Test Procedure for Piezoelectric Accelerometer Ice Thickness Measurement

Size: px
Start display at page:

Download "Test Procedure for Piezoelectric Accelerometer Ice Thickness Measurement"

Transcription

1 1 of 10 Test Procedure for Piezoelectric Accelerometer Ice Thickness Measurement Revision Date Reviewed By Approved By Summary: This document outlines the function test procedure for a piezoelectric accelerometer when used to determine the thickness of ice.

2 2 of 10 Table of Contents 1.0 Background Introduction Ice/Air Interface Test Assembly Ice/Air Interface Ice/Air Interface Ice/Air Interface Ice/Water Interface Test Assembly Ice/Water Interface Ice/Water Interface Ice/Water Interface Notes/Punch List Conclusions and Recommendations 10 Figures and Tables Figure 1: Electrical Component Arrangement 7 Figure 2: Ice Coupon on Testing Stand 8 Figure 3: Ice Coupon in Testing Basin 8 Table 1: Test Results 7

3 3 of 10 Equipment to be Tested: Chosen Transducer/Receiver for TekPoint s Ice Thickness Measuring Device Test Summary: This test will verify the function of a piezoelectric accelerometer when used to measure ice thickness. The device will be coupled with an ice coupon. Generated voltage peaks at specific time intervals will be used to calculate ice thickness values. These values will then be recorded and compared with actual ice thickness measurements. Testing Objectives: Verify the function of a piezoelectric accelerometer when used to determine ice thickness Equipment Required for Test Piezoelectric Accelerometer Oscilloscope Charge Amplifier DC Power Supply Connecting Cables/Wires Ice Coupon Testing Stand Testing Basin Salt Water Solution Small Impact Hammer Preconditions of Test: Oscilloscope has been adjusted Ice coupons have been cut to have uniform thicknesses of 2, 4, and 6

4 4 of Background A piezoelectric accelerometer is an instrument that utilizes the piezoelectric effect of certain materials to measure dynamic changes in mechanical variables. As with all transducers, piezoelectric accelerometers convert one form of energy into another and provide an electrical signal in response to a quantity, property, or condition that is being measured. The particular piezoelectric accelerometer used in this test is designed to measure sound vibrations in solids. Determining the time between an initial sound vibration and its reflection through a solid makes it possible to calculate material thickness. A charge amplifier will be used to amplify the accelerometer output and an oscilloscope will be used to display voltage change corresponding to the received sound waves. The speed of sound through ice can be calculated using: Where: E = Young's Modulus of ice (6 GPa) v = Piosson's Ratio of ice (.28) p = Density of ice (917 Kg/m3) CL = Longitudinal wave speed, calculated to be 2220 m/s, which agrees with published values of 1928 m/s (published values will be used for calculations) The ice thickness can be calculated using: Where: & = Minimum thickness of ice sheet (5 cm, from Design Parameters) = Maximum thickness of ice sheet (60 cm, from Design Parameters) = Longitudinal wave speed (1928 m/s) = Minimum response time of signal(calculated to be approximately 50micro s) = Maximum response time of signal(calculated to be approximately 625 micro s)

5 5 of Introduction This document defines the test procedure for evaluating the operation of a piezoelectric accelerometer when used to determine ice thickness. 3.0 Ice/Air Interface Test 3.1 Assembly 1 Connect all electrical components accordingly. Refer to Figure Ice/Air Interface 1 Place and center a 2 ice coupon on the testing stand. Refer to Figure 2. 2 Spray the ice coupons surface with the salt water solution 3 Place the piezoelectric accelerometer in the center of the ice coupon 4 Lightly tap the accelerometer with the small impact hammer to initiate sound vibrations 5 Collect data from oscilloscope and calculate ice thickness 6 Record collected data in Table 1 7 Remove transducer/receiver from ice coupon 8 Remove ice coupon from testing stand Ice/Air Interface 1 Place and center a 4 ice coupon on the testing stand. Refer to Figure 2 2 Repeat steps 2 through 8 in section 3.2

6 6 of Ice/Air Interface 1 Place and center a 6 ice coupon on the testing stand. Refer to Figure 2 2 Repeat steps 2 through 8 from section Ice/Water Interface Test 4.1 Assembly 1 Connect all electrical components accordingly. Refer to Figure 1 2 Fill the testing basin to the indicated level with tap water Ice/Water Interface 1 Place and center a 2 ice coupon in the testing basin. Refer to Figure 3 2 Spray the ice coupons surface with the salt water solution 3 Place the piezoelectric accelerometer in the center of the ice coupon 4 Lightly tap the accelerometer with the small impact hammer to initiate sound vibrations 5 Collect data from oscilloscope and calculate ice thickness 6 Record collected data in Table 1 7 Remove piezoelectric accelerometer from ice coupon 8 Remove ice coupon from testing basin Ice/Water Interface 1 Place and center a 4 ice coupon in the testing basin. Refer to Figure 3 2 Repeat steps 2 through 8 in section 4.2

