting conical target Absorbing target 8
9 Ultrasonic power standard by Calorimetry Ultrasonic power: 15 W~100 W Frequency: 1 MHz~3 MHz Relative expanded uncertainty: 9% 9
10 Background Ultrasonic power A key quantity related to thermal hazards on patients in medical fields Increased use of high power ultrasound for cancer therapy in medical filed It is requirement ultrasonic power standard in high power range for safety and security of patients. 10
11 Problem of RFB method RFB method can accurately measure ultrasonic power up to about 15 W. RFB method is not applicable to determination of high ultrasonic power Problem of receiving target Absorbing target : Thermal damage Reflecting conical target: Acoustic streaming 11
12 Termal damage of absorbing target Absorbing target thermal damage by high power ultrasound Conical target Absorbing target Thermal damaged at 20 W,5.9MHz (non-focused 20mmf beam) 5 12
13 Acoustic streamng on conical target Conical target Effect of acoustic streaming generated by high power ultrasound Because conical target is pushed laterally by acoustic streaming, the radiation force can not be accurately measured by the electronic balance. Conical target Acoustic streaming Transdcuer 13
14 Power [mw] Example of ultrasonic power measurement by absorbing and conical targets absorbing conical 5.9MHz Absorbing Thermal damage at absorbing target Saturation at conical target Conical Oscillator output voltage [mv] Transducer-Target distance 1mm Upside measurement potential at RFB method 15 W~20 W 14
15 Calorimetric method Requirement to develop as an alternative of RFB method for ultrasonic power in high power range. Development of an accurate measurement technique by calorimetric method Measurement device and calculation method Ultrasonic power of up to 100 W Comparison with RFB method, which is primary standard of NMIJ, between 10 W and 15 W 15
16 Conditions required for claorimetric water vessel a) The total ultrasonic energy should be converted to the temperature rise. b) Irradiated ultrasound should not return to the transducer transmission surface to avoid any change in characteristics of transducer. c) Heat loss should be minimized. 16
17 Calorimteric cylindrical water vessel Temperature sensor Air layer Transducer 17
18 Transducer output [V] Hydrophone output [V] Mesurement of ultrasound pulses in calorimetirc water vessel. 0.8 Transducer Pulse/Echo Propagating ultrasound Hydrophone Input signal Time [s] Ultrasound pules could be measured by hydrophone, periodically. Ultrasound waves do not reenter to the transducer transmission surface. 18
19 Heat-shielding material Polyethylene Reduction in uncertainty of measurement by eliminating effect of outdoor air Insulative case State of measurement 19
20 Air-backing ultrasonic transducer The transducer has low heat generation owing to low elastic and dielectric losses. PZT diameter: 20 mm Frequency: 1 MHz Outside diameter of transducer: 40 mm PZT (Fuji ceramics,c-213) 20mmf Ultrasonic transdcuer Ultrasound transducer cross-section view 20
21 Temperature (deg) New calorimetric method by NMIJ The method obtains ultrasonic power from water temperature before and after ultrasound irradiation for eliminating the effect of viscous heating T off T on t b t x t a Ultrasound ON Temperature rise T=T off- T on Ultrasound OFF Time (s) Example of temporal change in water temperature in calorimetric vessel. P T = C p t M x C p : Specific heat capacity M: Water mass T: Temperature rise t x : Ultrasound exposre The difference between the two extrapolated temperatures is defined as T 21
22 Effect of acoustic cavitation bubbles It is necessary to investigate the effect on dissolved oxygen (DO) level of water for precise measurement Experimental condition Frequency: 1 MHz, 2MHz, 3MHz Saturated water: DO level of 8mg/l Degassed water: DO level of 2mg/l Compared range: 10 W-15 W Initial temperature: 22.8 ±0.3 Experimentals are performed under the same conditions expect for DO level of water 22
23 Ultrasonic power (W) Ultrasonic power (W) Ultrasonic power (W) Saturated Degassed Applied voltage (V) Saturated Degassed Applied voltage (V) Saturated Degassed 1 MHz 2 MHz 3 MHz Applied voltage (V) There were significant different between ultrasonic power at all frequencies. Higher ultrasonic power was obtained with saturated water rather than with degassed water. Differences between two samples increased with voltage applied to transducer and decreased with ultrasound frequency. Difference of two samples might be due to acoustic cavitation 23
24 Sonochemiluminescence (SCL) We investigated the occurrence of acoustic cavitation bubbles using SCL. SCL is chemical reaction between luminol anion and reactive oxygen species generated by acoustic cavitation bubbles. SCL is blue glow. It can also be used for the qualitative assessment of amount generated cavitation by measuring SCL intensity. 24
