Skin Model and its impact on Digital Mammography
|
|
- Milton Potter
- 5 years ago
- Views:
Transcription
1 Skin Model and its impact on Digital Mammography Rodrigo T. Massera; Alessandra Tomal Institute of Physics "Gleb Wataghin University of Campinas Campinas, Brazil 1
2 Outline Motivation Methodology Results Mammography Dosimetry Mean Glandular Dose Implemented Models How? MGD vs Skin Model Differences Conclusions Summary 2
3 Introduction Why is dosimetry important in mammography? Population-based screening programs Use Ionizing Radiation Quality Control and Optimization 3
4 Mean Glandular Dose (MGD) Real Breast Incident Photons Skin Adipose Tissue Glandular Tissue
5 Mean Glandular Dose (MGD) Real Breast Energy Deposited: Glandular Tissue Directly measured? Monte Carlo simulation Skin Glandular Tissue Adipose Tissue
6 Mean Glandular Dose (MGD) Parameters to consider... X-ray spectrum; Geometric Model; Breast Thickness; Breast Composition (f g ); Skin Model? Previous Estimations: 5 mm Adipose Tissue ( Dance 1990) 4 mm Skin Tissue (Wu 1991/Boone 1999) Credits: Boone & Hernandez 2016, AAPM. Changing Perceptions and Updated Methods for Mammography Dosimetry Current Measures: 64% thinner Using breast-ct: 1.44 mm (Vedantham et al 2012) 1.45 mm (Huang et al 2008); + adipose layer 6
7 Objectives Study the impact of skin models on Mean Glandular Dose in Digital Mammography Adapt MC Code Analysis Geometry MGD calculus MGD X Skin Models 7
8 Outline Motivation Methodology Results Mammography Dosimetry Mean Glandular Dose Implemented Models How? MGD vs Skin Model Differences Conclusions Summary 8
9 Methodology Monte Carlo code: PENELOPE (2014) + peneasy (2015) Beam Parameters: Monoenergetic (8 60 kev) Polyenergetic (22 35 kv): Mo (Mo-Rh) Rh (Rh) W (Rh-Al-Ag) *X-ray spectra from Hernandez et al (2014) 9
10 Methodology Breast Model* t = 2 cm 8 cm Glandularity (f g ) = 1%-100% *Compositions from Hammerstein et al
11 Methodology Breast Model* t = 2 cm 8 cm Glandularity (f g ) = 1%-100% Skin shielding Models I. 5 mm adipose; II. 4 mm skin; III. 1,45 mm skin; IV. 1,45 mm skin + 2 mm adipose; V. 1,45 mm skin + 3,55 mm adipose; 11 *Compositions from Hammerstein et al 1979
12 Mean Glandular Dose (MGD) Monte Carlo Simulations How do we separate the deposited energy between tissues? Skin peneasy: E avg Homogeneous Glandular- Adipose Tissue Mixture Adipose Tissue 12
13 Mean Glandular Dose (MGD) MGD Weighing method (Dance 1990) Simulation starts: E gland =0 (I) (III) Photoelectric Interaction Type Simulation Ends: Return E gland Incoherent MGD = E gland Mass f g (II) nmgd = MGD K air 13
14 Code modifications... Dosimetry nmgd = MGD K air MGD Air Kerma from Primary Photons ~ simulations 14
15 Automatization with Python Windows/Linux full compatibility User Input Mono/Poly; Energy Range/Anode Filter; Breast Thickness range; Breast Composition; Skin Model; Detector Type; Antiscatter Grid Model; etc Python Script List of Simulations PENELOPE Parallel Simulations ~ simulations Return Data Python Data Collection and saving 15
16 Automatization with Python 3-10 min/simulation Uncertainty (1-0.25%) Processor i Ghz 16
17 Outline Motivation Methodology Results Mammography Dosimetry Mean Glandular Dose Implemented Models How? MGD vs Skin Model Differences Conclusions Summary 17
18 Results - Code Validation AAPM Report 195 (2015) <1% 18
19 Results - Code Validation Sarno et al PMB <4% 5 cm thick; 20% f g ; 1.45 mm skin; 19
20 Results: Skin shielding models Monoenergetic Beam 1% f g 100% f g 20
21 Results: Skin shielding models Monoenergetic Beam Depth Dose 18 kev 20% f g 2 cm breast 8 cm breast 21