7 7 of Ice/Water Interface 1 Place and center a 6 ice coupon in the testing basin. Refer to Figure 3 2 Repeat steps 2 through 8 in section 4.2 Ice/Air Coupon Thickness Recorded Thickness Actual Thickness Ice/Water Coupon Thickness Recorded Thickness Actual Thickness Table 1: Test Results Figure 1: Electrical Component Arrangement

8 8 of 10 Figure 2: Ice Coupon on Testing Stand Figure 3: Ice Coupon on Testing Basin

9 9 of Notes/Punch List

10 10 of Recommendations and Conclusions This test should provide a sufficient amount of data to determine weather or not the chosen technology is suitable for measuring ice thickness. Electrical components have been selected according to piezoelectric transducer type. The oscilloscope will be adjusted accordingly as well. Data collected from the success or failure of testing a piezoelectric accelerometer should not solely be used to evaluate accelerometer operation. Data collected from the test, and observations made during testing, can provide insight related to the feasibility of using sound to determine ice thickness. This procedure will also provide basic insight into how the prototype will operate and what changes could be made to further improve our concept.

Transducer/Receiver Function Test Procedure

Transducer/Receiver Function Test Procedure 1 of 10 Transducer/Receiver Function Test Procedure Revision Date Reviewed By Approved By Summary: This document outlines the function test procedure for the chosen transducer/receiver to be used in TekPoint

More information

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal

More information

Characteristics of Liquids Atomization Using Surface Acoustic Wave

Characteristics of Liquids Atomization Using Surface Acoustic Wave Characteristics of Liquids Atomization Using Surface Acoustic Wave Minoru Kurosawa, Akira Futami, and Toshiro Higuchi Dept. of Precision Machinery Engineering, The University of Tokyo, Hongo, Bunkyo-ku,

More information

STUDY ON STRUCTURE OF PZT PILES BASED TRANSDUCER FOR HARVESTING ENERGY FROM ASPHALT PAVEMENT

STUDY ON STRUCTURE OF PZT PILES BASED TRANSDUCER FOR HARVESTING ENERGY FROM ASPHALT PAVEMENT STUDY ON STRUCTURE OF PZT PILES BASED TRANSDUCER FOR HARVESTING ENERGY FROM ASPHALT PAVEMENT Hongduo Zhao * Professor, Key Laboratory of Road and Traffic Engineering of Ministry of Education, Tongji University,

More information

Descriptions of NDT Projects Fall 2004 October 31, 2004

Descriptions of NDT Projects Fall 2004 October 31, 2004 Descriptions of NDT Projects Fall 2004 October 31, 2004 Introduction There are two separate NDT labs in Magister: ULTRA for ultrasound and EDDY for eddy current. Both labs are equipped with mechanical

More information

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of

More information

Technical Tip: How to Connect and Troubleshoot Transducer Feedback. Product(s): iqpump1000 and iqpump Micro Drives

Technical Tip: How to Connect and Troubleshoot Transducer Feedback. Product(s): iqpump1000 and iqpump Micro Drives This Technical Tip assists users in applying Yaskawa iqpump1000 and iqpump Micro products to pump systems and provides useful information on: running without a transducer fixing Fdb-L or Feedback Loss

More information

Ultrasound Measurements and Non-destructive Testing Educational Laboratory

Ultrasound Measurements and Non-destructive Testing Educational Laboratory Session 3548 Ultrasound Measurements and Non-destructive Testing Educational Laboratory Vladimir Genis, Horacio Sosa Goodwin College of Professional Studies, Drexel University, Philadelphia, 19104 Emil

More information

GUIDANCE NOTE. Application of MEMS Ultrasonic Transducers to Flow Measurement.

GUIDANCE NOTE. Application of MEMS Ultrasonic Transducers to Flow Measurement. GUIDANCE NOTE Application of MEMS Ultrasonic Transducers www.tuvnel.com Application of MEMS Ultrasonic Transducers Introduction This Guidance Note provides a review of the key issues associated with the

More information

Ultrasonic Transducer. Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film

Ultrasonic Transducer. Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film Ultrasonic Piezoelectric ceramics / Piezoelectric polymer film / Piezoelectric thin film Ultrasonic Technology is gentle both to the Earth and humankind Diminution of negative influences to the environment,

More information

NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment

NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment Masahiro Yoshioka National Metrology Institute of Japan (NMIJ) National

More information

PHONOCARDIOGRAPHY (PCG)

PHONOCARDIOGRAPHY (PCG) PHONOCARDIOGRAPHY (PCG) The technique of listening to sounds produced by the organs and vessels of the body is called auscultation. The areas at which the heart sounds are heard better are called auscultation

More information

Producing and Detecting Sound

Producing and Detecting Sound Sound Producing and Detecting Sound What do you think? Read the two statements below and decide whether you agree or disagree with them. Place an A in the Before column if you agree with the statement