25 Relative brightness value (%) Relative brightness value (%) Relative brightness value (%) Saturated Degassed Applied voltage (V) Saturated Degassed Applied voltage (V) Saturated Degaased 1 MHz 2 MHz 3 MHz Applied voltage (V) Given the positive correlation between SCL intensity and ultrasonic power measured by calorimetry It is quite likely that acoustic cavitation bubbles accounts for the difference in measured ultrasonic power between two sample water. We infer that the effect is due to specific heat capacity of cavitation bubbles It is important to use degassed water for calorimetry. 25
26 Ultrasonic power (W) Ultrasonic power measurement by calorimetric method Calorimetric RFB Applied voltage (V) Calorimetric measurement system constructed by NMIJ could meaure ultrasonic power of up to 100 W. 26
27 Ultrasonic power (W) Comparison of calorimetric method with 15 RFB method Calorimetric RFB Applied voltage (V) The values obtained by the new calorimetric method are within 5% of those obtained by the RFB method. 27
28 Uncertainty budget Uncertainty factor Type 15 W 20 W 30 W 40 W 50 W 60 W 70 W 80 W 90 W 100 W Mass of water V Uncertainty on read of measuring cylinder B Density of waterρ Uncertainty on measuring cylinder B Remaining quality of measuring cylinder B Change of water density by temperature change B Specific heat capacity of water C p Change of specific heat capacity by temperature change B Uncertainty on water temperature measurement Effect on dissolved oxygen level B Uncertainty on thermometer B Variation of extrapolation B Uncertainty on ununiformity of temperature dispersion in water vessel B Uncertainty on outflow and inflow of heat Uncertainty on correction of outflow of heat A Inflow of heat B Measurement time t Uncertainty on read of measurement time B Measurement system Repetition of measurement A Combined standard uncertainty u Expanded uncertainty U(k=2)
29 Summary We explain about radiation force balance method and calorimetry in NMIJ. We created a stable and accurate method for the establishment of a high ultrasonic power standard. We plan to extend ultrasonic power range up to 200 W. 29
NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment
NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment Masahiro Yoshioka National Metrology Institute of Japan (NMIJ) National
More informationSound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound
CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal
More information39th Annual UIA Symposium. Metrology research for External Beam Cancer Therapy. K.-V. Jenderka
39th Annual UIA Symposium Metrology research for External Beam Cancer Therapy K.-V. Jenderka P B Physikalisch-Technische Bundesanstalt Braunschweig Department Sound www.ptb.de Topics Overview Therapeutic
More informationKRISS MEASUREMENT TECHNIQUE OF ULTRASOUND EXPOSURE PARAMETERS AND PREPARATION OF ULTRASOUND POWER KEY COMPARISON IN APMP
KRISS MEASUREMENT TECHNIQUE OF ULTRASOUND EXPOSURE PARAMETERS AND PREPARATION OF ULTRASOUND POWER KEY COMPARISON IN APMP Date: 22 nd Sep. 2015 Yong Tae KIM Center for Medical Metrology, KRISS INTRODUCTION
More informationThe current status of Acoustics, Ultrasound and Vibration measurement standards at NMIJ/ AIST
The current status of Acoustics, Ultrasound and Vibration measurement standards at NMIJ/ AIST National Metrology Institute of Japan (NMIJ) National Institute of Advanced Industrial Science and Technology
More informationCurrent Status of Ultrasonic Device Calibration in China
APMP TCAUV - Changsha Current Status of Ultrasonic Device Calibration in China YANG Ping, NIM E-mail: yangp@nim.ac.cn 1 2 3 5 6 Contents 1 Ultrasound Power 2.1 Ultrasound Field Hydrophone Calibration
More informationUnderwater Acoustic Measurements in Megahertz Frequency Range.
Underwater Acoustic Measurements in Megahertz Frequency Range. Current State and Prospects of Development in Russia Alexander M. Enyakov,, Many medical applications of underwater acoustic measurements
More informationUncertainty Evaluation of Effective Radiation Area of Ultrasound Transducers: Preliminary Results of Inmetro s Laboratory of Ultrasound
Uncertainty Evaluation of Effective Radiation Area of Ultrasound Transducers: Preliminary Results of Inmetro s Laboratory of Ultrasound A. V Alvarenga and R. Costa-Felix Laboratory of Ultrasound - Inmetro,
More informationSupplementary Movie Caption
Supplementary Movie Caption 1. Movie S1. Ultrasound-induced blood focusing in vitro (Fig.2b). 2. Movie S2. Acoustic canalization of blood flow in the gap between two capillaries (Fig. 2d). 3. Movie S3.