22 Results: Skin shielding models Polyenergetic Beam 2 cm 8 cm 20% f g 15% 16% 36% 34% 22
23 Results: Skin shielding models Polyenergetic Beam Mo/Mo 28 kv 2 cm breast 21% 23% 23
24 Results: Skin shielding models Polyenergetic Beam Mo/Mo 28 kv 21% 1% f g 17% 24
25 Results: Summary Polyenergetic Beam Skin Models 25
26 Outline Motivation Methodology Results Mammography Dosimetry Mean Glandular Dose Implemented Models How? MGD vs Skin Model Differences Conclusions Summary 26
27 Conclusions The Skin Model has a significant impact on MGD estimates; Skin model affects the MGD up to 37%; Larger variations: low energies; high glandularity, thin breasts Depth Dose : skin attenuation and Homogeneous Mixture Volume MGD Variation f g ( 50%) Breast Thickness ( 350%) Skin Model ( 40%) Reduce the uncertanties; Patient-specific dosimetry; Heterogeneous breast Anode/Filter ( 90%) Tube Potential ( 130%) 27
28 28 Acknowledgement Process 2016/ Process 2015/ Process / Lab Members and Alumni Collaborators Rodrigo T. Massera Bruno L. Rodrigues José Maria Fernandez-Varea
29 Our Institution Funded in 1966 Credits: Lucas Rodolfo de Castro Moura University of Campinas (UNICAMP): 1st in Latin America
30 Thank You! Rodrigo T. Massera MSc Student IFGW - UNICAMP 30
Patient Dosimetry in Mammography and Tomosynthesis:
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor
More informationDosimetry in digital mammography
Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol
More informationBreast CT and Dosimetry
2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Breast CT and Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair (Research) of Radiology Professor of
More informationRadiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist
Radiation Dosimetry in Digital Breast Tomosynthesis March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Syllabus 1. Introduction 2. Dosimetry in Mammography 3. Dosimetry
More informationMethod for determination of the mean fraction of glandular tissue in individual female breasts using mammography
INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 50 (2005) 5953 5967 PHYSICS IN MEDICINE AND BIOLOGY doi:10.1088/0031-9155/50/24/013 Method for determination of the mean fraction of glandular tissue in
More informationDirect half value layer measurements in mammography - is near enough good enough?
Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.
More informationEffect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis
Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More informationUsing Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography
Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,
More informationRadiation Exposure and Image Quality in X-Ray Diagnostic Radiology
Horst Aichinger Joachim Dierker Sigrid Joite-Barfuß Manfred Säbel Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Physical Principles and Clinical Applications Second Edition Springer
More informationPilot study of patient and phantom breast dose measurements in Bulgaria
Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast
More informationMammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433
- 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.
More informationΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ MONTE CARLO
ΠΑΝΕΠΙΣΤΗΜΙΟ ΠΑΤΡΩΝ ΤΜΗΜΑ ΙΑΤΡΙΚΗΣ ΤΜΗΜΑ ΦΥΣΙΚΗΣ ΙΑΤΜΗΜΑΤΙΚΟ ΠΡΟΓΡΑΜΜΑ ΜΕΤΑΠΤΥΧΙΑΚΩΝ ΣΠΟΥ ΩΝ ΣΤΗΝ ΙΑΤΡΙΚΗ ΦΥΣΙΚΗ ΜΕΛΕΤΗ ΑΠΟΡΡΟΦΩΜΕΝΗΣ ΟΣΗΣ ΣΕ ΜΕΓΕΘΥΝΤΙΚΕΣ ΛΗΨΕΙΣ ΣΤΗ ΜΑΣΤΟΓΡΑΦΙΑ ΜΕ ΤΕΧΝΙΚΕΣ ΠΡΟΣΟΜΟΙΩΣΗΣ
More informationImpact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography
Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.