More information

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics The Physics of Ultrasound Pipe Organ 10-8000 Emission Dog 452-1080 Man 85-1100 Spectrum Bat 10,000-120,000 Porpoise 7000-120,000 Claus G. Roehrborn Professor and Chairman 10 20 Cycles per second Reception

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) The story so far The heart pumps blood around the body. It has four chambers which contact

More information

HiFFUT A New Class of Transducer

HiFFUT A New Class of Transducer HiFFUT A New Class of Transducer Project Meeting April 2017 Dr Andrew Feeney Study of the electro-mechanical behaviour of flexural ultrasonic transducers This work will help us to design HiFFUTs for particular

More information

Heat Drain Device for Ultrasound Imaging Probes ESAOTE SPA

Heat Drain Device for Ultrasound Imaging Probes ESAOTE SPA Heat Drain Device for Ultrasound Imaging Probes ESAOTE SPA L. Spicci, G. Vigna TRANSDUCER STRUCTURE OVERVIEW Finished probe Complete transducer Silicon lens Filling resin INTRODUCTION Electrical driving

More information

HiFFUT A New Class of Transducer

HiFFUT A New Class of Transducer HiFFUT A New Class of Transducer Kick-Off Meeting Dr Andrew Feeney, Postdoctoral Research Fellow My Background I obtained a Master s degree in Mechanical Engineering from the University of Glasgow in 2010

More information

Preamble (disclaimer)

Preamble (disclaimer) Preamble (disclaimer) PHYSICS AND PRINCIPLES OF HEAD/NECK ULTRASOUND Joseph C. Sniezek, MD FACS LTC, MC, USA Otolaryngology/H&N Surgery Tripler Army Medical Center 1. I am not a physicist 2. ACS has recommended

More information

Ultrasonic Condition Monitoring MORE THAN JUST A LEAK DETECTOR

Ultrasonic Condition Monitoring MORE THAN JUST A LEAK DETECTOR Innovative Sound Solutions TM Ultrasonic Condition Monitoring MORE THAN JUST A LEAK DETECTOR Presented by Benjamin Fried May, 2006 CTRL Systems, Inc. 1004 Littlestown Pike, Suite H Westminster, MD 21157

More information

Exam Practice Guide. Units 1 & 2 Physics: Detailed Study 5 - Investigations: Medical physics Examination Questions

Exam Practice Guide. Units 1 & 2 Physics: Detailed Study 5 - Investigations: Medical physics Examination Questions Exam Practice Guide Units 1 & 2 Physics: Detailed Study 5 - Investigations: Medical physics Examination Questions Key Features: 22 original examination style questions on all examinable topics. Full solutions

More information

Ultrasonic Testing Level I:

Ultrasonic Testing Level I: Ultrasonic Testing Level I: 1- Sound Wave - Introduction - ASNT Level I - Sound Wave Propagation - Velocity / Frequency / Wave Length - Acoustic Impedance - Energy / Intensity 2- Ultrasound Wave Modes

More information

17.4 Sound and Hearing

17.4 Sound and Hearing You can identify sounds without seeing them because sound waves carry information to your ears. People who work in places where sound is very loud need to protect their hearing. Properties of Sound Waves

More information

Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer

Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer Byungsik YOON AUTHOR 1, Hee-Jong LEE CO-AUTHOR

More information

MSD P13038 Hearing Aid Design. Final Review May 10, 2013

MSD P13038 Hearing Aid Design. Final Review May 10, 2013 MSD P13038 Hearing Aid Design Final Review May 10, 2013 Agenda 5.10.13 Project Overview (2) Mechanical Needs & Spec Review (10) Electrical & Computer Needs & Spec Review (10) Budget Review (2) Schedule

More information

Sound from Left or Right?

Sound from Left or Right? Sound from Left or Right? Pre-Activity Quiz 1. How does our sense of hearing work? 2. Why do we have two ears? 3. How does a stethoscope work? (A device used by doctors to listen to the sound of your heart.)

More information

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 Principles of Ultrasound Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012 None Disclosures Outline Introduction Benefits and Limitations of US Ultrasound (US) Physics

More information

Ultrasonic Testing of Composite Structures

Ultrasonic Testing of Composite Structures I. Introduction Ultrasonic Testing of Composite Structures This section of this work defines ultrasound basic concepts and Ultrasonic Technique. It describes the details of how ultrasonic testing works,

More information

Non-Destructive Inspection of Polyethylene Fusions and Electrofusions Dr. Ken Oliphant, P.Eng. and Dalton Crosswell, JANA Corporation

Non-Destructive Inspection of Polyethylene Fusions and Electrofusions Dr. Ken Oliphant, P.Eng. and Dalton Crosswell, JANA Corporation Non-Destructive Inspection of Polyethylene Fusions and Electrofusions Dr. Ken Oliphant, P.Eng. and Dalton Crosswell, JANA Corporation Abstract For buried gas distribution pipelines, the majority of the