More informationUltrasonic Transducer. Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film
Ultrasonic Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film Ultrasonic Technology is gentle both to the Earth and humankind Diminution of negative influences to the environment,
More informationIEC 87: Ultrasonics Overview update July 2002
- 1 - CCAUV/02-07 IEC 87: Ultrasonics Overview update July 2002 Roy C Preston Secretary IEC TC 87: Ultrasonics British Electrotechnical Committee, British Standards Institution, 389 Chiswick High Road,
More information10ème Congrès Français d'acoustique Lyon, Avril 2010
1ème Congrès Français d'acoustique Lyon, 12-16 Avril 21 Stabilization of Cavitation Zone for Therapeutic Applications A. Moussatov, C. Lafon, J.-L. Mestas, S. Chesnais, J. Ngo, L. Somaglino, J.-Y. Chapelon
More informationAnnular Array Transducer and Matched Amplifier for Therapeutic Ultrasound
ARCHIVES OF ACOUSTICS 35, 4, 653 660 (2010) DOI: 10.2478/v10168-010-0049-6 Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound Wojciech SECOMSKI, Andrzej NOWICKI, Janusz WÓJCIK, Marcin
More informationSupplement (videos)
Supplement (videos) Ruben s tube (sound): http://www.youtube.com/watch?v=gpcquuwqayw Doppler US (diagnostic use): http://www.youtube.com/watch?v=fgxzg-j_hfw http://www.youtube.com/watch?v=upsmenyoju8 High
More informationNon-Destructive Inspection of Composite Wrapped Thick-Wall Cylinders
Non-Destructive Inspection of Composite Wrapped Thick-Wall Cylinders Jikai Du, John Feldhacker, Christopher Jerred and Fereidoon Delfanian May 17-19, 2010 Joint Armaments Conference, Exhibition and Firing
More informationThe current status of Acoustics, Ultrasound and Vibration measurement standards at NMIJ/ AIST
The current status of Acoustics, Ultrasound and Vibration measurement standards at NMIJ/ AIST National Metrology Institute of Japan (NMIJ) National Institute of Advanced Industrial Science and Technology
More informationULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah
ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),
More information2/11/2009. Evidence for the Molecular, Cellular and Physiological Effects of Microstreaming and Cavitation at MHz and khz Ultrasound Frequencies
Evidence for the Molecular, Cellular and Physiological Effects of Microstreaming and Cavitation at MHz and khz Ultrasound Frequencies ISEPA : Las Vegas : Feb 2009 Professor Tim Watson University of Hertfordshire
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-2-37 2001 AMENDMENT 1 2004-08 Amendment 1 Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring
More informationVISUALIZATION OF TRANSDUCER-PRODUCED SOUND FIELDS IN SOLIDS
VISUALIZATION OF TRANSDUCER-PRODUCED SOUND FIELDS IN SOLIDS Wolfgang Sachse* Department of Theoretical and Applied Mechanics Cornell University, Ithaca, New York - 14853 ABSTRACT Broadband ultrasonic pulses
More informationThe Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics
The Physics of Ultrasound Pipe Organ 10-8000 Emission Dog 452-1080 Man 85-1100 Spectrum Bat 10,000-120,000 Porpoise 7000-120,000 Claus G. Roehrborn Professor and Chairman 10 20 Cycles per second Reception
More informationCharacterization of Acoustic Cavitation from a Megasonic Nozzle Transducer for Photomask Cleaning
Putting Confidence in Ultrasound Characterization of Acoustic Cavitation from a Megasonic Nozzle Transducer for Photomask Cleaning Nicolas Candia (1), Claudio Zanelli (2), Johann Brunner (3), Joachim Straka
More informationANALYSIS OF ULTRASOUND FIELD PARAMETERS DURING SONICATION EXPERIMENTS IN VITRO INFLUENCE OF LABORATORY GLASS AND PLASTICS
ANALYSIS OF ULTRASOUND FIELD PARAMETERS DURING SONICATION EXPERIMENTS IN VITRO INFLUENCE OF LABORATORY GLASS AND PLASTICS Martin Snehota 1,2, Jaromir Vachutka 1, Ladislav Dolezal 1, Hana Kolarova 1,2,
More informationUltrasound Measurements and Non-destructive Testing Educational Laboratory
Session 3548 Ultrasound Measurements and Non-destructive Testing Educational Laboratory Vladimir Genis, Horacio Sosa Goodwin College of Professional Studies, Drexel University, Philadelphia, 19104 Emil
More informationStress Wave Focusing Transducers
UCRL-K-130697 PREPRINT Stress Wave Focusing Transducers Steven R. Visuri, Richard A. London, Luiz Da Silva This paper was prepared for submittal to Optical Society of America, Spring Topical Meetings Orlando,
More informationMEASUREMENTS OF ULTRASONIC CAVITATION IN LIQUIDS AND TISSUE MIMICKING MATERIALS
MEASUREMENTS OF ULTRASONIC CAVITATION IN LIQUIDS AND TISSUE MIMICKING MATERIALS Adriano Troia, Pietro De Nicola Metrology for Quality of Life, National Institute of Metrological Research, strada delle
More informationLesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.