More informationfor Mammography Dosimetry
Changing Perceptions and Updated Methods for Mammography Dosimetry John M. Boone, Ph.D., FAAPM, FSBI, FACR, FAIMBE Professor of Radiology & Biomedical Engineering University of California Davis Sacramento,
More informationBREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0
BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2
More informationDosimetry comparison of orthovoltage x-ray and 137 Cs irradiation of the murine bone marrow compartment Matthew Belley
Dosimetry comparison of orthovoltage x-ray and 137 Cs irradiation of the murine bone marrow compartment Matthew Belley NCHPS Fall Meeting October 9, 2015 Duke Medical Physics Disclaimer Financial support
More informationThis article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing
More informationBreast dosimetry system in screen/film mammography
Bull Fae Health Sei, Okayama Univ Med Seh, 10: 99-106, 2000 (Original article) Breast dosimetry system in screen/film mammography Sachiko GOTO, Yoshiharu AZUMA, Toshinori MARUYAMA, Yoshitada NAKAGIRI,
More informationImproving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations
Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,
More informationTechnical quality control in the Norwegian Breast Cancer Screening Program with focus on radiation doses
Technical quality control in the Norwegian Breast Cancer Screening Program with focus on radiation doses Ingrid Helen Ryste Hauge Department of Physics Faculty of Mathematics and Natural Sciences University
More informationReview of TG-186 recommendations
Review of TG-186 recommendations Implementation of advanced brachytherapy dose calculation algorithms beyond TG-43 Rowan M. Thomson Carleton Laboratory for Radiotherapy Physics Carleton University Ottawa
More informationMammography calibration qualities establishment in a Mo- Mo clinical system
Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733
More informationPATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME
Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY
More informationInvivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
More informationContrast Enhanced Spectral Mammography (CESM) Updates
Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da
More informationPhD Title: Improving QA/QC in mammography screening and breast in Montenegro
3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:
More informationThe ANDANTE project: a multidisciplinary approach to neutron RBE
The ANDANTE project: a multidisciplinary approach to neutron RBE Andrea Ottolenghi, Klaus Trott, Giorgio Baiocco, Vere Smyth Università degli Studi di Pavia, Italy On behalf of the ANDANTE project MELODI
More informationPatient effective dose evaluation for chest X-ray examination in three digital radiography centers
Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical
More informationA tissue-equivalent phantom series for mammography dosimetry
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOL. 5, NO. 4, FALL 2004 A tissue-equivalent phantom series for mammography dosimetry William P. Argo, 1 Kathleen Hintenlang, 2 and David E. Hintenlang 3 U.S.
More informationCT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center
CT Dose Estimation John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology 1 University of California Davis Medical Center CT Dose Estimation Introduction The CTDI Family of Metrics
More informationApplications of Monte Carlo simulations to radiation dosimetry
Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationEVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME
EVOLUTION OF THE EXCESS ABSOLUT RISK (EAR) IN THE VALENCIAN BREAST CANCER SCREENING PROGRAMME S. Ferrer (), M Ramos (), J.I. Villaescusa (), G Verdú (), M.D. Salas (3), M.D. Cuevas (4) () Departamento
More informationX- ray imaging. F. Grüner
X- ray imaging F. Grüner Röntgen s first medical imaging Improvements. System-specific image processing Adaptive auto-windowing algorithm Basic principles X- ray absorp1on/phase contrast imaging X- ray
More informationComparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
Iran. J. Radiat. Res., 21; 7 (4): 217-222 Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Downloaded from ijrr.com at 6:52 +33
More informationPGY-1. Resident Review Session Schedule
1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT
More informationMedical Physics 4 I3 Radiation in Medicine
Name: Date: 1. This question is about radiation dosimetry. Medical Physics 4 I3 Radiation in Medicine Define exposure. A patient is injected with a gamma ray emitter. The radiation from the source creates
More informationDosimetric characterization of surface applicators for use with the Xoft ebx system
Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal
More informationMonte Carlo Modelling: a reliable and efficient tool in radiation dosimetry
Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationIAEA TECDOC 1517 Quality Control in Mammography Software
IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.
More informationVenue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry
Venue: IEEE NSS/MIC/RTSD Conference, Seoul, South Korea, 27 th October 2013 Workshop: NWK3/RD1 Radiation Protection and Dosimetry M. Caresana a, A. Sashala Naik a,c, S. Rollet b, M. Ferrarini a,d a Polytechnic
More informationMetrology Laboratory of Ionizing Radiation
Metrology Laboratory of Ionizing Radiation Carlos Oliveira The Metrology Laboratory of Ionizing Radiation (LMRI) has been involved in activities related with scientific, technical and legal metrology.