More information

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques Engineering, 2013, 5, 551-555 doi:10.4236/eng.2013.56066 Published Online June 2013 (http://www.scirp.org/journal/eng) Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic

More information

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1 Q1.(a) When ultrasound is incident at an interface between two different media some energy is transmitted and some is reflected. The ratio of the reflected energy intensity I r to the incident energy intensity

More information

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods

More information

Ultrasonic Transducer for the Hydrothermal Method

Ultrasonic Transducer for the Hydrothermal Method Journal of the Korean Physical Society, Vol. 57, No. 4, October 2010, pp. 1122 1126 Ultrasonic Transducer for the Hydrothermal Method Peter Bornmann and Tobias Hemsel Mechatronics and Dynamics, University

More information

PLEASE DO NOT REMOVE THIS PAGE

PLEASE DO NOT REMOVE THIS PAGE Thank you for downloading this document from the RMIT Research Repository. The RMIT Research Repository is an open access database showcasing the research outputs of RMIT University researchers. RMIT Research

More information

General about Calibration and Service on Audiometers and Impedance Instruments

General about Calibration and Service on Audiometers and Impedance Instruments General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev.4 2008-05-21/JF General about Calibration and Service on Audiometers

More information

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah ULTRASOUND IMAGING EE 472 F2018 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Diagnostic Ultrasound: Physics and Equipment, 2nd ed., by Peter R. Hoskins (Editor), Kevin Martin (Editor),

More information

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM Basic of Ultrasound Physics E FAST & Renal Examination Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM What is Sound? Sound is Mechanical pressure waves What is Ultrasound? Ultrasounds are sound waves

More information

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016 Ultrasound Principles of Medical Imaging Prof. Dr. Philippe Cattin MIAC, University of Basel Contents Abstract 1 Image Generation Echography A-Mode B-Mode M-Mode 2.5D Ultrasound 3D Ultrasound 4D Ultrasound

More information

Terminology Tissue Appearance

Terminology Tissue Appearance By Marc Nielsen, MD Advantages/Disadvantages Generation of Image Ultrasound Machine/Transducer selection Modes of Ultrasound Terminology Tissue Appearance Scanning Technique Real-time Portable No ionizing

More information

General Treatment Equipment

General Treatment Equipment General Treatment Equipment We have quality products in General Treatment Equipment. All our models are efficient and display quality in Panchakarma (Ayurvedic) products. IMI - 3111: EXAMINATION COUCH

More information

Ultrasound Knobology

Ultrasound Knobology Ultrasound Knobology Raj Dasgupta MD, FACP, FCCP, FASSM Assistant Professor of Clinical Medicine Pulmonary / Critical Care / Sleep Medicine University of Southern California (USC) Objectives Physics of

More information

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE 4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE Purpose of experiment Determine the main characteristics of ultrasound waves, and the distances and positions of models using an ultrasonic echoscope.

More information

AVR Based Gesture Vocalizer Using Speech Synthesizer IC

AVR Based Gesture Vocalizer Using Speech Synthesizer IC AVR Based Gesture Vocalizer Using Speech Synthesizer IC Mr.M.V.N.R.P.kumar 1, Mr.Ashutosh Kumar 2, Ms. S.B.Arawandekar 3, Mr.A. A. Bhosale 4, Mr. R. L. Bhosale 5 Dept. Of E&TC, L.N.B.C.I.E.T. Raigaon,

More information

ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications

ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications ACOUSTORAPHY Acousto-Optical (AO) Ultrasonics & Its Applications International Workshop on Imaging NDE April 2007 IGCAR, kalpakkam, India Dr D S Dulay NDT Consultants Limited Middlemarch House, Siskin

More information

Iso-inertial dynamometer

Iso-inertial dynamometer Iso-inertial dynamometer A. Equipment set up B. Data acquisition and export C. Training mode 1 P a g e A. Equipment set up 1. Connect all the devices as depicted below: 2. Launch the software by clicking

More information

TC for Acoustics, Ultrasound and Vibration: Highlights Salvador Barrera-

TC for Acoustics, Ultrasound and Vibration: Highlights Salvador Barrera- TC for Acoustics, Ultrasound and Vibration: Highlights Salvador Barrera-Figueroa, TC-AUV Chair 1 Outline Overview of the TC-AUV Highlights: EMRP IND09 Traceable Dynamic Measurement of Mechanical Quantities

More information

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2 Page: 1 of 6 TABLE OF CONTENTS SECTION PAGE NO. 1.0 Introduction 2 2.0 Reference Documents 2 3.0 Summary of Practice 2 4.0 Significance & Use 3 5.0 Apparatus 3 6.0 Calibration & Adjustment of Apparatus

More information

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE 4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE Purpose of experiment Determine the main characteristics of ultrasound waves, and the distances and positions of models using an ultrasonic echoscope.