Lesson 03: Sound Wave Propagation and Reflection This lesson contains 15 slides plus 14 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 8 through 14 in
More informationThe Yesterday, Today and Tomorrow of Focused Ultrasound Surgery CHONGQING HAIFU (HIFU) TECHNOLOGY CO., LTD.
The Yesterday, Today and Tomorrow of Focused Ultrasound Surgery 1 Terms and Principle Evolution of surgery in the last 200 years Open Surgery Laparoscopic Surgery Focused Ultrasound Surgery 2 Terms and
More informationSam Howard PhD Tests & measurements, Onda Corporation. Timothy Demchek PhD - Human tissue heating, Indian State University
Should you be using Therapeutic Ultrasound and how can we improve our research Stephen J. Straub PhD, ATC Thanks Lenn Johns PhD - cell biologist, QU Sam Howard PhD Tests & measurements, Onda Corporation
More informationEssentials of Medical Ultrasound
Essentials of Medical Ultrasound Medical Methods Essentials of Medical Ultrasound: A Practical Introduction to the Principles, Techniques, and Biomedical Applications, Edited by Michael H. Repacholi and
More informationHeat Drain Device for Ultrasound Imaging Probes ESAOTE SPA
Heat Drain Device for Ultrasound Imaging Probes ESAOTE SPA L. Spicci, G. Vigna TRANSDUCER STRUCTURE OVERVIEW Finished probe Complete transducer Silicon lens Filling resin INTRODUCTION Electrical driving
More informationJAMES R. KNIGHT
JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops
More informationApplication of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System
Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of
More informationFeng Xiujuan National Institute of Metrology (NIM),China
The acoustic calibration service in transportation at NIM Feng Xiujuan National Institute of Metrology (NIM),China 1. Calibration requirements 2. Calibration service at NIM 2.1 Microphone 2.2 Ultrasonic
More informationSCIENCE (REVISED SYLLABUS) HIGHER LEVEL
J.37 PRE-JUNIOR CERTIFICATE EXAMINATION, 2010 SCIENCE (REVISED SYLLABUS) HIGHER LEVEL TIME : 2 HOURS INSTRUCTIONS 1. Write your name, school name and teacher s name in the boxes provided on this page.
More informationAutomatic quality control processing for detection of element/channel dropout in ultrasound transducers
Automatic quality control processing for detection of element/channel dropout in ultrasound transducers Paul L. Carson, Ph.D. University of Michigan, Department of Radiology, Ann Arbor, MI, USA Acknowledgements
More informationEmployer s Unit of Competence Ultrasonic testing of materials, products and plant
Employer s Unit of Competence Ultrasonic testing of materials, products and plant Document: AA064 Issue 2 May 2016 Image - if cover page required Supported by lead employer Overview This standard identifies
More informationPhysical Principles of Ultrasound
Physical Principles of Ultrasound Grateful appreciation to Richard A. Lopchinsky, MD, FACS and Nancy H. Van Name, RDMS, RTR, and MarleneKattaron, RDMS 2000 UIC All Rights Reserved. Course Objectives Identify
More informationMatthew Eschbach Robert Sterling Nesbitt Joseph Ouellette Sarah Roberge
Matthew Eschbach Robert Sterling Nesbitt Joseph Ouellette Sarah Roberge Bone Remodeling Osteocytes act as mechanoreceptors Biochemical reactions Osteoclasts resorb bone Osteoblasts rebuild bone Edentulism
More informationDEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS
DEVELOPMENT OF ULTRASONIC TESTING TECHNIQUE TO INSPECT CONTAINMENT LINERS EMBEDDED IN CONCRETE ON NUCLEAR POWER PLANTS H. Ishida 1, Y. Kurozumi 1, and Y. Kaneshima 2 1 Institute of Nuclear Safety System,
More informationTHE EFFECT OF ULTRASONIC IRRADIATION ON THE SURVIVAL OF DAPHNIA MAGNA
Exp. Biol. (1968), 49, 61-70 With 7 text-figura Printed in Great Britain 6l THE EFFECT OF ULTRASONIC IRRADIATION ON THE SURVIVAL OF DAPHNIA MAGNA BY P. N. T. WELLS Department of Zoology, University of
More informationUltrasonically Assisted Extraction (UAE) of Natural Products Some Guidelines for Good Practice and Reporting
Ultrasonically Assisted Extraction (UAE) of Natural Products Some Guidelines for Good Practice and Reporting Vinatoru, M. Postprint (accepted) version deposited in CURVE November 2014 Original citation:
More informationCan you believe that the ultrasonic waves are used for cleaning purpose???