More informationComparison Between Film-Screen and Digital Mammography for Woman Breast Cancer Screening: Mean Glandular Dose
Academic Journal of Cancer Research 7 (2): 162-167, 2014 ISSN 1995-8943 IDOSI Publications, 2014 DOI: 10.5829/idosi.ajcr.2014.7.2.83313 Comparison Between Film-Screen and Digital Mammography for Woman
More informationExamination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA
Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi
More informationCalibration coefficients of dosimeters used in mammography for various target/filter combinations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 Calibration coefficients of dosimeters used in mammography for various target/filter combinations Michiharu Sekimoto, 1a Yoh Katoh,
More informationDose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP
Iran. J. Radiat. Res., 2006; 4 (1): 15-19 Dose distribution and dosimetry parameters calculation of MED3633 palladium- source in water phantom using MCNP A.A. Mowlavi 1*,A. Binesh 2, H. Moslehitabar 3
More informationAsk EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina
More informationOptimizing non-pb radiation shielding materials using bilayers
Optimizing non-pb radiation shielding materials using bilayers J. P. McCaffrey, a E. Mainegra-Hing, and H. Shen Institute for National Measurement Standards, National Research Council of Canada, Building
More informationQuality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio
Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz
More informationClinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography
Diagn Interv Radiol 2007; 13:134 139 Turkish Society of Radiology 2007 BREAST IMAGING ORIGINAL ARTICLE Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationCURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC
CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine Acknowledgments John Boone, PhD Michael McNitt-Grey,
More informationHighlights and Conclusions Thematic Area 4: Radiation of Patients
Highlights and Conclusions Thematic Area 4: Radiation of Patients by Robert H. Corbett and Eliseo Vañó Main topics: Diagnostic Radiology Interventional Radiology Nuclear Medicine Radiotherapy New Techniques
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationOn the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams
On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,
More informationAn Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations
An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations Aiping Ding, X. George Xu Rensselaer Polytechnic Institute Troy, NY USA http://rrmdg.rpi.edu 1 Acknowledgements
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationCurrent Status of Electronic Brachytherapy Dosimetry
Current Status of Electronic Brachytherapy Dosimetry 2014 NCCAAPM Fall Meeting La Crosse, WI Wes Culberson, PhD, DABR University of Wisconsin Madison University of Wisconsin Medical Radiation Research
More informationDevelopment of a program for estimation of patient exposure in mammography
Developmt of a program for estimation of patit exposure in mammography Y. Asada 1, S. Suzuki, M. Kobayashi, S. Koga 1 Faculty of radiological technology, Fujita Health University, School of Health Scices,
More informationPlastic scintillation detectors for dose monitoring in digital breast tomosynthesis
Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis J. Antunes 1,2, J. Machado 1,2, L. Peralta 1,2, N. Matela 1,3 1 Departamento de Física da Faculdade de Ciências da Universidade
More informationMean glandular dose values used for the mammography screening program in Poland according to the type of image registration system
Original article NOWOTWORY Journal of Oncology 2017, volume 67, number 2, 115 120 DOI: 10.5603/NJO.2017.0018 Polskie Towarzystwo Onkologiczne ISSN 0029 540X www.nowotwory.edu.pl Mean glandular dose values
More informationFAX, A FEMALE ADULT VOXEL PHANTOM FOR RADIATION PROTECTION DOSIMETRY
EFFECTIVE DOSE RATIOS FOR THE TOMOGRAPHIC MAX AND FAX PHANTOMS ******************************************************** EFFECTIVE DOSE RATIOS FOR TOMOGRAPHIC AND STYLIZED MODELS FROM > EXTERNAL EXPOSURE
More informationPatient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of Low Dose Radiation Calculations
University of New Mexico UNM Digital Repository Nuclear Engineering ETDs Engineering ETDs Spring 4-9-2018 Patient Specific Radiation Doses From Projection Radiography Images: Improving the Accuracy of
More informationApplication of international standards to diagnostic radiology dosimetry
Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,
More informationApplication of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom
Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,
More informationImproving clinical brachytherapy: dosimetry & verification
1/XXX Improving clinical brachytherapy: dosimetry & verification Organ Frank motion Verhaegen, in radiotherapy Gabriel Fonseca, Murillo Bellezzo, Shane White, Guillaume Landry, Brigitte Reniers nada* MAASTRO
More informationMonte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.
Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In
More informationApplication of the UF Computational Phantoms for Operational Health Physics Dosimetry for NASA
Application of the UF Computational Phantoms for Operational Health Physics Dosimetry for NASA FCHPS & FLAAPM 2015 Fall Meeting November 13 th, 2015 Michelle Sands, M.S. University of Florida Introduction
More informationSimulations of Preclinical andclinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy. Using GATE
GATE Simulations of Preclinical andclinical Scans in Emission Tomography, Transmission Tomography and Radiation Therapy Using GATE Quick tour & Highlights! GATE Training, INSTN-Saclay, October 2015 Albertine
More informationALARA in mammography screening. Hilde Bosmans et al.