More information

FERROXCUBE. Multilayer suppressors EMI-suppression products. Supersedes data of February Sep 01

FERROXCUBE. Multilayer suppressors EMI-suppression products. Supersedes data of February Sep 01 FERROXCUBE DATA SHEET Supersedes data of February 2002 2004 Sep 01 MULTILAYER SUPPRESSORS are a powerful solution for EMI/RFI attenuation for electronic equipment. Supplied in seven standard sizes (0402,

More information

Science5 (SoundMulberry4th)

Science5 (SoundMulberry4th) Science5 (SoundMulberry4th) Name: Date: 1. The figure below shows a glass partly filled with water. Grace tapped the glass with a metal spoon and heard a sound. Which action would raise the pitch when

More information

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED from a CURRENT LOOP P18L TYPE USER S MANUAL 1 2 Contents 1. Application... 2 2. Basic requirements, operational safety... 2 3. Installation... 2 3.1. Assembly...

More information

Draft for comments only Not to be cited as East African Standard

Draft for comments only Not to be cited as East African Standard EAST AFRICAN STANDARD Poultry egg fertility tester Specification EAST AFRICAN COMMUNITY CD/K/636:2010 ICS 67.120.20 EAC 2010 First Edition 2010 CD/K/636:2010 Foreword Development of the East African Standards

More information

A STUDY ON HIGH FREQUENCY VIBRATION FOR TOOTH EXTRACTION

A STUDY ON HIGH FREQUENCY VIBRATION FOR TOOTH EXTRACTION Volume 119 No. 15 2018, 1507-1512 ISSN: 1314-3395 (on-line version) url: http://www.acadpubl.eu/hub/ http://www.acadpubl.eu/hub/ A STUDY ON HIGH FREQUENCY VIBRATION FOR TOOTH EXTRACTION Hepsiba.D 1 L.D.Vijay

More information

PHYS 1112L - Introductory Physics Laboratory II

PHYS 1112L - Introductory Physics Laboratory II PHYS 1112L - Introductory Physics Laboratory II Laboratory Advanced Sheet EKG Lab (IBEAM) 1. Objectives. The objectives of this laboratory are a. to study the electrical activity of human heart b. to observe

More information

Ultrasonic Testing. Basic Principles

Ultrasonic Testing. Basic Principles Ultrasonic Testing Ultrasonic Testing (UT) uses high frequency sound waves (typically in the range between 0.5 and 15 MHz) to conduct examinations and make measurements. Besides its wide use in engineering

More information

Spirometry air flow measurement using PVDF Film

Spirometry air flow measurement using PVDF Film Spirometry air flow measurement using PVDF Film Manisha R.Mhetre, H.K.Abhyankar Department of Instrumentation Engineering, Vishwakarma Institute of Technology, Pune - 411037, Maharashtra, India Email:

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

Ultrasonic Instruments for Non-Destructive Testing

Ultrasonic Instruments for Non-Destructive Testing Ultrasonic Instruments for Non-Destructive Testing SCOPE The SoniSys Out-of-Plane Ultrasonic System, OPUS, is a unique ultrasonic system providing non-destructive Z- direction analysis of materials including

More information

Proceedings Thin Film PZT Acoustic Sensor for Fully Implantable Cochlear Implants

Proceedings Thin Film PZT Acoustic Sensor for Fully Implantable Cochlear Implants Proceedings Thin Film PZT Acoustic Sensor for Fully Implantable Cochlear Implants Bedirhan İlik 1,2, Aziz Koyuncuoğlu 2, Hasan Uluşan 1, Salar Chamanian 1, Dilek Işık 2, Özlem Şardan-Sukas 2 and Haluk

More information

Hearing. istockphoto/thinkstock

Hearing. istockphoto/thinkstock Hearing istockphoto/thinkstock Audition The sense or act of hearing The Stimulus Input: Sound Waves Sound waves are composed of changes in air pressure unfolding over time. Acoustical transduction: Conversion

More information

(12) United States Patent

(12) United States Patent (12) United States Patent Kozári et al. USOO6309685B1 (10) Patent No.: US 6,309,685 B1 (45) Date of Patent: Oct. 30, 2001 (54) PROCEDURE AND INSTRUMENT FOR THE QUICK PICKLING OF MAINLY MEAT (76) Inventors:

More information

MATCHING LAYER DESIGN OF AN ULTRASONIC TRANS- DUCER FOR WIRELESS POWER TRANSFER SYSTEM

MATCHING LAYER DESIGN OF AN ULTRASONIC TRANS- DUCER FOR WIRELESS POWER TRANSFER SYSTEM MATCHIN LAYE DESIN OF AN ULTASONIC TANS- DUCE FO WIELESS POWE TANSFE SYSTEM unn Hang Electronics and Telecommunications esearch Institute, Multidisciplinary Sensor esearch roup, Daejeon, South Korea email:

More information

High resolution ultrasound scanner for skin imaging

High resolution ultrasound scanner for skin imaging High resolution ultrasound scanner for skin imaging Christine Turlat Sales Director Atys medical 17 Parc d Arbora 69510 SOUCIEU EN JARREST Atys company Principle of ultrasound imaging DERMCUP Normal image

More information

3/20/2017. Disclosures. Ultrasound Fundamentals. Ultrasound Fundamentals. Bone Anatomy. Tissue Characteristics

3/20/2017. Disclosures. Ultrasound Fundamentals. Ultrasound Fundamentals. Bone Anatomy. Tissue Characteristics Disclosures Images of ultrasound equipment in this presentation are not an endorsement Fundamentals of Musculoskeletal Ultrasound Physics and Knobology Shane A. Shapiro, M.D. Assistant Professor Orthopedic

More information

CONVENTIONAL AND DIGITAL HEARING AIDS

CONVENTIONAL AND DIGITAL HEARING AIDS CONVENTIONAL AND DIGITAL HEARING AIDS Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This Medical

More information

Sound and its characteristics. The decibel scale. Structure and function of the ear. Békésy s theory. Molecular basis of hair cell function.

Sound and its characteristics. The decibel scale. Structure and function of the ear. Békésy s theory. Molecular basis of hair cell function. Hearing Sound and its characteristics. The decibel scale. Structure and function of the ear. Békésy s theory. Molecular basis of hair cell function. 19/11/2014 Sound A type of longitudinal mass wave that

More information

Chapter 3 Biological measurement 3.1 Nerve conduction

Chapter 3 Biological measurement 3.1 Nerve conduction Chapter 3 Biological measurement 3.1 Nerve conduction Learning objectives: What is in a nerve fibre? How does a nerve fibre transmit an electrical impulse? What do we mean by action potential? Nerve cells

More information

Performance Evaluation of Smart Glove Applied to Experimental Rig to Control Human Hand Tremor for Parkinson disease

Performance Evaluation of Smart Glove Applied to Experimental Rig to Control Human Hand Tremor for Parkinson disease Performance Evaluation of Smart Glove Applied to Experimental Rig to Control Human Hand Tremor for Parkinson disease SUHAIL KAZI, AZIZAN, ZARHAMDY MD ZAIN and MUSA MAILAH Department of Applied Mechanics,

More information

DIAGNOSTIC TECHNIQUE OF ABNORMALITIES IN BALL BEARINGS WITH AN ULTRASONIC METHOD

DIAGNOSTIC TECHNIQUE OF ABNORMALITIES IN BALL BEARINGS WITH AN ULTRASONIC METHOD 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand DIAGNOSTIC TECHNIQUE OF ABNORMALITIES IN BALL BEARINGS WITH AN ULTRASONIC METHOD Akitoshi Takeuchi 1 1 Kochi

More information

MODELING USING THE FINITE ELEMENT METHOD OF AN ULTRASONIC STACK USED FOR ULTRASONIC WELDING STAKING PROCESS

MODELING USING THE FINITE ELEMENT METHOD OF AN ULTRASONIC STACK USED FOR ULTRASONIC WELDING STAKING PROCESS SISOM 2011 and Session of the Commission of Acoustics, Bucharest 25-26 May MODELING USING THE FINITE ELEMENT METHOD OF AN ULTRASONIC STACK USED FOR ULTRASONIC WELDING STAKING PROCESS Doru Virgil PĂUŞAN

More information

Implants for surgery Active implantable medical devices. Part 7: Particular requirements for cochlear implant systems

Implants for surgery Active implantable medical devices. Part 7: Particular requirements for cochlear implant systems Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 14708-7 First edition 2013-01-15 Implants for surgery Active implantable medical devices Part 7: Particular requirements for cochlear implant systems

More information

Development of an Electronic Glove with Voice Output for Finger Posture Recognition

Development of an Electronic Glove with Voice Output for Finger Posture Recognition Development of an Electronic Glove with Voice Output for Finger Posture Recognition F. Wong*, E. H. Loh, P. Y. Lim, R. R. Porle, R. Chin, K. Teo and K. A. Mohamad Faculty of Engineering, Universiti Malaysia

More information

Introduction to Nondestructive Testing. Outline

Introduction to Nondestructive Testing. Outline Introduction to Nondestructive Testing Outline Introduction to NDT Overview of Six Most Common NDT Methods Selected Applications Page 1 Definition of NDT The use of noninvasive techniques to determine

More information

Basic Physics of Ultrasound and Knobology

Basic Physics of Ultrasound and Knobology WELCOME TO UTMB Basic Physics of Ultrasound and Knobology By Daneshvari Solanki, FRCA Laura B. McDaniel Distinguished Professor Anesthesiology and Pain Medicine University of Texas Medical Branch Galveston,

More information

ESWL Based on Experience + Innovation

ESWL Based on Experience + Innovation Dreifachfokus DSR- Ortung Dreifachfokus DSR Localization Triple-focus Double-layer Piezo Technology Automatic Patient Positioning ESWL Based on Experience + Innovation Richard Wolf GmbH and ELvation Medical