Introduction (Ultrasonic Cleaning Unit) Can you believe that the ultrasonic waves are used for cleaning purpose??? Learning Objectives On completion of this chapter you will be able to: 1. Describe the
More informationCLOUD CAVITATION CONTROL FOR A THERAPEUTIC ULTRASOUND APPLICATION
CAV03-OS-2-1-008 Fifth International Symposium on Cavitation (CAV2003) Osaka, Japan, November 1-4, 2003 CLOUD CAVITATION CONTROL FOR A THERAPEUTIC ULTRASOUND APPLICATION Teiichiro Ikeda Department of Mechanical
More informationSection 1. Executive Summary
Section 1 Executive Summary The toolkit was written by public health scientists and is intended as a background document, current to 2012, to assist medical health officers and environmental health officers
More informationDescriptions of NDT Projects Fall 2004 October 31, 2004
Descriptions of NDT Projects Fall 2004 October 31, 2004 Introduction There are two separate NDT labs in Magister: ULTRA for ultrasound and EDDY for eddy current. Both labs are equipped with mechanical
More informationCharacteristics of Liquids Atomization Using Surface Acoustic Wave
Characteristics of Liquids Atomization Using Surface Acoustic Wave Minoru Kurosawa, Akira Futami, and Toshiro Higuchi Dept. of Precision Machinery Engineering, The University of Tokyo, Hongo, Bunkyo-ku,
More informationCOURSE DESCRIPTION FOR NONDESTRUCTIVE TESTING
COURSE DESCRIPTION FOR NONDESTRUCTIVE TESTING TULSA, OKLAHOMA INTRODUCTION TO NONDESTRUCTIVE TESTING QCT1817 COURSE DESCRIPTION: In this course students will learn about materials and processes, find basic
More informationTECHNICAL SPECIFICATION
TECHNICAL SPECIFICATION IEC TS 62462 First edition 2007-05 Ultrasonics Output test Guide for the maintenance of ultrasound physiotherapy systems Reference number IEC/TS 62462:2007(E) THIS PUBLICATION IS
More informationULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING
ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING H. J. Shin 1, Y. H. Jang 1, J. R. Kwan 2, H. D. Lee 3 1 INDE System Co., Ltd., Suwon, Kyunggi-do, 440-746,
More informationEffects of Magnetic field on Voltage response of PMT
Effects of Magnetic field on Voltage response of PMT Supervised by: Prof. Tsutomu Mibe Nitin, Parth, Sumukh KEK 31 st May 2017 Nitin, Parth, Sumukh (KEK) Effect of Magnetic field on PMT 31 st May 2017
More informationAn Energy Efficient Sensor for Thyroid Monitoring through the IoT
An Energy Efficient Sensor for Thyroid Monitoring through the IoT P. Sundaravadivel 1, S. P. Mohanty 2, E. Kougianos 3, U. Albalawi 4. NanoSystem Design Laboratory (NSDL, http://nsdl.cse.unt.edu) University
More informationPrimary level ultrasonic power measurement by radiation force balance (RFB) developed at National Institute of Standards (NIS) Egypt
IOSR Journal of Engineering (IOSRJEN) ISSN (e): 2250-3021, ISSN (p): 2278-8719 Vol. 07, Issue 10 (October. 2017), V1 PP 42-49 www.iosrjen.org Primary level ultrasonic power measurement by radiation force
More informationPhotoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar
Photoacoustic Imaging 1 Photoacoustic Imaging and Therapy in Biomedicine Nicholas Tobey and Grace Yook Optical Engineering Dr. Kasra Daneshvar July 16, 2010 Photoacoustic Imaging 2 Abstract When a pulsed
More informationAn Overview of Ultrasound Testing For Lesion Detection in Human Kidney