ALARA in mammography screening Hilde Bosmans et al. ALARA in the mammography Overview 1. Screening 2. The balance: 1. Image quality 2. Radiation dose 3. In practice Screening leeftijd Mammography Screening
More informationIPPE Iron shell transmission experiment with 14 MeV neutron source and its analysis by the Monte-Carlo method
IPPE Iron shell transmission experiment with 14 MeV neutron source and its analysis by the Monte-Carlo method S.P. Simakov 1,2, M.G. Kobozev 1, A.A. Lychagin 1, V.A. Talalaev 1, U. Fischer 2, U. von Möllendorff
More informationDosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom
Dosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom Institute of Physics E. Gaviola - CONICET & LIIFAMIRX - Laboratorio de Investigación
More informationValidation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons. Yann Perrot, LPC, CNRS/IN2P3
Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons Yann Perrot, LPC, CNRS/IN2P3 1 Projects @ PCSV 2 Projects @ PCSV Computation for life science
More informationThe biological effectiveness of low-energy photons
Onkologisches Zentrum Klinik für Strahlentherapie und Radioonkologie Medizinische Strahlenphysik The biological effectiveness of low-energy photons Elisabetta Gargioni University Medical Center Hamburg-Eppendorf
More informationRange Uncertainties in Proton Therapy
Range Uncertainties in Proton Therapy Harald Paganetti PhD Professor of Radiation Oncology, Harvard Medical School Director of Physics Research, Massachusetts General Hospital, Department of Radiation
More informationGATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT
GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS
More informationCone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood
Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex
More informationEstablishment of the new IEC mammography qualities in a clinical system used for instruments calibration
Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas
More informationCalibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System
Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of
More informationTesting of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution
Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi
More informationEstimating Patient Radiation Dose from Computed Tomography
Estimating Patient Radiation Dose from Computed Tomography C. Cagnon, J. DeMarco, E. Angel, M. McNitt-Gray UCLA David Geffen School of Medicine 1 Patient Dose from CT Advances in Technology... Helical
More informationState of the art and future development for standardized estimation of organ doses in CT
State of the art and future development for standardized estimation of organ doses in CT March 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist Imagination at work. Agenda Introduction Duke Florida
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationComparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5
Comparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5 S.M. Reda Department of Physics, Faculty of Science, Zagazig University,
More informationCalibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia
Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute
More informationGlandularity Estimation in Japanese Women by Using a Breast Model Made from Mammographic Findings of European Women
Jpn. J. Health Phys. 51 (4) 238 244 (2016) DOI: 10.5453/jhps.51.238 Original Paper Glandularity Estimation in Japanese Women by Using a Breast Model Made from Mammographic Findings of European Women Ai
More informationMammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences
1313 C. Kimme-Smith1 J. Wang2 N. DeBruhl1 M. Basic2 L. W. Bassett1 Received October 27, 1 993; accepted after revision January 25, 1994. Presented at the annual meeting of the American Roentgen Ray Society,
More informationOpportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI
Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI john.sandrik@med.ge.com with many thanks to Vince Polkus, Advanced Applications Product Mgr. 1 Content
More informationImpact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program
Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,
More informationDose standards for safe and secure breast cancer screening
Research paper Dose standards for safe and secure breast cancer screening - Establishment of X-ray dose standards for mammography and construction of a calibration service system- Takahiro Tanaka *, Tadahiro
More informationManik Aima, Larry A. DeWerd, Wesley S. Culberson
Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,
More informationA. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.
Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1
More informationDisclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.
Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives
More informationEffect of thyroid shielding during mammography: measurements on phantom and patient as well as estimation with Monte Carlo simulation
Pyka et al. European Radiology Experimental (2018) 2:14 https://doi.org/10.1186/s41747-018-0042-9 European Radiology Experimental ORIGINAL ARTICLE Effect of thyroid shielding during mammography: measurements
More informationFLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system
FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system Collaboration Project: Θ Θ Θ INFN, Milan; Medical Physics Department, Niguarda Ca'Granda Hospital, Milan; Elekta
More informationImplementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations
More informationCommissioning and Radiobiology of the INTRABEAM System
Commissioning and Radiobiology of the INTRABEAM System Susha Pillai and Junan Zhang Scheme INTRABEAM System. Physics Commissioning, QA, and Radiation Protection Radiobiology 1 Disclosure OHSU is an INTRABEAM
More information