More information

Evaluation of heavy wire bond tools with ceramic tip technology

Evaluation of heavy wire bond tools with ceramic tip technology Introduction F&K Physiktechnik GmbH, Potsdam Germany, develops and offers test and measurement equipment to characterize the ultrasonic components of wire and flip chip die bonders for microelectronic

More information

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D. Diagnostic Ultrasound Sutiporn Khampunnip, M.D. Definition of Ultrasound Ultrasound is simply sound waves, like audible sound. High-frequency sound and refers to mechanical vibrations above 20 khz. Human

More information

(12) Patent Application Publication (10) Pub. No.: US 2004/ A1

(12) Patent Application Publication (10) Pub. No.: US 2004/ A1 (19) United States US 2004OO39244A1 (12) Patent Application Publication (10) Pub. No.: US 2004/0039244A1 Kroll et al. (43) Pub. Date: Feb. 26, 2004 (54) METHOD FOR CREATING A COUPLING BETWEENA DEVICE AND

More information

Hearing and Sound Study Guide

Hearing and Sound Study Guide Hearing and Sound Study Guide I made the following website as a study tool to help you prepare for Thursday with videos and sample questions hearingandsound.weebly.com 1. Sounds make vibrations and vibrations

More information

Ultrasound Physics & Terminology

Ultrasound Physics & Terminology Ultrasound Physics & Terminology This module includes the following: Basic physics terms Basic principles of ultrasound Ultrasound terminology and terms Common artifacts seen Doppler principles Terms for

More information

Ultrasound Physics & Doppler

Ultrasound Physics & Doppler Ultrasound Physics & Doppler Endocrine University 2018 Mark Lupo, MD, FACE, ECNU Objectives Review the essential components of ultrasound physics in neck sonography Demonstrate the importance of ultrasound

More information

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary Ultrasound Physics and Knobology Alan Macfarlane Consultant Anaesthetist Glasgow Royal Infirmary RAPM 2009; 34: 40-46 Ultrasound Proficiency Understanding US image generation and device operation Image

More information

An Overview of Ultrasound Testing For Lesion Detection in Human Kidney

An Overview of Ultrasound Testing For Lesion Detection in Human Kidney Journal of Tomography System & Sensors Application Vol.1, Issue 1, June 2018 An Overview of Ultrasound Testing For Lesion Detection in Human Kidney Aina Fadhilah Abd Rahim 1, Zawin Najah Abd Halim 1, Jaysuman

More information

(12) United States Patent (10) Patent No.: US 6,691,578 B1

(12) United States Patent (10) Patent No.: US 6,691,578 B1 USOO6691578B1 (12) United States Patent (10) Patent No.: US 6,691,578 B1 Puskas (45) Date of Patent: Feb. 17, 2004 (54) MEASUREMENT SYSTEMS FOR 4,710,233 A 12/1987 Hohmann et al.... 205/701 ULTRASOUND

More information

NONDESTRUCTIVE EVALUATION OF PLATE TYPE NUCLEAR FUEL ELEMENTS DURING MANUFACTURING STAGE USING ULTRASONIC TEST METHOD

NONDESTRUCTIVE EVALUATION OF PLATE TYPE NUCLEAR FUEL ELEMENTS DURING MANUFACTURING STAGE USING ULTRASONIC TEST METHOD 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 NONDESTRUCTIVE

More information

COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION

COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION 034 - Proceeding of the Global Summit on Education (GSE2013) COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION ABSTRACT Nur Azah Hamzaid, Ammar

More information

OCCLUSION REDUCTION SYSTEM FOR HEARING AIDS WITH AN IMPROVED TRANSDUCER AND AN ASSOCIATED ALGORITHM

OCCLUSION REDUCTION SYSTEM FOR HEARING AIDS WITH AN IMPROVED TRANSDUCER AND AN ASSOCIATED ALGORITHM OCCLUSION REDUCTION SYSTEM FOR HEARING AIDS WITH AN IMPROVED TRANSDUCER AND AN ASSOCIATED ALGORITHM Masahiro Sunohara, Masatoshi Osawa, Takumi Hashiura and Makoto Tateno RION CO., LTD. 3-2-41, Higashimotomachi,

More information

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03

SOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 SOLUTIONS Homework #3 Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 Problem 1: a) Where in the cochlea would you say the process of "fourier decomposition" of the incoming

More information

Android based Monitoring Human Knee Joint Movement Using Wearable Computing

Android based Monitoring Human Knee Joint Movement Using Wearable Computing Android based Monitoring Human Knee Joint Movement Using Wearable Computing Abstract In today s fast moving lifestyle, incidents regarding health issues are surfacing every day. One of the major issues