Journal of Tomography System & Sensors Application Vol.1, Issue 1, June 2018 An Overview of Ultrasound Testing For Lesion Detection in Human Kidney Aina Fadhilah Abd Rahim 1, Zawin Najah Abd Halim 1, Jaysuman
More informationA PRELIMINARY ANALYSIS OF POSSIBILITIES OF COMPENSATING FAULTS OF LASER AND ULTRASONIC TECHNOLOGIES IN SURGERY. Tadeusz GUDRA, Sylwia MUC
ARCHIVES OF ACOUSTICS 32, 4 (Supplement), 117 122 (2007) A PRELIMINARY ANALYSIS OF POSSIBILITIES OF COMPENSATING FAULTS OF LASER AND ULTRASONIC TECHNOLOGIES IN SURGERY Tadeusz GUDRA, Sylwia MUC Wrocław
More informationULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS
1 OB/Gyn (Core) Ultrasound Principles of Ultrasound Physics and Instrumentation Nathan Pinkney, BS, CDOS Philadelphia College of Osteopathic Medicine 2016 ULTRASOUND CATEGORIES OF SOUND INFRASOUND = below
More informationEE 791 Lecture 2 Jan 19, 2015
EE 791 Lecture 2 Jan 19, 2015 Action Potential Conduction And Neural Organization EE 791-Lecture 2 1 Core-conductor model: In the core-conductor model we approximate an axon or a segment of a dendrite
More informationDevelopments in Ultrasonic Inspection II
Developments in Ultrasonic Inspection II An Ultrasonic Technique for the Testing of Plates Embedded in Concrete with Synthesis of Signals from a Multi-element Probe H. Ishida, Y. Kurozumi, Institute of
More informationMATCHING LAYER DESIGN OF AN ULTRASONIC TRANS- DUCER FOR WIRELESS POWER TRANSFER SYSTEM
MATCHIN LAYE DESIN OF AN ULTASONIC TANS- DUCE FO WIELESS POWE TANSFE SYSTEM unn Hang Electronics and Telecommunications esearch Institute, Multidisciplinary Sensor esearch roup, Daejeon, South Korea email:
More informationIntroduction to Biomedical Imaging
Alejandro Frangi, PhD Computational Imaging Lab Department of Information & Communication Technology Pompeu Fabra University www.cilab.upf.edu Basic principles. Comparison to X-rays Ultrasound > 20kHz
More informationInvestigation of Geometric Effect on the Ultrasonic Processing of Liquids
Clemson University TigerPrints All Theses Theses 5-2016 Investigation of Geometric Effect on the Ultrasonic Processing of Liquids Pravan Pasumarthi Clemson University, ppasuma@clemson.edu Follow this and
More informationFocused Ultrasound Body-shaping system Beijing Sanhe Beauty S & T Co.,Ltd
Focused Ultrasound Body-shaping system Beijing Sanhe Beauty S & T Co.,Ltd Copyright: www.sanhemedical.com Non-invasive Fat Cell Elimination for Body Shaping Salushape is a non invasive body shaping system
More informationBasic Physics of Ultrasound and Knobology
WELCOME TO UTMB Basic Physics of Ultrasound and Knobology By Daneshvari Solanki, FRCA Laura B. McDaniel Distinguished Professor Anesthesiology and Pain Medicine University of Texas Medical Branch Galveston,
More informationReport originating from the Frauenhofer-Institute of Construction Physics
IBP Report originating from the Frauenhofer-Institute of Construction Physics Abridged version from IBP-Report BS 113/84 An examination of the principle of operation behind the sonic wave device SK1 and
More informationUltrasonic transducers
Woodhead Publishing Series in Electronic and Optical Materials; Number 29 Ultrasonic transducers Materials and design for sensors, actuators and medical applications Edited by K. Nakamura Oxford Cambridge
More informationBasic Physics of Ultrasound in Transesophageal Echocardiography
SPECIAL ARTICLE IJUTPC Basic Physics of Ultrasound in Transesophageal Echocardiography Basic Physics of Ultrasound in Transesophageal Echocardiography 1 Mary Korula, 2 Ravi Hebballi 1 Senior Consultant,
More informationProceedings of Meetings on Acoustics
Proceedings of Meetings on Acoustics Volume 19, 2013 http://acousticalsociety.org/ ICA 2013 Montreal Montreal, Canada 2-7 June 2013 Biomedical Acoustics Session 1pBAb: Ultrasound Contrast Agents and Passive
More informationComparisons of Three Analytical Techniques for Measuring Absolute and Relative Temperature Changes During Ultrasound Treatment
Western Michigan University ScholarWorks at WMU Dissertations Graduate College 4-2011 Comparisons of Three Analytical Techniques for Measuring Absolute and Relative Temperature Changes During Ultrasound
More informationBasic Ultrasound Physics Board Review Questions
Basic Ultrasound Physics Board Review Questions Sidney K. Edelman, PhD ESP Ultrasound The Woodlands, TX Question 1 What is the wavelength of 2 MHz sound in soft tissue? 1. 1.54 mm 2. 0.75 mm 3. 0.75 cm
More informationTECHNICAL SPECIFICATION
TECHNICAL SPECIFICATION IEC TS 62462 First edition 2007-05 Ultrasonics Output test Guide for the maintenance of ultrasound physiotherapy systems Commission Electrotechnique Internationale International
More informationHigh power density prototype for high precision transcranial therapy
High power density prototype for high precision transcranial therapy M. Pernot a, R. Berriet b, J-F. Aubry a, O. Le Baron b, M. Tanter a, G. Fleury b, L. Chupin b, L. Gallet b, and M. Fink a a Laboratoire
More informationTissue Strain Analytics Virtual Touch Tissue Imaging and Quantification
Whitepaper Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification ACUSON S2000 Ultrasound System Answers for life. Page 1 Tissue Strain Analytics: Virtual Touch Tissue Imaging and Quantification
More informationHiFFUT A New Class of Transducer
HiFFUT A New Class of Transducer Project Meeting April 2017 Dr Andrew Feeney Study of the electro-mechanical behaviour of flexural ultrasonic transducers This work will help us to design HiFFUTs for particular
More informationUS - Ultrasonic Systems. Instrumentation and Applications
Coordinating unit: Teaching unit: Academic year: Degree: ECTS credits: 2017 230 - ETSETB - Barcelona School of Telecommunications Engineering 710 - EEL - Department of Electronic Engineering ERASMUS MUNDUS
More informationa series of in vitro experiments was conducted using monooctanoin (MO) and methyl tertiary butyl
A LOW FREQUENCY (220 khz) ULTRASOUND SYSTEM FOR ENHANCEMENT OF GALLSTONE DISSOLUTION N.T. Sanghvi, EJ. Fry, B.T. Burney,* S.L. Griffith, S.M. Wessling, T.D.Franklin, Jr. and R.E Morris+ Indianapolis Center
More informationFOCUSED ULTRASOUND STIMULATION OF THE PERIPHERAL NERVOUS SYSTEM: PHYSICAL BASIS AND PRACTICAL APPLICATIONS (REVIEW)
International Journal of Modern Physics: Advances in Theory and Application Vol. 1, Issue 1, 016, Pages 45-118 Published Online on September 4, 016 016 Jyoti Academic Press http://jyotiacademicpress.org
More informationProgress in Development of HIFU CMUTs for use under MR-guidance
Progress in Development of HIFU CMUTs for use under MR-guidance Serena H. Wong*, Ronald D. Watkins, Mario Kupnik*, Kim Butts Pauly, and B.T. Khuri-Yakub* Stanford University 450 Via Palou, Stanford, CA
More informationGUIDANCE NOTE. Application of MEMS Ultrasonic Transducers to Flow Measurement.