More information

ON A MECHATRONIC SYSTEM TO DETERMINE DYNAMIC PARAMETERS OF THE HUMAN LIMB DURING LOCOMOTION USED FOR ARTIFICIAL MUSCLES AND IMPLANT DESIGN

ON A MECHATRONIC SYSTEM TO DETERMINE DYNAMIC PARAMETERS OF THE HUMAN LIMB DURING LOCOMOTION USED FOR ARTIFICIAL MUSCLES AND IMPLANT DESIGN 6th International DAAAM Baltic Conference INDUSTRIAL ENGINEERING 24-26 April 2008, Tallinn, Estonia ON A MECHATRONIC SYSTEM TO DETERMINE DYNAMIC PARAMETERS OF THE HUMAN LIMB DURING LOCOMOTION USED FOR

More information

Portable Stimulator Design Aimed at Differentiating Facial Nerves From Normal Tissues

Portable Stimulator Design Aimed at Differentiating Facial Nerves From Normal Tissues Journal of Medical Systems, Vol. 28, No. 2, April 2004 ( C 2004) Portable Stimulator Design Aimed at Differentiating Facial Nerves From Normal Tissues S. Kara, 1,2 S. Kemaloğlu, 1 F. Şener, 1 M. Okandan,

More information

VENTRICULAR DEFIBRILLATOR

VENTRICULAR DEFIBRILLATOR VENTRICULAR DEFIBRILLATOR Group No: B03 Ritesh Agarwal (06004037) ritesh_agarwal@iitb.ac.in Sanket Kabra (06007017) sanketkabra@iitb.ac.in Prateek Mittal (06007021) prateekm@iitb.ac.in Supervisor: Prof.

More information

An Overview of BMIs. Luca Rossini. Workshop on Brain Machine Interfaces for Space Applications

An Overview of BMIs. Luca Rossini. Workshop on Brain Machine Interfaces for Space Applications An Overview of BMIs Luca Rossini Workshop on Brain Machine Interfaces for Space Applications European Space Research and Technology Centre, European Space Agency Noordvijk, 30 th November 2009 Definition

More information

Sound. Chapter Test A. Multiple Choice. 1 Pearson Education, Inc., or its affiliates. All rights reserved.

Sound. Chapter Test A. Multiple Choice. 1 Pearson Education, Inc., or its affiliates. All rights reserved. _ Sound Chapter Test A Multiple Choice Write the letter of the correct answer on the line at the left. _ 1. Sound is a disturbance that travels through a medium as a a. longitudinal wave. b. surface wave.

More information

Translation of Unintelligible Speech of User into Synthetic Speech Using Augmentative and Alternative Communication

Translation of Unintelligible Speech of User into Synthetic Speech Using Augmentative and Alternative Communication Translation of Unintelligible Speech of User into Synthetic Speech Using Augmentative and Alternative Communication S.Jasmine Vanitha 1, K. Manimozhi 2 P.G scholar, Department of EEE, V.R.S College of

More information

ascending phases began to diverge was taken to mark the onset of decay in the

ascending phases began to diverge was taken to mark the onset of decay in the 605 J. Physiol. (I954) I24, 605-6I2 THE DURATION OF THE PLATEAU OF FULL ACTIVITY IN FROG MUSCLE BY J. M. RITCHIE From the National Institute for Medical Research, Mill Hill, London, N.W. 7 (Received 26

More information

Ultrasonic Phased Array Testing of Complex Aircraft Structures

Ultrasonic Phased Array Testing of Complex Aircraft Structures ECNDT 2006 - Tu.1.1.2 Ultrasonic Phased Array Testing of Complex Aircraft Structures Ernst RAU, Ernst GRAUVOGL, Holger MANZKE, EADS Mil., Manching, Germany Philippe CYR, Olympus NDT, Québec, Canada 1.

More information

IT S A SECRET! (1 Hour) Addresses NGSS Level of Difficulty: 4 Grade Range: K-2

IT S A SECRET! (1 Hour) Addresses NGSS Level of Difficulty: 4 Grade Range: K-2 IT S A SECRET! (1 Hour) Addresses NGSS Level of Difficulty: 4 Grade Range: K-2 OVERVIEW In this activity, students explore how sound waves travel through various materials. They will build a sound transmission

More information

Introduction & Physics of ED Ultrasound. Objectives. What? - Limited Studies. Who? - ED Docs

Introduction & Physics of ED Ultrasound. Objectives. What? - Limited Studies. Who? - ED Docs Introduction & Physics of ED Ultrasound Martine Sargent, MD Ultrasound Director, Assistant Professor UCSF Department of Emergency Medicine San Francisco General Hospital & Trauma Center Objectives Who?

More information

Chapter 13 Physics of the Ear and Hearing

Chapter 13 Physics of the Ear and Hearing Hearing 100 times greater dynamic range than vision Wide frequency range (20 ~ 20,000 Hz) Sense of hearing Mechanical system that stimulates the hair cells in the cochlea Sensors that produce action potentials

More information