GUIDANCE NOTE Application of MEMS Ultrasonic Transducers www.tuvnel.com Application of MEMS Ultrasonic Transducers Introduction This Guidance Note provides a review of the key issues associated with the
More informationDEVELOPMENT OF ELECTROMAGNETIC ACOUSTIC TRANSDUCER (EMAT) PHASED ARRAYS FOR SFR INSPECTION
DEVELOPMENT OF ELECTROMAGNETIC ACOUSTIC TRANSDUCER (EMAT) PHASED ARRAYS FOR SFR INSPECTION QNDE 2013 Florian LE BOURDAIS and Benoit MARCHAND CEA LIST, Centre de Saclay F-91191 Gif-sur-Yvette, France OUTLINE
More informationFocused Ultrasound Body shaping system
Focused Ultrasound Body shaping system Beijing Sanhe Beauty S & T Co.,Ltd Copyright: www.sanhebeauty.com Fat Cell Destruction for Non-Invasive Body Sahping The Salushape system is a non invasive body shaping
More informationUltrasound in Medicine
Ultrasound in Medicine Experimental Equipment for Medical Education Universities Colleges Medical Schools Medical and Med-Technical Training Education can befun! WELCOME TO GAMPT Devices and accessories
More informationUltrasound Physics & Terminology
Ultrasound Physics & Terminology This module includes the following: Basic physics terms Basic principles of ultrasound Ultrasound terminology and terms Common artifacts seen Doppler principles Terms for
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationExam Practice Guide. Units 1 & 2 Physics: Detailed Study 5 - Investigations: Medical physics Examination Questions
Exam Practice Guide Units 1 & 2 Physics: Detailed Study 5 - Investigations: Medical physics Examination Questions Key Features: 22 original examination style questions on all examinable topics. Full solutions
More informationA computer-aided diagnostic scheme for classification of clustered microcalcifications in mammograms
ABSTRACTS OF RECENT Ph.D. THESES PERTINENT TO MEDICAL PHYSICS Appropriate thesis abstracts for this section should be submitted to Professor John Cameron, c/o Medical Physics, One Physics Ellipse, College
More informationFT-302 Force Transducer
Technical Note FT-302 LabScribe is a trademark of 2015 Overview The FT-302 is a high-sensitivity dual-range research grade force transducer designed to measure forces in the 0.005 to 10 gram and 0 to 100
More informationDevelopment of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion
Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods
More informationTherapeutic Ultrasound
THERAPEUTIC ULTRASOUND Therapeutic is a form of MECHANICAL energy Mechanical vibration at increasing frequencies is known as sound energy. Below 16Hz, these vibrations are not recognizable as sound The
More informationA Methodology for Ultrasound Product Development Applications in HIFU and High Frequency
A Methodology for Ultrasound Product Development Applications in HIFU and High Frequency Claudio I. Zanelli Samuel M. Howard Ultrasonic Industry Association 35 th Annual Symposium - San Diego, California
More informationPresents: Towards an implantable continuous glucose sensor. by Erik Johannessen. Illustration: Flemming Sveen
Presents: Towards an implantable continuous glucose sensor. by Erik Johannessen Illustration: Flemming Sveen Π = i c RT, i = 1, n 2, 3... n Π = osmotic pressure (Pa) i = Van t Hoff factor c n = molar concentration
More informationLaser enhanced high-intensity focused ultrasound thrombolysis: An in vitro study
Laser enhanced high-intensity focused ultrasound thrombolysis: An in vitro study Huizhong Cui and Xinmai Yang a) KU Bioengineering Research Center, Department of Mechanical Engineering, University of Kansas,
More informationWhite Paper. iqprobes
White Paper iqprobes Transducers are the core of Ultrasound technology. Integrating physics, electronics and geometrics in their design is the greatest engineering challenge of the signal processing chain.
More informationNeutron Measurements for Intensity Modulated Radiation Therapy
SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,
More informationTerminology Tissue Appearance
By Marc Nielsen, MD Advantages/Disadvantages Generation of Image Ultrasound Machine/Transducer selection Modes of Ultrasound Terminology Tissue Appearance Scanning Technique Real-time Portable No ionizing
More informationEstimating dynamic changes of tissue attenuation coefficient during high-intensity focused ultrasound treatment
Rahimian and Tavakkoli Journal of Therapeutic Ultrasound 2013, 1:14 RESEARCH Open Access Estimating dynamic changes of tissue attenuation coefficient during high-intensity focused ultrasound treatment
More informationMedical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.
Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation
More informationACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications
ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications International Workshop on Imaging NDE April 2007 IGCAR, kalpakkam, India Dr D S Dulay NDT Consultants Limited Middlemarch House, Siskin
More informationFlaw Assessment Using Shear wave Phased array Ultrasonic Transducer
18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer Byungsik YOON AUTHOR 1, Hee-Jong LEE CO-AUTHOR
More informationAvailable online at ScienceDirect. Physics Procedia 87 (2016 ) 35 41
Available online at www.sciencedirect.com ScienceDirect Physics Procedia 87 (2016 ) 35 41 44th Annual Symposium of the Ultrasonic Industry Association, UIA 44th Symposium, 20-22 April 2015, Washington,
More information